Medtronic RC 37612 User Manual

M927170A005_FC.fm UC200xxxxxx EN
4.625 x 6 inches (117 mm x 152 mm)
Activa® RC 37612
Multi-program rechargeable neurostimulator Ładowalny neurostymulator wieloprogramowy Multiprogramovatelný dobíjecí neurostimulátor
Мультипрограммный нейростимулятор с перезаряжаемой батареей Viacprogramový nabíjateľný neurostimulátor
Medtronic Confidential
Implant manual ■ Podręcznik implantacji ■ Implantační příručka Руководство по имплантации
! USA
M927170A005 Rev B
Rx only
Príručka k implantátu
Printing instructions: Refer to the “Implant manual” cate gory Table 1 in doc# A00002 for Neuro Core European Printing Instructions.
M927170A005_FC.fm UC200xxxxxx EN
4.625 x 6 inches (117 mm x 152 mm)
Medtronic Confidential
M927170A005 Rev B
Explanation of symbols on product or package labeling
Refer to the appropriate product for symbols that apply.
Conformité Européenne (European Conformity). This symbol means that the device fully complies with AIMD Directive 90/385/EEC (NB 0123) and R&TTE Directive 1999/5/EC. The use of this device might be subject to individual country licensing regimes in Europe.
Open here
Do not reuse
STERILE
EO
Sterilization: ethylene-oxide gas
Caution: consult accompanying documents
Use by
Manufacturing date
Serial number
+XXX °F
+XX °C
-XX °F
-XX °C
Storage temperature
For USA audiences only
37612 2007-09  English 1
2007-09
Filename Date Time UC200xxxxxx EN
4.625 x 6 inches (117 mm x 152 mm)
Medtronic Confidential
ImplantManual.xsl - IPGTemplate.fm
Version: 06-11-2008
M927170A005
Medtronic® and Activa® are registered trademarks of Medtronic, Inc. SoftStart/Stop™ is a trademark of Medtronic, Inc.
2 English  37612 2007-09
2007-09
Filename Date Time UC200xxxxxx EN
4.625 x 6 inches (117 mm x 152 mm)
Medtronic Confidential
ImplantManual.xsl - IPGTemplate.fm
Version: 06-11-2008
M927170A005
Table of contents
Description 5 Package contents 5 Patient identification card 5 Device specifications 5 Declaration of conformity 8 Instructions for use 9
Charging the neurostimulator battery 9 Verifying neurostimulator operation 9 Connecting the extension to the neurostimulator 10 Implanting the neurostimulator 11 Checking system integrity 12 Completing the implant procedure 13
Refer to the indications sheet for indications and related information.
Refer to the appropriate information for prescribers booklet for contraindications, warnings, precautions, adverse events summary, individualization of treatment, patient selection, use in specific populations, resterilization, and component disposal.
Refer to System Eligibility, Battery Longevity, Specifications reference manual packaged with the software application card for neurostimulator selection, battery longevity calculations and specific neurostimulator specifications.
! USA
Refer to the clinical summary booklet packaged with the neurostimulator for
information on the clinical study results of the neurostimulation system.
37612 2007-09  English 3
2007-09
Filename Date Time UC200xxxxxx EN
4.625 x 6 inches (117 mm x 152 mm)
Medtronic Confidential
ImplantManual.xsl - IPGTemplate.fm
Version: 06-11-2008
M927170A005
4 English  37612 2007-09
2007-09
Filename Date Time UC200xxxxxx EN
4.625 x 6 inches (117 mm x 152 mm)
Medtronic Confidential
ImplantManual.xsl - IPGTemplate.fm
Version: 06-11-2008
M927170A005
Description
The Medtronic Activa RC Model 37612 Neurostimulator is part of a neurostimulation system for deep brain stimulation.
Package contents
Neurostimulator
Torque wrench
Product literature
Warranty card (USA only)
Registration form
Patient identification card
Patient identification card
A patient identification card is packaged with this device. Advise the patient to carry the identification card at all times.
! USA
The patient identification card packaged with the device is temporary; a permanent card
will be mailed to the patient when Medtronic receives the registration form. The implant registration form registers the device warranties and creates a record of the device
in Medtronic’s implant data system.
