Medtronic SVDD303 Programming Guide

Page 1
SIGMA
300/200/100Software 1.1
Vision™Software Model 9963
Pacemaker Programming Guide
Page 2
Page 3
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
A guide to using the 9790 Series programmer for Medtronic.Sigma 300/200/100 Series pacemakers
Refer to the Medtronic.Sigma 300/200/100 Series Pacemaker Reference Guide for information on the pacemakers.
Page 4
The following terms are trademarks of Medtronic, Inc.
Marker Channel, Medtronic, Medtronic.Sigma, Medtronic.Vision, Auto-PVARP, Quick Look, Quick Secure, Sigma, Significant Events, and Vision.
Created by Cardiovascular Technical Communications
Medtronic, Inc. 1999
All Rights Reserved
Page 5

How to Use This Guide

Information is Contained in Two Guides

Product information about Medtronic.Sigma Series software and pacemakers is presented in two separate guides.
The Pacemaker Programming Guide (PPG) accompanies Medtronic.Sigma Series software and contains instructions on how to use the programmer and the programming software.
The Pacemaker Reference Guide (PRG) is a supplementary guide that provides detailed information on Medtronic.Sigma Series pacemakers.
Note:
Refer to the technical manual supplied with the pacemaker
for information about which features are supported by a Medtronic.Sigma pacemaker.

About this Guide

How to Use This Guide
This guide presents the following information to use the 9790 programmer.
How to setup and configure the programmer and access on-line help.
How to start a patient session, use the various follow-up features during the session, and properly end the session.
How to view and print the patient’s ECG and EGM waveform traces.
How to configure the pacemaker to collect diagnostic data and how to retrieve and view this information.
How to measure stimulation thresholds and sensing levels.
How to program parameter values and verify rate response parameters settings.
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
iii
Page 6
How to Use This Guide

About the Pacemaker Reference Guide

This supplementary guide describes in detail, how the pacemaker operates and specifies the capabilities of each model.
Describes the pacing modes, rate response options, special therapy features, telemetry types, and data collection options. In some cases, guidelines are given on how to configure the pacemaker operation.
Contains troubleshooting information for electrical and hemodynamic problems.
Specifies parameter and data collection capabilities, longevity projections, and mechanical and electrical specifications.
Provides general warning and cautions, potential interference sources, and general indications for pacing.
Contains a glossary of terms.
iv
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
Page 7
Table of Contents
1. Programmer Basics
Table of Contents
How to Use This Guide iii
Information is Contained in Two Guides iii About this Guide iii About the Pacemaker Reference Guide iv
Setting Up the 9790 Programmer 1-2
Programmer Setup Check List 1-2
Connecting an External Monitor/Recorder 1-4 Connecting the Programmer to Skin Electrodes 1-6 Positioning and Using the Programming Head 1-7
When to Position the Programming Head 1-7
Effect of the Programming Head on Pacemaker
Operation 1-9
Observing Magnet Mode Operation 1-10
Alternative PROGRAM and
INTERROGATE Buttons 1-10
Automatic Interrogation at the
Start of a Session 1-11 About the Display Screen 1-12
Features and Conventions of the Display Screen 1-12
Using the On-Screen Keyboard 1-18
Audible Tones 1-20 Programming Emergency Parameters 1-21 Recording an ECG Strip 1-22
About the ECG Recording 1-23 Using the On-line Help Feature 1-25
Entering the Help System 1-26
Using the Links in Help 1-27
Exiting the Help System 1-28
Searching for Information in Help 1-28
2. Conducting a Patient Session
Important Reminders About Medtronic.Sigma Pacemakers 2-2
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
v
Page 8
Table of Contents
New Programmer Features 2-2 Automated Monitoring and Collection of Patient
Diagnostic Data 2-3
Starting a Patient Session 2-4
The Starting Point of a Patient Session 2-4 The Tool Palette Between Sessions 2-5 About Automatic Interrogation 2-6 Procedure for Starting a Patient Session 2-7
Proceeding with Task Selection 2-10
The First Task Screen 2-10
Selecting Another Task or Function 2-10 Interrogating the Pacemaker 2-12 Taking a Quick Look at Pacemaker Operation 2-13
The Quick Look Screen 2-13
Viewing Data on the Quick Look Screen 2-14 Viewing Battery and Lead Measurements 2-17
Battery and Lead Measurements Screen 2-17
Pacemaker Battery Measurements 2-18
Lead System Measurements 2-19
Updating the Displayed Data 2-20 Checking the Present Parameter Settings 2-21
The Therapy Parameters Screen 2-21 Viewing Patient Information Stored in the Pacemaker 2-22 Recording an ECG Strip of Magnet Operation 2-23
The Magnet Test Setup Screen 2-23
Procedure for Conducting a Magnet Test 2-24
Collecting an ECG Strip 2-25
About the Collected ECG Strips 2-25 Checking the Patient’s Underlying Rhythm 2-26
The Underlying Rhythm Test Screen 2-26
Procedure for Checking the Patient’s Underlying
Rhythm 2-27
Collected ECG Strips 2-29 Printing Data Reports 2-30
Using the Print Button 2-30
Printing Additional Reports 2-32
Using the Continuous Print “Trace” Mode 2-33
vi
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
Page 9
Table of Contents
Printing Reports Held in the Print Queue 2-33
Ending a Patient Session 2-35
Saving a Session on Diskette 2-37 Options for Clearing Pacemaker Data 2-39
3. Viewing the Patient’s ECG and EGM Traces
Viewing the ECG and Other Rhythm Waveforms 3-2
Expanding the Live Rhythm Monitor Window 3-3 About the Waveform Traces 3-4
Adjusting and Configuring the Display 3-7
Adjusting Waveform Size 3-9 Selecting a Trace Source 3-9 Color Coding the Waveform Traces 3-11 Selecting a Different Sweep Speed 3-12 Programming a Different Telemetry Mode 3-13 Selecting the ECG Filter 3-14 Enabling Artifact Enhancement 3-15 Arranging the Waveforms 3-15
Freezing and Analyzing a Waveform Strip 3-18
Using the On-Screen Calipers 3-19 Adjusting the Frozen Traces 3-21 Displaying a Marker Channel Diagram 3-22 Saving a Frozen Waveform Strip 3-27 Printing the Frozen Strip 3-27
Recalling and Viewing Waveform Strips 3-28
4. Collecting Diagnostic Data
About Data Collected by the Pacemaker 4-2
Types of Collected Data 4-2 Important Points about Using the Data Function 4-4
Displaying Collected Data 4-7
About the Collected Data - Graphs and Tables Screen 4-7 Automatically Collected Data 4-8
Choosing Clinician-Selected Detailed Data Collection 4-17
Off 4-17 Custom Rate Trend (Rate versus Time) 4-17 High Rate Detail (Atrial or Ventricular) 4-21
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
vii
Page 10
Table of Contents
Programming Data Collection 4-23
Setup Options 4-23 Clearing Data From the Pacemaker 4-26
5. Evaluating Parameter Settings
Measuring Stimulation Thresholds 5-2
The Threshold Test Setup Screen 5-2
Setting Up a Threshold Measurement Test 5-3
Executing a Strength-Duration Threshold Test 5-6
Executing an Auto Decrement (Amplitude or Pulse Width)
Threshold Test 5-13
Conducting a Manual Threshold Test 5-15 Determining a Sensitivity Setting 5-18
The Sensing Test Setup Screen 5-18
Checking the Patient’s Intrinsic Rate 5-19
Executing an Automatic Sensing Test 5-20
Automatic Sensing Test Results 5-23
Executing a Manual Sensing Test 5-24 Using Temporary Programming to Evaluate
Parameter Settings 5-27
The Temporary Test Setup Screen 5-27
Procedure for Conducting a
Temporary Test 5-28
Programming a High Temporary Pacing Rate 5-30
viii
6. Programming Pacemaker Parameters
Programming Parameters 6-2
Selecting the Therapy Parameters Screen 6-2
Programming, Printing, and Clearing Parameter
Values 6-3
Parameter Restrictions 6-4
How to Program Parameters 6-6
How to Program Subordinate Parameters 6-8 Saving/Retrieving a Set of Parameter Values 6-13
Saving a Set of Parameter Values 6-14
Retrieving a Saved or Nominal Set of Parameter
Values 6-15
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
Page 11
Table of Contents
7. Rate Response Setup
Using Exercise Test to Verify Rate Response 7-2
Overview of the Exercise Test 7-2 Selecting the Exercise Test 7-3 Starting the Exercise Test 7-3 Evaluating the Results of the Exercise Test 7-5 Adjusting Rate Response 7-6
8. Miscellaneous Setup Options - Programmer and Pacemaker
Selecting System Operating Preferences 8-2 Adjusting Programmer Time and Date 8-4 Improving the Detection of Pacing Artifacts 8-5 Checking the Software Version 8-6 Starting the Demonstrations Option 8-7 Programming Patient Data into Pacemaker Memory 8-9 Connecting an External Printer 8-11
Printer Compatibility 8-11 Materials You Will Need 8-13 Connecting the Printer 8-14
A. Appendix: Parameter Values and Restrictions
Programmable Modes and Parameters A-2 Automatic Diagnostics A-13 Clinician-Selectable Diagnostics A-15 Programming Restrictions A-18
I. Index
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
ix
Page 12
Table of Contents
x
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
Page 13
Programmer Basics
This chapter provides an overview of setting up the programmer and covers information you should be familiar with before you begin a patient session.
This information includes such things as using the programming head and stylus, programming Emergency parameters, and recording an ECG strip. Sections include an introduction to features of the display screen and use of the on-line Help system.
1
Setting Up the 9790 Programmer
1-2
1
For more detailed information on setting up your programmer, refer to Programmer Description and Setup, a guide supplied with your 9790 series programmer.
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
Connecting the Programmer to Skin Electrodes
Positioning and Using the Programming Head
About the Display Screen
Programming Emergency Parameters
Recording an ECG Strip
Using the On-line Help Feature
1-25
1-6
1-7
1-12
1-21
1-22
1-1
Page 14
Programmer Basics

Setting Up the 9790 Programmer

Setting Up the 9790 Programmer
The following topics summarize the basic steps for setting up your 9790 series programmer. For more complete information on setting up your programmer, refer to the guide, Programmer Description and Setup, supplied with your programmer.

