Federal (USA) law restricts this device to sale by or on the order of a physician.
This instrument is to be used by authorized personnel only.
Device Tracking
(USA only, including US government-owned units) The Food and Drug Administration requires
defibrillator manufacturers and distributors to track the location of their defibrillators. If your defibrillator
has been sold, donated, lost, stolen, exported, or destroyed or if it was not obtained directly from
Medtronic Physio-Control Corp., please notify Physio at 1.800.442.1142, extension 4530.
Responsibility f or Information
It is the responsibility of our customers to ensure that the appropriate person(s) within their organization
have access to this information, including general safety information provided in Section 1.
European Union Contact
Medtronic Physio-Control
Medtronic Physio-Control
Medtronic Physio-ControlMedtronic Physio-Control
11811 Willows Road Northeast
Post Office Box 97006
Redmond, WA 98073-9706 USA
Telephone: 425.867.4000
Toll Free (USA only): 800.442.1142
Fax: 425.867.4121
Internet: www.physiocontrol.com
The following terms are used either in these operating instructions or on the Battery Support System 2
(BSS 2):
Danger: Immediate hazards that will result in serious personal injury or death.
Warning:Hazards or unsafe practices that could result in serious personal injury or death.
Caution: Hazardsor unsafe practices that could result in minor personal injury, product damage,or
property damage.
GENERAL WARNINGS AND CAUTIONS
The following are general warning and caution statements. Other specific warnings and cautions are
provided as needed in other sections of these operating instructions.
WARNINGS!
Shock or fire hazard.
Do not immerseany portion of this device in water or otherfluids. Avoid spilling any fluids on the device
or accessories. Do not clean with alcohol, ketones, or other flammable agents. Do not autoclave or
sterilize this device unless otherwise specified.
Possible fire or explosion.
Do not use this device in the presence of flammable gases or anesthetics. Place the BSS 2 in the
proper location as specified in these operating instructions.
Possible fire or explosion.
The BSS 2 is designed to be used with Physio-Control batteries only. Other manufacturers’ batteries
may overheat in this battery support system. Do not use with remanufactured or alternate source
batteries.
Possible loss of power during patient care.
Physio-Control has no information regarding the performance or effectiveness of its LIFEPAK
®
defibrillator/monitors if they are used with other manufacturers’ batteries or battery chargers. Using
other manufacturers’ batteries or battery chargers may result in device failure and may void warranty.
Use only Physio-Control batteries and the appropriate PHYSIO-CONTROL Battery Support System.
CAUTIONS!
Possible equipment damage.
This device may be damaged by mechanicalor physical abuse such as immersionin water or dropping.
If the device has been abused, remove it from use and contact a qualified service technician.
Possible or potential inaccurate battery charge level indicator.
Using the PHYSIO-CONTROL®Battery Support System (P/N 801807) or the two-well Battery Charger
(P/Ns 9-00284, 9-00288, and 801530) to charge and maintain a FASTPAK
®
2 battery will eventually
result in an inaccurate battery charge level indicator. Use only the BSS 2 (P/N 3010035)to charge and
maintain FASTPAK 2 batteries.
Possible equipment damage.
The supplied power cord is used to connect the battery support system to ac line voltage. Excessive
voltage may cause fuses in the BSS 2 to blow. Before connecting the battery support system, ensure
that you are using a grounded outlet of the correct voltage.
Note: The BSS 2 has no ON/O FF switch. When connecting the BSS 2 to ac line voltage, make sure
that the power cord is easily accessible and is not obstructed in any way. In case of device
malfunction or overheating, disconnect the device from the ac line voltage by removing the
power cord.
