Medtronic SC 37602, PC 37601 User Manual

Activa® SC
37602
Multi-program neurostimulator
Implant manual
  Rx only
Explanation of symbols on product or package labeling
Refer to the appropriate product for symbols that apply.
Open here
Do not reuse
STERILIZE
2
Do not resterilize
STERILE
EO
Sterilized using ethylene oxide
Consult instructions for use
Date of manufacture
Manufacturer
Use by
XXX °F
XX °C
-XX °F
-XX °C
Temperature limitation
Serial number
Conformité Européenne (European Conformity). This symbol means that the device fully complies with AIMD Directive 90/385/EEC (NB 0123).
EC
REP
Authorized representative in the European Community
For USA audiences only
Do not use if package is damaged
37602 2016-06-01  English 3
PIN No.
Pin number
4 English  37602 2016-06-01
Medtronic®, Activa®, and SoftStart/Stop® are trademarks of Medtronic, Inc., registered in the U.S. and other countries.
37602 2016-06-01  English 5
6 English  37602 2016-06-01
Table of contents
Description 9
Package contents 9
Patient identification card 9
Device specifications 9
Declaration of conformity 12
Instructions for use 13
Verifying neurostimulator operation 13 Connecting the extension to the neurostimulator 13 Implanting the neurostimulator 15 Checking system integrity 16 Completing the implant procedure 16
Refer to the indications sheet for indications and related information.
Refer to the appropriate information for prescribers booklet and any additional associated product information for contraindications, warnings, precautions, component disposal, and other important device therapy information.
Refer to the MRI Guidelines for Medtronic Deep Brain Stimulation Systems instructions for use manual for the MRI conditions and MRI-specific contraindication, warnings, and precautions for conducting an MRI scan.
Refer to the System Eligibility Battery Longevity reference manual for neurostimulator selection and battery longevity calculations.
! USA
Refer to the clinical summary booklet for information on the clinical study results of the neurostimulation system, individualization of treatment, adverse events summary, and use in specific populations.
37602 2016-06-01  English 7
8 English  37602 2016-06-01
Description
The Medtronic Activa SC Model 37602 Neurostimulator is part of a neurostimulation system for deep brain stimulation.
Package contents
Neurostimulator
Torque wrench
Product literature
Warranty card
Registration form
Patient identification card
Patient identification card
A patient identification card is packaged with this device. Advise the patient to carry the identification card at all times.
! USA
The patient identification card packaged with the device is temporary; a permanent
card will be mailed to the patient when Medtronic receives the registration form. The implant registration form registers the device warranties and creates a record of the
device in Medtronic’s implant data system.
Device specifications
The neurostimulator is a multi-program device that delivers stimulation through one lead. The stimulation settings are stored in programs. A program is a specific combination of pulse width, rate, and amplitude settings acting on a specific electrode combination. Up to two programs can be combined into a group. When using more than one program, the pulses are delivered sequentially—first a pulse from one program, then a pulse from the next program.
Pulse width, amplitude, and electrode polarity for each program within the group can have different values. Rate, rate limits, SoftStart/Stop and Cycling for each program within the group have the same values.
37602 2016-06-01  English 9
Table 1. Operating values for the Activa SC Model 37602 neurostimulator
Programmable parameter
Operating range and resolution
a
Number of defined groups 1 to 4 Number of programs per group 1 to 2
Electrode configuration
b
1 to 4 electrodes per lead as anode (+), cathode (–), or
Off; case defined as anode or Off Amplitude (voltage mode) 0 to 10.5 V with 0.05-V or 0.1-V resolution Amplitude (current mode) 0 to 25.5 mA with 0.1-mA resolution Amplitude – upper patient limit Tracking limit (by hemisphere): +0 to +2 (0.2 resolution);
+2 to +4.5 (0.5 resolution) Amplitude – lower patient limit Tracking limit (by hemisphere): –0 to –2 (0.2 resolution);
–2 to –4.0 (0.5 resolution); full range
c
Pulse width 60 to 450 µs (10-µs resolution) Pulse width – upper patient limit Tracking limit: +0 to +100 µs (10-µs resolution) Pulse width – lower patient limit Tracking limit: –0 to –100 µs (10-µs resolution) Rate (voltage mode) 2 to 250 Hz (resolution: 1 Hz from 2 Hz to 10 Hz, 5 Hz
from 10 Hz to 250 Hz)
d
Rate (current mode)
30 to 250 Hz (5-Hz resolution)
d
Rate – upper patient limit Tracking limit: +0 to +50 Hz (10-Hz resolution) Rate – lower patient limit Tracking limit: –0 to –50 Hz (10-Hz resolution) SoftStart/Stop Off, On: 1-, 2-, 4-, or 8-second ramp duration Cycling Off, On: 0.1 s to 24 hr (resolution: 0.1 s from 0.1 s to
1 s, 1 s from 1 s to 59 s, 1 min from 1 min to 59 min,
1 hr from 1 hr to 24 hr)
a
Interlocks will prevent the use of some parameter combinations.