Device specifications
The neurostimulator is a multi-programmable, rechargeable device that delivers stimulation through 1 or 2 leads. The stimulation settings are stored in programs. A program is a specific combination of pulse width, rate, and amplitude settings acting on a specific electrode combination. Up to four programs can be combined into a group, with a maximum of 2 programs per lead. When using more than one program, the pulses are delivered sequentially —first a pulse from one program, then a pulse from the next program.
Pulse width, amplitude, and electrode polarity for each program within the group can have different values. Rate, rate limits, ramping, and cycling for each program within the group have the same values.
37612 2007-09  English 5
2007-09
Filename Date Time UC200xxxxxx EN
4.625 x 6 inches (117 mm x 152 mm)
Medtronic Confidential
ImplantManual.xsl - IPGTemplate.fm
Version: 06-11-2008
M927170A005
Table 1. Operating values for the Activa RC Model 37612 neurostimulator
Programmable parameter
Operating range and resolution
a
Number of defined groups
1 to 4
b
Number of programs per group
1 to 4
b
Electrode configuration Up to 4 electrodes as anode, cathode, or OFF
Case defined as anode or OFF Amplitude (voltage mode) 0 to 10.5 V with 0.05 V or 0.1-V resolution Amplitude (current mode) 0 to 25.5 mA with 0.1-mA resolution Amplitude – upper patient limit By hemisphere:
Tracking limit: +0 to +2 (0.2 resolution); +2 to +4.5 (0.5
resolution) Amplitude – lower patient limit By hemisphere:
Tracking limit: -0 to -2 (0.2 resolution); -2 to -4.0 (0.5
resolution); full range
c
Pulse width 60 to 450 µs (10-µs resolution) Pulse width – upper patient limit Tracking limit: +0 to +100 µs (10-µs resolution) Pulse width – lower patient limit Tracking limit: -0 to -100 µs (10-µs resolution) Rate (voltage mode) 2 to 250 Hz (resolution: 1 Hz from 2 Hz to 10 Hz, 5 Hz from
10 Hz to 250 Hz)
d
Rate (current mode)
30 to 250 Hz (5-Hz resolution)
d
Rate – upper patient limit Tracking limit: +0 to +50 Hz (10-Hz resolution) Rate – lower patient limit Tracking limit: -0 to -50 Hz (10-Hz resolution) SoftStart/Stop OFF, ON: 1-, 2-, 4-, or 8-second ramp duration Cycling OFF, ON: 0.1 s to 24 hr (resolution: 0.1 s from 0.1 s to 1 s,
1 s from 1 s to 59 s, 1 min from 1 min to 59 min, 1 hr from
1 hr to 24 hr)
a
Interlocks will prevent the use of some parameter combinations.
b
No more than 16 programs may be defined within the 4 groups.
c
Full range = -10.5 V (voltage mode); -25.5 mA (current mode)
d
Rate limited to 125 Hz when two programs are active on a single lead.
6 English  37612 2007-09
2007-09
Filename Date Time UC200xxxxxx EN
4.625 x 6 inches (117 mm x 152 mm)
Medtronic Confidential
ImplantManual.xsl - IPGTemplate.fm
Version: 06-11-2008
M927170A005
Table 2. Physical characteristics of the Activa RC Model 37612 neurostimulator
a
Description Value
Connector type Octapolar, in-line 2.8-mm (0.110-in) spacing Height 54 mm (2.1 in) Length 54 mm (2.1 in) Thickness
case 9 mm (0.4 in)
connector 11 mm (0.4 in) Weight 40 g (1.6 oz) Volume
22 cm
3
Battery life 9 years
Power source
b
Lithium ion rechargeable battery
Storage temperature –18° to +52°C (0° to +126°F)
Serial number model designator
c
NKG
Radiopaque Identification (ID) code NKG
a
All measurements are approximate.
b
The neurostimulator is not shipped with a full battery charge. Refer to the charging system user manual for neurostimulator charging instructions.
c
The serial number is the model designator followed by a number. The clinician programmer displays the entire serial number beginning with the model designator.
