G SAFETY INFORMATION
Understand and follow operating instructions carefully.
G WARNING
■ Do not use test leads or the multimeter if they are damaged
■ Do not use the multimeter if the tester is not operating properly
■ To reduce risk of fire and shock do not expose this product to rain or
any moisture
■ When using test leads or probes, keep fingers behind the finger guards.
■ Remove test leads from meter before opening the battery door or meter
case.
■ Use the meter only as specified in this manual or the protection by the
meter might be impaired.
■ Always use proper terminals, switch position, and range for measurements.
■ Never attempt a voltage measurement with the test lead inserted into the
A input terminal.
■ Verify the meter’s operation by measuring a known voltage. If in doubt,
have the meter serviced.
■ Do not apply more than the rated voltage, as marked on meter, between
terminals or between any terminal and earth ground.
■ Do not attempt a current measurement when the open voltage is above the
fuse protection rating.
■ Suspected open circuit voltage can be checked with voltage function.
■ Only replace the blown fuse with the proper rating as specified in this
manual.
■ Use caution with voltages above 30 Vac rms, 42 Vac peak , or 60 Vdc.
These voltages pose a shock hazard.
■ To avoid false readings that can lead to electric shock and injury, replace
battery as soon as low battery indicator
■ Disconnect circuit power and discharge all high voltage capacitors before
testing resistance, continuity, diodes, or capacitance.
■ Do not use meter around explosive gas or vapour.
appears.
2
www.megger.com
G CAUTION
■ Disconnect the test leads from the test points before changing the position
of the function rotary switch.
■ Never connect a source of voltage with the function rotary switch
W , ,
■ Do not expose Meter to extremes in temperature or high humidity.
■ Never set the meter in power supply circuit in equipment that could result in damage the meter
and the equipment under test.
Symbols as marked on the meter and instruction card
d , a
,
A, b position.
A function to measure the voltage of a
da
F Risk of electric shock
G See instruction card
d DC measurement
t Equipment protected by double or reinforced insulation
Battery
a AC measurement
c Conforms to EU directives
Maintenance
Do not attempt to repair this Meter. It contains no user-serviceable parts. Repair or
servicing should only be performed by qualified personnel.
Cleaning
Periodically wipe the case with a dry cloth and detergent. Do not use abrasives or
solvents.
www.megger.com
3
AC / DC Voltage and Frequency
Continuity and Diode
4
www.megger.com
Resistance and Capacitance
Note – To improve the measurement accuracy of small value capacitor, record the
reading with the test leads open, then subtract the residual capacitance of the meter
and leads from measurement.
C
UNKNOWN
= C
MEASUREMENT
- C
RESIDUAL
MIN MAX Record
www.megger.com
5
DC / AC Current
Display Hold
6
www.megger.com
Manual Ranging and Auto Ranging
Auto Power Off (Battery Saver)
www.megger.com
7
Disable Auto Power Off
8
www.megger.com
Non-Contact Voltage Alert (VOLTSENSE)
1. Volt sense switch will be activated on any function or at OFF status.
2. Test leads are not used for the Volt sense test.
3. Press the Volt sense button. The display will blank out, a tone sounds
and the red LED lights up to verify the instrument is operational. The
Volt sense button must be held down to detect the presence of voltage
without use of the leads.
4. If a voltage of 50V to 600V (50 to 500Hz) is detected near the top of
meter a continuous tone sounds and the red LED near the top of meter
illuminates.
www.megger.com
9
Fuse Replacement
Refer to the following diagram to replace fuse.
Use only a fuse with the amperage, interrupt, voltage,
and speed rating specified.
G
˙ Fuse rating: 10A, 500V
10
www.megger.com
Battery Replacement
Refer to the following diagram to replace the batteries:
G
Replace the batteries as soon as the low batteries
Operating Ambient : Non-condensing ≤10 °C, 11 °C ~ 30 °C (≤80% R.H)
31 °C ~ 40 °C (≤75% R.H), 41 °C ~ 50 °C (≤45% R.H)
Storage Temperature : –20 °C to 60 °C , 0 to 80% R.H. when battery removed
from Meter.
