Megger DCM1500S User Manual

Page 1
DCM1500S
Digital Clamp Meter
User Guide
Page 2
This document is copyright of:
Megger Limited, Archcliffe Road, Dover, Kent CT17 9EN. ENGLAND
T +44 (0)1304 502101 F +44 (0)1304 207342 www.megger.com
Megger Ltd reserves the right to alter the specification of its products from time to time without notice. Although every effort is made to ensure the accuracy of the information contained within this document it is not warranted or represented by Megger Ltd. to be a complete and up - to - date description.
For Patent information about this instrument refer to the following web site:
megger.com/patents
DCM1500S User Guide
2
www.megger.com
Page 3
This manual supersedes all previous issues of this manual. Please ensure that you are using the most recent issue of this document. Destroy any copies that are of an older issue.
Declaration of Conformity
Hereby, Megger Instruments Limited declares that radio equipment manufactured by Megger Instruments Limited described in this user guide is in compliance with Directive 2014/53/EU. Other equipment manufactured by Megger Instruments Limited described in this user guide is in compliance with Directives 2014/30/EU and 2014/35/EU where they apply.
The full text of Megger Instruments EU declarations of conformity are available at the following internet address
megger.com/company/about-us/eu-dofc
www.megger.com DCM1500S User Guide
3
Page 4
Safety Information
G
Understand and follow the operating instructions carefully.
WARNING
G
Identify hazardous conditions and actions that could cause BODILY HARM or DEATH
 When using test leads or probes, keep your fingers behind the finger guards.
 Personal protective equipment should be used if there are ACCESSIBLE
HAZARDOUS LIVE PARTS in the installation where measurement is to be carried out.
 Remove test leads from meter before opening the battery door or meter case.
 Use the meter only as specified in this manual; otherwise, the protection provided
by the meter may be impaired.
 Always use the proper input terminals, switch position, and range for
measurements.
 Verify the meters operation by measuring a known voltage. If in doubt, have the
meter calibrated.
 Do not apply more than the rated voltage, as marked on the meter, between
terminals or between any terminal and earth.
 Use caution with voltages above 30 V AC rms, 42 V AC peak, or 60 V DC. These
voltages pose a shock hazard.
 To avoid false readings that can lead to electric shock and injury, replace the
battery as soon as the low battery indicator blinks.
 Disconnect circuit power and discharge all high-voltage capacitors before testing
resistance, continuity, diodes, or capacitance.
 Do not use the meter around explosive gas or vapour.
 To reduce the risk of fire or electric shock do not expose this product to rain or
moisture.
 Probe assemblies to be used for MAINS measurements shall be RATED as
appropriate for MEASUREMENT CATEGORY III or IV according to EN 61010­031 and shall have a voltage RATING of at least the voltage of the circuit to be measured.
DCM1500S User Guide
4
www.megger.com
Page 5
Safety Information
G
■ DO NOT USE the test leads if the internal white insulation layer is exposed.
■ DO NOT USE the test leads above maximum ratings of CAT Environment or
voltage and current that are indicated on the probe and probe tip guard.
■ Do not apply a current with a frequency that is higher than the frequency
response range speci󰘰ed in the Electrical Speci󰘰cations section.
■ Do not apply or remove the clamp or test leads on or around uninsulated
hazardous live conductors where a potential to cause electric shock, electrical
burns or arc 󰘱ash exists.
CAUTION
G
 Disconnect the test leads from the test points before changing the position of the
function rotary switch.
 Never connect a source of voltage with the function rotary switch in Ω,
position.
 Do not expose the meter to extremes of temperature or high humidity.
and
www.megger.com DCM1500S User Guide
5
Page 6
Symbols as marked on the Meter and Instruction manual
Risk of electric shock
See instruction manual
DC measurement
AC measurement
Both direct and alternating current
Equipment protected by double or reinforced insulation
Battery
Earth
Conforms to EU directives
Conforms to UKCA directives
Application around and removal from hazardous live conductors
Do not discard this product or throw away
Bluetooth®
CATIV
Measurement category IV: Equipment connected between the origin of the low-voltage mains supply outside the building and the consumer unit.
CATIII
Measurement category III: Equipment connected between the consumer unit and the electrical outlets.
CATII
Measurement category II: Equipment connected between the electrical outlets and the user’s equipment.
