Medtronic.Sigma™ 300/200/100 Series
Pacemaker Reference Guide
Caution: Federal law (USA) restricts this device to sale
by or on the order of a physician.
A guide to the
Medtronic.Sigma
300/200/100 Series
pacemakers
Refer to the Medtronic.Sigma™
300/200/100 Pacemaker
Programming Guide for information
on software and programming.
™
The following terms are trademarks of Medtronic, Inc.
Auto PVARP, Marker Channel, Medtronic, Medtronic.Sigma, Medtronic.Vision, Quick Look,
Quick Secure, Sigma, Significant Events, and Vision.
How to Use This Guide
Information is Contained in Two Guides
Product information about Medtronic.Sigma 300/200/100 Series
software and pacemakers is presented in two separate guides.
The Pacemaker Reference Guide (PRG) is a supplementary guide
that provides detailed information on Medtronic.Sigma
300/200/100 Series pacemakers.
The Pacemaker Programming Guide (PPG) accompanies
Medtronic.Sigma 300/200/100 Series software and contains
instructions on how to use the programmer and the programming
software.
Note:
Refer to the technical manual supplied with the pacemaker
for information about which features are supported by a
Medtronic.Sigma pacemaker.
How to Use This Guide
About this Guide
This supplementary guide describes in detail, how the pacemaker
operates and specifies the capabilities of each model.
■
Describes the pacing modes, rate response options, special
therapy features, telemetry types, and data collection options.
In some cases, guidelines are given on how to configure the
pacemaker operation.
■
Contains troubleshooting information for electrical and
hemodynamic problems.
■
Specifies parameter and data collection capabilities, longevity
projections, and mechanical and electrical specifications.
■
Provides general warnings and cautions, plus potential
interference sources.
■
Contains a glossary of terms.
Medtronic.Sigma 300/200/100 Series Pacemaker Reference Guide
iii
How to Use This Guide
About the Pacemaker Programming Guide
This guide presents the following information to use the 9790
programmer.
■
How to setup and configure the programmer and access
on-line help.
■
How to start a patient session, use the various follow-up
features during the session, and properly end the session.
■
How to view and print the patient’s ECG and EGM waveform
traces.
■
How to configure the pacemaker to collect diagnostic data
and how to retrieve and view this information.
■
How to measure stimulation thresholds and sensing levels.
■
How to program parameter values and verify rate response
parameters settings.
iv
Medtronic.Sigma 300/200/100 Series Pacemaker Reference Guide
Table of Contents
1. Pacing Modes
Ta b l e of C on te n ts
How to Use This Guideiii
Information is Contained in Two Guidesiii
About this Guideiii
About the Pacemaker Programming Guideiv
Introduction1-2
Pacing Mode Selection1-2
NBG Pacing Codes1-2
Further Information1-3
Mode Selection Decision Tree:1-4
Mode Pertinency Tables1-5
Indications and Usage1-7
Contraindications1-9
DDDR Mode1-10
DDD Mode1-11
DDIR Mode1-12
DDI Mode1-13
DVIR Mode1-14
DVI Mode1-15
VDD Mode1-16
AAIR / ADIR Modes1-17
AAI / ADI Modes1-18
VVIR / VDIR Modes1-19
VVI / VDI Modes1-20
AAT / VVT Modes1-21
DOOR / AOOR / VOOR Modes1-22
DOO / AOO / VOO Modes1-23
ODO / OAO / OVO Modes1-24
Medtronic.Sigma 300/200/100 Series Pacemaker Reference Guide
v
Ta b le of Co nt en ts
Overview2-4
How Activity Threshold Influences Rate2-5
Rate Response Operation2-7
How Activity Acceleration and Deceleration Influence
Rate2-10
Exercise Deceleration Operation2-12
Adjusting Rate Response with the Exercise Test2-13
Overview2-13
3. Pacemaker Timing
Rates3-2
Overview3-2
A–A and V–V Timing3-3
Lower Rate3-4
Operating Lower Rate3-5
Selecting a Lower Rate3-5
Sensor-Indicated Rate3-6
Sensor-Indicated Rate Effect on Other Intervals3-7
Upper Tracking Rate3-8
Upper Sensor Rate3-8
Programming Considerations and Restrictions3-9
Rate Limit3-9
