Megger 210170 Specifications

EXTENDED RANGE
Insulation Resistance Testers
Multiple test voltages
Guard terminal to eliminate surface leakage current
Voltage indicating range
Large easy-to-read scale
Insulation Resistance Testers
DESCRIPTION
These Major Megger®Extended Range Insulation Resistance Testers are the ideal instruments when high ranges are needed for measuring insulation resistance. Both provide excellent test voltage regulation and an external guard terminal to eliminate surface leakage current from the measurement. The instruments are all electronic; they generate a regulated dc high voltage and use low-zero-drift, high-accuracy circuits, with high current sensitivity. Designed for portability, each is enclosed in a molded, impact-resistance, flame-retardant case.
Both instruments are protected for connection to power distribution systems up to 300 V Line-Ground and 500 V Line-Line for Installation Category III. This relates to transient overvoltage likely to be found in fixed installation wiring.
Cat. No. 210170
This unit provides high-resistance readings up to 20,000 M. It features an analog meter movement calibrated directly in megohms, ohms and ac volts. Test voltages of 100, 250, 500 and 1000 volts dc are selected by a rotary switch. Insulation resistance measurements up to 20 G can be made on all voltage ranges.
The unit is also equipped with a 0- to 5000-ohm continuity range and a 0- to 600-volt ac or presence of dc voltage measuring range. Power is supplied by an ac generator, driven by a hand crank, through a gear train.
Cat. No. 210600
This digital, autoranging model provides insulation resistance readings up to 1999 M. It has a 0- to 600-volt ac measuring range. Test voltages of 100, 250, 500 and 1000 volts dc are selected by a rotary switch, and insulation resistance to 1999 Mautoranging can be made on the 500 and 1000 V ranges. It incorporates a 3-1/2 digit LCD calibrated directly in megohms, ohms and ac/dc volts.
APPLICATIONS
Major Megger testers measure the insulation resistance of wires, power cables, motors, generators, transformers, switchboards and electrical controls. Both models offer multiple test voltages for performing spot and step-voltage tests. Typical applications include:
Acceptance testing at time of installation to check conformance to specifications.
Routine preventive maintenance testing after installation.
Quality assurance testing by the manufacturer.
Diagnostic testing to isolate faulty components for repair.
Major Megger insulation testers can be used for detecting high- or low-resistance grounds, short circuits in apparatus, cables, wiring, etc. whether caused by moisture, oil, dirt, corrosion, damage to insulation or natural deterioration. They can also be used to determine the presence of moisture, solvents and semiconducting foreign materials in wires, cables and other conductors, and in built-up insulation systems such as those found in motor windings.
EXTENDED RANGE
Insulation Resistance Testers
Typically, these testers are used to take a series of measurements over a period of time which will show the gradual decline that takes place in the insulation during its operational life. Such monitoring enables the user or to anticipate future performance and to plan ahead for repairs. They are also used to show improvements in the insulation of motor, transformer and generator windings that result from drying-out procedures used after exposure to excessive humidity or water.
FEATURES AND BENEFITS
• Guard terminal to eliminate surface leakage current from the measurement
• High-resistance readings up to 20,000 M(analog model)
• Digital autoranging with readings up to 1999 M(digital model)
• Multiple test voltages
SPECIFICATIONS
Cat. No. 210170
Insulation Range
Accuracy: ±3% of scale length Test Voltage: ±5% Short-circuit Current: 220 mA nominal on all ranges Scale Length: 3.08 in. (78 mm)
Low Resistance Range
Accuracy: ±3% of scale length Test Voltage: 3 V ±5% Short-circuit Current: 30 mA ±10% Scale Length: 3.08 in. (78 mm)
Default Voltage Range
0 to 600 V ac/dc accuracy ±2.5% of scale length
Discharge
Capacitive circuits are automatically discharged when the “TEST” button is released following an insulation test.
Temperature Range
Operating: 14º F to 122º F (-10º C to +50º C) Storage: 4º F to 158º F (-15º C to +70º C)
Humidity
Operating: 70% RH at 68º F (20º C); 50% RH at 104º F (40º C) Storage: 95% RH at 95º F (35º C)
Cat. No. 210600
Insulation Range
Accuracy (73.4º F ±3.6º F):
<500 M: ±3% ±2 digits 500 to 1999 M: ±5% ±2 digits
Temperature Coefficient:
<500 M: <86º F, ±0.06% per ºF;
>86º F ±0.14% per ºF
500 to 1999 M: < 86º F ± 0.11% per ºF;
>86º F ±0.28% per ºF Test Voltage: ±5% at open circuit Short-circuit Current: 1.2 mA nominal
Hum Resistance
Within twice the normal accuracy if the noise current does not exceed 1 mA for the 1-kV range, or 0.5 mA for the other ranges, and is within 45 to 65 Hz.
Low Resistance Range
Accuracy: ±2% of reading ±1 digit Temperature Coefficient: ±0.06% per ºF Maximum Test Voltage: 4.3 V nominal Short-circuit Current: 2.75 mA nominal
Default Voltage Range
0 to 600 V ac/dc Accuracy: ±2% of reading ±1 digit at 45 to 65 Hz Temperature Coefficient: ±0.06% per ºF Input Impedance: 300 kto 1 Mdepending on input voltage
Display
3-1/2 digit LCD. Maximum reading 1999.
Battery Life
Typically 2100 x5 second operations
Discharge
Capacitive circuits are automatically discharged when the “TEST” button is released following an insulation test.
Temperature Range
Operating: 23º F to 122º F (-5º C to +50º C) Storage: 4º F to 158º F (-15º C to +70º C)
Humidity
Operating: 70% RH at 68º F (20º C); 60% RH at 95º F (35º C); 50% at 104º F (40º C) for measurements <200 M Storage: 95% RH at 95º F (35º C)
O/C Test Insulation
Voltage Resistance Resistance Power Voltage
Cat. No. (V dc) Range (MΩ) Range Source Range Dimensions Weight
210170 0 to 20,000 0 to 5000 Hand 8.25 H x 5 W x 5 D in.
(at 3 V dc) cranked 210 H x 125 W x 125 D mm 100 0 to 600 V ac 2.2 lb 250 0 to 199.9 at 0.01 to 1.999 k 6 AA or presence 7 H x 5 W x 5 D in. (.98 kg)
210600 500 all test voltages cells of dc 180 H x 125 W x 125 D mm
1000 2.00 to 19.99 k (1.5 V)
0 to 1999 at (at 4.3 V dc
500 and 1 kV Open Circuit)