Unsafe Voltage
G
 To alert you to the presence of a potentially hazardous voltage, when the tester
detects a voltage 30 V or a voltage overload (OL) in V, mV, PV. The V symbol is displayed.
DCM1500S User Guide
6
www.megger.com
Page 7
Features
 6000 Count LCD display
 Backlight
 VoltSeek (Non-Contact Voltage Detection)
 True RMS reading on AC
 Data logger (up to 4000 sample values)
 Manual save mode
 Bluetooth® communication
 Torch turns on when clamp jaws open
 1500 A AC/DC current measurement
 DC voltage measurement up to 2000 V using the PVHV lead set
 AV voltage measurement up to 1500 V using the PVHV lead set
 600 kΩ Resistance measurement / continuity beeper
 Frequency counter
 Capacitance/Diode test
 °C/°F Temperature function
 Inrush current
 DCA auto-zeroing button
 Max/Min hold
 Smart hold
 High frequency noise rejection filter
 Auto power off (APO)
 CAT IV 600 V/CAT III 1000 V safety standard
www.megger.com DCM1500S User Guide
7
Page 8
Unpacking and Inspection
Upon removing your new Solar/PV Clamp Meter from its packing, you should have the following items:
 Megger DCM1500S Solar/PV Clamp Meter
 Test lead set (one black, one red)
 Test probes (one black, one red)
 Crocodile clips (one black, one red)
 TP100 Temperature probe
 Megger PVHV1 Lead set (4mm plugs)
 Megger PVHV2 Lead set (MC4 PV plugs)
 User Manual
 Carry case
 Batteries
DCM1500S User Guide
8
www.megger.com
Page 9
The Meter Description
Function Button
1 JAW
2 VoltSeek LED
3 Push button
4 Trigger
5 Rotary switch to turn the instrument On / Off and select a function
6 6,000 count LCD display
7 Positive input terminal
8 Common (earth reference) input terminal
www.megger.com DCM1500S User Guide
9
Page 10
Making Basic Measurements
Preparation and Caution Before Measurement
Observe the rules of G Warnings and G Cautions.
G
CAUTION
G
When connecting the test leads to the DUT (Device Under Test) connect the common test lead before connecting the live test lead. When removing the test leads remove the live test lead before removing the common test lead.
Measuring Voltage
Turn the switch to select the measuring function.
DCM1500S User Guide
10
www.megger.com
Page 11
Measuring PV Voltage
PV String
Turn the switch and press the Function button to select AC/DC mode
CAUTION
G
This function is only available with the dedicated PVHV test lead sets.
The PVHV test leads are designed for use in environments that are not directly connected to the mains. Always select correct AC/DC mode to perform any high voltage measurement.
The meter will flash this symbol voltage is different and/or dangerous.
and the correct mode symbol (AC/DC) if the input
F
www.megger.com DCM1500S User Guide
11
Page 12
Measuring Current
CAT IV 600 V
G
CAT III 1000 V
Fixed barrier hand guard
Do not hold the meter
G
across the 󰄈xed barrier
Turn the switch and press the Function button to select AC/DC/Hz mode.
Note: The torch will turn on when the jaw is opened.
DCM1500S User Guide
12
www.megger.com
Page 13
Measuring Resistance/Continuity/Capacitance/Diode
Turn the switch and press the Function button to select the measuring function
www.megger.com DCM1500S User Guide
13
Page 14
Measuring Temperature °C / °F
Function
button
K Type
Temperature
Probe
Vent / Pipe
Turn the switch and press the Function button to select °C / °F mode.
DCM1500S User Guide
14
www.megger.com
Page 15
Measuring µA
(Example)
Flame sensor
Control module
Turn the switch and press the Function button to select AC/DC mode.
www.megger.com DCM1500S User Guide
15
Page 16
Using the Function Button
Switch
Position
Function
Press the Function button to change the function for each rotary switch position.
DCM1500S User Guide
16
www.megger.com
Page 17
Measuring Frequency
Backlight
Function
button
Press
> 2 sec
2 sec
Press Function button for over 2 seconds to turn Backlight on or off.
www.megger.com DCM1500S User Guide
17
Page 18
MIN/MAX
Press
>2 sec to exit the MIN/MAX mode
The MIN/MAX mode records the min and max input values.
