IMPORTANT INFORMATION ON DISASSEMBLY/ASSEMBLY
INSTRUCTIONS FOR CORTICAL WEDGE INSERTER, P/N 9010000570
Refer to the section entitled “Reprocessing - General Considerations” in the Medtronic Reusable
Instruments Instructions for Use prior to reprocessing this instrument. After the instrument is
disassembled, continue with Step 3 in the section entitled “Manual Cleaning Instructions - Cleaning
Instructions - Dedicated Cleaning Area.”
DISASSEMBLY
1. To remove the Inserter Shaft, rmly grip the Inserter Sleeve (9010000570-01) with one hand, and
the Inserter Knob (9010000572) with the other hand.

2. Turn the Inserter Knob counter clockwise repeatedly to remove.
3. Separate the Inserter Shaft from the Inner Threaded Shaft (9010000571).
ASSEMBLY
1. To assemble, reverse the steps in the previous section.