Megger CFL510F Operating Manual

M
CFL510F Cable Fault Locator
USER MANUAL
GSAFETY WARNINGS
Although this tester does not generate any hazardous voltages, circuits to which it can be connected could be dangerous due to electric shock hazard or due to arcing (initiated by short circuit). While every effort has been made by the manufacturer to reduce the hazard, the
own safety. For use on energised systems rated up to 300 V Installation Category III* the fused clip set Megger Part
Number 6111-218, must be used.
* Relates to the transient over voltages likely to be met in fixed wiring installations.
4mm plug to BNC Adaptor:
This 4mm plug to BNC adapter is intended for use with telecomm network cables only, it is not designed or intended for direct connection to an energised mains supply. However, in normal use it may be subject to telecom network voltages (TNV) as defined by IEC 60950 3rd edition (1999-04).
The BNC plug and socket are, by necessity, accessible. The outer sheath for this connector is normally at SELV levels, however under single fault conditions it may carry hazardous voltages. The operator must therefore verify that the accessible plug or socket is at SELV levels prior to touching, or alternatively wear appropriate insulated gloves.
The instrument should not be used if any part of it is damaged.
Test leads, probes and crocodile clips must be in good order, clean and with no broken or cracked
insulation.
Check that all lead connections are correct before making a test.
A Fused Lead Set must be used to connect to energised live systems. Refer to the accessories section for options.
Disconnect the test leads before accessing the battery compartment.
Refer to operating instructions for further explanation and precautions.
Safety Warnings and Precautions must be read and understood before the instrument is used.
They must be observed during use.
2
NOTE
THE INSTRUMENTS MUST ONLY BE USED BY SUITABLY TRAINED AND COMPETENT PERSONS.
CONTENTS
3
GSafety Warnings 3
Introduction 5
General description Battery replacement
User Controls and Display 8
Layout and functions
Operation 10
General Testing procedure Line Feed Connection to cable under test Measuring distance to fault Menu TX Null Velocity Factor Pulse widths Techniques to improve accuracy
Care and Maintenance 15
Specifications 16
Symbols used on the instrument are:
G Caution: Refer to accompanying notes. F Equipment protected throughout by
double or reinforced insulation.
t Instrument flash tested to 3.7kV rms. c Equipment complies with current EU
directives.
INTRODUCTION
Thank you for purchasing the CFL510F cable fault locator. Before attempting use of your new instrument please take the time to read this user guide, ultimately this will save you time, advise you of any precautions you need to take and could prevent damage to yourself and the instrument.
The CFL510F is an advanced instrument capable of identifying a wide range of cable faults. The instrument uses a technique called Pulse Echo (also known as Time Domain Reflectometry or TDR). A pulse is launched into a cable from one end. This can be on either a pair of conductors, or a conductor and the screen. The pulse travels down the cable at a velocity determined by the insulation between the conductors and this resistance to the flow of the pulse is characterised as impedance for the cable. Any changes in cable impedance will cause a proportion of the pulse to be reflected.
The pulse velocity is normally described as a fraction of the speed of light and is called the Velocity Factor. By measuring the time between the transmitted pulse and the reception of the reflected pulse, and multiplying this by the speed of light and the velocity factor, the actual distance to the reflection point can be established.
Reflections are caused by changes in the cables characteristic impedance, such as poor joints or discontinuities. Faults showing an impedance higher
than that of the cables normal impedance will cause a reflection of the same polarity, i.e.positive, whilst faults with an impedance lower than that of the cable will cause an inverse negative going reflection. Matched cable terminations absorb all the pulse hence no “end of cable” reflection will occur, the cable appearing endless. Open or short circuits will reflect all the pulse and a large reflection will be displayed. At an open or short circuit all the transmitted energy is reflected and the CFL510F will not ‘see’ the cable beyond that fault.
As a pulse travels down a cable, the size and shape of that pulse is gradually attenuated by the cable. The pulse reduces in amplitude and becomes more elongated or stretched. The level of attenuation (or losses) is determined by the cable type, the condition of the cable and any connections along its length. The limit of how far you can see is determined by the point beyond which you will not be able to see or distinguish a reflection. To help identify small reflections, especially at greater distance the CFL510F has an adjustable gain setting. By increasing the gain small reflections become more obvious.
The CFL510F can be used on any cable consisting of at least two insulated metallic elements, one of which may be the armouring or screen of the cable. The CFL510F has internal matching networks to allow testing of 25, 50, 75 and 100 cables. (These typically correspond to power, coaxial data and data/telecoms
4
cable). By selecting the CFL510F impedance closest to that of the cable under test, maximum power can be transmitted into the cable allowing long cables to be tested.