When the input goes below the recorded min value or above the recorded max value, the meter beeps and records the new value. Press Hold button to pause the recording.
DCM1500S User Guide
18
www.megger.com
Page 19
Volt Seek
Sensor position
Volt Seek
2 sec
Do not hold across this line when sensing the hazardous voltage
Press MIN/MAX button for over 2 seconds to enter/exit Volt Seek mode. Press MIN/MAX button to switch from high to low sensitivity.
WARNING
G
The Volt Seek LED indicates the electric field. If the Volt Seek LED is not on, voltage could still be present.
www.megger.com DCM1500S User Guide
19
Page 20
Smart Hold
The meter will beep continuously, and the display will flash if the measured signal is larger than the display reading by 50 counts.
DCA ZERO
Remove the clamp from the conductor before performing DCA ZERO.
Press the HOLD button for over 2 seconds to compensate for any residual magnetism.
Press
> 2 sec
Press
> 2 sec
DCM1500S User Guide
20
www.megger.com
Page 21
High Frequency Rejection Filter (HFR)
The High Frequency Rejection (HFR) mode uses a low pass filter when taking AC measurements. The cut-off frequency (-3 dB point) of the low pass filter is 800 Hz.
Press
WARNING
F
Hazardous voltages may still be present when the LCD reading is very low in HFR mode. Verify the voltage again without HFR mode activated.
www.megger.com DCM1500S User Guide
21
Page 22
INRUSH
When using the inrush current mode, select a suitable measurement range before pressing the HFR/INRUSH button and triggering the inrush current measurement feature.
Motor O󰓐
Press
> 2 sec
Inrush
2 sec
Motor On Time
Trigger
Point
Detecting
Press
> 2 sec
Inrush
2 sec
The current is > trigger level
DCM1500S User Guide
22
www.megger.com
Page 23
Auto Power Off
After user speci󰄈ed time
with no operation (Note Time
Setting of Auto Power O󰓐)
Wake up the meter by turning the switch or pressing any button.
Time Setting of Auto Power Off
Function
Button
Function
Button
Press and hold the function button whilst turning the meter on, the meter beeps 3 times prior to APO (Auto Power Off) activating.
Release the function button and press it again to select the desired time. The auto power off time can be set to 5 mins,10 mins, 20 mins, or disabled (OFF).
www.megger.com DCM1500S User Guide
23
Page 24
Testing LCD Display
Turn on the meter while keeping HOLD button pressed.
DCM1500S User Guide
24
www.megger.com
Page 25
Function of LOG Button
Press
Press the Bluetooth® button while powering-up to select the Logger mode, Manual Saving mode or Clear the memory.
www.megger.com DCM1500S User Guide
25
Page 26
Data Logger
The meter can store up to 4000 data entries in internal memory.
Press the Bluetooth® button for more than 2 seconds to activate data logger mode. The meter will enter time interval setting mode. Press the Bluetooth® button again to select the desired sample interval. This can be set to 1, 5, 10, 30 or 60 seconds.
Time interval setting mode
Press
> 2 sec
PressPress
Press
Start to record automatically after 2 seconds without operation
LOG icon 󰄉ashes
while logging
CAUTION
G
All stored data is cleared at the start of each data log or when Clr (Clear) or Save (Manual Saving) is selected. Data can be downloaded and saved for analysis by the Megger Link App.
DCM1500S User Guide
26
www.megger.com
Page 27
Manual Saving Mode
Flash
Press
> 2 sec
To save a reading
Show stored reading
amount for 2 seconds
If Manual Saving is selected, readings can be stored by pressing the Hold/Zero button when in Log mode. The sample location is displayed on each button press.
CAUTION
G
All stored data is saved until switching to data logger mode or executing the clear function.
www.megger.com DCM1500S User Guide
27
Page 28
Bluetooth®
The meter uses Bluetooth® low energy (BLE) V4.0 wireless technology to transfer the real­time reading and the stored data. The open-air communication range is up to 10 meters.
Download the “Megger Link” App. Turn on the Bluetooth function of the meter by pressing the Bluetooth button and open the Megger Link App to connect the DCM. The Bluetooth icon of the meter will flash whilst connecting and remain on the LCD once the connection is established.
