Medtronic W1DR01 Errata

Errata Sheet for the Model SW030 Software for the Medtronic 2090 CareLink Programmer and 29901 CareLink Encore Programmer
The following table lists certain conditions that may occur during use of the software and the resolution of these conditions.
Condition Observed Behavior Suggested Corrective Action
When Atrial Sensitivity is disabled in a dual chamber device, it is still possible to do an Atrium 50 HZ Burst EP Study test but the device will not detect the atrial arrhythmia.
The following list includes trademarks or registered trademarks of Medtronic in the United States and possibly in other countries. All other trademarks are the property of their respective owners.
Astra, Azure, CareLink, CareLink Encore, Medtronic, Medtronic CareLink, SureScan
Caution: Federal law (USA) restricts this device to sale by or on the order of a physician.
An atrial sensed event will not appear on the Live Rhythm Monitor screen. If an atrial lead is present, running the test would introduce an atrial arrhythmia that is not being detected and reported.
Determine whether a pin plug or atrial lead is being used by checking the Patient Information screen or running Lead Impedance Test. If a pin plug is being used, the patient is unaffected and no further action is needed. If an atrial lead is present, determine if Atrial Sensitivity should continue to be disabled. If not, program Atrial Sensitivity to a value other than Off to run the 50 HZ Burst test.
© 2016 Medtronic M970718A001 A 2016-11-15