Megger BMM2500 User Manual

This instrument is manufactured in the United Kingdom. The company reserves the right to change the specification or design without prior notice.
MEGGER is the registered Trade Mark of AVO INTERNATIONAL LIMITED. Copyright ©, AVO INTERNATIONAL LIMITED
Part No 6172-484 Edition 3 – Printed in England – 01HH
AVO INTERNATIONAL
Archcliffe Road PO Box 9007 4271 Bronze Way MEGGER SARL Dover Valley Forge Dallas 29 Allée de Villemomble Kent, CT17 9EN. PA 19484-9007 TX 75237-1017 93340 Le Raincy England. U.S.A. U.S.A. Paris, France
Tel: +44 (0) 1304 502100 Tel: +1 (610) 676-8500 Tel: +1 (800) 723-2861 (U.S.A. only) Tel: +33 (1) 43.02.37.54
Fax: +44 (0) 1304 207342 Fax: +1 (610) 676-8610 Tel: +1 (214) 330-3203 (International) Fax: +33 (1) 43.02.16.24
MEGGER
®
User Guide Guide de l’utilisateur B e d i e n u n g s a n l e i t u n g
I N S U L ATION MULTIMETERS
MEGGER BMM2500
CONTENTS
2
Symbols used on the instrument are:
Risk of Electric Shock
Refer to User Guide.
Equipment protected throughout by Double Insulation (Class II).
Equipment complies with current EU Directives.
Equipment must not be connected to installations >500V.
>500V
Safety Warnings 3
General Description 4
Continuity Testing 9 Continuity Bleeper 10 Millivolt Tests 11 Capacitance Tests 12 Milliamps Tests 12 Storing Results 13 Datalogging 16 Setup Modes 18
Specification 20 Repair and Warranty 25
3
SAFETY WARNINGS
Safety Warnings and Precautions must be read and understood before the instrument is used.
They must be observed during use.
The circuit under test must be de-energized and isolated before connections are made except for
voltage measurement.
Circuit connections must not be touched during a test.After insulation tests, capacitive circuits must be allowed to discharge before disconnecting the
test leads.
The Live Circuit Warning and Automatic Discharge are additional safety features and should not
be regarded as a substitute for normal safe working practice.
Replacement fuses must be of the correct type and rating. Failure to fit the correctly rated fuse
will result in damage to the instrument in the event of an overload.
Test leads, including crocodile clips, must be in good order, clean and have no broken or
cracked insulation.
Ensure that hands remain behind guards of probes/clips when testing.U.K. Safety Authorities recommend the use of fused test leads when measuring voltage on high
energy systems.
NOTE
T H E I N S T R U M E N T S M U S T O N LY B E U S E D B Y S U I TA B LY T R A I N E D A N D C O M P E T E N T P E R S O N S
. U S E R S O F T H I S E Q U I P M E N T
A N D
/O R T H E I R E M P L O Y E R S A R E R E M I N D E D T H AT HE A LT H A N D SA F E T Y LE G I S L AT I O N R E Q U I R E T H E M T O C A R R Y O U T VA L I D
R I S K A S S E S S M E N T S O F A L L E L E C T R I C A L W O R K S O A S T O I D E N T I F Y P O T E N T I A L S O U R C E S O F E L E C T R I C A L D A N G E R A N D R I S K
O F E L E C T R I C A L I N J U RY S U C H A S F R O M I N A D V E RT E N T S H O R T C I R C U I T S . W H E R E T H E A S S E S S M E N T S S H O W T H AT T H E R I S K I S S I G N I F I C A N T T H E N T H E U S E O F F U S E D T E S T L E A D S C O N S T R U C T E D I N A C C O R D A N C E W I T H T H E HSE G U I D A N C E N O T E G S 3 8
‘ EL E C T R I C A L TE S T EQ U I P M E N T F O R U S E B Y EL E C T R I C I A N S S H O U L D B E U S E D .
GENERAL DESCRIPTION
4
The MEGGER®BMM2500 Series instruments are
battery powered Insulation and Continuity testers, with a measurement capability from 0,01 Continuity to 200G Insulation.
