Medtronics NIM 2.0 User Manual

A basic guide to nerve integrity monitoring
NIM™ 2.0 SYSTEMS:
Protocol and Troubleshooting Guide
WELCOME
Medtronic ENT has developed this basic guide to nerve integrity monitoring as it applies to the NIM™ 2.0 System family to assist in simplifying setup procedures. (The NIM 2.0 System family includes the NIM-Response®2.0 and the NIM-Neuro®2.0.)
Each section gives procedure-specific protocols to follow and includes such information as: monitoring goals, electrode placement, impedance values, difference values, threshold and stimulation ranges, as well as samples of responses.
We hope that this booklet will be a helpful guide to follow and we appreciate your patronage.
Important: This document is not intended to replace the surgeon’s medical judgment or knowledge of neural anatomy and physiology. Nerve monitoring does not prevent the surgical severance of nerves. Nerve monitoring with the NIM 2.0 System family or other monitors is only a technical aid and cannot substitute for the skill, experience and anatomical knowledge of the surgeon.
The user should refer to Medtronic ENT’s Operations Manual for further instruction regarding any equipment referenced in this document. Requisite training and know-how for performing evoked EMG monitoring in surgical applications supplements the surgeon’s knowledge of nerve anatomy for preservation of nerve function during surgical procedures.
Table of Contents
Intracranial (Skull Base)...........................................................................................................2-3
Cerebellopontine Angle Tumor • Vestibular Schwannoma (acoustic neuroma) • Microvascular Decompression • Trigeminal Nerve Resection • Vestibular Nerve Section
Intratemporal..............................................................................................................................4-5
Facial Nerve Decompression • Mastoidectomy • Tympanoplasty • Cochlear Implantation • Translabrynthine Approach to Posterior Fossa • Labyrinthectomy
Extratemporal.............................................................................................................................6-7
Parotidectomy • Submandibular Gland Dissection • Head and Neck Dissection • Congenital Aural Atresia
Neck Dissections.......................................................................................................................8-9
Thyroidectomy • Parathyroidectomy • Radical Neck Dissection • Anterior Cervical Fusion • Substernal Goiter • Hemithyroidectomy
Monitoring Tips.......................................................................................................................10-13
Tips for Reducing Artifact • Verifying Stimulus Delivery • Examples of EMG Response: Stimulated EMG Response, Mechanically Evoked EMG Response, Train Response, Metal Artifact, Electrode Placement
Troubleshooting Guide .........................................................................................................14-15
1
2
Skull base procedures may place at risk that segment of the facial nerve (Cranial Nerve VII) from the brainstem through the internal auditory canal (IAC). Tumors may be located either adjacent to the brain stem or within the IAC and are often intimately involved with the facial nerve. In order to remove the tumor, there will be considerable drilling of bone which may produce heat that may affect the nerve. Nerve dissections often involve direct manipulation, stretching or traction of the facial nerve. The facial nerve is nonmyelinated at this section, hence lesser amounts of stimulation will produce an EMG effect.
If other cranial motor nerves are at risk, additional electrodes may be placed to monitor EMG activity in the appropriate muscle. Additional cranial nerves that are commonly monitored in skull base procedures include the trigeminal nerve (Cranial Nerve V) and the vagus nerve (Cranial Nerve X).
As a general rule, when shunting current directly through bone, you will need 1.0mA of stimulus for every millimeter of bone above the nerve. Shunting through soft tissue will usually require a setting of at least 0.8mA.
Monitoring Goal
Locate, identify, and map the nerve.
Monitor manipulation effect.
It is extremely valuable for the surgeon to verify nerve integrity prior to closing by stimulating proximal to the tumor site BEFORE and AFTER the surgery.
Cerebellopontine Angle Tumor
Vestibular Schwannoma (acoustic neuroma)
Microvascular Decompression
Trigeminal Nerve Resection
Vestibular Nerve Section
INTRACRANIAL (SKULL BASE)
Procedures
3
INTRACRANIAL (SKULL BASE)
Electrode Placement: Figure 1.
Procedure Specific Protocols
Suggested Electrodes Part No.
