The information contained in this document was accurate at time of publication. Medtronic Xomed reserves the right to
make changes in the product described in this manual without notice and without incorporating those changes in any
products already sold.
TABLE OF CONTENTS
Symbols ............................................................................................................................................ iii
When the System Arrives ................................................................................................................. iv
Warnings and Precautions ...........................................................................................................iv-vi
Customer Service............................................................................................................................. vii
A TTENTION, SEE INSTRUCTIONS FOR USE
CA T ALOG NUMBER
SERIAL NUMBER
DA TE OF MANUF ACTURE
USE BY DA TE
DO NOT REUSE
STERILIZED BY ETHYLENE OXIDE.
DO NOT USE IF P ACKAGE IS OPENED OR DAMAGED.
STERILIZED BY RADIA TION.
DO NOT USE IF P ACKAGE IS OPENED OR DAMAGED.
LOT NUMBER
ON (MAIN POWER)
OFF (MAIN POWER)
REVERSE
FOR WARD
ACC
OSCILLATION
FOOTSWITCH — V ARIABLE MODE
FOOTSWITCH — ST AR T/STOP
FOOTSWITCH
FOOTSWITCH DIRECTION - FOR W ARD
FOOTSWITCH DIRECTION - OSCILLA TE
FOOTSWITCH DIRECTION - REVERSE
TYPE B APPLIED P AR T
ACCESSOR Y CONNECTOR
MANUAL ST AR T/STOP
HANDPIECE
SKEETER® HANDPIECE
RF TRANSMITTER (INTERFERENCE MA Y OCCUR)
MENU
MENU
IRRIGANT
COOLANT
PUMP
LOCKED
UNLOCKED
CAUTION: PINCH HAZARD. KEEP FINGERS CLEAR OF ROLLERS.
iii
WHEN THE SYSTEM ARRIVES
UNPACKING AND INSPECTION
As the box is unpacked, check off the contents of the box against the items listed on the packing slip. If the contents
are incomplete or if there is damage, notify Customer Service. If the shipping container is damaged, or the cushioning
material shows signs of stress, notify the carrier as well.
Keep the shipping materials for carrier inspection.
After unpacking your new XPS® Model 3000 System, save the cartons and packing material. If the instrument is to
be shipped from one location to another, the carefully designed shipping package will provide proper protection.
WARNINGS AND PRECAUTIONS
It is important that the XPS® Model 3000 operator be familiar with this manual: its precautions, procedures and
safety issues. Three labels are used in this manual to identify important concerns, conditions, or procedures:
“WARNING”
Identifies conditions or practices that present a risk of injury to the patient and/or user.
"PRECAUTION / CAUTION"
Identifies conditions or practices that could result in damage to the equipment.
"NOTE"
Identifies special information allowing easier maintenance of equipment or to clarify or emphasize important
instructions.
WARNINGS
• Always inspect the components before and after use for any damage. If damage is observed, do not use damaged
part until it is replaced. Damaged parts may deposit metal shavings on surgical site.
• Do not modify accessories used with the handpiece. Performance could be diminished with modified accessories.
• Insertion of metal objects in blade or bur tip may cause the blade or bur to break leaving fragments in the wound.
The fragments may be difficult to remove, causing irritation, inflammation and foreign-body response at surgical
site.
• Do not use any parts other than Medtronic Xomed, Inc. system components as damage or substandard performance
could result.
• Blade and bur accessories are available for resection of soft tissue and bone for surgical procedures. Use of
accessories depends on the intended application and patient needs. Sharp-cutting powered accessories induce
bleeding and removal of significant tissue and bone.
• Do not attach unapproved components to the XPS® 3000 Console to avoid electrical macro shock.
• Carefully inspect burs prior to and following each use for excessive wear, fragmentation, eccentricities or other
defects. Do not use dull, damanged or bent burs. Use of dull burs can reduce the handpiece effectiveness and
cause the handpiece temperature to increase.
• Do not attempt to resharpen used burs. Worn burs should be replaced with new ones frequently to ensure
effective cutting and control of the drill.
• Excessive pressure applied to bur may cause bur fracture. Should a bur fracture during use, extreme care must be
exercised to ensure that all fragments of the bur are retrieved and removed from the patient. Unremoved bur
fragments may cause tissue damage to the patient.
• T est for bur wobble (eccentricity) at desired speed prior to use. Use a bur guard (Xcalibur only) if needed. Select
a new bur or reduce speed if wobble is observed.
• T est for bur wobble (eccentricity) at the desired speed prior to use. Select a new bur or reduce speed if wobble is
observed. Eccentricity of the bur can cause bur vibration and may result in excessive tissue and bone destruction
and hearing damage. Always examine the operation of each bur in the handpiece before each use. Operating
certain burs at high speed can cause vibration of the bur.
• Keep the cooling sleeve irrigated to prevent thermal injury to tissue.
• During procedures near nerves, keep bur and bur cannula away from tissue to minimize the potential for thermal
injury .
iv
WARNINGS (continued)
• For procedures near nerves, nerve monitoring should be used to alert the user of the potential for injury.
• Use care in application of the moving cutting end to only appropriate anatomical landmarks and the intended
surgical site when using XPS® accessories. The use of powered reciprocating/rotating instruments may result in
vibration-related injury. Use appropriate precautions.
®
• Employ visualization when using rotating XPS
accessories. Discontinue powered application in the event of
lack of visualization of the surgical site.
• Use methods at the operative site to control bleeding that do not compromise patient safety during at-risk
surgery.
• Use lock on StraightShot
®
M4 handpiece to prevent inadvertant rotation of blade or bur during use.
• When precise location of the blade tip is required, engage the rotation lock on the handpiece, then calibrate and
verify the blade tip on the Image Guided Surgery (IGS) system. Always lock the StraightShot® M4 handpiece
when driving non-rotatable blades to maintain their IGS calibration.
• Always keep the cutting tip of the accessory away from fingers and loose clothing. Prevent laceration of user and
cross-contamination through compromised glove.
• Do not change accessory with handpiece running to prevent laceration of user and cross-contamination through
compromised glove.
• Bending or prying may break the blade or bur, causing harm to patient or staf f.
• Discontinue use of curved bur if tip begins to wobble and replace bur to prevent unintended tissue removal from
patient.
• Do not use burs above the speed indicated on the bur label. Exceeding speed may cause the burs to break.
• Disposable devices are for single-use only. Do not attempt to sterilize disposable devices. The disposables are
packed sterile and not intended for repeat use. To prevent contamination use only once.
• Disconnect power to the XPS® 3000 Console before cleaning the unit to avoid electrical macro shock.
• After each procedure, properly clean all reusable system components.
• Sterilize and dry reusable device before storing the system. Decrease likelihood of cross-contamination with
timely sterilization.
• Remove and discard accessories following local regulations for proper disposal of contaminated materials.
• Properly dispose of single-use devices removed from sterile packages. Devices lost sterility upon removal from
packaging.
• All service must be performed by Medtronic Xomed-qualified personnel only .
WARNINGS BEFORE SURGERY
• Verify reusable device was sterilized prior to use. If not sterilized, do not use.
• Do not use accessory if package is opened or damaged. Broken seal offers no protection against crosscontamination.
• Do not operate the XPS® 3000 System in the presence of flammable anesthetics. Avoid potential ignition or
explosion of gases.
• Achieve electrical grounding reliability with proper connections. Connect the XPS® 3000 Console to hospital
grade receptacles only.
• Use adequate irrigation from a separate user-provided irrigating source. The use of a bur without irrigation may
cause an inordinate amount of heat buildup resulting in thermal injury to tissue.
• When not operating handpiece, ensure that handpiece rests on a non-conductive surface that provides containment
for handpiece and blade/bur. Avoid unintended thermal injury by an uncontained handpiece.
• When not operating handpiece, eliminate accidental foot control activation. Control energy to and through
handpiece to prevent unintended tissue, bone or nerve resection.
• This medical device complies with EN60601-1-2 safety standard for electromagnetic compatibility , requirements
and test. However, if this equipment is operated in the presence of high levels of electromagnetic interference
(EMI) or highly sensitive equipment, interference may be encountered and the user should take whatever steps
are necessary to eliminate or reduce the source of the interference. Diminished performance may lengthen operating
time for anesthetized patient.
PRECAUTIONS
• XPS® Blades should be operated in the oscillate mode only . Operating in the forward mode may cause damage to
the blade. XPS® burs should be operated in the forward mode only.
• The handpiece should be evaluated prior to each use for suitable operating condition.
• System components should be operated and inspected for damage prior to use. Do not use system if damage is
apparent.
v
PRECAUTIONS (continued)
• It is recommended that a secondary handpiece be available to minimize any downtime and inconvenience to the
surgical staff.
• The StraightShot® Magnum® II and StraightShot® M4 Microdebriders are intended to operate at speeds greater
than 6,000 rpm ONL Y when used with the XPS® High Speed bur line.
• Always ensure that the bur is securely engaged into the handpiece prior to operating the system.
• When operating or testing the Poweforma®, Xcalibur® and Visao™ drill handpieces, ensure the bur is properly
inserted and locked into the handpiece. Running the drill with the collet unlocked can damage the locking
mechanism.
• Improper priming of the Xcalibur® and Visao™ Hi-Speed Drills will result in excessive handpiece temperature.
• The Xcalibur® and V isao™ High-Speed W ater-Cooled Drills work only with the XPS® 3000. Use without water-bag
will damage the motor in the handpiece.
• Do not allow the console or footswitch to get wet. If liquid enters the console or footswitch, damage could occur.
• To prevent damage to curved blades and burs, disconnect suction tube prior to changing blade or bur during
procedure.
• DO NOT clean handpieces in ultrasonic cleaner or allow them to be fully immersed in any soaking solution. These
procedures may result in damaging the handpieces beyond repair.
• Do not sterilize the console or footswitch.
• Do not clean the handpiece in an ultrasonic cleaner or cold soak sterilize the handpiece in glutaraldehyde. Do not
immerse handpieces, motors, or handpiece cables in any solution, except as detailed in the Cleaning section of
this manual. These procedures void the warranty and may damage the handpiece beyond repair.
• Always handle handpieces with care to avoid damage.
• Do not use organic solvents such as acteone or isopropyl alcohol to clean the bur chuck. Use only an enzymatic
detergent and distilled water for cleaning after every case or as required.
• Remove the bur from the handpiece before sterilization.
• Sterilize immediately after cleaning. Do not store unless a drying cycle has been performed.
• Always wrap the electrical cord carefully inside the sterilization tray during sterilization to prevent damage to the
cord when closing the tray lid.
• Remove the handpiece from the sterilizer immediately after the sterilization cycle is complete.
• Temperatures higher than those stated may be used for handpiece sterilization when necessary to satisfy
governmental or health care facility requirements so long as the temperature does not exceed 149°C (300ºF).
Heating above 149°C (300ºF) may damage components and will void the warranty .
• Regardless of which type of steam sterilization is used, it is extremely important that the handpiece is rapidly and
completely dried before it is stored. Do not operate or store the handpiece unless a drying cycle has been
preformed. If a vacuum drying cycle is not used following steam sterilization, moisture may be trapped within the
handpiece causing corrosion and residue deposits, resulting in premature wear and a reduction in the functional
life expectancy of the handpiece. In addition, damage may result if the handpiece is operated with moisture in the
electronic connections.
• Store system in a clean dry place.
• Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and put into
service according to the EMC information provided in this Guide.
• Portable and mobile RF communications equipment can affect Medical Electrical Equipment.
• Use of accessories and cables other than those specified and sold by Medtronic Xomed may result in increased
emissions and decreased immunity of this unit.
• The XPS
®
3000 should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is
necessary , the XPS® 3000 should be observed to verify normal operation in the configuration in which it will be
used.
NOTICE
It is important that you read this entire manual
before you use the XPS
use these devices before you have read and are
thoroughly familiar with this document.
vi
®
Systems. It is unsafe to
CUSTOMER SERVICE
U.S. CUSTOMER SERVICE
General customer service and technical support are available toll-free:
800-874-5797 or 904-296-9600
Monday-Friday
8:00 AM - 6:00 PM E.S.T .
www.xomed.com
MICROELECTRONICS REPAIR
T echnical Support:
800-872-9877
904-296-6448 (F AX)
Monday - Friday
8:00 AM - 5:00 PM E.S.T .
Return Address:
Medtronic Xomed, Inc.
4102 Southpoint Blvd.
Jacksonville, FL 32216-0980 U.S.A.
Attention: Repair Department
CUSTOMER SERVICE INFORMATION
For further information regarding the use of this product or to report any problems, please contact Medtronic
Xomed using the appropriate information provided on the blue and white contact information card packaged with
each device; or contact your local distributor.
MEDTRONIC XOMED, INC. HELPLINE
Should you need immediate help with a technical question or guidance through the appropriate protocol, just call
the Medtronic Xomed Help Line at 1-800-874-5797.
NOTE:
When contacting our Customer Service and Technical Support, please have the appropriate product number,
product serial number, date of purchase, and nature of inquiry available.
Product Number
Serial Number
Date of Purchase
vii
DEVICE DESCRIPTION
SYSTEM DESCRIPTION
The XPS® 3000 System consists of a power control console, footswitches, connection cables, and assorted
handpieces to drive various burs, blades, drills, rasps, and cannulae.