Offering multi-voltage facilities, the instruments take full advantage of microprocessor technology
and feature a large liquid crystal display combining digital and analogue readings. The analogue dis­play has the benefit of indicating trends and fluctu­ations in readings, while the digital readout gives direct accurate results. The display is also backlit
giving clear visibility even in low light conditions.
The BMM2500 Series instruments have the unique
capability of being able to measure voltages down
to a resolution of 0,1mV. This gives the user the option to fit a wide variety of transducers to further
enhance the capabilities of the BMM Series instru­ments, eg temperature or humidity measurement.
A customised connector on the top of the instru­ment enables the optional M E G G E R®S P 1
Switched probe to be used for two handed probe
operation.
The 250V, 500V and 1000V ranges can be used to test electrical installations in compliance with BS7671 (16th Edition IEEE Wiring Regulations) IEC364 and HD384, since each range has a 1mA minimum test current at the minimum pass values of insulation specified in these documents. The 100V range (BMM2580) is ideal for testing telecommunications equipment which would be damaged by higher voltages. The 50V range (BMM2580) is useful for testing sensitive equip­ment, such as electronic components, and com­puter peripherals.
The BMM2500 series instruments have a current facility which enables up to 500mA to be mea­sured, this together with , V and mV ranges means that the instrument can realistically be used in situations where previously a multimeter would be needed. For higher currents an optional current clamp is available.
5
The BMM2500 series instruments have an RS232 interface built in, and will allow the stor­age and download of results to a PC. The instru­ments come with all the necessary software to download and tabulate the results. The down­load format is also AVO Powersuite compatible enabling instruments to form part of an integrat­ed test and certification system. The BMM2500 series instruments also have the capability to data-log results over an extended time period so that long term measurements of systems can be taken.
Designed to IEC1010-1 the BMM2500 Series are p rotected against connection to a 500V Category III supply. The instruments have a basic accuracy of ±2% at 20°C. The instruments are waterproof and dustproof to IP54. This helps maintain accuracy and ensures maximum relia­bility in harsh environments.
OPERATION
6
Refer to Safety Warnings before using the instrument
Testing is automatically inhibited if:
An external voltage >25V is present when switched to any insulation range position.
An external voltage >10V is present on all other ranges (excluding OFF/V/mV).
The external voltage is indicated on the display, on insulation ranges an audible bleeper will sound if a test is attempted.
Live Circuit Warning
When more than 25V is applied to the terminals in the insulation ranges, the instrument defaults to a voltmeter and gives an audible warning if a test is attempted. On all other switch positions except O F F / V / m V when more than 10V is applied the default voltmeter will be activated. Testing will be inhibited.
Voltage Testing on High Energy Systems
Use extreme care when using or measuring volt-
ages above 30V, particularly in high energy sys­tems. Fused test leads are available as optional accessories for local situations where increased protection is required.
Auto-shut Off
To conserve battery life, Auto-shut Off (pre c e d e d by a series of bleeps) operates after appro x . 1 0 minutes of instrument inactivity on insulation, 5 minutes on all other ranges. If the instrument is switched on while holding the k e y, the Auto s h u t - O f f time is extended to 60 minutes. To re s t o r e operation after Auto-shut Off, select O F F followed by the re q u i red switch position.
Note: It is recommended that the instrument is switched to the OFF position when not in use.
Backlight
The backlight is activated by pressing the
key. The backlight will remain illuminated for approx. one minute before automatically switching off to conserve battery life, alternative­ly the key can be re-pressed.
7
Insulation Tests (M) (See fig.1)
The insulation tests apply a known volt­age to the circ u i t under test and m e a s u re the re s u l t­ing leakage curre n t . The circuit under test must be com­pletely d e - e n e r- g i s e d and i s o l a t e d b e f o re test connec­tions are made.
Insulation tests are only initiated when the TEST but- ton is pressed.
1. Set the range switch to the test voltage required.
2. Connect the test leads, first to the instrument, and then to the isolated item under test.
3. Press the TEST button to activate the test voltage. Take the reading.
4. Release the TEST button at the end of the test. ( P ress the TEST button if the l t b f e a t u re is
enabled). The last reading will hold on the display.
5. Any capacitive circuits charged during a test will automatically discharge. If significant voltage remains the voltage warning will occur and the voltage present displayed.