The velocity factor of the CFL510F must be adjusted to match that of the cable under test, allowing an accurate distance measurement to be read directly from the instrument. Where the VF of a cable is not known, but the length is, the cursor can be set to the end of the cable and the VF on the CFL510F adjusted until the correct cable length is displayed.
Other configuration settings include changing the distance units between metres and feet, changing the propagation velocity units between a ratio and a distance per microsecond. Display contrast is fully adjustable to compensate for all viewing conditions. A backlight aids viewing in low ambient light conditions.
The instrument can be powered by manganese-alkali, nickel-cadmium or nickel-metal-hydride batteries. All cells must be of the same type.
5
When the low battery symbol appears in the display window the cells are nearly exhausted and should be replaced as soon as possible. Use alkaline cells IEC LR6 (AA) 1.5V or 1.2V rechargeable cells only.
To install or replace the cells, switch the instrument off. Disconnect the test leads, loosen the battery cover retaining screws and remove the cover. Lift out and disconnect the battery holder.
Replace the cells, ensuring that correct polarity is observed (shown on the battery holder).
Incorrect battery cell polarity can cause electrolyte leakage resulting in damage to the instrument
Refitting the battery holder is the reverse of removing it.
BATTERY FITTING AND REPLACEMENT
6
7
USER CONTROLS AND DISPLAY
The controls of the CFL510F have been arranged such that the instrument is easy to learn and use. The instrument controls consist of the following
Instrument Display:
The display shows the user the current settings of the instrument and the reflected energy trace from the cable connected.
Tx Null: This rotary control allows the user to match the internal balance circuit of the CLF510F to that of the cable under test. When correctly adjusted the majority of the displayed transmitted pulse can be nulled out, allowing cable features close to the start of the cable to be identified. Refer to section TX Null for further details.
Cursor Left: Moves the ‘distance’ cursor to the left. Auto repeats if held down.
Cursor Right: Moves the ‘distance’ cursor to the right. Auto repeats if held down
Transmission pulse
Range
Distance to cursor
Impedance
Gain
Reflection
Find key: Automatically moves the cursor to the first potential fault (change of impedance).
Power: Pressing this button will turn the instrument on or off depending on its current state.
Range: A bi-directional button allowing range adjustment from 10m (30ft) to 3km (10 000ft) in 6 steps. The bi-directional feature allows ranges to be selected forward and backward, without having to scroll through all the ranges in one direction
Gain: This control is used to increase or decrease the gain of the instrument. This helps the user identify faults over the entire cable length.
Velocity Factor: Adjusts the Velocity Factor (VF) of the instrument to match that of the cable under test.
IMPORTANT: The VF value must match the VF of the cable under test to give an accurate distance measurement.
Menu: Allows the user to change the units of measurement for Distance, Velocity Factor, Impedance and Test Rate. Refer to menu section later in this text.
Backlight: Pressing this button will toggle the backlight on and off. The backlight will automatically switch off after 1 minute
Contrast: Allows the user to manually correct the display contrast to suit local ambient lighting conditions.
Output Sockets: 4mm shrouded sockets designed to accept the leads supplied with the instrument.
Battery cover: Located on the back of the instrument and provides the user with access to the battery compartment. The cover must not be removed while the instrument is switched on or connected to a cable. The instrument must not be operated with the cover open.
USER CONTROLS AND DISPLAY
8
9
OPERATION
General Testing procedure
Ensure the correct test leads are firmly fitted into the sockets of the instrument.
Switch on the instrument. The CFL510F will display the start screen for a couple of seconds, followed by a trace. The instrument will be set to the range and velocity factor last used. If these settings are different for the cable under test (C.U.T) then use the
RANGE
and VF keys to set the correct values. Refer also to ‘Velocity factor’ later in this text.
Connection to Cable Under Test
Connect the test lead to the cable under test. Connection may be made to a live system with a voltage to earth (ground) less than 300V with an installation (over voltage) category of III or lower. This means that the instrument may be connected to any fixed wiring of a building installation, but not to primary supply circuits such as overhead cables.
A Fused Lead Set must be used to connect to energised live systems. Refer to the accessories section for options.
When selecting 25 cable impedance, an internal 50Hz/60Hz filter is automatically switched in.
Measuring distance to fault
Ensure that the whole length of the cable can be seen on the display, and that the range selected is correct.
Find key
The Find key searches for the furthest major change of impedance. This may be a fault or the end of the cable if this is a short or open circuit.