Temperature Coefficient : 0.15 x (Spec.Accy) / °C ,< 18 °C or > 28 °C
Measure : Samples 2 times per second nominal.
Altitude : 6561.7 ft (2000 m)
Safety : Complies with EN61010-1, UL61010-1, IEC 61010-1,
V/Ω :CATIII 600 V, CATII 1000 V.
A :CATIII 500 V
Pollution degree : 2
Power Requirements : 1.5 V x 2 IEC LR03, AM4 or AAA size
Dimensions (W x H x D) : 74 mm x 156 mm x 44 mm
Weight : (320 g) including battery.
Accessories : Battery (installed), Test leads and user manual
is displayed when the batteries voltage drops
12
www.megger.com
Electrical Specifications
Accuracy is ± (% reading + number of digits) at 23 °C ± 5 °C < 80% RH.
DC / AC Volts
RangeAC Accuracy
200.0 mV *Unspecified
2.000 V *±(1.5%+5 dgt) 50 Hz ~ 300 Hz
20.00 V ~ 200.0 V *
750 V AC / 1000 V DC
DC Accuracy : ±(0.5% + 2 dgt)
Over voltage protection : 1000 V DC or 750 V ACrms.
Input Impedance : 10 MΩ // less than 100 pF.
* CMRR / NMRR : (Common Mode Rejection Ratio)
(Normal Mode Rejection Ratio)
VAC : CMRR > 60 dB at DC, 50 Hz / 60 Hz
VDC : CMRR > 100 dB at DC, 50 Hz / 60 Hz
NMRR > 50 dB at DC, 50 Hz / 60 Hz
AC Conversion Type :
Average sensing rms indication.
AC conversions are ac-coupled, true rms responding, calibrated to the sine wave
input. * The minimum LCD reading is 1400 count in Auto Ranging Mode.
Crest Factor : C.F. = Peak / Rms
+ 1.5% addition error for C.F. from 1.4 to 3
+ 3% addition error for C.F. from 3 to 4
±(1.5%+5 dgt) 50 Hz ~ 500 Hz *
www.megger.com
13
DC / AC Current
RangeDC AccuracyAC AccuracyVoltage Burden
2.000 A
10.00 A **
Overload Protection :A input : 10 A (500 V) fast blow fuse
* AC Conversion Type : Conversion type and additional specification are same as DC/
AC Voltage.
** Ampere Testing Duty Ratio Table
AmpereTesting TimeRest Time
10 A1 min10 min
9 A2 min10 min
8 A3 min10 min
7 A4 min10 min
6 A5 min10 min
5 AContinually N/A
±(1.0% + 3 dgt)
±(1.5% + 5 dgt)
50 Hz ~ 500 Hz *
2 V max
Resistance
RangeAccuracyVoltage Burden
200.0 ~ 200.0 KΩ
2.000 MΩ ** ± (1.0 % + 3 dgt)
20.00 MΩ *± (1.5 % + 3 dgt)
Open circuit Voltage : -1.3 V approx.
* <100 dgt rolling.
* * The minimum LCD reading is 1400 count in Auto Ranging Mode.
± (0.7 % + 3 dgt)
**
2 V max
14
www.megger.com
Diode Check and Continuity
RangeResolutionAccuracy
10 mV±(1.5% + 5 dgt)*
* For 0.4 V ~ 0.8 V
Max.Test Current : 1.5 mA
Max. Open Circuit Voltage : 2 V
Overload Protection : 600 V rms.
Frequency
RangeSensitivityAccuracy
2000 Hz ~200.0 KHz> 1.5 Vac rms,
< 5 Vac rms
2.000 MHz
~ 20.00 MHz
Overload Protection : 600 V rms.
Minimum pulse width : > 25 ns
Duty cycle limits : > 30% and <70%
> 2 Vac rms,
< 5 Vac rms
Frequency :
0.01% ±1 digit
Capacitance
RangeAccuracyOverload Protection
2.000 nF ~ 200.0 μF
2.000 mF *
* < 10 dgt of reading rolling.
