Megger MFT1700, MFT1710, MFT1720, MFT1730, MFT1800 User Manual

...
1
M
MFT1700 series
Multifunction testers
User manual
2
Contents
SAFETY WARNINGS .................................................................................................................................................................3
1. Introduction .............................................................................................................................................................................4
2. Overview .................................................................................................................................................................................4
2.1 Front panel and controls .................................................................................................................................................6
2.2 Waste electrical and electronic equipment .......................................................................................................................6
2.3 Battery and Fuse Location, fitting and replacement .........................................................................................................6
3. Operation .......................................................................................................................................................................7
3.1 General operation – all models ........................................................................................................................................7
3.2 Mode button functions ...................................................................................................................................................7
3.3 Test inhibit ......................................................................................................................................................................7
4. Voltage, frequency, Current and temperature measurement .........................................................................................................8
4.1 Making a voltage measurement ......................................................................................................................................8
5. Continuity / resistance measurement ............................................................................................................................................10
5.1 Nulling test lead resistance (up to 9.99ohms) ................................................................................................................10
5.2 Making a CONTINUITY measurement ..............................................................................................................................11
5.3 Storing / downloading results (MFT1730 only) .................................................................................................................11
5.4 Continuity Buzzer ON/OFF ...............................................................................................................................................11
5.5 Switch probe (SP5) ..........................................................................................................................................................11
5.6 Buzzer threshold .............................................................................................................................................................12
5.7 Measurement methods and sources of error ...................................................................................................................12
6. Insulation resistance ...............................................................................................................................................................12
6.1 Making an INSULATION measurement.............................................................................................................................12
6.2 Insulation test lock ..........................................................................................................................................................13
6.3 Measurement methods and sources of error ...................................................................................................................13
7. Loop Impedance testing ...........................................................................................................................................................13
7.1 Range selection and test leads ........................................................................................................................................14
7.2 Making a loop impedance measurement .........................................................................................................................15
7.3 Phase to Neutral (or Phase to Phase) testing ....................................................................................................................19
7.4 Prospective Fault Current and Short Circuit calculation (PFC & PSCC) ..............................................................................20
7.5 Zmax test mode (MFT1720 and 1730 only) .....................................................................................................................20
7.6 MAKING A Zmax LOOP IMPEDANCE MEASUREMENT .....................................................................................................20
7.7 MAKING AN R1+R2 LOOP IMPEDANCE MEASUREMENT .................................................................................................21
7.8 Measurement methods and sources of error ...................................................................................................................22
8. Residual Current Device testing ................................................................................................................................................22
8.1 Making an rcd measurement...........................................................................................................................................23
8.2 Selecting RCD type .........................................................................................................................................................23
8.3 ½ x I RCD Current Rating (No-trip test) .........................................................................................................................23
8.4 1 x I RCD Current Rating (Tripping test on 30mA RCD) ..................................................................................................24
8.5 5 x I RCD Current Rating (Tripping test on 30mA RCD) ..................................................................................................25
8.6 RampTest .......................................................................................................................................................................25
8.7 Type A (DC Sensitive) RCD test ......................................................................................................................................26
8.8 Type B (Pure DC) RCD test ...............................................................................................................................................26
8.9 Variable RCDs (not MFT1710) .........................................................................................................................................27
8.10 AUTO RCD testing ..........................................................................................................................................................27
8.11 3 Phase RCD testing ........................................................................................................................................................27
8.12 Touch voltage display ......................................................................................................................................................29
8.13 Measurement methods and sources of error ...................................................................................................................29
8.14 Useful information ..........................................................................................................................................................29
9. Earth resistance measurement..................................................................................................................................................29
9.1 Connection terminals ......................................................................................................................................................29
9.2 Touch voltage limit ..........................................................................................................................................................29
9.4 Making a measurement – Three terminal resistance measurement ..................................................................................30
10. SETUP OPTIONS .......................................................................................................................................................................31
11. Warning messages ...................................................................................................................................................................32
11.1 Startup warnings ............................................................................................................................................................32
11.2 Battery ............................................................................................................................................................................32
11.3 Battery charger ...............................................................................................................................................................32
11.4 Fuse warning ..................................................................................................................................................................32
11.5 Invalid rotary switch setting .............................................................................................................................................32
11.6 Continuity test ................................................................................................................................................................32
11.7 Insulation test .................................................................................................................................................................32
11.8 RCD Test .........................................................................................................................................................................32
11.9 RCD range selection errors ..............................................................................................................................................32
11.10 Loop Test ......................................................................................................................................................................32
11.11Earth test ........................................................................................................................................................................32
11.12 Test will not start ..........................................................................................................................................................32
3
Appendix A – Sending, Storing, Deleting and Recalling Test Results .....................................................................................................33
Appendix B – Downloading data using Bluetooth® .............................................................................................................................35
Appendix C – Installation category definitions .....................................................................................................................................36
Appendix D – Safe working practice ....................................................................................................................................................36
Appendix E – Cleaning and maintenance ............................................................................................................................................37
Appendix F – Earth resistance testing – Basic principles .......................................................................................................................37
General Specification ..........................................................................................................................................................................38
Repair and Warranty ...........................................................................................................................................................................41
G SAFETY WARNINGS
Safety warnings and precautions must be read and understood before the instrument is used. They must be observed during use.