When found the event is displayed and the cursor placed on the leading edge. The gain, zoom and cursor position may need to be adjusted for accurate measurement. Note that Find does not adjust the TxNull or Velocity Factor.
To enable ‘difficult to see’ faults to be identified, the gain of the instrument can be adjusted. With the gain at minimum small changes of cable impedance may be overlooked. By increasing the gain the fault becomes more obvious.
If no faults are obvious, adjust the gain until any major reflections appear. Open and short circuits should easily be identified. Partial faults can be less obvious.
If no significant reflections can be seen use the “TX Null” to minimise the effect of the output pulse on any potential “near end” faults (refer to section on TX Null).
In the event of no reflections being visible, increase gain until any reflection can be easily identified.
(If no reflections can be seen, try shorting or earthing the far end of the cable to ensure that you are “seeing” the whole length of the cable).
The cursor can also be moved using the LEFT and RIGHT cursor keys. Move the cursor to the beginning of the reflection. The distance to the fault can then be directly read from the display.
NOTE: The distance calculation is performed using the velocity factor set in the CFL510F. If this velocity factor is not correct for the circuit under test, the displayed distance will be incorrect.
Opposite are shown two typical trace displays. The top shows an open circuit at 2000m away; the second a short circuit at 2000m away.
OPERATION
10
Instrument Features Menu
The Menu key allows the user to change the units of measurement for Distance, Velocity factor, Impedance and Test Rate.
By pressing the
Menu key the CFL510F will display:
To select a function use the GAIN key to scroll up or down the menu.
To change the units us the LEFT or RIGHT cursor keys.
OPERATION
11
Available options Description Options
Distance Unit The Units in which the distance to m (metres) or ft(feet).
fault is displayed.
VF Format CFL510F velocity factor RATIO, m/us or ft/us
(which should be adjusted to match the cable under test)
Impedance Output impedance of the CFL510F 25/50 75/100
Test Rate Refresh rate 1/second or 3/second
Distance Unit m VF Format RATIO Impedance 75 Test rate 1s
TX Null
Without the “TX Null” control, the transmitted pulse would be visible at the beginning of the trace, swamping any reflections within the pulse length (the dead zone). The ‘TX Null’ circuit matches the characteristic impedance of the cable under test to produce an equivalent pulse. Subtracting this equivalent pulse from the transmitted pulse effectively removes the dead zone from the display and allows “near end” reflections to be seen.
NOTE: In some cases, it may not be possible to completely null the transmitted pulse.
Velocity Factor
The velocity factor is used by the instrument to convert the measured time for a pulse to be reflected, into a distance. It can be displayed as a ratio of the speed of light (eg 0.660 = 66% of the speed of light), or as a distance per microsecond in ft/us or m/us.
If the exact length of cable is known and the reflection from the cable end is visible then an accurate velocity factor can be determined:
Locate the reflection caused by the end of the
known length of cable with the instrument set on the shortest possible range to see the end of the cable.
Locate the start of this reflection as described in
the Operation section of this manual.
Adjust the velocity factor until the correct cable
length is shown.
Note the VF value for future reference.
The measurement of the distance to the fault can now be made with more confidence. The ability of the instrument to accurately measure the distance to a cable feature relies on the velocity factor being correct. Any errors in the velocity factor are directly related to distance measurement errors.
OPERATION
12
OPERATION
13
Pulse Widths
As the RANGE and IMPEDANCE matching of the CFL510F are adjusted so the duration of the transmitted pulse changes. The pulse widths range from 7ns to 3µs to overcome signal attenuation and enable the instrument to see further down a length of cable. The greater the range selected on the CFL510F, the wider the transmitted pulse.
The accuracy of the “Distance to fault” is not affected by the length of the pulse. However, if two or more features exist close together (excluding open or short circuits), then the second or subsequent feature may be partially masked by the reflection from the first fault. Hence, for potential multiple features, the instrument should be used with the shortest suitable range, and so the smallest pulse width, that can see both features.
For output pulse characteristics, refer to output pulse data in the CFL510F Specification at the end of this guide.
Techniques
To improve on the accuracy of the measurement, numerous techniques can be used, depending on the situation encountered. Not every situation can be described, but the following points are effective and the most common and easily implemented methods.
Test the cable from both ends
When fault finding a cable it is good practice to take measurements from both ends. Particularly in the case of open circuit faults, when the true end of the cable is not visible. If the measurement is made from both ends, then the combined answer should equal the expected length of the cable. Even when the true end of the cable is visible on the display, the reflections after the fault may be too obscure to analyse clearly. In this case, measurement from both ends yields a clearer picture as well as improved accuracy.