Possible Atrial Competition at High Rates3-10
Mean Atrial Rate3-10
AV Intervals3-12
Overview3-12
Selecting PAV and SAV3-14
Rate Adaptive AV3-16
Overview3-16
Programming for Rate Adaptive AV3-16
RAAV Operations3-17
RAAV Operation in the DDDR and DDD Modes3-19
RAAV and Sick Sinus Syndrome3-19
Medtronic.Sigma 300/200/100 Series Pacemaker Reference Guide
Single Chamber Atrial Blanking3-22
Refractory Periods3-23
Overview3-23
Post-Ventricular Atrial Refractory Period3-23
Auto PVARP (Tracking Modes)3-25
Auto PVARP (Nontracking Modes)3-26
Total Atrial Refractory Period (TARP)3-27
Ventricular Refractory Period3-28
Atrial Refractory Period (Single Chamber)3-29
Noise Reversion3-30
Preventing Noise Sensing3-32
Spontaneous PVARP Extension3-32
High Rate Atrial Tracking3-33
Overview3-33
2:1 Block3-33
Pacemaker Wenckebach3-34
High Rate Operation in the DDDR Mode3-36
4. Lead / Cardiac Tissue Interface
Selecting Pacing Parameters4-2
Overview4-2
Selecting Pacing Polarity4-2
Muscle Stimulation with Unipolar Pacing4-3
Bipolar Pacing Polarity Confirmation4-3
Determining Stimulation Threshold at Implant4-3
Verifying Stimulation Threshold at Follow-up4-4
Selecting Output Parameters4-4
For Further Information4-5
Selecting Sensing Parameters4-6
Overview4-6
Selecting Sensing Polarity4-6
Bipolar Sensing Polarity Confirmation4-7
Determining Sensing Threshold(s) at Implant4-7
Verifying Sensing Threshold(s) at Follow-up4-8
Selecting Sensitivity Settings4-8
Effects of Myopotentials During Unipolar Pacing4-9
For Further Information4-10
Ta b l e of C on te n ts
Medtronic.Sigma 300/200/100 Series Pacemaker Reference Guide
vii
Ta b le of Co nt en ts
Lead Monitor4-11
Overview4-11
How Lead Monitoring Works4-11
Lead Impedance Data4-13
Automatic Lead Impedance
(Chronic Lead Trend)4-13
Clinician-Selected Lead Impedance Detail4-14
For Further Information4-14
Transtelephonic Follow-up Features4-15
Overview4-15
The Threshold Margin Test (TMT)4-15
Threshold Margin Test Operation4-15
Enhanced Transtelephonic Monitoring4-17
For Further Information4-17
5. Special Therapy Options
Mode Switch and Diagnostic5-2
Overview5-2
How Atrial Tachyarrhythmia is Defined5-3
How Atrial Tachyarrhythmia is Detected5-3
Switching to Non-Atrial Tracking Mode5-3
Switching Back to Atrial Tracking Mode5-5
Mode Switching Interruption5-5
Programming Restrictions5-5
Recording Mode Switch Episode Data5-6
Non-Competitive Atrial Pacing5-7
Overview5-7
How NCAP Affects Atrial Timing5-7
How NCAP Affects Ventricular Timing5-8
NCAP Availability5-9
For Further Information5-9
PMT Intervention5-10
Overview5-10
How the Pacemaker Defines PMT5-10
PMT Therapy Intervention5-10
Automatic Therapy Suspension5-11
Interactions with Other Features5-12
viii
Medtronic.Sigma 300/200/100 Series Pacemaker Reference Guide
Patient Intervention for PMT5-12
For Further Information5-12
PVC Response5-13
Overview5-13
How the Pacemaker Defines a PVC5-13
Extending PVARP5-14
Interaction with Other Features5-14
PVCs Automatically Counted5-15
For Further Information5-15
Ventricular Safety Pacing5-16
Overview5-16
How VSP Operates5-17
Sleep Function5-18
Overview5-18
How the Sleep Function Works5-18
Interrupting the Sleep Function5-19
Programming Considerations5-19
Evaluating Sleep Function Operation5-19
Single Chamber Hysteresis5-20
Overview5-20
How Hysteresis Works5-20
Programming Considerations5-21
Interactions with Sleep Function5-21
Ta b l e of C on te n ts
6. Telemetry Data
Establishing Telemetry6-2
For Further Information6-2
Parameter Summary6-3
Overview6-3
Parameters Reported6-3
Possible Variation from Programmed Values6-4
For Further Information6-4
Patient Information6-5
Overview6-5
Parameters Reported6-5
Battery and Lead Information6-6
Overview6-6
Medtronic.Sigma 300/200/100 Series Pacemaker Reference Guide
ix
Ta b le of Co nt en ts
Telemetered Data6-6
Conditions and Variance in Measurements6-7
Chronic Lead Impedance Trend6-7
For Further Information6-7
Marker Channel Telemetry6-8
Overview6-8