6. Remove the test leads only when no voltage is indicated.
Locking Test Button (ltb)
When it is desired to do a long insulation test, the test can be ‘locked on’ by pressing the XXXXX key while the test button is held down. The warning will appear on the display and both buttons may be released whilst the test continues. The next press of the test button will terminate the test.
Note: There is a short delay on the first operation of ‘1000 V’ range, each time the range is selected. This is to prevent accidental application of 1kV.
The M range features a leakage current display. Leakage current is the value of current that flows during the insulation test. To view the leakage cur­rent press the XXXXX key. To view insulation resis­tance press the XXXXX key again.
FIG.1
8
Good procedure whilst Insulation Testing
Care must be taken when taking measurements greater than a few G. The leads must be clean
dry and in good condition. They must also not be allowed to tangle. It is also advisable that the
switched probe SP1 is not used as the accuracy at
high value measurements is not guaranteed. The instrument should also be clean and dry with par­ticular attention paid to the terminals. Also attempt to reduce any leakage that may give erroneous
results on the item under test.
Polarisation Index Testing
Polarization Index (P I) is the term applied to the
Dielectric Absorption Ratio when resistance values
a r e measured after 1 minute and again after 10 min-
utes. Polarization Index is then the resistance value
after 10 minutes divided by the resistance value
after 1 minute. The test can be run at any voltage.
M o re detailed information on P I Testing and value assessment can be found in AVO International pub­lications listed in the Accessories page.
Automatic Discharge
When the TEST button is released after an insula­tion test (or re-pressed if ltb feature is enabled), a 200k load is automatically switched across the terminals to discharge the item under test. Any voltage present will be indicated on the display so that the discharge can be monitored.
Typical Terminal Voltage Characteristics
Continuity Testing () (See fig.2)
The continuity tests are activated when the probes make contact of less than a few k. The test oper­ates without the need to press the TEST button. When the test leads are removed the reading will hold for a few seconds and then reset. To recall the last result press the XXXXX key. This range is not suitable for diode testing since the automatic con­tact detector will not be activated when connected to a diode.
1. Set the selector switch to .
2. Connect the test leads.
3. The test will activate automatically.
4. After the test probes are disconnected, the reading will be held for a few seconds.
Zeroing of Test Lead Resistance
The resistance of the test leads can be nulled on the continuity range (up to 9,99Ω). The null information is retained in non-volatile memory and so will be re m e m b e red when the instrument is switched off .
1. Select the Continuity range.
2. Short the test leads across a known good conductor using prods.
3. When the reading has stabilised, press the TEST button. The z e ro offset symbol
will appear.
4. To release the zero offset press the test button again.
The continuity range f e a t u res a range lock f a c i l i t y. To LOCK the continuity range press the
key, the LOCK symbol will appear. To scroll
through the available ranges press the key. To de-select the LOCK feature hold the key down.
9
FIG.2
10
Possible sources of error
Measurements and results can be effected by the following:
The impedance of operating circuits connect­ed in parallel
Impedance such as inductors that vary during the measurement
A poor connection to the circuit under test.
Continuity Bleeper
The continuity bleeper sounds continuously when
less than 5 is detected. Short bleeps will sound
for resistances lower than a few k and above 5.
1. Set the selector switch to
2. Connect the test leads.
Display: Audible:
<5 continuous bleep <3k short bleep >3k no bleep
Resistance Tests (k)
This is a low voltage (5 V) low current (25 µA) test for sensitive electronic equipment. It operates in
the same way as the continuity ranges.
1. Set the selector switch to k.
2. Connect the test leads.
3. The test will activate automatically. The k range features a range lock facility. To
LOCK the k range press the key, the LOCK symbol will appear. To scroll through the available ranges press the key. To de-select the L O C K f e a t u re hold the key down.
The resistance range is protected by a high imped­ance method and therefore if the instrument is connected to a live circuit the fuse will not blow as on the insulation, continuity and buzzer ranges. The instrument will merely indicate the applied over-voltage.
Diode Testing
This range can also be used for diode testing, the positive terminal being the source of the test cur­rent. A forward biased semi-conductor junction will typically measure 15 to 30k, and a reverse junc­tion much higher. These features together with the small test current and wide measurement range
11
(0,01k to 10000k) make the resistance range very useful for general purpose testing.