XPS® 3000 CONSOLE — Provides power to the selected handpiece. In addition to the StraightShot® M4
Microdebrider and V isao™ High-Speed Otologic Drill, the system console also powers the StraightShot® Magnum
and Magnum® II handpiece, Powerforma® (High Speed Drill), Xcalibur® Hi-Torque and Hi-Speed Drills, and Skeeter
(microdrill) handpieces. It includes an integrated irrigation pump for irrigation of blades and burs and a second
pump (optional) for motor coolant of the V isao™ High-Speed and Xcalibur® Hi-Speed Otologic Drills.
FOOTSWITCH — The Single Function footswitch controls speed; the Multifunction footswitch controls
handpiece speed, rotation mode, and footswitch mode.
XPS® STRAIGHTSHOT® M4 MICRODEBRIDER — Features a finger wheel that allows the user to rotate the
cutting tip of specially designed straight and curved rotatable blades. It is also compatible with non-rotating blades
and burs. It includes a locking mechanism to prevent inadvertent rotation of non-rotating blades. The lightweight
design combines high power and precision performance and can operate in forward mode of 12,000 rpm for bur
operation and 5,000 rpm in oscillate for blade operation. The StraightShot® M4 also provides irrigation tubing
management grooves to keep tubing out of the surgeon’s way during a procedure.
STRAIGHTSHOT® MAGNUM® II MICRODEBRIDER HANDPIECE — Features a locking mechanism and allows
adjustability of blade or bur tip alignment. The lightweight design combines high power and precision performance
and can operate in forward mode of 12,000 rpm for bur operation and 5,000 rpm oscillate for blade operation.
STRAIGHTSHOT® MAGNUM® HANDPIECE — Features a locking mechanism and allows adjustability of blade
or bur tip alignment. The lightweight design combines high power and precision performance.
VISAO™ HIGH-SPEED OTOLOGIC DRILL — The lightweight design is used in otologic bone drilling operations.
It may be used with either standard burs or with V isao™ High-Speed Curved Burs.
®
®
XCALIBUR® HI-TORQUE AND HI-SPEED (W A TER-COOLED) OTOLOGIC DRILLS — Has two motor
assemblies, one hi-torque and one hi-speed (Water-Cooled); and one straight and one angled handpiece attachment. It is used in otologic bone drilling operations.
SKEETER® UL TRA-LITE OTO-TOOL — A slender , lightweight drill handpiece and burs specifically used in
middle ear surgical procedures, including stapes footplate procedures. The Skeeter® may be powered from the XPS
3000 console, XPS® 2000 console, or from a battery powered variable speed foot control.
POWERFORMA® OTOLOGIC DRILL — Straight and angled drill handpieces for otologic bone drilling. The
Powerforma® connects, via an adapter plug, to the XPS® 3000 and XPS® 2000 consoles.
®
1
INDICATIONS
The XPS® 3000 is intended for the incision and removal of soft and hard tissue or bone in general otorhinolaryngology, head and neck, and otoneurological sur gery . An integral pump is provided for irrigation and a second integral
pump may be provided for handpiece cooling.
Otology / neurotology indications include aural atresia, cholesteatoma, cochleostomy, development of a suture
tunnel for cochlear implant fixation, drainage of petrous apex cyst from endaural and middle-fossa approach,
endolymphatic hydrops, exostosis lesion removal, facial nerve decompression, mastoidectomy, mastoidotomy,
ossicular chain reconstruction (OCR), otosclerosis, removal of ear tumors including acoustic neuroma, tympanoplasty and vestibular neurectomy.
Sinus indications include septoplasty, removal of septal spurs, polypectomy, antrostomy, ethmoidectomy/
sphenoethmoidectomy, frontal sinus trephination and irrigation, frontal sinus drill out, endoscopic DCR, transsphenoidal procedures, maxillary sinus polypectomy, circumferential maxillary antrostomy, choanal atresia,
sphenoidectomy , and medial, lateral, and posterior frontal sinusotomy .
Nasopharyngeal/laryngeal indications include adenoidectomy , tracheal procedures, laryngeal polypectomy, laryngeal lesion debulking, including the surgical management of recurrent respiratory papillomatosis (RPP), and tonsillectomy, tonsillotomy for obstructive tonsillar disease and removal of endobronchial lesions.
Head and neck (ENT) indications include soft tissue shaving, rhinoplasty (narrowing of the bony vault and revision
of the bony pyramid), removal and shaping of bone during rhinoplasty procedures, removal of adipose tissue (lipo
debridement) in the maxillary and mandibular regions of the face, removal of acoustic neuroma, and incision and
removal of soft tissue during plastic, reconstructive, and/or aesthetic surgery.
The XPS
and hard tissue or bone is required. These include spinal and small and large joint arthroscopic procedures.
®
3000 system is indicated for use in orthopedic surgical procedures where the cutting and removal of soft
Magnum® Sterilization Tray
IV Pole Basket for Footswitch
Model 3000 System User’s Guide
Model 3000 System Repair Manual
StraightShot® M4 Sterilization Tray
Hi-Speed Extended 75 mm Bur Guard, Without Irrigation
®
Hi-T orque Extended Bur Guard, W ith Irrigation
®
Hi-Speed Extended 75 mm Bur Guard, With Irrigation
®
Hi-Speed Extended 64 mm Bur Guard, Without Irrigation
®
Hi-Speed Extended 64 mm Bur Guard, With Irrigation
®
Sterilization Tray
®
Hi-Speed Drill Sterilization Tray
Irrigation Sleeve
™
Extended Bur Guard, Without Irrigation
™
Extended Bur Guard, With Irrigation
™
Sterilization Tray
®
For a complete list of system components and accessories
contact your local Medtronic Xomed representative
or call Customer Service at 1-800-874-5797 or 1-904-296-9600.
3
XPS® 3000 CONSOLE DESCRIPTION
FRONT CONSOLE
1. SPEED DISPLA Y – When the system is running, the actual operating speed of the handpiece displays. When
inactive, the display shows the top speed set by the operator.
2. ROTATION INDICA TOR – Displays rotation mode (forward, reverse, or oscillate). Lights indicate selected
rotation. Rotation may be selected at either the multi-function footswitch or with the center control button in the
normal mode.
3. PROCEDURE AND CURRENT OPERATIONS DISPLA Y – In operating mode, displays procedure-blade/bur
setting and irrigation pump rate. In procedure selection/menu mode, it displays a choice of procedures, allowing
the user to choose suggested operational settings for the current procedure.
4. CONTROL BUTTON CLUSTER – Operates in two modes, Normal and Procedure. In Normal operation, center
button controls direction (forward and oscillate), top and bottom buttons control speed (increase and decrease),
and left and right button control rate of irrigation (decrease and increase). Depressing the menu button activates
the Procedure Mode. Arrow buttons control the cursor and the center button is the Select button. Operation of
the speed and procedure functions is described below.
5. HANDPIECE CONNEC-TORS – Connect the appropriate handpiece to the
console here. (Large connector for all handpieces
except for Skeeter® handpiece.)
6. FOOTSWITCH CON-NECTOR – Connect the
footswitch to the console
here. During standard operation the footswitch
must be connected to operate the active handpiece.
7. HANDPIECE IRRIGA-TION PUMP – Pumps
fluid through a tubing set
to the blade or bur.
8. XCALIBUR® HI-SPEED
WATER COOLANT
PUMP –
Pumps cooling fluid for
the W ater -Cooled drill
motor assembly (available only on REF#
1897102).
Figure 1 Front Console
4
REAR CONSOLE
1. MANUAL ST AR T STOP BUTTON – In the case of footswitch failure intraoperatively , button allows
circulating nurse to manually turn handpiece operation on and off.
2. ENDO-SCRUB® CABLE CONNECTOR – Accepts accessory cable which powers the Endo-Scrub® 2.
3. FUSE ACCESS – REPLACE ONL Y WITH 250V, 4.0A, T ype F , 5 x 20mm fuses.
4. POWER SWITCH – Press the power switch to the “|” position to turn on the system. Press the switch to the
“O” position to turn the system off.
5. POWER CORD CONNECTOR – Hospital grade power cord connects here.
6. DB-9 CONNECTOR – Connector for future data exchange.
Figure 2 Rear Console
5
XPS® 3000 SET-UP AND USE
CAUTION
Inspect components for damage before use. Do not operate if components are damaged.
CABLE CONNECTIONS
To connect the cables to the console:
1. Insert the cable connectors by aligning the red dot on the cable connector with the red dot on the console
connector.
2. Push connectors together. They should lock in place.
To disconnect the cables from the console:
1. Grasp the cable connector and gently pull back on the collar of the connector.
2 . The connector will disconnect.
CAUTION
Grasp the cable connector; do not pull on the cable.
CONSOLE SET-UP
LANGUAGE SELECTION
The system is provided with a one pump console (1897101) or two pump console (1897102) defaulted to the
customer’s preferred language. Each console is capable of displaying six (6) languages as indicated below:
One Pump - 1897101
English
French
Italian
German
Spanish
Dutch
One Pump - 1897101
English
Spanish
Danish
Finnish
Swedish
Portuguese
T wo Pumps - 1897102
English
French
Italian
German
Spanish
Dutch
T wo Pumps - 1897102
English
Spanish
Danish
Finnish
Swedish
Portuguese
To change the language selection, simultaneously press the left and right arrow key on the control button cluster.
Next, a menu will appear on the display showing the language choices. Use the arrow keys to move the cursor to the
desired language and press the center key to select.
PRECAUTIONS
• System components should be operated and inspected for damage prior to use. DO NOT USE SYSTEM IF
DAMAGE IS APPARENT .
• Do not allow the console to get wet. If liquid enters the console, damage could occur.
WARNING
• Only connect the console to hospital grade receptacles to ensure electrical grounding reliability.
1. Plug the power cord into the console and into the wall outlet.
2. Connect the handpiece (s) to the appropriate console connector (s).
3. Connect the footswitch cable to the console. See next section for instructions on footswitch operation.
4. Turn the console on by pressing the power switch on the back panel. The console will display default
settings for the handpiece speed, direction of cutter rotation, and type of surgical procedure.
5 . Figures 4 and 5 shows the default settings in Normal and Procedure Mode.
6
Increase Speed
Select Menu Mode
Increase Irrigation
and Prime
Select
Scroll UpReturn to
Standard Mode
Scroll
Right
Decrease Irrigation
Decrease Speed
Figure 3-A Control Button Cluster-
Standard Mode
Direction Change
Scroll LeftScroll Down
Figure 3-B Control Button Cluster-
Procedure Mode
6. The control button cluster can control speed, direction, irrigation rate, and, in Procedure Mode (depress
menu button), give the operator suggested operating speeds for different procedures.
7 . Speeds may be adjusted by pressing top and bottom buttons to increase and decrease speed. Pump flow rate
may be adjusted by pressing right (increase) and left (decrease) buttons; press and hold right button for
prime. Direction (forward, reverse, oscillate) can be changed by pressing the center button.
Default speeds are listed below (* Not applicable).
FWD6,000 rpm12,000 rpm
Xcalibur® Hi-TorqueFWD or REV10,000 rpm52,000 rpm
Xcalibur® Hi-SpeedFWD or REV10,000 rpm80,000 rpm
8. ROTA TION MODE — Changed either by pressing the Rotation Control button (in the center of the control
button cluster) on the console or, if the multifunction footswitch is used, depressing and releasing the left
pedal.
7
0
Figure 4 Default Displays and Settings
rpm
MIDDLE EAR
1200
SKEETER
FORWARD
IRRIGATION OFF
No
Yes
HA NDPIECE
INSERT HANDPIECE
DETE CTE D ?
WHI C H
HA NDPIECE?
SKEETER
DRILL
HI-SPD
MAGNUM
DRILL
& M4
M AGNUM II
rpm
MASTOID
43000
rpm
MASTOID
60000
rpm
SINUS
BUR
FORWARD
IRRIGATION 40%
BUR
FORWARD
IRRIGATION 40%
5000
BLADE
OSCILLATE
IRRIGATION 40%
rpm
SINUS
3000
BLADE
OSCILLATE
Defaul t
Settings
Displays and
8
IRRIGATION 40%
Figure 5 Menu Flow Diagram
rpm
1500
ADENOID/TONSIL
RADENO ID
OSCILLATE
rpm
PLASTICS
1800
MICRO-PLANER
OSCILLAT E
IRRIGATION 25%
rpm
PLASTICS
4000
IRRIGATION 45%
HYDRO
HYDR OBRASION
FORWAR D
IRRIGATION 25%
RHINOLIPO
rpm
5000
rpm
PLASTICS
RHINOPLASTY
FORWARD
IRRIGA TION OFF
SINUS
Diagram
ADENOID
& M4
Magnum II
10000
HI-SPEED BUR
FORWARD
IRRIGAT ION 55%
Menu Flow
<<PLASTICS>>
>RHINOPLASTY EXIT
LIPOPLASTY
HY DROBRASION
PLASTICSLARYNGEAL
MENU
Display
Microresector?