Programmer Setup Check List

Verify that the components illustrated below are properly connected to the programmer.
Selector Pen
disconnect the stylus.
1-2
Note:
Programming Head
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
(Stylus)
Do not
ECG Cable
Figure 1-1.
The Programmer
Page 15
Programmer Basics
Setting Up the 9790 Programmer
Figure 1-2.
Figure 1-3.
Power Cord
Programming Head
Figure 1-4.
ECG Cable
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
1-3
Page 16
Programmer Basics
Setting Up the 9790 Programmer

Connecting an External Monitor/Recorder

To connect an external monitor or recorder to your programmer, you need the optional adaptor shown below. Connect the adaptor to the analog output port on the right side of the programmer.
Calibrate Button
w
Warning:
Peripheral equipment connected to the 9790 programmer must be certified according to the applicable IEC standards (IEC 950 for data processing equipment and IEC 601-1 for medical equipment). The system formed by connecting peripheral equipment to the 9790 programmer must comply with IEC 601-1-1 for medical electrical systems. It is the responsibility of the person who connects the peripheral equipment to comply with IEC standards. It is the responsibility of the user to keep peripheral equipment that is certified to IEC 950 at least two meters away from the patient. Contact the peripheral equipment manufacturer for information about IEC certification.
Connector Type: Output Signal:
(switchable)
ABCD
±
1V
BNC
±1 Volt or ±5mV
Signal Output Adaptor
±
5mV
Connecting Optional Adaptor to the Analog
Output Port on the Programmer.
Figure 1-5.
1-4
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
Adaptor Output Signals
(after model selection):
- Patient’s ECG (upper most ECG
A
on display)
- Telemetered EGM (upper most EGM
B
on display)
- Telemetered EGM (other EGM
C
if programmed)
- Marker Channel Telemetry
D
Connecting the Optional External Adaptor
Page 17
Programmer Basics
Setting Up the 9790 Programmer
Using the Calibrate Button
Selecting the Calibrate button located next to the analog output port adds a reference signal (as shown in Figure 1-6 and Figure 1-7) to the trace of Marker Channel and EGM telemetry.
The Marker Channel signal shows the relative marker amplitudes, which are not annotated with character codes on an external device. The EGM calibration signal acts as a voltage reference for the displayed EGM.
Atrial Pace (AP)
Atrial Sense (AS)
Atrial Refractory Sense (AR)
Vent. Refractory Sense (VR)
Vent. Sense (VS)
Vent. Pace (VP)
Figure 1-6.
5 mV
1 mV
Figure 1-7.
Marker Channel Signals
EGM Calibration Signals
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
1-5
Page 18
Programmer Basics

Connecting the Programmer to Skin Electrodes

Connecting the Programmer to Skin Electrodes
At the start of a patient session, the programmer must be connected to skin electrodes on the patient. The ECG display and measurement functions will not operate without detection of the surface ECG signal.
Use an electrode lead to connect each skin electrode to the appropriate port on the ECG cable. Typical electrode placement is shown below. Use standard procedures for attaching disposable skin electrodes to the patient.
RA (R)
RL (N)
Figure 1-8.
Refer to the guide, Programmer Description and Setup, supplied with your 9790 programmer for more details on this procedure.
Connecting the Programmer to Skin Electrodes
LA (L)
C (C)
LL (F)
( ) = IEC Coding
1-6
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
Page 19
Programmer Basics

Positioning and Using the Programming Head

Positioning and Using the Programming Head
In many of the procedures described in the following chapters, you will be directed to position the programming head over the patient’s pacemaker.

When to Position the Programming Head

You must position the programming head over the patient’s pacemaker whenever a procedure initiates communication between the programmer and pacemaker. Examples are:
At the start of a patient session when you select the
[Auto-Identify] or [Start] button. The programmer
automatically initiates an interrogation that takes a number of seconds to complete.
Prior to executing any command that results in a telemetry or programming transmission.
To view or record Marker Channel or EGM telemetry.
Caution:
Do not position the programming head over an implanted pacemaker during electrocautery or defibrillation procedures.
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
1-7
Page 20
Programmer Basics
Positioning and Using the Programming Head
How to Position the Programming Head
The programming head must be properly positioned as described below. An incorrectly positioned programming head can result in the failure of a transmitted command and the loss of telemetry.
1.
Hold the programming head directly against the patient’s skin with the face of the programming head parallel to the pacemaker.
1-8
Figure 1-9.
2.
Position the programming head so that the amber light in
Positioning the Programming Head
the light array goes out and one or more of the green lights come on. Move the head a little in each direction to find the position that lights the greatest number of green lights. This is the optimum position.
Light Array
Figure 1-10.
Green
Amber/Green
Light Array Indicator
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
Page 21
The present pacing mode shows in the status bar at
the top of the screen.
Programmer Basics
Positioning and Using the Programming Head
Effect of the Programming Head on Pacemaker Operation
If model selection and application (software) loading are complete, positioning the programming head over a Medtronic.Sigma pacemaker does not put the pacemaker in the magnet mode of operation, as is the case with previous pacemaker models. The programmer automatically sends a Cancel Magnet command to the pacemaker, which causes it to continue to operate as programmed.
An exception to this operation can occur if the programming head does not establish a telemetry link with the pacemaker (because of strong electrical interference or because the programming head is improperly positioned). In such cases, positioning the programming head would result in magnet mode operation until a telemetry link is established. You can easily verify the present pacing mode by observing the Status Line at the top of the screen.
Pacemaker operation returns to its programmed state about 2 seconds after you remove the programming head from its position over the pacemaker.
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
1-9
Page 22
Programmer Basics
Positioning and Using the Programming Head

Observing Magnet Mode Operation

To observe Magnet mode operation during a patient session, you must conduct the Magnet test described in Chapter 2. Initiating the Magnet test results in a Threshold Margin Test (TMT) and causes the pacemaker to operate in the Magnet mode.
Note:
Between patient sessions before you select the pacemaker, positioning the programming head over a Medtronic.Sigma pacemaker will result in magnet mode operation.
Alternative
INTERROGATE
To initiate the Program and Interrogate commands, you have the option to use the on-screen the Program head.
[Interrogate]
Note:
The
[P] or [I] button on the programming head is active only
when its counterpart is displayed as an active button on the display screen.
PROGRAM
Buttons
[P] and Interrogate [I] buttons on the programming
Button
Figure 1-11.
Alternative Program and Interrogate
Programming Head Buttons
and
[Program] and [Interrogate] buttons or
Light Array
[Program]
Button
1-10
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
Page 23
Programmer Basics
Positioning and Using the Programming Head

Automatic Interrogation at the Start of a Session

At the start of a session when you select [Start] or [Auto Identify], the programmer automatically interrogates the patient’s pacemaker for all the data contained within the pacemaker. For this process, you must position the programming head and hold it steady in place until the interrogation is complete.
At the bottom center of the screen you will see an indicator showing the progress of the interrogation. Because this interrogation retrieves all the data stored in the pacemaker, including diagnostic data, the process may take a number of seconds.
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
1-11
Page 24
Programmer Basics

About the Display Screen

About the Display Screen
The programmer display screen is an interactive device that not only displays information in the form of both text and graphics, but functions as a control panel by displaying buttons and menu options that you can select using the stylus.
Features and Conventions of the Display Screen
This section describes the features and conventions of the display screen. The Therapy Parameters screen below shows the main elements of the typical screen.
Status Bar
Live Rhythm
Monitor Window
1-12
Task area
Active Field Button Tool PaletteCommand Bar
Figure 1-12.
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
Main Elements of a Display Screen Example
Page 25
Status Bar
Programmer Basics
About the Display Screen
Note:
For information on changing the language in the screens (for example, from English to German), see “Selecting System Operating Preferences” on page 8-2.
The Status Bar
The status bar at the top of the screen shows:
The present pacing mode.
When any one of a number of test conditions is occurring.
The pacemaker model.
The Live Rhythm Monitor Window
Live Rhythm Monitor
Window
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
This window is a partial view of the full-screen display of ECG, Marker Channel, and telemetered EGM waveform traces. You can expand this window to its full size by selecting the small square button in the upper-right corner of the window or by selecting the
[Adjust...] button.
Heart rate and rate interval show if the programming head is positioned over the pacemaker.
Annotations above the waveform trace show the point of programmed parameter changes.
1-13
Page 26
Programmer Basics
About the Display Screen
The waveform trace or traces that show in this window depend on the selected task screen and how traces have been arranged in the full-screen view. Refer to “Viewing the ECG and Other Rhythm Waveforms” on page 3-2 for information about the live rhythm monitor.
1-14
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
Page 27
Programmer Basics
About the Display Screen
The Task Area
The portion of screen between the live rhythm monitor window at the top and the command bar at the bottom changes according to the task or function you select. The example in Figure 1-12 shows the Therapy Parameters screen for programming pacemaker parameter settings. This task area would appear much different if you selected, for example, the Threshold Test Setup screen.
The Command Bar
The bar at the bottom of the screen always shows the command buttons for programming “emergency” parameters settings, interrogating the pacemaker, and ending the patient session.
For complete information on these functions, see “Programming Emergency Parameters” on page 1-21, “Interrogating the Pacemaker” on page 2-12, and “Ending a Patient Session” on page 2-35.
Active fields show as
unshaded areas on
the screen.
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
Active Fields
Unshaded areas, or boxes, appearing in the task area are “active fields” that respond to the stylus.
Selecting a value, word, name, or phrase that appears in an active field opens a menu or window of alternative options for whatever is represented in that field.
1-15
Page 28
Programmer Basics
S
About the Display Screen
electing an active field
opens a menu or window of
alternative options, or in
some cases, an
on-screen keyboard.
For example, touching the mode value “DDDR” in the Mode field with the tip of the stylus opens a window of mode options. Selecting any one of these options replaces the original DDDR value with the selected one.
Selecting some fields (those with terms ending in an ellipsis, such as “Rate Response...”) opens a window displaying additional fields. Some fields that require entry of information, such as patient data, open an on-screen keyboard. How to use this keyboard is described later in this section.
Buttons
Buttons like those shown below let you operate the programmer using the stylus. You can “press” a button by touching it with the tip of the stylus.
Buttons having a less
distinct shaded label are
not presently active.
1-16
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
Buttons may directly execute a command, such as the
[Program]
button, or they may open a window that prompts another action. Usually such buttons have a label ending with an ellipsis, such as the
[Save...] or [Get...] buttons shown above.
Page 29
Buttons
Icons
Programmer Basics
About the Display Screen
A procedure may instruct you to “press and hold” a button. In such cases you should touch the tip of the stylus to the button and continue to maintain pressure against the button until it is time to “release” the button.
The Tool Palette
The collection of buttons and icons along the edge of the screen is referred to as the “tool palette.” These are the controls you will use to choose the task or function screen you want displayed. Once you have started a patient session, the tool palette is always displayed, making it quick and easy to move to the desired task or function.
Each of the icons acts like a button. To select an icon, touch the icon with the stylus. The “<” symbol adjacent to four of the icons indicates that selecting one of these icons opens a menu of related options. The icons without the < symbol directly open a task screen.
Refer to Table 2-2 on page 2-11 for a brief explanation of the purpose of each button and icon in the tool palette.
Note:
The Demo icon is available only when the programmer is operating in the Demo mode.
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
1-17
Page 30
Programmer Basics
About the Display Screen

