(347) Output voltage is not within specified limits
(361) Overvoltage
(364) Measured capacitances don't match
(365) Specimen capacitance below limit
(366) Specimen capacitance above limit
(367) Measured DC current > MaxDCCurrent
(368) Measured hum current > MaxHumCurrent
IDAX is an insulation diagnostic system for investigations/analysis of dielectric materials,
normally insulators. The measurement method used is dielectric spectroscopy, i.e. measurement
of the dielectric properties of the material as a function of the frequency, and in some cases
also as a function of the voltage. By studying the dielectric material properties as functions of
frequency it is possible to make a distinction between different types of phenomena. For
example, it is possible to separate polarization loss from leakage currents.
The system applies a sinusoidal voltage with desired frequency over the sample. This voltage
will generate a current in the sample. By accurately measuring the voltage and the current, the
sample impedance can be calculated. Depending upon the actual sample model, various
parameters can be calculated from the impedance, such as capacitance, loss, resistance etc. This
procedure can then be repeated at the specified frequencies and voltage levels and even more
information on the sample can be revealed. More information can be found in the chapter
"Measurement technique".
The system is fully computerized and most of the measurements can be performed
automatically.
More technical and safety details are given in sections:
Specifications
System and accessories
Safety precautions
1.1.1
1.1.1.1
Specifications
The Insulation Diagnostic System IDAX measures the insulation impedance, at different
frequencies, for insulation materials present in high and medium voltage transmission and
distribution networks. The measurement results are used for insulation diagnosis.
The IDAX uses voltages up to 200 Vpeak (~140 V RMS). The voltage range can be increased by
adding an external high voltage unit (the high voltage unit is specified separately).
IDAX-300 Specifications
General
The Insulation Diagnostic System IDAX-300 measures the insulation impedance, at different
frequencies, for insulation materials present in high and medium voltage transmission and
distribution networks. The measurement results are used for insulation diagnosis.
The IDAX-300 uses voltages up to 200 Vpeak (~140 V RMS). The voltage range can be increased
by adding an external high voltage unit (the high voltage unit is specified separately). The
IDAX-300 is operated using an external laptop/PC.
The IDAX-300 uses a three electrode set-up, which allows for measurements of non-grounded
as well as grounded objects with and without guard.
The IDAX-300 system applies sinusoidal shaped voltages at different frequencies across the test
object and simultaneously measures the current through it. The output voltage is either taken
from a 10 V
A "Warning" statement denotes hazardous conditions and actions that can cause bodily
harm or death.
A "Caution!" statement denotes conditions and actions that can damage the IDAX or the
equipment being tested.
Warning!
Always follow local safety regulations when working on high voltage equipment.
Make certain all personnel working with IDAX have been trained in its correct use
and that all applicable safety precautions are taken.
BEFORE connecting this unit to the mains outlet user must verify that only a single
protective ground systems exists AND no measurable voltage potential exists
between the mains outlet protective ground and the test object protective ground. If
a voltage potential is found between mains outlet protective ground and the test
object protective ground then additional measures described in local safety standards
should be applied to ensure personal safety.
The measurement system can generate hazardous and even lethal voltages.
Do not attempt to service the IDAX yourself as opening or removing covers may
expose you to dangerous voltage.
Do not use any accessories that are not intended for use together with the IDAX.
Unplug the IDAX from the mains supply when it is left unattended or not in use.
Before cleaning, unplug the IDAX from the mains supply. Use a damp cloth for
cleaning. Do not use liquid or aerosol cleaners.
Caution!
Make sure that the mains voltage selected on the selector switch located on the back
panel corresponds to the voltage level before connecting the mains.
Refer all servicing to qualified service personnel.
If you need to return your IDAX, please use either the original crate or one of
equivalent strength.
The IDAX-206 come with the measurement and display software already installed on the
internal computer. On the IDAX-300 and if you want to use the IDAX-206 with an external PC
you have to install the software.
To start installing the software insert the USB memory stick and run the program IDAXInstallXXXX.exe from it.
When first starting the IDAX software after installing it make sure the IDAX206/300 is
connected and turned on. For the IDAX-206 also check that the switch in the rear panel is set to
EXT USB. When the software starts it will look for the IDAX hardware and connect to it
automatically.