Default Handpiece
No
Yes
<<PROCEDURES>>
>SINUS PLASTICS
LA RYNGEA L AD ENOID
EX IT
EXIT
SINUS
<<SINUS>>
<<LARYNGEAL>>
> BLADE EXIT
BUR
>TRICUT
SKIMMER
HI-SPEED BUR
HI-SPEED BUR
BLADE
EXIT
Magnum
rpm
SINUS
6000
BUR
6000
& M4
Magnum II
Magnum
5000
3000
BUR
FORWARD
IRRIG ATION 55%
rpm
SINUS
BUR
FORWARD
IRRIGAT I ON 5 5%
rpm
SINUS
BLADE
OSCILLAT E
IRRIGA T ION 40%
rpm
SINUS
BLADE
OSCILLATE
IRRIGA T ION 40%
T RICUTSKIMMER
EXIT
AVAILABLE
NO MENU OPTIONS
500
rpm
LARYNGEAL
SKIMMER
OSCILLATE
IRRIGATION DRIP
rpm
LARYNGEAL
1200
TRICUT
OSCILLAT E
IRRIGATION DRIP
9
XPS® MULTIFUNCTION FOOTSWITCH, 1895400
The multifunction footswitch allows control of handpiece speed, rotation mode,
and footswitch mode. The footswitch cable connects with the console by a push
connector.
1. Right Pedal — The right pedal provides control of handpiece speed.
When in the VARIABLE operating mode, handpiece speed increases as
the pedal is depressed. In the START/STOP operating mode, the handpiece automatically operates at the MAX SET SPEED displayed on the
console when the pedal is depressed. The right pedal also activates the
irrigation pump if connected.
2. Left Pedal — Depressing and releasing the left pedal changes the
ROT A TION direction (oscillate, forward, or reverse) of the handpiece.
Note: Rotation mode may only be changed when the handpiece is
not running.
3. Top Button — Depressing and releasing this button changes
the FOOTSWITCH MODE (V ariable or Start/S top).
PRECAUTIONS
• DO NOT immerse the footswitch.
• DO NOT sterilize the footswitch.
Figure 6-A
XPS
1895400
3
1
2
3
1
2
®
Multifunction Footswitch,
XPS® MULTIFUNCTION FOOTSWITCH, SINGLE PEDAL, 1895420
The multifunction footswitch allows control of handpiece speed, rotation
mode, and footswitch mode. The footswitch cable connects with the console
by a push connector.
1. Pedal — The pedal provides control of handpiece speed. When in the
VARIABLE operating mode, handpiece speed increases as the pedal is
depressed. In the START/STOP operating mode, the handpiece
automatically operates at the MAX SET SPEED displayed on the console when the pedal is depressed. The pedal also activates the irrigation pump if connected.
2. Left Button — Depressing and releasing this button changes the
ROT ATION direction (oscillate, forward, or reverse) of the handpiece.
Note: Rotation mode may only be changed when the handpiece is
not running.
3. Right Button — Depressing and releasing this button changes the FOOTSWITCH MODE (V ariable or Start/
Stop).
PRECAUTIONS
• DO NOT immerse the footswitch.
• DO NOT sterilize the footswitch.
XPS® SINGLE FUNCTION
FOOTSWITCH, 1895410
2
→
3
←
↑
1
Figure 6-B
XPS® Multifunction Footswitch,
Single Pedal, 1895420
10
The single function footswitch operates only in a variable mode. It
controls handpiece speed, which increases as the pedal is depressed.
The pedal also activates the irrigation pump if connected. The
footswitch cable connects with the console by a push connector.
Figure 7
XPS® Single Function Footswitch,
1895410
XPS® STRAIGHTSHOT® M4 AND MAGNUM® II
MICRODEBRIDER SYSTEM SET-UP AND USE
The StraightShot® M4 and Magnum® II Microdebrider Drill Systems are intended for use with the XPS® Model 3000
Console, although they will operate with the XPS® 2000 Console. Likewise, the Straightshot® Magnum® will operate
with the XPS® 3000 Console. The StraightShot® M4 and Magnum® II default at 5,000 rpm oscillate and 6,000 rpm
forward. The StraightShot® Magnum® defaults at 3,000 rpm oscillate and 6,000 rpm forward. However, the
StraightShot® M4 and Magnum® II have top operating speeds of 5,000 rpm oscillate and 12,000 rpm in forward while
the StraightShot® Magnum® has top operating speeds of 3,000 rpm oscillate and 6,000 rpm in forward. If the
StraightShot® M4 and Magnum® II are connected to the XPS® 2000 Console they will operate at the same speeds as
the StraightShot® Magnum®. The multifunction or single function footswitches control the operating speeds of the
handpiece.
The StraightShot® M4 and Magnum® II Microdebrider Systems support a number of Otorhinolaryngology indications listed under “Indications.” Many disposable blades and burs are available for this purpose. Additionally , the
XPS® 3000 console offers a procedure setting screen which recommends and defaults to suggested operating
speeds for each procedure. The default settings can be adjusted. Speed can be adjusted, within allowable ranges,
using the top and bottom buttons in the control button cluster. Direction (forward and oscillate) can be adjusted
using the middle button. Pump flow rate for irrigation (bottom pump) is controlled with left and right buttons in the
control button cluster.
Pressing the “menu” button accesses procedure menus. After pressing the menu button, top, bottom, right, and left
buttons control the cursor on the display. The center button acts as the select button.
The StraightShot® M4 and Magnum® II Microdebrider Systems consist of the following main components:
• XPS® Model 3000 Console with integrated irrigation pump.
• StraightShot® M4 or Magnum® II Microdebrider Handpiece (to use the Image Guided capabilities of the
Image Guided version of this handpiece, refer to the following User’s Guides for your IGS system: LandmarX
ENT Image Guidance Systems, Treon® ENT Image Guidance System with LandmarX® Software, LandmarX
Evolution® ENT Image Guidance System, LandmarX Evolution® Plus ENT Image Guidance System, LandmarX
Element System and iNav™ Element System ).
• Footswitch
• XPS® Microdebrider blades and burs.
BEFORE SURGERY
WARNINGS
• Refer to “Warnings Before Surgery” in the front section of this manual.
The following instructions are recommended for set-up and use of the StraightShot® M4 and Magnum® II
Microdebrider Systems with the XPS® Model 3000 Console.
1. Inspect components for damage and determine if system is ready to use.
2 . On IV pole, mount XPS® Model 3000 console and plug unit into power source.
3 . Connect gray accessory cable (REF # 1895505) between Endo-Scrub® 2 and XPS® Model 3000 console if Endo-
Scrub® 2 is being used.
4. On the rear of the console, turn the power switch to the ON position.
5 . W ith tubing set (REF # 1895522) purchased separately or included with straight XPS® blades, spike the irriga-
tion bag, then place and secure tubing set into the lower pump in console (top pump is intended for cooling
discussed in drill set-up section).
6. Transferring sterilized StraightShot® M4 or Magnum® II handpiece plug/cord from scrub to circulator, plug
StraightShot® M4 or Magnum® II into the large handpiece port on the XPS® 3000 console.
®
®
11
7 . Connect footswitch into the XPS® Model 3000 and place under operating table.
8 . Place desired XPS® blade or bur in the StraightShot® M4 or Magnum® II collet mechanism as described in the
next section.
9 . Attach standard suction tubing to exhaust port of StraightShot® M4 or Magnum® II and to dedicated suction
canister.
10. Clip irrigation tubing, suction tubing, and handpiece cable together with clips enclosed with XPS® blade
tubing set, or with StraightShot® M4. Clips are permanently attached to electrical cable.
11. T o prime the irrigation tubing, press and hold the right arrow key . The pump will run in Prime mode until
the right arrow key is released.
12. After pump has been primed, set irrigation pump rate to desired setting.
13. If handpiece speed other than default is desired, increase and decrease with top and bottom buttons.
DO NOT use burs above the speed indicated on the bur label.
14. If direction other than oscillate is desired, change between forward and oscillate by using the center button.
15. If suggested procedure settings are desired, press menu button, use top, bottom, left, and right buttons to
scroll through the menu to pick procedure and blade. Press center button to select an option. See menu flow
diagram, Figure 5.
The following set-up is recommended for the StraightShot
StraightShot® to
XPS® Irrigation Tubing
REF # 1895522
XPS® I.V . Pole Basket for
Footswitch REF # 1897510
I.V. Pole
®
M4 and Magnum® II Microdebrider System:
XPS® 3000 Console, One or Two Pumps
REF # 1897101 or 1897102
Power Cord Connected to
Standard AC Wall Power
Standard Operating Room
Suction Tubing Connected to
Dedicated Wall Suction
StraightShot® M4 Handpiece
REF#1898200T (shown) or
StraightShot® Magnum® II
Handpiece
REF # 1897200, 1897201,
or 1897200T
12
XPS® Blade or Bur
(see XPS® brochure for REF #)
XPS® Footswitch (1895400 shown)
REF # 1895400, 1895410, or 1895420
Figure 8 XPS® StraightShot® M4 and Magnum® II Microdebrider System
XPS® STRAIGHTSHOT® M4 AND MAGNUM® II
HANDPIECE ACCESSORIES
WARNINGS
• Use adequate irrigation from a separate user-provided irrigating source. The use of a bur without irrigation may
cause an inordinate amount of heat buildup resulting in thermal injury to tissue.
• Disposable devices are for single-use only. Do not attempt to sterilize disposable devices. The disposables are
packed sterile and not intended for repeat use. To prevent contamination use only once.
• Use care in application of the moving cutting end to only appropriate anatomical landmarks and the intended
surgical site when using XPS
vibration-related injury. Use appropriate precautions.
• Employ visualization when using rotating XPS® accessories. Discontinue powered application in the event of
lack of visualization of the surgical site.
• Use methods at the operative site to control bleeding that do not compromise patient safety during at-risk
surgery.
• Always inspect the components before and after use for any damage. If damage is observed, do not use damaged
part until it is replaced. Damaged parts may deposit metal shavings on surgical site.
• Do not use any parts other than Medtronic Xomed, Inc. system components as damage or substandard performance could result.
• Remove and discard accessories following local regulations for proper disposal of contaminated materials.
• Do not modify accessories used with the handpiece. Performance could be diminished with modified accessories.
• Insertion of metal objects in blade or bur tip may cause the blade or bur to break leaving fragments in the wound.
The fragments may be difficult to remove, causing irritation, inflammation and foreign-body response at surgical
site.
• Always keep the cutting tip of the accessory away from fingers and loose clothing. Prevent laceration of user and
cross-contamination through compromised glove.
• Do not change accessory with handpiece running to prevent laceration of user and cross-contamination through
compromised glove.
• Discontinue use of curved bur if tip begins to wobble and replace bur to prevent unintended tissue removal from
patient.
• Do not use accessory if package is opened or damaged. Broken seal offers no protection against cross-contamination.
• Bending or prying may break the accessory, causing harm to patient or staff.
• Accessories are available for resection of soft tissue and bone for surgical procedures. Use of accessories
depends on the intended application and patient needs. Sharp-cutting powered accessories induce bleeding and
removal of significant tissue and bone.
• Insertion of metal objects in blade tip may cause the blade to break leaving fragments in the wound. The
fragments may be difficult to remove, causing irritation, inflammation and foreign-body response at surgical site.
• Use lock on StraightShot® M4 handpiece to prevent inadvertant rotation of blade or bur during use.
• When precise location of the blade tip is required, engage the rotation lock on the handpiece, then calibrate and
verify the blade tip on the Image Guided Surgery (IGS) system. Always lock the StraightShot® M4 handpiece
when driving non-rotatable blades to maintain their IGS calibration.
®
accessories. The use of powered reciprocating/rotating instruments may result in
13
BLADE OR BUR INSTALLATION
1. Using thumb, depress the locking collar (A) on the front
of the handpiece.
2 . Insert blade or bur with a slight rotating motion until blade
or bur is seated. For rotatable straight blades, orient the
irrigation barb at the 3 o’clock position for right-handed
surgeons and 9 o’clock for left-handed surgeons. For rotatable curved blades, orient the irrigation barb at 3 o’clock.
T o fully seat an M4 rotatable blade, adjust the finger wheel
with small back-and-forth motions.
3. Align blade or bur tip opening to desired position.
4 . Release the locking collar .
5 . Pull on the blade or bur to ensure engagement and visually
check to make sure distal tip of inner blade is in contact with the distal tip of the outer cannula.
6 . For non-rotating blades, orient the blade or bur to desired position (usually 12 o’clock for the cutting surface and
6 o’clock for the irrigation barb) and locking mechanism (D).
TUBING ATTACHMENT
1. Adjust the clamp on the irrigation tubing to the off position
and spike the bag of irrigant.
2. Push the free end of the irrigation tubing onto the irrigation
port on the blade (B) until tubing passes barb on the port and
is secure.
3. Attach suction tubing securely to the suction port on the
handpiece (C).
4. Secure the suction and irrigation tubing to the handpiece
cable with the clip (if provided) and the white clamps included in the blade packaging.
Figure 9 Blade or Bur Installation
A
Figure 10 Tubing Attachment
C
B
D
XPS® STRAIGHTSHOT® M4 AND MAGNUM® II
SURGICAL PRECAUTIONS
• XPS® Blades should be operated in the oscillate mode only . Operating in the forward mode may cause damage to
the blade. XPS® burs should be operated in the forward mode only.
• Be sure the blade is fully engaged in the Microdebrider and verify the tip is fully engaged with the outer cannula
prior to use.
