DCM1500S User Guide
2
www.megger.com
Page 3
This manual supersedes all previous issues of this manual. Please ensure that you are
using the most recent issue of this document. Destroy any copies that are of an older
issue.
Declaration of Conformity
Hereby, Megger Instruments Limited declares that radio equipment manufactured by
Megger Instruments Limited described in this user guide is in compliance with Directive
2014/53/EU. Other equipment manufactured by Megger Instruments Limited described
in this user guide is in compliance with Directives 2014/30/EU and 2014/35/EU where they
apply.
The full text of Megger Instruments EU declarations of conformity are available at the
following internet address
megger.com/company/about-us/eu-dofc
www.megger.comDCM1500S User Guide
3
Page 4
Safety Information
G
Understand and follow the operating instructions carefully.
WARNING
G
Identify hazardous conditions and actions that could cause BODILY HARM or DEATH
When using test leads or probes, keep your fingers behind the finger guards.
Personal protective equipment should be used if there are ACCESSIBLE
HAZARDOUS LIVE PARTS in the installation where measurement is to be carried
out.
Remove test leads from meter before opening the battery door or meter case.
Use the meter only as specified in this manual; otherwise, the protection provided
by the meter may be impaired.
Always use the proper input terminals, switch position, and range for
measurements.
Verify the meters operation by measuring a known voltage. If in doubt, have the
meter calibrated.
Do not apply more than the rated voltage, as marked on the meter, between
terminals or between any terminal and earth.
Use caution with voltages above 30 V AC rms, 42 V AC peak, or 60 V DC. These
voltages pose a shock hazard.
To avoid false readings that can lead to electric shock and injury, replace the
battery as soon as the low battery indicator blinks.
Disconnect circuit power and discharge all high-voltage capacitors before testing
resistance, continuity, diodes, or capacitance.
Do not use the meter around explosive gas or vapour.
To reduce the risk of fire or electric shock do not expose this product to rain or
moisture.
Probe assemblies to be used for MAINS measurements shall be RATED as
appropriate for MEASUREMENT CATEGORY III or IV according to EN 61010031 and shall have a voltage RATING of at least the voltage of the circuit to be
measured.
DCM1500S User Guide
4
www.megger.com
Page 5
Safety Information
G
■ DO NOT USE the test leads if the internal white insulation layer is exposed.
■ DO NOT USE the test leads above maximum ratings of CAT Environment or
voltage and current that are indicated on the probe and probe tip guard.
■ Do not apply a current with a frequency that is higher than the frequency
response range specied in the Electrical Specications section.
■ Do not apply or remove the clamp or test leads on or around uninsulated
hazardous live conductors where a potential to cause electric shock, electrical
burns or arc ash exists.
CAUTION
G
Disconnect the test leads from the test points before changing the position of the
function rotary switch.
Never connect a source of voltage with the function rotary switch in Ω,
position.
Do not expose the meter to extremes of temperature or high humidity.
and
www.megger.comDCM1500S User Guide
5
Page 6
Symbols as marked on the Meter and Instruction manual
Risk of electric shock
See instruction manual
DC measurement
AC measurement
Both direct and alternating current
Equipment protected by double or reinforced insulation
Battery
Earth
Conforms to EU directives
Conforms to UKCA directives
Application around and removal from hazardous live conductors
Do not discard this product or throw away
Bluetooth®
CATIV
Measurement category IV: Equipment connected between the origin of the low-voltage
mains supply outside the building and the consumer unit.
CATIII
Measurement category III: Equipment connected between the consumer unit and the
electrical outlets.
CATII
Measurement category II: Equipment connected between the electrical outlets and the
user’s equipment.
Unsafe Voltage
G
To alert you to the presence of a potentially hazardous voltage, when the tester
detects a voltage ≥30 V or a voltage overload (OL) in V, mV, PV. The V symbol is
displayed.
DCM1500S User Guide
6
www.megger.com
Page 7
Features
6000 Count LCD display
Backlight
VoltSeek (Non-Contact Voltage Detection)
True RMS reading on AC
Data logger (up to 4000 sample values)
Manual save mode
Bluetooth® communication
Torch turns on when clamp jaws open
1500 A AC/DC current measurement
DC voltage measurement up to 2000 V using the PVHV lead set
AV voltage measurement up to 1500 V using the PVHV lead set
Upon removing your new Solar/PV Clamp Meter from its packing, you should have the
following items:
Megger DCM1500S Solar/PV Clamp Meter
Test lead set (one black, one red)
Test probes (one black, one red)
Crocodile clips (one black, one red)
TP100 Temperature probe
Megger PVHV1 Lead set (4mm plugs)
Megger PVHV2 Lead set (MC4 PV plugs)
User Manual
Carry case
Batteries
DCM1500S User Guide
8
www.megger.com
Page 9
The Meter Description
Function Button
1 JAW
2 VoltSeek LED
3 Push button
4 Trigger
5 Rotary switch to turn the instrument On / Off and select a function
6 6,000 count LCD display
7 Positive input terminal
8 Common (earth reference) input terminal
www.megger.comDCM1500S User Guide
9
Page 10
Making Basic Measurements
Preparation and Caution Before Measurement
Observe the rules of G Warnings and G Cautions.