QR Code for the Megger Link App on the Google Play Store
DCM1500S User Guide
28
QR Code for the Megger Link App on the Apple App Store
www.megger.com
Page 29
Default Temperature Units Setting
OR
Press and hold the MIN/MAX button whilst turning the meter on to select °C or °F.
www.megger.com DCM1500S User Guide
29
Page 30
Low Battery and Battery Replacement
Replace the battery as soon as the low battery indicator appears to avoid false readings. Refer to the following figure to replace the batteries
CAUTION
G
Remove test leads from meter before opening the battery cover or meter case.
DCM1500S User Guide
30
www.megger.com
Page 31
Specifications
General Specifications
Display count:
Overrange display:
Measurement frequency :
Maximum Conductor Size of JAW:
Dimensions (W x H x D):
Weight (approx.):
Low Batteries Indication
Power requirements :
Battery life :
Environmental Conditions
Indoor Use Only
Pollution degree:
Maximum Operating Altitude:
Operating Temperature & Relative Humidity :
Storage temperature :
Vibration:
Drop Protection :
Safety :
6,000 counts.
«OL» or «-OL»
3 times per second
42mm diameter
62 x 254 x 41 mm
480 g (including battery)
If battery voltage drops below the operating voltage, will flash.
AA size battery x 2 (R6, LR6, 15D, 15A)
Approximately 200 hours (based on Alkaline batteries without using the backlight).
2
2,000m (6562 ft)
-10°C ~ 10°C,
-10°C ~ 30°C, ≤80% RH. 30°C ~ 40°C, ≤75% RH. 40°C ~ 50°C, ≤45% RH.
-20 to +60°C, 0 to 80% RH. (batteries not fitted)
Random Vibration per MIL-PRF-28800F Class 2
1.2 m (4 ft) drop to hardwood on concrete floor
EN 61010-1, EN 61010-2-032, EN 61010-2-033 for CAT III 1000 V, CAT IV 600 V, EN 61326-1
www.megger.com DCM1500S User Guide
31
Page 32
Electrical Specifications
Accuracy is given as ± (% of reading + counts of least significant digit) at 23°C ± 5°C, with relative humidity less than 80% and is specified for 1 year after calibration.
Temperature coefficient:
AC Function :
0.1 x (Specified accuracy) / °C, < 18 °C, > 28 °C
 ACV and ACA specifications are AC coupled,
true RMS.
 Accuracy is unspecified for a Square Wave.
 For non-sinusoidal waveforms, Additional
Accuracy by Crest Factor (C.F.): Add 3.0% for C.F. 1.0 ~ 2.0 Add 5.0% for C.F. 2.0 ~ 2.5 Add 7.0% for C.F. 2.5 ~ 3.0
 Max. Crest Factor of Input Signal:
3.0 @ 3000 counts
2.0 @ 4500 counts
1.5 @ 6000 counts
 Frequency Response is specified for sine
waveform. LCD displays 0 counts when the reading < 20 counts.
DC mV
Range OL Reading Resolution Accuracy
600.0 mV 660.0 mV 0.1 mV ± (0.7% + 5D)
Input Impedance :
Overload Protection:
10 MΩ
AC/DC 1,000 V
DC Voltage
Range OL Reading Resolution Accuracy
600.0 V 660.0 V 0.1 V
1000 V 1100 V 1 V
Input Impedance :
Overload Protection:
DCM1500S User Guide
32
10 MΩ
AC/DC 1,000 V
± (0.7% + 2D)
www.megger.com
Page 33
AC Voltage
Range OL Reading Resolution Accuracy
600.0 V 660.0 V 0.1 V
1000 V 1100 V 1 V
± (1.0% + 5D)
Input Impedance :
Frequency Response :
Overload Protection :
10MΩ // less than 100pF
45 ~ 400Hz (Sine Wave)
AC/DC 1,000 V
PV DC Voltage (using PVHV1 or PVHV2 lead set)
Range OL Reading Resolution Accuracy
600.0 V 660.0 V 0.1 V
2000 V 2200 V 1 V
Input Impedance :
Overload Protection :
10MΩ
AC/DC 1,000 V
± (2.0% + 5D)
PV AC Voltage (using PVHV1 or PVHV2 lead set)
Range OL Reading Resolution Accuracy
600.0 V 660.0 V 0.1 V
1500 V 1600 V 1 V
Input Impedance :
Frequency Response :
Overload Protection :
10MΩ
45 ~ 400Hz (Sine Wave)
AC/DC 1,000 V
± (2.0% + 5D)
AC/DC µA
Range OL Reading Resolution Accuracy
400.0 µA 440.0 µA 0.1 µA
4000 µA 4400 µA 1 µA
Input Impedance :
Frequency Response :
Overload Protection :
www.megger.com DCM1500S User Guide
Approx. 2.2 kΩ
45 ~ 400Hz (Sine Wave)
AC/DC 1,000 V
± (1.0% + 3D)
33
Page 34
AC/DC Current
Range OL Reading Resolution Accuracy
60.00 A 66.00 A 0.01 A
600.0 A 660.0 A 0.1 A
1500 A 1550 A 1 A
Add 10 digits to the accuracy when <5.0 A. Add 0.5% to the accuracy when >1000 A. Add 1% to the accuracy when >100 Hz.
