Medtronic SVDD303 Programming Guide

SIGMA
300/200/100Software 1.1
Vision™Software Model 9963
Pacemaker Programming Guide
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
A guide to using the 9790 Series programmer for Medtronic.Sigma 300/200/100 Series pacemakers
Refer to the Medtronic.Sigma 300/200/100 Series Pacemaker Reference Guide for information on the pacemakers.
The following terms are trademarks of Medtronic, Inc.
Marker Channel, Medtronic, Medtronic.Sigma, Medtronic.Vision, Auto-PVARP, Quick Look, Quick Secure, Sigma, Significant Events, and Vision.
Created by Cardiovascular Technical Communications
Medtronic, Inc. 1999
All Rights Reserved

How to Use This Guide

Information is Contained in Two Guides

Product information about Medtronic.Sigma Series software and pacemakers is presented in two separate guides.
The Pacemaker Programming Guide (PPG) accompanies Medtronic.Sigma Series software and contains instructions on how to use the programmer and the programming software.
The Pacemaker Reference Guide (PRG) is a supplementary guide that provides detailed information on Medtronic.Sigma Series pacemakers.
Note:
Refer to the technical manual supplied with the pacemaker
for information about which features are supported by a Medtronic.Sigma pacemaker.

About this Guide

How to Use This Guide
This guide presents the following information to use the 9790 programmer.
How to setup and configure the programmer and access on-line help.
How to start a patient session, use the various follow-up features during the session, and properly end the session.
How to view and print the patient’s ECG and EGM waveform traces.
How to configure the pacemaker to collect diagnostic data and how to retrieve and view this information.
How to measure stimulation thresholds and sensing levels.
How to program parameter values and verify rate response parameters settings.
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
iii
How to Use This Guide

