Medtronic SFR3-4-20-05 Instructions for Use

INSTRUCTIONS FOR USE
Solitaire™ Platinum
Revascularization Device
DWGS71064-010 Rev. B (04/2019)
TABLE OF CONTENTS
Solitaire™ Platinum Revascularization Device
English . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Symbol Glossary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19
2
English en
4
Instructions for Use
Solitaire™ Platinum Revascularization Device
PRECAUTIONS
The Solitaire™ Platinum Revascularization Device should only be used by physicians trained in interventional neuroradiology and treatment of ischemic stroke.
Operators should take all necessary precautions to limit X-ray radiation doses to patients and themselves by using sucient shielding, reducing uoroscopy times, and modifying X-ray technical factors whenever possible.
Carefully inspect the sterile package and the Solitaire™ Platinum Revascularization Device prior to use to verify that neither has been damaged during shipment. Do not use kinked or damaged components.
The Solitaire™ Platinum Revascularization Device is not to be used after the expiration date imprinted on the product label.
Refer to the appropriate intravenous tissue plasminogen activator (IV t-PA) manufacturer labeling for indications, contraindications, warnings, precautions, and instructions for use.
Initiate mechanical thrombectomy treatment as soon as possible.
For indication 3, endovascular therapy with the device should be started within 16 hours of symptom onset.
For indication 3, users should validate their imaging analysis techniques to ensure robust and consistent results for assessing core infarct size.
DESCRIPTION
The Solitaire™ Platinum Revascularization Device is designed to restore blood ow by removing thrombus in patients experiencing ischemic stroke due to large intracranial vessel occlusion. The device is designed for use in the neurovasculature such as the internal carotid artery, M1 and M2 segments of the middle cerebral artery, basilar, and the vertebral arteries.
Figure 1:
Solitaire™ Platinum Revascularization Device
5
3
1
2
INDICATIONS
1. The Solitaire™ Platinum Revascularization Device is indicated for use to restore blood ow in the neurovasculature by removing thrombus for the treatment of acute ischemic stroke to reduce disability in patients with a persistent, proximal anterior circulation, large vessel occlusion, and smaller core infarcts who have rst received intravenous tissue plasminogen activator (IV t-PA). Endovascular therapy with the device should be started within 6 hours of symptom onset.
2. The Solitaire™ Platinum Revascularization Device is indicated to restore blood ow by removing thrombus from a large intracranial vessel in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for IV t-PA or who fail IV t-PA therapy are candidates for treatment.
3. The Solitaire™ Platinum Revascularization Device is indicated for use to restore blood ow in the neurovasculature by removing thrombus for the treatment of acute ischemic stroke to reduce disability in patients with a persistent, proximal anterior circulation, large vessel occlusion of the internal carotid artery (ICA) or middle cerebral artery (MCA)-M1 segments with smaller core infarcts (<70 cc by CTA or MRA, <25 cc by MR-DWI). Endovascular therapy with the device should start within 6-16 hours of time last seen well in patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy.
CONTRAINDICATIONS
Use of the Solitaire™ Platinum Revascularization Device is contraindicated under these circumstances.
Patients with known hypersensitivity to nickel-titanium.
Patients with stenosis and/or pre-existing stent proximal to the thrombus site that may preclude safe recovery of the Solitaire™ Platinum Revascularization Device.
Patients with angiographic evidence of carotid dissection.
POTENTIAL COMPLICATIONS
Possible complications include, but are not limited to the following:
Adverse reaction to antiplatelet/ anticoagulation agents or contrast media
Air Embolism
Allergic reactions Perforation or dissection of the vessel
Arteriovenous Fistula Persistent neurological decits
Brain Edema Post procedure bleeding
Change in mental status Pseudo aneurysm formation
Device(s) deformation, collapse, fracture or malfunction
Distal embolization including to a previously uninvolved territory
Hematoma and hemorrhage at puncture site
Infection Thrombosis (acute and subacute)
Inammation Vascular occlusion
Intracranial Hemorrhage Vasoconstriction (Vasospasm)
Ischemia
Neurologic deterioration including stroke progression, stroke in new vascular territory, and death
The risk of complication of radiation exposure (e.g., alopecia, burns ranging in severity from skin reddening to ulcers, cataracts, and delayed neoplasia) increases as the procedure time and the number of procedures increase
WARNINGS  ALL INDICATIONS
1. Proximal marker
2. Distal markers
3. Introducer sheath
4. Push wire
5. Fluorosafe marker
Model
Recommended Vessel
Diameter1 (mm)
min. max. (mm) (inch) (cm) (mm) (mm) (mm) (cm) Distal Prox. (mm)
SFR3-4-20-05
SFR3-4-20-10
SFR3-4-40-10
SFR3-6-20-10
SFR3-6-24-06
SFR3-6-40-10
1
Select a Solitaire™ Platinum Revascularization Device based on the sizing recommendations in Table 1 and based on the smallest vessel diameter at thrombus site.