Voltage Tests (V)
If >1V a.c. or d.c. is present at the terminals the measured voltage is indicated on the display. The voltage display will function within specification even if the fuse has blown. If the voltmeter operation is in question, test the voltmeter on a known source.
1. Set the selector switch to V.
2. Connect the test leads.
3. After a short settling time, the reading will be displayed automatically.
To view the frequency of the a.c. V being measure d p ress the k e y. The frequency will be displayed in the range 16Hz-460Hz. To view a.c. V press the
key again.
Millivolt Tests(mV)
The measured a.c. or d.c. voltage is indicated on the display.
1. Set the selector switch to mV.
2. Select either a.c. or d.c. mV using the key.
3. Connect the test leads.
4. After a short settling time, the reading will be
displayed automatically.
Note: Live circuit warning does not function on mV range.
Zeroing of d.c. mV (no a.c. mV zero facility)
To zero the d.c. mV range, short the leads togeth­er in the d.c. mV position, wait for the reading to settle and then press the TEST button. Up to 9,9mV can be zeroed on the d.c. mV range. The symbol will appear to indicate the zero has been adjusted.
1. Select the d.c. mV range.
2. Short the test leads together.
3. When the reading has stabilised, press the
TEST button. The zero offset symbol will appear.
4. To release the zero offset press the TEST but-
ton again. To view the frequency of the a.c. mV being mea­sured press the key. The frequency will be displayed in the range 16Hz-460Hz. To view a.c.
12
mV press the key again. For inputs less than 10mV frequency is not displayed.
Capacitance Tests (uF) (BMM2580 only)
The measured capacitance is indicated on the display.
1. Set the selector switch to uF.
2. Connect the test leads to the circuit under test.
3. After a short settling time, the reading will be displayed automatically.
Zeroing of uF
To zero the uF range, disconnect the leads, wait for
the reading to settle and then press the TEST but- ton. Up to 10,0nF can be zeroed on the uF range.
The symbol will appear to indicate the zero has been adjusted.
1. Select the uF range.
2. Disconnect the test leads from the circ u i t under test.
3. When the reading has stabilised, press the TEST button. The zero offset symbol will appear.
4. To release the zero offset press the test button again.
The range is suitable for the testing of discrete components and short low interference level signal lines. If electrolytic capacitors are being tested then the red lead should be connected to +ve of the capacitor. This range is not suitable for check­ing capacitance of signal lines which are subject to high levels of a.c. interference. When the test is started – – – will show on the dis­play, if there is excessive noise this symbol will remain or flash indicating that there is too much noise for a result to be reached.
Milliamps Tests (mA)
Because of the low source impedance associated with current measuring this test has an added fea­ture ensuring that when the range is first entered the default voltmeter is visible. Testing will be inhib­ited if more than 25V is present at the terminals. To start testing the TEST button should be pressed and held down for approx. 2 seconds to activate the mA range. Once activated, the TEST button no longer needs to be used and the measured value will be displayed automatically. To switch the display between a.c. and d.c. press the
13
key.
1. Set the selector switch to mA.
2. Connect the test leads.
3. Select either a.c. or d.c. m A using the
k e y.
4. Press and hold down the TEST button for
approx. 2 seconds.
5. After a short settle time, the reading will be displayed automatically.
To view the frequency of the a.c. mA being mea-
sured press the key. The frequency will be displayed in the range 16Hz-460Hz. To view a.c. mA press the key again. For inputs less than 10mA frequency is not displayed.
Storing Results on M and (RCL)
After an insulation test or continuity test the result is displayed on the screen, this may be saved with additional information. A circuit number (1-99) may be assigned and circuits may be grouped using the distribution board feature. In this way, when down­loading to AVO PowerSuite, the results can be eas­ily split into different test schedules. When the results are displayed or printed, a change in the distribution board is indicated.
Changing Distribution Boards (DB)
Before a test the distribution board number may be changed as follows:-
1. Move the rotary selector switch to the RCL position. The code rcl is displayed.
2. Press the key. The currently selected DB
code is displayed, e.g. d01.
3 . This number may be changed using the
and keys to display the re q u i red number.