1-2Batt ery Support Syste m 2 Operatin g Instruction s
Page 9
Safety Information
SYMBOLS AND ABBREVIATIONS
The symbols and abbreviations below may be found inthese operating instructions, on the BSS 2, or on
FASTPAK and LIFEPAK batteries:
Attention, consult accompanying documents
RBRC
RBRC
Ni-Cd
Recycle NiCd battery
Recycle PB battery
LOT
or
YYWW
YYYY
Recycle this product
Rechargeable battery: recycle battery
Direct Current (DC)
Alternating Current (AC)
Positive terminal
Negative terminal
Fuse
Type B, isolation (no patient connection)
Indoor use only
Manufacturing date (batteries)
Manufacturing date (BSS 2)
NRTL/C
Canadian Standards Association certification for United States (Nationally
Recognized Test Laboratory) and Canada
Marking of conformity according to the Medical Device Directive 93/42/EEC
Hz
V
A
W
Battery Support System 2 Operating Instructions1-3
Slow Blow fuse
Light Emitting Diode
Nickel Cadmium chemistry battery
Part Number
Serial Number
Sealed Lead Acid chemistry battery
DOCUMENT CONVENTIONS
Text references to device buttons and labels are indicated in capital letters, for example:
CHARGE.
1-4Batt ery Support Syste m 2 Operatin g Instruction s
Page 11
2
BASIC ORIENTATION
This section provides a basic orientation to the PHYSIO-CONTROL
Battery Support System 2.
PHYSIO-CONTROL Battery Support System 2page 2-2
The FASTPAK and LIFEPAK Family of Batteries2-2
Unpacking and Inspecting2-3
Controls, Indicators, and Connectors2-3
Inserting Batteries2-4
®
Battery Support System 2 Operating Instructions2-1
The PHYSIO-CONTROL Battery Support System 2 (BSS 2) provides charging, conditioning, and shelf
life testing for the Physio-Control FASTPAK and LIFEPAK batteries.
The BSS 2:
• Provides a clear indication that a battery is charging, ready for use, or needs to be discarded.
• Can charge, condition, or shelf life test up to three FASTPAK and LIFEPAK SLA batteries at one time.
• Is provided with a d etachable power cord that connects to a grounded ac outlet.
• Has a wall mount bracket (optional) that permits horizontal or vertical mounting.
THE FASTPAK AND LIFEPAK FAMILY OF BATTERIES
WARNING!
Possible explosion.
Attempting to charge a LIFEPAK 5 battery in the BSS 2 may cause the battery to explode. Do not
charge LIFEPAK 5 batteries in the BSS 2.
All the batteries shown in Figure 2-1 can be used in the LIFEPAK®12 defibrillator/monitor series, but
only the FASTPAK batteries can be used in other LIFEPAK defibrillator/monitors. The LIFEPAK
batteries can only be used in the LIFEPAK 12 defibrillator/monitor.
FASTPAK, FASTPAK 2, and LIFEPAK NiCd batteries have a nickel c admium (NiCd) chemistry and
LIFEPAK SLA batteries have a sealed lead acid (SLA) chemistry. In addition, the FASTPAK 2 and
LIFEPAK NiCd batteries have a charge level indicator (fuel gauge).
Charge FASTPAK batteries in either the Battery Support System (BSS) (P/N 801807), the BSS 2
(P/N 3010035), or the LIFEPAK 12 defibrillator when it is powered by the AC or DC Power Adapter.
FASTPAK batteries require periodic conditioning and shelf life testing to optimize performance.
Condition and shelf life test FASTPAK batteries in the BSS or BSS 2.
Charge the FASTPAK 2 battery in the BSS 2 or in the LIFEPAK 12 defibrillator when it is powered by
the AC or DC Power Adapter. Using the BSS (P/N 801807) or the two-well Battery Charger (P/Ns 900284, 9-00288, and 801530) to charge and maintain a FASTPAK 2 battery will eventually result in an
inaccurate battery charge level indicator. FASTPAK 2 batteries require periodic c onditioning and shelf
life testing to optimize performance. Condition and shelf life test FASTPAK 2 batteries only in the
BSS 2.
Charge LIFEPAK NiCd and SLA batteries either in the BSS 2 or in the LIFEPAK 12 defibrillator w hen it
is powered by the AC or DC Power Adapter. LIFEPAK NiCd batteries require periodic conditioning and
shelf life testing to optimize performance. Periodic conditioning and shelf life testing may optimize SLA
battery performance. Condition and shelf life test LIFEPAK NiCd and SLA batteries in the BSS 2.