Standard Marker Channel Telemetry6-8
Intracardiac Electrograms 6-9
Overview 6-9
Intracardiac Electrogram Recording6-10
Uses for the Intracardiac Electrogram6-11
For Further Information 6-11
Extended Telemetry6-12
Overview 6-12
Additional Battery Drain6-12
7. Miscellaneous Operations
Magnet Mode Operation 7-2
Overview 7-2
Magnet Mode Operation 7-2
Threshold Margin Test7-3
Transtelephonic Monitor Feature 7-3
Special Operation with Extended Telemetry7-4
For Further Information 7-4
Temporary Programming 7-5
Overview 7-5
Temporarily Programmable Parameters7-5
Temporary Refractory Period Settings 7-6
For Further Information 7-6
Electrical Reset 7-7
Overview 7-7
Reset Parameter Values7-7
Elective Replacement Indicator (ERI) 7-8
Overview 7-8
Basis for Setting ERI 7-8
ERI Verification7-8
Emergency Pacing 7-9
x
Medtronic.Sigma 300/200/100 Series Pacemaker Reference Guide
Ta b l e of C on te n ts
8. Diagnostics
Introduction to Diagnostics8-2
Automatic Diagnostics8-3
Clinician-Selected Diagnostics8-4
Battery / Lead Data 8-5
Suspending and Clearing of Data 8-5
For Further Information 8-6
Atrial and Ventricular Rate Histograms8-7
Automatic Data Collection8-8
Retrieving the Atrial and Ventricular Rate Histograms8-9
For Further Information 8-9
AV Conduction Histograms8-10
Automatic Data Collection8-11
Retrieving the AV Conduction Histogram 8-12
For Further Information 8-12
Atrial and Ventricular High Rate Diagnostics8-13
Automatic Data Collection8-13
Programmable Data Collection8-14
How High Rate Episodes Are Defined 8-15
Limitation to Detect High Rate Atrial Events8-16
Retrieving Atrial and Ventricular
High Rate Diagnostics8-17
For Further Information 8-17
Custom Rate Trend (Rate Versus Time)8-18
Identifying the Cause of an Electrical Problem 9-3
Correcting an Electrical Problem 9-5
Troubleshooting Hemodynamic Problems9-6
Medtronic.Sigma 300/200/100 Series Pacemaker Reference Guide
xi
Ta b le of Co nt en ts
Defining a Hemodynamic Problem 9-6
Identifying the Cause of a Hemodynamic Problem 9-6
Correcting a Hemodynamic Problem 9-8
Handling, Storage, and Resterilization9-9
Handling and Storage9-9
Resterilization9-9
Pacemaker Longevity 9-10
Background9-10
Elective Replacement Indicator 9-10
Time from ERI to Cessation of Pacing9-10
Distinguishing ERI from Electrical Reset9-11
For Further Information 9-11
Replacing the Pacemaker 9-12
For Further Information 9-12
Patient Information and Service 9-13
Patient Registration Information9-13
Establishing a Patient Record 9-13
Assistance Information9-14
For Further Information 9-14
xii
A. Pacemaker Description
Model Number Designator A-2
Radiopaque Codes A-3
Physical DimensionsA-4
Connector Dimensions A-5
Programmable Modes and ParametersE-2
Temporary Parameters E-11
Timing Reference E-13
Dual Chamber Timing SummaryE-17
Ta b l e of C on te n ts
F. Warnings and Precautions, and EMI
Special Notice F-2
WarningsF-4
Programming and Pacemaker OperationF-4
Pacemaker Dependent Patients F-4
Precautions F-6
Programming and Pacemaker OperationF-6
Rate IncreasesF-8
Unipolar SensingF-8
Implantable DefibrillatorF-9
Potential ComplicationsF-10
Potential EventsF-12
Hospital or Medical Environment InterferenceF-13
Therapeutic Diathermy F-13
Medtronic.Sigma 300/200/100 Series Pacemaker Reference Guide
xiii
Ta b le of Co nt en ts
Magnetic Resonance Imaging F-13
Electrosurgical Cautery F-13
External Defibrillation F-15
Diagnostic X-ray and Fluoroscopic Radiation F-15
LithotripsyF-15
Radiofrequency AblationF-16
Home and Job Environment InterferenceF-17
High Voltage Power Transmission Lines F-17
Communication EquipmentF-17
Commercial Electrical Equipment F-17
Home AppliancesF-17
Electronic Article Surveillance (EAS) F-18
Cellular PhonesF-18
G. Glossary
H. Index
xiv
Medtronic.Sigma 300/200/100 Series Pacemaker Reference Guide
Understanding
Pacemaker Operation
Chapters 1 - 9 provide detailed
information about the operation of
the Sigma 300/200/100 Series
pacemakers.