• Paired Subdermal Electrodes 8227410
• Prass Paired Electrodes, 18mm 8227304
Suggested Stimulator Probes Part No.
• Prass Standard Monopolar Probe 8225101
• Prass Bipolar Probe 8225451
Suggested Stimulus Instruments Part No.
• Kartush Dissection Set (KSD) 1352400
• Neurotologic Dissection Set (NSD) 1353400
Typical Electrode Readings Channel Values Difference Values
Subdermal needle < 10 KΩ < 1 KΩ Prass paired needle < 25 KΩ < 5 KΩ
Event Threshold Range
Event Threshold Range is 50uV-100uV. You may wish to increase threshold in case of an EMG train response.
Stimulation Range
Use 0.8mA to begin, in general, unless otherwise directed by the surgeon at the start of surgery when the monopolar probe is in use with goal of mapping the nerve. Increase the stimulus setting until an EMG response has been elicited. At this stage, there will be tissue and bone through which the current must travel. Once the nerve has been located, reduce the stimulus level. There is rarely a need to use more than
1.0mA in the cerebellar pontine angle (CPA) because it is possible to place a stimulator directly on the nerve. The Kartush Bipolar Probe is good to use when the nerve is exposed because current is not shunted and low stimulus levels can be used.
Operative Side
Orbicularis Oculi m.
Orbicularis Oris m.
Ground
Stim Return (+)
4
The facial nerve may be at risk during surgeries involving the temporal bone because the exact location of the nerve varies by patient. This is especially true for revision cases. These procedures, generally, are “quiet surgeries” without a lot of artifact.
Intraoperative facial nerve monitoring (IFNM) allows the surgeon to avoid the nerve. In procedures with considerable bone drilling, heating may affect the nerve. In nerve decompressions, IFNM can aid in pinpointing that portion of the nerve to be decompressed.
Monitoring Goal
Locate, identify, and map the nerve.
Monitor manipulation effect.
It is extremely valuable for the surgeon to verify nerve integrity prior to closing by stimulating after the surgery.
Facial Nerve Decompression
Mastoidectomy
Tympanoplasty
Cochlear Implantation
Translabrynthine Approach to Posterior Fossa
Labyrinthectomy
INTRATEMPORAL
Procedures
5
INTRATEMPORAL
Electrode Placement: Figure 2.
Procedure Specific Protocols
Suggested Electrodes Part No.
• Paired Subdermal Electrodes 8227410
• Prass Paired Electrodes, 18mm 8227304
Suggested Stimulator Probes Part No.
• Prass Standard Monopolar Probe 8225101
• Kartush Side-by-Side Bipolar Probe 8225401
Suggested Stimulus Instruments Part No.
• Kartush Dissection Set (KSD) 1352400
• Neurotologic Dissection Set (NSD) 1353400
Typical Electrode Readings Channel Values Difference Values
Subdermal needle < 10 KΩ < 1 KΩ Prass paired needle < 25 KΩ < 5 KΩ
Event Threshold Range
Event Threshold Range is 50uV-100uV. May decrease to 50uV to see earliest response.
Stimulation Range
Use 0.8mA at the beginning (mapping) of the procedure or as directed by the surgeon. Hold the probe tip perpendicular to the tissue for approximately one second to elicit a response.
There is rarely a need to use more than 1.0mA in the cerebellar pontine angle (CPA) because it is possible to place a stimulator directly on the nerve. Use lower stimulus levels as directed by the surgeon and increase slowly until a response is obtained.
Operative Side
Orbicularis Oculi m.
Orbicularis Oris m.
Ground
Stim Return (+)
6
Most common in this group for facial nerve monitoring is the parotidectomy. The facial nerve enters the parotid gland as one bundle, then splits into five different branches within the body of the parotid gland. To monitor a given nerve branch, electrodes must be placed in the muscle that it innervates. The nerve is myelinated now, hence more stimulus is usually necessary to evoke a response.
In general, parotidectomies are “noisy surgeries” due to potential manipulation of the nerve during blunt dissection. Also, the electrodes are in close proximity to the surgical site.
Monitoring Goal
Locate, identify, and map the nerve and branches.