2-2Batt ery Support Syste m 2 Operatin g Instruction s
Page 13
Basic Orientation
UNPACKING AND INSPECTING
Remove the BSS 2, ac power cord, and product literature from the shipping container. Carefully inspect
the BSS 2 and the power cord for any signs of damage during shipping. Verify the receipt of everything
noted in the packing list. Save the shipping container and the foam inserts in case the BSS 2 needs to
be returned.
CONTROLS, INDICATORS, AND CO NNECTORS
Figure 2-2, Figure 2-3, and Table 2-1 provide an overview of the controls, indicators, and connectors for
the BSS 2 and FASTPAK 2 batteries.
9
8
7
6
5
4
3
Figure 2-2
Table 2-1 Controls, indicators, and connectors
1
2
3
4
5
6
Controls, indicators, and connectors
SERVICE
POWER
SHELF LIFE
CONDITION
CHARGE
DISCARD
Indicates BSS 2 needs service.
Indicates power is on.
Amber LED indicates battery is undergoing a shelf life test.
Amber LED indicates battery is conditioning.
Amber LED indicates battery is charging.
Indicates battery should be removed from use and discarded/
recycled.
READY
7
Indicates battery is ready for use.
12
10, 11
1
2
8+ /– TerminalsTransfer charge current onto battery.
Battery Support System 2 Operating Instructions2-3
9Communication PinsProvide digital communication between the BSS 2 and the battery.
10AC Power Input
Receptacle for ac power cord.
Receptacle
11FusesTwo fuses help protect the BSS 2 from current surges.
12Power CordCable for operation from an ac outlet.
14
Figure 2-3
Battery fuel gauge
13
13Fuel Gauge ButtonLights the fuel gauge indicators.
14Fuel Gauge
Indicators
One to four lights indicate the relative charge of the battery,
with four lights indicating maximum charge.
One light flashing: charge battery.
Two alternately flashing lights: battery requires conditioning.
No lights: battery has 0% charge or needs to be discarded.
INSERTING BATTERIES
CAUTION!
Failure to charge batteries.
Battery pins in the battery support system may be damaged if the batteries are dropped or forced into
the battery wells. Inspect the battery pins routinely for signs of damage.
Routinely inspect batteries for damage or leakage. Discard/recycle damaged or leaking batteries. To
insert a battery into a battery well:
1 Inspect the +/- terminals (pins) in the battery well for signs of damage.
2 Inspect the battery well for signs of damage or distortion.
3 Align the battery so that the battery clip is over the pins in the battery well.
4 Insert the end of the battery opposite the battery clip into the battery well.
5 Firmly press the other end into the battery well until it clicks into place.
To remove the battery, press the battery clip in and lift the battery out of the battery well.
2-4Batt ery Support Syste m 2 Operatin g Instruction s
Page 15
3
BATTERY MAINTENANCE
This section describes how to use the PHYSIO-CONTROL Battery
Support System 2 to maintain FASTPAK and LIFEPAK batteries.
Battery Maintenance Warningpage 3-2
Battery Maintenance Guidelines and Procedures3-2
Receiving New Batteries3-4
Storing Batteries3-4
Discarding/Recycling Batteries3-4
Battery Support System 2 Operating Instructions3-1
Possible loss of power and delay of therapy during patient care.
Using an improperly maintained battery to power the defibrillator may cause power failure without
warning. Use the appropriate PHYSIO-CONTROL Battery Support System to charge and condition
batteries.
BATTERY MAINTENANCE GUI DELINES AND PROCEDURES
To maximize BSS 2 operation and battery performance and life, observe these guidelines:
• Place the BSS 2 in the proper location
• Charge batteries at room temperature
• Rotate batteries
• Condition batteries every three m onths
• Perform shelf life tests every six months (optional for LIFEPAK SLA)
Place the BSS 2 in the Proper Location
To help ensure proper BSS 2 functioning:
• Place in a dry, well-ventilated area
• Keep at room temperature
• Keep batteries in wells
• Do not place in direct sunlight
• Do not place near a heat s ource or an air conditioner
• Keep fan vent on back panel unobstructed
Charge Batteries at Room Temperature
Charging batteries at room temperature, 20°C to 25°C (68°F to 78°F), is preferred to maximize battery
performance and life. The extreme temperature range for charging batteries is 5°C to 35°C
(41°F to 95°F). Batteries charged outside this temperature range may not reach a full charge (even if
charging time is increased) and irreversible cell damage could occur.
