Symptom Cause Solution
TROUBLESHOOTING GUIDE
Please refer to the NIM-Response® 2.0 System User’s Guide (82-50651) or the NIM-Neuro® 2.0
System User’s Guide (82-50650) for complete operating instructions.
No visual display or audio alarms
at power-up.
Touching the screen has
unexpected results.
Electrode impedance is too high.
>10K
Ω for subdermal electrodes
>25K
Ω for Prass Paired electrodes
>10K
Ω for EMG tube
>40K
Ω for hookwire electrodes
Electrode impedance = 0.0K
Ω.
Channel button is flashing.
Electrode reading is “— K
Ω”
or “OFF.”
Electrode difference is greater
than 2KΩ (subdermal electrodes)
or 10K
Ω (Prass Paired electrodes).
Electrosurgical interference.
Interference with anesthesia
equipment.
Excessive muting.
Inadequate muting.
Power cord not connected to outlet or to
the NIM™ 2.0 system.
Power switch not turned on.
Touch screen out of calibration.
Electrode dislodged from patient, but not
completely out.
High resistance in electrode.
Electrode pin not firmly inserted into Patient
Interface.
Positive and negative electrodes touching
below surface of skin.
Extremely low impedance, particularly in EMG
tubes.
Electrode laying on skin surface.
Electrode placement insecure.
Dirty electrode tip.
Electrode cable broken.
Electrode pin disconnected from
Patient Interface.
Dirty electrode.
Mismatched pair.
Unequal placement.
Muting Probe not connected.
Muting Probe input insufficient.
Electrosurgical grounding inadequate.
Source of interference unidentified.
NIM 2.0 system or Patient Interface cable too
close to ESU or its cables.
Lead checking current near anesthesia electrodes.
Unit receiving excessive signal into the Muting
Probe or electrode leads.
Signal from ESU is inadequate to cause muting.
Plug in power cord.
Turn power switch on.
Turn unit off, then press the screen until the screen
calibration test is displayed. Follow the instructions on
the screen to recalibrate.
Insert dislodged electrode; tape down in place.
Remove and replace with new electrode.
Check connection at Patient Interface box.
Remove and relocate electrodes.
Use “tap test” near electrodes to evoke EMG or artifact.
If activity is noted on channel in question, proceed.
Re-insert electrode in question.
Remove and replace electrode in question.
Check connection to Patient Interface box.
Remove and replace electrode for appropriate channel
with highest impedance reading first.
Remove and replace electrode in question.
• Check Muting Probe connections.
• Move input to “MORE MUTE”.
• Check electrosurgical grounding pad on patient.
• Identify source of interference; then eliminate or
separate from the NIM 2.0 system.
• Maintain separation between electrosurgical cable and
the NIM 2.0 system.
• For less coupling, coil up the Muting Probe next to the
NIM 2.0 system.
Have anesthesia try alternate electrode channel.
Turn stimulator to 0.0mA when not stimulating.
Move the Muting Probe connector to a lower number
unit until it stops muting. If it still mutes in position “1,”
disconnect the muting detector completely.
Move the Muting Probe connector to a higher number
until it mutes. If it still does not mute in position “4,”
loop the ESU cable and clip the muting detector over
the doubled cable.