4. The number can be accepted by pressing the
key, or aborted by pressing the EXIT button.
5. When the number is saved the code Std is
displayed (accompanied by a long bleep) to confirm that the data has been saved.
14
Testing may now continue with all subsequent
results associated with the new distribution board
number.
Storing a result
On completion and display of the measurement:-
1. Press and hold the key. After about 1 second, a bleep will be heard. For both Continuity and Insulation, a code, as given in the following table is displayed. This code is used to describe the circuit tested and can accordingly be modified by the user.
2. The code may be changed by pressing the
and keys.
3. The code may be accepted by pressing the key, or aborted by pressing the EXIT button.
4. The circuit number is then displayed as 2 digits e.g. c01.
Note:- Many different tests may be saved under
the same circuit number.
5. The circuit number may be changed by pressing the and keys to display an appropriate number.
6. The number can be accepted and the results saved by pressing the key, or the
procedure aborted by pressing the EXIT but­ton.
7. When the result is saved, the code Std is
displayed (accompanied by a long bleep) to confirm that the data has been saved. The display of FULL indicates that there is no more test storage. Approx. 300 results can be stored in memory.
Test to be Saved Display code Meaning
Continuity r1 Single CCT
r2 Single CCT rr1 Ring CCT rr2 Ring CCT rrn Ring CCT r12 R1+R2 ReturnCCT
Insulation n_e N-PE
L_n L-N L_E L-PE L_L L-L
15
Delete all data
1. Move the rotary selector switch to the RCL
position. The code rcl is displayed.
2. Press the and keys together. The
code dEL is displayed.
3. Confirm that the data is no longer required by pressing the key or abort by pressing any
other key. The code rcl is displayed.
Print Results (see Setup Modes)
1. Connect printer and the instrument with a ser-
ial printer lead.
2. Move the rotary selector switch to the RCL position. The code rcl is displayed.
3. Commence the printout by pressing the TEST button. Abort at any time by pressing and hold­ing the k e y . The code rc l is displayed.
Retrieve Stored Results
It is possible to view previously stored test results by switching the rotary switch to the RCL position.
1. Move the rotary selector switch to the RCL
position. The code rcl is displayed.
2. Select the re q u i red distribution board by
pressing the and keys.
The distribution board numbers are shown in the order that the results were stored. A long bleep is sounded when the end of the list is reached.
3. Press the key to list the circuit numbers used in the currently displayed distribution board or press the EXIT button to return to the RCL display.
4. Select the required circuit number by pressing the and keys. The circuit numbers are shown in the order that the results were stored. A long bleep is sounded when the end of the list is reached.
5. Press the EXIT button to return to the distribu­tion board selection screen, or press the
XXXXXkey to show the stored test codes. The
following codes are used to identify test results:-
Code Meaning
Con Continuity Test InS Insulation Test
6. Select the re q u i red test by pressing the
and keys. The tests are shown in
16
the above order. Hold a key down to auto­repeat. A long bleep is sounded when the end of the list is reached.
7. Press the EXIT button to return to the circuit
number selection screen, or press the key to scroll through the stored test results, together with any additional connection information.
Download to PC
The BMM series has been designed to be used
with AVO Powersuite for Windows which will accept the test results and enable the production of various certificates, including Periodic
Inspection and Completion. The CD supplied with
the instrument contains AVO Download Manager Program. This enables stored and data-logged
results to be downloaded to a PC, the creation of
simple test report files which may be exported to other applications and used to create data back­ups or reports/certificates. Download manager also enables certain changes to be made to your instrument setup such as changing the 2nd lan-
guage of the printouts. The CD contains a user
guide giving you full instructions on the use of Download Manager.
Cable Configuration
Normally, a double-ended 9-way ‘D’ female socket lead suitable for connecting PC to PC is required. This lead should not exceed 3m in length. A lead is available as an accessory, or one can be made up as follows:–
Datalogging
Datalogging means the automatic recording of measurements at regular intervals over an extend­ed period, for viewing at a later time. Results are held in the internal memory of the tester, but can only be extracted by means of a PC connected via the RS232 port.