Measurement Technique
Short overview of measurement principle is given in this section.
How impedance is measured Gives basic principle of IDAX operation
Sample modelingDescribes theoretical models used for presenting measurement
results
How Impedance is Measured
IDAX measures impedance. By measuring the impedance at one point, i,e, at a specific
frequency and amplitude, parameters such as resistance, capacitance and loss can be calculated.
The impedance of a sample is measured by applying a voltage across the sample. This voltage
will generate a current through the sample. By accurately measuring the voltage and the
current, the impedance can be calculated, see illustration below.
Computer with
DSP-board
Voltage source
Control voltage
Sample
ZU
I
Measured voltage
ElectrometerVoltmeter
AV
Measured current
Fig.1. Measurement of electrical impedance.
The impedance is calculated using Ohm's law:
where Z, U and I are complex entities.
The voltage (and the current) is generated by a voltage source. There are currently two internal
voltage sources available in the IDAX system, which can deliver a maximum peak output of
10 Vpeak and 200 Vpeak, respectively. The voltage is measured by means of a voltmeter and
the current is measured by an ammeter or electrometer which acts as a current-to-voltage
converter. The analogue signals (voltages) are then converted to digital samples of the signals
that are used in subsequent calculations.
1.4.2
Sample Modelling
The impedance, Z, can be presented directly or by using different impedance models. Two ways
of presenting Z directly are the polar and the rectangular, as follows:
Polar:
Rectangular:
Two simple models which are usually used in circuit analysis, although more seldom in
insulation analysis, are capacitance, C, and resistance, R. The equivalent RC circuit models
available are series and parallel models (see Fig.1) calculated as follows:
a) b)
Fig.1. Equivalent circuit models: a) series RC circuit, b) parallel RC circuit.
Parallel:
Series:
where w = 2pf and f is frequency.
Another model, more often used in insulation diagnostics, is the complex capacitance model
describing the insulation impedance as a complex capacitance, where the imaginary part of the
capacitance represents the losses. The complex capacitance model is defined as follows:
The DC' is defined as the capacitance, C', with an arbitrary constant k (usually negative) added.
The aim of this parameter is to make it possible to distinguish between small changes in
capacitance in graphical presentation.
A model, very often used in insulation diagnostics, is a description of the insulation impedance
as a capacitance combined with a dissipation factor, tand, or a power factor (PF or cosj). The
capacitance, tand and cosj/PF are defined as follows:
C, PF, Tan-Delta:
If tand and cosj (PF) are small, then tand » cosj = PF (E.g. tand=0.1 corresponds to PF=0.0995)
Insulation diagnostics is based on material characterization and therefore material models are
often used. To be able to define material parameters from measured impedance Z the
geometry of the sample, described in terms of the geometrical capacitance C0, has to be
defined. In the illustration below, a vacuum (or air-filled) capacitor of defined geometry is
shown. Since no "material" is between the electrodes, the capacitance of a) is the geometrical
capacitance.
UC
I
a)b)c)
Material parameter models based on a geometrical capacitance C0 and material parameters.
In the above illustration b) and c) a material is inserted between the electrodes and it will
influence the current, I, flowing in the circuit. The influence of the material can be described by
different parameters using either a dielectric model or a conductive model. In the dielectric
model the "material capacitance", the permittivity, is a complex function describing both the
capacitance and the loss. Whereas in the conductive model the capacitance is described by a
permittivity and the loss by a conductivity (or resistivity). The dielectric and resistive models are
derived as follows:
If geometrical capacitance, C0, is unknown it can be set by the user by entering a permittivity
(dielectric constant), e', for the material. Entering a permittivity, e', will let the system calculate
an approximate C0 making use of the material models available. Yet, one must be aware that
the accuracy of the absolute values are limited by the accuracy of the entered permittivity. If
the geometrical capacitance is unknown and an approximate permittivity is given, the
geometrical capacitance is calculated as follows:
The impedance value, Z, used in the calculation is the first measurement point in the actual
measurement.
Getting Started with IDAX
This describes how to perform a diagnostic measurement on a specific test specimen. More
details about the software functions can be found in IDAX System Control and IDAX Results.