• To prevent damage to curved blades, disconnect suction tube prior to changing blade or bur during procedure.
• The StraightShot® Magnum® II and StraightShot® M4 Microdebriders are intended to operate at speeds greater
than 6,000 rpm ONL Y when used with the XPS® High Speed bur line.
14
T able. Using Wheel and Lock on S traighShot® M4 Handpiece linked to an Image Guided Surgery
(IGS) System.
Non-Rotatable Blades,
Including Laryngeal
M4 Rotatable Blades
Location of tip must be known
during procedure.
1. Adjust blade tip to desired
location.
2. Activate LOCK.
3. Mount IGS attachement to
handpiece using Allen wrench.
4. Calibrate blade tip on IGS.
5. Wheel is not usable.
1. Adjust blade tip to desired
location.
2. Mount IGS attachment to
handpiece using Allen wrench.
3. Calibrate blade tip on IGS.
V ariation preferred during procedure.
Tip Location
1. Do not LOCK handpiece.
2. Do not refer to IGS for
indication of location.
3. Wheel is usable to vary tip
location.
1. Release blade from collet.
2. Rotate blade into one of four
present orientations.
3. Lock blade into collet.
Blade-window
Not applicable.
1. Do not LOCK
handpiece.
2. Mount IGS
attachment ot
handpiece using
Allen wrench.
3. Calibrate blade tip
on IGS.
4. Wheel is usable to
vary window
location.
AFTER SURGERY
BLADE OR BUR REMOV AL
1 . If the irrigation pump is not used, turn off the irrigation flow using the roller clamp on the irrigation tubing.
2. Remove and discard the tubing following local regulations for proper disposal of contaminated materials.
3 .Depress the locking collar of the handpiece and pull the blade or bur out of the handpiece and discard following
local regulations for proper disposal of contaminated materials.
15
VISAO™ AND XCALIBUR® DRILL SYSTEM
SET-UP AND USE
The Visao™ and Xcalibur® Hi-Speed Drills are used with the XPS® Model 3000 console with two integral pumps (one for
handpiece cooling, the other for irrigation). The Xcalibur
®
Hi-T orque Drill System is intended for use with the XPS® Model
3000 Console, or the XPS® Model 2000 console and the optional XPS® Irrigator Pump, for the controlled dissection and
removal of bone during ENT surgery.
Speed of the XPS® 3000 Drill System is set by use of the top and bottom buttons on the control button cluster on the console,
and controlled by using the XPS
®
single function or multifunction footswitch. Rate of Irrigation is controlled by left and right
(decrease and increase) buttons on the control button cluster.
The Visao™ High-Speed Otologic Drill operates with Medtronic Xomed straight burs and V isao™ High Speed Curved Burs to
meet the needs of the surgeon, eliminating the need for handpiece attachments.
The Xcalibur® drills have the ability to operate on the XPS® Model 3000 with interchangeable Hi-Speed water-cooled and HiTorque motors and straight or angled handpieces to meet the needs of the surgeon. Note that only the Xcalibur® Hi-Speed
Straight or Xcalibur® Angled Handpieces will fit onto the Xcalibur® Hi-Speed (W ater-Cooled) Motor Assembly .
The Visao™ and Xcalibur® Drill’s bur notch mechanism incorporates Medtronic Xomed surgical burs of diameter .092 inches
(2.34mm) that are at least 1.73 inches (44mm) long.
The XPS® 3000 Drill System is comprised of the following main components:
• XPS Model 3000 Console with Optional Coolant Pump
• Footswitch
• Either: Visao™ High-Speed Otologic Drill
• Or: Xcalibur Hi-Speed and Hi-Torque Electric Motors,
Xcalibur Hi-Speed and Hi-T orque, Straight or Angled Handpiece Attachments.
The following set-up is recommended for the Hi-T orque Motor:
I.V . Pole
Irrigation Tubing Set
REF # 3318503
XPS® 3000 Console, One or Two Pumps
REF # 1897101 or 1897102
Xcalibur® Angled Hi-Speed
Handpiece Att achment REF # 3333755
or Xcalibur® Angled Hi-Torque Handpiece
Attachment REF # 3333750
Xcalibur® Straight Hi-Speed Hand piece
Attachment REF # 3333705 or Xcalibur
Straight Hi-Torque Handpiece Attachment
REF # 3333700
XPS® I.V . Pole Basket for
Footswitch REF # 1897510
®
Figure 11 Xcalibur® Hi-T orque Drill System
16
Lower Pump
for Bur
Irrigation
Medtronic Xomed
Surgical Bur
(see Surgical Bur
Catalog for REF #)
Xcalibur® Hi-T orque Drill Motor
REF # 3333850 (shown with
angled handpiece attachment)
Irrigation Tubing Clips
(irrigation tubing to cable,
provided with irrigation tubing)
Power Cord Connected to
Standard AC Wall Power
XPS® Footswitch (1895400 shown)
REF # 1895400, 1895410, or
1895420
The following set-up is recommended for the V isao™ High Speed Otologic Drill (W ater-Cooled) and Xcalibur
Hi-Speed (W ater-Cooled) Motor:
®
Irrigation Tubing Set
REF # 3318503
Hi- Speed Drill Coolant
Tubing Set REF # 3318506
(with dual cannulated
spike and squeezable drip
chamber)
T op Pump for
Motor Cooling
Lower Pump
for Bur Irrigation
I.V . Pole
Straight Hi-S peed Hand piece
Attachment REF # 3333705
(Xcalibur
®
only)
Medtronic Xomed Surgical Bur
(see Surgical Bur Catalog for REF #)
XPS® 3000 Console, Two
Pumps REF # 1897102
Medtronic Xomed Surgical Bur (see
Surgical Bur Catalog for REF #)
Angled Hi-Speed Handpiece
Attachment REF # 3333755
(Xcalibur® only)
Xcalibur® Hi-Speed Motor
(Water-Cooled) REF
# 3333840 (shown with
straight handpiece attachment)
XPS® I.V . Pole Basket for
Footswitch REF # 1897510
Figure 12 Visao™ High-Speed Otologic Drill or Xcalibur® Hi-Speed Drill System
Coolant Tubing Clips (coolant tubing
& irrigation tubing to cable, provided
with coolant tubing)
Power Cord to
Standard AC Wall Power
XPS® Footswitch (1895400 shown)
REF # 1895400, 1895410, or 1895420
17
BEFORE SURGERY
WARNINGS
• Refer to “Warnings Before Surgery” in the front section of this manual.
™
The following instructions are recommended for set-up and use of the V isao
High-Speed Otologic Drill, Xcalibur
Hi-Speed (W ater-Cooled) Drill system and/or the Xcalibur® Hi-T orque Drill System with the XPS® Model 3000:
1. Inspect components for damage and determine if system is ready to use.
2 . On IV pole, mount XPS® Model 3000 console and plug unit into power source.
3. On the rear of the console, turn the power switch to the ON position.
4 . Plug the connector end of the motor cord into the XPS® Model 3000 console.
5 . Connect footswitch into XPS® Model 3000 console and place under operating table.
6 . If irrigation is desired, connect Irrigation T ubing (REF # 3318503) to irrigation port on V isao™ Irrigation Sleeve
(REF # 3334610) or, V isao™ Extended Bur Guard with Irrigation (REF # 3334635) or, Xcalibur® Irrigation Sleeve
(REF # 3333610) or, Xcalibur® Hi-Speed Extended 75 mm Bur Guard with Irrigation (REF # 3333635) or, Xcalibur
®
Hi-Speed Extended 64 mm Bur Guard with Irrigation (REF # 3333655) or, Xcalibur® Hi-Torque Extended Bur
Guard with Irrigation (REF # 3333630). Spike irrigation bag and place the short piece of silicone tubing through
the lower pump. Close pump head.
7 . If the V isao
™
High-Speed Otologic Drill (REF # 3334800 or REF # 3334800T) or Xcalibur® Hi-Speed, Water-
Cooled, Motor Assembly (REF # 3333840) is used, attach each of the tubes from the Coolant Tubing (REF #
3318506) to the ports on the handpiece. Spike a second large bag of irrigation fluid (sterile water , or saline if
sterile water is not available). Place the short piece of silicone tubing through the coolant pump (top pump on
the XPS® Model 3000 console).
8 . Clip Irrigation T ubing to V isao™ motor cable clips or Xcalibur® motor cable using clips provided with tubing set.
9. If the Visao™ Drill is used, both the irrigation tubing and coolant tubing can be retained in the clips on the
handpiee cable. If the Hi-speed (Water-Cooled) motor is used, clip Hi-Speed Coolant Tubing and Irrigation
Tubing (if used) to Xcalibur® Motor Cable using clips provided with the tubing set. (The clips in the Irrigation
Tubing will not be needed.)
10 . T o prime the irrigation tubing, press and hold the right arrow key . The pump will run in Prime mode until the
right arrow key is released. This function runs simultaneously with the coolant pump.
11 . To prime the coolant pump, press and hold the right arrow key . The pump will run in Prime mode until the right
arrow key is released. This function runs simultaneously with the irrigation pump. Alternatively , ensure the
handpiece bur lock is closed (align small dots on handpiece), then depress the foot pedal momentarily (1-2
seconds). After the foot pedal is released, the cooling pump will continue to run for approximately one minute.
Continue to prime the cooling system until all bubbles are removed.
12. Select drill speed settings by depressing the top or bottom buttons (increase and decrease speed) in the control
button cluster.
13 . Unlock the Visao™ or Xcalibur® Drill handpiece collet by rotating lock ring 1/4 turn, insert desired bur (rotating
to engage the base of the bur completely into position in the handpiece), then rotate the lock ring 1/4 turn back
to lock the bur into the handpiece (aligning the small dots on the handpiece).
14. Always tug bur for security. Repeat step 13 if necessary .
15. IMPORTANT: Test drill by depressing foot pedal prior to use. The system is now ready to use. Note that the
cooling pump will continue to run for about one (1) minute after your foot is removed from the foot pedal.
®
18
VISAO™ AND XCALIBUR® BUR INSTALLATION AND TUBING
ATTACHMENT
WARNINGS
• Do not change accessory with handpiece running to prevent laceration of user and cross-contamination through
compromised glove.
• Do not use burs above the speed indicated on the bur label. Exceeding speed may cause the burs to break.
• Carefully inspect burs prior to and following each use for excessive wear, fragmentation, eccentricities or other
defects. Do not use dull, damanged or bent burs. Use of dull burs can reduce the handpiece effectiveness and
cause the handpiece temperature to increase.
• Do not attempt to resharpen used burs. Worn burs should be replaced with new ones frequently to ensure
effective cutting and control of the drill.
• Excessive pressure applied to bur may cause bur fracture. Should a bur fracture during use, extreme care must be
exercised to ensure that all fragments of the bur are retrieved and removed from the patient. Unremoved bur
fragments may cause tissue damage to the patient.
• Test for bur wobble (eccentricity) at desired speed prior to use. Use a bur guard (Xcalibur
a new bur or reduce speed if wobble is observed.
• Test for bur wobble (eccentricity) at the desired speed prior to use. Select a new bur or reduce speed if wobble is
observed. Eccentricity of the bur can cause bur vibration and may result in excessive tissue and bone destruction
and hearing damage. Always examine the operation of each bur in the handpiece before each use. Operating
certain burs at high speed can cause vibration of the bur.
Visao™ High-Speed Curved Burs Only
• Keep the cooling sleeve irrigated to prevent thermal injury to tissue.
• During procedures near nerves, keep bur and bur cannula away from tissue to minimize the potential for thermal
injury.
• For procedures near nerves, nerve monitoring should be used to alert the user of the potential for injury.
®
only) if needed. Select
BUR ATTACHMENT
1 . Using thumb and index finger, twist handpiece locking mechanism counter clockwise to unlock handpiece.
2 .Insert desired Medtronic Xomed Surgical Bur into handpiece, giving bur a 1/4 turn as it is seated
(or until it seats fully into the handpiece).
3 . Lock and secure bur by twisting handpiece locking mechanism clockwise.
4. Tug on bur to ensure bur is secured and locked into the handpiece.
CURVED BUR INST ALLA TION (VISAO™ HIGH SPEED CURVED BURS ONL Y)
1 .Using thumb and index finger, twist the handpiece locking collar to the unlocked position.
2. Insert the curved bur onto the nose of the handpiece loosely and rotate the curved bur to the desired angular
position.
3 .Firmly push the curved bur hub onto the handpiece until it snaps in place. Gently twist the curved bur hub back
and forth to ensure it is completely seated in the angular detents.
4 .Using thumb and index finger, twist the handpiece locking collar to the locked position.
5. Push the bur head into the handpiece until it snaps in place.
6 .Gently tug on the bur head to ensure it is locked in place.
TUBING A TT ACHMENT - IRRIGA TION
1 . During initial set-up, ensure drill irrigation tubing is clipped into place along the motor electrical cable using clips
provided with tubing set.
2 . Insert tubing from the Visao™ or Xcalibur® Drill handpiece on to the V isao™ Irrigation Sleeve (REF # 3334610) or,
the Visao™ Extended Bur Guard with Irrigation (REF # 3334635) or, Xcalibur® Irrigation Sleeve Assembly (REF #
3333610) or, Xcalibur® Hi-Speed Extended 75 mm Bur Guard with Irrigation (REF # 3333635) or, Xcalibur® Hi-Speed
Extended 64 mm Bur Guard with Irrigation (REF # 3333655) or, Xcalibur® Hi-T orque Extended Bur Guard with
Irrigation (REF # 3333630).