Signal Insulation Tester 9-way ‘D’ 25-way ‘D’
Rx 2 3 2 Tx 3 2 3 DSR 6 4 20 GND 5 5 7
17
Storage v. Datalogging Comparison.
By storage, we mean the saving to memory of results one-by-one as tests are performed. Each result has to be individually saved after associating it with user-selectable connection data and circuit number, along with a previously chosen distribu­tion board number. However, when data logging is running, results are automatically and continuous­ly saved to memory and no other information is recorded.
Storage and logging are mutually exclusive func­tions. The instrument cannot be set up to perform both operations at any given time. If a result is stored, then all logged data is erased, and vice versa. Also, only one set of logged data can be held in memory. A new logging session will erase the previous data. This differs in behaviour from the storage function, in which data from successive tests accumulates in the memory until it is full. Another difference between storage and logging is that the former applies only to insulation and con­tinuity testing, whereas all types of measurement can be logged (buzzer range excluded).
Starting a logging session
Once logging has been enabled and the interval set up, a session of data-logging can begin. (see Setup Modes)
1. Hold down the key and turn the rotary switch to the desired function.
2. Release the key. The message log will be seen. To confirm that logging is required, p ress the display key and a confirming
message will be shown briefly.
3. Pressing EXIT will cancel logging and off will be displayed.
If data-logging is turned on, it will commence as soon as the test starts. That is, immediately on volts, millivolts, ohms, k and capacitance func­tions, or when the test button is pressed on insula­tion and milliamps ranges. On milliamps or milli­volts functions, press theXXXXXkey if required, after logging has started.
Stopping a logging session
While a logging session is running, it can be stopped at any time by turning the range switch. Any data logged up to that point will be retained.
18
Alternatively, logging can be left running and will stop automatically a short while after the memory becomes full.
The number of results which can be logged is about 300, after which the message ‘full’ will be flashed on the display for a few minutes. The instrument will then shut down.
Other considerations
Note that the batteries may not last through the whole logging session, depending on their condi­tion and on the log interval which has been set. If the batteries should fail, any results already logged will be retained. Note also, that if the TEST button has been set for non-latching operation, this will be temporarily overridden during logging of insulation results, giving a press-to-start, pre s s - t o - s t o p action. The TEST button’s mode of operation will return to its original setting after the logging ses­sion has finished.
Viewing logged data.
Logged results can only be retrieved via the instru­ment’s RS232 port, using a PC connected via a
serial cable. The disk supplied with the instrument carries a programme capable of handling the data.
The method is similar to that for ‘downloading’ stored results. For further information see the sec­tion entitled “Download to PC”. Note that on milli­volts d.c. and milliamps d.c. measurements, the frequency result is always set to zero. On millivolts a.c. and milliamps a.c. measurements, the fre­quency result is set to zero if the frequency could
not be measured.
Setup Modes
Serial Printer configuration
The printer should be set to 9600 baud, 8 bit data, no parity and 1 stop bit. This instrument uses a special isolated serial interface which is powered from the PC or printer. In the unlikely event that your PC or printer is not able to power the inter­face, it will be necessary to provide an additional supply. Contact AVO Product Support for details.
Printer Setup Mode
The instrument cannot respond to a busy signal
19
given by a printer, and therefore waits at the end of
each line. This time (“Printer delay time”) and the
printer report language can be changed.
1. Press and hold the key then turn the
rotary selector switch from the OFF position to the RCL position. The code Pdt is displayed.
2. Release the key.
To change the Printer speed
1. Toggle the key to scroll through and
display the code Pdt
2. Press the key. The current setting is displayed.
3. Toggle the and keys until the required setting is displayed.
4. To save the new setting, press the key.
The bleeper sounds and std is displayed. To
abort the new setting, press the EXIT button.
To select the printer language
1 . P ress the key to scroll through and dis-
play the code l n g.
2 . P ress the key. The current printer re p o r t
language is displayed as 1 (English) or 2 (as given on the type label on the User Guide cover).
Note: Language 2 can be changed – see below.
3. Press the key until the required language setting is displayed
4. To save the new setting, press the key.
The bleeper sounds and is Std displayed. To abort the new setting, press the EXIT button.
Changing the Second Printer Language
Language 2 can be changed using AVO Download Manager Program supplied with your instrument. Full instructions are contained in the User Guide with the CD.