Using the On-Screen Keyboard

Certain fields on the screen allow you to enter data, such as the patient’s name or chart number. Selecting such a field automatically displays the on-screen keyboard shown below. By touching the letter or character buttons with the stylus, you can use this display feature like an actual keyboard.
Text Entry Window
Space Bar
Figure 1-13.
Keyboard Screen
Cursor
Position Keys
The function of the on-screen keyboard buttons are very similar to the keys on a computer keyboard or typewriter.
1-18
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
Page 31
Programmer Basics
About the Display Screen
Table 1-1.
Button or Feature Function
Text Entry Window
(See Figure 1-13)
Space Bar
(See Figure 1-13)
Shows the text as you enter it using the keyboard. You can enter only as many characters as can fit in the selected field.
Clears all characters from the text entry window.
Closes the Keyboard screen without changing the selected field.
Deletes the character to the left of the cursor in the text entry window.
Has no effect.
Locks all characters into upper case until the
key is selected again. While the
[Caps]
key is down, the operate as described below except that characters will shift to lower case.
Shifts all of the characters into upper case and replaces the top row of numbers with commonly used symbols. This shift cancels automatically after you select a character or symbol.
Locks all of the characters into upper case and replaces the top row of numbers with commonly used symbols. Press this key again to return to number keys and lower case.
Shifts the keyboard to a limited set of characters. Selecting a character or symbol cancels
Inserts a blank space.
Keyboard Buttons
and
[Shift]
key.
[Alt]
[Shift Lock]
[Caps]
keys
Moves the cursor one space to the left or the right.
Closes the Keyboard screen and enters the text in the text entry window into the selected field.
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
1-19
Page 32
Programmer Basics
About the Display Screen

Audible Tones

Certain events in the operation of the programmer result in an audible signal. The following tones alert the user to the success or failure of an action.
A two-tone beep (low-to-high) indicates confirmation of an Interrogate or a Program command.
A single low-tone beep indicates that an Interrogate, Program, or Emergency command was not confirmed. It can also indicate that the selected command cannot be executed.
A single, short beep coincides with pressing the Interrogate or the Program button. It also occurs upon automatic identification of the pacemaker.
1-20
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
Page 33

Programming Emergency Parameters

Emergency
+
Programming Emergency Parameters
The Emergency programming command is a safety feature that overrides all other functions and immediately programs the pacemaker to preset emergency values intended to provide pacing support under a variety of conditions. This programming cancels any temporary function in effect and restores Magnet mode operation.
To Program Emergency Parameters
1.
Position the programming head over the pacemaker.
2.
Take either of the following actions:
Press the square red button on the left side of the display panel.
Or, select the on-screen [Emergency] button in the lower left corner of the screen
.
Programmer Basics
Red Button
Emergency Button
3.
Hold the programming head steady until a confirmation message appears. If programming is not confirmed, verify that the programming head is properly positioned and then reselect the
[Emergency] button or the square red button.
Emergency values are programmed settings that provide higher­than-normal energy output. It is not intended that the pacemaker be left at these settings. Refer to Sigma Pacemaker Reference Guide for a list of Emergency parameters settings for Medtronic.Sigma pacemakers.
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
1-21
Page 34
Programmer Basics

Recording an ECG Strip

Recording an ECG Strip
At any time during a patient session, you can initiate a continuous, real-time ECG recording as described below.
To Start an ECG Recording
Press the desired paper speed button (
12.5 mm/sec
25 mm/sec
12.5 mm/sec
Paper Advance
25 mm/sec or
).
1-22
Figure 1-14.
To Stop the Recording
Press the same paper speed button again.
Before you tear off the ECG strip, press the button to advance the strip to a perforation.
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
ECG Chart Recorder Control Buttons
[Paper Advance]
Page 35
Annotations
ECG Trace
Programmer Basics
Recording an ECG Strip

About the ECG Recording

Because the printed recording provides a higher resolution, it may show artifacts and events that do not appear on the display.
Annotation of Executed Commands – Information on the ECG printout includes an indication of when certain commands to the pacemaker occurred. When confirmation of the command is received, the command name is printed at the appropriate point in the margin above the waveform grid. A recording made during use of the following test functions will show the programmed test values as they are programmed: Magnet, Underlying Rhythm, Threshold, Sensing, and Temporary.
Marker Telemetry
EGM Telemetry
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
Figure 1-15.
Example ECG Printout with Marker and EGM Telemetry
Marker Channel and EGM Telemetry – If the programming head is positioned over the patient’s pacemaker, the recording will include a trace or traces of the telemetry being received from the pacemaker. In this example, the patient’s ECG is accompanied by Marker and EGM telemetry.
1-23
Page 36
Programmer Basics
Recording an ECG Strip
ECG and EGM Trace Adjustment – The ECG and EGM are recorded according to control settings accessible from the Adjust window (see “Adjusting and Configuring the Display” on page 3-7). The ECG source (Lead I, II, or III) printed is the ECG trace appearing first in the live rhythm monitor window on the programmer screen. When the telemetry mode is Dual EGM, atrial and ventricular EGMs appear.
Missing Markers – A programming command or interrogation momentarily interrupts the transmission of Marker Channel telemetry. This interruption can result in missing markers. The point at which the command occurred is marked above the ECG trace by a “down” and/or “up” arrow. The down arrow (▼) indicates a transmitted command from the programmer; the up arrow (▲) indicates a telemetry response from the pacemaker.
1-24
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
Page 37

Using the On-line Help Feature

The On-line Help system provides summaries on-screen about many of the features and operations of the programmer. (Refer to the manuals that apply to the pacemaker for more complete information.) When you request Help, a Help window opens (covering much of the workspace temporarily). If the initial Help window does not have the information you were seeking, you can find another one that does by using the buttons and icons listed in Table 1-2.
When you are ready to leave the On-line Help system, simply close the Help window. You will return to the workspace that was in view before you requested Help.
Programmer Basics
Using the On-line Help Feature
Table 1-2.
Button or Icon What It Does
[Previous]
[Topics]
button Displays the Help window last viewed or closes
button Closes the Help window.
[Close]
button Displays a list of terms discussed in Help. From
[Index]
button Displays a list of all available Help topics. From
Getting Around in the Help System
Specific Help — If this icon is displayed within a screen or window, select it to get specific help related to that screen or window.
This icon is displayed only if Specific Help is available.
General Help — Select this button at any time to enter Help, starting with a Topics list.
the Help window first opened.
this list, you can select a Help topic to learn about one of the terms.
this list you can select any topic.
Link icon — Indicates that more Help is available. Select this symbol to jump to a related topic.
Some pictures may have links also.
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
1-25
Page 38
Programmer Basics
Using the On-line Help Feature
Table 1-2.
Button or Icon What It Does
(underlined phrase) Definition available — Select the phrase to see
Scroll bar Use this if a Help topic extends beyond the
Getting Around in the Help System
a pop-up window that defines the phrase. Select
[Previous]
window size.
to close the pop-up window.

Entering the Help System

Figure 1-16 shows ways to enter the Help system:
If the [?] icon is present, select it for specific help about the screen or window in which it appears. This icon appears near the window’s title bar when specific help is available.
Select the [Help…] button, which is always present. This leads you directly to a topics list so that you can search for information. You can also search by using a Help index.
Once you have entered the Help system, you can use the built-in links to jump from the current Help window to other related Help windows.
1-26
[Help…]
Title bar for window having
button for
general Help
icon for
[?]
specific Help
specific Help
Figure 1-16.
Entering On-line Help
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
Page 39
Programmer Basics
Using the On-line Help Feature

Using the Links in Help

Figure 1-17 shows the links that allow you to see additional Help windows:
The basic link, whose icon indicates that more Help is available. Select it to jump to a related topic.
The definition link, whose symbol is an underlined phrase, indicates that there is a pop-up window to define the phrase. Select the phrase to open the pop-up window. Select outside the pop-up window (or select
Note:
Some pictures may have links also. Each of these pictures has instructions on how to select the links and where they lead.
If you have viewed a series of topics and wish to return to a topic viewed earlier in that series, you can use the go backward through the series of topics one at a time.
Pop-up definition window Link to definition window Scroll Bar
[Previous]) to close it.
[Previous] button to
Topic window
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
Help push buttons Link to another
Figure 1-17.
Typical Help Windows (Topic and Definition)
Help topic
1-27
Page 40
Programmer Basics
Using the On-line Help Feature

Exiting the Help System

When you are ready to leave Help and return to the programmer workspace, select the
[Close] button. The Help window closes,
and the programmer screen displays the same information that it did when you entered Help.
Note:
Except for the
[Emergency] button, no other functions on
the programmer screen are usable when Help is open. To use other programmer functions, you must first exit the Help system.