3. Rotate the irrigation sleeve to the desired surgeon position.
4 . Spike the irrigation bag.
5. Install the tubing into the pump.
6 .Prime the tubing set.
7. Test tubing flow and set-up by depressing footswitch.
19
TUBING A TT ACHMENT - COOLING (W A TER-COOLED MOTOR)
1 . During initial set-up, ensure water-cooled drill coolant tubing is clipped into place along the motor electrical cable.
For Xcalibur®, use clips provided with tubing set. (The irrigation tubing, if used, can be attached using these same
clips, and the clips in the irrigation tubing set will not be needed.)
2. Connect tubing set to the cooling ports of the water-cooled motor.
3 . Spike a large bag of irrigation fluid (preferably sterile water, but saline may be used).
4. Place the short section of silicone tubing through the cooling pump (upper pump).
5. Prime the cooling tubing set while squeezing drip chamber a few times. Continue to prime until air bubbles no
longer occur at the coolant bag. IMPORTANT: Ensure all air is purged out of the motor . Roll the motor back and
forth during priming to encourage air out of the motor’s cooling jacket.
VISAO™ AND XCALIBUR® SURGICAL PRECAUTIONS
• Always ensure that the bur is securely engaged into the handpiece prior to operating the system.
• When operating or testing the Poweforma®, Xcalibur® and Visao™ drill handpieces, ensure the bur is properly
inserted and locked into the handpiece. Running the drill with the collet unlocked can damage the locking
mechanism.
• Improper priming of the Xcalibur® and V isao™ High-Speed Drills will result in excessive handpiece temperature.
• The Visao™ and Xcalibur® Hi-Speed W ater-Cooled Drills work only with the XPS® 3000. Use without water-bag
will damage the motor in the handpiece.
• IMPORTANT: Always examine operation of each bur in a handpiece before each use. Operating certain burs at
high speeds can cause vibration of the bur. The following are suggestions for reducing bur vibration:
> Reduce handpiece operating speed.
> Use only burs that are rated for High Speed operation.
> Use Visao™ Extended Bur Guards (REF # 3334625 or 3334635) with 64 mm or 75 mm burs in the V isao™ High-
Speed Otologic Drill.
> Use Xcalibur® Hi-Speed Extended 64 mm Bur Guards (REF # 3333645 or 3333655) with 64 mm burs in the
Xcalibur® Hi-Speed Drill.
> Use Xcalibur® Hi-Speed Extended 75 mm Bur Guards (REF # 3333625 or 3333635) with 75 mm burs in the
Xcalibur® Hi-Speed Drill.
> Use Xcalibur® Hi-Torque Extended Bur Guards (REF # 3333620 or 3333630) with 75 mm burs in the Xcalibur
Hi-T orque Drill.
> Operate handpiece at 50% speed when using burs 75 mm or longer .
> Select a new bur .
®
20
POWERFORMA® DRILL SYSTEM SET-UP AND USE
The Powerforma® Drill System is intended for use with the XPS® Model 3000 console, or the XPS® Model 2000
console and optional XPS
The primary application of the Powerforma
®
Irrigator Pump, for the controlled dissection and removal of bone during ENT surgery .
®
Drill System is mastoid bone drilling. Speed of the Powerforma® Drill
System is set by use of the top and bottom buttons on the control button cluster on the console, and controlled by
using the XPS® single function or multifunction footswitch. Rate of Irrigation is controlled by left and right (decrease and increase) buttons on control button cluster.
The Powerforma® Drill System’s bur notch mechanism incorporates Medtronic Xomed sur gical burs of diameter .092
inches (2.34 mm) that are at least 1.73 inches (44 mm) long.
The Powerforma® Drill System is comprised of the following main components:
• XPS® Model 3000 Console
• Straight or Angled Handpiece
• Footswitch
• Adapter for Powerforma® Handpiece
The following set-up is recommended for the Powerforma® Drill System:
I.V . Pole
Powerforma® to XPS® Irrigator
Tubing REF # 1895520
XPS® 3000 Console, One or Two Pumps
REF # 1897101 or 1897102
Powerforma® Straight Handpiece
REF # 3327700
Powerforma® Angled Handpiece
REF # 3327750
Adapter for Powerforma
Handpiece
REF # 1994025
XPS® I.V. Pole Basket for
Footswitch REF # 1897510
Medtronic Xomed
Surgical Bur
(see Surgical Bur
Catalog for REF #)
Powerforma® Tubing Clips
REF# 3318600
®
Irrigation Tubing
REF # 3318501
or 3318502
Figure 13 Powerforma® Drill System
Power Cord Connected to
Standard AC Wall Power
XPS® Footswitch (1895400 shown)
REF # 1895400, 1895410, or 1895420
21
BEFORE SURGERY
WARNINGS
• Refer to “Warnings Before Surgery” in the front section of this manual.
The following instructions are recommended for set-up and use of the Powerforma
3000:
1. Inspect components for damage and determine if system is ready to use.
2 . On IV pole, mount XPS® Model 3000 console and plug unit into power source.
3. On the rear of the console, turn the power switch to the ON position.
4. Attach the XPS® to Powerforma® Tubing (REF # 1895520) to the Powerforma® Irrigation Tubing (REF # 3318501)
which should already be attached to the Powerforma® Handpiece.
®
Drill System with XPS® Model
5 . Clip the Powerforma® Irrigation Tubing (REF # 3318501) along the handpiece electrical cord with Powerforma
Tubing Clips (REF # 3318600).
6 . Transfer the connector end of the handpiece electrical cable, along with the irrigation tubing, to the circulating
nurse. Be careful to maintain sterility .
7 . Spike the irrigation bag and place the XPS® to Powerforma® Tubing (REF # 1895520) through the lower pump.
Close the pump head.
8 . Connect footswitch into XPS® Model 3000 console and place under operating table.
9 . To prime the irrigation tubing, press and hold the right arrow key . The pump will run in Prime mode until the
right arrow key is released.
10. After tubing has been primed, set irrigation pump rate to desired setting.
11. Select drill speed settings by depressing the top or bottom buttons (increase and decrease speed) in the control
button cluster.
12. To insert a bur , pull back on the bur release button with your thumb, and slide the bur shaft into the handpiece
chuck. T wist the bur a quarter -turn (1/4) or until it slides fully into place. Allow the bur release button to slide
forward, which locks the bur into place.
13. Always tug bur to ensure it is securely locked in place. Repeat step 12 if necessary.
®
22
14. Test drill by depressing foot pedal. The system is now ready to use.
15. To remove the bur, pull back on the bur release button, and pull the bur out of the chuck.
POWERFORMA® BUR INSTALLATION AND TUBING ATTACHMENT
WARNINGS
• Ensure bur is properly secured in the Powerforma® handpiece prior to surgeon activation.
• DO NOT change burs with handpiece running.
• Use adequate irrigation. The use of a bur without irrigation may cause an inordinate amount of heat build-up
resulting in thermal injury to tissue.
• Test for bur wobble (eccentricity) at desired speed prior to use. Select a new bur or reduce speed if wobble is
observed.
• Excessive pressure applied to bur may cause bur fracture which may cause injury.
• DO NOT use dull, damaged, or bent burs. Change burs frequently for optimal performance. Use of dull burs can
reduce handpiece cutting effectiveness and cause handpiece temperature to increase.
BUR ATTACHMENT
1. Using your thumb, pull back on the bur release button.
2 .Insert desired Medtronic Xomed surgical bur into handpiece, giving bur a 1/4 turn as it is seated
(or until it seats fully into the handpiece).
3 .Allow the bur release button to slide forward.
4. Tug on bur to ensure bur is secured and locked into the handpiece.
TUBING A TT ACHMENT - IRRIGA TION
1 .During initial set-up, ensure drill irrigation tubing is clipped into place on drill using clips provided with tubing
set.
2 . Connect tubing set from the Powerforma® Drill handpiece into XPS® Powerforma® tubing set (REF # 1895520).
3 . Install XPS® to Powerforma® tubing set (REF # 1895520) into the lower pump. Close the pump head.
4 .Prime the tubing set.
5. Test tubing flow and set-up by depressing footswitch.
POWERFORMA® SURGICAL PRECAUTIONS
• Excessive noise from the bur when drilling close to the cochlea or ossicular chain may cause hearing damage.
• Excessive force may lead to bur bending, breakage or excess vibration. If the bur fractures during use, the
resultant sharp surfaces may cause lacerations of the body tissue. A bur which fractures inside a patient may also
cause tissue injury and necessitate emergency procedures for extraction.
• Eccentricity of the bur can cause bur vibration and may result in excess tissue and bone destruction and hearing
damage.
• When operating or testing the Powerforma® drill handpiece, ensure bur is properly inserted into the handpiece.
Running the handpiece with an unsecured bur can cause damage. Running a handpiece without a bur is not
recommended.
• During procedures, operating room technicians and nurses should always check bur function, fit, tightness, and
security in the handpiece before handing off to surgeon for use.
• Always examine operation of each bur in a handpiece before each use. Operating certain burs at high speeds can
cause vibration of the bur. The following are suggestions for reducing bur vibration:
> Reduce handpiece operating speed.
> Operate handpiece at 50% speed when using burs 75 mm or longer.
> Select a new bur.
23
SKEETER® ULTRA-LITE OTO-TOOL SYSTEM
SET-UP AND USE
The Skeeter® Ultra-Lite Oto-Tool System is a slender, lightweight drill handpiece with burs, specifically used in
middle ear surgical procedures, including stapes footplate surgery. The Skeeter® handpiece and burs may be used
with the XPS® Model 3000 via direct connection with the Skeeter® handpiece connector. The lightweight Skeeter
Ultra-Lite Oto-Tool Handpiece weighs 57 grams. The drill shaft diameter is approximately 2mm and is angled
approximately 15 degrees from the plane of the handpiece. The shaft angulation and small diameter maximize
visualization of the surgical field during drill use.
The Skeeter® Ultra-lite Oto-T ool Handpiece can be powered by three (3) 9 volt batteries housed in the Accelerator
Foot Control. However, when used with the XPS® Model 3000 console, the handpiece plugs directly into the
console and is operated using the XPS® single function or multifunction footswitch. The footswitch allows variable
speed operation of the handpieces ranging from very slow to high speed in excess of 12,000 rpm.
The Skeeter® is used with the Oto-Flex bur. The Oto-Flex bur designed for use with the Skeeter® are composed of a
flexible stainless steel shaft with a bur and a PTFE bearing at one end and the handpiece engagement at the other.
The inside of the PTFE bearing is lightly coated with silicone spray to reduce operating friction.
COLOR CODING
For ease of identification of size, all Oto-Flex Burs are color coded. Diamond Burs are further differentiated from
Carbide Burs by a white band on the shaft of each Diamond Bur . A Color Code Chart is conveniently provided in the
base of the Oto-T ool System Sterilization Case. The bur size identification color code is as follows:
®
Violet:0.5 mmY ellow:0.8 mmBrown:1.8 mm
Blue:0.6 mmOrange: 1.0 mmRed:2.3 mm
Green:0.7 mmGray:1.4 mmBlack:Specialty
For a complete listing of Skeeter Ultra-Lite Oto-Tool System, consult the “System Component and Accessories”
section of the Skeeter User’s Guide (68E1279).
The Skeeter® Ultra-Lite Oto-T ool System consists of the following main components:
• XPS® Model 3000 Console
• Skeeter® Ultra-Lite Oto-Tool Handpiece
• Footswitch
• Oto-Flex burs
24
The following set-up is recommended for the Skeeter® Ultra-Lite Oto-T ool System:
I.V . Pole
XPS® 3000 Console, One or Two Pumps,
REF # 1897101 or 1897102
Skeeter® Ultra-Lite Oto-Tool
REF # 3055601
Skeeter® Oto-Flex Bur
(see Surgical Bur Catalog for REF #)
XPS® I.V. Pole Basket for Footswitch
REF # 1897510
Figure 14 Skeeter® Ultra-Lite Oto-Tool System
Power Cord to Standard
AC Wall Power
XPS® Footswitch (1895400 shown)
REF # 1895400, 1895410, or 1895420
25
BEFORE SURGERY
WARNINGS
• Refer to “Warnings Before Sur gery” in the front section of this manual.
• DO NOT attempt to resharpen used burs. Worn burs should be replaced with new ones frequently to ensure
effective cutting and control of the drill.
• Carefully inspect burs both prior to and following each use for signs of excessive wear, fragmentation, eccentricities or other defects. Replace any suspicious bur with a new one prior to use.
• Should a bur fracture in use, extreme care must be exercised to ensure that all fragments of the bur are retrieved
and removed from the patient. Unremoved bur fragments may cause tissue damage to the patient.
PRECAUTIONS
• Always handle handpieces with care to avoid damage.
• Make certain the appropriate bur is securely fastened into the handpiece prior to use. USE ONLY MEDTRONIC
XOMED ROT ARY BURS INTENDED FOR USE WITH THE SKEETER
• Excessive noise from the bur when drilling close to the cochlea or ossicular chain may cause hearing damage.
