Medtronic Strata Valve Reference Manual

Strata® Valve
ADJUSTMENT REFERENCE GUIDE
Strata® Valve Adjustment Tools
Preoperative Procedures
READING THE PERFORMANCE LEVEL
1 2
1. Position the Locator Tool above the valve while it is still in the sterile
packaging, so the blue arrows are pointing in the same direction as the arrow on the valve. Ensure that the white circular portion of the valve is located between the two lines labeled “Valve Mechanism.”
2. Set the Indicator Tool (compass) into the Locator Tool while aligning the red bands on the tools. Read the current Performance Level setting of the valve, indicated by the blue triangle mark on the Indicator Tool.
ADJUSTING THE PERFORMANCE LEVEL
3 4
3. Keeping the Locator Tool in position, remove the Indicator Tool. Place the Adjustment Tool (magnet) into the Locator Tool with the blue triangle pointing at the current Performance Level setting.
4. Turn the Adjustment Tool to the desired Performance Level setting.
5 6
5. Holding the Locator Tool in place, remove the Adjustment Tool
by withdrawing it straight up and place at least 60 cm (24”) away from the valve.
6. To conrm that the desired Performance Level setting change has been accomplished, repeat Step 2.