CAUTION!
Possible battery damage.
Charging a battery at temperaturesbelow 5°C (41°F) or above 35°C (95°F) prevents the battery from
reaching full capacity and may cause irreversible cell damage.
Fully charge SLA batteries between uses. If SLA batteries are not 100% recharged between uses,
sulfation (lead sulfate buildup on electrode surfaces inside the battery) can occur. Sulfation reduces
battery capacity and may result in premature battery failure.
To charge a battery:
1 Insert a battery into a BSS 2 battery well.
2
lights when a battery reaches full charge. The battery may remain in the battery well until
READY
CHARGE
needed. The BSS 2 supplies a trickle charge thatprevents overcharging and maintainsthe battery at
peak capacity.
Note:
CONDITION
LED lightsand the conditioning proced ure begins automaticallyif a FASTPAK 2,
LIFEPAK NiCd, or LIFEPAK SL A battery requires conditioning.
Note: Withdrawal of the battery from the charger prior to a
inadequate charge.
Note: If
DISCARD
lights, remove the battery from use and discard/recycle (see page 3-4).
LED lights and charging begins automatically.
indication may result in
READY
3-2Batt ery Support Syste m 2 Operatin g Instruction s
Page 17
Battery Maintenance
Rotate Batteries
Rotate all batteries in active use so that they are used with equal frequency.
Condition Batteries Every Three Months
Voltage depression is a condition that reduces battery performance. When NiCd batteries repeatedly
receive a shallow discharge (that is, not allowed to drain completely between charging cycles), voltage
depression occurs. This condition is often mistakenly called “memory.” Voltage depression can usually
be reversed by conditioning the battery every three months.
Conditioning is a series of charge/deep discharge cycles performed in the BSS 2 to measure and
optimize battery capacity.
To condition a battery:
1 Insert the battery into any battery well.
CONDITION
LIFEPAK NiCd, or LIFEPAK SLA battery requires conditioning. Allow the FASTPAK 2 or LIFEPAK
batteries to complete the condition cycle.
2 Press
3
READY
CONDITION
lights when a battery is fully recharged. The BSS 2 automatically recharges a battery that
for other FASTPAK batteries.
passes the conditioning process.
LED lights automatically when a FASTPAK 2,
Note: If
DISCARD
lights, remove the battery from use and discard/recycle.
Note: If a power failure occurs during battery conditioning, the BSS 2 interrupts conditioning and
reverts to
CHARGE
mode once power is restored. Battery conditioning may not have been
completed. Repeat conditioning process.
Perform Shelf Life Testing Every Six Months
Batteries self-discharge when not in use. A new NiCd battery self-discharges approximately 10% of its
capacity in the first dayand 1% of its capacity every day thereafter when stored at room temperature. In
10 days, a new NiCd battery loses approximately 20% of its capacity.
SLA batteries have alow self-discharge rate. A new SLA battery self-discharges approximately 0.1% of
its capacity each day when stored at room temperature. In 10 days, a new SLA battery loses
approximately 1.0% of its capacity.
The actual rate of battery self-discharge depends on:
• Battery age
• Temperature
• Frequency of use
• Length of time in storage
• Physical condition
These factors can combine to significantly increase the battery discharge rate. For example, an older
NiCd battery stored at higher temperatures may have an accelerated self-discharge rate much greater
than 1% a day.
The self-discharge rate increases as the battery ages.
The shelf life test evaluates the self-discharge rate of a battery. Any battery that fails the shelf life test
(
DISCARD
To perform a shelf life test:
1 Complete the conditioning procedure.
2 Remove the battery from the BSS 2 and store for 7 days at room temperature.
3 After storage, insert the battery into a battery well and press
4 Verify that
lights) should be removed from use and discarded/recycled.