n The circuit under test must be switched off, de-energised and isolated before test connections are made when carrying out insulation
and continuity tests.
n Continuity of protective conductors and earthed equipotential bonding of new or modified installations must be verified before
carrying out an earth fault loop impedance test, RCD or earth testing
n Do not touch circuit connections and exposed metalwork of an installation or equipment under test. Under fault conditions the
system earth could become hazardous live.
n Do not touch the earth stakes, test leads and their terminations (including connections to the earthing system under test) if an
installation earth fault can arise unless adequate precautions are taken.
n The ‘live circuit warning’ and ‘automatic discharge’ functions are additional safety features and should not be regarded as a
substitute for normal safe working practices.
n Do not move the rotary switch positions while a test is in progress. n Do not operate the instrument or connect it to any external system if it shows any visible signs of damage or if it has been stored
for prolonged periods in unfavourable conditions.
n Do not operate the instrument or connect it to any external system if the battery compartment or casing is open or any parts of
the case (including keypad, selector switch, display window, etc.) are missing.
n Always disconnect the instrument from all systems while batteries are being changed or the fuse replaced n Do notreplace the rechargeable cells in the MFT1730 with non-rechargeable “dry” cells and attempt to charge the cells.
This can cause explosion or fire.
n Do not operate the charging equipment supplied with the MFT1730 in damp or wet environments or outside. All test
leads must be removed from the instrument while charging.
n After insulation tests, capacitive circuits must be allowed to discharge before disconnecting test leads. Locking the insulation test
ON should only be used where there is no risk of a circuit holding a charge.
n The instrument should not be used if any part of it is damaged. n Test leads, probes and crocodile clips must be in good order, clean and with no broken or cracked insulation. n All test leads supplied with the instrument form part of the measuring circuit of the instrument. They must not be modified or
changed in any way, or be used with any other electrical instrument or appliance.
n A plug severed from the power cord MUST be destroyed, as a plug with bare conductors is hazardous in a live socket outlet. n Ensure that hands remain behind guards of probes/clips when testing. n U.K. Safety Authorities recommend the use of fused test leads when measuring voltage on high energy systems. n Replacement fuses must be of the correct type and rating. n Failure to fit the correctly rated fuse will result in damage to the instrument in the event of an overload. n Special precautions are necessary when operating in situations where “live” earths may be encountered: isolation switches and
fuses (not supplied with this instrument) must be used.
n Special precautions are necessary when working near high tension systems (MV and HV): rubber gloves and shoes (not supplied
with this instrument) should be worn.
n Special precautions are necessary when working in wet conditions or in agricultural areas: observe the local safety standards and
take all necessary special precautions applicable to the particular location and do not touch the test leads with bare hands.
LIVE EARTH SAFETY PRECAUTIONS
A ‘Live’ earth is one that carries current from the mains supply, or could do so under fault conditions. The following warnings apply in addition to those listed previously.
n All persons involved must be trained and competent in isolation and safety procedures for the system to be worked on. They
must be clearly instructed not to touch the earth electrode, test stakes, test leads, or their terminations if any ‘Live’ earths may be
encountered. It is recommended that they wear appropriate rubber gloves, rubber soled shoes, and stand on a rubber mat.
n The earth electrode under test should be isolated from the circuit it is protecting before testing commences. If this is not possible,
ART (attached Rod Technique) may be used to measure electrode resistance.
n The instrument terminals should be connected to the system under test through isolation switches that are rated to handle the
likely maximum fault voltages and currents that could be encountered at the installation. The isolation switch must be open whilst any personal contact is made with the remote test stakes, or the connecting leads, e.g. when changing their position.
n The instrument terminals should be connected to the system under test through fuses that are rated to handle the likely maximum
fault voltages and currents that could be encountered at the installation.