G
CAUTION
G
When connecting the test leads to the DUT (Device Under Test) connect the common test
lead before connecting the live test lead. When removing the test leads remove the live
test lead before removing the common test lead.
Measuring Voltage
Turn the switch to select the measuring function.
DCM1500S User Guide
10
www.megger.com
Page 11
Measuring PV Voltage
PV String
Turn the switch and press the Function button to select AC/DC mode
CAUTION
G
This function is only available with the dedicated PVHV test lead sets.
The PVHV test leads are designed for use in environments that are not directly
connected to the mains. Always select correct AC/DC mode to perform any high voltage
measurement.
The meter will flash this symbol
voltage is different and/or dangerous.
and the correct mode symbol (AC/DC) if the input
F
www.megger.comDCM1500S User Guide
11
Page 12
Measuring Current
CAT IV 600 V
G
CAT III 1000 V
Fixed barrier hand guard
Do not hold the meter
G
across the xed barrier
Turn the switch and press the Function button to select AC/DC/Hz mode.
Note: The torch will turn on when the jaw is opened.
DCM1500S User Guide
12
www.megger.com
Page 13
Measuring Resistance/Continuity/Capacitance/Diode
Turn the switch and press the Function button to select the measuring function
www.megger.comDCM1500S User Guide
13
Page 14
Measuring Temperature °C / °F
Function
button
K Type
Temperature
Probe
Vent / Pipe
Turn the switch and press the Function button to select °C / °F mode.
DCM1500S User Guide
14
www.megger.com
Page 15
Measuring µA
(Example)
Flame sensor
Control
module
Turn the switch and press the Function button to select AC/DC mode.
www.megger.comDCM1500S User Guide
15
Page 16
Using the Function Button
Switch
Position
Function
Press the Function button to change the function for each rotary switch position.
DCM1500S User Guide
16
www.megger.com
Page 17
Measuring Frequency
Backlight
Function
button
Press
> 2 sec
2 sec
Press Function button for over 2 seconds to turn Backlight on or off.
www.megger.comDCM1500S User Guide
17
Page 18
MIN/MAX
Press
>2 sec to exit
the MIN/MAX
mode
The MIN/MAX mode records the min and max input values.
When the input goes below the recorded min value or above the recorded max value, the
meter beeps and records the new value. Press Hold button to pause the recording.
DCM1500S User Guide
18
www.megger.com
Page 19
Volt Seek
Sensor position
Volt Seek
2 sec
Do not hold
across this line
when sensing the
hazardous voltage
Press MIN/MAX button for over 2 seconds to enter/exit Volt Seek mode. Press MIN/MAX
button to switch from high to low sensitivity.
WARNING
G
The Volt Seek LED indicates the electric field. If the Volt Seek LED is not on, voltage could
still be present.
www.megger.comDCM1500S User Guide
19
Page 20
Smart Hold
The meter will beep continuously, and the display will flash if the measured signal is larger
than the display reading by 50 counts.
DCA ZERO
Remove the clamp from the conductor before performing DCA ZERO.
Press the HOLD button for over 2 seconds to compensate for any residual magnetism.
Press
> 2 sec
Press
> 2 sec
DCM1500S User Guide
20
www.megger.com
Page 21
High Frequency Rejection Filter (HFR)
The High Frequency Rejection (HFR) mode uses a low pass filter when taking AC
measurements. The cut-off frequency (-3 dB point) of the low pass filter is 800 Hz.
Press
WARNING
F
Hazardous voltages may still be present when the LCD reading is very low in HFR mode.
Verify the voltage again without HFR mode activated.
www.megger.comDCM1500S User Guide
21
Page 22
INRUSH
When using the inrush current mode, select a suitable measurement range before pressing
the HFR/INRUSH button and triggering the inrush current measurement feature.