Data-logging
Enabling/disabling data-logging: The factory setting is that logging is disabled and storage is enabled.
1. Turn the range knob to ‘RCL’ while holding down the key. Pdt will be displayed.
2. Press the key until log is showing
3. Press the key to show the current status of the logging function, i.e on or off.
4. Pressing the key toggles the setting between on and off.
5. Press the key to accept the setup, or the
EXIT button to quit.
20
Setting the data-logging interval.
1. While log is being displayed, press the
key. Int will appear on the display.
2. Press the key to see the value of the log­ging interval (in seconds)
3. Set the interval in 10 second steps with the and keys (max. value 1990 sec-
onds, min. value 10 seconds)
4. Press the key to accept the new value,
or the EXIT button to quit.
5. Press the EXIT button again to escape from
the logging setup screens. The display will s h o w rc l.
Using the MEGGER SP1 Switched Probe
Operation
The MEGGER SP1 is an accessory for designated
MEGGER installation test instruments. When fitted
in the specially designed connector, in place of the existing ‘Low’ lead, the SP1 acts as a remote test button to operate the instrument and as a ‘Low’
probe. This simplifies instrument control and two­handed probing. The SP1 is suitable for use with MEGGER insulation test instruments up to 1kV
output test voltage.
Safety
Meets the safety requirements for double insula­tion to IEC1010-2-031 (1995), EN61010-2-031 (1995), IEC 1010-1 (1995), EN61010-1 (1995) Category III* 300 V phase to earth and 500V phase to phase. The probe is fitted with an internal, non­replaceable fuse, to protect the user should the probe be used accidentally in conjunction with a testlead in the low terminal.
*Relates to transient overvoltage likely to be found in fixed installation wiring.
Do not use the probe if any part of it is damaged.
Battery Replacement
When the low battery symbol appears, the cells are nearly exhausted and should be replaced as soon as possible. Use Alkaline cells IEC LR6 (AA) or NiCd rechargeable. To install or replace the cells, disconnect the test leads, switch the instru­ment to OFF and loosen the captive screws on the rear of the battery compartment. Remove the cover and disconnect the battery holder from the battery leads. Ensure that the replacement cells are
21
fitted with the correct polarity in accordance with the label in the battery holder. Reconnect the bat­tery holder to the battery leads. Replace and re­secure the battery compartment cover. Remove the cells if the instrument is not going to be used for an extended period of time.
Fuse Checking and Replacement
To check the instrument fuse, switch to an insula­tion range and press the TEST button. The symbol
XXXX will appear if the fuse is ruptured. To replace
the fuse, disconnect the test leads, switch the instrument OFF and loosen the captive screws holding the battery compartment cover in place. Remove the cover and replace the fuse. Replace and re-secure the battery compartment cover.
Specification
22
(All quoted accuracies are at +20ºC.)
INSULATION RANGES
Nominal Test 250V,500V,1000V (BMM2500)
Voltage (d.c.): 50V,100V,250V,500V,1000V (BMM2580)
Test voltage accuracy: +15% maximum on open circ u i t Short circuit curre n t : < 2 mA
Test Current on load: 1mA at min. pass value of insulation
specified in BS7671, HD384 and IEC 364, 2mA max.
Accuracy:
(BMM2500)
Range Full Scale Accuracy
1000V 20G ± 2% ± 2 digits ± 0,2% per G 500V 10G ± 2% ± 2 digits ± 0,4% per G
250V 5G ± 2% ± 2 digits ± 0,8% per G
(BMM2580)
Range Full Scale Accuracy
1000V 200G ± 2% ± 2 digits ± 0,2% per G 500V 100G ± 2% ± 2 digits ± 0,4% per G
250V 50G ± 2% ± 2 digits ± 0,8% per G
100V 20G ± 2% ± 2 digits ± 2,0% per G 50V 10G ± 2% ± 2 digits ± 4,0% per G
Note: Above specifications only apply when high quality
silicone leads are being used.
