MEDTRUM TECHNOLOGIES JN022, MD1026, MD1027 User Manual

B ureau Veritas
2. Select a temp basal type, temp rate or percent and duration, then
Tap Next to review the temporary basal rates set.
Note: If Percent is selected, you can set the temp basal rate, not exceeding the
Max Basal Rate, between 0 and 200% with an increment of 1%. If Rate (U/H) is selected, you can set the temp basal between 0 and the Max Basal Rate with an increment of 0.05 U/H.
Note: You can set the duration between 30 min and 24h with an increment of
30 min.
3. Make sure that the temp basal is correct in this Temp Basal Review, then
Slide to activate.
5.5.2 Cancel a Temp Basal
1. Go to the Cancel Temp Basal screen. Select Cancel Temp Basal.
135
B ureau Veritas
Main MenuBasalCancel Temp Basal
2. Slide to stop temp basal delivery, or tap to continue delivering.
Note: If you suspend insulin delivery while a temp basal rate is active, the temp
basal rate will be canceled.
5.6 Preset Temp Basal
With the preset temp basal feature, you can program temp basal rates for recurring short-term situations. You can set up to seven preset temp basal rates: Heavy Ex, Medium Ex, Light Ex, Sick, Temp 1, and Temp 2 and Temp 3.
5.6.1 Preset Temp Basal Setup
1. Go to the Preset Temp Setup screen.
136
B ureau Veritas
Main MenuSettingsInsulin PumpBasal SetupPreset Temp Setup
2. Select a preset temp basal you want to edit. Choose the temp basal type
(rate or percent).
3. Set the duration and rate/percentage of the preset temp basal. Tap Save
to save settings.
5.6.2 Activate a Preset Temp Basal
You must set up a preset temp basal before you can activate it.
1. Go to the Preset Temp Basal screen.
Main MenuBasalPreset Temp Basal
The programmed preset temp basal types are displayed on this screen. If you have not set up any preset temp basal rate, this screen shows No Presets.
2. Select the preset temp basal you want to activate.
137
B ureau Veritas
3. Confirm your preset temp basal settings.
4. Slide to activate.
5.7 Reminder
5.7.1 Bolus Reminder
When you fail to deliver a Bolus at between time point A and time point B, you will receive a Reminder at time point B.
1. Time Point A
1
You can add, delete, or review Reminders when the Bolus Reminder option is turned on. Go to the Bolus Reminder screen.
Main MenuSettingsReminders Bolus Reminder
2
2. Time Point B
Add Reminder Tap + Add time segment to add one Reminder, setting the start and end time.
138
B ureau Veritas
Note:
(1) The end time should be at least 30 min later than the start time. You can program up to four bolus Reminders.
(2) The Reminders will be saved automatically.
Delete Reminder Slide from right to left on one segment, tap Delete to delete this segment.
5.7.2 BG Reminder
After you deliver a bolus, you may want to check your BG. The BG Reminder is an optional feature that reminds you to check your BG after a bolus.
Go to the BG Reminder Setup screen.
Main Menu SettingsRemindersBG Reminder
139
B ureau Veritas
If you have BG Reminder turned on, the BG REMINDER DURATION screen appears when you set Bolus.
It allows you to set the time before you are reminded to check your blood glucose after a bolus.
The time ranges from 00:30 to 05:00 with an increment of 30 minutes. The default time is 00:30.
You can also turn off the BG reminder after each bolus.
You can accept or modify the time before you are reminded.
5.8 Pump History
5.8.1 Pump History
The Pump History displays the delivery history (bolus, basal and total daily
140
B ureau Veritas
delivery history) and alert history (pump alerts and alarms). Go to the Pump History screen.
Main MenuHistoryPump History
5.8.1.1 Delivery History
You can select one day to review its delivery graph. It displays the summary of basal, bolus and total delivery for one day.
1. Tap the date to switch
between records of different dates
2. Tap the info icon to
review legend meanings.
3. Tap the “Basal, Bolus,
Total “summary chart at the bottom of Delivery History screen to see details.
(1) The legend
141
B ureau Veritas
Legend Abbreviation
Basal Temp Basal
(2) Basal History
This screen displays most detailed Basal information.
Normal Bolus Extended Bolus
Auto Suspend
Manual Suspend
Stop
New Patch
Significance Basal Rate infusion curve
Temp Basal Rate infusion curve
Normal Bolus delivery icon Extended Bolus delivery icon
This tag appears when any of the following alarms occurs: AUTO OFF, PREDICTIVE LOW SUSPEND, LOW SUSPEND, EXCEEDS MAX TDD, EXCEEDS MAX 1HR DELIVERY ALARM. Manually suspend all insulin delivery
Including deactivate patch, discard patch, and alarms: OCCLUSION DETECTED, PATCH EXPIRED, PATCH ERROR, PATCH BATT DEPLETED, PUMP BASE ERROR, EMPTY RESERVOIR.
When you activate a new patch, this icon appears.
142
B ureau Veritas
(3) Bolus history
The summary information includes:
The start time of this bolus; Status of bolus: completed, canceled, delivering; Bolus type; Amount of bolus delivered |Amount of bolus programmed.
Bolus Type:
N: Normal Bolus E: Extended Bolus C: Combo Bolus Normal: Normal Bolus by Manual Bolus Extended: Extended Bolus by Manual Bolus
143
B ureau Veritas
Combo: Combo Bolus by Manual Bolus Calc-N: Normal Bolus by Bolus Calculator Calc-E: Extended Bolus by Bolus Calculator Calc-C: Combo Bolus by Bolus Calculator
Tap record line to view more detailed information. See Chapter “Advanced Pump Features” for more information.
(4) Daily Totals
This screen displays most detailed Daily Totals information.
N represents Normal Bolus. E represents Extended Bolus. C represents Combo Bolus. FoodBolus 7.30U #1 means that there is one food bolus doses in the
selected day with a total amount of 7.30U.
CorrBolus 2.10U #1 means that there is one correction bolus doses in the
selected day with a total amount of 2.10U.
Food+Corr 0.00U #0 means that there is no bolus dose that both covers
carbs and corrects glucose in the selected day.
ManualBo 9.70U #2 means that there are two manual bolus doses in the
selected day with a total amount of 9.70U.
144
B ureau Veritas
5.8.1.2 Alert History
Go to the pump Alert History screen.
Main MenuHistoryPump HistoryAlert History
Tap the date to switch between records of different dates. Tap each Alert/Alarm to view alert detail information. Tap to return to the previous menu.
See Section “Alert Icons” in Chapter How to us the PDM for more information about how to address alarms and alerts and the meanings of different alarm/alert icons.
5.8.2 Pump Summary History
5.8.2.1 Summary History: Insulin History
This screen displays the insulin delivery summary history. Go to the Insulin History screen.
Main MenuHistorySummary History Insulin History
145
B ureau Veritas
Daily Insulin: Total daily dose of insulin delivered per day. Daily Basal: Average daily dose and percentage of insulin delivered as Basal. Daily Bolus: Average daily dose and percentage of insulin delivered as Bolus. Daily Carbs: Average daily amount of carbs.
5.8.2.2 Summary History: Bolus History
This screen displays the Bolus summary history.
Main MenuHistorySummary History Bolus History
Carbs Bolus Only: Average daily dose and the number of times of Food Bolus only on the selected days.
BG Correction Only: Average daily dose and the number of times of BG correction Bolus only on the selected days.
5.9 Troubleshooting Pump issues
Can I take a sauna with a Patch Pump on?
No. Firstly, the operating temperature range for the Patch Pumps is +5°C ~ +40°C. Secondly, if you take a sauna, insulin will be absorbed faster into your body, and
your blood glucose can fluctuate.
Can I dive with a Patch Pump on?
No.
146
B ureau Veritas
Your Patch Pump is waterproof to a depth of 2.5 meters (8 feet) for up to 60 minutes (IPX8).
It means the maximum pressure the device can tolerate equals the pressure in
2.5m deep in STILL water instead of flowing water. It is OK to take a shower or go swimming with the devices on, but if you go diving,
the water pressure may be too high for the devices.
I didn’t see an alert message, but it appeared in History.
If one of the following alerts happened, the PDM would beep/vibrate and display a message first, and if you missed that alert, later when you checked the PDM, the condition that triggered the alert had changed (for example, your glucose level returned to the target range), then you wouldn't see any message on the screen, you would only find it in History.
Alert Alert change EXCEEDS MAX TDD After insulin delivery automatically starts
again, alert is switched to BASAL RESUMED.
EXCEEDS MAX 1HR DELIVERY After insulin delivery automatically starts
again, alert is switched to BASAL RESUMED.
If one of the following alerts happened, the PDM would beep/vibrate and display a message first, and if you missed that alert, later when you checked the PDM, the alert may have escalated to another alert/alarm, and you will ONLY see the message of the escalated alert/alarm. The first alert will appear in History.
Alert Alert escalation LOW RESERVOIR EMPTY RESERVOIR PATCH EXP ADVISORY PATCH EXP IN 1 HOUR, then PATCH EXPIRED
AUTO OFF ALERT AUTO OFF
Lights on the Patch Pump
Once you connect the pump base with a new Reservoir Patch, you will see the indicator light flashing in the order of blue, green, yellow, and red. When you are activating the new patch, you will see the green light flashing until the basal pattern is activated.