NOTE
THE INSTRUMENT MUST ONLY BE USED BY SUITABLY TRAINED AND COMPETENT PERSONS
Users of this equipment and/or their employers are reminded that Health and Safety Legislation requires them to carry out valid risk assessments of all electrical work so as to identify potential sources of electrical danger and risk of electrical injury such as inadvertent short circuits. Where the assessments show that the risk is significant then the use of fused test leads constructed in accordance with the HSE guidance not GS38 ‘Electrical Test Equipment for use by Electricians’ should be used .
4
This instrument is internally protected against electrical damage when used for the purposes of testing low voltage electrical installations as defined herein. If used in a manor other than those defined in this user guide the protection capabilities could be impaired with poten­tial risk to the operator and the instrument.
Symbols used on the instrument are:
G Caution: Refer to accompanying notes
Maximum 300 V a.c. CAT IV to earth
Maximum nominal system voltage of 600 V
Instrument protected by 2 x F2A 600 V 50 kA fuses
c Equipment complies with current EU Directives
This equipment should be recycled as electronic waste
Equipment complies with ‘C tick’ requirements
12 Vdc charger socket
1. Introduction
Congratulations on your purchase of a genuine Megger Multifunction tester. The MFT1700 series Multi-function tester is a compact instrument designed to perform all of the functions required by the electrical contractor to fully test domestic, commercial and industrial wiring. Specially designed to comply with U.K., European and other International wiring regulations and standards, the MFT1700 may be used on all single and three phase systems with rated voltages up to 300 V a.c. rms to earth (ground).
2. Overview
2.1 Front panel and controls
5
2.1.1 Display
Display symbols
Symbol Meaning Symbol Meaning
Test function locked on (also used to indicate a change is saved in setup)
G
Warning triangle – instruction to refer to this user guide
Test lead null active
Fuse blown
Battery indicator
Touch voltage limit (and Earth test voltage) set to 50 V (change setup)
NiMH
Battery type set to rechargeable NiMH - Change in setup section 10
Buzzer enabled
>100 V
Indicates that the ground noise voltage exceeds the instrument measurement capability (test is inhibited)
RCD test in AUTO mode
Rp (Rs)
Potential stake (P stake) resistance exceeds range for accurate measurement
Type AC RCD selected
Rc (RH)
Current stake (C Stake) resistance exceeds range for accurate measurement
Type A RCD selected
V
Ground noise voltage exceeds range for accurate measurement of resistance
Type S RCD (Type AC)
V
VCLAMP error
Type S RCD (Type A)
I
ICLAMP error
Type B RCD selected
Bluetooth enabled
Fast or Full RAMP test selected
T
Instrument is too hot, allow to cool
Instrument is running a test
Earth loop noise detected.
N<->L
Live and netural connections reversed
Reference loop measurement
Loop measurement with Zref value automatically deducted
ZMAX
Loop maximum measurement
6
2.2 Waste electrical and electronic equipment
WEEE
The crossed out wheeled bin placed on Megger products is a reminder not to dispose of the product at the end of its life with general waste.
Megger is registered in the UK as a producer of electrical and electronic equipment. The Registration No is WEE/HE0146QT
2.3 Battery and Fuse Location, fitting and replacement
Battery type: 6 x 1.5 V Alkaline LR6 (AA) or NiMH HR6 rechargeable Fuse type: 2 x 2 A (F) HBC 50 kA 600 V
Battery condition is shown by the following display symbols:
Where NiMH rechargeable batteries are fitted, the battery condition display can be adjusted accordingly. Refer to section 10 SETUP OP­TIONS to change between alkaline and rechargeable batteries.
When set to NiMH batteries, the battery indicator in the display wil show NiMH under the battery status symbol as below: (Feature avail­able on all models).
To replace batteries or fuse:
Switch off the instrument. Disconnect the instrument from any electrical circuits. Remove the battery cover from the base.