Monitor manipulation effect.
It is extremely valuable for the surgeon to verify nerve integrity prior to closing by stimulating proximal to the tumor site BEFORE and AFTER the surgery.
Parotidectomy
Submandibular Gland Dissection
Head and Neck Dissection
Congenital Aural Atresia
EXTRATEMPORAL
Procedures
7
EXTRATEMPORAL
Electrode Placement: Figure 3.
Procedure Specific Protocols
Suggested Electrodes Part No.
• Paired Subdermal Electrodes 8227410
• Prass Paired Electrodes, 18mm 8227304
Suggested Stimulator Probes Part No.
• Prass Standard Monopolar Probe 8225101
• Kartush Side-by-Side Bipolar Probe 8225401
Typical Electrode Readings Channel Values Difference Values
Subdermal needle < 10 KΩ < 1 KΩ Prass paired needle < 25 KΩ < 5 KΩ
Event Threshold Range
Event Threshold Range is 100uV-150uV.
Stimulation Range
Use 0.8mA at the beginning of the procedure (mapping) or as directed by the surgeon. Increase rapidly until a response is evoked. May need to increase to 3.0mA.
The best guideline for setting the stimulus intensity level is to use the lowest amount of stimulation needed to produce an EMG response large enough for monitoring. Be prepared to stimulate to as high as 3mA if necessary. Apply probe perpendicular to tissue for approximately one second to elicit a response.
Operative Side
Frontalis m.
Orbicularis Oculi m.
Orbicularis Oris m.
Mentalis m.
Ground
Stim Return (+)
8
Intraoperative monitoring of the Recurrent Laryngeal Nerve – branch of the Vagus (Cranial Nerve X). This surgery may be “noisy” due to potential manipulation of the nerve during blunt dissection. Also, the EMG tube is in close proximity to the surgical site which may produce some artifact.
Tips for Monitoring during Anterior Cervical Surgery
The EMG Endotracheal Tube should be utilized as for thyroidectomy. It may not be possible to directly stimulate the Recurrent Laryngeal Nerve as it may not be exposed. However, the NIM 2.0 System family will passively monitor traction and pressure status of the nerve and is especially helpful while the Thompson Retractor is being positioned and extended.
Monitoring Goal
Map the nerve, monitor manipulation effect, and verify nerve integrity.
It is extremely valuable for the surgeon to verify nerve integrity prior to moving to the contralateral lobe dissection and prior to closing. Stimulate proximally and distally to the tumor site.
Thyroidectomy
Parathyroidectomy
Radical Neck Dissection
Anterior Cervical Fusion
Substernal Goiter
Hemithyriodectomy
NECK DISSECTIONS
Procedures
9
NECK DISSECTIONS
NIM™ Standard Electrode Placement: Figure 4b.
Procedure Specific Protocols
Suggested Electrodes Part No.
NIM™ EMG Reinforced Endotracheal Tubes 8229XXX See catalog for choice of sizes.
• Usually use one size larger than normal.
• Use only non anaesthetic lubricants like K-Y®Jelly.
• Place EMG tube in the midline & with electrodes in contact with vocal cords. Secure tube to midline of mouth (not to the side). Temporarily deflate cuff for adjusting position. Complete instructions are provided in the package insert.
• Visualize electrodes contacting true vocal cords.
Suggested Stimulator Probes Part No.
• Prass Standard Monopolar Probe 8225101
• Kartush Side-by-Side Bipolar Probe 8225401
Typical Electrode Readings Channel Values Difference Values
EMG tube < 10 KΩ < 1 KΩ
Event Threshold Range
Event Threshold Range is 100uV-150uV.
Stimulation Range
The Stimulation Range is approximately 1.0mA. The best guideline for setting the stimulus intensity level is to use the lowest amount of stimulation needed to produce an EMG response large enough for monitoring. If stimulation levels are too high, nerve fibers of adjacent branches may be stimulated. Be prepared to stimulate at a higher level if necessary. Apply probe perpendicularly to tissue for about a second to elicit a response.