Searching for Information in Help

If you do not find the information you were looking for in the current Help screen, there are several methods for searching for it:
Topics List — Select the [Topics] push button to open a window having two columns. The column on the left lists categories of Help topics. One of these categories is always selected, and the topics for that category are listed in the right­hand column. You can view:
1-28
any of the topics listed (on the right) by selecting the link
icon next to its name, or
another category by selecting its name (on the left).
Figure 1-18 shows what happens when you select a different category. In this example, the category “Initial Help” is selected first. The right-hand column lists topics included in this category. Suppose that you are searching for Help on one of the pacing therapies. If you select “Pacemaker Features” as a new category, notice that the right-hand column changes. You can then select a link icon for the specific topic you would like to view.
Help Index — Select the [Index] button to open a window that lists indexed terms in alphabetical order. Like a book index, the terms are not limited to titles in Help. You can go to a Help window explaining any of the indexed terms by selecting the link icon next to that term.
Figure 1-19 shows some typical index entries.
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
Page 41
Programmer Basics
Using the On-line Help Feature
Unexpanded
Help categories
To open a
Help topic,
select its
link icon
To expand a new Help category,
select its name
Figure 1-18.
Topics List for Help
Expansion of a Help category
To open a Help topic,
select the link icon for its
Index listing.
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
Figure 1-19.
Help Index
1-29
Page 42
Programmer Basics
Using the On-line Help Feature
1-30
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
Page 43
Conducting a Patient Session
This chapter describes how to begin a patient session and how to execute some of the typical tasks you might use to evaluate operation of the patient’s pacing system. Information about printing, saving, and transferring session data is also included.
Important Reminders About Medtronic.Sigma Pacemakers
Starting a Patient Session
Proceeding with Task Selection
Interrogating the Pacemaker
Taking a Quick Look at Pacemaker Operation
Viewing Battery and Lead Measurements
2-2
2-4
2-10
2-12
2-13
2-17
2
Checking the Present Parameter Settings
Viewing Patient Information Stored in the Pacemaker
Recording an ECG Strip of Magnet Operation
Checking the Patient’s Underlying Rhythm
Printing Data Reports
Ending a Patient Session
2-21
2
2-22
2-23
2-26
2-30
2-35
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
2-1
Page 44
Conducting a Patient Session
Important Reminders About Medtronic.Sigma Pacemakers
Important Reminders About Medtronic.Sigma Pacemakers
This section lists some of the important features of the Medtronic.Sigma pacing system.
New features of the programmer software
Automated monitoring and collection of diagnostic data by the pacemaker

New Programmer Features

Auto Cancel Magnet – Cancels the magnet automatically when you place the programmer head over the pacemaker. With Magnet Test, you can view and record magnet operation.
Full Auto-Interrogation – Occurs automatically at the start of a session.
2-2
Full-size Reports – You can choose to connect an external printer for printing full-size reports.
Live Rhythm Waveform Display – You can quickly tailor a multiple-trace display of the patient’s ECG, atrial and ventricular EGMs, and Marker Channel signals.
Saved Session Data – You can save data from a session on a diskette, which allows you to import session data into a data base.
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
Page 45
Conducting a Patient Session
Important Reminders About Medtronic.Sigma Pacemakers
Automated Monitoring and Collection of Patient Diagnostic Data
Diagnostic monitors collect data automatically in the background:
Rate and AV Conduction Histograms
Atrial High Rate Episodes (including Mode Switch Episodes) and Ventricular High Rate Episodes
Chronic Lead Impedance Trend
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
2-3
Page 46
Conducting a Patient Session

Starting a Patient Session

Starting a Patient Session
Because the programmer collects and stores data on a session-by­session basis, it is important to correctly start and end each session. This section describes how to start a patient session. To end a session, refer to “Ending a Patient Session” on page 2-35.

The Starting Point of a Patient Session

A patient session always begins at the Select Model screen. See “Procedure for Starting a Patient Session” on page 2-7. If you are between patient sessions, you can access other screens by using the icons and buttons described in Table 2-1.
The Select Model screen appears:
Upon completion of the self test when you turn the programmer on.
After you end a patient session.
2-4
Figure 2-1.
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
Select Model Screen
Page 47
Conducting a Patient Session
Starting a Patient Session
Notes:
If the Select Model screen does not look like this example and you see a
[Medtronic] button, select the button to display this
screen.
The [Nominals] button does not apply to Medtronic.Sigma pacemakers. This button allows you to set up Site Nominals if you select a pacemaker with this feature.

The Tool Palette Between Sessions

Table 2-1.
Tool Palette Tool Selecting the Tool (Button or Icon)… Reference
Freezes a segment of the live rhythm display.
Note:
saved) between patient sessions. Markers and EGM traces are not present between patient sessions.
Accesses the rhythm strips saved during a patient session.
Note:
sessions.
Opens a window of options for adjusting the live rhythm display.
Note:
patient session.
Displays the screen for selecting a pacemaker model and starting a patient session.
Displays a queue of print requests from previous sessions as well as frozen waveform reports requested between sessions.
Displays the programmer setup options.
Preferences Time and Date Artifact Detection Software Demonstrations
Tool Palette Between Patient Sessions
page 3-18
A frozen strip can be viewed and printed (but not
page 3-28
No saved strips are available between patient
page 3-7
Additional adjustment options are present during a
page 2-4
page 2-33
page 8-2 page 8-4 page 8-5 page 8-6 page 8-7
Note:
When some functions are active on the display, selecting a tool button or icon will have
no effect. Closing the active window restores operation of the tool palette.
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
2-5
Page 48
Conducting a Patient Session
Starting a Patient Session

About Automatic Interrogation

At the start of a patient session when you select the [Auto-Identify] button or the Session” on page 2-7), the programmer automatically attempts to interrogate the patient’s pacemaker to retrieve all the data that might be needed during the session.
Note:
must position the programming head over the pacemaker and continue to hold it in place until the interrogation is complete.
The interrogation time may extend to 55 seconds because of the amount of information stored in the pacemaker. A status gauge at the bottom of the screen shows progress of the interrogation. The
[Stop] button lets you cancel the interrogation once it has started.
Note:
button, you cannot proceed with session activities until the interrogation is 100% complete.
[Start] button (see “Procedure for Starting a Patient
To take advantage of this automatic interrogation, you
Except for use of the
[Emergency] button or the [Stop]
2-6
You can choose to stop interrogation and continue with other activities (not using the automatic interrogation feature). To do this, select the the process is complete. In this case, when data is needed, the programmer will display a message prompting you to interrogate the pacemaker.
You also can manually interrogate the pacemaker at any time during the patient session (see “Interrogating the Pacemaker” on page 2-12).
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
[Stop] button or lift the programming head before
Page 49
Conducting a Patient Session
Starting a Patient Session
Warning Messages – As a result of an interrogation, a warning message box may be displayed. Examples of these include:
ERI (Elective Replacement Indicator)
POR (Power On Reset) or Full Electrical Reset
Interrogation interrupted or unsuccessful
You must acknowledge the message before proceeding with other activities. You can attempt to clear an ERI or POR message by following the instructions in the message. Otherwise, you can close the message box and clear the condition later. See “Resetting ERI or Electrical Reset” on page 6-11.

Procedure for Starting a Patient Session

From the Select Model screen, you can start a patient session by either of two methods:
By selecting the [Auto-Identify] button, which automatically identifies and selects the patient’s pacemaker model.
By selecting the “Medtronic.Sigma Series” selection displayed on the screen and then selecting the
Note:
To review software screens without interrogating a
[Start] button.
pacemaker, use the demonstrations option (See “Starting the Demonstrations Option” on page 8-7). Starting a session without interrogating a pacemaker provides only a limited number of screens for review.
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
2-7
Page 50
Conducting a Patient Session
Starting a Patient Session
To Start a Session Using Auto-Identify
1.
2.
3.
Display the Select Model screen:
If the programmer is not operating, turn it on. The power
switch is on the left side and near the back. The Select Model screen appears after a short, self-test sequence. (Select the
If the programmer is operating, but the Select Model
screen is not displayed, choose the Select Model icon from the tool palette at the side of the screen.
If the Select Model icon is not in the tool palette, the steps
to properly end the previous patient session have not been executed. Refer to “Ending a Patient Session” on page 2-35.
Position the programming head over the patient’s pacemaker and continue to hold it steady.
Select the
After 40 to 50 seconds of internal software loading, the programmer displays the first task screen. Continue to hold the programming head in place until the initial interrogation process is 100% complete.
[Medtronic] button if it appears on the screen.)
[Auto-Identify] button at the bottom of the screen.
2-8
Refer to “Proceeding with Task Selection” on page 2-10 for information on selecting the functions or tasks you have planned for the session.
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
Page 51
Categories
Models
[Start]
button
Conducting a Patient Session
Starting a Patient Session
To Start a Session by Selecting the Pacemaker Model
1.
Display the Select Model screen. (Refer to the previous procedure.)
2.
Select the appropriate category to view pacemaker models.
3.
Select “Medtronic.Sigma Series” from the list of models. (The specific model number will appear on the status bar at the top of the display screen during the patient session.)
4.
If you want the programmer to automatically interrogate the patient’s pacemaker, position the programming head.
5.
Select the
6.
After 40 to 50 seconds of internal software loading, the programmer displays the first task screen. Continue to hold the programming head in place until the initial interrogation process is 100% complete.
Refer to the next topic for information on selecting the functions or tasks you have planned for the session.
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
[Start] button.
2-9
Page 52
Conducting a Patient Session

Proceeding with Task Selection

Proceeding with Task Selection
After you select the [Auto-Identify] or [Start] button to begin a patient session, a short period of internal software loading occurs before the first task screen appears.

The First Task Screen

The first task screen to appear is the Quick Look screen. Refer to “Taking a Quick Look at Pacemaker Operation” on page 2-13 for information about the Quick Look screen.