For battery replacement:
a) Remove old cells and refit new batteries following correct polarity as marked on the battery holder. c) Replace the battery cover.
Incorrect battery cell polarity can cause electrolyte leakage, resulting in damage to the instrument.
For fuse replacement
a) Withdraw each fuse in turn and check for failure. The blown fuse must be replaced with a 2 A (F) HBC 50 kA 600 V fuse.
Rechargeable batteries and battery charging
The MFT1730 is supplied with rechargeable NiMH cells. These batteries can be charged in the instrument, using the supplied Megger Charger.
To charge the batteries:
Ensure fitted batteries are of the rechargeable NiMH type.
Connect the 12 V DC plug of the charger to the socket on the front of the MFT marked
Warning: Whenever battery cells are being recharged, there should be no connections to the instrument terminals and the instrument should be switched off.
Warning: Do not attempt to recharge non-rechargeable (Primary) cells in the MFT1730. Doing so may result in instrument damage and may cause personal injury.
Ensure ambient temperatures are between 4 ºC and 40 ºC while charging the MFT.
Note: The crossed out wheeled bin placed on the batteries is a reminder not to dispose of them with general waste at the end of their life.
Spent Alkaline and NiMH batteries are classified as portable batteries and should be disposed of in the UK in accordance with Local Authority requirements. For disposal of batteries in other parts of the EU contact your local distributor. Megger is registered in the UK as a producer of batteries.
The Registration number is BPRN00142
7
3. Operation
3.1 General operation – all models
3.1.1 Switching on
Turn the rotary knob away from the off position.
The instrument will perform internal self tests then display the appropriate test screen, depending on the position of the function knobs.
3.1.2 Switching off
Turn the primary function knob to the OFF position.
The instrument will automatically turn itself off after 20 minutes* of inactivity. Press any button or turn either of the rotary knobs to turn back on. * 2 minute option in SETUP, refer to section 10.
3.1.3 Backlight
Press the backlight button.
The backlight will operate for 20 seconds.
3.1.4 Test buttons
Test buttons are duplicated on the left and right. Both buttons perform the same function except when the
is displayed, in this case the right hand buttons perform a scrolling function. The left RED button also performs storage and lock functions on the MFT1730.
3.1.4 Test button lock
To lock the test button hold down either of the RED test lock buttons with the
symbol, whilst holding down the test but-
ton. If
is displayed the right hand buttons perform a scrolling function.
3.2 Mode button functions
The function of the mode button is dependent on the test function selected:
3.3 Test inhibit
Each test mode has conditions under which testing will be inhibited, as below:
3.3.1 Insulation
Detection of a circuit voltage above 50 V (a warning is displayed at 25 V)
3.3.2 Continuity
Detection of a circuit voltage above that used by the instrument will inhibit testing.
Test selected Function Options Comments
V/ºC Volts (mv model specific) Temperature requires suitable transducer
Continuity
RLO
Buzzer ENABLE/DISABLE
Buzzer ON Buzzer OFF
Buzzes on <2 Ω May be changed in SETUP. Refer to Section 10.
Insulation
RSIO
Buzzer ENABLE/DISABLE
Buzzer ON Buzzer OFF
Buzzes on <1 MΩ May be changed in SETUP. Refer to Section 10.
Loop impedance (Z)
L-PE
3Lo - 3 wire no trip 2Hi – 2 wire high current 2Lo – 2 wire no trip
3Lo 2Hi 2Lo
3Lo default measurement 2Lo not available if 3 wire connection is detected
RCD
0º /180º selection
0º 180º
(Press and release)
RCD Type
AC A AC(s) A(s) B
(Press and HOLD) Type B available on MFT1730 only
EARTH (RE) Touch voltage limit 50 V/ 25 V
(Press and release) MFT1730 only
SETUP
Refer to instrument setup section 10.
8
3.3.3 Earth loop impedance
Touch voltage exceeds 50 V (or 25 V depending on instrument configuration) Supply voltage over range or under range Supply frequency out of specification
3.3.4 RCD testing
Touch voltage detected or predicted to exceed 50 V (or 25 V depending on instrument configuration) Supply voltage over range or under range Supply Frequency out of specification
3.3.5 Earth testing
External voltage greater than 25 V present Leads not corrected connected as per the test requirements Potential stake not within range (Rp) Current stake not within range (Rc)
Other conditions that will inhibit testing include:
3.3.6 Battery exhausted
All testing will be inhibited in the event of a flat battery, refer to section 2.3.