About the Pacemaker Reference Guide

This supplementary guide describes in detail, how the pacemaker operates and specifies the capabilities of each model.
Describes the pacing modes, rate response options, special therapy features, telemetry types, and data collection options. In some cases, guidelines are given on how to configure the pacemaker operation.
Contains troubleshooting information for electrical and hemodynamic problems.
Specifies parameter and data collection capabilities, longevity projections, and mechanical and electrical specifications.
Provides general warning and cautions, potential interference sources, and general indications for pacing.
Contains a glossary of terms.
iv
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
Table of Contents
1. Programmer Basics
Table of Contents
How to Use This Guide iii
Information is Contained in Two Guides iii About this Guide iii About the Pacemaker Reference Guide iv
Setting Up the 9790 Programmer 1-2
Programmer Setup Check List 1-2
Connecting an External Monitor/Recorder 1-4 Connecting the Programmer to Skin Electrodes 1-6 Positioning and Using the Programming Head 1-7
When to Position the Programming Head 1-7
Effect of the Programming Head on Pacemaker
Operation 1-9
Observing Magnet Mode Operation 1-10
Alternative PROGRAM and
INTERROGATE Buttons 1-10
Automatic Interrogation at the
Start of a Session 1-11 About the Display Screen 1-12
Features and Conventions of the Display Screen 1-12
Using the On-Screen Keyboard 1-18
Audible Tones 1-20 Programming Emergency Parameters 1-21 Recording an ECG Strip 1-22
About the ECG Recording 1-23 Using the On-line Help Feature 1-25
Entering the Help System 1-26
Using the Links in Help 1-27
Exiting the Help System 1-28
Searching for Information in Help 1-28
2. Conducting a Patient Session
Important Reminders About Medtronic.Sigma Pacemakers 2-2
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
v
Table of Contents
New Programmer Features 2-2 Automated Monitoring and Collection of Patient
Diagnostic Data 2-3
Starting a Patient Session 2-4
The Starting Point of a Patient Session 2-4 The Tool Palette Between Sessions 2-5 About Automatic Interrogation 2-6 Procedure for Starting a Patient Session 2-7
Proceeding with Task Selection 2-10
The First Task Screen 2-10
Selecting Another Task or Function 2-10 Interrogating the Pacemaker 2-12 Taking a Quick Look at Pacemaker Operation 2-13
The Quick Look Screen 2-13
Viewing Data on the Quick Look Screen 2-14 Viewing Battery and Lead Measurements 2-17
Battery and Lead Measurements Screen 2-17
Pacemaker Battery Measurements 2-18
Lead System Measurements 2-19
Updating the Displayed Data 2-20 Checking the Present Parameter Settings 2-21
The Therapy Parameters Screen 2-21 Viewing Patient Information Stored in the Pacemaker 2-22 Recording an ECG Strip of Magnet Operation 2-23
The Magnet Test Setup Screen 2-23
Procedure for Conducting a Magnet Test 2-24
Collecting an ECG Strip 2-25
About the Collected ECG Strips 2-25 Checking the Patient’s Underlying Rhythm 2-26
The Underlying Rhythm Test Screen 2-26
Procedure for Checking the Patient’s Underlying
Rhythm 2-27
Collected ECG Strips 2-29 Printing Data Reports 2-30
Using the Print Button 2-30
Printing Additional Reports 2-32
Using the Continuous Print “Trace” Mode 2-33
vi
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
Table of Contents
Printing Reports Held in the Print Queue 2-33
Ending a Patient Session 2-35
Saving a Session on Diskette 2-37 Options for Clearing Pacemaker Data 2-39
3. Viewing the Patient’s ECG and EGM Traces
Viewing the ECG and Other Rhythm Waveforms 3-2
Expanding the Live Rhythm Monitor Window 3-3 About the Waveform Traces 3-4
Adjusting and Configuring the Display 3-7
Adjusting Waveform Size 3-9 Selecting a Trace Source 3-9 Color Coding the Waveform Traces 3-11 Selecting a Different Sweep Speed 3-12 Programming a Different Telemetry Mode 3-13 Selecting the ECG Filter 3-14 Enabling Artifact Enhancement 3-15 Arranging the Waveforms 3-15
Freezing and Analyzing a Waveform Strip 3-18
Using the On-Screen Calipers 3-19 Adjusting the Frozen Traces 3-21 Displaying a Marker Channel Diagram 3-22 Saving a Frozen Waveform Strip 3-27 Printing the Frozen Strip 3-27
Recalling and Viewing Waveform Strips 3-28
4. Collecting Diagnostic Data
About Data Collected by the Pacemaker 4-2
Types of Collected Data 4-2 Important Points about Using the Data Function 4-4
Displaying Collected Data 4-7
About the Collected Data - Graphs and Tables Screen 4-7 Automatically Collected Data 4-8
Choosing Clinician-Selected Detailed Data Collection 4-17
Off 4-17 Custom Rate Trend (Rate versus Time) 4-17 High Rate Detail (Atrial or Ventricular) 4-21
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
vii
Table of Contents
Programming Data Collection 4-23
Setup Options 4-23 Clearing Data From the Pacemaker 4-26
5. Evaluating Parameter Settings
Measuring Stimulation Thresholds 5-2
The Threshold Test Setup Screen 5-2
Setting Up a Threshold Measurement Test 5-3
Executing a Strength-Duration Threshold Test 5-6
Executing an Auto Decrement (Amplitude or Pulse Width)
Threshold Test 5-13
Conducting a Manual Threshold Test 5-15 Determining a Sensitivity Setting 5-18
The Sensing Test Setup Screen 5-18
Checking the Patient’s Intrinsic Rate 5-19
Executing an Automatic Sensing Test 5-20
Automatic Sensing Test Results 5-23
Executing a Manual Sensing Test 5-24 Using Temporary Programming to Evaluate
Parameter Settings 5-27
The Temporary Test Setup Screen 5-27
Procedure for Conducting a
Temporary Test 5-28
Programming a High Temporary Pacing Rate 5-30
viii
6. Programming Pacemaker Parameters
Programming Parameters 6-2
Selecting the Therapy Parameters Screen 6-2
Programming, Printing, and Clearing Parameter
Values 6-3
Parameter Restrictions 6-4
How to Program Parameters 6-6
How to Program Subordinate Parameters 6-8 Saving/Retrieving a Set of Parameter Values 6-13
Saving a Set of Parameter Values 6-14
Retrieving a Saved or Nominal Set of Parameter
Values 6-15
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
Table of Contents
7. Rate Response Setup
Using Exercise Test to Verify Rate Response 7-2
Overview of the Exercise Test 7-2 Selecting the Exercise Test 7-3 Starting the Exercise Test 7-3 Evaluating the Results of the Exercise Test 7-5 Adjusting Rate Response 7-6
8. Miscellaneous Setup Options - Programmer and Pacemaker
Selecting System Operating Preferences 8-2 Adjusting Programmer Time and Date 8-4 Improving the Detection of Pacing Artifacts 8-5 Checking the Software Version 8-6 Starting the Demonstrations Option 8-7 Programming Patient Data into Pacemaker Memory 8-9 Connecting an External Printer 8-11
Printer Compatibility 8-11 Materials You Will Need 8-13 Connecting the Printer 8-14
A. Appendix: Parameter Values and Restrictions
Programmable Modes and Parameters A-2 Automatic Diagnostics A-13 Clinician-Selectable Diagnostics A-15 Programming Restrictions A-18
I. Index
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
ix
Table of Contents
x
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
Programmer Basics
This chapter provides an overview of setting up the programmer and covers information you should be familiar with before you begin a patient session.
This information includes such things as using the programming head and stylus, programming Emergency parameters, and recording an ECG strip. Sections include an introduction to features of the display screen and use of the on-line Help system.
1
Setting Up the 9790 Programmer
1-2
1
For more detailed information on setting up your programmer, refer to Programmer Description and Setup, a guide supplied with your 9790 series programmer.
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
Connecting the Programmer to Skin Electrodes
Positioning and Using the Programming Head
About the Display Screen
Programming Emergency Parameters
Recording an ECG Strip
Using the On-line Help Feature
1-25
1-6
1-7
1-12
1-21
1-22
1-1
Programmer Basics