• Excessive force may lead to bur bending, breakage or excess vibration. If the bur fractures during use, the
resultant sharp surfaces may cause lacerations of the body tissue. A bur which fractures inside a patient may
also cause tissue injury and necessitate emergency procedures for extraction.
• Eccentricity of the bur can cause bur vibration and may result in excess tissue and bone destruction and
hearing damage.
• DO NOT attempt to bend needle nose or effect repairs to the system.
®
UL TRA-LITE OTO-TOOL.
The following instructions are recommended set-up and use of the Skeeter
®
Ultra-Lite Oto-Tool with the XPS® 3000:
1. Inspect components for damage and determine if system is ready to use.
2 . On IV pole, mount XPS® Model 3000 console and plug unit into power source.
3. On the rear of the console, turn the power switch to the ON position.
4. After removal of the handpiece from the sterilization tray, transfer the connector from the sterile field to the
circulating nurse.
5. The cable is plugged into the appropriate (smallest) connector in the console.
6 . Connect footswitch into XPS® Model 3000 console and place under operating table.
7. Load the desired bur for the procedure into the handpiece by inserting the bur shaft through the distal end of
the handpiece with a slight twisting motion while simultaneously depressing the bur release button. The bur is
locked into place when a “click” is noted. Locking of the bur should be checked prior to use by firmly pulling
on the bur after the “click” is noted. T o remove the bur from the handpiece, depress the Bur Release Button on
the handpiece and carefully pull the bur out.
8. Select drill speed settings by depressing the top and bottom buttons (increase and decrease speed) in the
control button cluster.
9. Tug the bur to ensure its security in the handpiece.
10. Test drill by depressing foot pedal. The system is now ready to use.
26
CLEANING AND STERILIZATION GUIDELINES
The following cycles are compatible with the each handpiece as listed below; however, it is the r esponsibility of
each health care facility to validate the sterilization process with their particular equipment and procedures to
assure sterility.
PRECAUTION
• DO NOT cold soak sterilize the handpiece in glutaraldehyde. This will void the W arranty.
WARNINGS
• Disconnect power before cleaning the unit.
• Inspect components for any damage before and after each use. If you see damage, do not use the system until it
is repaired.
• After cleaning and sterilization, verify functionality by operating the equipment prior to the surgical procedure.
AFTER SURGERY
®
XPS
3000 Console
• DO NOT immerse the console.
• Turn the power off and unplug console before cleaning.
Footswitches
• DO NOT immerse the footswitch.
XPS® StraightShot® M4, StraightShot® Magnum® II and StraightShot® Magnum® Handpieces
• DO NOT immerse the handpieces.
• Remove and discard the tubing and blade/bur following local regulations for proper disposal of
contaminated materials.
Visao™ High-Speed Otologic Drill
• DO NOT immerse the handpiece.
• Remove bur from handpiece prior to cleaning.
Xcalibur® Hi-T or que Handpiece
• DO NOT immerse the handpiece attachments or the motor assembly .
• Remove bur from handpiece prior to cleaning.
• DONOT insert cleaning plug in the front of the motor assembly before sterilization.
• DO insert cleaning plug in the front of the motor assembly before cleaning.
Xcalibur® Hi-Speed Handpiece
• DO NOT immerse the handpiece attachments or the motor assembly .
• Remove bur from handpiece prior to cleaning.
• DONOT insert cleaning plug in the front of the motor assembly before sterilization.
• DO insert cleaning plug in the front of the motor assembly before cleaning.
Powerforma® Handpiece
• Remove bur from handpiece prior to cleaning.
• DO NOT immerse the handpiece.
Skeeter® Handpiece
• Remove bur from handpiece prior to cleaning.
• DO NOT immerse the handpiece.
Oto-Flex Burs
• DO NOT immerse the burs.
27
AFTER SURGERY (continued)
XPS Disposable Blades and Burs
• If the irrigation pump is not used, turn off the irrigation flow using the roller clamp on the irrigation tubing.
• Remove and discard the tubing following local regulations for proper disposal of contaminated materials.
• Depress the locking collar of the handpiece and pull the blade or bur out of the handpiece and discard following
local regulations for proper disposal of contaminated materials.
Medtronic Xomed Reusable Burs for Mastoid Drilling
• Initiate cleaning immediately after surgery (see cleaning section).
CLEANING
The following cycles are compatible with the each handpiece as listed below; however, it is the r esponsibility of
each health care facility to validate the sterilization process with their particular equipment and procedures to
assure sterility.
PRECAUTIONS
• DO NOT use organic solvents such as acetone or isopropyl alcohol to clean the bur chuck.
• USE ONLY an enzymatic detergent and distilled water for cleaning after every case or as required.
• DO NOT clean the handpieces in an ultrasonic cleaner or allow them to be fully immersed in any soaking
solution. These procedures may result in damaging the handpieces beyond repair.
XPS ®3000 Console
• Wipe down the console with standard disinfectant.
• Dry the console with a clean, non-abrasive cloth. Wipe dry before storing.
• Be sure to dry off the console after cleaning. Moisture inside the console could cause damage.
Footswitches
• DO NOT immerse the footswitch.
• Clean the outside of the footswitch by wiping with standard disinfectant.
XPS® StraightShot® M4, StraightShot® Magnum® II and StraightShot® Magnum® Handpieces
• DO NOT immerse the handpiece.
• Wipe the handpiece and cable with disinfectant applied to a clean, non-abrasive cloth.
• Gently clean the handpiece with a moistened soft bristle brush or pipe cleaner, making sure to clean all passages. Use an enzymatic detergent solution to loosen and remove collected tissues from the unit.
• Hold the handpiece with the front end pointed downward during rinsing.
• Dry the handpiece and cable with a lint-free towel. Make sure to dry off the electrical connection on the cable
ends.
• Apply a small amount of silicone spray into front-end collet and outside of handpiece.
Wipe dry before storing.
• Sterilize the handpieces immediately after cleaning.
Visao™ High-Speed Otologic Drill, Ir rigation Sleeves, and Bur Guar ds
• DO NOT immerse the handpiece.
• After surgery, clean the irrigation sleeves and bur guards with an enzymatic detergent solution. Wipe the
handpiece and cable with disinfectant applied to a clean, non-abrasive cloth.
• A chuck brush cleaner (REF# 3112500) or an appropriately sized small (plastic bristle) bore brush may be
inserted into the distal end of the Visao™ handpiece, irrigation sleeves and bur guards to assist in removing
fluids, tissue, or bone fragments.
• Rinse out the distal end of the handpiece. Shake excess water from the handpiece.
• Ensure all water is drained from the cooling housing. If saline was used for cooling during surgery, use distilled
water to rinse the housing prior to draining.
• Using distilled water, rinse saline from the irrigation nozzles. Drain the nozzle of all water.
• Sterilize the handpiece, irrigation sleeves, and bur guards immediately after cleaning.
28
CLEANING (continued)
Xcalibur® Hi-T or que and Hi-Speed Straight and Angled Handpiece Attachments, Irrigation
Sleeves, and Bur Guards
• DO NOT immerse the straight and angled handpiece attachments.
• After surgery, clean the irrigation sleeves and bur guards with an enzymatic detergent solution. Wipe the
straight and angled handpiece attachments with disinfectant applied to a clean non-abrasive cloth.
• A chuck brush cleaner (REF# 3112500) or an appropriately sized small (plastic bristle) bore brush may be
inserted into the distal end of the Xcalibur
in removing fluids, tissue, or bone fragments.
• Rinse out the distal end of the handpiece. Shake excess water from the handpiece attachments.
• Using distilled water, rinse saline from the irrigation nozzles. Drain the nozzle of all water .
• Sterilize the handpiece, irrigation sleeves, and bur guards immediately after cleaning.
Xcalibur® Hi-T or que Motor
• DO NOT immerse the motor assembly.
• Ensure that the cleaning plug is inserted into the front of the motor assembly .
• After surgery, wipe down motor cables and motor housing with disinfectant.
• If using a dishwasher, place handpiece in a vertical position (motor up) to allow water drainage.
• Sterilize the motor assembly immediately after cleaning.
Xcalibur® Hi-Speed Motor
• DO NOT immerse the motor assembly.
• Ensure that the cleaning plug is inserted into the front of the motor assembly .
• After surgery, wipe down motor cables and motor housing with disinfectant.
• Ensure all water is drained from the motors cooling housing. If saline was used for cooling during surgery , use
distilled water to rinse the housing prior to draining.
• If using a dishwasher, place handpiece in a vertical position (motor up) to allow water drainage.
• Sterilize the motor assembly immediately after cleaning.
®
handpiece attachments, irrigation sleeves and bur guards to assist
Powerforma® Handpiece
• DO NOT immerse the handpiece.
• IMPORT A NT NOTE: These procedures are critical to continued optimal performance. O.R. Personnel
should be familiar with the cleaning solutions and should use them routinely each time the handpiece is
used.
CLEANING CAUTIONS
• Always hold the handpiece with the nose pointing down to prevent any moisture from
reaching the motor .
• If using a dishwasher, place handpiece in a vertical position (motor up) to allow water drainage.
• DO NOT LUBRICA TE, SOAK, WASH OR SUBMERGE the handpiece in detergent or any solution.
• Always keep the handpiece as dry as possible to prevent corrosion, especially during sterilization and storage.
CLEANING HANDPIECE BUR CHUCK
• Set the maximum speed to the 25% setting.
• CAUTION: Cleaning at speeds over 25% may damage handpiece.
• Remove the bur and insert only the nose of the handpiece into 1 oz. of an enzymatic detergent solution.
• Run the handpiece for approximately 30 seconds.
• Rinse the bur chuck by running the handpiece in distilled water for approximately 30 seconds.
• Dry the inside of the bur chuck with a chuck brush (pipe cleaner, REF # 31 12500).
• Wipe off the handpiece with disinfectant and a soft cloth to remove blood and particles.
• Cover the nose of the handpiece with the nose protector. This attachment shields the chuck and irrigation tip.
• Sterilize the handpieces immediately after cleaning.
CLEANING TUBING
• Using distilled water, flush saline from reusable tubing and the tubing that runs through the handpiece. Clear
the tubing of all water by running in the pump until all water is removed. Autoclave the reusable tubing along
with the handpiece.
• Sterilize the tubing, along with the handpiece, immediately after cleaning.
OR
• If using disposable tubing, flush the portion of tubing that runs through the handpiece, and clear it of all water.
Then discard the disposable tubing.
• Use a fresh tubing set for the next surgical procedure.
29
CLEANING (continued)
Skeeter® Handpiece
• The handpiece may be cleaned and carefully rinsed with an enzymatic detergent. DO NOT FULL Y IMMERSE
OR UL TRASONICALL Y CLEAN THIS INSTRUMENT . The cannulated needle nose should be cleaned after
each use and prior to sterilization by immersing the distal end of the handpiece in an enzymatic detergent
solution to the level of the Bur Release Button. DO NOT USE ANY CLEANING INSTRUMENTS IN THE
CANNULA TED SHAFT OF THE HANDPIECE.
• To remove occasional residual buildup on handpiece cable connector, use a soft brush and isopropyl alcohol.
Rinsing:
• Immerse the distal end of the handpiece in distilled water up to the Bur Release Button and use a gentle swirling
motion to flush away residual cleaning solution. Avoid water accumulation in the motor housing by shaking
excess water droplets out with a downward motion.
Lubrication:
• Silicone spray should be sprayed into the cannulated shaft of the handpiece prior to sterilization. Apply
silicone spray until surplus silicone lubricant is noted on the outside of the Bur Release Button. Wipe away
excess lubricant from the handpiece. Following this procedure will insure that the bur release mechanism is well
lubricated for proper functioning.
• Sterilize the handpieces immediately after cleaning.
Oto-Flex Burs
Cleaning:
• The Skeeter® Ultra-Lite Oto-T ool Systems accepts Oto-Flex Burs. All Oto-Flex Burs may be cleaned either by
using an enzymatic detergent or by ultrasonic means. Rinse with distilled water to remove cleaning solution
residues and allow burs to dry. Following cleaning, apply a light coating of silicone spray in the following
manner: grasp the PTFE bearing and rotate the bur to assure application of the silicone spray inside the bearing.
Maintenance:
• During routine use normal wear may become evident on the bur and bur shaft. Examine burs under magnification for evidence of wear following each cleaning. Minor bowing of the bur shaft may be straightened by using
the fingers to rub and pull on the small shaft. Care should be exercised not to crimp or bend the small shaft
during handling or when straightening the shaft. Frequent replacement of burs is recommended to assure
optimum cutting performance and PTFE bearing concentricity. Burs exhibiting the following conditions should
be replaced: 1) nicks on cutting surfaces, 2) noticeable wear on PTFE bearings, 3) severe bends or crimps on bur
shaft, or 4) bur not running concentrically.
30
XPS® Disposable Blades and Burs
• Do Not reuse. Remove and discard following local regulations for proper disposal of contaminated
materials.
Medtronic Xomed Reusable Burs for Mastoid Drilling
CAUTION
• Burs should be removed from the handpiece for sterilization and drying.
• Monitor bur edges by viewing under microscope or magnifying glass. Dull burs increase pressure on the
handpiece. The motor may overheat and be damaged if continually overloaded.
Cleaning:
1 .Promptly and thoroughly rinse burs with deionized water after each use.