4. Voltage, frequency, current and temperature measurement
4.1 Making a voltage measurement
.1 Set the Main rotary range knob to volts (The position of the right hand rotary range knob does not matter)
.2 Using two test leads, connect test leads to the L1 (+ve) and L2 (-ve) terminals
OR if Using the Mains plug Lead SAI10:
a. For Live to Neutral measurements, connect the Red connector to the L1 terminal and the Blue connector to the L2
terminal
b. For Live to Earth measurements connect the Red test to the L1 terminal and the Green connector to the L2 terminal
9
Connection (a) Connection (b)
Note: When connecting all three test leads (eg Phase, Neutral and Earth) or the mains plug test lead, the voltage displayed is the high-
est of the three possible voltages.
On models with a dedicated mV range this is selected using the Mode button to select mV mode.
4.2 Frequency measurement
.1 Automatically displayed when connecting to a live circuit as per 4.1 above
4.3 Phase rotation (Not MFT1710)
Display of Phase rotation is Automatic when all three test leads are connected to the 3 phase supply as below: .1 Set the main rotary range knob to volts (The position of the right hand rotary range knob does not matter)
.2 Using three test leads, connect test leads to the L1 to Phase1, L2 to Phase 2 and L3 to Phase 3. The MFT will display L1 L2 L3 or L1 L3 L2 depending on the direction of phase rotation.
Normal rotation Reverse rotation
4.4 Leakage current measurement
Leakage current measurement uses the optional accessory current clamp (ICLAMP).
.1 On the MFT1720, MFT1730, set the primary range knob to clamp position
.
.2 Connect to ICLAMP (part no. ICLAMP) to the ICLAMP
socket on the MFT.
.3 Connect the clamp to the circuit conductor. The instrument will display the ac current flowing in the conductor.
4.5 Temperature measurement (not on MFT1710)
.1 Connect the thermocouple transducer to the L1 (+ve) and L2 (-ve) terminals
.2 Press the Mode button to select ºC. (Pressing the mode button will cycle round the V, mV and ºC measurement modes)
The display will show the temperature at the tip of the temperature probe.
10
4.6 Switch probe
In the V/mV/ºC mode all measurements except temperature can be made with the remote switch probe. Tests are automatic and do not require the test button to be pressed.
.1 Connect the switch probe to the switch probe socket. The probe replaces the standard RED test lead and can now be used as a normal test probe.
5. Continuity / resistance measurement
IMPORTANT
The continuity test will auto-range from 0.01 Ω to 99.9 kΩ. Circuits up to 2 Ω will be tested at >200 mA. To change the test current, go to seciton 10 - Setup.
The continuity test is automatic. The test starts as soon as the leads are connected to a circuit. The TEST button is ONLY used to null the lead set.
Warning: Prior to any continuity testing, ensure the circuits under test are isolated and not live.
SETUP allows the follow configuration options:
n Positive test current n Bi-Directional test current
Bi-directional test current allows the automatic testing of the circuit in both directions and the highest measured value being displayed See section 10 SETUP OPTIONS.
5.1 Nulling test lead resistance (up to 9.99 ohms)
Before starting a continuity test, the test lead resistance should be nulled such that it does not add extra resistance to the circuit being measured. Once nulled it does not need repeating for each test. Periodically it should be checked to ensure nothing has changed.
The “Lead Null” value is retained even when the tester is switched off.
To null test leads:
.1 Short test probes or clips together and press the TEST button. The null symbol
will be displayed to indicate
lead null is active.
Lead null OFF
This null value is stored until the TEST button is pressed again.
.2 To cancel the LEAD NULL, separate the test leads and press the TEST button.
11
5.2 Making a CONTINUITY measurement
.1 Set the Primary (Left) range knob to
range. (The position of the right hand rotary range knob must not be in
the
position).
.2 Connect two test leads to the L1 (+ve) and L2 (-ve) terminals on the instrument. A continuity measurement is made automatically.
NOTES: Measurements are prevented when: A resistance of > 99,9 kΩ is present Circuit voltages in excess of 4 V are detected.
5.3 Storing / downloading results (MFT1730 only)
For full details see Appedix A and B. Once the display shows a value it will automatically be logged into temporary memory. Unless stored, this will be over written by the next measurement. To store this result or to send it to a PowerSuite compatible device, refer to Appedix A and B.