Frequency Response :
Overload Protection :
(Sine Wave) 45 ~ 400 Hz for ≤1000 A 45 ~ 65 Hz for >1000 A
AC/DC 1500 A
± (2.0% + 5D)
Frequency
Range OL Reading Resolution Accuracy
100.00 Hz 100.00 Hz 0.01 Hz
1000.0 Hz 1000.0 Hz 0.1 Hz
10.000 kHz 10.000 kHz 0.001 kHz
± (0.3% + 3D)
Minimum Sensitivity :
Minimum Frequency :
Overload Protection :
> 5 V (for ACV 1 Hz ~ 10 kHz) > 8 A (for ACA 1 Hz ~ 1 kHz)
1 Hz
AC/DC 1000 V and 1500 A
HFR (High Frequency Rejection)
Available for ACV and ACA. Add ± 4% to specified accuracy of each function and each range for 45 Hz to 200 Hz. Accuracy is unspecified for > 200 Hz.
Cut-off Frequency (-3 dB): 800 Hz
Inrush Current
Available for ACA and Flexible Current Probe. Trigger level: 50d. Add ± 3% to specified accuracy of each function and each range.
DCM1500S User Guide
34
www.megger.com
Page 35
Resistance
Range OL Reading Resolution Accuracy
600.0 Ω 660.0 Ω 0.1 Ω ± (0.9% + 5D)
6.000 kΩ 6.600 kΩ 0.001 kΩ
60.00 kΩ 66.00 kΩ 0.01 kΩ
600.0 kΩ 660.0 kΩ 0.1 kΩ
To obtain a more accurate reading, the lead resistance should be measured by connecting the probes/clips together and the result deducted from any resistance measurement taken.
Overload Protection :
AC/DC 1000 V
± (0.9% + 2D)
Continuity
Built-in buzzer sounds when measured resistance is less than 20Ω and the sound is off when measured resistance is more than 200 Ω, between 20 Ω to 200 Ω the buzzer may be either on or off.
Continuity Indicator :
Response Time of Buzzer :
Overload Protection :
2.7 kHz Tone Buzzer
< 100 ms
AC/DC 1000 V
Diode
Range OL Reading Resolution Accuracy
1.500 V 1.550 V 0.001 V ± (0.9% + 2D)
Open Circuit Voltage :
Overload Protection :
Approx. 1.8 V
AC/DC 1000 V
Capacitance
Range OL Reading Resolution Accuracy
100.0 µF 110.0 µF 0.1 µF
1000 µF 1100 µF 1 µF
Overload Protection :
www.megger.com DCM1500S User Guide
AC/DC 1000 V
± (1.9% + 2D)
35
Page 36
VoltSeek
Voltage Range of High
80 V ~ 1000 V (At the top edge of the jaw)
Sensitivity :
Voltage Range of Low
160 V ~ 1000 V (At the top edge of the jaw)
Sensitivity :
Temperature
Range OL Reading Resolution Accuracy
-40.0 °C – 400.0 °C 440.0 °C 0.1 °C ± (1% + 20D)
-40.0 °F – 752.0 °F 824.0 °F 0.1 °F ± (1% + 36D)
The accuracy does not include the accuracy of the thermocouple probe. Accuracy specification assumes surrounding temperature stable to ±1 °C. For surrounding temperature changes of ± 2 °C rated accuracy applies after 2 hours.