2
Select a Solitaire™ Platinum Revascularization Device usable length that is at least as long as the length of the thrombus.
2.0 4.0 0.5 0.021 180 4.0 20.0 31.0 <130 3 1 5
2.0 4.0 0.5 0.021 180 4.0 20.0 31.0 <130 3 1 10
2.0 4.0 0.5 0.021 180 4.0 40.0 50.0 <130 3 1 10
2.0 5.5 0.7 0.027 180 6.0 20.0 31.0 <130 4 1 10
2.0 5.5 0.7 0.027 180 6.0 24.0 37.0 <130 4 1 6
2.0 5.5 0.7 0.027 180 6.0 40.0 47.0 <130 4 1 10
A. Stent diameter B. Stent length C. Length from distal marker to proximal marker D. Length from distal tip to uorosafe marker E. Marker to marker length
Solitaire™ Platinum Revascularization Device:
Product Specifications and Recommended Sizing Guidelines
Minimum Microcatheter ID
Push Wire
Length
The appropriate anti-platelet and anti-coagulation therapy should be administered in accordance with standard medical practice.
Table 1.
Stent
Diameter
Stent Length
2
Proximal
Marker to
Distal Marker
Length
Length from
Distal tip to
fluorosafe
marker
Radiopaque
Markers
Radiopaque
Stent Markers
Spacing
3
WARNINGS  ALL INDICATIONS
Administer IV t-PA as soon as possible for all patients who are indicated to receive the drug. Do not cause delays in this therapy.
Per IV t-PA manufacturer labeling, IV t-PA should be administered within 3 hours of stroke symptom onset
(IV t-PA use beyond 3 hours is not approved in the United States).
Do not torque the Solitaire™ Platinum Revascularization Device.
For vessel safety, do not perform more than three recovery attempts in the same vessel using Solitaire™ Platinum Revascularization Devices.
For device safety, do not use each Solitaire™ Platinum Revascularization Device for more than three ow restoration recoveries.
For each new Solitaire™ Platinum Revascularization Device, use a new microcatheter.
Solitaire™ Platinum Revascularization Device does not allow for electrolytic detachment.
To prevent device separation:
Do not oversize device.
Do not recover (i.e. pull back) the device when encountering excessive resistance. Instead, resheath
the device with the microcatheter and then remove the entire system under aspiration. If resistance is encountered during resheathing, discontinue and remove the entire system under aspiration.
Do not treat patients with known stenosis proximal to the thrombus site.
This device is supplied STERILE for single use only. Do not reprocess or re-sterilize. Reprocessing and re-sterilization increase the risks of patient infection and compromised device performance.
WARNINGS  INDICATION 1 & 3 ONLY
The safety and eectiveness has not been established for the Solitaire™ Platinum device to reduce disability in patients with the following:
Posterior circulation occlusions
More distal occlusions in the anterior circulation
Large core infarct (ASPECTS ≤7)
CLINICIAN USE INFORMATION
Materials Required
The following parts are required to use the Solitaire™ Platinum Revascularization Device:
Solitaire™ Platinum Revascularization Device with the SFR3-4-20-05, SFR3-4-20-10 and SFR3-4-40-10 reference numbers should be introduced only through a size “18” microcatheter with a minimum inside diameter of 0.021 inches.
Solitaire™ Platinum Revascularization Device with the SFR3-6-20-10, SFR3-6-24-06 and SFR3-6-40-10 reference numbers should be introduced only through a microcatheter with a minimum inside diameter of 0.027 inches.