±(1.9% + 8 dgt)600 V rms
www.megger.com
15
CATIV Measurement category IV: Equipment connected between the origin of
the low-voltage mains supply outside the building and the
consumer unit.
CATIII Measurement category III: Equipment connected between the consumer
unit and the electrical outlets.
CATII Measurement category II: Equipment connected between the electrical
outlets and the user’s equipment.
WEEE Directive
The crossed out wheeled bin symbol on the instrument and on the batteries is a
reminder not to dispose of them with general waste at the end of their life.
Megger is registered in the UK as a Producer of Electrical and Electronic equipment.
The registration no is; WEE/DJ2235XR.
Users of Megger products in the UK may dispose of them at the end of their useful
life by contacting B2B Compliance at www.b2bcompliance.org.uk or by telephone
on 01691 676124.
Users of Megger products in other parts of the EU should contact their local Megger
company or distributor.
Battery Disposal
Batteries in this product are classified as Portable Batteries under the Batteries
Directive. Please contact Megger Ltd for instructions on the safe disposal of these
batteries.
For disposal of batteries in other parts of the EU contact your local distributor.
Megger is registered in the UK as a producer of batteries.
The registration number is BPRN01235.
For Further information see www.megger.com
16
www.megger.com
Limited Warranty
This meter is warranted to the original purchaser against defects in material and
workmanship for 1 year from the date of purchase. During this warranty period,
manufacturer will, at its option, replace or repair the defective unit, subject to
verification of the defect or malfunction.
This warranty does not cover fuses, disposable batteries, or damage from abuse,
neglect, accident, unauthorised repair, alteration, contamination, or abnormal
conditions of operation or handling.
Any implied warranties arising out of the sale of this product, including but not
limited to implied warranties of merchantability and fitness for a particular purpose,
are limited to the above. The manufacturer shall not be liable for loss of use of the
instrument or other incidental or consequential damages, expenses, or economic loss,
or for any claim or claims for such damage, expense or economic loss. Some states
or countries laws vary, so the above limitations or exclusions may not apply to you.
Megger Limited
Archcliffe Road
Dover, Kent
CT17 9EN
Tel: +44 (0) 1304 502 101
Fax: +44 (0) 1304 207 342
www.megger.com
www.megger.com
17
18
www.megger.com
mM
AVO210
Digitale Multimeter
Bedienungsanleitung
G SICHERHEITSINFORMATIONEN
Verstehen und befolgen Sie die Gebrauchsanweisung sorgfältig.
G WARNHINWEIS
■ Verwenden Sie weder das Multimeter noch Messleitungen,
wenn sie schadhaft sind.
■ Verwenden Sie das Multimeter nicht, wenn das Prüfgerät nicht
ordnungsgemäß funktioniert.
■ Um die Gefahr von Bränden oder elektrischen Schlägen zu verringern,
darf dieses Gerät weder Regen noch Feuchtigkeit ausgesetzt werden.
■ Halten Sie bei Verwendung von Messleitungen oder Sonden die Finger
stets hinter dem Fingerschutz.
■ Entfernen Sie vor dem Öffnen der Batteriefachabdeckung oder des
Metergehäuses die Messleitungen vom Meter.
■ Messgerät nur für die in der Bedienungsanleitung beschriebenen Arbeiten verwenden, da ansonsten die Schutzfunktion des Geräts
beeinträchtigt werden kann.
■ Verwenden Sie für Ihre Messungen die Anschlüsse, Schalterstellung und den Bereich stets korrekt.
■ Versuchen Sie niemals eine Spannungsmessung, wenn die Messleitung am Anschluss A angeschlossen ist.
■ Überprüfen Sie die Funktionsfähigkeit des Messgeräts, indem Sie eine Ihnen bekannte Spannung messen. Bringen Sie das Messgerät im
Zweifelsfall zur Wartung.
■ Zwischen den Anschlüssen bzw. zwischen den Anschlüssen und der Erde darf nicht mehr als die Nennspannung gemäß Kennzeichnung am
Messgerät anliegen.
■ Versuchen Sie keine Strommessung, wenn die offene Spannung über dem Schutzwert der Sicherung liegt.