Motor O
Press
> 2 sec
Inrush
2 sec
Motor OnTime
Trigger
Point
Detecting
Press
> 2 sec
Inrush
2 sec
The current is > trigger level
DCM1500S User Guide
22
www.megger.com
Page 23
Auto Power Off
After user specied time
with no operation (Note Time
Setting of Auto Power O)
Wake up the meter by turning the switch or pressing any button.
Time Setting of Auto Power Off
Function
Button
Function
Button
Press and hold the function button whilst turning the meter on, the meter beeps 3 times
prior to APO (Auto Power Off) activating.
Release the function button and press it again to select the desired time. The auto power
off time can be set to 5 mins,10 mins, 20 mins, or disabled (OFF).
www.megger.comDCM1500S User Guide
23
Page 24
Testing LCD Display
Turn on the meter while keeping HOLD button pressed.
DCM1500S User Guide
24
www.megger.com
Page 25
Function of LOG Button
Press
Press the Bluetooth® button while powering-up to select the Logger mode,
Manual Saving mode or Clear the memory.
www.megger.comDCM1500S User Guide
25
Page 26
Data Logger
The meter can store up to 4000 data entries in internal memory.
Press the Bluetooth® button for more than 2 seconds to activate data logger mode.
The meter will enter time interval setting mode. Press the Bluetooth® button again to
select the desired sample interval. This can be set to 1, 5, 10, 30 or 60 seconds.
Time interval setting mode
Press
> 2 sec
PressPress
Press
Start to record automatically after
2 seconds without operation
LOG icon ashes
while logging
CAUTION
G
All stored data is cleared at the start of each data log or when Clr (Clear) or Save
(Manual Saving) is selected. Data can be downloaded and saved for analysis by the
Megger Link App.
DCM1500S User Guide
26
www.megger.com
Page 27
Manual Saving Mode
Flash
Press
> 2 sec
To save a reading
Show stored reading
amount for 2 seconds
If Manual Saving is selected, readings can be stored by pressing the Hold/Zero button
when in Log mode. The sample location is displayed on each button press.
CAUTION
G
All stored data is saved until switching to data logger mode or executing the clear
function.
www.megger.comDCM1500S User Guide
27
Page 28
Bluetooth®
The meter uses Bluetooth® low energy (BLE) V4.0 wireless technology to transfer the realtime reading and the stored data. The open-air communication range is up to 10 meters.
Download the “Megger Link” App. Turn on the Bluetooth function of the meter by
pressing the Bluetooth button and open the Megger Link App to connect the DCM.
The Bluetooth icon of the meter will flash whilst connecting and remain on the LCD once
the connection is established.
QR Code for the Megger Link App
on the Google Play Store
DCM1500S User Guide
28
QR Code for the Megger Link App
on the Apple App Store
www.megger.com
Page 29
Default Temperature Units Setting
OR
Press and hold the MIN/MAX button whilst turning the meter on to select °C or °F.
www.megger.comDCM1500S User Guide
29
Page 30
Low Battery and Battery Replacement
Replace the battery as soon as the low battery indicator appears to avoid false readings.
Refer to the following figure to replace the batteries
CAUTION
G
Remove test leads from meter before opening the battery cover or meter case.
DCM1500S User Guide
30
www.megger.com
Page 31
Specifications
General Specifications
Display count:
Overrange display:
Measurement frequency :
Maximum Conductor Size of
JAW:
Dimensions (W x H x D):
Weight (approx.):
Low Batteries Indication
Power requirements :
Battery life :
Environmental Conditions
Indoor Use Only
Pollution degree:
Maximum Operating Altitude:
Operating Temperature &
Relative Humidity :
Storage temperature :
Vibration:
Drop Protection :
Safety :
6,000 counts.
«OL» or «-OL»
3 times per second
42mm diameter
62 x 254 x 41 mm
480 g (including battery)
If battery voltage drops below the operating
voltage, will flash.
AA size battery x 2 (R6, LR6, 15D, 15A)
Approximately 200 hours (based on Alkaline
batteries without using the backlight).
-20 to +60°C, 0 to 80% RH.
(batteries not fitted)
Random Vibration per MIL-PRF-28800F
Class 2
1.2 m (4 ft) drop to hardwood on concrete
floor
EN 61010-1, EN 61010-2-032,
EN 61010-2-033 for CAT III 1000 V, CAT IV
600 V, EN 61326-1
www.megger.comDCM1500S User Guide
31
Page 32
Electrical Specifications
Accuracy is given as ± (% of reading + counts of least significant digit) at 23°C ± 5°C,
with relative humidity less than 80% and is specified for 1 year after calibration.