Compatibility testing has been performed with the Rebar™ microcatheter. Refer to the instructions supplied with all interventional devices and materials to be used in conjunction with the Solitaire™ Platinum Revascularization Device for their intended uses, contraindications and potential complications.
Other accessories for performing a procedure and NOT supplied; to be selected based on the physician’s experience and preferences:
Minimum 5F guide catheter (NOTE: Users should select a guide catheter with appropriate support to deliver interventional devices. For a guide catheter a minimum inside diameter 0.061’’ is recommended. For a balloon guide catheter ensure to use 8-9F with minimum inside diameter 0.075”)
Microcatheter (refer to Table 1)
Guidewire
Aspiration Device (60 cc syringe or aspiration pump/system, such as Riptide™ Aspiration System) NOTE: The Riptide™ Aspiration System has been bench tested with the Solitaire™ Revascularization Device as an aspiration device during retrieval to the guide catheter. Other comparable aspiration pump systems have been reported in clinical literature, although these devices have not been specically tested with the Solitaire™ Revascularization Devices.
Saline solution/heparin-saline continuous ush set
Rotating Hemostasis Valve (RHV)
Infusion stand
Femoral arterial lock
PREPARATION AND PROCEDURE
Preparation
1. Administer anti-coagulation and anti-platelet medications per standard institutional guidelines.
2. Aided by angiographic radiography, determine the location and size of the area to be revascularized.
3. Select a Solitaire™ Platinum Revascularization Device per Table 1.
4. To achieve optimal performance of the Solitaire™ Platinum Revascularization Device and to reduce the risk of thromboembolic complications, maintain continuous ushing action between a) the femoral arterial sheath and the guide catheter, b) the microcatheter and the guide catheter and c) the microcatheter and the push wire and the Solitaire™ Platinum Revascularization Device. Check all connections to make sure that during the continuous ush no air enters the guide catheter or the microcatheter.
5. Position a suitable guide catheter as close to thrombus site as possible employing a standard method. The guide catheter should be appropriately sized to retrieve clot if so desired in subsequent steps. Connect a RHV to the tting of the guide catheter, and then connect a tube to the continuous ush.
6. With the aid of Table 1, select a microcatheter suitable for advancing the Solitaire™ Platinum Revascularization Device.
7. Connect a second RHV to the tting of the microcatheter and then connect a tube to the continuous ush.
8. Set the ush rate per standard institutional guidelines.
9. With the aid of a suitable guide wire, advance the microcatheter until the end of the microcatheter is positioned distal to the thrombus so that the usable length portion of the Solitaire™ Platinum Revascularization Device will extend past each side of the thrombus in the vessel when fully deployed. Tighten the RHV around the microcatheter.
Delivering the Solitaire™ Platinum Revascularization Device
10. Insert the distal end of the introducer sheath partially into the RHV connected to the microcatheter. Tighten the RHV and verify that uid exits the proximal end of the introducer sheath.
11. Loosen the RHV and advance the introducer sheath until it is rmly seated in the hub of the microcatheter. Tighten the RHV around the introducer sheath to prevent back ow of blood, but not so tight as to damage the Solitaire™ Platinum Revascularization Device during its introduction into the microcatheter. Conrm that there are no air bubbles trapped anywhere in the system.
12. Transfer the Solitaire™ Platinum Revascularization Device into the microcatheter by advancing the push wire in a smooth, continuous manner. Once the exible portion of the push wire has entered the microcatheter shaft, loosen the RHV and remove the introducer sheath over the proximal end of the push wire. Once completed, tighten the RHV around the push wire. Leaving the introducer sheath in place will interrupt normal infusion of ushing solution and allow back ow of blood into the microcatheter.
13. Visually verify that the ushing solution is infusing normally. Once conrmed, loosen the RHV to advance the push wire.
14. Once the Solitaire™ Platinum Revascularization Device has been advanced to the uorosafe marker band begin uoroscopic imaging. With the aid of uoroscopic monitoring, carefully advance the Solitaire™ Platinum Revascularization Device until its distal markers line up at the end of the microcatheter. The Solitaire™ Platinum Revascularization Device should be positioned such that the usable length portion of the device will extend past each side of the thrombus in the vessel when the device is fully deployed.