It is recommended to start using IDAX by going through the procedure in this section in order
to become familiar with the instrument and the software. This procedure is also a good test to
perform whenever a confirmation of the proper functioning of the instrument is required. This
test is executed using the standard equipment delivered with IDAX.
Before starting to use IDAX certain preparatory procedures must be followed.
1. Before connecting IDAX instrument power cord to the mains outlet:
read the safety precautions
§
connect the transparent ground lead to a protective earth (ground) close to measurement
§
specimen and to panel of instrument.
2. Connect IDAX instrument to a mains outlet which meets the requirements in User's Manual
and local safety regulations, and switch on IDAX instrument using the mains switch POWER
on the front panel.
Preparatory Procedure: Specimen
1. All specimen capacitances must be discharged before connecting IDAX leads. This means that
all specimen electrodes must be short circuited and grounded. The electrodes must remain
grounded until IDAX leads are connected. Below a general picture of ground connections is
shown.
1.5.1
2. In the case of Test Capacitor, keeping the electrodes grounded while connecting IDAX leads
is complicated. Therefore, in this particular case with Test Capacitor, it is permitted to remove
the ground connections before connecting IDAX.
Measurement
Make sure that IDAX is properly connected to mains outlet.
1.
Turn on IDAX using the mains switch POWER on the front panel. The IDAX performs a
2.
self-test and start-up procedure.
Activate the Results window if not active after start-up.
3.
In Results create new object selecting Object from menu File / New. Click "Browse" and
4.
select !!Test Capacitor.obj template and enter arbitrary name as Object Name, for example
"Test" and click OK.
By default all measurement templates listed as Associated Measurement Templates will be
5.
available for performing the measurements. It is possible to deselect one or more
measurement templates from Associated Measurement Templates. As shown in picture
below, four Measurement Templates !!C10 (GST-Guard), !!C12 (UST), !!C12, C10, C10+C12
and !!C20 (GST-Guard) are associated with the object "Test" for IDAX-206, similar procedure
as IDAX-300.
Fill in available fields with relevant information.
Create new measurement sequence selecting Measurement Sequence from menu File /
New. Arbitrary name can be entered as Measurement Sequence, for example "Getting
Started" and press OK. New measurement sequence with corresponding Measurement Plan
files will appear in Test Browser.
IDAX User's Manual14
Fill in available fields with relevant information.
8.
1.5.1.1
Test Sequence for IDAX-300
This is Getting Started test sequence when using IDAX-300
The specimen to be measured using the built-in capacitance box built into the system that
terminates at the front panel. The respective capacitances are:
C10 = ~2.5 nF
C12 = ~4.7 nF
Warning!
The measurement system can generate hazardous and even lethal voltages.
In Test Browser select Measurement Plan file !!C10 (GST-Guard), carefully read and follow the
instructions. Alternatively, follow the instructions below.
Make sure that preparation procedures have been followed (see also "Ungrounded Specimen
1.
Test (UST) Measurement").
1.5.1.2
Connect the generator output to the upper capacitance connector.
2.
Connect Current 1 and Sense 1 to the lower capacitance connector.
3.
Remove possible ground connections from the connection pins.
4.
Turn on the output switch activate the output and then start the measurement.
5.
Wait until measurement is finished or abort earlier by stopping the measurement.
In case of a two winding transformer the set-up of IDAX-206 (IDA 200) and IDAX-300 are very
similar; the leads need to be moved between test 2 and test 3.
In case of a three winding transformer the IDAX-300 second input channel is utilized (standard);
the leads need to be moved between test 2 and test 3, between test 4 and test 5. If use
IDAX-206 (IDA 200) on a three winding transformer, the leads need to be moved between every
test.
If using IDAX-300 with two electrometers (option), the two set-ups using same cable set-up can
be measured simultaneously, e.g. test 1 and test 2 (test 3 and test4; test 5 and test 6).
Note that Test 6, CTH (Energize T and measure H), in many cases are replaced by CHT (Energize
H and Measure T):
· IDAX-300: CHT,UST-2, H, L, T (same cable set-up as Test 1 and Test 2)
· IDAX-206: CHT,UST, H, T, L shorted to tank
During Measurement
When the measurement has been started the measurement file icon changes to an icon
corresponding to the data file.