Selecting Another Task or Function

To proceed with the session, select the desired task or function from the button and icon options grouped along the edge of the screen (see Table 2-2 on page 2-11). This group of buttons and icons is referred to as the “tool palette.” It is always available (except during the execution of certain functions) so that you can quickly and easily display a desired task or function screen.
2-10
Note:
When some functions are active on the display, selecting a tool button or icon will have no effect. Closing the active window restores operation of the tool palette.
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
Page 53
Conducting a Patient Session
Proceeding with Task Selection
Table 2-2.
Tool Palette Tool Selecting the Tool (Button or Icon)… Reference
Freezes a segment of the live rhythm display. page 3-18
Accesses the rhythm strips saved since the start of the session.
Opens a window of options for adjusting the live rhythm display.
Accesses the available Help topics Chapter 1
Displays the options for retrieving information about the patient’s pacemaker and its operation and for setting up or clearing the data collection functions.
Quick Look - Initial Interrogation Graphs and Tables Battery and Lead Measurements Data Collection Setup/Clear
Displays the parameter programming screen. page 6-2
Displays the pacing system test options.
Magnet Underlying Rhythm Threshold Exercise Sensing
Temporary
Displays the following session report options:
Available Reports Print Queue
Displays patient information: page 2-22
Tool Palette During a Patient Session
page 3-28
page 3-7
page 2-13 page 4-7 page 2-17 page 4-23
page 2-23 page 2-26 page 5-2 page 7-2 page 5-18 page 5-27
page 2-32 page 2-33
Displays case studies in Demo mode
Note:
Demo diskette is needed for Demo mode. Otherwise, icon is not visible.
page 8-7
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
2-11
Page 54
Conducting a Patient Session

Interrogating the Pacemaker

Interrogating the Pacemaker
At the start of the patient session (as described in “Procedure for Starting a Patient Session” on page 2-7) the programmer performs an automatic interrogation to retrieve all information stored in the pacemaker.
You can also manually interrogate the pacemaker at any time during the patient session. In some cases, a pop-up window may request that you do so.
To Interrogate the Pacemaker
1.
Collected data can
be interrogated
only once during a
patient session.
Select the screen or press the Interrogate head.
2.
From the window of options, select the type of information you want to retrieve. You can select more than one option.
[Interrogate] button at the bottom center of the
[I] button on the programming
2-12
The word “(AGAIN)” indicates that a previous interrogation has already retrieved this data.
3.
Position the programming head and select the or press the programming head Interrogate
Hold the programming head steady until the interrogation is 100% complete as shown by the status gauge at the bottom of the screen. The
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
[Stop] button lets you cancel the interrogation.
[Start] button
[I] button.
Page 55
Conducting a Patient Session

Taking a Quick Look at Pacemaker Operation

Taking a Quick Look at Pacemaker Operation
Quick Look provides a summary or overview of pacemaker operation by displaying the essential information on one screen. This data was retrieved during the initial interrogation of the session. It includes some of the data accumulated by the pacemaker since the last patient session.

The Quick Look Screen

The Quick Look screen appears automatically when you start a patient session or when you choose to display it.
Quick Look - Initial Interrogation
Graphs and Tables
Battery and Lead Measurements
Data Collection Setup/Clear
Figure 2-2.
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
Pacemaker Information on the Quick Look Screen
2-13
Page 56
Conducting a Patient Session
Taking a Quick Look at Pacemaker Operation

Viewing Data on the Quick Look Screen

The Quick Look screen displays the following data and data access options.
Remaining Longevity – This is an estimate of the time in months remaining until pacemaker replacement is required. This estimate is based on the programmed parameter settings and events recorded by the pacemaker since the last patient session. Note the following explanation of the longevity estimates, which you can view by selecting the symbol next to the displayed estimate.
2-14
Atrial Lead – The data under this heading includes programmed values for atrial amplitude, pulse width, and sensitivity. The measured impedance of the atrial pacing lead appears also.
Ventricular Lead – Data under this heading includes programmed values for ventricular amplitude, pulse width, and sensitivity. The measured impedance of the ventricular pacing lead appears also.
Mode/Rates – Under this heading are listed the programmed values for the pacing mode, lower rate, and upper rate.
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
Page 57
Conducting a Patient Session
Taking a Quick Look at Pacemaker Operation
(Number of) Total Events – This heading includes the number of heart beats (events) recorded since the last patient session. The data lists the percentage of these recorded beats that fall into each of the applicable event sequence categories (see Table 2-3).
To see Rate Histograms, select the Histogram icon next to the data. For more information on these histograms, refer to “Heart Rate Histograms” on page 4-9.
Note:
The data under the Total Events and Significant Events headings show data collected by the pacemaker since the last patient session. This data is automatically cleared from pacemaker memory after the session has ended. After the data is cleared, it cannot be recalled.
Significant Events – The information displayed in the field under this heading summarizes the results of diagnostic data collection since the last patient session. This field lists the number of significant events recorded by the various diagnostic monitoring functions (see Table 2-4).
To view a graph or table showing the details associated with a significant event, select the event and then select the
[Open Data]
button. Refer to “Displaying Collected Data” on page 4-7 for information about viewing the data recorded by the various monitoring functions.
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
2-15
Page 58
Conducting a Patient Session
Taking a Quick Look at Pacemaker Operation
Table 2-3.
Event Sequence Categories
Dual Chamber Pacing Modes (and ADI, ADIR, VDI, VDIR, ODO)
AS-VS Atrial Sense - Ventricular Sense
AS-VP Atrial Sense - Ventricular Pace
AP-VS Atrial Pace - Ventricular Sense
AP-VP Atrial Pace - Ventricular Pace
VDD Pacing Mode
AS-VS Atrial Sense - Ventricular Sense
AS-VP Atrial Sense - Ventricular Pace
Single Chamber Pacing Modes
Paced Atrial or ventricular chamber Pace
Sensed Atrial or ventricular chamber Sense
Table 2-4.
Significant Event Monitoring Functions
Functions and Criteria Used for Significant Event Reporting
Ventricular High Rate Episodes One or more episodes detected.
Atrial Polarity Switch
a
Lead impedance outside Notify if > and Notify if < settings.
Ventricular Polarity Switch
a
Lead impedance outside Notify if > and Notify if < settings.
2-16
Electrical Reset Monitor Occurrence of reset conditions.
ERI Monitor Occurrence of ERI conditions.
Mode Switch Episodes One or more episodes detected
Atrial High Rate Episodes One or more episodes detected.
a
Lead impedance reported if lead polarity is switched
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
Page 59
Conducting a Patient Session

Viewing Battery and Lead Measurements

Viewing Battery and Lead Measurements
By selecting the Battery and Lead Measurements screen, you can view information about the pacemaker battery and the lead system based on real-time measurements and calculations made at the time of pacemaker interrogation.

Battery and Lead Measurements Screen

The information on this screen provides a detailed status of the pacemaker battery and the output conditions pertaining to the pacing lead system.
Quick Look - Initial Interrogation
Graphs and Tables
Battery and Lead Measurements
Data Collection Setup/Clear
Figure 2-3.
Note:
The values measured for the pacemaker battery and the lead system can change from one measurement to the next.
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
The Battery and Lead Measurements Screen
2-17
Page 60
Conducting a Patient Session
Viewing Battery and Lead Measurements

Pacemaker Battery Measurements

Battery Status – Displays an “OK” or “Replace Pacer” message based on battery voltage and internal impedance measurements.
Remaining Longevity – Calculated estimate of the time (in months) remaining until pacemaker replacement will be required. This calculated estimate is based on the programmed parameter settings and event data accumulated by the pacemaker since the previous patient session.
For an on-screen explanation of the longevity estimates, select the information ( ) button appearing after the “Remaining Longevity” heading.
2-18
Caution:
estimated remaining longevity. For this decision, use only the elective replacement indicators or the “Replace Pacer” battery status message.
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
Elective replacement should not be based on the
Page 61
Conducting a Patient Session
Viewing Battery and Lead Measurements
Voltage – The battery voltage obtained during the previous measurement.
Current – Present current drain on the pacemaker battery averaged over a pacing cycle.
Impedance – Internal electrical impedance of the pacemaker battery.

Lead System Measurements

Amplitude – Present programmed amplitude of a pacing pulse.
Pulse Width – Present programmed pulse width setting.
Output Energy – The output energy contained in a single pacing
pulse.
Measured Current – Current in the pacing lead during delivery of a pacing pulse.
Measured Impedance – Electrical impedance presented by the pacing lead and electrode/tissue interface.
Pace Polarity – The present lead electrode configuration (unipolar or bipolar) used for pacing.
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
2-19
Page 62
Conducting a Patient Session
Viewing Battery and Lead Measurements

Updating the Displayed Data

You can update the data shown on the screen by the following interrogation procedure.
To Update the Pacemaker Battery and Lead Data
1.
Position the programming head and hold it in place.
2.
Select the
As indicated by the pop-up window, the interrogation and measurement process takes a few seconds.
[Measure Again] button.
2-20
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
Page 63
Conducting a Patient Session

Checking the Present Parameter Settings

Checking the Present Parameter Settings
The first step to viewing the parameter settings to which the patient’s pacemaker is presently programmed is to display the Therapy Parameters screen shown below.

The Therapy Parameters Screen

This screen is used to view permanent parameters or to program them to the desired settings.
An ellipsis (...) appearing
after an option indicates
that selecting that option
will display additional
parameters.
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
Figure 2-4.
The parameter values displayed on this screen are the parameter settings to which the patient’s pacemaker is presently programmed. If the field for a parameter or option displays an ellipsis (e.g., Rate Response…), there are subordinate parameters that are not displayed on this screen.
For more information, see “Programming Parameters” on page 6-2. Table 6-1 lists subordinate therapy parameters.
The Therapy Parameters Screen
2-21
Page 64
Conducting a Patient Session
Viewing Patient Information Stored in the Pacemaker
Viewing Patient Information Stored in the Pacemaker
The Medtronic.Sigma pacemakers can store patient related information that you can view and print during a patient session. This information typically is programmed into the pacemaker at the time of implantation, but it can be revised at any time (see “Programming Patient Data into Pacemaker Memory” on page 8-9).
2-22
Figure 2-5.
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
Viewing the Patient Information Screen
Page 65
Conducting a Patient Session

Recording an ECG Strip of Magnet Operation

Recording an ECG Strip of Magnet Operation
With the Medtronic.Sigma pacemakers, positioning the programming head does not cause the pacemaker to operate in its magnet mode as has been the rule with previous Medtronic pacemakers (see “New Programmer Features” on page 2-2). To record or view magnet operation for Sigma models, you must use the Magnet test.