WARNING  ALL INDICATIONS
If excessive resistance is encountered during the delivery of the Solitaire™ Platinum Revascularization Device, discontinue the delivery and identify the cause of the resistance. Advancement of the Solitaire™ Platinum Revascularizaton Device against resistance may result in device damage and/or patient injury.
Deploying the Solitaire™ Platinum Revascularization Device
15. Loosen the RHV around the microcatheter. To deploy the Solitaire™ Platinum Revascularization Device, x the push wire to maintain the position of the device while carefully withdrawing the microcatheter in the proximal direction.
Figure 2:
Solitaire™ Platinum Revascularization Device Deployment
16. Retract the microcatheter until it is just proximal to the proximal marker of the Solitaire™ Platinum Revascularization Device. Tighten the RHV to prevent any movement of the push wire. The usable length of the deployed Solitaire™ device should extend past each side of the thrombus.
17. Tighten the RHV around the microcatheter. Angiographically assess the revascularization status of the treated vessel.
Revascularization Device Recovery
18. If using balloon guide catheter, inate guide catheter balloon to occlude vessel as specied in Balloon Guide Catheter labeling.
19. Ensure the proximal marker on the Solitaire™ Platinum Revascularization Device is within the microcatheter as in Figure 3 below. Loosen the RHV around the microcatheter enough for movement while still maintaining a seal.
Figure 3:
Solitaire™ Platinum Revascularization Device Recovery Position
4
20. To retrieve thrombus, slowly withdraw the microcatheter and the Solitaire™ Platinum Revascularization Device as a unit to the guide catheter tip while applying aspiration to the guide catheter with the aspiration device. Never advance the deployed Solitaire™ Platinum Revascularization Device distally.
NOTE: Ensure microcatheter covers proximal marker.
21. Apply vigorous aspiration to the guide catheter using the aspiration device and recover the Solitaire™ Platinum Revascularization Device and microcatheter inside guide catheter. Continue aspirating guide catheter until the Solitaire™ Platinum Revascularization Device and microcatheter are nearly withdrawn from the guide catheter.
NOTE: If withdrawal into the guide catheter is dicult, deate balloon (if using balloon guide catheter) and then simultaneously withdraw guide catheter, microcatheter and Solitaire™ Platinum Revascularization Device as a unit through the sheath while maintaining aspiration. Remove sheath if necessary.
WARNINGS  ALL INDICATIONS
If excessive resistance is encountered during recovery of the Solitaire™ Platinum Revascularization Device, discontinue the recovery and identify the cause of the resistance.
For vessel safety, do not perform more than three recovery attempts in the same vessel using the Solitaire™ Platinum Revascularization Device.
22. Open the guide catheter RHV to allow the microcatheter and the Solitaire™ Platinum Revascularization Device to exit without resistance. Use care to avoid interaction with the site of the intervention and to prevent air from entering the system.
23. Aspirate the guide catheter to ensure the guide catheter is clear of any thrombus material.
24. Deate guide catheter balloon if using balloon guide catheter.
25. If additional ow restoration attempts are desired with:
a. A new Solitaire™ Platinum Revascularization Device, then:
i. Repeat the steps described above starting with the “Preparation” section.
b. The same Solitaire™ Platinum Revascularization Device, then:
i. Clean the device with saline solution.
NOTE: Do not use solvents or autoclave.
ii. Carefully inspect the device for damage. If there is any damage, do not use the device and use a new
Solitaire™ Platinum Revascularization Device for subsequent ow restoration attempts following the steps described above starting with the “Preparation” section. Use of a damaged device could result in additional device damage or patient injury.
WARNING  ALL INDICATIONS
For device safety, do not use each Solitaire™ Platinum Revascularization Device for more than three ow restoration recoveries.
Solitaire™ Platinum Revascularization Device Re-Sheathing
If resheathing of the Solitaire™ Platinum Revascularization Device is necessary (e.g. repositioning), follow these steps:
WARNING  ALL INDICATIONS
Advancing the microcatheter while the device is engaged in clot may lead to embolization of debris.
Do not advance the microcatheter against any resistance.
Do not reposition more than two times.
