Medtronic X0610003 Instructions for Use

Crescent™ Spinal System 0381444E Rev. C
2020-09-30
IMPORTANT INFORMATION ON THE CRESCENT™ SPINAL SYSTEM
PURPOSE
This device is a fusion device intended for stabilization and to promote bone fusion during the normal healing process following surgical correction of disorders of the spine. The product should be implanted only by a physician thoroughly knowledgeable in the implant's material and surgical aspects and instructed as to its mechanical and material applications and limitations.
DESCRIPTION
Implied warranties of merchantability and fitness for a particular purpose or use are specifically excluded.
INDICATIONS
The Crescent™ Spinal System is indicated for interbody fusion with autogenous bone and/or allograft bone graft comprised of cancellous and/or corticocancellous bone graft, and/or demineralized allograft bone with bone marrow aspirate in patients with Degenerative Disc Disease (DDD) at one or two levels from L2 to S1. These DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non­operative treatment. Implants are to be used with autogenous bone and/or allograft bone graft comprised of cancellous and/or corticocancellous bone graft, and/or demineralized allograft bone with bone marrow aspirate. These devices are intended to be used with supplemental fixation instrumentation cleared by the FDA for use in the lumbar spine.
CONTRAINDICATIONS
This device is not intended for cervical spine use. Contraindications include:
Infection local to the operative site.Signs of local inflammation.Fever or leukocytosis.Morbid obesity.Pregnancy.Mental illness.Any other condition which would preclude the potential benefit of spinal implant surgery, such as the presence of tumors or
congenital abnormalities, fracture local to the operating site, elevation of sedimentation rate unexplained by other diseases, elevation of white blood count (WBC), or a marked left shift in the WBC differential count.
Suspected or documented allergy or intolerance to composite materials.Cases not needing a fusion.Cases not described in the indications.Patients unwilling to cooperate with postoperative instructions.Patients with a known hereditary or acquired bone friability or calcification problem should not be considered for this type of
surgery.
Pediatric cases, nor where the patient still has general skeletal growth.Spondylolisthesis unable to be reduced to Grade 1.Cases where implant components selected for would be too large or too small to achieve a successful result.Cases that require the mixing of metals from two different components or systems.Patients having inadequate tissue coverage over the operative site or inadequate bone stock or quality.Patients in which implant use would interfere with anatomical structures or expected physiological performance.Prior fusion at the level to be treated.
Nota bene: although not absolute contraindications, conditions to be considered as potential factors for not using this device include:
Severe bone resorption.
Osteomalacia.Severe osteoporosis.
POTENTIAL ADVERSE EVENTS
Adverse effects may occur when the device is used either with or without associated instrumentation. Risk of adverse effects as a result of movement and non-stabilization may increase in cases where associated complementary support is not employed. Potential adverse events include:
Implant migration.Breakage of device.Foreign body reaction to implants including possible tumor formation, auto immune disease, and/or scarring.Pressure on the surrounding tissues or organs.Loss of proper spinal curvature, correction, height, and/or reduction.Infection.Bone fracture or stress shielding at, above, or below the level of surgery.Non-union (or pseudoarthrosis).Loss of neurological function, appearance of radiculopathy, dural tears, and/or development of pain.Neurovascular compromise including paralysis, temporary or permanent retrograde ejaculation in males, or other types of
serious injury.
Cerebral spinal fluid leakage.Haemorrhage of blood vessels and/or hematomas.Discitis, arachnoiditis, and/or other types of inflammation.Deep venous thrombosis, thrombophlebitis, and/or pulmonary embolus.Bone graft donor site complication.Inability to resume activities of normal daily living.Early or late loosening or movement of device.Urinary retention, loss of bladder control, or other types of urological system compromise.Scar formation possibly causing neurological compromise or compression around nerves and/or pain.Fracture, damage, or penetration of any spinal bone (including the sacrum, pedicles, and/or vertebral body) and/or bone
graft or bone graft harvest site at, above, and/or below the level of surgery.