Setting Up the 9790 Programmer

Setting Up the 9790 Programmer
The following topics summarize the basic steps for setting up your 9790 series programmer. For more complete information on setting up your programmer, refer to the guide, Programmer Description and Setup, supplied with your programmer.

Programmer Setup Check List

Verify that the components illustrated below are properly connected to the programmer.
Selector Pen
disconnect the stylus.
1-2
Note:
Programming Head
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
(Stylus)
Do not
ECG Cable
Figure 1-1.
The Programmer
Programmer Basics
Setting Up the 9790 Programmer
Figure 1-2.
Figure 1-3.
Power Cord
Programming Head
Figure 1-4.
ECG Cable
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
1-3
Programmer Basics
Setting Up the 9790 Programmer

Connecting an External Monitor/Recorder

To connect an external monitor or recorder to your programmer, you need the optional adaptor shown below. Connect the adaptor to the analog output port on the right side of the programmer.
Calibrate Button
w
Warning:
Peripheral equipment connected to the 9790 programmer must be certified according to the applicable IEC standards (IEC 950 for data processing equipment and IEC 601-1 for medical equipment). The system formed by connecting peripheral equipment to the 9790 programmer must comply with IEC 601-1-1 for medical electrical systems. It is the responsibility of the person who connects the peripheral equipment to comply with IEC standards. It is the responsibility of the user to keep peripheral equipment that is certified to IEC 950 at least two meters away from the patient. Contact the peripheral equipment manufacturer for information about IEC certification.
Connector Type: Output Signal:
(switchable)
ABCD
±
1V
BNC
±1 Volt or ±5mV
Signal Output Adaptor
±
5mV
Connecting Optional Adaptor to the Analog
Output Port on the Programmer.
Figure 1-5.
1-4
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
Adaptor Output Signals
(after model selection):
- Patient’s ECG (upper most ECG
A
on display)
- Telemetered EGM (upper most EGM
B
on display)
- Telemetered EGM (other EGM
C
if programmed)
- Marker Channel Telemetry
D
Connecting the Optional External Adaptor
Programmer Basics
Setting Up the 9790 Programmer
Using the Calibrate Button
Selecting the Calibrate button located next to the analog output port adds a reference signal (as shown in Figure 1-6 and Figure 1-7) to the trace of Marker Channel and EGM telemetry.
The Marker Channel signal shows the relative marker amplitudes, which are not annotated with character codes on an external device. The EGM calibration signal acts as a voltage reference for the displayed EGM.
Atrial Pace (AP)
Atrial Sense (AS)
Atrial Refractory Sense (AR)
Vent. Refractory Sense (VR)
Vent. Sense (VS)
Vent. Pace (VP)
Figure 1-6.
5 mV
1 mV
Figure 1-7.
Marker Channel Signals
EGM Calibration Signals
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
1-5
Programmer Basics

Connecting the Programmer to Skin Electrodes

Connecting the Programmer to Skin Electrodes
At the start of a patient session, the programmer must be connected to skin electrodes on the patient. The ECG display and measurement functions will not operate without detection of the surface ECG signal.
Use an electrode lead to connect each skin electrode to the appropriate port on the ECG cable. Typical electrode placement is shown below. Use standard procedures for attaching disposable skin electrodes to the patient.
RA (R)
RL (N)
Figure 1-8.
Refer to the guide, Programmer Description and Setup, supplied with your 9790 programmer for more details on this procedure.
Connecting the Programmer to Skin Electrodes
LA (L)
C (C)
LL (F)
( ) = IEC Coding
1-6
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
Programmer Basics

Positioning and Using the Programming Head

Positioning and Using the Programming Head
In many of the procedures described in the following chapters, you will be directed to position the programming head over the patient’s pacemaker.