37612 2007-09  English 7
2007-09
Filename Date Time UC200xxxxxx EN
4.625 x 6 inches (117 mm x 152 mm)
Medtronic Confidential
ImplantManual.xsl - IPGTemplate.fm
Version: 06-11-2008
M927170A005
Table 3. Material of components in the Activa RC Model 37612 package
Components Material Material contacts
human tissue
Neurostimulator
Case Titanium, parylene Yes Connector block Polysulfone, silicone rubber,
silicone medical adhesive
Yes
Grommets, seals Silicone rubber Yes Setscrews Titanium alloy Yes Adhesive Silicone medical adhesive Yes
Torque wrench
Handle Polyetherimide Yes Shaft Stainless steel Yes
Declaration of conformity
Medtronic declares that this product is in conformity with the essential requirements of Directive 1999/5/EC on Radio and Telecommunications Terminal Equipment and Directive 90/385/EEC on Active Implantable Medical Devices.
For additional information, contact the appropriate Medtronic representative listed on the inside back cover of this manual.
8 English  37612 2007-09
2007-09
Filename Date Time UC200xxxxxx EN
4.625 x 6 inches (117 mm x 152 mm)
Medtronic Confidential
ImplantManual.xsl - IPGTemplate.fm
Version: 06-11-2008
M927170A005
Instructions for use
Implanting physicians should be experienced in stereotactic and functional neurosurgery and have expertise with functional stereotactic neurosurgical treatment of movement disorders and should be thoroughly familiar with all product labeling.
w Warning: DO NOT use the recharger on an unhealed wound. The recharger, antenna,
and holster are not sterile, and contact with the wound can cause an infection.
# Caution: Advise patients to charge the neurostimulator on a regular basis, taking into
consideration the therapeutic parameters and battery usage of the patient, to prevent the battery from overdischarging. If the neurostimulator battery is allowed to overdischarge, the patient cannot charge the neurostimulator; however, the clinician may be able to restore the battery function using the Physician Recharge Mode on the recharger (refer to the troubleshooting section of the software manual).
Allowing the neurostimulator battery to overdischarge will permanently affect the neurostimulator in one of the following ways:
Battery function is restored; however, charging sessions may be more frequent
because battery capacity has been reduced.
Battery function is not restored and the neurostimulator must be surgically replaced.
Battery function is not restored when:
the neurostimulator battery is permanently damaged.the neurostimulator battery has been overdischarged and restored twice
before. The third time the battery is overdischarged, the neurostimulator will reach end of service. Surgery is required to replace the neurostimulator.
# Cautions:
When using sharp instruments near the neurostimulator, be extremely careful to
avoid nicking or damaging the case, the insulation, or the connector block. Damaging the neurostimulator may require surgical replacement.
Do not use saline or other ionic fluids at connections, which could result in a short
circuit.
Charging the neurostimulator battery
Charge the neurostimulator battery before opening the package. For charging instructions, refer to the charging system user manual.
Verifying neurostimulator operation
Before opening the sterile neurostimulator package, verify that the neurostimulator is operable by using the clinician programmer to interrogate the neurostimulator and read the neurostimulator battery charge level. (Refer to the software manual for instructions on how to read the battery charge level.)
37612 2007-09  English 9
2007-09
Filename Date Time UC200xxxxxx EN
4.625 x 6 inches (117 mm x 152 mm)
Medtronic Confidential
ImplantManual.xsl - IPGTemplate.fm
Version: 06-11-2008
M927170A005
# Caution: Do not implant a neurostimulator if it was dropped onto a hard surface from a
height of 30 cm (12 in) or more, because the neurostimulator may be damaged and fail
to operate properly.
Note: The neurostimulator pocket may be flushed with an antibiotic solution; do not submerge the neurostimulator in fluid.
Connecting the extension to the neurostimulator
# Caution: Before connecting components, wipe off any body fluids and dry all
connections. Fluids in the connections may result in stimulation at the connection site,
intermittent stimulation, or loss of stimulation.
1. Wipe the extension connectors with sterile gauze. If necessary, use sterile (United States
Pharmacopeia [USP]) water or a nonionic antibiotic solution.
2. Make sure the connector block receptacles are dry and clean.
3. Insert each extension connector into the appropriate neurostimulator socket until it is
seated fully within the connector block (Figure 1).
Notes:
During insertion, some resistance is typical.
To retract the setscrews, insert the torque wrench into the self-sealing grommet and rotate the setscrews counterclockwise; however, do not remove the setscrews from the connector block.
