IMPORTANT INFORMATION ON DISASSEMBLY/ASSEMBLY
INSTRUCTIONS FOR GUIDE HANDLE, P/N EX0711424
Refer to the section entitled “Reprocessing – General Considerations” in the Medtronic Reusable Instruments Instructions for Use prior to reprocessing this instrument. After the instrument is disassembled, continue with step 3 in the section entitled “Manual Cleaning Instructions-Cleaning Instructions-Dedicated Cleaning Area”.
NOTE: Reassemble the instrument prior to sterilization.
ASSEMBLY INSTRUCTIONS
1. Align handle and sleeve.
2. Insert distal tip of handle at knurled end of the sleeve.
1800 Pyramid Place Memphis, TN 38132 Telephone 800 933 2635 (In U.S.A.) 901 396 3133 (Outside U.S.A.) Fax 901 396 0356
IMPORTANT INFORMATION ON DISASSEMBLY/ASSEMBLY
INSTRUCTIONS FOR GUIDE HANDLE, P/N EX0711424
Refer to the section entitled “Reprocessing – General Considerations in the Medtronic Reusable Instruments Instructions for Use prior to reprocessing this instrument. After the instrument is disassembled, continue with step 3 in the section entitled Manual Cleaning Instructions-Cleaning Instructions-Dedicated Cleaning Area.
NOTE: Reassemble the instrument prior to sterilization.
ASSEMBLY INSTRUCTIONS
ENGLISH
2014-01-23
1. Align handle and sleeve.
2. Insert distal tip of handle at knurled end of the sleeve.
3. Advance sleeve up and thread clockwise onto the handle.
DISASSEMBLY INSTRUCTIONS
1. Turn sleeve counter-clockwise to unthread sleeve from the handle.
2. Advance sleeve down shaft of handle.
3. Remove sleeve from handle.