
Disassembly/Assembly M708348B192E Rev. B
Disassembly/Assembly M708348B192 Rev. B
Instructions for
GUIDE HANDLE,
P/N EX0711424
Instructions for
GUIDE HANDLE, P/N EX0711424
2014-01-23
Medtronic Sofamor Danek USA, Inc.
1800 Pyramid Place
Memphis, TN 38132
Telephone 800 933 2635 (In U.S.A.)
901 396 3133 (Outside U.S.A.)
Fax 901 396 0356
ENGLISH
IMPORTANT INFORMATION ON DISASSEMBLY/ASSEMBLY
INSTRUCTIONS FOR GUIDE HANDLE, P/N EX0711424
Refer to the section entitled “Reprocessing – General Considerations” in the Medtronic Reusable
Instruments Instructions for Use prior to reprocessing this instrument. After the instrument is
disassembled, continue with step 3 in the section entitled “Manual Cleaning Instructions-Cleaning
Instructions-Dedicated Cleaning Area”.
NOTE: Reassemble the instrument prior to sterilization.
ASSEMBLY INSTRUCTIONS
1. Align handle and sleeve.
2. Insert distal tip of handle at knurled end of the sleeve.

Medtronic Sofamor Danek USA, Inc.
1800 Pyramid Place
Memphis, TN 38132
Telephone 800 933 2635 (In U.S.A.)
901 396 3133 (Outside U.S.A.)
Fax 901 396 0356
IMPORTANT INFORMATION ON DISASSEMBLY/ASSEMBLY
INSTRUCTIONS FOR GUIDE HANDLE, P/N EX0711424
Refer to the section entitled “Reprocessing – General Considerations” in the Medtronic Reusable
Instruments Instructions for Use prior to reprocessing this instrument. After the instrument is
disassembled, continue with step 3 in the section entitled “Manual Cleaning Instructions-Cleaning
Instructions-Dedicated Cleaning Area”.
NOTE: Reassemble the instrument prior to sterilization.
ASSEMBLY INSTRUCTIONS
ENGLISH
2014-01-23
1. Align handle and sleeve.
2. Insert distal tip of handle at knurled end of the sleeve.
3. Advance sleeve up and thread clockwise onto the handle.
DISASSEMBLY INSTRUCTIONS
1. Turn sleeve counter-clockwise to unthread sleeve from the
handle.
2. Advance sleeve down shaft of handle.
3. Remove sleeve from handle.
©2014 MEDTRONIC SOFAMOR DANEK USA, Inc. All rights reserved.