Retropulsed graft.Herniated nucleus, disc disruption or degeneration at, above, or below the level of surgery.Loss of or increase in spinal mobility or function.Reproductive system compromise, including sterility, loss of consortium, and sexual dysfunction.Development of respiratory problems (e.g. pulmonary embolism, bronchitis, pneumonia, etc.).Change in mental status.Cessation of any potential growth of the operated portion of the spine.Death.
WARNINGS AND PRECAUTIONS
Do not reuse or reprocess devices labeled as single use devices. Reuse or reprocessing of a single use devices may compromise the structural integrity and the intended function of the device which could result in patient injury. A successful result is not always achieved in every surgical case. This fact is especially true in spinal surgery where other patient conditions may compromise results. Use of this product without autogenous bone and/or allograft bone graft comprised of cancellous and/or corticocancellous bone graft, and/or demineralized allograft bone with bone marrow aspirate or in cases that do not develop a union will not be successful.
Preoperative and operating procedures, including knowledge of surgical techniques, good reduction, and correct selection and placement of implants are important considerations in successful use of the system. Further, proper selection and compliance of the patient will greatly affect results. Patients who smoke were shown to have a reduced incidence of bone fusion. These patients should be advised of this fact and warned of this consequence. Obese, malnourished, and/or alcohol/drug abuse patients and those with poor muscle and bone quality and/or nerve paralysis are also poor candidates for spinal fusion.
Patients with a previous spinal surgery at the levels to be treated may have different clinical outcomes compared to those without a previous spinal surgery.
Physician note: although the physician is the learned intermediary between the company and the patient, the important medical information in this document should be conveyed to the patient.
For US Audiences Only
Caution: Federal law (USA) restricts these devices to sale by or on the order of a physician.
MR CONDITIONAL
The Crescent™ Spinal System was determined to be MR-Conditional based on non-clinical testing and engineering rationales. A patient with this device can be safely scanned immediately after device placement under the following conditions:
STATIC MAGNETIC FIELD
Static magnetic field of 1.5 Tesla and 3-Tesla.
Maximum spatial gradient magnetic field of 3000-Gauss/cm (30 T/m) or less.Maximum whole body average specific absorption rate (SAR) of 2.0 W/kg or less underNormal Operating Mode, for 15 minutes of scanning per pulse sequence.
Under the scan conditions defined, a worst-case interbody fusion device representative of the Crescent™ Spinal System produced a maximum temperature rise of 4.4°C after 15 minutes of continuous scanning.
In non-clinical testing, the image artifact caused by a worst-case interbody fusion device representative of the Crescent™ Spinal System extends approximately 20 mm from the device when imaged with a gradient echo pulse sequence in a 3.0 Tesla MR system. Therefore, optimization of MR imaging parameters to compensate for the presence of this device may be necessary.
If Crescent™ Spinal System is used in connection with any device which is not MR Conditional, please be advised that this combination has not been tested in the MR environment and, therefore, higher heating and possible injury to the patient may occur. The presence of other implants or the health state of the patient may require a modification of the MR conditions.
IMPLANT SELECTION
Selection of proper size, shape, and design of the implant for each patient is crucial to success of the procedure. Surgical implants are subject to repeated stresses in use and their strength is limited by the need to adapt the design to the human anatomy. Unless great care is taken in patient selection, placement of the implant, and postoperative management to minimize stresses on the implant, such stresses may cause material fatigue and consequent breakage or loosening of the device before the fusion process is complete, which may result in further injury or the need to remove the device prematurely.
DEVICE FIXATION
Installation and positional adjustment of implants must only be done with special ancillary instruments and equipment supplied by Medtronic. In the interests of patient safety, it is therefore recommended that Medtronic implants are not used with devices from any other source.
Never, under any circumstances, reuse a Crescent™ Spinal System device. Even when a removed device appears undamaged, it may have small defects or internal stress patterns that may lead to early breakage.
PREOPERATIVE
Only patients that meet the criteria described in the indications should be selected.Patient conditions and/or predispositions such as those addressed in the contraindications should be avoided.Care should be taken when handling and storing of the device. They should not be scratched or damaged. Devices should
be protected during storage especially from corrosive environments.
The surgeon should be familiar with the various devices before use and should personally verify all devices are present
before surgery.