Pacing Modes
Rate Response
Pacemaker Timing
Lead/Cardiac Tissue Interface
Special Therapy Options
Telemetry Data
Miscellaneous Operations
Diagnostics
Troubleshooting The
Pacing System
Pacing Modes
This chapter provides information about the modes
available with the pacemaker.
1
1
Introduction
Mode Selection
Decision Tree:
Mode Pertinency Tables
Indications and Usage
Contraindications
DDDR Mode
DDD Mode
DDIR Mode
DDI Mode
DVIR Mode
DVI Mode
Medtronic.Sigma 300/200/100 Series Pacemaker Reference Guide
1-2
1-4
1-9
1-10
1-11
1-12
1-13
1-14
1-15
1-7
1-5
VDD Mode
AAIR / ADIR Modes
AAI / ADI Modes
VVIR / VDIR Modes
VVI / VDI Modes
AAT / VVT Modes
DOOR / AOOR / VOOR
Modes
DOO / AOO / VOO Modes
ODO / OAO / OVO Modes
1-16
1-17
1-18
1-19
1-20
1-21
1-22
1-23
1-24
1-1
Pacing Modes
Introduction
Introduction
Pacing Mode Selection
This chapter provides an introduction to pacemaker modes as an
aid to pacing mode selection. The chapter is organized as follows:
Mode Selection Decision Tree – This decision tree, based on the
1991 ACC/AHA guidelines for pacemaker implantation,
1
provides a simple means of identifying pacing modes appropriate
for given indications.
Mode Pertinency Tables – These tables show which features and
parameters apply to each commonly used pacing mode.
Indications and Contraindications – These sections provide the
indications, usage, and contraindications for using the
Medtronic.Sigma pacemakers.
Mode Descriptions – These descriptions indicate the modes
available with the pacemaker and how they operate.
NBG Pacing Codes
The pacemaker modes are defined in the NBG code.2 Each fiveletter NBG code describes a specific type of operation for
implantable pacemakers. For simplicity, this manual uses only the
first three or four letters, such as DDD, DDIR, DVIR. Figure 1-1
describes the first four letters of the NBG code.
1-2
1
Dreifus LS, Fisch C, Griffin JC, et al. Guidelines for implantation of cardiac
pacemakers and antiarrhythmia devices. A report of the American College of
Cardiology/American Heart Association Task Force on Assessment of
Diagnostic and Therapeutic Cardiovascular Procedures (Committee on
Pacemaker Implantation).
18: 1-13.
2
Bernstein A., et al., “The NASPE/BPEG Pacemaker Code,” PACE, 10(4), Jul-Aug
1987. (“NBG” stands for The North American Society of Pacing and
Electrophysiology [NASPE] and the British Pacing and Electrophysiology
Group [BPEG] Generic. NBG’s five-letter code supersedes the ICHD Code.
Journal of the American College of Cardiology.
Medtronic.Sigma 300/200/100 Series Pacemaker Reference Guide
1991;
Pacing Modes
Introduction
CHAMBER PACED
V = Ventricle
A = Atrium
D = Dual Chamber
S = Single Chamber
O = None
CHAMBER SENSED
V = Ventricle
A = Atrium
D = Dual Chamber
S = Single Chamber
O = None
Figure 1-1.