If there is no response, shunt stimulus in the area just below the Adam’s apple (cricothyroid), through the muscle. If there is response, the tube is probably too low through the vocal cords and will need to be repositioned.
NIM Contact™ Tube Electrode Placement: Figure 4.
MAINTAIN PATIENT’S EMG TUBE MIDLINE
Operative Side
Right Vocal Fold
Left Vocal Fold
MAINTAIN PATIENT’S EMG TUBE MIDLINE
Operative Side
Right Vocal Fold
Left Vocal Fold
Ground
Stim Return (+)
Ground
Stim Return (+)
10
Verifying Stimulus Delivery
The “Stimulus Tone” or “Stimulus Voice” can be heard if stimulus is flowing to the surgical site from the probe tip. Stimulus delivery can also be confirmed by comparing the Stimulus setting with the Stimulus Measure readings (mA). The stimulus measure reading will be found on the top right-hand side of the screen. The value should be approximately the same as the stimulus setting.
Tips for Reducing Artifact
• Keep the NIM 2.0 Systems away from the electrosurgical unit and other electrical equipment.
• Ensure electrode cables are not intertwined with stim anode return or with stimulator probe cable.
• Utilize a bipolar stimulating probe as soon as the nerve is exposed to minimize artifact. Current is flowing in a very small area and stim levels can be reduced accordingly. Also, dry the surgical area before stimulating, as fluid can shunt current. Keep bipolar tips dry to prevent shunting.
• Minimize excessive stimulating current. Each patient’s tissue conducts electrical energy differently. To decrease artifactual interference, it is better to establish effective levels of stimulating current as soon as possible to keep the current to a minimum. Begin stimulation with the lowest level,
0.05mA, with the Event Threshold at
100uV. Increase stimulus intensity fairly quickly until raw EMG sound is heard. This is the lowest level of stimulation which will evoke a response. Increase stimulus until an audible event tone occurs (the four-times-per-second tone we expect to hear when stimulating the nerve). Set the stimulation to at least this value.
• Note the reading next to the amplitude shown on the screen and adjust the Event Threshold slightly lower than that value. This will effectively screen out unwanted artifactual EMG activity.
• As an option in head and neck dissections, you may change the Stimulus Filter on the menu. For otology and skull base procedures, verify that the artifact delay marker is set to 3.1 milliseconds.
• If necessary, owing to frequent artifacts tones, disable the “Tone Audio”. Yo u
must now rely on raw EMG only.
MONITORING TIPS
11
MONITORING TIPS
Procedure Specific Protocols
Example of EMG Responses
Interpretation of events by the surgeon is enhanced by knowing the characteristics of the waveform, peak-to-peak amplitude, strength of the audio signal, and the surgical context.
Stimulated EMG Response
Cause: Electrical stimulation Sound: Precisely timed clicks, “machine gun”, four times per second
Mechanically-Evoked EMG Response
Cause: Direct surgical manipulation Sound: Few clicks synchronous with manipulation
12
MONITORING TIPS
Train Response
Cause: Traction, Pressure, Irrigation, Caloric (hot or cold) Sound: Repetitive asynchronous clicks, “popping corn”
Static Metal-to-Metal Artifact
Cause: Metal-to-metal (“banging of instruments”) Sound: “Pop”
References
This selection of reference clinical articles is provided for additional background material related to the monitoring procedures herein. The health care professional should seek and review all other clinical reference materials as dictated by an individual patient’s clinical condition.
1. Neuromonitoring in Otology and Head and Neck Surgery, Jack Kartush, MD and Kenneth R. Bouchard, Ph.D., Raven Press, New York, 1992
2. Handbook of Intraoperative Monitoring, Douglas L. Beck, M.A., Singular Publishing Group, Inc., 1994
3. Intraoperative Neurophysiologic Monitoring. Aage R. Miller, Harwood Academic Publishers, 1995
13
MONITORING TIPS
Facial Cranial Nerve VII - 2 Ch.
Facial Cranial Nerve VII - 4 Ch.
Extraocular Cranial Nerve III, IV, VI
Trigeminal Cranial Nerve V
Ground
Stim Return (+)
Operative Side
Glossopharyngeal m.