b
In current mode, a maximum of 2 electrodes (including the case) can be configured: one as anode (+) and one as cathode (–).
c
Full range = –10.5 V (voltage mode); –25.5 mA (current mode).
d
Rate limited to 125 Hz when two programs are active on a single lead.
10 English  37602 2016-06-01
Table 2. Physical characteristics of the Activa SC Model 37602 neurostimulator
a
Description Value
Connector type
Quadrapolar, two bore
b
Height 55 mm (2.2 in) Length 60 mm (2.4 in) Thickness 11 mm (0.4 in) Weight 45 g (1.6 oz) Volume
28 cm
3
Power source
4.5 Amp hours, 3.2 V HCSVO
c
cell
Temperature limitation –18 °C to +52 °C (0 °F to +126 °F)
Serial number model designator
d
NLA
Radiopaque identification (ID) code NLA Transmitter
Carrier frequency 175 kHz Output level (at 300 m) –48.7 dBuV/m
a
All measurements are approximate.
b
Compatible with two-pronged extension.
c
Hybrid combined silver vanadium oxide.
d
The serial number is the model designator followed by a number. The clinician programmer displays the entire serial number beginning with the model designator.
37602 2016-06-01  English 11
Table 3. Material of components in the Activa SC Model 37602 package
Components Material Material contacts
human tissue
Neurostimulator
Case Titanium Yes Connector block Polyurethane, silicone rubber,
silicone medical adhesive
Yes
Grommets, seals Silicone rubber Yes Setscrews Titanium alloy Yes Adhesive Silicone medical adhesive Yes
Torque wrench
Handle Polyetherimide Yes Shaft Stainless steel Yes
Declaration of conformity
Medtronic declares that this product is in conformity with the essential requirements of Directive 90/385/EEC on Active Implantable Medical Devices.
For additional information, contact the appropriate Medtronic representative listed on the inside back cover of this manual.
12 English  37602 2016-06-01
Instructions for use
Implanting physicians should be experienced in stereotactic and functional neurosurgery and deep brain stimulation procedures, as well as thoroughly familiar with all product labeling.
Cautions:
When using sharp instruments near the neurostimulator, be extremely careful to avoid nicking or damaging the case or the connector block. Damaging the neurostimulator may require surgical replacement.
Do not use saline or other ionic fluids at connections, which could result in a short circuit.
Verifying neurostimulator operation
Before opening the sterile neurostimulator package, verify that the neurostimulator is operable by using the clinician programmer to interrogate the neurostimulator and read the neurostimulator battery service life. (Refer to the software manual for instructions on how to read the battery service life.)
Caution: Do not implant a neurostimulator if it was dropped onto a hard surface from a height of 30 cm (12 in) or more, because the neurostimulator may be damaged and fail to operate properly.
Note: The neurostimulator pocket may be flushed with an antibiotic solution; do not submerge the neurostimulator in fluid.
Connecting the extension to the neurostimulator
Caution: Before connecting components, wipe off any body fluids and dry all
connections. Fluids in the connection may result in stimulation at the connection site, intermittent stimulation, or loss of stimulation.
1. Wipe the extension connector pins with sterile gauze. If necessary, use sterile (United States Pharmacopeia [USP]) water or a nonionic antibiotic solution.