A yellow (orange) light indicates an alert, while a red light indicates an alarm.
147
B ureau Veritas
If the PDM is away from the Patch Pump, how will the basal rate be delivered?
The selected basal pattern is stored in the pump base, which means that even if the PDM is away, the basal pattern will continue as planned.
Can I fill the patch with insulin when the patch is on body?
NO. NEVER DO THAT. Insulin can go directly into your body, which is very dangerous.
No magnetic objects around when activating (priming)
When you are filling the Reservoir Patch, make sure that it is at least 30 cm (12 inches) from any magnetic objects, such as magnets, mobile phones, tablets, other Reservoir Patches, TVs, refrigerators, and sound options. The Patch Pump will detect the volume of insulin in the reservoir once it is filled, and if the Patch Pump is in a magnetic field, the volume detected can be inaccurate.
148
B ureau Veritas
149
B ureau Veritas
150
B ureau Veritas
6 How to use CGM system (Optional)
6.1 Glucose Alerts
Set your Low and High glucose alerts before using the sensor. See Section
“Glucose Alerts” in Chapter “How to use (P)LGS” for more information.
6.2 Change Sensor
Your Sensor gives glucose readings for up to fourteen days. When a Sensor expires or fails, your Sensor session ends automatically, and PDM displays no more glucose readings. You must remove the Sensor and disconnect the Transmitter.
6.2.1 Disconnect Sensor from Your PDM
Go to Disconnect Sensor screen.
Main MenuSensorDisconnect Sensor
Note: The Disconnect Sensor option is only available when a Sensor is currently
connected to the PDM.
151
B ureau Veritas
6.2.2 Remove the Current Sensor and Disconnect the
Transmitter
1. Gently peel the adhesive pad off your skin in one continuous movement
to remove the Sensor and Transmitter.
2. Pinch the ribbed release tabs on the sides of the Sensor support mount,
and gently pull the Transmitter away from the Sensor support mount.
3. Discard the Sensor support mount and reuse the Transmitter.
Note: Do Not discard your Transmitter. It is reusable and rechargeable. Note: Make sure that you completely disconnect the Transmitter from the
Sensor when you do. Do NOT store the Transmitter connecting a Sensor or a USB charging cable on which may kill the Transmitter battery.
6.2.3 Charge the Transmitter
The Transmitter is charged via a USB charging cable which is plugged into a USB
2.0/3.0 port or a power adapter with a rated voltage of DC 5V and a rated current higher than DC 1000mA. The device with the USB port and the power adapter must comply with EN 60950-1 or EN 60601-1.
152
B ureau Veritas
The battery must be fully charged the first time you use the Transmitter, which may take up to 2 hours. It is recommended to recharge the Transmitter after each Sensor session. If a Transmitter is stored for two months, you must fully charge the Transmitter battery to ensure it works properly.
The indicator light will flash when the Transmitter is being charged, and go off when the Transmitter is fully charged.
Note: We recommend that your Transmitter is only charged by an intended and
qualified operator.
6.2.4 Add the Transmitter SN
Any time when you switch to a new Transmitter and/or a PDM you must add the Transmitter SN.
Main MenuSensorTransmitter SN
1. Tap Connect Sensor if you have set the Transmitter SN.
Note: Don’t forget to update the SN if you change to a new Transmitter. Note: You can only change the Transmitter SN when there is no Sensor
connected. You can find the Transmitter SN on the product box or on the back of the
Transmitter.
153
B ureau Veritas
2. You can either enter SN manually or search for the SN if it is the first time you enter the SN.
3. You can only enter SN manually if you want to update the SN.
154
B ureau Veritas
Enter SN manually
Tap --------- or the existing Transmitter SN, you will see the following screen. Then enter the SN to your PDM and press Done.
Search for the SN
If you select Search, make sure that your Transmitter is connected to a new Sensor and move the PDM closer to your CGM before searching. See Section
“Insert a New Sensor” for more information.
If you tap Search in Sensor, you will see the following message when you search for the SN.
155
B ureau Veritas
If your PDM finds one Transmitter, the Transmitter SN appears on the screen. Confirm it once it matches the SN printed on your Transmitter. If it is correct, tap OK.
If your PDM finds multiple Transmitters, tap OK to go back to Sensor Menu, then select “---------” to enter the SN manually.
156
B ureau Veritas
If your PDM does not find a Transmitter, make sure that your Transmitter is connected to a new Sensor, move the PDM closer to your CGM, and enter the SN manually.
6.2.5 Insert a New Sensor
6.2.5.1 Select an Insertion Site
When choosing the location for the Sensor, consider the following:
That you can comfortably reach the Sensor. That you apply the Sensor to a flat area of skin with adequate
subcutaneous fat.
That the area stays flat during normal daily activities without bending or
creasing.
When choosing the location for the Sensor, avoid the following:
Areas that are constrained by clothing, such as the belt line or waist. Curved or rigid areas due to muscle or bone. Areas that involve rigorous movement during exercise. Areas of skin with scars, tattoos, or irritation. 5.0 cm (2 inches) around the navel. Areas with excess hair. Within 7.5 cm (3 inches) of an insulin pump infusion site or manual
injection site.
Shown here are the best body areas (shaded) for Sensor insertion.
157
B ureau Veritas
Front Back Front Back
If you choose an insertion site on your abdomen (buttock for children), apply the Sensor horizontally. If you choose an insertion site on your upper arm, apply
the Sensor vertically. Have a rotation schedule for choosing a new site. Using the same site too often
might not allow the skin to heal and can possibly cause scarring or skin irritation.
6.2.5.2 Prepare the Insertion Site
1. Wash your hands thoroughly with soap and water and wait for them dry
up.
2. Wipe the selected insertion area with rubbing alcohol and wait for the area
to dry up. This may help prevent infection. Do NOT insert the Sensor until the cleaned area is dry. This will make the Sensor adhesive stay on the skin more firmly.
Warning: If the Sensor dislodges because the Sensor support adhesive fails to
adhere to the skin, you may get false or no readings. Improper site selection and improper site preparation may result in poor adhesion.
6.2.5.3 Unpack the Glucose Sensor
Open the Sensor package by peeling off the paper on the back of the package. Pay attention to the following:
Warning: Do NOT use a Sensor if its sterile package has been damaged or
opened, or the Sensor has expired, or the Sensor is damaged in any way.
Note: Wash your hands with soap and water and let them dry before opening
the Sensor package and handling the Sensor. After opening the package, avoid touching any Sensor surface that will be in contact with the body, i.e., adhesive
158
B ureau Veritas
surface. You may contaminate the insertion site and suffer an infection if you have unclean hands while inserting the Sensor.
6.2.5.4 Remove the Protective Liner from the Sensor Support Mount
Bend the two-piece protective liner slightly on the edge so you can see the seam between the two pieces. Hold the inserter part of the Sensor, and try not to touch the adhesive surface. Remove the liners from the Sensor support mount one after another.
6.2.5.5 Locate the Sensor Support Mount
If you are inserting the Sensor on your abdomen or lower back, place the Sensor horizontally on your skin.
If you are inserting the Sensor on your upper arm or thigh, place the Sensor vertically on your skin.
Move your fingers around the adhesive pad to secure it to your skin.
6.2.5.6 Remove the Safety Lock
Hold the Glucose Sensor with one hand. Firmly squeeze the two release tabs of the safety lock with your thumb and index finger of the other hand, as you lift the safety lock away from the inserter. Keep the safety lock, you will need it later.
159
B ureau Veritas
6.2.5.7 Insert the Sensor
Hold the inserter as shown below and press the two buttons at the same time. You might feel a slight pinch as the Sensor is placed just under your skin.
6.2.5.8 Remove the Inserter
Pinch and hold the ribbed release tabs on the sides of the Sensor support mount with one hand, twist the inserter about 40° in the direction (anticlockwise) shown with the other hand, until the orange triangle marked on the inserter lines up with the orange line on the Sensor support mount, and then lift the inserter vertically away from the mount. Only the Sensor support mount will be left on your body.
160
B ureau Veritas
6.2.5.9 Check the Sensor Support Mount
Confirm that the Sensor support mount remains tightly adhered to your skin by sliding your finger along the edges of the adhesive pad and examine for any gaps in adhesion.
Warning: If bleeding occurs at the insertion site, do not attach the Transmitter
to the Sensor. Apply steady pressure using a sterile gauze or clean cloth for up to 3 minutes. If bleeding stops, attach the Transmitter to the Sensor. If bleeding continues, remove the Sensor, treat the site as necessary, and insert a new Sensor at a different site.
Warning: Check the insertion site frequently for infection or inflammation
redness, swelling or pain. Remove the Sensor and seek professional medical help if one of these conditions occurs.
6.2.5.10 Discard the Sensor Inserter Safely
Attach the safety lock on the inserter to cover its opening and conceal the needle inside. Follow local waste disposal regulations when discarding the inserter. We recommend discarding the Sensor inserter into a sharps container or a puncture-proof container with a tight lid.
6.2.6 Attach Your Transmitter
Note: If you are changing Sensor, make sure that your Transmitter was
disconnected from the old Sensor at least one minute before being connected to the new Sensor.