The size of device for the case should be determined prior to surgery. An adequate inventory of implant sizes should be
available at the time of surgery, including sizes larger and smaller than those expected to be used.
Unless supplied sterile, all devices should be cleaned and sterilized before use. Additional sterile components should be
available in case of any unexpected need.
INTRAOPERATIVE
Instructions in any available Crescent™ Spinal System surgical technique manual should be carefully followed.At all times, extreme caution should be used around the spinal cord and nerve roots. Damage to nerves will cause loss of
neurological functions.
Breakage, slippage, or misuse of instruments or implants may cause injury to the patient or operative personnel.To ensure proper fusion below and around the location of the fusion, autogenous bone and/or allograft bone graft
comprised of cancellous and/or corticocancellous bone graft, and/or demineralized allograft bone with bone marrow aspirate must be used.
When used via a posterior approach, supplemental posterior instrumentation is recommended. Posterior supplemental
fixation is limited to Medtronic posterior instrumentation systems.
Bone cement should not be used because this material may make removal of these components difficult or impossible.
Heat generated from the curing process may damage or deform PEEK devices.
POSTOPERATIVE
The physician’s postoperative directions and warnings to the patient and the corresponding patient compliance are extremely important.
Detailed instructions on use and limitations of the device should be given to the patient. The patient must be warned that
loosening and/or breakage of the device are complications which may occur as result of early or excessive weight-bearing muscular activity or sudden jolts or shock to the spine.
The patient should be advised not to smoke or consume excess alcohol during period of the bone fusion process.The patient should be advised of the inability to bend at the point of spinal fusion and taught to compensate for this
permanent physical restriction in body motion.
It is important that immobilization of union is established and confirmed by roentgenographic examination. If a non-union
develops or if components loosen, migrate, and/or break, the devices should be revised and/or removed immediately before serious injury occurs.
Crescent™ Spinal System implants are interbody devices and are intended to stabilize the operative area during the fusion
process.
Retrieved devices should be treated in such a manner that reuse in another surgical procedure is not possible.
PACKAGING
Packages for each of the components should be intact upon receipt. If a loaner or consignment system is used, all sets should be carefully checked for completeness, and all devices including instruments should be carefully checked to ensure there is no damage prior to use. Damaged packages or products should not be used and should be returned to Medtronic.
STERILIZATION
Only sterile products should be placed in the operative field. Unless marked sterile and clearly labeled as such in an unopened sterile package provided by Medtronic, implants used in surgery must be sterilized by the hospital prior to use. Unless specified elsewhere, these products are recommended to be steam sterilized by the hospital using one of the following sets of process parameters:
Method Cycle Temperature Exposure time Dry time
Steam Pre-Vacuum 270°F (132°C) 4 Minutes 30 Minutes
Steam Gravity 250°F (121°C) 60 Minutes 30 Minutes Steam* Pre-Vacuum* 273°F (134°C)* 20 Minutes 30 Minutes Steam* Gravity* 273°F (134°C)* 20 Minutes 30 Minutes
Note: because of the many variables involved in sterilization, each medical facility should calibrate and verify the sterilization process (e.g. temperatures, exposure times) used for their equipment.
*For outside the US, some non-U.S. health care authorities recommend sterilization according to these parameters to minimize the risk of transmission of Creutzfeldt-Jakob disease, especially of surgical instruments that could come into contact with the central nervous system.
PRODUCT COMPLAINTS
To report any product problems, contact Medtronic.
FURTHER INFORMATION
Recommended directions for use of this system (surgical operative techniques) are available at no charge upon request. If further information is required, contact Medtronic.
©2020 Medtronic Sofamor Danek USA, Inc. All rights reserved.
Medtronic Sofamor Danek USA, Inc.
1800 Pyramid Place Memphis, TN 38132 Telephone: 800 933 2635 (USA)
901 396 3133 (Outside USA)
Fax: 901 396 0356
EXPLANATION OF SYMBOLS
CAUTION: Federal law (USA) restricts these devices to sale by or on the order of a physician.
Do not re-use
Batch code
Manufacturer
Date of Manufacture
Catalogue number
For US audiences only
Sterilized using irradiation
Use-by date
Consult instructions for use at this website.
MR Conditional
Loading...