1. Loosen the RHV around the microcatheter and around the push wire. With the aid of uoroscopic monitoring, hold the push wire rmly in its position to prevent the Solitaire™ Platinum Revascularization Device from moving.
2. Carefully re-sheath the Solitaire™ Platinum Revascularization Device by advancing the microcatheter over the Solitaire™ Platinum Device until the distal markers of the Solitaire™ Platinum Device line up at the end of the microcatheter as illustrated in Figure 4 below. If significant resistance is encountered during the re-sheathing process, stop immediately and proceed to the section above entitled “Revascularization Device Recovery”.
Figure 4:
Solitaire™ Platinum Revascularization Device Re-sheathing
HOW SUPPLIED
The Solitaire™ Platinum Revascularization Device is provided sterile for single patient use only.
Sterile: This device is sterilized with Ethylene Oxide. Non-pyrogenic. Contents: One (1) Solitaire™ Platinum Revascularization Device. Storage: Store product in a dry, cool place.
SAFETY AND EFFECTIVENESS INFORMATION  INDICATION 1
SWIFT PRIME Clinical Study
SWIFT PRIME (Solitaire™ FR or Solitaire™ 2 With the Intention For Thrombectomy as Primary Endovascular Treatment for Acute Ischemic Stroke) is a global, multicenter, two-arm, prospective, randomized, open, blinded endpoint (PROBE) clinical study comparing neurological disability outcomes (dened by mRS) in Acute Ischemic Stroke (AIS) patients who are treated with either IV t-PA alone or IV t-PA in combination with Solitaire™ FR or Solitaire™ 2 mechanical thrombectomy intervention. Subjects receiving IV t-PA within 4.5 hours of symptom onset were randomized 1:1 to mechanical thrombectomy with Solitaire™ within 6 hours of onset, or to continuation with IV t-PA alone. Within this group, the Analysis cohort was dened as those subjects who were administered IV t-PA within 3 hours of symptom onset.
Inclusion Criteria
Subjects were considered eligible for the study if they satised all of the following inclusion criteria:
Subject or subject’s legally authorized representative has signed and dated an Informed Consent Form
Age 18 – 80
Clinical signs consistent with acute ischemic stroke
Pre-stroke Modied Rankin Score less than or equal to 1
National Institute of Health Stroke Scale (NIHSS) score of at least 8 and less than 30 at the time of randomization
Initiation of IV t-PA within 4.5 hours of onset of stroke symptoms
Thrombolysis in Cerebral Infarction (TICI) 0 to 1 ow in the intracranial internal carotid artery, M1 segment of the MCA, or carotid terminus conrmed by CT or MR angiography that is accessible to the Solitaire™ FR or Solitaire™ 2 device
Subject is able to be treated within 6 hours of stroke symptoms onset and within 1.5 hours from CTA or MRA to groin puncture
Subject is willing to conduct follow-up visits
Exclusion Criteria
Subjects were considered ineligible for study participation if they met any of the following exclusion criteria:
Subject who is contraindicated to IV t-PA as per local national guidelines
Females who are pregnant or lactating
Rapid neurological improvement prior to randomization suggesting resolution of signs/symptoms of stroke
Known serious sensitivity to radiographic contrast agents
Known sensitivity to Nickel, Titanium metals or their alloys
Current participation in another investigational drug or device study
Known hereditary or acquired hemorrhagic diathesis, coagulation factor deciency
Renal failure as dened by a serum creatinine greater than 2.0 mg/dl (or 176.8 µmol/l) or Glomerular Filtration Rate [GFR] less than 30
Requires hemodialysis or peritoneal dialysis, or who have contraindication to an angiogram
Life expectancy less than 90 days
Clinical presentation suggests a subarachnoid hemorrhage, even if initial CT or MRI scan is normal
Suspicion of aortic dissection
Co-morbid disease or condition that would confound the neurological and functional evaluations or compromise survival or ability to complete follow-up assessments
Currently uses or has a recent history of illicit drug(s) or abuses alcohol
Known history of arterial tortuosity, pre-existing stent, and/or other arterial disease which would prevent the device from reaching the target vessel and/or preclude safe recovery of the device
Additionally, subjects were also considered ineligible for study participation if they met any of the following imaging exclusion criteria:
CT or MRI evidence of hemorrhage on presentation, mass eect or intracranial tumor (except small
meningioma), cerebral vasculitis, basilar artery (BA) occlusion or posterior cerebral artery (PCA) occlusion, carotid dissection, or complete cervical carotid occlusion requiring stenting at the time of the index procedure
CT showing hypodensity or MRI showing hyperintensity involving greater than 1/3 of the middle cerebral
artery (MCA) territory (or in other territories, >100 cc of tissue) on presentation
Baseline non-contrast CT or DWI MRI evidence of a moderate/large core dened as extensive early
ischemic changes of Alberta Stroke Program Early CT score (ASPECTS) less than 6. Patients enrolled under RAPID were excluded based on the following:
a) MRI- or CT-assessed core infarct lesion greater than 50 cc; or b) Severe hypoperfusion lesion (10 sec or more Tmax lesion larger than 100 cc; or c) Ischemic penumbra < 15 cc and mismatch ratio ≤1.8.