SHELF LIFE
SHELF LIFE
LED lights.
within 3 seconds.
Battery Support System 2 Operating Instructions3-3
lights when the battery passes the shelf life test and is fully recharged. The battery may then
READY
be returned to use.
Note: If
DISCARD
lights, remove the battery from use and discard/recycle.
Note: If a power failure occurs during shelf life test, the BSS 2 interrupts theshelf life testand reverts
to
CHARGE
mode onc e power is restored. Shelf life test may not have been completed. Repeat the
shelf life test.
RECEIVING NEW BATTERIES
When newly-purchased batteries are received, charge each new battery. Because batteries selfdischarge during storage, a new battery may not be fully charged when it is received.
STORING BATTERIES
WARNING!
Possible loss of power during patient care.
Stored batteries lose charge. Failure to charge a stored battery before use may cause device power
failure without warning. Always charge a stored battery before placing it in active use.
Store batteries in or out of the BSS 2 except when performing the shelf life test. During storage,
batteries still require routine maintenance (refer to the conditioning procedure, page 3-3, and the shelf
life test, page 3-3).
Storing an SLA battery that is less than 100% charged can result in permanent damage.
When storing batteries:
• Fully charge LIFEPAK SLA batteries before storing.
• Store batteries between 4.4°C and 26.7°C (40°F to 80°F). Lower temperatures reduce the battery
self-discharge rate. Higher temperatures increase the self-discharge rate.
• Do not freeze batteries. Damage to the battery may result.
• Charge stored batteries before placing in use.
DISCARDING/RECYCLING BATTERIES
Properly maintained FASTPAK batteries have a useful life of approximately two years. Properly
maintained FASTPAK 2 and LIFEPAK NiCd batteries have a useful life up to five years. Properly
maintained LIFEPAK SLA batteries have a useful life of up to three years. A FASTPAK, FASTPAK 2,
LIFEPAK NiCd, or LIFEPAK SLA battery has reached the end of useful life if one or more of the
following circumstances occur:
• Battery fails conditioning or shelf life test.
• There is physical damage to the battery case.
• The battery is leaking.
• The BSS 2 indicates
DISCARD
Note: The FASTPAK 2, LIFEPAK NiCd, and LIFEPAK SLA batteries have internal parameters that
establish limits for useful life. If these parameters are exceeded , the BSS 2 will indicate
when battery is inserted into the battery well.
3-4Batt ery Support Syste m 2 Operatin g Instruction s
during any battery maintenance procedure.
DISCARD
Page 19
Battery Maintenance
To promote awareness of battery recycling, Physio-Control FASTPAK and LIFEPAK batteries are
marked with one of these symbols:
RBRC
RBRC
Ni-Cd
When a FASTPAK or LIFEPAK battery reaches the end of its useful life, recycle the battery as follows:
Battery Recycling in the USA
Recycle batteries by participating with Medtronic Physio-Control in a national batteryrecycling program.
Contact your local Physio representative to obtain shippinginstructions and shipping containers. Do not
return your batteries to the Medtronic Physio-Control offices in Redmond, Washington, unless
instructed to do so by your Physio representative.
Battery Recycling Outside the USA
Recycle batteries according to national and local regulations. Contact your local Physio representative
for assistance.
Battery Support System 2 Operating Instructions3-5
This section describes the PHYSIO-CONTROL Battery Support
System 2 maintenance.
The BSS 2 Service Indicatorpage 4-2
Cleaning4-2
Fuse Replacement4-2
Troubleshooting4-3
Replacement Parts and Accessories4-4
Specifications4-4
Warranty Policy4-6
Product Recycling I nformation4-6
Battery Support System 2 Operating Instructions4-1
Do not disassemble the battery support system. It contains no operator-serviceable components and
dangerous high voltages may be present. Contact a qualified service technician for repair.
When the BSS 2 ac power cord is inserted into an ac power receptacle, the BSS 2 performs a series of
self-diagnostic tests. Power-up testing is indicated by all the BSS 2 indicator lights briefly illuminating. If
self-testing is successful all the indicators (except
self-test,
SERVICE
lights. Remove the BSS 2 from use and contact a qualified service technician for
repair.