Before attaching the Transmitter to the Sensor, you must have the Transmitter battery fully charged and the PDM set up.
Snap the Transmitter into the Sensor support mount until the two flexible arms fit into the notches on the Transmitter. The indicator light will flash green after
161
B ureau Veritas
successful connection, three times after properly connected and another six times after successful system check.
Note: Make sure that you hear a click when you snap the Transmitter in place.
If it is not fully snapped in, electrical connection and waterproof can be compromised, which can lead to inaccurate Sensor glucose readings.
Tape the Sensor Support Mount (Optional) The Sensor support mount should stay on your skin using its own adhesive. But, if you find that the Sensor support mount is not adhering well during daily activities, you can use medical tape for extra support. Only tape over the white adhesive pad on all sides for even support. Do NOT tape over the Transmitter or any of the plastic parts on the Sensor support mount.
6.2.7 Connect Sensor to Your PDM
1. Go to Connect Sensor screen.
Main MenuSensorConnect Sensor
162
B ureau Veritas
Note: The Connect Sensor option is only available when no Sensor is currently
connected to the PDM.
2. Make sure that your Transmitter is connected to a Sensor and that your Transmitter SN is found or entered, and then continue by tapping Next.
3. If the sensor is calibration-free, enter the sensor code on the sensor label
which is unique for each sensor. Once the sensor code is entered successfully, the calibrations aren’t required.
Or, skip the sensor code input step and go to Connection screen. The sensor need to be calibrated twice (once every 12 hours) on the first day and then once every 24 hours since the second day.
4. When finished, the following screen appears.
163
B ureau Veritas
Note:
If you want to remove a Sensor before its expiration, disconnect it from your PDM first before you connect a new Sensor. When you connect a new Sensor directly, a “SENSOR RECONNECTED” message will appear on your PDM.
6.3 Calibrate Your Sensor
Each time the PDM prompts you with the message “METER BG NOW” or “SENSOR CAL REMINDER”, you must enter a BG measurement to calibrate your
Sensor.
Go to the Sensor Calibration screen.
Main MenuSensorSensor Calibration
164
B ureau Veritas
Note: If your sensor is not a calibration-free sensor, or you skip the sensor
code input step, you must calibrate your sensor at least twice (once every 12 hours) on the first day and then once every 24 hours since the second day. If
you have entered sensor code successfully, the system won’t require
calibration. But you can calibrate the sensor if you want.
Note: Calibration is unavailable under the following circumstances:
Sensor disconnected from the PDM Sensor warm-up Within 15 min after the alert SENSOR CAL ERROR Poor RF communication between the Transmitter and the PDM No Readings
6.3.1 Enter Your Meter BG
Here you can enter your present blood glucose measured by a finger prick blood glucose meter.
1. Go to the Enter BG screen.
Main MenuSensorSensor Calibration
165
B ureau Veritas
Note: Please enter the exact blood glucose value of a carefully performed
fingerstick displayed on your blood glucose meter within five minutes.
2. Tap Done to confirm your fingerstick, then tap Yes button to start
calibration.
3. When finished, the following screen appears.
166
B ureau Veritas
6.3.2 Set Calibration Repeat
Go to the Cal Repeat screen.
Main MenuSettingsCGM SystemCal Repeat
After you receive and clear a “METER BG NOW” alert, PDM will repeat the
alert until you enter a new blood glucose measurement. You can turn Cal Repeat on/off. If Cal Repeat is on, you can set the repeat time
of “METER BG NOW” alert from 5 min to 1h with an increment of 5 min.
6.3.3 Calibration Reminder
Calibration reminder enables you to get reminded a certain time before the due time of next calibration.
1. Go to the Cal Reminder screen.
167
B ureau Veritas
Main MenuSettingsReminders Cal Reminder
2. You can turn Cal Reminder on/off.
Note: If Cal Reminder is on, you can set the time between 5 min and 6 h with
an increment of 5 min.
6.4 CGM System settings
Go to the CGM System screen.
Main MenuSettingsCGM System
6.4.1 CGM Feature on/off
The CGM feature must be turned on to receive Sensor data.
1. Select CGM System in the Settings menu.
Main MenuSettingsCGM System
168
B ureau Veritas
2. You can turn on or off the CGM feature.
3. After you turn on the CGM System, the Transmitter SN menu appears.
6.4.2 Set the Transmitter SN
Tap Settings on the Main Menu to enter the Settings screen. Tap CGM System to enter the CGM settings screen. Turn the CGM System feature on.
Tap Transmitter SN to add this Transmitter to your PDM. You can use your PDM to search for your Transmitter (only for the first time), or you can enter the SN printed on your Transmitter manually.
You can also enter your new Transmitter SN in CGM System menu. See “Add the
Transmitter SN” for more information.
169
B ureau Veritas
6.4.3 Graph Range
You can set the time range of sensor graph in horizontal screen as 3, 6, 12, 24 hours. The default range is 12 hours.
6.4.4 Cal Repeat
See Section Calibrate Your Sensorin this chapter for more information.
6.4.5 Alert Silence
Go to the Alert Silence screen.
Main MenuSettingsCGM SystemAlert Silence
Warning: Muting the alarms is not recommended when you are unable to
interact with your PDM (for instance, when you are asleep).
170
B ureau Veritas
Interacting with your PDM includes activities such as pressing the power button and checking the screen.
With the Alert Silence feature you can keep glucose alerts silent for a specified time of 30 minutes to 24 hours.
There are five Alert Silence options: Off— This means all glucose alerts are turned on: the PDM will beep or
vibrate if any Sensor alert occurs.
Low The PDM will not beep or vibrate if a low alert (LOW GLUCOSE,
RAPID FALL or LOW PREDICTED) occurs during the specified time.
High The PDM will not beep or vibrate if a high alert (HIGH GLUCOSE,
RAPID RISE or HIGH PREDICTED) occurs during the specified time.
High and Low The PDM will not beep or vibrate if a high/low alert
(HIGH/LOW GLUCOSE, RAPID RISE/FALL, HIGH/LOW PREDICTED) occurs during the specified time.
All — The PDM will not beep or vibrate if “LOST SENSOR”, “SENSOR CAL
REMINDER”, “METER BG NOW”, “SENSOR EXP IN 6 HOURS”, “SENSOR EXP IN 2 HOURS”, “SENSOR EXP IN 30 MINS”, “SENSOR EXPIRED”, or any of
the high/low alert occurs during the specified time.
See “Status Bar Icons” and Chapter “Safety System and Alarms/Alerts” for more information.
6.4.6 Sensor Expired
Go to the Sensor Expired screen.
Main MenuSettingsCGM SystemSensor Expired
171
B ureau Veritas
For MD1026, the expiration date is fixed as 14 days.
In both settings, “SENSOR EXP IN 6 HOURS” alert, “SENSOR EXP IN 2 HOURS” alert, “SENSOR EXP IN 30 MIN” and “SENSOR EXPIRED” alert will respectively
appear.
6.5 Sensor History
6.5.1 Sensor History
Your PDM stores detailed Sensor history to help you keep track of your glucose readings and Sensor conditions.
Go to the Sensor History screen.
Main MenuHistorySensor History
6.5.1.1 Data History
1. Select Data History in the Sensor History screen. The Data History screen shows all of the Sensor sessions that have recently
occurred. Each line shows the Sensor session start date and duration (day/hour/minute). For example, the record 28-08-2018 5/21/8 means the Sensor was started on 28-08-2018 and has been used for 5 days 21 hours and 8 minutes.
172
B ureau Veritas
2. Select a Sensor session and you will see the last day’s Sensor history data. The Y-axis of the Sensor graph is featured by four values: 5, 10, 15, 20 mmol/L
(90, 180, 270, 360 mg/dL). The X-axis of the Sensor graph presents a period of 24 hours.
The Sensor graph can be switched to a landscape screen display. Long tap the Sensor graph for 1 second and the display will turn horizontal.
Note:
Tap the Sensor graph and move the cursor to spot the glucose values. Use
the left and right arrow button to do fine adjustment for choosing the time. The time interval between two values is 2 minutes.
The time a new Sensor is applied will be marked with a green square tag
“ ”. Readings during warm-up phase will not be displayed but marked as “warm-up”.
173
B ureau Veritas
Glucose value or special status will always be shown in the area below,
between the left and right arrow button. Special status includes: calibration error (ERR), no readings (???), warm-up phase (Warm-up), Sensor glucose
is above 22.2 mmol/L or 400mg/dL (HIGH) and Sensor glucose is below 2.2 mmol/L or 40mg/dL (LOW).
After the warm-up phase, the values before the first calibration are marked
as “BG“.
When the Sensor calibration expires, the reading values will be underlined. Calibration will be marked with a red dot “ ”. In the landscape screen display, tap the Home Key to return to the Home
Screen.
In the following situations, you cannot enter landscape screen by long-
pressing the graph
when no Sensor is connected. when the data is being recovered after reconnection.
3. Tap the date and you will see a list of dates within that session.
4. Select a date and you will see the 24-hour Sensor trend graph of that day.
6.5.1.2 Calibration History
Select Calibration History in the Sensor History screen. The Calibration History screen displays the calibration history.
174
B ureau Veritas
6.5.1.3 Alert History
Select Alert History in the Sensor History screen. The Alert History screen shows you all of the Sensor alerts that have recently occurred.