Temperature coefficient:
AC Function :
0.1 x (Specified accuracy) / °C,
< 18 °C, > 28 °C
ACV and ACA specifications are AC coupled,
true RMS.
Accuracy is unspecified for a Square Wave.
For non-sinusoidal waveforms, Additional
Accuracy by Crest Factor (C.F.):
Add 3.0% for C.F. 1.0 ~ 2.0
Add 5.0% for C.F. 2.0 ~ 2.5
Add 7.0% for C.F. 2.5 ~ 3.0
Max. Crest Factor of Input Signal:
3.0 @ 3000 counts
2.0 @ 4500 counts
1.5 @ 6000 counts
Frequency Response is specified for sine
waveform. LCD displays 0 counts when the
reading < 20 counts.
DC mV
RangeOL ReadingResolutionAccuracy
600.0 mV660.0 mV0.1 mV± (0.7% + 5D)
Input Impedance :
Overload Protection:
10 MΩ
AC/DC 1,000 V
DC Voltage
RangeOL ReadingResolutionAccuracy
600.0 V660.0 V0.1 V
1000 V1100 V1 V
Input Impedance :
Overload Protection:
DCM1500S User Guide
32
10 MΩ
AC/DC 1,000 V
± (0.7% + 2D)
www.megger.com
Page 33
AC Voltage
RangeOL ReadingResolutionAccuracy
600.0 V660.0 V0.1 V
1000 V1100 V1 V
± (1.0% + 5D)
Input Impedance :
Frequency Response :
Overload Protection :
10MΩ // less than 100pF
45 ~ 400Hz (Sine Wave)
AC/DC 1,000 V
PV DC Voltage (using PVHV1 or PVHV2 lead set)
RangeOL ReadingResolutionAccuracy
600.0 V660.0 V0.1 V
2000 V2200 V1 V
Input Impedance :
Overload Protection :
10MΩ
AC/DC 1,000 V
± (2.0% + 5D)
PV AC Voltage (using PVHV1 or PVHV2 lead set)
RangeOL ReadingResolutionAccuracy
600.0 V660.0 V0.1 V
1500 V1600 V1 V
Input Impedance :
Frequency Response :
Overload Protection :
10MΩ
45 ~ 400Hz (Sine Wave)
AC/DC 1,000 V
± (2.0% + 5D)
AC/DC µA
RangeOL ReadingResolutionAccuracy
400.0 µA440.0 µA0.1 µA
4000 µA4400 µA1 µA
Input Impedance :
Frequency Response :
Overload Protection :
www.megger.comDCM1500S User Guide
Approx. 2.2 kΩ
45 ~ 400Hz (Sine Wave)
AC/DC 1,000 V
± (1.0% + 3D)
33
Page 34
AC/DC Current
RangeOL ReadingResolutionAccuracy
60.00 A66.00 A0.01 A
600.0 A660.0 A0.1 A
1500 A1550 A1 A
Add 10 digits to the accuracy when <5.0 A. Add 0.5% to the accuracy when >1000 A.
Add 1% to the accuracy when >100 Hz.
Frequency Response :
Overload Protection :
(Sine Wave) 45 ~ 400 Hz for ≤1000 A
45 ~ 65 Hz for >1000 A
AC/DC 1500 A
± (2.0% + 5D)
Frequency
RangeOL ReadingResolutionAccuracy
100.00 Hz100.00 Hz0.01 Hz
1000.0 Hz1000.0 Hz0.1 Hz
10.000 kHz10.000 kHz0.001 kHz
± (0.3% + 3D)
Minimum Sensitivity :
Minimum Frequency :
Overload Protection :
> 5 V (for ACV 1 Hz ~ 10 kHz)
> 8 A (for ACA 1 Hz ~ 1 kHz)
1 Hz
AC/DC 1000 V and 1500 A
HFR (High Frequency Rejection)
Available for ACV and ACA. Add ± 4% to specified accuracy of each function and each
range for 45 Hz to 200 Hz. Accuracy is unspecified for > 200 Hz.
Cut-off Frequency (-3 dB): 800 Hz
Inrush Current
Available for ACA and Flexible Current Probe. Trigger level: ≥50d.
Add ± 3% to specified accuracy of each function and each range.
DCM1500S User Guide
34
www.megger.com
Page 35
Resistance
RangeOL ReadingResolutionAccuracy
600.0 Ω660.0 Ω0.1 Ω± (0.9% + 5D)
6.000 kΩ6.600 kΩ0.001 kΩ
60.00 kΩ66.00 kΩ0.01 kΩ
600.0 kΩ660.0 kΩ0.1 kΩ
To obtain a more accurate reading, the lead resistance should be measured by
connecting the probes/clips together and the result deducted from any resistance
measurement taken.