# Caution: Do not insert the extension connector into the connector block if the
setscrews are not sufficiently retracted. If the setscrews are not retracted, the setscrews may damage the extension and the extension will not be seated fully into the connector block.
Socket II (Electrodes 8-15)
Socket I (Electrodes 0-7)
Extension 2
Extension 1
Figure 1. Insert the extension connectors into the neurostimulator.
10 English  37612 2007-09
2007-09
Filename Date Time UC200xxxxxx EN
4.625 x 6 inches (117 mm x 152 mm)
Medtronic Confidential
ImplantManual.xsl - IPGTemplate.fm
Version: 06-11-2008
M927170A005
Note: Insert a connector plug (from an accessory kit) into unused neurostimulator socket.
4. For each extension or plug, fully insert the torque wrench (packaged with the
neurostimulator) into each self-sealing grommet of the connector block and tighten each setscrew (Figure 2).
# Cautions:
To prevent undertightening the neurostimulator setscrews, do not use the
torque wrench from the extension kit. Undertightening may result in insufficient electrical contact within the connector block, which may cause intermittent stimulation.
Be sure the torque wrench is fully inserted into the self-sealing grommet. If the
torque wrench is not fully inserted, the setscrew may be damaged, resulting in intermittent or loss of stimulation.
Before tightening setscrews, ensure that the extension connector is inserted
into the connector block to prevent damaging the extension.
Verify that each leaf of the self-sealing grommet is closed after the torque
wrench is withdrawn. If fluid leaks through a grommet seal that is not fully closed, the patient may experience shocking, burning, or irritation at the neurostimulator implant location, or intermittent stimulation, or loss of stimulation.
Figure 2. Tightening the setscrews in the self-sealing grommet.
Implanting the neurostimulator
1. Place the neurostimulator into the subcutaneous pocket with the Medtronic logo facing
outward, away from muscle tissue, and ensure that the extension is not bent sharply.
37612 2007-09  English 11
2007-09
Filename Date Time UC200xxxxxx EN
4.625 x 6 inches (117 mm x 152 mm)
Medtronic Confidential
ImplantManual.xsl - IPGTemplate.fm
Version: 06-11-2008
M927170A005
# Cautions:
Ensure that the neurostimulator is placed no deeper than 1 cm (0.4 in) below
the skin and is parallel to the skin. If the neurostimulator is too deep or is not parallel to the skin, recharge may be inefficient or unsuccessful.
Position the neurostimulator with the Medtronic logo facing outward. If
implanted with the Medtronic logo facing inward, the neurostimulator will be difficult to charge.
Do not coil excess extension in front of the neurostimulator. Wrap excess
extension around the perimeter (Figure 3) of the neurostimulator to minimize subcutaneous pocket depth, help minimize potential damage during neurostimulator replacement surgery, help minimize potential kinking of the extension, and minimize interference with telemetry and recharge operation. Excess extension should not exceed two wraps around the perimeter of the neurostimulator. Extension lengths requiring more than two wraps can interfere with telemetry.
Figure 3. Wrap excess extensions around the perimeter of the neurostimulator.
2. Use the suture holes in the connector block to secure the neurostimulator to the muscle
fascia with nonabsorbable silk.
Checking system integrity
# Caution: To use the nonsterile clinician programmer in a sterile field, place a sterile
barrier between the patient and the programming head to prevent infection. Do not
sterilize any part of the clinician programmer. Sterilization may damage the programmer.
Note: The neurostimulator should be in the pocket during system interrogation for integrity to ensure proper readings.
1. To ensure proper connection of each extension to the neurostimulator, use the clinician
programmer to program the basic stimulation parameters, check the battery status, and
check the electrode impedances to rule out a short or open circuit.
2. If the system integrity test results are not acceptable, refer to "Connecting the extension
to the neurostimulator" on page 10.
12 English  37612 2007-09
2007-09
Filename Date Time UC200xxxxxx EN
4.625 x 6 inches (117 mm x 152 mm)
Medtronic Confidential
ImplantManual.xsl - IPGTemplate.fm
Version: 06-11-2008
M927170A005
Completing the implant procedure
1. Close and dress all incisions.
2. Ensure that a patient control device is given to the patient.
3. Complete the device tracking and patient registration paperwork and return the
documents to Medtronic.