Select an alert record to view the details. Tap to return to the previous menu.
See Section “Alert Icons” and Chapter “Safety System and Alarms/Alerts” for more information about how to address the alarms and alerts.
6.5.2 Summary History: Sensor History
This screen displays the SG readings summary history. Go to Sensor History screen.
Main MenuHistorySummary HistorySensor History
175
B ureau Veritas
Average SG: Average SG readings of the selected days. Time in target range: The percentage of the duration in which SG reading is in the target range (3.9 - 10.0 mmol/L or 70 - 180 mg/dL). Time above range: The percentage of the duration in which SG reading is above the target range (10.0 mmol/L or 180 mg/dL). Time below range: The percentage of the duration in which SG reading is below the target range (3.9 mmol/L or 70 mg/dL).
6.6 Troubleshooting CGM issues
Can I take a sauna with the CGM System on?
No. Firstly, the operating temperature range for the Transmitter is +5°C ~ +40°C. Secondly, if you take a sauna, your blood glucose can fluctuate.
Can I dive with a Sensor on?
No. Your Sensor (including the installed Transmitter) is waterproof to a depth of 2.5
meters (8 feet) for up to 60 minutes (IPX8). It means the maximum pressure the device can tolerate equals the pressure in
2.5m deep in STILL water instead of flowing water. It is OK to take a shower or go swimming with the devices on, but if you go diving,
the water pressure may be too high for the devices.
176
B ureau Veritas
I didn’t see an alert message, but it appeared in History.
If one of the following alerts happened, the PDM would beep/vibrate and display a message first, and if you missed that alert, later when you checked the PDM, the condition that triggered the alert had changed (for example, your glucose level returned to the target range), then you wouldn't see any message on the screen, you would only find it in History.
1. LOW GLUCOSE
2. HIGH GLUCOSE
3. LOW PREDICTED
4. HIGH PREDICTED
5. RAPID RISE
6. RAPID FALL
7. ALERT SILENCE
8. SENSOR ERROR
9. BELOW 3.1 mmol/L (56 mg/dL)
10. LOST SENSOR If the following alert happened, the PDM would beep/vibrate and display a
message first, and if you missed that alert, later when you checked the PDM, the alert may have escalated to another alert/alarm, and you will ONLY see the message of the escalated alert/alarm. The first alert will appear in History.
Alert Alert Escalation SENSOR EXP IN 6 HOURS
SENSOR EXP IN 2 HOURS, then SENSOR EXP IN 30 MIN, at last SENSOR EXPIRED
Charging the Transmitter
We recommend that you charge the Transmitter after each Sensor session, or make sure that at least 1 minute has passed before you attach the Transmitter to a new Sensor.
Green lights after installing the Transmitter
After you install the Transmitter, the green light on the Transmitter will flash 3 times immediately indicating that the Transmitter is properly connected with the Sensor, and flash another 6 times within one minute indicating that the system check has completed.
177
B ureau Veritas
Some Sensor readings missing on the Sensor Trend Screen
If the PDM is too far away from the Transmitter, or the Bluetooth communication between the Transmitter and the PDM is temporarily interrupted, some Sensor readings might be missing in the Sensor Trend Graph screen.
Solution: Move the PDM close to the Transmitter, and wait for a while. The data will be recovered automatically.
What to do when a “Lost Sensor” alert happens
Move the PDM closer. If the PDM cannot connect with the Transmitter in 10 minutes, keep the Sensor in, disconnect the Sensor from the PDM menu, and connect again.
178
B ureau Veritas
179
B ureau Veritas
180
B ureau Veritas
7 How to use (P)LGS (Optional)
7.1 (P)LGS settings
The Glucose Alerts and Low Glucose Suspend/Predictive Low Glucose Suspend (Low Suspend/ Pre Low Suspend) functions are included under the EasyLoop Menu. The (Pre) Low Suspend function is available when CGM and Insulin Pump systems are both online. The glucose limits for alerts of Low Suspend and Pre Low Suspend are the same.
Tap EasyLoop on the Home Menu Screen to enter the EasyLoop screen.
7.1.1 Glucose Alerts
When the glucose alerts feature is turned on, the system can send you glucose alerts including High/Low Glucose, High/Low Predicted and Rate Alerts.
Main MenuEasyLoopGlucose Alerts
181
B ureau Veritas
1. You can turn on or off the Glucose Alerts.
2. Tap to save the settings.
7.1.1.1 High/Low Limits
You need to set the high and low Glucose Limits recommended by your healthcare provider after you turn the glucose alerts On. Your recommended glucose limits may vary throughout the day, you can set up to eight pairs for different time periods.
Go to the Glucose Limits screen.
Main menuEasyLoopGlucose AlertsGlucose Limits
182
B ureau Veritas
1. Add segments The starting time of the first segment is fixed to be 00:00 or 12:00A. Add time segments by choosing from 00:30-23:30 or 12:30A-11:30P, with an
increment of 00:30. You will be reminded if the time segment to be set already exists. When the
time segments are successfully set, they will be listed chronologically. If you only set one segment, the glucose limits of this segment will be applied
for 24 hours. You can set up to 8 segments with the Low and High limits for each during real-
time monitoring.
Note:
In the time segments, only the segment starting from 0:00 cannot be
deleted. You can always edit the input in each segment. The Low Limit
183
B ureau Veritas
rage is 2.8-5.0mmol/L (50-90 mg/dL), the High Limit range is 5.5-
22.2mmol/L (100-400 mg/dL), both with an increment of 0.1mmol/L (1mg/dL). The High Limit value is always larger than the Low Limit value.
In the first segment, the default Low Limit is 4.4mmol/L, the default
High Limit is 13.3mmol/L.
2. Delete Segments
Slide from right to left on one segment, tap Delete to delete this segment.
3. Tap to save the settings.
7.1.1.2 Predictive Alerts
The predictive alerts calculate when you are going to reach your Low or High Glucose Limits, and then send you an alert before you reach those limits. A predictive alert informs you that if your Sensor glucose keeps falling or rising at the current rate, you will reach your Glucose Limit in the number of minutes you set before.
Go to the Predictive Alerts screen.
Main menuEasyLoopGlucose Alerts Predictive Alerts
1. You can tap to turn on/off the Predictive Alerts
184
B ureau Veritas
2. Tap the blue plus/ minus sign to set the predictive alert time. You will be reminded of a predicted high or low glucose value some time (the predictive alert time) in advance.
Note: You can set the time between 5 min and 30 min with an increment of 5
min.
3. Tap to save the settings.
7.1.1.3 Rate Alerts
There are two types of rate alerts: Rapid Fall for Sensor glucose decreasing at or faster than your pre-selected
rate
Rapid Rise for Sensor glucose increasing at or faster than your pre-selected
rate
Go to the Rate Alerts screen.
Main menuEasyLoopGlucose Alerts Rate Alerts
185
B ureau Veritas
Go to the Rise screen.
1. You can tap to turn on/off the Rise Alerts.
2. You can choose a relative mild or an acute rising rate. See Section “Sensor
Status” for more information.
186
B ureau Veritas
3. You can also set the rate between 0.065 mmol/L/min and 0.275 mmol/L/min (1.1 mg/dL/min and 5.0 mg/dL/min) with an increment of 0.005 mmol/L/min (0.1 mg/dL/min).
Tap the blue plus/minus sign to set the rise alert. You will be reminded when your SG is rising rapidly.
4. Tap to save the settings. Go to the Fall screen.
1. You can tap to turn on/off the Fall Alerts.
2. You can choose a relative mild or an acute falling rate. See Section “Sensor
Status” for more information.
187
B ureau Veritas
You also can set the custom rate between 0.065 mmol/L/min and 0.275 mmol/L/min (1.1 mg/dL/min and 5.0 mg/dL/min) with an increment of 0.005 mmol/L/min (0.1 mg/dL/min).
Tap the blue rate value to set the fall alert time. You will be reminded when your SG is falling rapidly.
3. Tap to save the settings.
7.1.1.4 Repeat
You can set the amount of time between alerts after the first alert. After you
receive and clear “HIGH/LOW GLUCOSE”, “RAPID RISE/FALL” or “HIGH/LOW PREDICTED”, the alert will repeat in accordance with your settings until the
condition that caused the alert is resolved.
188
B ureau Veritas
Note: You can turn on or off alert. Note: You can set the Repeat time between 5 min and 3h with an increment of
5 min.
7.1.2 (Pre) Low Suspend
Go to the Low Suspend screen.
Main menuEasyLoop(Pre) Low Suspend
Note: If you turn the Low Suspend on, the Predictive Low Suspend feature will
appear.
7.1.2.1 Low Suspend
The Low Glucose Suspend feature is only available when both a Patch Pump and a Glucose Sensor are in use. The factory setting for this feature is off. If you turn it on, your PDM will automatically suspend insulin delivery and give an alarm
189
B ureau Veritas
when your Sensor glucose is at or below the low suspend limit, and resume basal insulin when the risk of low glucose no longer exits. This feature can be used as a safe guard against excessive insulin delivery. You may choose to program this feature based on the lowest acceptable Sensor glucose. Discuss what settings are best for you with your healthcare provider.