Overload Protection :
AC/DC 1000 V
± (0.9% + 2D)
Continuity
Built-in buzzer sounds when measured resistance is less than 20Ω and the sound is off
when measured resistance is more than 200 Ω, between 20 Ω to 200 Ω the buzzer may
be either on or off.
Continuity Indicator :
Response Time of Buzzer :
Overload Protection :
2.7 kHz Tone Buzzer
< 100 ms
AC/DC 1000 V
Diode
RangeOL ReadingResolutionAccuracy
1.500 V1.550 V0.001 V± (0.9% + 2D)
Open Circuit Voltage :
Overload Protection :
Approx. 1.8 V
AC/DC 1000 V
Capacitance
RangeOL ReadingResolutionAccuracy
100.0 µF110.0 µF0.1 µF
1000 µF1100 µF1 µF
Overload Protection :
www.megger.comDCM1500S User Guide
AC/DC 1000 V
± (1.9% + 2D)
35
Page 36
VoltSeek
Voltage Range of High
80 V ~ 1000 V (At the top edge of the jaw)
Sensitivity :
Voltage Range of Low
160 V ~ 1000 V (At the top edge of the jaw)
Sensitivity :
Temperature
RangeOL ReadingResolutionAccuracy
-40.0 °C – 400.0 °C440.0 °C0.1 °C± (1% + 20D)
-40.0 °F – 752.0 °F824.0 °F0.1 °F± (1% + 36D)
The accuracy does not include the accuracy of the thermocouple probe.
Accuracy specification assumes surrounding temperature stable to ±1 °C.
For surrounding temperature changes of ± 2 °C rated accuracy applies after 2 hours.
Overload Protection :
AC/DC 1000 V
Specification PVHV1 (4mm plugs) and PVHV2 (MC4) test leads
Input Impedance :
Overvoltage Category :
Pollution Degree :
Exposed probe tip length :
Environmental ratings :
Altitude :
Safety Standard :
10 MΩ
CAT II 1000 V AC ,1500 V DC CAT III 1000 V,
CAT IV 600 V.
2
Test probe: 18 mm to 3.5 mm
(0.70 inch to 0.14 inch)
-10 °C to 45 °C (-4 °F to 113 °F), 80% R.H.
2000 m (6,562 ft)
EN61010-031
Product manufactured in Taiwan
DCM1500S User Guide
36
www.megger.com
Page 37
Maintenance
Do not attempt to repair this meter. It contains no user-serviceable parts. Repair or
servicing should only be performed by qualified personnel.
Cleaning
Periodically wipe the case with a dry cloth and detergent, do not use abrasives or solvents
WEEE Directive
The crossed out wheeled bin symbol on the instrument and on the batteries is a reminder
not to dispose of them with general waste at the end of their life. Megger is registered in
the UK as a Producer of Electrical and Electronic equipment. The registration No is; WEE/
DJ2235XR.
Users of Megger products in the UK may dispose of them at the end of their useful life by
contacting B2B Compliance at www.b2bcompliance.org.uk or by telephone on 01691
676124.
Users of Megger products in other regions should contact their local Megger office or
distributor.
Battery Disposal
The batteries in this product are classified as Portable Batteries under the Batteries
Directive. Please contact either Megger Ltd, your local Megger office or distributor for
instructions on the safe disposal of these batteries. Megger is registered in the UK as a
producer of batteries. The registration number is BPRN01235.
For further information see www.megger.com
www.megger.comDCM1500S User Guide
37
Page 38
Warranty (3 years)
This meter is warranted to the original purchaser against defects in material and
workmanship for 3 year from the date of purchase.
During this warranty period, the manufacturer will, at its option, replace or repair the
defective unit, subject to verification of the defect or malfunction.
This warranty does not cover fuses, disposable batteries, or damage from abuse, neglect,
accident, unauthorised repair, alteration, contamination, or abnormal conditions of
operation or handling.
Any implied warranties arising out of the sale of this product, including but not limited to
implied warranties of merchantability and fitness for a particular purpose, are limited to
the above. The manufacturer shall not be liable for loss of use of the instrument or other
incidental or consequential damages, expenses, or economic loss, or for any claim or
claims for such damage, expense or economic loss. Some states or countries laws vary, so
the above limitations or exclusions may not apply to you.
Megger Limited
Archcliffe Road
Dover
Kent
CT17 9EN