Measuring Range:
0,01M to 200G (0-100G on analogue scale). EN61557 Operating range: 0,10M to 1,00G
Leakage Current: 10% ±3digits
CONTINUITY
Measuring Range: 0,01 to 99,9 (0 to 10 on
analogue scale)
EN61557 Operating range: 0,10 to 99,9 Accuracy: ±2% ±2 digits Open circuit voltage: 5V ±1V Test current: 210mA ±10mA (0-2) Zero offset at probes: 0,10 typical
Lead resistance zeroing: Up to 9,99 Noise rejection: 1V rms 50/60Hz Buzzer: Operates at less than 5
(approx).
RESISTANCE
Measuring Range: 0,01k to 9,99M (0 to
100M on analogue scale)
Accuracy: ±3% ±2 digits Open circuit voltage: 5V ±1V Short circuit current: 25µA ±5 µA
23
VOLTAGE
Measuring Range: ±1V to ±500V (0 to 1000V
on analogue scale)
Accuracy: 0-500V d.c. ±2% ±3 digit
0-500V a.c (50/60Hz) ±2% ±3 digits 0-500V 400Hz a.c. ±5% ±3 digits
Input resistance: approx 200k. Detector Threshold: 1V
MILLIVOLTS
Measuring Range: ±0,1mV to ±1999mV
(0 to 1000mV on analogue scale)
Accuracy:
0,1mV to 10mV d.c. or a.c. (50/60Hz) ±2% ±5 digits 10mV to 1999mV d.c. or a.c. (50/60Hz) ±2% ±3 digits 0,1mV to 10mV a.c. (16-460Hz) ±5% ±7 digits 10mV to 1999mV a.c. (16-460 Hz) ±5% ±5 digits
d.c. milliVolts zeroing: Up to 9,9mV Input resistance: >3M
CAPACITANCE
Measuring Range: 0,1nF to 9,99uF Accuracy: ±3% ±2 digits ±0,2nF uF zeroing: Up to 10nF
MILLI-AMPS
Measuring Range: 0,1mA to 500mA (0 to 1000mA on
analogue scale)
Accuracy:
0,1mA to 10mA d.c. or a.c. (50/60 Hz) ±2% ±5 digits 10mA to 500mA d.c. or a.c. (50/60 Hz) ±2% ±3 digits 0,1mA to 10mA a.c. (16-460 Hz) ±5% ±7 digits 10mA to 500mA a.c. (16-460 Hz) ±5% ±5 digits
FREQUENCY
Measuring range: 16Hz to 460Hz Accuracy: ±1% ±1digit
24
Basic and service errors for
Insulation and Resistance ranges.
The basic error is the maximum inaccuracy of the instru­ment under ideal conditions, whereas the service error is the maximum inaccuracy taking into effect of battery volt-
age, temperature, interference, and system voltage and frequency, where applicable. After determining the service error, we can then calculate the measurement range. This is the range of measurement over which the error in ser­vice is less than 30% of the reading. Digital instruments
are affected by the number of digits error – for example a value 0,10 measured with the continuity range may give a display in the range 0,07 to 0,13 which is a maximum error of 30%. Therefore the measurement range measur­ing low resistance is 0,10 to 99,9. When checking that
a measurement does not exceed a limit, the service error needs to be taken into account and these tables enable this to be done quickly and easily. These will guarantee that the value being measured is greater than or less than
the limit value specified as appropriate.
Insulation Resistance – M
Limit Min. Indicated Reading Limit Min. Indicated Reading 0,10 0,14 2,00 2,12 0,20 0,25 3,00 3,16 0,30 0,35 4,00 4,20 0,40 0,46 5,00 5,24 0,50 0,56 10,00 10,8 0,60 0,66 20,00 21,2 0,70 0,77 30,00 31,6 0,80 0,87 40,00 42,0 0,90 0,98 50,00 52,4 1,00 1,08 100,00 94,0
Continuity Resistance –
Limit Max. Indicated Reading Limit Max. Indicated Reading 0,10 0,06 2,00 1,88 0,20 0,15 3,00 2,84 0,30 0,25 4,00 3,80 0,40 0,34 5,00 4,76 0,50 0,44 10,00 9,56 0,60 0,54 20,00 18,8 0,70 0,63 30,00 28,4 0,80 0,73 40,00 38,0 0,90 0,82 50,00 47,6 1,00 0,92 100,00 92,0
Loading...
+ 58 hidden pages