2. Make sure the connector block receptacles are dry and clean.
3. Insert the appropriate extension connector pins into the appropriate neurostimulator
socket until they are seated fully within the connector block (Figure 1). If resistance is felt while inserting the connector pins, use the torque wrench (packaged with the neurostimulator) to retract the setscrews.
Note: To retract the setscrews, insert the torque wrench into the self-sealing grommet and rotate the setscrews counterclockwise; however, do not remove the setscrews from the connector block.
37602 2016-06-01  English 13
Cautions:
Do not insert the extension connector into the connector block if the setscrews are not sufficiently retracted. If the setscrews are not retracted, the extension connector pins may damage the setscrews and the extension connector pins will not be seated fully into the connector block.
Limit counter-clockwise rotations of the neurostimulator setscrews when retracting them. Too many counter-clockwise rotations may disengage the setscrew from the connector block.
Figure 1. Insert the extension connector pins fully into the neurostimulator.
4. Fully insert the torque wrench into each self-sealing grommet of the connector block and tighten each setscrew (Figure 2).
Note: The torque wrench must be oriented to the same angle as the setscrew (Figure 2).
14 English  37602 2016-06-01
Cautions:
Be sure the torque wrench is fully inserted into the self-sealing grommet. If the torque wrench is not fully inserted, the setscrew may be damaged, resulting in intermittent or loss of stimulation.
Before tightening the setscrews, ensure that the extension connector pins are inserted into the connector block to prevent damaging the connector block.
Verify that each leaf of the self-sealing grommet is closed after the torque wrench is withdrawn. If fluid leaks through a grommet seal that is not fully closed, the patient may experience shocking, burning, or irritation at the neurostimulator implant location, or intermittent stimulation or loss of stimulation may occur.
Figure 2. Tightening the setscrews in the self-sealing grommet.
Implanting the neurostimulator
1. Place the neurostimulator into the subcutaneous pocket with the Medtronic logo
facing outward, away from muscle tissue, and ensure that the extension is not bent sharply.
Cautions:
Ensure that the neurostimulator is placed no deeper than 4 cm (1.5 in) below the skin and is parallel to the skin. If the neurostimulator is too deep or is not parallel to the skin, telemetry may be unsuccessful.
Do not coil excess extension in front of the neurostimulator. Wrap excess extension around the perimeter (Figure 3) of the neurostimulator to minimize subcutaneous pocket depth, help minimize potential damage during neurostimulator replacement surgery, help minimize potential kinking of the extension, and minimize interference with telemetry. Excess extension should not exceed two wraps around the perimeter of the neurostimulator.
37602 2016-06-01  English 15
Extension lengths requiring more than two wraps can interfere with telemetry.
Figure 3. Wrap excess extension around the perimeter of the neurostimulator.
2. Use the suture holes in the connector block to secure the neurostimulator to the muscle fascia with nonabsorbable silk.
Checking system integrity
Caution: To use the nonsterile clinician programmer in a sterile field, place a sterile
barrier between the patient and the programming head to prevent infection. Do not sterilize any part of the clinician programmer. Sterilization may damage the programmer.
Note: The neurostimulator should be in the pocket during system interrogation for integrity to ensure proper readings.
1. To ensure proper connection of the extension to the neurostimulator, use the clinician programmer to program the lead configuration and basic stimulation parameters, check the battery status, and check the electrode impedances to rule out a short or open circuit.
2. If the system integrity test results are not acceptable, refer to "Connecting the extension to the neurostimulator" on page 13.
Completing the implant procedure
1. Close and dress all incisions.
2. Ensure that a patient control device is given to the patient.
3. Complete the device tracking and patient registration paperwork and return the
documents to Medtronic.