Glossopharyngeal Cranial Nerve IX
Spinal Accessory Cranial Nerve XI
Hypoglossal Cranial Nerve XII
Procedure Specific Protocols
Electrode Placement
The surgeon will insert electrodes into the appropriate muscle location innervated by the monitored nerve. Additionally, a ground electrode (green) and a stim return (white) are needed to complete the electrode setup.
Vagus Cranial Nerve X
-(VI N.)
-(IV N.)
-(III N.)
Operative Side
+(VI N.) Sup. oblique m.
+(IV N.) Lat. rectus m.
+(III N.) Inf. rectus m.
Ground
Stim Return (+)
Operative Side
Frontalis m.
Orbicularis Oculi m.
Orbicularis Oris m.
Mentalis m.
Ground
Stim Return (+)
Operative Side
Temporalis m.
Masseter m.
Ground
Stim Return (+)
Operative Side
Orbicularis Oculi m.
Orbicularis Oris m.
Ground
Stim Return (+)
Operative Side
Hypoglossal m.
Ground
Stim Return (+)
MAINTAIN PATIENT’S EMG TUBE MIDLINE
Operative Side
Right Vocal Fold
Left Vocal Fold
Ground
Stim Return (+)
MAINTAIN PATIENT’S EMG TUBE MIDLINE
Operative Side
Right Vocal Fold
Left Vocal Fold
Ground
Stim Return (+)
Operative Side
Sternocleidomastoid m.
Trapezius m.
Ground
Stim Return (+)
Symptom Cause Solution
TROUBLESHOOTING GUIDE
Please refer to the NIM-Response® 2.0 System User’s Guide (82-50651) or the NIM-Neuro® 2.0 System User’s Guide (82-50650) for complete operating instructions.
No visual display or audio alarms at power-up.
Touching the screen has unexpected results.
Electrode impedance is too high. >10K
Ω for subdermal electrodes
>25K
Ω for Prass Paired electrodes
>10K
Ω for EMG tube
>40K
Ω for hookwire electrodes
Electrode impedance = 0.0K
Ω.
Channel button is flashing. Electrode reading is “— K
Ω
or “OFF.”
Electrode difference is greater than 2KΩ (subdermal electrodes) or 10K
Ω (Prass Paired electrodes).
Electrosurgical interference.
Interference with anesthesia equipment.
Excessive muting.
Inadequate muting.
Power cord not connected to outlet or to the NIM™ 2.0 system.
Power switch not turned on.
Touch screen out of calibration.
Electrode dislodged from patient, but not completely out.
High resistance in electrode.
Electrode pin not firmly inserted into Patient Interface.
Positive and negative electrodes touching below surface of skin.
Extremely low impedance, particularly in EMG tubes.
Electrode laying on skin surface. Electrode placement insecure.
Dirty electrode tip. Electrode cable broken.
Electrode pin disconnected from Patient Interface.
Dirty electrode. Mismatched pair. Unequal placement.
Muting Probe not connected. Muting Probe input insufficient. Electrosurgical grounding inadequate. Source of interference unidentified.
NIM 2.0 system or Patient Interface cable too close to ESU or its cables.
Lead checking current near anesthesia elec­trodes.
Unit receiving excessive signal into the Muting Probe or electrode leads.
Signal from ESU is inadequate to cause muting.
Plug in power cord.
Turn power switch on.
Turn unit off, then press the screen until the screen calibration test is displayed. Follow the instructions on the screen to recalibrate.
Insert dislodged electrode; tape down in place.
Remove and replace with new electrode.
Check connection at Patient Interface box.
Remove and relocate electrodes.
Use “tap test” near electrodes to evoke EMG or artifact. If activity is noted on channel in question, proceed.
Re-insert electrode in question.
Remove and replace electrode in question.
Check connection to Patient Interface box.
Remove and replace electrode for appropriate channel with highest impedance reading first. Remove and replace electrode in question.
• Check Muting Probe connections.
• Move input to “MORE MUTE”.
• Check electrosurgical grounding pad on patient.
• Identify source of interference; then eliminate or separate from the NIM 2.0 system.