5.4 Continuity Buzzer ON/OFF
Whilst in the continuity range, press the MODE button
. This will toggle the buzzer ON and OFF.
Buzzer ON = Buzzer OFF = No symbol
5.5 Switch probe (SP5)
In the CONTUNUITY/RESISTANCE mode all measurements can be made with the remote switch probe (SP5). Tests are automatic and do not require the TEST button to be pressed.
.1 Connect the switch probe to the switch probe socket L1 (+ve). The switch probe replaces the standard RED test lead. Test as in 5.2 above.
12
5.6 Buzzer threshold
If the measured resistance is less than the buzzer threshold, the buzzer will sound. The resistance at which the buzzer stops sounding can be changed to meet individual test requirements. Refer to the SETUP section 10 of this guide.
Selectable limits of 0.5 Ω, 1 Ω, 2 Ω, 5 Ω, 10 Ω, 20 Ω, 50 Ω, 100 Ω. (depending on model) are available. This setting is stored even when the instrument is switched off.
5.7 Measurement methods and sources of error
Method of measurement
The 2-wire lead set must be used for this measurement. A d.c voltage of nominally 4,4 V with a current limit of >200 mA is used to measure resistance less than 2 Ω.
Possible sources of error
Measurement results can be affected by the following:
n The presence of circuits connected in parallel. n Presence of AC voltages on the circuit being measured n A poor connection to the circuit under test n Incorrectly nulled test leads n Use of fused leads
6. Insulation resistance
IMPORTANT:
The insulation test is protected by a live circuit warning. Detection of a voltage over 50 V will inhibit testing. This applies whether or not the insulation test is locked on.
6.1 Making an INSULATION measurement
.1 Set the left hand rotary range knob to the
insulation test voltage required:
.2 Connect two test leads to the L1 (+ve ) and L2 (-ve) terminals on the instrument.
13
.3 To start test, press and hold either of the TEST buttons,
or , on the instrument.
Release the test button after the displayed reading has settled. Circuit will now discharge safely.
Note: A 1000 V warning is displayed whenever the 1000 V range is selected for the first time and the TEST button is pressed.
6.2 Insulation test lock
To lock an insulation test ON, hold down either of the TEST buttons followed by either of the RED LOCK buttons.
To release the “Locked on” insulation test, press the TEST button.
Warning: The test voltage will be permanently present on the test probes or crocodile clips when in the locked position.
Warning: Auto discharge - Auto discharge facility automatically and safely discharges the circuit at the completion of an
insulation test.
Live circuit warning - operates when connected to Live circuits > 25 V. Testing is still permitted.
Test inhibited - Live circuits greater than 50 V will inhibit testing.
6.3 Measurement methods and sources of error
Method of measurement
The selected dc test voltage (current limited to less than 2 mA d.c).is applied to the circuit under test and the resistance is calculated from measurements of the resulting voltage and current.
Capacitive circuits can take some time to charge. This is displayed as an increasing voltage that takes longer to reach its maximum than normal.
The reading is stable with a circuit capacitance less than 5 µF.
7. Loop impedance testing
IMPORTANT
This measurement requires both selector knobs to be set to the Loop testing mode (GREEN RANGES) on the MFT1720 and 1730 or just the left knob on the 1710.
This is a live circuit test. All precautions relevant to working on live circuits, to ensure the safety of the operator and any other personnel should be in place.
Overview of the LOOP IMPEDANCE test
A Loop impedance test is the measurement of the impedance of a circuit whilst the circuit is electrically live. Unlike a continuity test, a loop impedance test applies a load to the circuit and measures the change in the circuit voltage, from which the loop “resistance” is calculated.
For those circuits protected by an RCD the load that is connected Phase to Earth must be small enough not to trip the RCD. Consequently there must be many tests performed to establish the loop impedance of the circuit. These are automatically performed and the end result is displayed.
Test Lead Null:
The MFT does not need the resistance of the test leads to be nulled for this test. They are already calibrated into the measurement circuit at 0.07 Ω. However if using fused leads or 3rd party test leads, the resistance of these leads may be different. In this case they can be measured using the continuity test and the resistance can be compensated for in the SETUP options, see section 10.
Loading...
+ 29 hidden pages