The Magnet Test Setup Screen

From the Magnet Test Setup screen, you can start and stop magnet operation in the pacemaker. Options let you collect an ECG strip of magnet operation and, if desired, non-magnet operation.
Magnet
Underlying Rhythm
Threshold
Exercise
Temporary
Sensing
Figure 2-6.
Magnet Test Setup Screen
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
2-23
Page 66
Conducting a Patient Session
Recording an ECG Strip of Magnet Operation
Note:
At any time during a Magnet test, lifting the programming head from over the patient’s pacemaker for at least 2 seconds restores operation of the pacemaker to its permanent status. This action should be taken in the event of programmer malfunction, loss of power, or the absence of an appropriate command confirmation.

Procedure for Conducting a Magnet Test

To Conduct a Magnet Test
1.
Display the Magnet Test Setup screen (see Figure 2-6 on page 2-23).
2.
To start magnet operation, position the programming head and select the
[START Magnet] button.
2-24
Observe that the live rhythm display shows the point at which magnet operation begins (“DOO, 85 ppm” for example). At this point, the pacemaker performs a Threshold Margin Test.
If you have chosen to collect an ECG strip (see “Collecting an ECG Strip” on page 2-25), a pop-up window shows progress during the collection process. Select the button if you want to stop the ECG collection before it completes. Selecting this button does not stop magnet operation.
3.
To stop magnet operation, select the
A pop-up window gives you the option to collect an ECG strip of non-magnet operation. Select magnet strip or select
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
[No] to close the window.
[STOP Magnet] button.
[Stop Collection]
[Yes] to collect a non-
Page 67
Conducting a Patient Session
Recording an ECG Strip of Magnet Operation

Collecting an ECG Strip

The programmer automatically collects an ECG strip during the Magnet test unless you cancel this option.
A check mark (✔) appearing in the Magnet Strip check box indicates that a strip will be collected. This is the default status when you open the screen.
To change the length of the strip, select the time field and choose the collection time you desire.
If you do not want automatic strip collection, select the Magnet Strip check box to clear the check mark.

About the Collected ECG Strips

If you have chosen to collect a Magnet strip or both a Magnet and Non-Magnet strip, the strips are stored by the programmer for viewing and printing. A Magnet Strip (and Non-Magnet Strip) icon appears at the bottom of the screen to indicate when strips are available for viewing and printing.
To View a Collected ECG Strip
Select the Magnet Strip or the Non-Magnet Strip icon near the bottom of the screen. Refer to “Recalling and Viewing Waveform Strips” on page 3-28 for information about using the strip viewing feature.
To Print a Magnet Test Report
Select the printing options. Refer to “Printing the Frozen Strip” on page 3-27 for information about printing waveform strips and reports.
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
[Print...] button and then choose the desired
2-25
Page 68
Conducting a Patient Session

Checking the Patient’s Underlying Rhythm

Checking the Patient’s Underlying Rhythm
To evaluate a patient’s underlying rhythm or determine the patient’s intrinsic heart rate, use the Underlying Rhythm test.
Magnet
Underlying Rhythm
Threshold
Exercise
Temporary
Sensing
Caution:
The use of this test function is intended for diagnostic and test purposes. It should be used only under conditions of careful patient monitoring and control.

The Underlying Rhythm Test Screen

The Underlying Rhythm Test screen provides two means for evaluating the patient’s underlying rhythm: 1) the Inhibit test and
2) the Manual Rate Decrease test.
2-26
Figure 2-7.
Underlying Rhythm Test Screen
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
Page 69
Conducting a Patient Session
Checking the Patient’s Underlying Rhythm
Note:
At any point during use of the Underlying Rhythm test, lifting the programming head away from the site of the patient’s pacemaker for at least 2 seconds will restore normal pacemaker operation. In the event of a programmer malfunction or loss of power, lift the programming head immediately.
Procedure for Checking the Patient’s Underlying Rhythm
Note:
The Inhibit mode basically shuts the pacemaker off for the duration of the test.
To Check Underlying Rhythm Using Inhibit
1.
Display the Underlying Rhythm Test screen (see Figure 2-7 on page 2-26).
2.
To stop the output of pacing stimuli:
a. Position the programming head. b. Press and hold the
Pacemaker blanking periods are shortened during this test to increase the period during which cardiac events can be sensed.
3.
To restore pacing, release the
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
[INHIBIT Press and Hold] button.
[INHIBIT Press and Hold] button.
2-27
Page 70
Conducting a Patient Session
Checking the Patient’s Underlying Rhythm
To Check Underlying Rhythm Using Manual Rate
Decrease
1.
Display the Underlying Rhythm Test screen (see Figure 2-7 on page 2-26).
2.
Choose the desired test pacing mode and the starting value for Lower Rate:
a. Select the Mode field to display the test mode options. b. Select the test pacing mode from the options displayed. c. Select the or button to adjust the Test Value for
Lower Rate to a setting close to the patient’s present pacing rate.
3.
Select the rate.
[START Test] button to engage the test mode and
2-28
4.
Select the button to gradually decrease Lower Rate until the patient’s underlying rhythm emerges. (Select the
button to increase Lower Rate as desired.)
5.
To end the test, select the
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
[STOP and Restore] button.
Page 71
Conducting a Patient Session
Checking the Patient’s Underlying Rhythm

Collected ECG Strips

During either type of Underlying Rhythm test, the programmer automatically collects and saves a 10-second ECG strip. A Test Strip icon appears near the bottom of the screen (after the test has ended) to indicate its availability.
To View a Collected ECG Strip
Select the Test Strip icon near the bottom of the screen. Refer to “Recalling and Viewing Waveform Strips” on page 3-28 for information about using the strip viewing feature.
To Print an Underlying Rhythm Test Report
Select the printing options. Refer to “Printing the Frozen Strip” on page 3-27 for information about printing waveform strips and reports.
[Print…] button and then choose the desired
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
2-29
Page 72
Conducting a Patient Session

Printing Data Reports

Printing Data Reports
You can print the data generated during a patient session using either of the following methods.
Using the Print Button – On most task screens, the button is available when there is a report available to print. Selecting this report associated with the current task screen.
Activating the “Trace Mode” – If you select the Trace printing mode, data is automatically printed as soon as the task data is available. This data is not printed in a report format.
Additional reports are available for printing from the Available Reports window. This window lists reports that are not associated with a specific task. Test results may also be available from this window.
As described in this section, you have the option to print full­page-sized reports. For this, the programmer must be connected to a compatible external printer. Refer to “Connecting an External Printer” on page 8-11 for instructions on connecting an external full-page printer.
[Print…] button allows you to print a formatted data
[Print…]

Using the Print Button

The [Print…] button appears or becomes active as soon as the execution of a task or function generates the data needed for the associated report. The the task screen.
[Print…] button appears near the bottom of
[Print…]
or becomes active when
2-30
button appears
task-related data has
been generated.
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
Selecting this button displays options for printing only the report associated with the current task screen.
Page 73
Conducting a Patient Session
To Print a Report Using the Print Button
1.
Select the
2.
From the Print Options window, make the following selections:
a. Select the Number of Copies field if you want to change
b. Select the desired paper size. Full Size requires that a
[Print…] button.
the number of copies to be printed. You can print from 1to 4 copies.
compatible, external printer be connected to the programmer (see “Connecting an External Printer” on page 8-11). Programmer Strip will print the report on the programmer’s strip chart printer.
Printing Data Reports
Note:
You can choose Full Size now but delay the printing until another time when the external printer is connected.
c. To print on an externally connected printer, the printer
type or “driver” listed in the Printer field must be compatible with the input requirements of the connected printer. Verify that the correct printer is selected.
3.
Select the desired print command:
To print the report immediately, select the
button.
To print the report later from the Print Queue window,
select
[Print Later]. See “Printing Reports Held in the Print
Queue” on page 2-33.
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
[Print Now]
2-31
Page 74
Conducting a Patient Session
Printing Data Reports

Printing Additional Reports

During a patient session, the programmer automatically accumulates data not associated with a specific task.
At any time during the session, you can view the list of additional reports that are available for printing and select the ones you want to print immediately or save in the print queue for printing later in the session or after the session has ended.
To Print From the List of Available Reports
1.
Available Reports
Print Queue
Select
Reports > Available Reports.
2-32
Note:
Device Memory Report is intended for use only by
Medtronic technical support personnel.
2.
From the displayed list of available reports, select the report you want to print.
3.
Select the
4.
From the Print Options window, choose the desired print options (see “Using the Print Button” on page 2-30).
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
[Print…] button.
Page 75
Check Box
Conducting a Patient Session
Printing Data Reports

Using the Continuous Print “Trace” Mode

If you want data to be printed as it becomes available during the patient session, you can activate the Trace printing mode. With this mode activated, data is printed as it is generated.
To Print Data Automatically Using the Trace Mode
1.
Select
Reports > Available Reports to display the Available
Reports window.
2.
From the Available Reports window, select the Print Continuously check box near the bottom of the window.
Appearance of a check mark (✔) indicates that the Trace printing mode is active.
3.
To turn off the Trace printing mode, select the check box again to clear the check mark.