Overload Protection :
AC/DC 1000 V
Specification PVHV1 (4mm plugs) and PVHV2 (MC4) test leads
Input Impedance :
Overvoltage Category :
Pollution Degree :
Exposed probe tip length :
Environmental ratings :
Altitude :
Safety Standard :
10 MΩ
CAT II 1000 V AC ,1500 V DC CAT III 1000 V, CAT IV 600 V.
2
Test probe: 18 mm to 3.5 mm (0.70 inch to 0.14 inch)
-10 °C to 45 °C (-4 °F to 113 °F), 80% R.H.
2000 m (6,562 ft)
EN61010-031
Product manufactured in Taiwan
DCM1500S User Guide
36
www.megger.com
Page 37
Maintenance
Do not attempt to repair this meter. It contains no user-serviceable parts. Repair or servicing should only be performed by qualified personnel.
Cleaning
Periodically wipe the case with a dry cloth and detergent, do not use abrasives or solvents
WEEE Directive
The crossed out wheeled bin symbol on the instrument and on the batteries is a reminder not to dispose of them with general waste at the end of their life. Megger is registered in the UK as a Producer of Electrical and Electronic equipment. The registration No is; WEE/ DJ2235XR.
Users of Megger products in the UK may dispose of them at the end of their useful life by contacting B2B Compliance at www.b2bcompliance.org.uk or by telephone on 01691
676124.
Users of Megger products in other regions should contact their local Megger office or distributor.
Battery Disposal
The batteries in this product are classified as Portable Batteries under the Batteries Directive. Please contact either Megger Ltd, your local Megger office or distributor for instructions on the safe disposal of these batteries. Megger is registered in the UK as a producer of batteries. The registration number is BPRN01235.
For further information see www.megger.com
www.megger.com DCM1500S User Guide
37
Page 38
Warranty (3 years)
This meter is warranted to the original purchaser against defects in material and workmanship for 3 year from the date of purchase.
During this warranty period, the manufacturer will, at its option, replace or repair the defective unit, subject to verification of the defect or malfunction.
This warranty does not cover fuses, disposable batteries, or damage from abuse, neglect, accident, unauthorised repair, alteration, contamination, or abnormal conditions of operation or handling.
Any implied warranties arising out of the sale of this product, including but not limited to implied warranties of merchantability and fitness for a particular purpose, are limited to the above. The manufacturer shall not be liable for loss of use of the instrument or other incidental or consequential damages, expenses, or economic loss, or for any claim or claims for such damage, expense or economic loss. Some states or countries laws vary, so the above limitations or exclusions may not apply to you.
Megger Limited Archcliffe Road Dover Kent CT17 9EN
Tel: +44 (0) 1304 502 101 Fax: +44 (0) 1304 207 342
www.megger.com
DCM1500S User Guide
38
www.megger.com
Page 39
www.megger.com DCM1500S User Guide
39
Page 40
Manufacturing sites
Megger Limited Archcliffe Road Dover Kent CT17 9EN ENGLAND
T. +44 (0)1 304 502101 F. +44 (0)1 304 207342
Megger USA - Dallas 4545 West Davis Street Dallas TX 75211-3422 USA
T. 800 723 2861 (USA only) T. +1 214 333 3201
F. +1 214 331 7399 E. USsales@megger.com
Megger GmbH Weststraße 59 52074 Aachen, Germany
T. +49 (0) 241 91380 500
Megger AB Rinkebyvägen 19, Box 724, SE-182 17 Danderyd
T. +46 08 510 195 00 E. seinfo@megger.com
Megger USA - Valley Forge Valley Forge Corporate Center 2621 Van Buren Avenue Norristown Pennsylvania, 19403 USA
T. +1 610 676 8500 F. +1 610 676 8610
Megger USA - Fort Collins 4812 McMurry Avenue Suite 100 Fort Collins CO 80525 USA
T. +1 970 282 1200
This instrument is manufactured in Taiwan.
The company reserves the right to change the specification or design without prior notice.
Megger is a registered trademark
The Bluetooth® word mark and logos are registered trademarks owned by Bluetooth SIG, Inc and is used under licence.
Part No: DCM1500S_1013-357_UG_EN_V02