Evidence that suggests, in the opinion of the investigator, the subject is not appropriate for mechanical
thrombectomy intervention
5
Study Endpoints
The primary eectiveness endpoint of the study is 90-day global disability assessed via the blinded evaluation of modied Rankin Scale (mRS). Secondary clinical ecacy endpoints of the study are:
Death due to any cause at 90 days.
Functional independence as dened by mRS score ≤ 2 at 90 days.
Change in NIHSS score at 27 ± 6 hours post randomization. The secondary technical ecacy endpoints of the study are:
Volume of cerebral infarction as measured by a CT or MRI scan at 27 ± 6 hours post randomization.
Reperfusion measured by reperfusion ratio on CT or MRI scan 27 ± 6 hours post randomization.
Arterial revascularization measured by TICI 2b or 3 following device use.
Correlation of RAPID-assessed core infarct volume with 27 ± 6 hours post randomization stroke infarction in subjects who achieved TICI 2b-3 reperfusion without intracranial hemorrhage.
Results
A total of 196 subjects were randomized into the SWIFT PRIME Study (98 in each group). The SWIFT PRIME study allowed IV t-PA use beyond 3 hours, although IV t-PA is not approved in the United States beyond 3 hours. Patients treated with IV-tPA beyond 3 hours did not factor strongly in the evaluation of the Solitaire™ 2 revascularization device and have been excluded from the analyses. The resulting Analysis Cohort consists of 161 subjects (84 in the IV t-PA plus Solitaire™ group and 77 with IV t-PA only). Additionally, 17 subjects in the IV t-PA plus Solitaire™ group were excluded from the primary and secondary ecacy endpoint analyses. These 17 subjects either received carotid stenting and/or angioplasty or were treated in a manner inconsistent with the Solitaire™ Instructions for Use. Therefore, the primary and secondary ecacy endpoint analyses cohort consists of 144 subjects.
The proportion of patients functionally independent (mRS 0-2) at the 90-day visit was higher in the IV t-PA plus Solitaire™ device group. A summary of the subject disposition is presented below.
Table 2. Summary of Subject Disposition (Analysis Cohort)
Subject Disposition IV t-PA Only IV t-PA + Solitaire™ Overall
Attended 90 day visit 63/77 (81.8%) 60/67 (89.6%) 123/144 (85.4%)
Died prior to 90 day visit 10/77 (13.0%) 6/67 (9.0%) 16/144(11.1%)
Early Terminated
Subject Lost to Follow-Up 3/77 (3.9%) 0/67 (0.0%) 3/144 (2.1%)
Subject Withdrew Consent 1/77 (1.3%) 1/67 (1.5%) 2/144 (1.4%)
Investigator Withdrew the Subject 0/77 (0.0%) 0/67 (0.0%) 0/144 (0.0%)
Total 77 (100.0%) 67 (100.0%) 144 (100.0%)
Standard summary statistics were calculated for all study variables and subject data were analyzed according to the group to which they were randomized. For continuous variables, statistics included means, standard deviations, medians and ranges. Categorical variables were summarized in frequency distributions.
For the primary ecacy endpoint, statistical signicance was declared using bounds predened in the group sequential analysis plan, which accounts for multiplicity due to interim analyses. Elsewhere, one-sided statistical tests having p-values less than 0.025 were deemed signicant while two-sided tests having p-values less than 0.05 were deemed signicant.