Printing Reports Held in the Print Queue

Whenever you make a report print request by selecting the
[Print Now] button or the [Print Later] button, control of the print
job is passed on to the print queue.
By displaying the Print Queue window, you can check the status of any print job and either print or delete those print jobs that are being held. All print jobs held for printing are displayed and up to 25 print jobs printed already are reported.
Note:
The print queue is still available after you end the patient session. Between sessions the Print Queue window indicates that only reports from a Kappa pacemaker session are available, however, reports from the previous Kappa and Sigma sessions are available.
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
2-33
Page 76
Conducting a Patient Session
Printing Data Reports
To Print or Delete Reports From the Print Queue
1.
Available Reports
Print Queue
2.
Select
Reports > Print Queue.
From the list of reports, select the report you want to print or delete.
Caution:
3.
Select the button for the desired action:
–Select
–Select
Once deleted, a report cannot be recovered.
[Print] to print the report.
[Delete] to delete the report from the print queue
Report list.
2-34
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
Page 77

Ending a Patient Session

Because the programmer collects and stores data on a session-by­session basis, it is important to correctly end a session when you finish. You should not begin using the programmer with a different patient until you have completed the procedure described in this section.
Note:
If the programmer detects a serial number that is different from the one acquired during the initial interrogation, it will automatically end the current session.
Once you end a session, you can still access the Print Queue window to print reports from the session. Prior to ending the session, you have the options to:
Save the session data on a diskette.
Change the way the pacemaker clears collected data.
To End a Patient Session
Conducting a Patient Session
Ending a Patient Session
1.
Select the bottom of the screen.
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
[End Session] button in the command bar at the
2-35
Page 78
Conducting a Patient Session
Ending a Patient Session
2.
3.
4.
Note the “Warnings!” field at the top of the End Session window. It informs you of any action you should take or any information you should know before you end the session. To return to the session, select the
[Cancel] button.
Before ending the session, consider the options available from the End Session window.
To save the session data on a diskette, see “Saving a Session on Diskette” below.
To change the settings for how collected data is cleared from the pacemaker when you end the session, see “Options for Clearing Pacemaker Data” on page 2-39.
To accomplish either of these tasks, proceed to its appropriate procedure. Do not select the
[End Now] button in step 4.
To end the session now and return to the Select Model screen, select the
[End Now] button.
2-36
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
Page 79
Conducting a Patient Session
Ending a Patient Session

Saving a Session on Diskette

Before you end a patient session, you can save data from the session on a diskette. Table 2-5 lists the data saved. Data is saved as a text file that can be imported into a word processing or data base program.
Table 2-5.
Feature Name Information Exported
Therapy Parameters
Patient Information Final programmed values
Battery and Lead Measurements Last measured values
Threshold Tests Last results for each test type
Sensing Tests Last results for each test type
Exercise Test Last results for each test type
Automatic Diagnostics
To Save Session Data on a Diskette
1.
Insert a 3.5-inch computer diskette into the disk drive on the
Saved Session Data
Initial interrogated values Final programmed values
conducted (for each chamber tested)
conducted (for each chamber tested)
conducted Event Counters
Atrial High Rate Episodes (including Mode Switch Episodes) Ventricular High Rate Episodes
right side of the programmer.
Use a formatted diskette that is IBM-compatible. It can have a capacity of either 720 Kb (DS, DD) or 1.44 Mb (DS, HD).
2.
From the End Session window (see step 1 on page 2-35), select the
[Save Session] button.
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
2-37
Page 80
Conducting a Patient Session
Ending a Patient Session
3.
Note:
If the diskette is full, the programmer will notify you that you need to insert a different diskette.
In the pop-up window, select the
[Start] button.
This starts the transfer of data to the diskette. This process takes a few seconds as indicated by a progress gauge on the screen.
2-38
Note:
After session data has been saved, you can change the options for clearing data (see “Options for Clearing Pacemaker Data” below). However, you cannot cancel the End Session sequence and return to the session.
4.
When the transfer of data is complete, select the button to end the patient session. Remove the diskette from the programmer.
Note:
The programmer automatically assigns a unique file name so that it cannot overwrite any existing session file. To determine which file was saved for a particular session, use an IBM-compatible computer to view the diskette file directory. The directory lists the date and time each file was created.
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
[End Now]
Page 81
Conducting a Patient Session
Ending a Patient Session

Options for Clearing Pacemaker Data

All diagnostic data collected in the pacemaker except Chronic Lead Impedance Trend is cleared one hour after you end a patient session. Collection of new data then starts. You can restart a session within one hour of ending it and still retrieve the collected data.
The End Session window provides access to options that let you make the following changes to the default settings:
You can choose to have the Chronic Lead Trend data also cleared when you end the session.
Note:
Chronic lead trend data typically should not be cleared
unless the pacing lead has been replaced.
You can choose to program either of the following options under “Pacemaker Data” instead of the default setting “1 hr after session end”:
Selecting the “Now” option causes the pacemaker to clear
the collected data and start collection of new data immediately when you end the session. New data collection starts when you lift the programming head.
Selecting the “Don’t Clear” option allows the pacemaker
to retain the collected diagnostic data and resume data collection after you end the session.
To Change a Clear Data Setting
1.
From the End Session window (see step 1 on page 2-35), select the field labeled “Clear Data in Pacemaker and Restart Collection.” This action opens the Clear Data window.
2.
To change the setting for clearing Chronic Lead Trend data, complete a, b, and c, below. Otherwise, skip to step 3.
a. Select the appropriate Lead Trend field.
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
2-39
Page 82
Conducting a Patient Session
Ending a Patient Session
Atrial Lead
Trend Field
Ventricular Lead
Trend Field
3.
b. From the option window, select the “Clear” option.
Selecting “Clear” results in a message window. Select the
[OK] button to continue.
c. If applicable, repeat steps 2a and 2b for the other Lead
Trend setting.
To change when the pacemaker data is cleared and collection restarts, complete a, b, and c (if “Now” is selected) below. Otherwise, skip to step 4.
2-40
a. From the Clear Data window, select the field labeled
“Clear Data in Pacemaker and Restart Collection.”
Pacemaker
Data Field
b. From the option window, select the “Now” option or the
“Don’t Clear” option.
c. If you chose the “Now” option, select
pop-up window to proceed.
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
[Continue] from the
Page 83
Conducting a Patient Session
Ending a Patient Session
4.
To end the patient session, position the programming head and execute one of the following two procedures.
Procedure a — If you performed step 3 by selecting the “Now” option, the End Session screen displays the
[Clear and End] button. Select the [Clear and End] button to
program the selected options and end the session.
Diagnostic data collection starts when you lift the programming head.
Procedure b — If you did not perform step 3 or if you performed step 3 and chose the “Don’t Clear” option, proceed as follows:
a. Select the
options. (Selecting the
[PROGRAM] button to program the selected
[Undo Pending] button cancels the
selected options.
)
The selected Chronic Lead Trend data is now cleared from pacemaker memory. It can no longer be retrieved.
b. Select the c. Select the
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
[Close] button.
[End Now] button to end the patient session.
2-41
Page 84
Conducting a Patient Session
Ending a Patient Session
2-42
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
Page 85
Viewing the Patient’s ECG and EGM Traces
This chapter describes how to use the features of the live rhythm window to view the patient’s cardiac signals and monitor pacemaker operation. Also covered is the waveform Freeze feature, which lets you freeze a 15-second strip of the patient’s rhythm signals.
For instructions on using the programmer’s ECG chart printer, refer to“Recording an ECG Strip” on page 1-22.
3
Viewing the ECG and Other Rhythm Waveforms
Adjusting and Configuring the Display
Freezing and Analyzing a Waveform Strip
Recalling and Viewing Waveform Strips
3-7
3-18
3-28
3
3-2
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
3-1
Page 86
Viewing the Patient’s ECG and EGM Traces

Viewing the ECG and Other Rhythm Waveforms

Viewing the ECG and Other Rhythm Waveforms
During a patient session, the programmer continuously displays the patient’s ECG (or other selected waveform trace) in the live rhythm monitor window at the top of the screen.
The size of the window automatically changes between the single­trace size shown above and the dual-trace size shown below, depending on the selected task screen. The dual-trace window is displayed for test functions, in which case, the traces displayed are the patient’s ECG and the EGM in the chamber being tested.
3-2
The instructions on the following pages describe how you can:
Expand the monitor window to its full-screen size.
Adjust the size, position, and color (programmers with color display panels) of the waveform traces.
Choose a different sweep speed, display pacing artifacts or turn them off, engage an interference filter.
Program pacemaker telemetry to a different EGM mode.
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
Page 87
Viewing the Patient’s ECG and EGM Traces
Viewing the ECG and Other Rhythm Waveforms
Expanding the Live Rhythm Monitor Window
To view all of the available waveform traces, you can expand the ECG window to its full size with the touch of a button. The expanded screen covers the task area of the current screen.
To Expand the Monitor Window to Its Full Size
Select the square button in the upper-right corner of the partial-view window.
This action expands the live rhythm monitor window to its maximum size.
Figure 3-1.
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
Live Rhythm Monitor Window - Expanded View
3-3
Page 88
Viewing the Patient’s ECG and EGM Traces
Viewing the ECG and Other Rhythm Waveforms
To Return to the Partial-View Window
Select the minimize button in the upper-right corner of the full-view window.
This action collapses the monitor window to its previous size and displays the task screen as it was prior to expanding the monitor window.
3-4
Figure 3-2.
Live Rhythm Monitor Window - Partial View