It is also good practice to follow the cable route with a cable tracer, as not all cable runs will be straight. It can save a great deal of time if the exact route of the cable is known as faults are often found at these points and can be accredited to third party intervention.
Care and Maintenance
Other than replacing the batteries, the instrument has no user serviceable parts. In case of failure it should be returned to your supplier or an approved Megger Limited repair agent.
Cleaning the instrument should only be done by wiping with a clean cloth dampened with soapy water or Isopropyl Alcohol (IPA).
Except where otherwise stated, this specification applies at an ambient temperature of 20C.
General
Ranges: 10m, 30m, 100m, 300m, 1000m, 3000m
(30ft, 100ft, 300ft, 1000ft, 3000ft, 10000ft)
Accuracy: 1% of range pixel at 0.67VF
[
Note- The measurement accuracy is for the indicated
cursor position only and is conditional on the velocity factor being correct.]
Resolution: 1% of range.
Input Protection: This instrument complies with
IEC61010-1 for connection to live systems up to 300V CAT III. Fused leads must be used if the voltage between the terminals exceeds 300V.
Output pulse: 5 volts peak to peak into open circuit. Pulse widths determined by range and cable impedance:
25 50 75 100
10m
7ns 7ns 7ns 7ns
30m 20ns 20ns 20ns 20ns
100m 100ns 60ns 100ns 100ns
300m 300ns 120ns 170ns 300ns
1000m 1000ns 520ns 680ns 1000ns
3000m 3000ns 2020ns 2340ns 3000ns
Gain: Set for each range with four
user selectable steps.
Velocity Factor: Variable from 0.01 to 0.99
in steps of 0.01
Impedance matching: 25, 50, 75, 100
Tx Null: 0 to 120
Refresh Rate: Selectable from the menu.
Once per second or three times a second.
Power Down: Automatic after 5 minutes with
no key press.
OPERATION
14
SPECIFICATIONS
15
Backlight: Stays on for 1 minute when activated.
Batteries: Six LR6 (AA) type batteries, Manganese-
alkali or nickel-cadmium or nickel-metal­hydride cells
Nominal voltage:
9V for Alkali or 7.2V for NiCad.
Battery consumption:
100mA nominal, 140mA with backlight. (20/30 hours continuous use depending on backlight use)
Safety: This instrument complies with IEC61010-1
for connection to live systems up to 300V CAT III. Fused leads must be used if the voltage between the terminals exceeds 300V.
EMC: Complies with Electromagnetic
Compatibility Specifications (Light industrial) BS EN 61326-1, with a minimum performance of ‘B’ for all immunity tests.
Mechanical
The instrument is designed for use indoors or outdoors and is rated to IP54.
Case Dimensions: 230 mm long (9 inches)
115 mm wide (4.5 inches) 48 mm deep (2 inches)
Instrument weight: 0.6kg (1.32lbs)
Case material: ABS
Connectors: Two 4mm-safety terminals.
Test Lead: 2 metres long consisting of
2 x 4mm shrouded connector to miniature crocodile clips
Display: 128 64 pixel Graphics LCD.
Environmental: Operational Temperature:
-15C to +50C (5F to 122F) Storage Temperature:
-20C to 70C (-4F to158F)
ORDERING INFORMATION
16
CFL510F Included Accessories
Test & Carry case with strap 6420-125 Miniature Clip Test Lead Set 6231-652 User Guide 6172-659
Optional Accessories
Fused prod and clip set 6111-218
For use on energised systems rated up to 300V CATIII use the fused prod and clip set Megger Part Number 6111-218, must be used.
17
18
M
Megger Limited Archcliffe Road, Dover Kent CT17 9EN England T (0) 1 304 502101 F (0) 1 304 207342
Megger
4271 Bronze Way, Dallas, Texas 75237-1088 USA T 1 800 723 2861 T 1 214 333 3201 F 1 214 331 7399
Megger Z.A. Du Buisson de la Couldre 23 rue Eugène Henaff 78190 TRAPPES France T +33 (1) 30.16.08.90 F +33 (1) 34.61.23.77
OTHER TECHNICAL SALES OFFICES Toronto CANADA, Sydney AUSTRALIA, Madrid SPAIN, Mumbai India and the Kingdom of BAHRAIN.
Megger products are distributed in 146 countries worldwide.
This instrument is manufactured in the United Kingdom. The company reserves the right to change the specification or design without prior notice.
Megger is a registered trademark
Part No. 6172-660 V03 Printed in England 1103 www.megger.com
Loading...