2. Soak in lukewarm*, mild* enzymatic detergent, and deionized water for a minimum of two minutes. Clean
ultrasonically in a lukewarm* solution of mild* detergent and deionized water for at least 30 seconds. Rinse
thoroughly with deionized water and wipe dry .
(*less than 43°C; pH 7.0 - 8.5)
Additional cleaning methods:
• . Presoaking in 3% hydrogen peroxide.
• . Additional thorough scrubbing of burs with a soft instrument brush while submerged in cleaning solution may
be warranted to remove blood and tissue.
STERILIZATION
The following cycles are compatible with the each handpiece as listed below; however, it is the r esponsibility of
each health care facility to validate the sterilization process with their particular equipment and procedures to
assure sterility.
WARNINGS
• DO NOT modify accessories used with any handpiece.
• Insertion of metal objects in blade or bur window may cause the blade or bur to break leaving fragments in the
wound which may be difficult to remove.
• Bending or prying may break the blade or bur, causing harm to patient or staff.
PRECAUTIONS
• DO NOT immerse the handpieces, as damage may result.
• After sterilizing, check electrical connectors for moisture. If moisture is present do not operate, as damage may
result. Do not operate or store the handpiece unless a drying cycle has been performed.
• DO NOT cold soak sterilize the handpiece in glutaraldehyde. This will void the warranty .
• Regardless of which type of steam sterilization is used, it is important that all electrical connectors of the handpiece be free of any moisture in the internal connection features. Damage may result to the instrument if it is
operated with moisture in the electrical connectors. Do not operate or store the handpiece unless a drying cycle
has been performed.
• Remove the handpiece from the sterilizer immediately after the sterilization cycle is complete.
• Increase temperatures higher than those stated when necessary to satisfy governmental or health care facility
requirements so long as the temperature does not exceed 149° C (300° F). Heating above 149° C (300° F) may
damage the handpiece and will void the warranty.
• DO NOT use organic solvents such as acetone or isopropyl alcohol to clean the bur chuck.
• Use only a mildly alkaline, low-sudsing detergent (Medtronic Xomed Cleaning Solution is recommended) and
distilled water for cleaning after every case or as required.
• Remove the bur from the Drill handpieces before sterilization.
• Store in a clean dry place.
• DO NOT clean handpieces in ultrasonic cleaner or allow them to be fully immersed in any soaking solution.
These procedures my result in damaging the handpieces beyond repair.
• DO NOT attempt to bend needle nose or effect repairs to either system.
• Remove the bur from the handpiece before sterilization.
• Sterilize immediately after cleaning. DO NOT STORE UNLESS A DRYING CYCLE HAS BEEN PERFORMED.
• Always wrap the electrical cord carefully inside the sterilizer tray during sterilization to prevent damage to the
cord when closing the tray lid.
• Regardless of which type of steam sterilization is used, it is extremely important that the handpiece is rapidly andcompletely dried before it is stored. If a vacuum drying cycle is not used following steam sterilization, moisture
may be trapped within the handpiece causing corrosion and residue deposits, resulting in premature wear and a
reduction in the functional life expectancy of the handpiece.
XPS® 3000 Console
• DO NOT attempt to sterilize.
Footswitches
• DO NOT attempt to sterilize.
31
STERILIZATION (continued)
XPS® StraightShot® M4, StraightShot® Magnum® II and StraightShot® Magnum® Handpieces
Steam S terilization — Gravity: W rapped — 121°C to 123°C (250°F to 254°F) for 40 minutes plus 8 minutes
minimum vacuum drying cycle.
Steam S terilization — Pre-Vac: Wrapped — 132°C to 135°C (270°F to 275°F) for 6 minutes minimum steam
cycle, plus 8 minutes minimum vacuum drying cycle.
Flash Autoclaving: AORN recommends for unwrapped flash cycles, a minimum of 10 minutes at 132°-135°C
(270°-275°F) in a gravity-displacement sterilizer, or 4 minutes at 132°-135°C (270°-275°F) in a pre-vacuum sterilizer. Medtronic Xomed recommends a minimum 8 minutes vacuum drying cycle.
Caution: Failure to perform the recommended vacuum drying cycle may result in premature wear of the handpiece.
STERRAD Sterilization: STERRAD compatible.
Ethylene Oxide: EtO compatible with the following cycle parameters:
T emperature: 54°C to 55°C (129°F to 131°F)
Relative Humidity: 35% minimum
V acuum Set Point: 1.3 psia
Preconditioning Time: 30 minutes
Minimum aeration period of twenty-four (24) hours at room temperature is recommended. Follow parameters
provided by the manufacturers of sterilizer equipment for each cycle. Alternately, mechanical aeration for
twelve (12) hours at 50°C ( 122°F), or 8 hours at 60°C (140 °F) is acceptable.
No matter which type of steam sterilization is used, it is extremely important that the handpiece be rapidlya nd completely dried before storage to prevent corrosion and residue deposits in the bearings and motor.
T emperature: 54°C to 55°C (129°F to 131°F)
Relative Humidity: 35% minimum
Chamber EtO Concentration: 575 to 625 mg/L
Exposure Duration: 2 hours
Steam S terilization — Gravity: W rapped — 121°C to 123°C (250°F to 254°F) for 40 minutes plus 8 minutes
minimum vacuum drying cycle.
Steam S terilization — Pre-Vac: Wrapped — 132°C to 135°C (270°F to 275°F) for 9 minutes minimum steam
cycle, plus 8 minutes minimum vacuum drying cycle.
Flash Autoclaving: AORN recommends for unwrapped flash cycles, a minimum of 10 minutes at 132°-135°C
(270°-275°F) in a gravity-displacement sterilizer, or 4 minutes at 132°-135°C (270°-275°F) in a pre-vacuum sterilizer. Medtronic Xomed recommends a minimum 8 minutes vacuum drying cycle.
Warning: Remove cleaning plug from the front of the motor assembly during sterilization.
Caution: Failure to perform the recommended vacuum drying cycle may result in premature wear of the hand-
piece.
STERRAD Sterilization: STERRAD compatible. (DO NOT USE STERRAD ON THE VISAO™ OR XCALIBUR
IRRIGA TION SLEEVE, THE VISAO™ OR XCALIBUR® EXTENDED BUR GUARDS WITH IRRIGA TION,
OR THE POWERFORMA® HANDPIECE.)
Ethylene Oxide: Handpieces are EtO compatible (see Preconditioning/Sterilization Parameters above). See
steam sterilization instructions for the reusable tubing used with the Powerforma Handpieces. Minimum aeration period of twenty-four (24) hours at room temperature is recommended. Follow parameters provided by the
manufacturers of sterilizer equipment for each cycle. Alternately , mechanical aeration for twelve (12) hours at
50°C (122 °F), or 8 hours at 60°C (140 °F) is acceptable.
®
32
No matter which type of steam sterilization is used, it is extremely important that the handpiece be rapidly andcompletely dried before storage to prevent corrosion and residue deposits in the bearings and motor . Use of a
minimum 8 minutes vacuum drying cycle is recommended.
STERILIZATION (continued)
Skeeter® Handpiece
Gravity:
Wrapped; 132°C to 135°C (270°F to 275°F) for 25 minutes, plus 8 minutes minimum vacuum drying cycle.
OR
Wrapped; 121°C to 135°C (250°F to 275°F) for 30 minutes, plus 8 minutes minimum vacuum drying cycle.
Pre-Vac: Wrapped; 132°C to 135°C (270°F to 275°F) for 6 minutes minimum steam cycle, plus 8 minutes
minimum vacuum drying cycle.
Flash Autoclaving: Unwrapped: A minimum of 3 minutes at a maximum temperature of 140°C (284°F) advised
only when emergency action is required (e.g.; a break in sterile technique). Immediately following the surgical
procedure, the handpiece should be thoroughly cleaned and sterilized with a drying cycle, preferably with
vacuum.
Ethylene Oxide: EtO compatible (see Preconditioning/Sterilization Parameters above). Minimum aeration period of twenty-four (24) hours at room temperature is recommended. Follow parameters provided by the manufacturers of sterilizer equipment for each cycle. Alternately , mechanical aeration for twelve (12) hours at 50°C
(122°F), or 8 hours at 60°C (140°F) is acceptable.
STERRAD Sterilization: STERRAD compatible.
No matter which type of steam sterilization is used, it is extremely important that the handpiece be rapidly andcompletely dried before storage to prevent corrosion and residue deposits in the bearings and motor .
Oto-Flex Burs
Medtronic Xomed drill systems and burs may be sterilized by any one of several properly validated methods.
The method of sterilization selected by individual health care facilities is dependent upon the capabilities of
each facility . IT IS THE USER’S RESPONSIBILITY TO VALIDATE THE STERILIZA TION PROCESS
TO ENSURE STERILITY. The following guidelines are recommended and are compatible with the burs.
STEAM STERILIZA TION
Gravity:
Wrapped; 132°C to 135°C (270°F to 275°F) for 25 minutes, plus 8 minutes minimum vacuum drying cycle.
OR
Wrapped; 121°C to 123°C (250°F to 253°F) for 30 minutes, plus 8 minutes minimum vacuum drying cycle.
Prevac: Wrapped: 132°C to 135°C (270°F to 275°F) for 6 minutes minimum steam cycle, plus 8 minutes minimum
vacuum drying cycle.
Flash Autoclaving: Unwrapped: A minimum of 3 minutes at a maximum temperature of 140°C (284°F) ADVISED
ONL Y WHEN EMERGENCY ACTION IS REQUIRED (e.g., a break in sterile technique). IMMEDIATELY
following the surgical procedure, the bur should be thoroughly cleaned and sterilized with a drying cycle,
preferably with vacuum. Flash sterilization should follow 2000 AORN Standards, Recommended Practices,
and Guidelines.
Ethylene Oxide Sterilization: EtO compatible (see Preconditioning/S terilization Parameters above). A minimum
aeration period of twenty-four (24) hours is recommended following EO sterilization. The parameters established by the sterilizer equipment manufacturer should be carefully followed for each cycle.
STERRAD Sterilization: STERRAD compatible.
XPS® Disposable Blades and Burs
• Limit use of disposable blades, burs and tubing to only one time, unless otherwise marked. Do not attempt to
sterilize disposable parts. Disposable, single-use parts are packaged sterile and are not intended for repeated
use.
33
STERILIZATION (continued)
Medtronic Xomed Reusable Burs for Mastoid Drilling
• Before sterilization, carefully inspect the bur tips, bur flutes or diamond tips under a microscope or magnifying
glass for any irregularities or eccentricities. Discard any burs which show signs of damage or wear .
• DO NOT COLD SOAK IN GLUTARALDEHYDE, CHLORINE, OR AMMONIUM SOLUTIONS OR DRY
HEA T STERILIZE AS DAMAGE T O THE BURS MA Y OCCUR.
The following cycle has been validated by Medtronic Xomed and is compatible with the products; however, it
is the responsibility of each health care facility to validate the sterilization process with their particular
equipment and procedures to assure sterility.
Steam Sterilization – Gravity: Wrapped – 121°C to 123°C (250°F to 253°F) for 40 minutes plus 8 minutes
minimum vacuum drying cycle.
Steam S terilization – Pre-V ac: W rapped - 132°C to 135°C (270°F to 275°F) for 9 minutes minimum steam cycle,
plus 8 minutes minimum vacuum drying cycle.
STERILIZATION REFERENCES
• AAMI Standards and Recommended Practices — Volume 1.1: Sterilization; Part 1-Good Hospital Practices,
Association for the Advancement of Medical Instrumentation, 1996. PH: 703-525-4890.
• AAMI Standards and Recommended Practices — Volume 1.2: Sterilization; Part 2-Hospital Equipment andIndustrial Process Contr ol, Association for the Advancement of Medical Instrumentation, 1996.
PH: 703-525-4890.
• 2000 AORN Standards,Recommended Practices, and Guidelines (Denver: AORN (Association of periOperative
Registered Nurses), 2000). PH: 303-755-6300.
• EN 550 Sterilization of Medical Devices, Validation and Routine control of EtO Sterilization.
• EN 554 Sterilization of Medical Devices, Validation and Routine control of Sterilization by Moist Heat.
34
TROUBLESHOOTING
SYSTEM MALFUNCTIONS
CODEERROR
2......................... Handpiece Fault
3......................... +5V Failure
4......................... +12V Failure
5......................... 48V Failure
7......................... Watchdog T imer Failure
IF ......... Any of the above error codes are displayed...
• Cycle power; i.e. turn the power off and back on.
• Call Customer Service.
IF .........Console lights fail to illuminate...
• Ensure that the power switch on the rear of the console is in the “|” position.
• Inspect power cord connections, and ensure that power cord is firmly seated.
• Inspect two fuses next to the power connector, and if a fuse appears to be blown, remove and replace
it with a 4.0A, 250V, Type F fuse.
BLADE OR BUR MALFUNCTIONS
IF ......... Upon inspection, blade or bur appears to be damaged or defective...
• Remove and replace defective blade or bur.
IF ......... Blade or bur is not firmly seated...
• Pull back locking collar and reseat the blade or bur.
IF ......... Blade opening is obstructed...
• Remove blade from surgical site and submerse the blade tip in sterile water with suction connected to the
handpiece to evacuate the obstruction, or use stylette to clear blade. The suction tube may be removed
from the handpiece and inserted over the end of the blade and blade activated to evacuate the obstruction from the tip of the blade.
