The information presented in this manual is believed to be adequate for the
intended use of the product. If the product or its individual instruments are
used for purposes other than those specified herein, confirmation of their
validity and suitability must be obtained from Megger. Refer to the warranty
information below. Specifications are subject to change without notice.
WARRANTY
Products supplied by Megger are warranted against defects in material and
workmanship for a period of one year following shipment. Our liability is
specifically limited to replacing or repairing, at our option, defective
equipment. Equipment returned to the factory for repair must be shipped
prepaid and insured. Contact your MEGGER representative for instructions
and a return authorization (RA) number. Please indicate all pertinent
information, including problem symptoms. Also specify the serial number and
the catalog number of the unit. This warranty does not include batteries, lamps
or other expendable items, where the original manufacturer’s warranty shall
apply. We make no other warranty. The warranty is void in the event of abuse
(failure to follow recommended operating procedures) or failure by the
customer to perform specific maintenance as indicated in this manual.
M
Valley Forge Corporate Center
2621 Van Buren Ave
Norristown, PA 19403-2329
610-676-8500 (Telephone)
610-676-8610 (Fax)
www.megger.com
Table of Contents
1.0 10 Ampere Systems .....................................................................................1
Section A INTRODUCTION............................................................1
Section B SAFETY PRECAUTIONS .............................................3
Section C RECEIVING INSTRUCTIONS ..................................5
Section D SPECIFICATIONS...........................................................7
Section E CIRCUIT DESCRIPTION...........................................15
Section F CONTROLS AND CONNECTIONS.....................19
Section G OPERATING PROCEDURE....................................23
Section H APPLICATIONS.............................................................27
Section I MAINTENANCE.............................................................31
Section J FIELD REPLACEABLE PARTS LIST.....................33
Section K WARRANTY & REPAIR.............................................35
Figure 1: Catalog No. 247000 DLRO instrument without charger
AVTM24-1J Rev A April 2006
iii
M
AVTM24-1J Rev A April 2006
iv
1.0
10 Ampere Systems
Section A
INTRODUCTION
Part 1.0 of this manual applies to a series of instruments;
the principal models are Catalog Numbers 247000,
247001, 247002. Part 2.0 describes the 100 Ampere
systems including catalog numbers 247100, 247101 and
247120.
The Catalog No. 247000 Digital Low Resistance
Ohmmeter is a portable, rugged instrument that provides a
direct-reading digital display of resistance in five ranges. It
is powered by self-contained rechargeable batteries. A
battery charger, operating from J1SV 50/60 Hz, is
furnished in a separate package.
The Catalog No. 247001 is similar to Catalog No. 247000
except that it has a built-in, line-operated charger.
AVTM24-1J Rev A April 2006
1
M
The Catalog No. 247100 is a 100 Ampere Test System
composed of a Catalog No. 247000 Measuring Module,
and a Cat. No. 247120 100-amp power supply. The
Catalog No. 24701 is also a 100 ampere test system.
The ohmmeter has four terminals for a Kelvin connection
to the sample under test, and it is recommended that
Biddle test leads shown in Bulletin 24-1 be used with this
instrument.
Panel features include a two-position on-off switch that
allows continuous or momentary operation, and on the
rechargeable battery models a meter to indicate relative
state of charge of the internal batteries.
A prop is provided to keep the lid in a partly open position
so that it can be used as a sunshade when reading the
digital display in bright sunlight.
The Catalog No. 247002 is similar to Catalog No. 247001
except that it has an extra 600 µΩrange.
AVTM24-1J REV A April 2006
2
Section B
SAFETY PRECAUTIONS
SAFETY IS THE RESPONSIBILITY OF THE USER
La seguridad es la responsabilidad del operador
WARNING
F
F
Do NOT connect this instrument to energized
circuits.
The Digital Low Resistance Ohmmeters have been
designed and constructed to meet the requirements of
ANSI/ISDA S82.01-1994. There is no shock hazard from
the instrument itself and it may be used freely in wet or
outdoor environments. The line-operated battery charger
and the A.C. supply, however, should be used and
maintained by trained personnel who are familiar with the
usual precautions for handling line-operated equipment.
DO NOT USE IN EXPLOSIVE ATMOSPHERE (This
includes poorly ventilated battery rooms and
enclosures.)
AVTM24-1J REV A April 2006
3
M
M
AVTM24-1J REV A April 2006
4
Section C
RECEIVING INSTRUCTIONS
READ COMPLETELY BEFORE OPERATING
Your Digital Low Resistance Ohmmeter has been
thoroughly tested and inspected to rigid specifications
before being shipped and is ready for use. Check the
equipment received against the packing list. Notify
Megger, Norristown, PA 19403 of any shortage of
materials. The instrument should be examined for damage
received in transit. If any damage is found, file a claim with
the carrier at once and notify Megger or its nearest
representative giving a detailed description of the damages
observed.
Charge batteries overnight before initial use of instrument.
Refer to Section G for charging procedure. Check for
correct operation of the instrument by connecting it to a
resistor of known value. See Section G, Operating
Procedure.