■ Die vermutete Leerlaufspannung kann mit der Spannungsfunktion geprüft werden.
■ Ersetzen Sie durchgebrannte Sicherungen nur durch Sicherungen mit den in diesem Handbuch angegebenen Werten.
20
www.megger.com
■ Seien Sie vorsichtig bei Spannungen über 30 VAC rms, 42 VAC Spitze oder 60 VDC. Bei diesen Spannungen besteht die
Gefahr eines Stromschlags.
■ Um Falschmessungen zu vermeiden, die zu elektrischem Schlag oder Verletzungen führen können, wechseln Sie die Batterie, sobald die
Batterieanzeige
■ Trennen Sie das Gerät vom Netz und entladen Sie alle Hochspannungskondensatoren, bevor Sie Widerstand, Durchgang,
Dioden oder Kapazität messen.
■ Verwenden Sie das Messgerät nicht bei explosivem Gas oder Dampf
angezeigt wird
.
G Vorsicht
■ Trennen Sie die Messleitungen von den Prüfpunkten, bevor Sie die Stellung des Drehschalters verändern.
■ Schließen Sie niemals eine Spannungsquelle an, wenn sich der
Drehschalter in der Stellung
■ Setzen Sie das Messgerät niemals extremen Temperaturen oder hoher Luftfeuchtigkeit aus.
■ Stellen Sie das Messgerät niemals auf die Funktion Spannung in einer Stromversorgung zu messen, welche zu Schäden am
Messgerät und dem zu prüfenden Gerät führen könnten.
W , ,
d , a
,
A, b befindet.
A fum die
da
Symbole, wie sie auf dem Messgerät und in der
Bedienungsanleitung erscheinen
F Gefahr eines elektischen Schlags
G Siehe beiliegende Hinweise
d DC Messung
t Geräte durch doppelte Isolation geschützt
Batterie
a AC Messung
c Das Gerät entspricht den aktuellen EU-Richtlinien
www.megger.com
21
Wartung
Versuchen Sie nicht, dieses Messgerät zu reparieren. Es enthält keine Bauteile, die
vom Anwender gewartet werden können. Reparatur oder Wartung dürfen nur von
qualifizierten Mitarbeitern durchgeführt werden.
Reinigung
Das Gehäuse von Zeit zu Zeit mit einem trockenen Tuch und Reinigungsmittel
abwischen. Keine scheuernden Mittel oder Lösungsmittel verwenden.
22
www.megger.com
AC/DC-Spannung und Frequenz
Durchgang und Dioden
www.megger.com
23
Widerstand und Kapazität
Hinweis – Um die Messgenauigkeit bei Kondensatoren mit kleinen Werten zu
verbessern, erfassen Sie den Messwert bei offenen Messleitungen und subtrahieren
Sie dann die Restkapazität von Messgerät und Messleitungen.
C
UNBEKANNT
= C
MESSUNG
– C
REST
MIN/MAX-Aufzeichnung
24
www.megger.com
Strom (AC/DC):
Daueranzeige
www.megger.com
25
Manueller und automatischer Messbereich
Autom. Abschaltung (Stromsparmodus)
26
www.megger.com
Autom. Ausschalten deaktivieren
www.megger.com
27
Berührungslose Warnung vor Spannung
(Spannungsfühler)
1. Spannungsfühler-Schalter wird bei jede Funktion oder im ausgeschalteten
Zustand aktiviert.
2. Zur Spannungsfühler-Messung werden keine Messleitungen verwendet.
3. Drücken Sie den Spannungsfühler-Taster. Die Anzeige erlischt, ein Ton
erklingt und die rote LED leuchtet, um die Betriebsbereitschaft des
Geräts anzuzeigen. Der Spannungsfühler-Taster muss gedrückt gehalten
werden, um ohne Messleitungen Spannung zu messen.
4. Wenn eine Spannung von 50 V bis 600 V wird (50 Hz bis 500 Hz) im
oberen Bereich des Messgeräts erkannt wird, erklingt ein Dauerton und
die rote LED im oberen Bereich des Messgeräts leuchtet.
28
www.megger.com