• Maintain separation between electrosurgical cable and the NIM 2.0 system.
• For less coupling, coil up the Muting Probe next to the NIM 2.0 system.
Have anesthesia try alternate electrode channel. Turn stimulator to 0.0mA when not stimulating.
Move the Muting Probe connector to a lower number unit until it stops muting. If it still mutes in position “1,” disconnect the muting detector completely.
Move the Muting Probe connector to a higher number until it mutes. If it still does not mute in position “4,” loop the ESU cable and clip the muting detector over the doubled cable.
TROUBLESHOOTING GUIDE
No response to direct stimulation.
Unexpected responses when not directly stimulating nerve.
Inadequate stimulus intensity.
Paralyzing anesthetic in use.
White stimulation (+) electrode has fallen out or is not connected.
Probe not connected.
Patient safety fuse blown.
Not holding probe on nerve long enough.
Nerve not contacted.
Volume control too low. Event threshold set too high.
Excessive current shunting in surgical field.
No electrodes in innervated muscle. Nerve not stimulable.
Unexplained continuous “train” EMG response.
Nerve or monitoring area being stimulated or manipulated by thermal or mechanical means.
Metal-to-metal discharge artifact.
Intertwined recording electrode and stimulator wires.
Inadvertent manipulation of electrode wires, Patient Interface cable, or recording area on patient.
Electrical interference from other equipment.
Increase stimulus intensity.
Eliminate paralyzing anesthetic.
• Check that Stimulus Measure is approximately the same value as the Stimulus setting. Re-insert electrode in question.
• Secure all recording and stimulating electrode connections and check the impedance values.
Check stimulator anode(+) and cathode (-) connections.
Check fuse in Patient Interface box (32 mA, x 250V). Replace if necessary.
Hold probe tip to nerve for at least 1 S.
• Check stimulator tip for obstruction. Replace if necessary.
• Check location of stimulation.
Check and correct all settings: volume, event threshold, stimulus intensity.
Remove fluids from surgical stimulating area.
Place channel electrodes in muscle to be monitored.
Identify and eliminate possible source of “train” stimulation:
• Cold irrigation
• Laser heat
• Retraction on nerve or muscles being recorded
• Patient waking from anesthesia
• Nerve drying
• Ultrasonic aspirator
Identify and eliminate source of inadvertent manipulation.
Determine response type from waveform pattern on 50 mS screen.
Disentangle recording electrode and stimulator cables.
Check area near recording electrodes for excessive stretching from tape, drapes, etc.
Check for intermittent stimulation from anesthesiologist (i.e., hand-held electrical stimulator).
Move NIM 2.0 system away from source of interference.
Make sure Patient Interface cable and electrode wires do not cross other electrical equipment or cables.
Symptom Cause Solution
Medtronic ENT Medtronic USA, Inc.
6743 Southpoint Drive North Jacksonville, FL 32216 USA www.MedtronicENT.com Toll free: (800) 874-5797 Fax: (800) 678-3995
890403 06.07 2007-341 Rev 1
This literature is intended for the exclusive use of physicians. Rx only. ©2007, Medtronic, Inc.
®
Registered mark of Medtronic, Inc. ™Trademark of Medtronic, Inc.
K-Y
®
is a registered mark of Johnson & Johnson Medical, Inc.
International Telephone Numbers
Australia: 1-800-668-670 Belgium: 32-2456-09-09 Canada: 1-800-217-1617 China: 86-21-50800998 France: 33-470-679-800 Germany: 49-211-5293-209 Hong Kong: 852-2919-1312 India: 91-22-26836733 Italy: 39-02-24137-324 Japan: 81-6-4795-1506
Korea: 82-2-3404-3600 Lebanon: 961-1-370-670 Luxembourg: 32-2456-09-09 Netherlands: 31-45-566-8371 Poland: 48-22-465-6942 Singapore: 65-6776-6255 Spain: 34-91-625-05-40 UK: 44-1923-205-166 USA: 1-904-296-9600
For further information, please call Medtronic ENT at 800-874-5797 or 904-296-9600. You may also consult our website at ww.MedtronicENT.com.
Loading...