DDDR
NBG Pacing Codes
MODE OF RESPONSE
T = Triggered
I = Inhibited
D = Double (Both)
O = None
PROGRAMMABLE/
RATE RESPONSE
P = Programmable
M = Multiprogrammable
C = Communicating
R = Rate Responsive
O = None
Further Information
The mode descriptions in this chapter provide only a basic
overview of each mode. For further details on the rate response,
timing and therapy capabilities refer to “Rate Response” on
page 2-1, “Pacemaker Timing” on page 3-1, and “Special Therapy
Options” on page 5-1, respectively.
Medtronic.Sigma 300/200/100 Series Pacemaker Reference Guide
1-3
Pacing Modes
Mode Selection Decision Tree:
Mode Selection Decision Tree:
Figure 1-2 shows a basic decision tree used to select the pacing mode. In the shaded boxes the
preferred mode(s) is listed and the alternate mode(s) appears below the dashed line.
Symptomatic
Bradycardia
No
Complete or
transient AV
block
s
S
I
n
A
o
i
o
n
p
d
e
q
u
d
r
e
s
e
e
a
t
?
1st, 2nd, or
3rd degree
normal
f
u
n
c
t
a
NoNo
1st, 2nd, or
3rd degree
block, sick
sinus
syndrome
DDDRDDDR
conduction
e
n
y
l
t
block,
sinus
DDD
VDD
Is AV
intact?
Ye s
conduction,
No
tachyarrhythmias
Ye s
Normal AV
conduction
Is SA node
function presently
adequate?
Normal AV
sick sinus
syndrome
(no SVTs)
AAIR
DDDR
Are atrial
present?
Normal AV
conduction
DDD
Ye s
Ye sN o
Is AF chronic?
paroxysmal SVT
Is AV conduction
NoYesYe s
Sick sinus
syndrome,
chronic AF
VVIR
VVI
Sick sinus
syndrome,
paroxysmal
SVT, 3rd
degree
block
DDDR with
mode
switching
Sick sinus
syndrome
intact?
Sick sinus
syndrome,
paroxysmal
SVT, 1st, 2nd
degree block
or normal
conduction
DDIR with
automatic
PVARP
1-4
Figure 1-2.
Mode Selection Tree
Medtronic.Sigma 300/200/100 Series Pacemaker Reference Guide
Pacing Modes
Part# 197839001 Rev B
Mode Pertinency Tables
✓
✓✓✓✓
–
✓✓
✓✓✓ ✓✓
✓✓✓✓ ✓
Mode Pertinency Tables
Table 1-1 and Table 1-2 show which pacing parameters and features apply to each pacing mode as indicated by black
check marks. Dashes indicate parameters that are programmable when mode switch is active. These parameters are
programmable. Asynchronous modes are not shown in these tables.
Medtronic.Sigma 300/200/100 Series Pacemaker Reference Guide
✓✓✓
RhythmaPMT Intervention
✓✓✓✓✓
✓✓✓✓✓✓
PVC Response
Ventricular Safety Pacing
Special Pacing
OperationsaSingle Chamber
✓✓✓✓✓✓✓✓✓✓✓✓✓✓✓
Hysteresis
Sleep Function
Rate ResponsebRate Response Curve
✓✓
✓✓
✓✓
–
–
✓
✓
Activity Threshold
✓✓
–
–
✓
✓
Activity Acceleration
See Chapter 5 for operational descriptions of special therapy options.bSee Chapter 2 for operational descriptions of rate response features.
Activity Deceleration
a
Indications and Usage
Medtronic.Sigma pacemakers are indicated for the following:
■
Rate adaptive pacing in patients who may benefit from
increased pacing rates concurrent with increases in activity.
■
Accepted patient conditions warranting chronic cardiac
pacing which include:
–Symptomatic paroxysmal or permanent second or third-
degree AV block.
–Symptomatic bilateral bundle branch block.
–Symptomatic paroxysmal or transient sinus node
dysfunctions with or without associated AV conduction
disorders.
–Bradycardia-tachycardia syndrome to prevent
symptomatic bradycardia or some forms of symptomatic
tachyarrhythmias.
Pacing Modes
Indications and Usage
Medtronic.Sigma pacemakers are also indicated for dual chamber
and atrial tracking modes in patients who may benefit from
maintenance of AV synchrony. Dual chamber modes are
specifically indicated for treatment of conduction disorders that
require restoration of both rate and AV synchrony, which include:
■
Various degrees of AV block to maintain the atrial
contribution to cardiac output.