Periodically inspect the BSS 2 for damage or cracks. Check that it is completely closed. Inspect the
power cord for damage, cracks, or bent pins.
The BSS 2 does not contain any operator-serviceable components. If the BSS 2 requires service,
contact a qualified service technician. When calling Medtronic Physio-Control to request service, have
the following information available: model number, serial number, and a description of the problem. If
the device must be sent to a service center or factory, repack it in the original shipping container to
prevent shipping damage.
Test modes, service protocols, fault codes, and other maintenance procedures are described in the
Battery Support System 2 Service Manual. In the USA, call 1.800.442.1142 for technical consultation,
manuals, or parts. Outside the USA, contact your local Medtronic Physio-Control service office.
) extinguish. If the BSS 2 fails any power-up
POWER
CLEANING
Clean the BSS 2 case, battery contacts, batteries, and ac power cord with a damp spongeor towel. The
following agents may be used:
• Mild soap and water
• Quaternary ammonium compounds
• Isopropyl alcohol
• Peracetic (peroxide) acid solutions
CAUTION!
Possible equipment damage.
Do not clean any part of the battery support system or ac power cord with bleach, bleach dilution,
ketones, solvents, flammable agents,
autoclave, or gas-sterilize the battery support system.
phenolic compounds, or abrasive agents. Do not steam,
FUSE REPLACEMENT
The BSS 2 has two fuses that help provide protection against over current. To replace the fuses, open
the fuse carrier door (located in the power input module) with a flat-bladed screw driver. Verify that the
replacement fuses are the same type and rating as listed in the specifications in Table 4-3.
When replacing fuses in the BSS2, useonly 5 x 20mm T 250V fuses approvedto IEC 127-2, sheet 1 or
2, such as:
• Schurter FST
• Bussman GDC
• Littelfuse 218
Note: Use of other fuses may cause premature failure of mains fuse.
4-2Batt ery Support Syste m 2 Operatin g Instruction s
Page 22
BSS 2 Maintenance
TROUBLESHOOTING
Refer to the troubleshooting tips in Table 4-1. If these troubleshooting tips do not resolve your question,
remove the BSS 2 from active use and contact a qualified service technician for service and repair. To
contact Medtronic Physio-Control in the USA, call 1.800.442.1142. Outside the USA, contact your local
Medtronic Physio-Control sales or service office.
Table 4-1 Troubleshooting tips for the BSS 2
ObservationPossible CauseCorrective Action
POWER
ac power cord is inserted into an
ac power receptacle.
CHARGE/CONDITION
light when a FASTPAK battery is
installed.
READY
of charging.
DISCARD
Battery powers devices for less
than the expected time.
SHELF LIFE
does not light whenthe
Power cord not properly
connected.
Make sure that the power cord is
securely connected at both ends.
Blown fuses.Replace fuses using the
proceduredescribedonpage 4-2
with the type of fuses specified in
Table 4-3.
does not
Inoperative ac outlet or tripped
circuit breaker in building.
Battery malfunction.Install a differentbattery in the
Remove from use and contact a
qualified service technician.
same battery well. If
CONDITION
lights, discard the
CHARGE/
original battery.
BSS 2 malfunction.Remove from use and contact a
qualified service technician.
does not light after hours
Battery does not charge to full
capacity (battery requires
Perform the conditioning
procedure (page 3-3).
conditioning).
Battery charged at excessively
high or low temperatures.
Perform the conditioning
procedure (page 3-3) at the
correct temperature.
illuminates.Battery is at end of useful life.Remove from use and discard/
recycle.
Battery needs conditioning.Perform the conditioning
procedure (page 3-3). If the
battery is still not performing to
expectations, perform a shelf life
test (page 3-3).
LED does not light.Conditioned battery was not
removedand stored prior to shelf
life test.
Condition, remove, and store
battery for 7 days prior to shelf
life testing.