About the Waveform Traces

The programmer can display traces from up to six signal sources. The ECG Leads (I, II, and III) are always available if the ECG leads are connected. Marker Annotation and Marker Intervals are displayed if the programming head is positioned over the pacemaker. These signal sources are labeled 1 - 5 in Figure 3-3.
The source of the sixth trace, also shown in Figure 3-3 as Ventricular EGM, is programmable to any one of four different telemetry options. (See “Programming a Different Telemetry Mode” on page 3-13.)
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
Page 89
Viewing the Patient’s ECG and EGM Traces
Viewing the ECG and Other Rhythm Waveforms
1
ECG signals detected via
skin electrodes attached
to the patient.
Sense and pace markers
received via telemetry
from the pacemaker.
Intervals measured
between the sense and
pace markers.
Programmable source.
Presently set to
Ventricular EGM.
2
3
4
5
6
Figure 3-3.
Waveform Trace Breakdown
The traces in Figure 3-3 are shown in a separated fashion for the purpose of explanation. Typically, the Marker Annotation and Marker Intervals traces are superimposed on an ECG or EGM trace to facilitate interpretation. Instructions on how to arrange the traces are covered on page 3-9 and page 3-15.
Note:
A programming command or interrogation momentarily interrupts the transmission of marker annotations. This interruption can result in missing markers on the trace display.
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
3-5
Page 90
Viewing the Patient’s ECG and EGM Traces
Viewing the ECG and Other Rhythm Waveforms
Table 3-1.
Trace Description ECG Lead I
ECG Lead II ECG Lead III
Marker Annotation
ECG signals are detected via skin electrodes attached to the patient. The programmer must be connected to these electrodes with the ECG cable as described in “Connecting the Programmer to Skin Electrodes” on page 1-6.
a
Marker annotations depict pacemaker operation by showing events as they occur within the pacemaker. These annotations, which are coded as shown below, are intended to facilitate ECG interpretation. The typical position of the Marker Annotation trace is superimposed on an ECG trace.
Dual Chamber Modes
- Atrial pace
AP
- Atrial sense
AS
- Atrial sense during refractory
AR
- Ventricular pace
VP
- Ventricular sense
VS
- Ventricular sense during refractory
VR
Other Markers
- Triggered pace (AAT or VVT pacing mode)
TP
- Mode Switch episode (marks the beginning and end)
MS
- Error marker (indicates a marker signal that could not be decoded)
ER
Marker Intervals
a
The programmer automatically measures the interval between pace and sense markers and displays these intervals (in milliseconds) as one of the traces. For dual chamber and ADI, ADIR, VDI, VDIR modes, the trace displays the A–V interval and the V–V interval. For single chamber modes, the trace displays the A–A or V–V interval depending on the chamber being paced.
Waveform Trace Information
Single Chamber Models
-Pace
P
-Sense
S
- Sense during refractory
SR
Programmable Trace
(
Telemetry Mode
a
)
The Telemetry Mode feature described on page 3-13 lets you program the pacemaker to transmit any one of four types of EGM signals to be displayed in the live rhythm monitor window. Refer to the instructions on page 3-13 for a description of each of these four EGM options.
- Atrial EGM
- Ventricular EGM
- Dual EGM
- Summed EGM
The Telemetry Mode feature is not available when the pacemaker is
Note:
at elective replacement indicator. Also, the Telemetry Mode feature can be programmed Off.
a
Since these traces depend on telemetry received from the pacemaker, they are not displayed unless the programming head is positioned over the pacemaker.
3-6
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
Page 91
Viewing the Patient’s ECG and EGM Traces

Adjusting and Configuring the Display

Adjusting and Configuring the Display
The Adjust window contains controls that let you tailor the live rhythm display to your specific needs. Note that once certain display attributes have been adjusted, the programmer will maintain these settings from one patient session to another.
To open the Adjust window, select the of the tool palette. This action expands the live rhythm display window (if it is not already expanded) and displays the Adjust window.
Figure 3-4.
Live Rhythm Display Adjust Window
[Adjust] button near the top
Refer to the pages that follow for instructions on using each of the controls in the Adjust window.
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
3-7
Page 92
Viewing the Patient’s ECG and EGM Traces
Adjusting and Configuring the Display
Adjust Size
Adjust Source
Adjust Color
Sweep speed
Telemetry Mode
ECG Filter
Table 3-2.
The Adjust Window Options
Select this button to adjust the amplitude of the waveforms (see “Adjusting Waveform Size” on page 3-9).
Select this button if you want to change the source of a particular trace (see “Selecting a Trace Source” on page 3-9).
Select this button if you want to change the color of one or more of the waveform traces (see “Color Coding the Waveform Traces” on page 3-11). Option is only available on programmers with color display panels.
Select the Sweep speed field to display the sweep speed options (12.5, 25, 50, and 100 mm/sec). To change the sweep speed, select the desired option (see “Selecting a Different Sweep Speed” on page 3-12).
Select the Telemetry Mode field to display the options for programming the type of waveform telemetry transmitted by the pacemaker (see “Programming a Different Telemetry Mode” on page 3-13).
Select this checkbox to turn the ECG filter on or off. A checkmark turned on. The filter may improve clarity of the ECG in the presence of interference (see “Selecting the ECG Filter” on page 3-14).
indicates that the filter is
(✓)
3-8
Show Artifacts
Cleanup
Normalize
OK
Select this checkbox to enable or disable the pacing artifact enhancement function. A checkmark enabled (see “Enabling Artifact Enhancement” on page 3-15).
Select this button to equalize the spacing between the traces
Select this button to equalize the trace spacing and to adjust the size of each trace to the default setting.
Select this button to close the Adjust window.
indicates that the function is
(✓)
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
Page 93
[Size]
button
Viewing the Patient’s ECG and EGM Traces
Adjusting and Configuring the Display

Adjusting Waveform Size

To Adjust the Size or (Amplitude) of a Waveform
1.
Select the
2.
For the trace you want to adjust, alternately press and release the increase button or decrease button to change the waveform size.
[Size] button in the Adjust window.
To return to the default setting, press the This action resets the size of all traces to the default setting and equalizes the spacing between traces.
[Normalize] button.

Selecting a Trace Source

You can change the order in which traces appear in the monitor window. One method is by changing the “source” of one or more of the traces as described below. Also, you can use the stylus to “drag” a trace to the desired location as described on page 3-15.
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
3-9
Page 94
Viewing the Patient’s ECG and EGM Traces
Adjusting and Configuring the Display
To Change the Source of a Trace
1.
Select the
[Source] button in the Adjust window.
[Source]
Source field
button
Note that traces may be superimposed on one another. In this example above, both the Lead II ECG and Marker Annotation signal make up the first trace. Also note that the name of only one of the trace sources appears at the left edge of the screen.
To display the name of the other trace source, “tap” the trace (source) name at the left edge of the screen. It will change to show the source of the other trace.
If you want to change the position of a superimposed trace, the source of the trace you want to move must show at the left edge of the screen. If it does not, alternate the source names as described in the paragraph above.
2.
Select the source field for the trace you want to change. A drop-down list appears displaying source options.
3-10
3.
From the drop-down menu of Source options, select the desired source.
Selecting a source option causes the source of the trace presently displayed to change positions with the trace source you selected.
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
Page 95
[Color]
Color field
button
Viewing the Patient’s ECG and EGM Traces
Adjusting and Configuring the Display

Color Coding the Waveform Traces

To Change the Color of a Waveform Trace
1.
Select the
2.
Select the color field for the trace you want to change.
[Color] button in the Adjust window.
3.
From the drop-down menu of Color options, select the desired color.
Note:
The option to color code trace waveforms is available only on programmers with color displays.
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
3-11
Page 96
Viewing the Patient’s ECG and EGM Traces
Adjusting and Configuring the Display

Selecting a Different Sweep Speed

From the Adjust window, you can set the trace sweep speed to one of four settings: 12.5, 25, 50, and 100 mm/sec. The default setting is 25 mm/sec.
To Change the Trace Display Sweep Speed
1.
Select the Sweep Speed field in the Adjust window.
Sweep Speed field
2.
From the drop-down menu of choices, select the desired speed.
Menu of choices
3-12
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
Page 97
Viewing the Patient’s ECG and EGM Traces
Adjusting and Configuring the Display

Programming a Different Telemetry Mode

In contrast to the traces that automatically appear in the live rhythm monitor window (the ECG, Marker Annotation, and Marker Intervals traces), the EGM trace can be programmed to any one of four Telemetry Mode options from the Adjust window.
To Program a Different Telemetry Mode
1.
Assure that the programming head is positioned over the patient’s pacemaker. Observe that marker annotations are appearing on the screen.
2.
Select the Telemetry Mode field in the Adjust window.
Telemetry Mode field
Menu of options
3.
From the drop-down menu of options, select the desired Telemetry Mode.
The Telemetry Mode options are as follows:
Atrial EGM – The intracardiac signal detected in the atria by the atrial pacing lead.
Ventricular EGM – The intracardiac signal detected in the ventricles by the ventricular pacing lead.
Dual EGM – Display of both the atrial and ventricular EGM traces. Since the pacemaker transmits both EGMs over a single telemetry channel in a “time sharing” fashion, resolution of the EGM signals in this case is reduced.
Summed EGM – The atrial and ventricular EGM signals combined (added together) into a single trace.
Off – No EGM trace is displayed.
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
3-13
Page 98
Viewing the Patient’s ECG and EGM Traces
Adjusting and Configuring the Display

Selecting the ECG Filter

Select the ECG Filter checkbox to turn the ECG filter on or off. A checkmark (✓) indicates that the filter is turned on.
The filter may improve clarity of both the displayed and printed ECG in the presence of interference. The filter affects the ECG detection bandwidth as follows:
Filter Off (no checkmark) – Bandwidth = 0.05 to 100 Hz Filter On (checkmark) – Bandwidth = 0.5 to 40 Hz
ECG Filter
checkbox
Show Artifacts
checkbox
3-14
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
Page 99
Viewing the Patient’s ECG and EGM Traces
Adjusting and Configuring the Display

Enabling Artifact Enhancement

Select the Show Artifacts checkbox to enable or disable the pacing artifact enhancement function. A checkmark (✓) indicates that the function is enabled.
The two trace illustrations in Figure 3-5 show how an ECG trace appears with and without this feature enabled.
Artifact Enhancement Enabled (Checkmark)
Artifact Enhancement Disabled (No Checkmark)
Figure 3-5.
Examples of Artifact Enhancement

Arranging the Waveforms

In addition to arranging the waveform traces by selecting the source as described on page 3-9, you can use the stylus to “drag” a trace to the desired position. The procedure in the example below shows how to move the Marker Annotation trace from its position over the ECG Lead II trace to a position over the Ventricular EGM trace.
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
3-15
Page 100
Viewing the Patient’s ECG and EGM Traces
Adjusting and Configuring the Display
To Move a Trace Using the Stylus
1.
First decide which trace you want to reposition. If the trace name is not displayed (Marker Annotation in this case), tap the name of the superimposed trace to display the hidden name.
2.
Press and hold the stylus against the name of the trace you want to move.
3-16
3.
Without lifting the stylus, drag the box now appearing around the trace name to the desired location.
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
Loading...