16 English  37602 2016-06-01
37602 2016-06-01  English 17
Contacts:
Asia:
Medtronic International Ltd. Tel. 02919-1300 Fax 02891-6830
Medtronic Asia Ltd. Tel. (02)-548-1148
Fax (02)-518-4786
Australia:
Medtronic Australasia Pty. Ltd. 97 Waterloo Road North Ryde, NSW 2113 Australia Tel. +61-2-9857-9000 Fax +61-2-9878-5100 Toll-free 1-800-668-670
Austria:
Medtronic Österreich GmbH Tel. 01-240440 Fax 01-24044-100
Belgium:
Medtronic Belgium S.A. Tel. 02-456-0900 Fax 02-460-2667
Canada:
Medtronic of Canada Ltd. Tel. (1-905)-460-3800 Fax (1905)-826-6620
Czech Republic:
Medtronic Czechia s.r.o. Tel. 2-965-795-80 Fax 2-965-795-89
Denmark:
Medtronic Danmark A/S Tel. 45-32-48-18-00 Fax 45-32-48-18-01
Finland:
Medtronic Finland Oy/LTD Tel. (09)-755-2500 Fax (09)-755-25018
France:
Medtronic France S.A.S. Tel. 01-5538-1700 Fax 01-5538-1800
Germany:
Medtronic GmbH Tel. (02159)-81490 Fax (02159)-8149100
Greece:
Medtronic Hellas S.A. Tel. 210-67-79-099 Fax 210-67-79-399
Hungary:
Medtronic Hungária Kft. Tel. 1-889-06-00 Fax 1-889-06-99
Ireland:
Medtronic Ireland Ltd. Tel. (01)-890-6522 Fax (01)-890-7220
Italy:
Medtronic Italia SpA Tel. 02-241371 Fax 02-241381
Tel. 06-328141 Fax 06-3215812
Japan:
Medtronic Japan Tel. 03-6776-0017 Fax 03-6774-4645
Latin America:
Medtronic, Inc. Tel. (1305)-500-9328 Fax (1786)-709-4244
Norway:
Medtronic Norge AS Tel. 67-10-32-00 Fax 67-10-32-10
Poland:
Medtronic Poland Sp. z.o.o. Tel. (022)-465-69-00 Fax (022)-465-69-17
Portugal:
Medtronic Portugal, Lda. Tel. 21-724-5100 Fax 21-724-5199
Russia:
Medtronic Russia Tel. (8495) 580-7377 Fax (8495) 580-7378
Slovakia:
Medtronic Slovakia, o.z. Tel. 0268 206 911 Fax 0268 206 999
Spain:
Medtronic Ibérica, S.A. Tel. 91-625-0400 Fax 91-650-7410
Sweden:
Medtronic AB Tel. 08-568-585-00 Fax 08-568-585-01
Switzerland:
Medtronic (Schweiz) AG Tel. 031-868-0100 Fax 031-868-0199
The Netherlands:
Medtronic B.V. Tel. (045)-566-8000 Fax (045)-566-8668
Turkey:
Medtronic Turkey Tel. +90 216 636 1000 Fax +90 216 636 1008
U.K.:
Medtronic U.K. Ltd. Tel. 01923-212213 Fax 01923-241004
USA:
Medtronic, Inc. Tel. (1-763)-505-5000 Fax (1-763)-505-1000 Toll-free: (1-800)-328-0810
Manufacturer
Medtronic, Inc. 710 Medtronic Parkway, Minneapolis, MN 55432-5604, USA www.medtronic.com Tel. +1-763-505-5000 Fax +1-763-505-1000
Authorized Representative
EC
REP
in the European Community
Medtronic B.V. Earl Bakkenstraat 10, 6422 PJ Heerlen, The Netherlands Tel. +31-45-566-8000 Fax +31-45-566-8668
Europe/Africa/Middle East Headquarters
Medtronic International Trading Sàrl Route du Molliau 31, Case Postale 84 CH - 1131 Tolochenaz, Switzerland www.medtronic.eu Tel. +41-21-802-7000 Fax +41-21-802-7900
Asia-Pacific
Medtronic International Ltd. Suite 1106-11, 11/F, Tower 1, The Gateway, 25 Canton Road, Tsimshatsui, Kowloon, Hong Kong Tel. +852-2919-1300 Fax +852-2891-6830
Contacts for specific countries are listed inside this cover.
*M939238A038*
© Medtronic, Inc. 2016 All Rights Reserved