AVO410
Digital Multimeter
AVO410
Digital Multimeter
 6000 count backlit digital display
 True RMS reading on AC mode
 1000 V DC / 750 V AC ranges
 10 A AC / DC ranges
 CAT IV 600 V
DESCRIPTION
The Megger AVO410 digital multimeter has been designed for the contracting electrician and has the additional features that also make the instrument suitable for wide range of applications and users.
The instrument offers AC and DC voltage and current measurements as well as resistance, frequency and capacitance ranges. True RMS readings on the AC functions are standard on the AVO410 and the instrument features a CATIV 600 V safety rating meaning the instrument is suitable for industrial applications.
The slim, compact case has a tough rubberised holster that provides that extra degree of protection from the extreme conditions found in industrial environments. The style of the case and positioning of the function switch and buttons means the unit sits comfortably in the palm for single handed use.
Continuous references to the user guide have been avoided by the AVO410 utilising simplified functions.
The display features a back light that allows measurements to be made in poorly lit areas.
The AVO410 test leads are supplied with silicon cable and have GS38 compliant shrouded tips on the prods.
Auto-ranging
When first selected, all functions are auto-ranging. A range button on the AVO410 allows multiple manual range selection on each function; a feature that is generally welcomed by many users.
Minimum / Maximum measurements
The instrument has a MIN MAX function that allows the user to switch between minimum and maximum measurements. The display does not have to be continually monitored to capture a momentary increase or fall in readings.
Data hold
This function allows a displayed result to be frozen on the display which avoids having to remember a measurement value. The hold function can be nested within the MIN MAX feature which stops the AVO410 continuously updating the minimum and maximum measurement values.
Voltage measurements
Both AC and DC voltage measurements up to 750 V and 1000 V respectively are possible with the AVO410, the AC reading being a true RMS value.
Current measurements
For current measurements up to 10 A, a separate fused terminal is provided to protect both user and instrument from excess current.
RS232
The AVO410 has optically isolated RS232 interface that allows the user to connect to a PC via a USB port for data acquisition and analysis. (Optional software is required for this function).
Continuity / diode testing
The continuity function features a buzzer and provides the user both optical and audio indication of identifying and confirming continuity between two points. This function also allows forward and reverse bias testing of diode and semiconductor junctions.
Resistance, capacitance and frequency
Resistance can measured directly on the ohms range from 0 to 60 MΩ with capacitance measurements from 0 to 6.000 mF. In addition, frequency measurements from 0 to 60 MHz are possible.
Page 2
SPECIFICATIONS
Display 6000 counts updates 1.5/sec.
Polarity Automatic, positive implied,
negative indicated
Over-range indication “OL” or “-OL”
Battery indicator Displayed when the battery
voltage drops below operating voltage
Auto power down Approx 10 minutes
Operating ambient Non-condensing ≤10 °C,
11 °C ~ 30 °C (≤80% R.H) 31 °C ~ 40 °C (≤75% R.H), 41 °C ~ 50 °C (≤45% R.H)
Storage temperature range and humidity
-20 °C to 60 °C, 0 to 80% R.H. when battery removed from meter
Temperature co-efficient 0.15 x (Spec.Acc’y) / °C,
<18 °C or >28 °C
Safety The instrument complies with
IEC61010 CAT IV 600 V
Power supply Standard 9 V battery PP3, NEDA
1604, IEC6F22, JIS006P
Battery life Alkaline 300 hours
Dimensions 76 mm x 158 mm x 38 mm
without holster 82 mm x 164 mm x 44 mm with holster
Weight 522 g
ELECTIRCAL SPECIFICATIONS
DC/AC volts
Range DC accuracy AC accuracy
600.0 mV ±(0.5% + 2 digits) 50 Hz/60 Hz sine wave only for 600.0 mV range, ±(0.9% +5 digits) 50 Hz ~ 500 Hz *
6.000 ±(0.5% + 2 digits) 50 Hz/60 Hz sine wave only for 600.0 mV range, ±(0.9% +5 digits) 50 Hz ~ 500 Hz *
60.00 V ±(0.5% + 2 digits) 50 Hz/60 Hz sine wave only for 600.0 mV range, ±(0.9% +5 digits) 50 Hz ~ 500 Hz * 600 V ±(0.5% + 2 digits) 50 Hz/60 Hz sine wave only for 600.0 mV range, ±(0.9% +5 digits) 50 Hz ~ 500 Hz * DC 1000 V ±(0.5% + 2 digits) 50 Hz/60 Hz sine AC 750 V wave only for 600.0 mV range, ±(0.9% +5 digits) 50 Hz ~ 500 Hz *
* The basic accuracy is specified for a sine wave below 4000 counts. Over 4000 counts, add 0.6% to the accuracy. For non-sine waves below 2000 counts, refer to the following for accuracy: ±1.5% addition error for C.F from 1.4 to 3
Protection 1000 V DC or AC
Input impedance 10 MΩ // less than 100 pF
Common mode rejection ratio / normal mode rejection ratio
(Common mode rejection ration/ normal mode rejection ratio) VAC: CMRR >60 dB at DC, 50 Hz/60 Hz VDC: CMRR >100 dB at DC, 50 Hz/60 Hz NMRR: >50 dB at DC, 50 Hz/60 Hz
2
Page 3
AC conversion type AC conversions are AC coupled
True RMS responding, calibrated to the sine wave input.
