Medtronic Sonicision Quick Reference Manual

QUICK REFERENCE GUIDE
Sonicision™ Curved Jaw Cordless Ultrasonic Dissection System
TABLE OF CONTENTS
IN THE OPERATING ROOM
4 Assembly
5 Changing a Battery Pack During a Procedure
theBattery Pack
7 Device Status and Troubleshooting
8 Charger Indications andTroubleshooting
CLEANING AND STERILIZATION
10 Cleaning the Reusable Components
12 Sterilizing the Reusable Components
2
IN THE OPERATING ROOM
3
ASSEMBLY
1
SCRUBBED PERSON
Slide the generator into the opening at the back of the dissector until it makes contact with the front of the dissector
opening.
2
SCRUBBED PERSON
Hold the dissector shaft­rotation wheel. Turn the torque knob clockwise until it clicks twice.
Tip: Listen and feel for the two clicks. This
will conrm that the
generator is properly attached to the dissector.
After assembly
To ensure the device meets required essential performance, it should be tested after assembly as follows:
With the clamping jaws open, test the minimum power
mode (energy button depressed to rst position) and maximum power mode (energy button fully depressed).
3
SCRUBBED PERSON
Place the sterile battery insertion guide onto the open dissector handle
and hold rmly.
Tip: Make sure the battery insertion guide is completely seated on the dissector. Improper placement will make it
dicult to insert the
battery
.
If the device generates dierent pulsating tones for each
power mode and a green LED is present, assembly is complete and the device is ready to use.
If inaudible or undierentiated tones or a non-green LED
are present, troubleshooting must be performed.
4
CIRCULATING PERSON
Orient the battery as shown on the battery insertion guide. Insert the battery pack through the battery insertion guide into the handle of the dissector.
Tip: The scrubbed person should provide additional support by holding the battery insertion guide onto the dissector during
battery insertion.
5
CIRCULATING PERSON
After the battery pack is seated in the dissector, remove the now contaminated battery insertion guide from the
sterile eld.
6
SCRUBBED PERSON
Do not touch the battery pack. Close the battery compartment door until there is tactile feedback and an audible click.
Tip: Press down on the top of the battery compartment door directly over the battery door latch for easiest
closure.
4
CHANGING A BATTERY PACK DURING A PROCEDURE
1
SCRUBBED PERSON
Hold the device with the battery compartment door up. Press the battery door release button to
open the compartment.
2
SCRUBBED PERSON
Place the new sterile battery insertion guide over the exposed battery pack. Seat the guide onto the open dissector handle
and hold rmly.
Tip: Ensure the battery pack handle is facing towards the jaws of the device and away from the battery door hinge. Do not attempt to place the battery insertion guide while the handle is in the improper orientation. If the battery handle is facing the wrong direction, tilt the jaws of the instrument down
towards the oor. This will
cause the handle to fall into proper orientation.
3
CIRCULATING PERSON
Grasp the battery pack handle and pull the battery pack through the insertion guide, out of the dissector, and into a clean, nonsterile environment.
Tip 1: Make sure the battery insertion guide is completely seated on the dissector. Improper placement will make
the battery dicult
to remove.
Tip 2: The scrubbed person should hold the battery insertion guide
rmly on the dissector.
4
CIRCULATING PERSON
Orient the new battery pack as shown on the battery insertion guide. Insert the battery pack through the battery insertion guide into the handle of the dissector.
Tip: The scrubbed person should provide additional support by holding the battery insertion guide onto the dissector during
battery insertion.
5
CIRCULATING PERSON
After the battery pack is seated in the dissector handle, remove the now contaminated battery insertion guide from the
sterile eld.
6
SCRUBBED PERSON
Do not touch the battery pack. Close the battery compartment door until there is tactile feedback and an audible click.
Tip: Press down on the top of the battery compartment door directly over the battery door latch for
easiest closure.
After changing the battery pack
With the clamping jaws open, test the minimum power mode (energy button
depressed to rst position) and maximum power mode (energy button fully depressed).
If the device generates dierent pulsating tones for each power mode and a
green LED is present, assembly is complete and the device is ready to use.
If inaudible or undierentiated tones or a non-green LED are present,
troubleshooting must be performed.
5
Loading...
+ 9 hidden pages