Note: The low suspend limit between 2.8 mmol/L and 5.0 mmol/L (50 mg/dL
and 80 mg/dL) based on the Glucose Low Limit settings. See Section “High/Low Limits” for more information.
Triggering Conditions for Low Suspend
The Sensor glucose value is at or below the low suspend limit.
Time of Suspension
Once Low Suspend is triggered, the period of suspension will last for at least 30 minutes unless you manually resume basal insulin. The maximum suspension time is 2 hours. After 2 hours of suspension, basal insulin will be resumed unconditionally.
Triggering Conditions for Automatic Resumption of Basal (from 30 min to 2 h after suspension)
Both of the following two conditions must be met for the system to resume basal insulin automatically.
The Sensor glucose value is at least 0.8 mmol/L (15 mg/dL) higher than the
low suspend limit.
The Sensor glucose value is predicted to be at least 1.7 mmol/L (30 mg/dL)
higher than the low suspend limit in half an hour.
Alarm Response
If the Low Suspend alarm is not cleared within 10 minutes, a siren will sound with the following Reminder.
190
B ureau Veritas
If the Low Suspend alarm is not cleared during suspension and insulin is resumed within 2 hours, the following Reminder will appear.
If the Low Suspend alarm is not cleared during suspension and insulin is automatically resumed after 2 hours, the siren will continue, and the following emergency message will appear.
191
B ureau Veritas
If the Low Suspend alarm is cleared during suspension, a Reminder will appear when insulin is automatically resumed.
For information on when the Low Suspend feature is unavailable, refer to
“Predictive Low Suspend”.
7.1.2.2 Predictive Low Suspend
The Predictive Low Glucose Suspend feature is available only when the Low Suspend feature is turned on and available. The factory setting for the Predictive Low Suspend feature is turned off. If you turn it on, your PDM will automatically suspend insulin delivery and give an alarm when your Sensor glucose is predicted to reach the low suspend limit in a set period of time, and resume basal insulin when the risk of low glucose no longer exits. This feature can be used as a safeguard against excessive insulin delivery. Discuss what settings are best for you with your healthcare provider.
Go to Pre Low Suspend screen.
192
B ureau Veritas
Main MenuEasyLoop(Pre) Low Suspend
Note: You can set the Time before Low between 5 min and 40 min with an
increment of 5 min. The factory default is 30 min.
Triggering Conditions for Predictive Low Suspend (from 30 min to 2 h after suspension)
Both of the following two conditions must be met to start Predictive Low Suspend.
The Sensor glucose value is at or within 3.9 mmol/L (70 mg/dL) above the
low suspend limit.
The Sensor glucose value is predicted to fall at or within 0.8 mmol/L (15
mg/dL) above the low suspend limit in the set period of time and the rate of glucose change is negative.
Time of Suspension
Once Predictive Low Suspend is triggered, the period of suspension will last for at least 30 minutes unless you manually resume basal insulin. The maximum suspension time is 2 hours. After 2 hours of suspension, basal insulin will be resumed unconditionally.
Triggering Conditions for Automatic Resumption of Basal
Both of the following two conditions must be met for the system to resume basal insulin automatically.
The Sensor glucose value is at least 0.8 mmol/L (15 mg/dL) higher than the
low suspend limit.
193
B ureau Veritas
The Sensor glucose value is predicted to be at least 1.7 mmol/L(30 mg/dL)
higher than the low suspend limit in half an hour.
Reminder of Resumption
Whether a Predictive Low Suspend alert is cleared or not, the same Reminder will appear when insulin is automatically resumed.
When the Low Suspend Feature and Predictive Low Suspend feature are Unavailable
After insulin delivery is resumed from Low Suspend or Predictive Low Suspend, the Low Suspend feature and Predictive Low Suspend feature will be unavailable for 30 minutes.
7.2 Summary History: Low Suspend History
Go to Low Suspend History screen.
Main MenuHistorySummary HistoryLow Suspend History
194
B ureau Veritas
This screen displays the (Pre) Low Suspend summary history. # of LGS: Average daily number of Suspend due to LGS. # of PLGS: Average daily number of Suspend due to PLGS. Time in suspend: Average daily duration suspended due to LGS or PLGS.
7.3 Troubleshooting Low Suspend issues
I didn’t see an alert message, but it appeared in History.
If one of the following alerts happened, the PDM would beep/vibrate and display a message first, and if you missed that alert, later when you checked the PDM, the condition that triggered the alert had changed (for example, your glucose level returned to the target range), then you wouldn't see any message on the screen, you would only find it in History.
Alert Alert change LOW SUSPEND After insulin delivery automatically starts again, alert is
switched to BASAL RESUMED. PRE LOW SUSPEND
After insulin delivery automatically starts again, alert is
switched to BASAL RESUMED.
195
B ureau Veritas
196
B ureau Veritas
8 Safety System and Alarms/Alerts
8.1 Safety System
Your A7+ TouchCare® System automatically performs a series of safety checks. The PDM sounds an alert or alarm and displays an on-screen message to let you know of an abnormal condition.
If you have more than one notification, you need to clear the first notification to see the next one.
Your alarm settings and alarm/alert history of the last 90 days are stored in the PDM even if the battery is depleted and will be restored once the PDM is properly charged. When the PDM battery is empty, new alarm/alert might not be successfully recorded.
Note: Do NOT set alarm (time point, limit value etc.) beyond the thresholds or
in a way that makes the safety system useless. Talk with your healthcare provider to see which settings are best for you.
Note: Your PDM and Pump consumes battery power when notifying you of
alerts, alarms, and reminders. If you do not acknowledge a notification, the PDM battery power drops fast as the notifications repeat and progress. This will result in reduced battery life and the CHARGE PDM NOW/PATCH BATT DEPLETED” Alarm or “PDM BATTERY LOW/PATCH BATTERY LOW” alert will appear sooner than expected.
8.2 Safety Checks
A single fault condition will cause the pump to suspend insulin delivery. Maximum infusion with a single fault condition is 0.05U.
8.3 Alarms
Alarms are triggered by serious or potentially serious conditions. You must respond to the alarm by taking appropriate action in order to clear the alarm condition.
For example:
197
B ureau Veritas
When Alarm “PATCH EXPIREDoccurs, the Lock Screen and Alarm screen display the following screen.
Alarm on the Lock Screen Alarm on the unlocked screen
If it is a high priority alarm, the PDM will display an alarm message with instructions and icon (a red triangle with three exclamation marks) in Alarm screen.
If it is a medium priority alarm, the PDM will display an alarm message with instructions and icon (a red triangle with two exclamation marks) in Alarm screen.
PDM Alarms in different audio modes:
Audio Mode medium priority alarm
Audio PDM emits ten beeps every twenty seconds.
Vibrate
Audio and Vibrate
Audio off / Vibrate off
PDM emits one-pulse vibration every twenty seconds
PDM emits three beeps and one-pulse vibration every twenty seconds
PDM emits one-pulse vibration every twenty seconds
198
B ureau Veritas
Patch Pump Alarms of different priorities in different audio modes:
Audio Mode high priority alarm
Audio
Vibrate
Audio and Vibrate
Audio off / Vibrate off
Audio Mode
Audio
Vibrate
Audio and Vibrate
Audio off /Vibrate off
PDM emits ten beeps every ten seconds.
PDM emits one-pulse vibration every ten seconds
PDM emits ten beeps and one-pulse vibration every ten seconds
PDM emits ten beeps every ten seconds
high priority alarm medium priority alarm
Patch Pump emits three beeps every one minute. Patch Pump emits three­pulse vibration every one minute. Patch Pump emits three beeps and three-pulse vibration every one minute. Patch Pump emits three­pulse vibration every one minute.
medium priority alarm
PDM emits ten beeps every twenty seconds.
PDM emits one-pulse vibration every twenty seconds
PDM emits three beeps and one-pulse vibration every twenty seconds
PDM emits one-pulse vibration every twenty seconds
Patch Pump emits three beeps every one minute. Patch Pump emits three­pulse vibration every one minute. Patch Pump emits three beeps and three-pulse vibration every one minute. Patch Pump emits three­pulse vibration every one minute.
199
B ureau Veritas
Alarm sound wave:
Icon Sound wave Significance
PDM emits ten beeps/vibrations each time.
PDM emits Three beeps/vibrations each time.
Patch Pump emits Three beeps/vibrations each time.
Patch Pump emits Three beeps/vibrations each time.
8.3.1 PDM Alarms
If a PDM alarm is not cleared within 10 minutes, your PDM will make a siren sound until the alarm is cleared.
PDM Message Priority Reason Actions to Take
Tap to clear it. PDM ERROR Remove device. Call customer support.
PDM ERROR The PDM has restarted. Change patch.
CHARGE PDM NOW Charge PDM now.
A PDM error is detected.
A PDM software error is detected and the PDM has restarted, but no settings have been changed.
The PDM battery is depleted.
Remove Pump and
Sensor.
Contact customer
support immediately.
Check blood glucose.
Tap to clear it.
Remove the Patch
Pump and change a
Reservoir Patch.
If the problem occurs
repeatedly, please
contact customer
support.
Tap to clear it.
Charge PDM battery.
200
B ureau Veritas
8.3.2 Pump Alarms
When a Pump alarm occurs: Indicator light: The indicator light on the pump flashes red once per second until
the alarm is cleared.