Crest factor C.F. = Peak/rms
DC/AC current
Diode check
Range Resolution Accuracy Diode 10 mV ±1.5% + digits*
* For 0.4 V ~ 0.8 V
Range DC AC Voltage accuracy accuracy burden
600.0 μA ± (1.0% + 2 digits) N/A <4 mV/μA 6000 μA ± (1.0% + 2 digits) N/A <4 mV/μA
6.000 A ± (1.0% + 2 digits) ±(1.5% +6 dgt) 2 V max 50 Hz ~ 500 Hz *1
10.00 A ± (1.0% + 2 digits) ±(1.5% +6 dgt) 2 V max 50 Hz ~ 500 Hz *1
Overload protection A input 10 A (500 V) fast blow fuse μA input 600 V rms
*1) AC conversion type
Conversion type and additional specification are the same as DC/AC voltage.
Resistance range
Range Accuracy Overload Protection
600.0 Ω *2 ± (0.7% + 2 digits) 600 V rms
6.000 KΩ ± (0.7% + 2 digits) 600 V rms
60.00 KΩ ± (0.7% + 2 digits) 600 V rms
600.0 KΩ ± (0.7% + 2 digits) 600 V rms
6.000 MΩ ± (1.0% + 2 digits) 600 V rms
60.00 MΩ*1 ± (1.5% + 2 digits) 600 V rms
Max. test current 1.5 mA Max. open circuit voltage 3 V Overload protection 600 V rms
Continuity
Built-in buzzer will sound when the resistance is less than 500 Ω approx. Response time is 100 ms approx.
Frequency measurement range
Range **Sensitivity Overload protection 6000 Hz 100 mV rms* Frequency: 0.1% ± 1 digit
60.00 KHz 100 mV rms* Frequency: 0.1% ± 1 digit
600.0 KHz 100 mV rms* Frequency: 0.1% ± 1 digit
6.000 MHz 250 mV rms Frequency: 0.1% ± 1 digit
60.0 MHz 1 V rms Frequency: 0.1% ± 1 digit
Overload protection 600 V rms * Less than 20 Hz, the sensitivity is 1.5 V rms ** Max. sensitivity <5 V ac rms
Accuracy
Range Accuracy
6.000 nF ± (1.9%) +8 digits)
60.00 nF ± (1.9%) +8 digits)
600.0 nF ± (1.9%) +8 digits)
6.000 μF ± (1.9%) +8 digits)
60.00 μF ± (1.9%) +8 digits)
600.0 μF ± (1.9%) +8 digits)
6.00 mF* ± (1.9%) +8 digits)
Overload protection 600 V rms
* <100 digit of reading rolling
Open circuit voltage -1.3 V approx.
*1<100 digit rolling *2 <10 digit rolling
ORDERING INFORMATION
Description Order Code
AVO410 digital multimeter CAT IV 600 V 1001-613
Included accessories
Test leads and probes
Optional accessories
Pouch 2007-366
UK
Archcliffe Road Dover CT17 9EN England T +44 (0) 1304 502101 F +44 (0) 1304 207342 UKsales@megger.com
UNITED STATES
4271 Bronze Way Dallas TX 75237-1019 USA T 800 723 2861 (USA only) T +1 214 333 3201 F +1 214 331 7399 USsales@megger.com
Auto power OFF (APO)
If idle for more than 10 minutes
OTHER TECHNICAL SALES OFFICES
Valley Forge USA, College Station USA, Sydney AUSTRALIA, Danderyd SWEDEN, Ontario CANADA, Trappes FRANCE, Oberursel GERMANY, Aargau SWITZERLAND, Kingdom of BAHRAIN, Mumbai INDIA, Johannesburg SOUTH AFRICA, Chonburi THAILAND
CERTIFICATION ISO
Registered to ISO 9001:2000 Cert. no. Q 09290
AVO410_DS_en_V03
www.megger.com