Megger AVO210 Specifications

AVO210
Digital multimeter
AVO210
Digital multimeter
 2000 count digital display
 1000 V DC / 750 V AC ranges
 10 A AC / DC ranges
 Resistance, frequency and capacitance ranges
 Non-contact voltage sense feature
 CAT III 600 V
DESCRIPTION
The AVO210 is a general purpose multimeter suitable for electricians, heating engineers and alarm technicians. The additional features make the instrument useful for a wide range of applications.
The instrument offers AC and DC voltage and current measurements as well as resistance, frequency and capacitance ranges.
The testleads included with the AVO210 have GS38 compliant shrouded tips.
Auto-ranging
When first selected, all functions are auto-ranging. A range button on the AVO210 allows multiple manual range selection on each function.
Minimum / Maximum measurements
The instrument has a MIN MAX function that allows the user to switch between minimum and maximum measurements. The display does not have to be continually monitored to capture a momentary increase or fall in readings.
Data Hold
This function allows a displayed result to be frozen which avoids having to remember a measurement value. The hold function can be nested within the MIN MAX feature which stops the AVO210 continuously updating the minimum and maximum values.
Voltage measurements
Both AC and DC voltage measurements up to 750 V and 1000 V respectively are possible with the AVO210.
Current measurements
A separate fused terminal is provided for current measurements up to 10 A for both AC and DC.
Continuity / diode testing
The continuity function features a buzzer and provides the user both optical and audible indication of identifying and confirming continuity between two points. This function also allows forward and reverse bias testing of diode and semiconductor junctions.
Voltsense function
The AVO210 has a built in non-contact voltage sensor fitted in the top of the instrument that is activated by the Voltsense button.
Resistance, capacitance and frequency
Resistance can be measured directly on the ohms range from 0 to 20 MΩ with capacitance measurements from 0 to 2.000 mF. In addition, frequency measurements from 0 to 20 MHz are possible.
SPECIFICATIONS
Display 2000 counts
Polarity Automatic, positive implied, negative indicated
Over-range indication “OL” or “-OL”
Battery indicator “<” is displayed when the batteries voltage drops below operating voltage
Auto power down Approx 10 minutes
Operating ambient Non-condensing ≤10 °C, 11 °C ~ 30 °C (≤80% R.H) 31°C ~ 40 °C (≤75% R.H), 41 °C ~
50 °C (≤45% R.H)
Storage temperature range and humidity
Temperature co-efficient 0.15 x (Spec. Accy.) / °C, < 18 °C or > 28 °C
Sample rate Samples 2 times per second nominal
Maximum altitude 6561.7 ft (2000 m)
Safety
Complies with EN61010-1, UL61010-1, IEC 61010-1, V/Ω CAT III 600 V, CATII 1000 V A CAT III 500 V
Pollution degree 2
Power supply 1.5 V x 2 LR03 or AAA size
Battery life Alkaline 250 hours
Dimensions 74 mm x 156 mm x 44 mm
Weight 320 g
ELECTRICAL
Accuracy is ± (% reading + number of digits) at 23°C ± 5°C < 80%RH.
AC/DC volts
Range AC Accuracy
200.0 mV Unspecified
2.000 V * ±(1.5%+5 dgts) 50 Hz ~ 300 Hz
20.00 V ~ 200.0 V * ±(1.5%+5 dgts) 50 Hz ~ 500 Hz * 750 V AC / 1000 V DC ±(1.5%+5 dgts) 50 Hz ~ 500 Hz *
-20 °C to 60 °C, 0 to 80% R.H. when battery removed from meter
AVO210
Digital multimeter
DC Accuracy : ±(0.5% + 2 dgts) Over voltage protection : 1000 V DC or 750 V AC rms. Input Impedance : 10 MΩ // less than 100 pF. * CMRR / NMRR : (Common Mode Rejection Ratio) (Normal Mode Rejection Ratio)
VAC : CMRR > 60 dB at DC, 50 Hz / 60 Hz VDC : CMRR > 100 dB at DC, 50 Hz / 60 Hz NMRR > 50 dB at DC, 50 Hz / 60 Hz
AC conversion type
Average sensing rms indication.
AC conversions are ac-coupled, true rms responding, calibrated to the sine wave input. * The minimum LCD reading is 1400 count in Auto Ranging Mode. Crest Factor : C.F. = Peak / Rms + 1.5% addition error for C.F. from 1.4 to 3 + 3% addition error for C.F. from 3 to 4
DC/AC current
Range DC Accuracy AC Accuracy Voltage Burden
2.000 A ±(1.0% + 3 dgts) ±(1.5% + 5 dgts) 2 V max
10.00 A ** ±(1.0% + 3 dgts) 50 Hz ~ 500 Hz* 2 V max
2
Overload protection
A input: 10 A (500 V) fast blow fuse * AC Conversion Type : Conversion type and additional specification are same as DC/AC voltage.
** Ampere Testing Duty Ratio Table Ampere Testing Time Rest Time
10 A 1 min 10 min 9 A 2 min 10 min 8 A 3 min 10 min 7 A 4 min 10 min 6 A 5 min 10 min 5 A Continually N/A
Resistance measurements
Range Accuracy Voltage Burden
200.0 ~ 200.0 KΩ ** ±(0.7% + 3 dgts) 2 V max
2.000 MΩ ** ±(1.0% + 3 dgts) 2 V max
20.00 MΩ * ±(1.5% + 3 dgts) 2 V max * <100 dgt rolling.
* * The minimum LCD reading is 1400 count in Auto Ranging Mode.
Open circuit voltage -1.3 V approx.
Diode check
Range Resolution Accuracy
(Diode symbol) 10 mV ±(1.5% + 5 dgts)*
Frequency
Range Sensitivity Accuracy
2000 Hz ~ 200.0 KHz >1.5 Vac rms, <5 Vac rms Frequency: 0.01%±1 digit
2.000 MHz ~ 20.00 MHz >2 Vac rms, <5 Vac rms Frequency: 0.01%±1 digit
Capacitance
Range Accuracy Overload Protection
2.000 nF ~ 200.0 μF ±(1.9% + 8 dgts) 600 V rms
2.000 mF * ±(1.9% + 8 dgts) 600 V rms
Input impedance 10 MΩ // less than 100 pF.
AVO210
Digital multimeter
ORDERING INFORMATION
Description Order Code
AVO210 digital multimeter CAT III 600 V 1000-969
Included accessories
Test leads
Optional accessories
Pouch 2007-366
UK
Archcliffe Road Dover CT17 9EN England T +44 (0) 1304 502101 F +44 (0) 1304 207342 UKsales@megger.com
UNITED STATES
4271 Bronze Way Dallas TX 75237-1019 USA T 800 723 2861 (USA only) T +1 214 333 3201 F +1 214 331 7399 USsales@megger.com
OTHER TECHNICAL SALES OFFICES
Valley Forge USA, College Station USA, Sydney AUSTRALIA, Danderyd SWEDEN, Ontario CANADA, Trappes FRANCE, Oberursel GERMANY, Aargau SWITZERLAND, Kingdom of BAHRAIN, Mumbai INDIA, Johannesburg SOUTH AFRICA, Chonburi THAILAND
3
CERTIFICATION ISO
Registered to ISO 9001:2008 Cert. no. Q 09290
Registered to ISO 14001-2004 Cert. no. EMS 61597
AVO210_DS_en_V01
www.megger.com
Megger is a registered trademark
Loading...