Note: If a Pump alarm is not cleared within 10 minutes, both your PDM and
Patch Pump will make a siren sound until the alarm is cleared. The following table lists high priority alarm messages.
PDM Message Priority Reason Actions to Take
OCCLUSION DETECTED Delivery stopped. Change Patch now.
Pump occlusion is detected.
Tap to clear it. Change Patch. Check blood glucose.
PATCH ERROR Delivery stopped. Change Patch now.
PUMP BASE ERROR Remove Pump. Call customer support.
The following table lists medium priority alarm messages.
PDM Message Priority Reason Actions to Take
AUTO OFF Delivery suspended. No status received.
A Reservoir Patch error is detected.
A Pump Base error is detected.
The PDM has not received a pump status during the time limit set.
Tap to clear it. Change Patch. Check blood glucose.
Tap to clear it. Remove Pump. Contact customer support immediately. Check blood glucose.
Tap to clear it. Resume basal delivery. Check blood glucose and treat it as necessary. Check Pump history.
201
B ureau Veritas
PDM Message Priority Reason Actions to Take PATCH EXPIRED
Delivery will stop. Change Patch now.
PATCH BATT DEPLETED Delivery stopped. Change Patch now.
EXCEEDS MAX TDD Exceeds max TDD. Delivery stopped.
EXCEEDS MAX 1HR DELIVERY Exceeds 1 hour max. Delivery stopped.
EMPTY RESERVOIR Delivery stopped. Change Patch now.
LOW SUSPEND Low glucose suspend activated.
The current Reservoir Patch has reached the end of its 3-day operating life.
The Reservoir Patch battery is depleted.
You have attempted to deliver more insulin than expected based on your Daily Max setting.
You have attempted to deliver more insulin than expected based on your Hour Max setting.
There is no insulin in the reservoir.
The last Sensor glucose reading is at or below the Low Glucose Suspend Limit set.
Tap to clear it.
Change Patch.
Check blood glucose.
Tap to clear it.
Change Patch.
Check blood glucose.
Tap to clear it.
Check blood glucose.
Resume basal delivery.
Check bolus history and
reevaluate your need
for insulin.
Continue to monitor
blood glucose.
Tap to clear it.
Check blood glucose.
Resume basal delivery.
Check bolus history and
reevaluate your need
for insulin.
Continue to monitor
blood glucose.
Tap to clear it.
Change Patch.
Check blood glucose.
Tap to clear it.
Check blood glucose
and treat it as
necessary.
202
B ureau Veritas
If the following alarm is not cleared within 10 minutes, only your PDM will make a siren sound until the alarm is cleared.
PDM Message Priority Reason Actions to Take PUMP OUT OF
RANGE Low Suspend failed. Move PDM close to Pump. Pre Suspend failed. Move PDM close to Pump.
Low Suspend or Predictive Low Suspend failed because the PDM had lost communication with the Patch Pump.
Tap to clear it. Move PDM close to Pump.
8.4 Alerts
Alerts are triggered by conditions that may require your attention. Alerts are less serious than alarms. You must respond to an alert by pressing buttons and/or taking actions.
For example: When Alert “PATCH BATTERY LOWoccurs, the Lock Screen and Alert screen display the following screen.
Alert in Lock Screen Alert after unlock in Alert screen
203
B ureau Veritas
The PDM displays an alert message with instructions and icon (an empty triangle with exclamation mark) in Alert screen.
CGM Alerts and PDM Alerts in different audio modes:
Audio Mode Alert
Audio PDM emits two beeps every three minutes.
Vibrate PDM emits one-pulse vibration every three minutes.
Audio and Vibrate
Audio off /Vibrate off
Patch Pump Alerts in different audio modes:
Audio Mode Alert
Audio PDM emits two beeps every three minutes. Vibrate
Audio and Vibrate Audio off /Vibrate off no beeping, no vibration
Audio Mode Alert
Audio
Vibrate
Audio and Vibrate
Audio off /Vibrate off
PDM emits two beeps and one-pulse vibration every three minutes.
no beeping, no vibration
PDM emits one-pulse vibration every three minutes.
PDM emits two beeps and one-pulse vibration every three minutes.
Patch Pump emits three beeps every three minutes.
Patch Pump emits three-pulse vibration every three minutes.
Patch Pump emits three beeps and three­pulse vibration every three minutes.
Patch Pump emits three-pulse vibration every three minutes.
204
B ureau Veritas
The sound wave of every alert beep:
Icon Sound wave Significance
Your PDM emits two beeps every time.
8.4.1 PDM Alerts
The following table lists alert messages for PDM.
PDM Message Priority Reason Actions to Take
PDM BATTERY LOW Low PDM battery. Charge battery soon.
The PDM battery is low.
Tap to clear it. Charge PDM battery soon.
8.4.2 Pump Alerts
When a Pump alert occurs: Indicator light: The indicator light on the Patch Pump flashes yellow once every
two seconds until the alert is cleared. The following table lists alert messages for Patch Pump.
PDM Message
END OF SUSPEND Delivery suspended at [].
LOW RESERVOIR [] remaining. Change Patch.
AUTO OFF ALERT Delivery stops if not cleared in 15 min.
Priority Reason Actions to Take
Insulin delivery has been suspended for more than 15 minutes. The insulin level in the Reservoir Patch has reached the set low limit.
The PDM did not receive a Pump status during the time limit set.
205
Tap to clear it. Resume basal delivery.
Tap to clear it. Change Patch soon.
Tap to clear it. Check blood glucose. Check Pump history.
B ureau Veritas
PDM Message PATCH EXP
ADVISORY Patch expiration in [] hours. PATCH EXP IN 1 HOUR Patch expiration in 1h. Change Patch soon.
PATCH BATTERY LOW No bolus allowed. Change Patch soon.
PRE LOW SUSPEND Delivery suspended. Predictive low glucose.
The response mode of the following alert is same as PDM alert.
PDM Message
PUMP RESTARTED Patch changed? For help call the CC.
Priority Reason Actions to Take
The Reservoir
Priority Reason Actions to Take
Patch will expire within the set time limit.
The Reservoir Patch will expire in less than 1 hour.
The Patch battery is running low. No bolus can be delivered. Basal delivery can only last about 30 minutes. The Sensor glucose may reach the Low Glucose Suspend Limit in the length of time set.
Pump restarted without Patch deactivation.
Tap to clear it. Change Patch soon.
Tap to clear it. Change Patch soon.
Tap to clear it. Change Reservoir Patch soon.
Tap to clear it. Check blood glucose and treat it as necessary.
Tap to clear it. Check if a new Patch is connected, and follow the instructions in this User Guide. Call customer support if you have any questions.
8.4.3 CGM Alerts
If you set audio option to Audio off/Vibrate off, your PDM will neither beep nor vibrate for all CGM alerts, except:
206
B ureau Veritas
When “BELOW 3.1 mmol/L(BELOW 56 mg/dL)” occurs, your PDM emits three­pulse vibration every three minutes. If not cleared within 9 minutes, your PDM will make a siren sound until the alert is cleared.
When “TRANSMITTER ERROR”, “CHARGE TRANSMITTER”, “SENSOR EXPIRED”, or “SENSOR FAILURE” occurs, your PDM emits three-pulse vibration every three
minutes. The following table lists alert messages for CGM.
PDM Message Priority Reason Actions to Take
TRANSMITTER BATTERY LOW Charge Transmitter soon. CHARGE TRANSMITTER Charge Transmitter now. TRANSMITTER ERROR Call customer support.
NO READINGS Check or change Sensor.
SENSOR EXPIRED Sensor session ended. Change Sensor.
SENSOR FAILURE Sensor session ended. Change Sensor.
The Transmitter battery is close to running out of power. The Transmitter battery is depleted. (MD1026) A Transmitter error is detected.
The Sensor signals are abnormal.
The current Sensor has reached its 14­day operating life.
The Sensor is not functioning properly.
Tap to clear it. Charge Transmitter soon.
Tap to clear it. Charge Transmitter.
Tap to clear it. Call customer support. Tap to clear it. Check if the Sensor gets bumped or dislodged, make sure that the Sensor is inserted correctly or change Sensor.
Tap to clear it. Change Sensor.
Tap to clear it. Change Sensor.
207
B ureau Veritas
PDM Message Priority Reason Actions to Take
METER BG NOW Enter a new meter BG for calibration or tap OK to clear the alert.
SENSOR CAL ERROR Enter a meter BG after 15 minutes.
LOW GLUCOSE Glucose level below Low Limit.
HIGH GLUCOSE Glucose level above High Limit.
LOW PREDICTED Glucose may reach Low Limit in [ ] min.
HIGH PREDICTED Glucose may reach High Limit in [ ] min.
A meter BG is needed immediately to calibrate the Sensor. The Sensor hasn't been calibrated properly.
The last Sensor glucose reading is at or below the Low Glucose Limit.
The last Sensor glucose reading is at or above the High Glucose Limit.
The Sensor glucose may reach Low Glucose Limit in the length of time.
The Sensor glucose may reach High Glucose Limit in the length of time.
Tap to clear it. Enter new meter BG for calibration or tap OK to clear the alert.
Tap to clear it. Enter meter BG after 15 minutes.