1. By pressing <F4> on the keyboard, Results window can be activated if not already active and
the measurement progress can be viewed when selecting Graph tab. The graph shows the
dissipation factor for the Test Capacitor in the frequency range 0.1 - 1000 Hz .
2. By pressing <F5> on the keyboard, an oscilloscope will be displayed on the screen. The Ch 0
(red) signal in the oscilloscope displays the applied voltage over the sample, Ch 1 (white) is
the output voltage from the first electrometer and Ch 2 (blue) is the output voltage from the
second electrometer. The output voltage from the electrometer is a function of the current
through the sample.
3. The instrument stops automatically after the last measurement point is completed, however,
the measurement can be interrupted earlier by a stop command. You can press stop button
or, on the IDAX-206, the OUTPUT OFF button. On the IDAX-206 a lit green READY LED on
the front panel indicates that the measurement is stopped or finished.
By default the measurement data is presented using a model defined in the measurement plan.
However, additional models can be added while inspecting the results (see Selecting Models for
details).
Depending on which parameters are of interest to the user, the measurement data can be
presented using different models. E.g. in the tand model the dissipation and/or power factor as
well as capacitance can be displayed in the graph as a function of frequency. See also how to
select models in sections "Configuration / Models" and "Sample Modelling".
The results can be viewed using three additional tabs: Graph, Table and 50/60 Hz data as
shown in pictures below.
2. In the Graph tab you can select the desired parameters. For example, the dielectric
parameter e'.
3. In Results window menu the choices File / Export, File / Print and File / Report allow for
exporting the measurement data and creation of reports.
All measurements are stored on the IDAX 206 system hard disk or the measurement computer's
hard disk on an IDAX-300 system and can be opened and analysed at any time. For comparison
purposes, many measurements can be loaded into the same Results window.
IDAX System Control
The IDAX System Control performs a measurement according to commands pre-arranged in a
Measurement plan or command file (C-file), using programming syntax described in IDAX
Command and Variable References. The Measurement plan used is available in Results tab
Measurement plan.
Editing of measurement plans and displaying the results.
Window is activated after up-start by default
Comments
Adding comments to the actual measurement.
Messages
Displaying messages from the software during a
measurement.
Oscilloscope
Viewing the actual voltage and current curve forms.
In addition to the IDAX System Control program window, a few more windows are accessible
from here.
1.6.1
1.6.2
1.6.3
Message Window
Messages from IDAX system to the user will be displayed in the Messages window. For example,
if a measurement was aborted prematurely by the program, the reason for it will be displayed
here.
Connect and Disconnect
In the File menu you have the choice to connect (if not connected) and disconnect (if
connected) to the IDAX unit. If you are not connected and the software needs to connect to the
IDAX unit it will be done automatically.
Oscilloscope
The Oscilloscope displays the signals present on channel 0 (white, Ch0) and channel 1 (red, Ch1)
inputs of the analogue IO unit.
The Ch0 and Ch1 check boxes select the channels that are displayed. Trace 0, white, is
proportional to the sample voltage. Trace 1, red, shows the electrometer output voltage, which
is a function of the input current.
1.6.4
Horizontal control
The oscilloscope time base can be set either on auto scale or manual. On auto scale, the time
base is set from the generator signal from the DSP. In the manual mode, the time base is set by
the Time/div slide bar. The time above the slide bar refers to the time per division, the full
oscilloscope window width consists of ten divisions.
Roll
The Roll check box can be used for slow signals that the oscilloscope may otherwise have
difficulties to accurately lock.
Vertical control
The gain of the input channels can be adjusted with the two vertical slide bars next to the Grid
check box at the oscilloscope window. The voltage range is displayed beside the Ch0 and Ch1
check boxes. The offset of the displayed signals can be adjusted by placing the mouse pointer
inside the oscilloscope window and dragging it up and down while holding the mouse button.
Grid
The Grid check box displays or hides the grid.
Changing Language
The language in which IDAX user interface (IDAX software) is displayed can be changed. The
following languages are available:
English
§
Swedish
§
German
§
French
§
Spanish
§
Language is changed selecting menu Configuration / Change Language in IDAX System
Control, which opens a file opening dialog. The selected language file, *.lng, is loaded and it
will be used after restarting the IDAX program. Each language has its own language file, *.lng,
and all language files are located in the directory D:\IDA\Language files\.