For adverse event reporting, the primary analysis is based on subject counts, not event counts. Both subject counts and event counts are presented in tabular summaries of results.
NOTE: The SFR3-4-20-05, SFR3-4-20-10, SFR3-4-40-10, SFR3-6-20-10, SFR3-6-24-06 and SFR3-6-40-10 models were not evaluated as part of the SWIFT PRIME study.
Table 3. Primary Effectiveness Endpoint (Analysis Cohort)
mRS Score at 90 days IV t-PA Only* IV t-PA + Solitaire™ p-value
0.0007
0 7.9% (6/76) 16.4% (11/67)
1 11.8% (9/76) 31.3% (21/67)
2 17.1% (13/76) 14.9% (10/67)
3 18.4% (14/76) 11.9% (8/67)
4 18.4% (14/76) 13.4% (9/67)
5/6 26.3% (20/76) 11.9% (8/67)
Missing data at 90-day was imputed using LOCF (except baseline data not used)
*One subject from the IV t-PA only group withdrew consent 27 hours post randomization. No 90 day mRS Score was available for this subject.
0 1 2 3 4 5 & 6
Solitaire™ + IV tPA (N =
67)
IV tPA (N=76*)
81612
*
One subject from the IV t-PA only group withdrew consent 27 hours
post randomizaon. No 90 day mRS Score was available for this subject.
31
17
Paents (%)
15
1218132612
18
Figure 5: Modified Rankin Shift at 90 days (Analysis Cohort)
Table 4. Secondary Clinical Efficacy Endpoints (Analysis Cohort)
Secondary Clinical Efficacy
Endpoint
IV t-PA Only*
IV t-PA +
Solitaire™
Mortality at 90 days 10/76 (13.2%) 6/67 (9.0%) 0.65 (0.22-1.89) 0.596
Functional Independence (mRS
0-2) at 90 days
28/76 (36.8%) 42/67 (62.7%) 2.88 (1.46-5.68) 0.003**
Odds Ratio/
Difference
[95% CI]
p-value
Change in NIHSS at 27 hours Post randomization
N 76 66 <0.001**
Mean ± SD -4.3 ± 6.4 -9.9 ± 7.2 -5.6 (-7.9, -3.3)
Median -3.0 -9.5
(min, max) (-24.0, 9.0) (-27.0,10.0)
*One subject from the IV t-PA only group withdrew consent 27 hours post randomization. No secondary clinical ecacy endpoints were available for this subject.
**These p-values are for informative purposes only. Based on the pre-dened hierarchical testing of the secondary endpoints, these particular endpoints were not achieved.
Table 5. Secondary Technical Efficacy Endpoints (Analysis Cohort)
Secondary Technical Efficacy Endpoint
IV t-PA Only IV t-PA + Solitaire™
Infarct Volume at 27 hours Post-Randomization (cc)**
N 77 66
Mean ± SD 68.9 ± 75.5 51.1 ± 86.1 N/A*
Median 46.5 19.2
(min, max) (0.0, 406.6) (0.0,530.5)
Reperfusion Ratio at 27 hours Post-Randomization**
N 44 45
Mean ± SD 61.6 ± 56.1 87.9 ± 37.2 N/A*
Median 84.0 100.0
(min, max) (-200.0, 100.0) (-94.0,100.0)
TICI 2b-3 Following Device Use**
% (n/N) N/A 90.2% (55/61) N/A
*Wilcoxon rank-sum test due to non-normality of data.
** Per Core Laboratory assessed data
Difference
[95% CI]
Table 6. Primary Safety Endpoints (Analysis Cohort)
Safety Endpoint IV t-PA Only IV t-PA + Solitaire™ Odds Ratio
Primary Safety Variables
Any serious adverse event* 33.8% (26/77) 31.0% (26/84) 0.88 (0.45-1.70)
Symptomatic ICH at 27 hours**
NA denotes not applicable
*Per Clinical Events Committee adjudication
**Per Core Laboratory assessed data
3.9% (3/77) 0.0% (0/84) NA
6
Loading...
+ 14 hidden pages