Tap to clear it. Check blood glucose and treat it as necessary. Continue to monitor blood glucose. Tap to clear it. Check blood glucose and treat it as necessary. Continue to monitor blood glucose. Tap to clear it. Check blood glucose and treat it as necessary. Continue to monitor blood glucose. Tap to clear it. Check blood glucose and treat it as necessary. Continue to monitor blood glucose.
208
B ureau Veritas
PDM Message Priority Reason Actions to Take
Tap to clear it.
RAPID RISE Sensor glucose is rising rapidly.
RAPID FALL Sensor glucose is falling rapidly.
BELOW 3.1 mmol/L Sensor glucose below
3.1 mmol/L. (BELOW 56 mg/dL Sensor glucose below 56 mg/dL.)
SENSOR EXP IN 6 HOURS Change Sensor in 6 hours.
SENSOR EXP IN 2 HOURS Change Sensor in 2 hours.
SENSOR EXP IN 30 MIN Change Sensor in 30 minutes.
The Sensor glucose is rising at a rate that is faster than the set Rise limit.
The Sensor glucose is falling at a rate that is faster than the set Fall limit.
The last Sensor glucose reading is at or below
3.1 mmol/L. (The last Sensor glucose reading is at or below 56 mg/dL.) The current Sensor session has 6 hours left until its period ends. The current Sensor session has 2 hours left until its period ends. The current Sensor session has 30 minutes left until its period ends.
Monitor trend and glucose level. Follow instructions from your healthcare provider. Tap to clear it. Monitor trend and glucose level. Follow instructions from your healthcare provider.
Tap to clear it. Check blood glucose and treat it as necessary. Continue to monitor blood glucose.
Tap to clear it. Change Sensor in 6 hours.
Tap to clear it. Change Sensor in 2 hours.
Tap to clear it. Change Sensor in 30 minutes.
209
B ureau Veritas
PDM Message Priority Reason Actions to Take
The PDM has LOST SENSOR Move PDM close to Transmitter.
If Alert Silence is turned on, the PDM gives no beep or vibration when an alert occurs. The ALERT SILENCE message will be displayed on your PDM screen instead, and you can check the alert in Sensor alert history. See Chapter “How
to use CGM system” for more information.
PDM Message Priority Reason Actions to Take
ALERT SILENCE Alerts have occurred. Check Sensor history.
Note:
If the Audio is on and the Alert Silence is off, the audio off icon will not appear in the top right corner of this screen.
not received a
signal from the
Transmitter for
10 minutes.
Sensor alerts have occurred during silence mode.
Tap to clear it. Move PDM close to Transmitter.
Tap to clear it. Check the Sensor Alert History. Take action based on the alert occurred.
If the Audio and Alert Silence are on, the temporary audio off icon” ”will appear in the top right corner of this screen.
210
B ureau Veritas
If the Audio is off, the audio off icon” will appear in the top right corner of this screen.
8.5 Reminding messages
Reminding messages are automatically displayed to remind you of a condition, function or event. Reminding messages include the notifications you get after setting reminders and the low-priority reminding notifications. A Message requires you to press buttons to clear it and/or to take action if necessary.
For example: When Message “BASAL RESUMEDoccurs, the Lock Screen and Message screen display the following screen.
211
B ureau Veritas
Message in Lock Screen Message after unlock in Alert screen
Audio/vibration: Your PDM emits two beep and/or one vibration every three minutes, three times in total.
8.5.1 PDM reminding messages
Condition PDM Message Reason
CHECK
SETTINGS
ALARM
CLOCK
HIGH BG
LOW BG Treat low BG. Monitor BG.
Check all settings.
Alarm Clock.
Treat high BG. Monitor BG.
An error might have occurred to your settings. An alarm clock is set at this time.
The blood glucose entered is higher than 13.9 mmol/L (250 mg/dL). The blood glucose entered is lower than 3.9 mmol/L (70 mg/dL).
8.5.2 Pump reminding messages
Condition PDM Message Reason
BG Reminder is turned on to
CHECK BG Check your BG.
BOLUS
REMINDER
Bolus is not delivered in specified period.
remind you to check meter BG after a bolus. Bolus Reminder is turned on to remind you to deliver a bolus within a specific period.
212
B ureau Veritas
Condition PDM Message Reason
ACTIVE
BASAL
EMPTY
BASAL
RESUMED
Your active Basal is 0.00 U/H.
Check BG. [ ] Basal active.
The selected basal rate or temp basal rate is 0.00 U/H.
The previously suspended basal rate is automatically resumed.
8.5.3 CGM reminding messages
Condition PDM Message
SENSOR CAL
REMINDER
SENSOR CAL
FAILED
Enter a new meter BG for CAL by [ ].
Sensor calibration failed. Please retry to calibrate later.
A meter BG must be entered by the time shown to calibrate Sensor.
A meter BG is needed a few minutes later to calibrate the Sensor.
Reason
SENSOR
RECONNECT
ED
Old sensor disconnected. New sensor has been connected.
213
The old sensor is disconnected and a new sensor is directly connected.
B ureau Veritas
214
B ureau Veritas
9 Manufacturer’s Declaration
9.1 Electromagnetic Emissions
Emissions Test Compliance
RF emissions EN 60601-1-2:2007+AC:2010, IEC 60601-1-2:2007, CISPR 11:2009+A1:2010 and IEC 60601-1-2:2014 RF emissions EN 60601-1-2:2007+AC:2010, IEC 60601-1-2:2007, CISPR 11:2009+A1:2010 and IEC 60601-1-2:2014
9.2 Electromagnetic Immunity
Group 1
Class B
Immunity Test IEC 60601 Test
Level The A7+ TouchCare® System is intended for use in the electromagnetic environment specified below. The customer or user of the A7+ TouchCare® System should assure that it is used in such an electromagnetic environment.
Electrostatic Discharge IEC 61000-4-2
Power Port, Signal and Iterconnecting Cable
±2kV, ±4kV, ±8kV contact discharge ±2kV,±4kV,±8k V, ±15kV air discharge
Table 5 of IEC 60601-1:2014
±2Kv, 100Hz,
Compliance
Level
±2kV, ±4kV, ±8kV contact discharge ±2kV,±4kV, ±8kV, ±15kV air discharge
During the test, the EUT can
Electromagnetic
Environment
For home healthcare environment and professional healthcare facility environment
The network power supply should have the quality used in a typical
215
B ureau Veritas
Immunity Test IEC 60601 Test
Level The A7+ TouchCare® System is intended for use in the electromagnetic environment specified below. The customer or user of the A7+ TouchCare® System should assure that it is used in such an electromagnetic environment.
IEC 61000-4­4:2012
Surge
IEC 61000-4­5:2005
GB/T 17626.11 Votage dips and interruptions to AC Power Port IEC 61000-4­11:2014
Power Frequency
for AC power port ±0.5kV, ±1kV (different mode) ±0.5kV, ±1kV, ±2kV (common mode)
0%UT;0.5T (0°, 45°, 90°, 135°, 180°, 225°, 270° and 315°) 0%UT;1T(0°) 70%UT 20T(0°); 0%UT 250T(0°)
Table 4 of IEC 60601-1­2:2014 30A/m,
Compliance
Level
operate as intended
During the test, the EUT can operate as intended
0.5T(10ms); 1T (20ms); 25T(500ms) ; 250T(5s)
30A/m Suitable for most
Electromagnetic
Environment
commercial or hospital environment. The network power supply should have the quality used in a typical commercial or hospital environment.
The network power supply should have the quality used in a typical commercial or hospital environment. If the A7+ user needs continuous operation during a power outage, it is recommended that the A7+ be powered by an uninterruptible power supply or battery.
environments, if there is no industrial magnetic
216
B ureau Veritas
Immunity Test IEC 60601 Test
Level The A7+ TouchCare® System is intended for use in the electromagnetic environment specified below. The customer or user of the A7+ TouchCare® System should assure that it is used in such an electromagnetic environment.
magnetic fields IEC 61000-4-8
Proximity fields from RF wireless communication equipment IEC 61000-4­3:2006+A1+A2 RF electromagnetic field immunity test IEC 61000-4­3:2006+A1+A2
50HZ and 60HZ
Table 9 of IEC 60601-1­2:2014
IEC 61000-4­3:2006+A1+A2 10V/m (for home healthcare environment and professional healthcare facility environment) 80 MHz~2.7 GHz
Compliance
Level
During the test, the EUT can operate as intended
10V/m (for home healthcare environme nt) 3V/m (for professiona l healthcare facility environme nt) 80 MHz~2.7 GHz
Electromagnetic
Environment
equipment nearby, the magnetic field strength will not exceed 400A/m Recommended separation distance d = [12/E1] P 80 MHz to 800 MHz d = [23/E1] P 800 MHz to 6 GHz Where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is there commended separation distance in meters (m). Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, should be less than the compliance
217
B ureau Veritas
Immunity Test IEC 60601 Test
Level The A7+ TouchCare® System is intended for use in the electromagnetic environment specified below. The customer or user of the A7+ TouchCare® System should assure that it is used in such an electromagnetic environment.
Note: UT means to the AC network voltage before the test voltage is
applied.
Note: At 80 MHz and 800 MHz, the higher frequency range applies. Note: These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption, and reflection from structures, objects and people.