AVTM24-1J REV A April 2006
5
M
Catalog No. 247000 with Charger Catalog No. 247001
Catalog No. 247101with Charger
Figure 2: Various DLRO Unit Combinations
AVTM24-1J REV A April 2006
6
Section D
SPECIFICATIONS
Ranges, Resolution, Accuracy
& Test Current
TEST
RANGE
0.000 -5.999 mO 10 1 µO ¼ % Rdg + 1 lsd* ½ % Rdg + 1lsd*
00.00 -59.99 mO 1 10 µO ¼ % Rdg + 1 lsd* ½ % Rdg + 1lsd*
000.0 -599.9 mO 0.1 100 µO ¼ % Rdg + 1 lsd* ½ % Rdg + 1lsd*
0.000 -5.999 O 0.01 1 µO ¼ % Rdg + 1 lsd* ½ % Rdg + 1lsd*
00.00 -59.99 O 0.001 10 µO ¼ % Rdg + 1 lsd* ½ % Rdg + 1lsd*
CURRENT
AMPERES
(±20%)
RESOLUTION
*(Least Significant Digit)
Zero Offset: Typically 0-1 count over 15-35°C.
Full accuracy can be realized by using
the Forward-Reverse switch to
average the readings. An internal zero
adjustment is also provided.
LIMIT OF ERROR
1 year
15-35°C
1 year
0-50°C
On-Off
Control:
Response
Time
AVTM24-1J REV A April 2006
Toggle switch energizes all circuits.
Has momentary and lock positions.
(after On-Off switch is closed):
2 seconds to final reading.
7
M
Interference
Effects:
Error in item under test caused by
line- frequency current of 0.1X test
current: ±l lsd
Error caused by 5-gauss line-
frequency
magnetic field: ±2 lsd
Error caused by connection to
ground or voltage to ground
(assuming battery operation): None
Input
Protection:
Effect of
Inductive Test
Item:
Temperature
Range:
One volt peak may be applied
between any two of the Rx terminals.
None except time constant delay. No
damage caused by inductive kick, but
see precautions in Section H.
Operating: 0-50°C
(line supply, 0-40°C)
Storage: -40°C to + 60°C
Display: Red LED, 4 digits 0.5” high, decimal
point, and negative (-) sign. Upper
right hand segments of digits flash on
overrange input
AVTM24-1J REV A April 2006
8
Specifications
Calibration
Adjustments:
Test Lead
Resistance
Internal zero and span adjustments
for all ranges.
Potential Leads (P1. P2):
No limitations.
Requirements:
Current Leads (C1, C2): each lead 20
milliohms nominal. Deviations affect
test current but do not affect accuracy
unless they are large compared to the
measuring range.
Shows state of charge of batteries.
Switch selects either battery for test.
Meter:
AVTM24-1J REV A April 2006
9
M
Battery
Charger:
Separate charger unit plugs into
receptacle on measuring unit. Full
charge in 14 hours. Charging may be
continued while instrument is
operating. Fused, with UL listed 3wire line cord and plug. Power
requirements, 0.22 amps at 115 volts,
50/60 Hz; or 0.165 amps at 230 volts,
50/60 Hz.
247000 11 5.01 8-7/8W x 8-1/8D x 8-1/8H
247000-47 - - 8-1/8W x 8-1/8D x 7-3/8H
247001 12.1 5.5 12W x 9 5/8D x 6 3/8H
247002 12.2 5.5 12W x 9 5/8D x 6 3/8H
AVTM24-1J REV A April 2006
Meets all applicable requirements
of ANSI/ISA S82.01-1994.
Highest voltage existing (except in
charger primary and line supply
primary) is 6 volts.
The measuring unit and charger
have rugged molded case with
handles and removable hinged
lids. The measuring unit lid can be
used as a sun shield. The charger
case has similar construction with
additional space for storing test
leads.
5/16” diameter studs with wing
nuts for spade or ring lugs.
Dimensions and Weights:
cm
22.4 x 20.6 x 20.6(Instrument)
22.4 x 20.6 x 18.8 (Charger)
30 x 24.5 x 16.2
30 x 24.5 x 16.2
11
M
CATALOG
247000 Instrument with Internal Rechargeable
247000-3 Same as Cat. No. 247000 except with
247000-47 Same as Cat. No. 247000 except with
247000-3-47 Same as Cat. No. 247000-47 except with
Catalog Numbers and Unit Combinations
for 10-Ampere Systems
No.
DESCRIPTION
Batteries and separate 115V 50/60 Hz
Charger Unit. Range: 0 to 60 Ohms in
5 ranges, 0 to 5.999 Milliohms lowest
range (Note 1).
100A adapter plug.
230V-50/60 Hz charger.
100A adapter plug.
247001 Same as Cat. No. 247000 except in single
case.
247002 Same as Cat. No. 247001 except with an
extra 600 µΩ range.
AVTM24-1J REV A April 2006
12
Specifications
Catalog Numbers and Unit Combinations
for 10-Ampere Systems
CATALOG
No.
DESCRIPTION
247100 See Part 2.0
247101 See Part 2.0
247120 See Part 2.0
NOTES:
Note 1: Leads not included. (See Bulletin 24-1 for
recommended test leads.)
See Page 36 for information concerning 100-Ampere
systems.
AVTM24-1J REV A April 2006
13