IF ......... Blade or bur is leaking irrigant...
• Check for proper blade/bur insertion by pulling back locking collar, and reseating blade/bur .
• Check blade for obstruction. If visible, evacuate obstruction by one of the above methods.
• Inspect suction connection on handpiece and at suction cannister, and if suction tubing is
disconnected, connect it.
• Remove and inspect suction tubing, and if obstructed, remove obstruction.
IF ......... Blade or bur continues to malfunction...
• Remove and replace blade or bur.
• Call Customer Service.
IF ......... Burs wobble in Xcalibur® or Powerforma® Handpieces...
• Reduce handpiece operating speeds.
• Use burs that are rated for the console speed selected.
• If necessary, use extended bur guards with Xcalibur® drills for 64 mm and 75 mm burs.
• Operate handpiece at 50% of full speed for 64 mm and 75 mm burs.
IF ......... Console lights illuminate but handpiece fails to rotate...
IF ......... Handpiece defaults to incorrect speed...
HANDPIECE MALFUNCTIONS
• Inspect handpiece cable connections, and ensure that handpiece cable connector is firmly seated.
• If handpiece continues to malfunction, call Customer Service.
• Dry handpiece per sterilization cycle recommendations and try again. If handpiece continues to
malfunction, call Customer Service.
FOOTSWITCH MALFUNCTIONS
IF ......... Console displays operation speed percentages and handpiece does not operate properly...
• Inspect footswitch cable connection, and ensure that footswitch cable connector is firmly seated.
IF ......... Footswitch continues to malfunction...
• Disconnect footswitch and operate handpiece with “Manual Start Stop” button on the rear console.
• Call Customer Service.
IF ......... Handpiece operates without activating the footswitch...
• Call Customer Service.
36
LIMITED WARRANTY
A. This LIMITED WARRANTY provides assurance for the customer who purchases an XPS® Model 3000 System (herein-
after the “Product”) that should the Product fail to function to Medtronic Xomed’s published specifications during the
term of this LIMITED WARRANTY (one year from the date of shipment for new Product, 90 days from date of shipment
for refurbished or used Product), Medtronic Xomed will either replace, repair, or issue a credit (adjusted to reflect the age
of the Product) for the Product or any portion thereof. This LIMITED WARRANTY is extended only to the buyer
purchasing the Product directly from Medtronic Xomed or from its affiliate or its authorized distributor or representative.
B . To qualify for this LIMITED WARRANTY, the following conditions must be met:
(1) The Product must be used on or before its “Use By” or “Use Before” date, if applicable.
(2) The Product must be used in accordance with its labeling and may not be altered or subjected to misuse, abuse,
accident or improper handling.
(3 ) Medtronic Xomed must be notified in writing within thirty (30) days following discovery of a defect.
(4 ) The Product must be returned to Medtronic Xomed within thirty (30) days of Medtronic Xomed receiving notice as
provided for in (3) above.
(5 ) Upon examination of the Product by Medtronic Xomed, Medtronic Xomed shall have determined that: (i) the
Product was not repaired or altered by anyone other than Medtronic Xomed or its authorized representative, (ii) the
Product was not operated under conditions other than normal use, and (iii) the prescribed periodic maintenance and
services have been performed on the Product.
C .This LIMITED W ARRANTY is limited to its express terms. THIS LIMITED W ARRANTY IS IN LIEU OF ALL OTHER
W ARRANTIES, EXPRESSED OR IMPLIED WHETHER ST A TUTOR Y OR OTHERWISE, INCLUDING ANY IMPLIED W ARRANTY OF MERCHANT ABILITY OR FITNESS FOR A PAR TICULAR PURPOSE. In no event shall
Medtronic Xomed be liable for any consequential, incidental, prospective or other similar damage resulting from a
defect, failure, or malfunction of the Product, whether a claim for such damage is based upon the warranty, contract,
negligence or otherwise.
D. The exclusions and limitations set out above are not intended to, and should not be construed so as to, contravene
mandatory provisions of applicable law . Users may benefit from statutory warranty rights under legislation governing
the sale of consumer goods. If any part or term of this LIMITED WARRANTY is held by any court of competent
jurisdiction to be illegal, unenforceable, or in conflict with applicable law , the validity of the remaining portion of the
LIMITED WARRANTY shall not be affected, and all rights and obligations shall be construed and enforced as if this
LIMITED WARRANTY did not contain the particular part or term held to be invalid.
CAUTION
Applicable law may restrict the sale, distribution or use of this device to, by or on the order of a licensed medical practitioner.
RETURNS AND/OR REPAIRS
Contact Medtronic Xomed Customer Service at (800-874-5797) to obtain a Return Goods Authorization number (RGA#) prior
to shipping the Product to Medtronic Xomed. Please have the original invoice number or purchase order number available
to assist in verifying warranty information. The RGA# should be prominently displayed on the box and included on all
paperwork enclosed with the return and/or repair. All Product returned to Medtronic Xomed should be safely packed in
protective wrapping.
Customer must supply the Purchase Order number; the correct shipping and billing address; and either a completed Repair
Order Form or a statement of the problem or reason for return.
37
TECHNICAL SPECIFICATIONS
HANDPIECES - XPS® StraightShot® M4 Microdebrider /
StraightShot® Magnum® II / LandmarX® Magnum® II / XPS
StraightShot® III
®
Part No.1898200TXPS® StraightShot® M4 Microdebrider
1897200StraightShot
1897200TLandmarX® Magnum® II Handpiece
1897201XPS
Speed500-5,000 rpm oscillate
500-12,000 rpm forward
Size14.3 cm length x 1.8 cm width (1898200T)
17 cm length x 1.6 cm diameter (1897200, 1897200T , and 1897201)
Weight228 g1898200T
240 g1897200
254 g1897200T
240 g1897201
Duty CycleThe XPS
StraightShot® III Handpieces under full load are rated for intermittent operation per the
following:
Maximum On Time 60 seconds
Minimum Off Time 30 seconds
®
StraightShot® M4, StraightShot® Magnum® II, LandmarX® Magnum® II, and XPS
operation per the following:
Maximum On Time 60 seconds
Minimum Off Time 30 seconds
MPS® POWERFORMA® DRILL
Part No.3327700Straight Handpiece
3327750Angled Handpiece
Speed10,000-52,000 rpm forward/reverse
Size19.1 cm length x 2.9 cm diameter3327700
19.7 cm length x 2.9 cm diameter3327750
Weight165 g3327700
182 g3327750
Duty CycleThe Powerforma® Handpieces under full load are rated for intermittent operation per the
following:
Maximum On Time 60 seconds
Minimum Off Time 30 seconds
SKEETER® ULTRA-LITE OTO-TOOL
®
Drill under full load are rated for intermittent
Part No.3055601
Speed12,000 rpm forward
Size17 cm length x 1.6 cm diameter
Weight57 g
Duty CycleContinuous run
CONSOLE
Part No.1897101XPS® Model 3000 Console W ith Irrigation (one pump)
ElectricalInput Power100-240 V AC, 47-63 Hz, 4 A
Size20 cm Width x 28 cm Height x 18.5 cm Depth (1897101 and 1897102)
Weight4.5 kg1897101
ClassClass I Medical Device per EN 60601-1
1897102XPS® Model 3000 Console W ith Coolant and Irrigation (two pumps)
5.0 kg1897102
39
XPS® MULTIFUNCTION FOOTSWITCH
Part No. 1895400
Size18 cm Length x 18.5 cm Width x 6.3 cm Height
Weight1.8 kg
Protection Against Water Ingress IPX8
XPS® MULTIFUNCTION FOOTSWITCH, SINGLE PEDAL
Part No. 1895420
Size18 cm Length x 18 cm Width x 3.8 cm Height
Weight1.8 kg
Protection Against Water Ingress IPX8
XPS® SINGLE FUNCTION FOOTSWITCH
Part No. 1895410
Size11.5 cm Length x 8.9 cm W idth x 5.6 cm Height
Weight0.8 kg
Protection Against Water Ingress IPX8
COMPLIANCE
Meets requirements of the following standards:
EN 60601-1:1990 with A1 and A12:1993, A2: 1995 and A13:1996
IEC 60601-1, Second Edition, 1988 with Amendment 1, 1991 and Amendment 2, 1995
UL 2601-1 2nd Ed., 1997
CAN/CSA-C22.2 No. 601.1-M90
IEC 60601-1-2:2001
EN 60601-1-2:2001
EC Certification: EC Cert. No. G1 00 12 41921 002
CB T est Certificate: US/801/ITS-M1
40
Guidance and manufacturer’s declaration – electromagnetic immunity - Part I
Guidance and manufacturer’s declaration – electromagnetic immunity - Part I
The XPS® 3000 is intended for use in the electromagnetic environment specified below.
The customer or the user of the XPS
Immunity test
Electrostatic discharge
(ESD)
IEC 61000-4-2
Electrical fast transient/
burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and voltage
variations on power
supply lines
IEC 61000-4-11
®
3000 should assure that it is used in such an environment.
IEC 60601 test level
±6 kV contact
±8 kV air
±2 kV for power supply lines
±1 kV for input/output lines
±1 kV differential mode
±2 kV common mode
T for 0.5 cycle
<5 % U
40 % UT for 5 cycles
70 % UT for 25 cycles
<5 % UT for 5 sec
Compliance level
±6 kV contact
±8 kV air
±2 kV for power supply lines
±1 kV for input/output lines
±1 kV differential mode
±2 kV common mode
<5 % UT for 0.5 cycle
40 % UT for 5 cycles
70 % UT for 25 cycles
<5 % UT for 5 sec
Electromagnetic
environment - guidance
Floors should be wood,
concrete, or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be at
least 30 %.
Mains power quality should
be that of a typical commercial or hospital environment.
Mains power quality should
be that of a typical commercial or hospital environment.
Mains power quality should
be that of a typical commercial or hospital environment.
If the user of the XPS® 3000
requires continuous operation
during power mains interruptions, it is recommended that
the XPS® 3000 be powered
from an uninterruptible
power supply or a battery.
Power frequency (50/60
3 A/m
3 A/m
Hz) magnetic field
IEC 61000-4-8
Note: UT is the a.c. mains voltage prior to application of the test level.
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical commercial or hospital environment.
41
Guidance and manufacturer’s declaration – electromagnetic emissions
Guidance and manufacturer’s declaration – electromagnetic emissions
Emissions test
RF emissions
CISPR 11
The XPS® 3000 is intended for use in the electromagnetic environment specified below.
The customer or the user of the XPS
Compliance
Group 1
®
3000 should assure that it is used in such an environment.
Electromagnetic environment - guidance
®
The XPS
3000 uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Class B
Class A
The XPS
®
3000 is suitable for use in all establishments other than
domestic and those directly connected to the public
low-voltage power supply network that supplies buildings for
domestic purpose.
V oltage fluctuations
Complies
IEC 61000-3-3
Recommended separation distances between portable and mobile RF communications equipment
and the XPS® 3000
Recommended separation distances between portable and mobile RF communications equipment and the XPS® 3000
The XPS® 3000 is intended for use in the electromagnetic environment in which radiated RF disturbances are controlled. The
customer or the user of the XPS® 3000 can help prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the XPS® 3000 as recommended below , according to the
maximum output power of the communications equipment.
Rated maximum power of transmitter
Separation distance according to frequency of transmitter
(meters)
W
0.01
0.1
1
10
100
150 kHz to 80 MHz
d = 1.2 P
√
0.12
0.38
1.2
3.8
12
80 MHz to 800 MHz
d = 1.2 P
0.12
0.38
1.2
3.8
12
800 MHz to 2.5 GHz
d = 2.3 P
√√
0.23
0.73
2.3
7.3
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be
estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
42
Guidance and manufacturer's declaration - electromagnetic immunity - Part II
Guidance and manufacturer’s declaration – electromagnetic immunity - Part II
Immunity test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
The XPS® 3000 is intended for use in the electromagnetic environment specified below.
The customer or the user of the XPS
IEC 60601 test level
®
3000 should assure that it is used in such an environment.
Compliance level
Electromagnetic environment - guidance
Portable and mobile RF communications equipment should be used no closer to any part of the
®
XPS
3000, including cables, than the recommended separation distance calculated from the
equation applicable to the frequency of the
transmitter.
Recommended separation distance
3 Vrms
3 Vrms
d = 1.2 P
√
150 kHz to 80 MHz
3 V / m
80 MHz to 2.5 GHz
3 V / m
d = 1.2 P 80 MHz to 800 MHz
√
d = 2.3 P 800 MHz to 2.5 GHz
√
where P is the maximum output power rating of
the transmitter in watts (W) according to the
transmitter manufacturer and d is the recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
should be less than the compliance level in each
frequency range.
b
Interference may occur in the vicinity of equipment marked with the following symbol:
a
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios,
amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured
field strength in the location in which the XPS
should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such
as re-orienting or relocating the XPS
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
®
3000.
®
3000 is used exceeds the applicable RF compliance level above, the XPS® 3000
RECOMMENDED ENVIRONMENTAL CONDITIONS
Operating
T emperature:................................... +10°C to +40°C (+50°F to +104°F)
Humidity: ........................................ 30% to 75% RH