Note: The table is per IEC (EN) 60601-1-2 Edition 3.
Field strengths
A. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcasts and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the A7+ TouchCare® System is used exceeds the applicable RF compliance level above, the A7+ TouchCare®
Compliance
Level
Electromagnetic
Environment
level in each frequency range. Interference may occur in the vicinity of equipment marked with the following symbol:
218
B ureau Veritas
System should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re­orienting or relocating the A7+ TouchCare® System. B. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m.
Electrostatic discharge
Although your A7+ TouchCare® System is designed to be unaffected by typical levels of electrostatic discharge (ESD), very high levels of ESD can result in reset of the A7+ TouchCare® System. If PDM restarts, please verify the PDM settings to ensure all settings are set to the desired values. If Pump Restarted occurs, please change a new patch. If CGM restarts, please recharge the transmitter and change a new sensor.
For more information on changing a new patch, see Chapter “How to use Patch Pump. For more information on changing a new sensor, see Chapter “How to use CGM system (Optional). For more information on re-entering your PDM settings, see Section Settings in Chapter How to use the PDM.
If you are unable to re-enter your PDM settings, change a new patch or sensor, or otherwise believe there is a problem with your device, contact your local representative.
219
P2.1d =
P2.1d =
P3.2d=
B ureau Veritas
Recommended separation distances between portable and mobile RF
communications equipment and the A7+ TouchCare® System
The A7+ TouchCare® System is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or user of the A7+ TouchCare® System can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the A7+ TouchCare® System as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output power of transmitter (W)
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73 1 1.2 1.2 2.3 10 3.8 3.8 7.3 100 12 12 23 For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where p is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note: At 80 MHz and 800 MHz, the separation distance for the higher
frequency range applies.
Note: These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects and people.
Separation distance according to the frequency of transmitter (m)
150 kHz~80 MHz
80 MHz~800 MHz
800 MHz~2.5 GHz
Warning:
1. EMC (Electromagnetic Compatibility) information in this guide should be referred before installing and using the A7+ TouchCare® System.
2. The A7+ TouchCare® System is not designed to be used in an environment with high voltage, high-intensity magnetic field, where the intensity of EM DISTURBANCES is high.
3. Portable RF Communications equipment should be used no closer than 30 cm (12 inches) to any part of the Medtrum products. Otherwise, degradation of the performance of this equipment could result.
220
B ureau Veritas
4. It should be avoided to use this equipment adjacent to or stacked with other medical equipment, because it could result in improper operation. If such use is necessary, this equipment and the other medical equipment should be observed to verify that they are operating normally.
5. Other cables and accessories may negatively affect EMC performance.
221
B ureau Veritas
222
LOT
REF
STERILEEO
STERILE R
SN
0197
B ureau Veritas
10 Appendix I: Symbols and Icons
10.1 Product Label Symbols
Symbol Meaning Symbol Meaning
Do NOT use if
Lot number
package is damaged
Reference number
Manufacturer
Use by: (yyyy­mm-dd)
Caution: See Instructions for use
Storage temperature
Do NOT reuse
CE mark by notified body
Waste Electrical and Electronic Equipment
IPX8
Sterilized using ethylene oxide
Sterilized using radiation
Follow instructions for use
Radio communication
Waterproof to 2.5 m for 1 hour
Device serial number
Type BF equipment (Protection from electrical shock)
Authorized representative in the European Community
223
B ureau Veritas
Symbol Meaning Symbol Meaning
Protection Against Insertion
IP22
of Large Objections and Dripping Water IEC 60529
10.2 PDM Icons
Icon Meaning Icon Meaning
High priority
alarm
Alert
Audio
temporary off
Pump RF signal
Charging
00:00 a
Medium priority
alarm
Audio off
Time
Battery
Charged
224
B ureau Veritas
225
B ureau Veritas
226
B ureau Veritas
11 Appendix II: Technical Information
11.1 PDM Specifications
Model: FM-018 Size: 76.2 x 48.4 x 9.375mm Weight: 42.4 g Screen: 2.4 in Operating Temperature Range: +5°C ~ +40°C Operating Relative Humidity Range: 20%~ 90%RH Operating Atmospheric Pressure: 700~1060 hPa Storage Temperature Range: -10°C ~ +55°C Storage Relative Humidity Range: 20%~ 90%RH Storage Atmospheric Pressure: 700~1060 hPa Classification: Internally powered, Continuous operation Battery: Built-in 3.8 V polymer lithium ion battery Power: 5.0VDC, 1.0A Battery Life: Approximately 1 week once fully charged. Data Storage: Automatically stores the previous 90 days’ data Wireless Communication Distance: 10 m with the Transmitter, 4 m with the
insulin pump
Alarm Type: Visual, audible and vibratory Volume: 52.3 dB(A) measured from 1 m distance Limited Warranty: 4 years Dust-proof and Waterproof Rating: IP22
11.2 Patch Pump Specifications
Model:
Pump Base: JN-022 Reservoir Patch: MD-JN-011
Size: 56.5mm x 33.3mm x 13.3 mm Weight: 21.5 g (without insulin) Operating Temperature Range: +5 °C ~ +40 °C Operating Relative Humidity Range: 20%~90%RH Operating Atmospheric Pressure: 700~1060 hPa
227
B ureau Veritas
Storage Temperature Range: -10°C ~ +55°C Storage Relative Humidity Range: 20%~90%RH Storage Atmospheric Pressure: 700~1060 hPa Classification: Internally powered, Type BF applied parts, Continuous
operation
Battery: Powered by two button batteries (1.5 V) Wireless Communication Distance: 4 m Waterproof Rating: IPX8 (2.5 m, 60 min) Limited Warranty of Pump Base: 1 year Shelf Life of Reservoir Patch: 2 years Sterilization Method of Reservoir Patch: By EO gas Reservoir Volume: 200 U (2 mL) (1 U=10 μL) Insulin Type Used: U-100 Basal Rate Range: 0.00~ 25 U/h (increment: 0.05 U/h) Bolus Range: 0.05 ~ 30 U (increment: 0.05 U) Bolus Delivery Rate: 0.05 U/2 s Maximum Infusion Pressure and Occlusion Pressure Threshold: 15 psi Maximum Time to Occlusion Alarm:
Basal Delivery (0.1 U/h): < 30 h Basal Delivery (1 U/h): < 3 h Bolus Delivery (3 U at 1.5 U/min): < 120 s
Bolus Volume after Occlusion Release: < 3 U Delivery Accuracy:
Basal: +/- 5% (at rates: 0.1~ 25 U/h) Bolus: +/- 5% (for all set values: 0.05 ~ 30 U) Accuracy Test Results (test cycle: 29 H, delivery rate: 1.0 U/H, average error:
0.40%):
228
B ureau Veritas
Note: The Patch Pump may not be able to achieve the above measurement
accuracy under certain circumstances such as vigorous exercise, or abnormal operating conditions.
229
B ureau Veritas
230
B ureau Veritas
11.3 Transmitter Specifications
Model: MD1026 Size: 36.1 mm x 19.4 mm x 7.2 mm Weight: 3.57 g Operating Temperature Range: +5°C ~+40°C Operating Relative Humidity Range: 20%~90%RH Operating Atmospheric Pressure: 700~1060 hPa Storage Temperature Range: -10°C~+55°C Storage Relative Humidity Range: 20%~90%RH Storage Atmospheric Pressure: 700~1060 hPa Battery: Built-in 3.7 V polymer lithium ion battery Waterproof Rating: IPX8 (2.5 m, 60 min) Classification: Type BF equipment, Continuous operation Data Storage: Automatically stores the previous 14 days’ data Wireless Communication Distance: 10 m Limited Warranty: 1 year
11.4 Glucose Sensor Specifications
Model: MD3026 Storage Temperature Range: +2°C~+30°C Storage Relative Humidity Range: 20%~90%RH Storage Atmospheric Pressure: 700~1060 hPa Glucose Range: 2.2~22.2mmol/L (40~400mg/dL) SterilizationMethod: By radiation Sensor Life: Up to 14 days
11.5 CGM System Accuracy
A multi-center, randomized clinic study is designed to determine the Sensor accuracy in adults with Type 1 or Type 2 diabetes. In-clinic testing consisted of frequent venous blood sample testing (by Yellow Springs Instrument 2300 STAT Plus™ Glucose Analyzer, YSI) on a random day in the Sensor life. The accuracy is based on the percentage of CGM glucose readings that are within ± 20%, 30% or 40% of YSI values at glucose values at or above (>=) 100 mg/dL (5.6
231
B ureau Veritas
mmol/L), or within ±20 mg/dL (1.1 mmol/l), 30 mg/dL (1.7 mmol/L) or 40 mg/dL (2.2 mmol/L) of YSI values at glucose values below (<) 100 mg/dL (5.6 mmol/L).
Table. Percentage of CGM Glucose Readings within ±20%/20 mg/dL , ±30%/30 mg/dL, or ±40%/40 mg/dL of the YSI; Calibrating every 12 hours, Abdomen insertion site.
Number of Matched Pairs CGM-YSI
1678 91% 97% 99%
±20%/20 mg/dL ±30%/30 mg/dL
±40%/40 mg/dL
232
B ureau Veritas
233
B ureau Veritas
234
Loading...