Battery Support System 2 Operating Instructions4-3
In the USA, call 1.800.442.1142 to order parts and accessories. Outside the USA, contact your local
Medtronic Physio-Control sales or service office.
Table 4-2 Replacement Parts and Accessories
DescriptionPart Number
Battery Support System 2 Operating Instructions3010515
Battery Support System 2 Service Manual3011409
FASTPAK 2 Battery3009375
FASTPAK Battery (white)9-10424-18
FASTPAK Battery (gold)9-10424-19
LIFEPAK NiCd Battery3009376
LIFEPAK SLA Battery3009378
AC Input Power Cord803650
BSS 2 Wall Mount Bracket3010932
SPECIFICATIONS
Table 4-3 lists the specifications for the Battery Support System 2.
Table 4-3 BSS 2 specifications
FunctionDescription
General Specifications:
Dimensions31.5 cm x 37.6 cm x 9.7 cm (12.4 inches x 14.8 inches x 3.8 inches)
Weight8.2kg (< 18 pounds) (excluding batteries)
Number of battery wells3. Charging, conditioning, and shelf life test capabilityin all battery
Power requirements:
Military
FusesTwo fuses in the power input module (5 x 20 mm T 250V, Low or High
Environmental Specifications:
Liquid IngressIPX0 per IEC529. Not protected againstingression of fluids. Indoor
Table 4-4 FASTPAK and LIFEPAK Battery Spec ificat ions
FASTPAK and FASTPAK 2
Battery Type
Weight
Voltage
Capacity
Charge Time (with fully depleted battery)
Conditioning Time
Charging Temperature Range
Operating Temperature Range
Nickel cadmium
0.7kg(1.5lbs)
12Vdc
1.2 amp hours
1.5 hours
7 hours typical, 8 hours maximum
+5° to 35°C (41° to 95°F)
0° to 50°C (32° to 122°F)
Long Term (>1 day) Storage Temperature
Range
0° to 35°C (32° to 95°F)
LIFEPAK NiCd
Battery Type
Weight
Voltage
Nickel cadmium
0.77 kg (1.7 lbs)
12Vdc
*
Capacity
Charge Time (fully depleted battery)
Conditioning Time
Charging Temperature Range
Operating Temperature Range
Long Term (>1 day) Storage Temperature
Range
LIFEPAK SLA
Battery Type
Weight
Voltage
Capacity
Charge Time (fully depleted battery)
Conditioning Time
Charging Temperature Range
1.7 amp hours
2.25 hours
8 hours typical, 10 hours maximum
+5° to 35°C (41° to 95°F)
0° to 50°C (32° to 122°F)
0° to 35°C (32° to 95°F)
Sealed lead acid
1.4kg(3.0lbs)
12Vdc
2.5 amp hours
6 hours typical, 12 hours maximum
28 hours typical, 56 hours maximum
+5° to 35°C (41° to 95°F)
Battery Support System 2 Operating Instructions4-5
Operating Temperature Range
Long Term (>1 day) Storage Temperature
Range
All specifications are at 20°C unless otherwise stated.
*
0° to 50°C (32° to 122°F)
0° to 35°C (32° to 95°F)
WARRANTY POLICY
The BSS 2 is warranted against all defects in materials and workmanship for a period of one year from
the date of delivery.
All batteries supplied by M edtronic Physio-Control for LIFEPAK defibrillator/monitor products are
warranted for a period of one year. If Physio receives notice of a battery defect during the warranty
period, Physio will replace the battery upon verification of the defect.
Use of other manufacturers’ batteries and accessories with Medtronic Physio-Control defibrillator/
monitor products may void Safety Agency Certifications and warranty.
PRODUCT RECYCLING INFORMATION
Recycle the device at the end of its useful life.
Recycling Assistance
The device should be recycled according to national and local regulations. Contact your local
Physio representative for assistance.
Preparation
The device should be clean and contaminant-free prior to being recycled.
Recycling of Disposable Electrodes
After using disposable electrodes, follow your local clinical procedures for recycling.
Packaging
Packaging should be recycled according to national and local regulations.
4-6Batt ery Support Syste m 2 Operatin g Instruction s
Page 26