When to Position the Programming Head

You must position the programming head over the patient’s pacemaker whenever a procedure initiates communication between the programmer and pacemaker. Examples are:
At the start of a patient session when you select the
[Auto-Identify] or [Start] button. The programmer
automatically initiates an interrogation that takes a number of seconds to complete.
Prior to executing any command that results in a telemetry or programming transmission.
To view or record Marker Channel or EGM telemetry.
Caution:
Do not position the programming head over an implanted pacemaker during electrocautery or defibrillation procedures.
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
1-7
Programmer Basics
Positioning and Using the Programming Head
How to Position the Programming Head
The programming head must be properly positioned as described below. An incorrectly positioned programming head can result in the failure of a transmitted command and the loss of telemetry.
1.
Hold the programming head directly against the patient’s skin with the face of the programming head parallel to the pacemaker.
1-8
Figure 1-9.
2.
Position the programming head so that the amber light in
Positioning the Programming Head
the light array goes out and one or more of the green lights come on. Move the head a little in each direction to find the position that lights the greatest number of green lights. This is the optimum position.
Light Array
Figure 1-10.
Green
Amber/Green
Light Array Indicator
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
The present pacing mode shows in the status bar at
the top of the screen.
Programmer Basics
Positioning and Using the Programming Head
Effect of the Programming Head on Pacemaker Operation
If model selection and application (software) loading are complete, positioning the programming head over a Medtronic.Sigma pacemaker does not put the pacemaker in the magnet mode of operation, as is the case with previous pacemaker models. The programmer automatically sends a Cancel Magnet command to the pacemaker, which causes it to continue to operate as programmed.
An exception to this operation can occur if the programming head does not establish a telemetry link with the pacemaker (because of strong electrical interference or because the programming head is improperly positioned). In such cases, positioning the programming head would result in magnet mode operation until a telemetry link is established. You can easily verify the present pacing mode by observing the Status Line at the top of the screen.
Pacemaker operation returns to its programmed state about 2 seconds after you remove the programming head from its position over the pacemaker.
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
1-9
Programmer Basics
Positioning and Using the Programming Head

Observing Magnet Mode Operation

To observe Magnet mode operation during a patient session, you must conduct the Magnet test described in Chapter 2. Initiating the Magnet test results in a Threshold Margin Test (TMT) and causes the pacemaker to operate in the Magnet mode.
Note:
Between patient sessions before you select the pacemaker, positioning the programming head over a Medtronic.Sigma pacemaker will result in magnet mode operation.
Alternative
INTERROGATE
To initiate the Program and Interrogate commands, you have the option to use the on-screen the Program head.
[Interrogate]
Note:
The
[P] or [I] button on the programming head is active only
when its counterpart is displayed as an active button on the display screen.
PROGRAM
Buttons
[P] and Interrogate [I] buttons on the programming
Button
Figure 1-11.
Alternative Program and Interrogate
Programming Head Buttons
and
[Program] and [Interrogate] buttons or
Light Array
[Program]
Button
1-10
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
Programmer Basics
Positioning and Using the Programming Head

Automatic Interrogation at the Start of a Session

At the start of a session when you select [Start] or [Auto Identify], the programmer automatically interrogates the patient’s pacemaker for all the data contained within the pacemaker. For this process, you must position the programming head and hold it steady in place until the interrogation is complete.
At the bottom center of the screen you will see an indicator showing the progress of the interrogation. Because this interrogation retrieves all the data stored in the pacemaker, including diagnostic data, the process may take a number of seconds.
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
1-11
Programmer Basics

