Tr eo n® Treatment Guidance System Manual 2-17
System Overview
System Electromagnetic Emissions and Immunity Declarations
Table 2- 6. Guidance and Manufacturer's Declaration - Electromagnetic Immunity IEC 60601-1-2: 2001, Table 202
The Treon® Treatment Guidance System is intended for use in the electromagnetic environment specified below. The customer
or the user of the Treon
®
Treatment Guidance System should assure that it is used in such an environment.
Immunity Test IEC 60601 Test
Level
Compliance Level Electromagnetic Environment - Guidance
Electrostatic Discharge
(ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV Air
± 6 kV contact
± 8 kV Air
Floors should be wood, concrete, or ceramic tile. If
floors are covered with synthetic material, the relative
humidity should be at least 30%.
Electrical Fast
Transient/Burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for input/output
lines
± 2 kV for power
supply lines
± 1 kV for input/output
lines
Mains power quality should be that of a typical
commercial or hospital environment.
Surge
IEC 61000-4-5
± 1 kV Differential
Mode
± 2 kV Common Mode
± 1 kV Differential
Mode
± 2 kV Common Mode
Mains power quality should be that of a typical
commercial or hospital environment.
Voltage Dips, Short
Interruptions, and
Voltage Variations on
Power Supply Input
Lines
IEC 61000-4-11
< 5% UT
(> 95% dip in UT)
for 0.5 cycle
< 5% UT
(> 95% dip in UT)
for 0.5 cycle
Mains power quality should be that of a typical
commercial or hospital environment. If the user of the
Treon
®
Treatment Guidance System requires
continued operation during power mains
interruptions, it is recommended that the Treon®
Treatment Guidance System be powered from an
uninterruptible power supply or a battery.
40% UT
(60% Dip in UT)
for 5 cycles
40% UT
(60% Dip in UT)
for 5 cycles
70% UT
(30% dip in UT)
for 25 cycles
70% UT
(30% dip in UT)
for 25 cycles
<5% UT
(> 95% Dip in UT)
for 5 sec
<5% UT
(> 95% Dip in UT)
for 5 sec
Power Frequency
(50/60 Hz) Magnetic
Field
IEC 61000-4-8
3 A/m 3 A/m Power Frequency Magnetic Fields should be at levels
characteristic of a typical location in a typical
commercial or hospital environment.
Note: UT is the a.c. mains voltage prior to application of the test level.