■
VVI intolerance (e.g., pacemaker syndrome) in the presence of
persistent sinus rhythm.
Medtronic.Sigma Series pacemakers are also indicated for VDD
modes in patients having adequate atrial rates and the following
indications:
Medtronic.Sigma 300/200/100 Series Pacemaker Reference Guide
1-7
Pacing Modes
Indications and Usage
■
Requirements for ventricular pacing when adequate atrial
rates and adequate intracavitary atrial complexes are present.
This includes the presence of complete AV block when atrial
contribution is needed for hemodynamic benefit or when
pacemaker syndrome has existed or is anticipated.
■
Normal sinus rhythm and normal AV conduction in patients
needing ventricular pacing intermittently.
1-8
Medtronic.Sigma 300/200/100 Series Pacemaker Reference Guide
Contraindications
Medtronic.Sigma pacemakers are contraindicated for the
following applications:
■
Dual chamber atrial pacing in patients with chronic refractory
atrial tachyarrhythmias.
■
Asynchronous pacing in the presence (or likelihood) of
competitive paced and intrinsic rhythms.
■
Unipolar pacing for patients with an implanted cardioverterdefibrillator (ICD) because it may cause unwanted delivery or
inhibition of ICD therapy.
Pacing Modes
Contraindications
Medtronic.Sigma 300/200/100 Series Pacemaker Reference Guide
1-9
Pacing Modes
DDDR Mode
DDDR Mode
In the DDDR mode, the pacemaker tracks the faster of the intrinsic
atrial rate or the sensor-indicated rate. If the intrinsic rate is faster,
the DDDR mode provides atrial synchronous pacing; otherwise,
AV sequential pacing occurs at the sensor-indicated rate.
■
Rate limits for atrial tracking (Upper Tracking Rate)1and
sensor tracking (Upper Sensor Rate) are separately
programmable.
■
The AV intervals that follow sensed atrial events (SAV) and
paced atrial events (PAV) are separately programmable, and
they can be programmed to shorten with increasing rates
(Rate Adaptive AV).
■
A nonrefractory sensed event in either chamber inhibits
pacing in that chamber. A ventricular nonrefractory sensed
event in the VA interval that is not preceded by an atrial sense
(AS or AR) is a pacemaker-defined PVC, and starts a new VA
interval.
The Total Atrial Refractory Period (TARP) may limit the tracking rate to a lesser
Sensor-indicated
Interval
A
S
V
S
A
P
value.
Medtronic.Sigma 300/200/100 Series Pacemaker Reference Guide
A
S
V
P
V
P
200 ms
DDD Mode
Pacing Modes
DDD Mode
The DDD mode provides atrial synchronous pacing in the
presence of intrinsic atrial activity; otherwise, AV sequential
pacing occurs at the Lower Rate.
■
Each atrial paced or nonrefractory atrial sensed event starts an
AV interval and a lower rate interval. The AV intervals that
follow sensed atrial events (SAV) and paced atrial events
(PAV) are separately programmable, and the SAV may be
optionally programmed to shorten with increasing rate (Rate
Adaptive AV).
■
A ventricular paced event may track atrial sensed events up
to the programmed Upper Tracking Rate.
■
A ventricular nonrefractory sensed event in the VA interval
1
that is not preceded by an atrial sense (AS or AR) is a
pacemaker-defined PVC, and starts a new VA interval.
The Total Atrial Refractory Period (TARP) may limit the tracking rate to a lesser
value.
Medtronic.Sigma 300/200/100 Series Pacemaker Reference Guide
Lower Rate Interval
A
S
V
S
200 ms
A
P
1-11
Pacing Modes
DDIR Mode
DDIR Mode
The DDIR mode provides dual chamber, sensor-driven,
atrioventricular (AV) sequential pacing for heart rate variation
without atrial tracking.
■
Atrial pacing occurs at the sensor-indicated rate, with
ventricular pacing at the end of the PAV interval unless
inhibited.
■
An atrial event sensed outside the PVARP will inhibit a
scheduled atrial stimulus but will not start an AV interval.
That is, ventricular paced events after such sensed atrial
events occur at the sensor-indicated rate. The following
ventriculoatrial (VA) interval may be extended slightly to
avoid an increasing atrial paced rate.
■
A ventricular nonrefractory sensed event in the VA interval
starts a new VA interval.