About the Display Screen

About the Display Screen
The programmer display screen is an interactive device that not only displays information in the form of both text and graphics, but functions as a control panel by displaying buttons and menu options that you can select using the stylus.
Features and Conventions of the Display Screen
This section describes the features and conventions of the display screen. The Therapy Parameters screen below shows the main elements of the typical screen.
Status Bar
Live Rhythm
Monitor Window
1-12
Task area
Active Field Button Tool PaletteCommand Bar
Figure 1-12.
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
Main Elements of a Display Screen Example
Status Bar
Programmer Basics
About the Display Screen
Note:
For information on changing the language in the screens (for example, from English to German), see “Selecting System Operating Preferences” on page 8-2.
The Status Bar
The status bar at the top of the screen shows:
The present pacing mode.
When any one of a number of test conditions is occurring.
The pacemaker model.
The Live Rhythm Monitor Window
Live Rhythm Monitor
Window
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
This window is a partial view of the full-screen display of ECG, Marker Channel, and telemetered EGM waveform traces. You can expand this window to its full size by selecting the small square button in the upper-right corner of the window or by selecting the
[Adjust...] button.
Heart rate and rate interval show if the programming head is positioned over the pacemaker.
Annotations above the waveform trace show the point of programmed parameter changes.
1-13
Programmer Basics
About the Display Screen
The waveform trace or traces that show in this window depend on the selected task screen and how traces have been arranged in the full-screen view. Refer to “Viewing the ECG and Other Rhythm Waveforms” on page 3-2 for information about the live rhythm monitor.
1-14
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
Programmer Basics
About the Display Screen
The Task Area
The portion of screen between the live rhythm monitor window at the top and the command bar at the bottom changes according to the task or function you select. The example in Figure 1-12 shows the Therapy Parameters screen for programming pacemaker parameter settings. This task area would appear much different if you selected, for example, the Threshold Test Setup screen.
The Command Bar
The bar at the bottom of the screen always shows the command buttons for programming “emergency” parameters settings, interrogating the pacemaker, and ending the patient session.
For complete information on these functions, see “Programming Emergency Parameters” on page 1-21, “Interrogating the Pacemaker” on page 2-12, and “Ending a Patient Session” on page 2-35.
Active fields show as
unshaded areas on
the screen.
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
Active Fields
Unshaded areas, or boxes, appearing in the task area are “active fields” that respond to the stylus.
Selecting a value, word, name, or phrase that appears in an active field opens a menu or window of alternative options for whatever is represented in that field.
1-15
Programmer Basics
S
About the Display Screen
electing an active field
opens a menu or window of
alternative options, or in
some cases, an
on-screen keyboard.
For example, touching the mode value “DDDR” in the Mode field with the tip of the stylus opens a window of mode options. Selecting any one of these options replaces the original DDDR value with the selected one.
Selecting some fields (those with terms ending in an ellipsis, such as “Rate Response...”) opens a window displaying additional fields. Some fields that require entry of information, such as patient data, open an on-screen keyboard. How to use this keyboard is described later in this section.
Buttons
Buttons like those shown below let you operate the programmer using the stylus. You can “press” a button by touching it with the tip of the stylus.
Buttons having a less
distinct shaded label are
not presently active.
1-16
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
Buttons may directly execute a command, such as the
[Program]
button, or they may open a window that prompts another action. Usually such buttons have a label ending with an ellipsis, such as the
[Save...] or [Get...] buttons shown above.
Buttons
Icons
Programmer Basics
About the Display Screen
A procedure may instruct you to “press and hold” a button. In such cases you should touch the tip of the stylus to the button and continue to maintain pressure against the button until it is time to “release” the button.
The Tool Palette
The collection of buttons and icons along the edge of the screen is referred to as the “tool palette.” These are the controls you will use to choose the task or function screen you want displayed. Once you have started a patient session, the tool palette is always displayed, making it quick and easy to move to the desired task or function.
Each of the icons acts like a button. To select an icon, touch the icon with the stylus. The “<” symbol adjacent to four of the icons indicates that selecting one of these icons opens a menu of related options. The icons without the < symbol directly open a task screen.
Refer to Table 2-2 on page 2-11 for a brief explanation of the purpose of each button and icon in the tool palette.
Note:
The Demo icon is available only when the programmer is operating in the Demo mode.
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
1-17
Programmer Basics
About the Display Screen

Using the On-Screen Keyboard

Certain fields on the screen allow you to enter data, such as the patient’s name or chart number. Selecting such a field automatically displays the on-screen keyboard shown below. By touching the letter or character buttons with the stylus, you can use this display feature like an actual keyboard.
Text Entry Window
Space Bar
Figure 1-13.
Keyboard Screen
Cursor
Position Keys
The function of the on-screen keyboard buttons are very similar to the keys on a computer keyboard or typewriter.
1-18
Medtronic.Sigma 300/200/100 Series Pacemaker Programming Guide
Loading...
+ 214 hidden pages