37612 2007-09  English 13
2007-09
Filename Date Time UC200xxxxxx EN
4.625 x 6 inches (117 mm x 152 mm)
Medtronic Confidential
ImplantManual.xsl - IPGTemplate.fm
Version: 06-11-2008
M927170A005
14 English  37612 2007-09
2007-09
Filename Date Time UC200xxxxxx EN
4.625 x 6 inches (117 mm x 152 mm)
Medtronic Confidential
ImplantManual.xsl - IPGTemplate.fm
Version: 06-11-2008
M927170A005
Objaśnienia symboli znajdujących się na etykietach produktu lub opakowania
Aby określić, które symbole mają zastosowanie w danym przypadku, należy zapoznać się z oznaczeniami odpowiedniego produktu.
Conformité Européenne (zgodność z normami Unii Europejskiej). Ten symbol oznacza, że urządzenie spełnia wszystkie wymagania Dyrektywy AIMD 90/385/EWG (NB 0123) oraz Dyrektywy R&TTE 1999/5/WE. Stosowanie niniejszego urządzenia może podlegać ograniczeniom licencyjnym, obowiązującym w poszczególnych krajach europejskich.
Otwierać tutaj
Produkt do jednorazowego zastosowania
STERILE
EO
Sterylizacja: gazowa — tlenkiem etylenu
Przestroga: należy zapoznać się z dołączoną dokumentacją
Termin ważności
Data produkcji
Numer seryjny
+XXX °F
+XX °C
-XX °F
-XX °C
Temperatura przechowywania
Dotyczy tylko odbiorców w USA
37612  Polski 1
2007-09
Filename Date Time UC200xxxxxx EN
4.625 x 6 inches (117 mm x 152 mm)
Medtronic Confidential
ImplantManual.xsl - IPGTemplate.fm
Version: 08-14-2008
M927170A005
Medtronic® i Activa® są zastrzeżonymi znakami towarowymi firmy Medtronic, Inc. SoftStart/Stop™ jest znakiem towarowym firmy Medtronic, Inc.
2 Polski  37612
2007-09
Filename Date Time UC200xxxxxx EN
4.625 x 6 inches (117 mm x 152 mm)
Medtronic Confidential
ImplantManual.xsl - IPGTemplate.fm
Version: 08-14-2008
M927170A005
Spis treści
Opis urządzenia 5 Zawartość opakowania 5 Karta identyfikacji pacjenta 5 Dane techniczne urządzenia 5 Deklaracja zgodności 8 Instrukcja użytkowania 9
Ładowanie baterii neurostymulatora 9 Kontrola działania neurostymulatora 10 Podłączanie przedłużacza do neurostymulatora 10 Implantacja neurostymulatora 12 Sprawdzenie integralności układu 13 Zakończenie procedury implantacji 13
Wskazania i związane z nimi informacje znajdują się na karcie wskazań.
Informacje dotyczące przeciwwskazań, ostrzeżeń, środków ostrożności, zdarzeń niepożądanych, indywidualizacji leczenia, doboru pacjentów, zastosowania w odniesieniu do poszczególnych grup pacjentów, powtórnej sterylizacji oraz utylizacji elementów znajdują się w broszurze "Informacje dla lekarza zlecającego".
Informacje na temat doboru neurostymulatora, obliczania czasu funkcjonowania baterii oraz danych technicznych konkretnego modelu neurostymulatora znajdują się w podręczniku "Dobór systemu, czas funkcjonowania baterii i dane techniczne" dołączonym do karty oprogramowania programatora.
! USA
Informacje na temat wyników badań klinicznych dotyczących systemu neurostymulacji znajdują się w broszurze "Podsumowanie wyników klinicznych" dołączonej do neurostymulatora.
37612  Polski 3
2007-09
Filename Date Time UC200xxxxxx EN
4.625 x 6 inches (117 mm x 152 mm)
Medtronic Confidential
ImplantManual.xsl - IPGTemplate.fm
Version: 08-14-2008
M927170A005
4 Polski  37612
2007-09
Filename Date Time UC200xxxxxx EN
4.625 x 6 inches (117 mm x 152 mm)
Medtronic Confidential
ImplantManual.xsl - IPGTemplate.fm
Version: 08-14-2008
M927170A005
Opis urządzenia
Neurostymulator Medtronic Activa RC model 37612 jest częścią systemu neurostymulacji służącego do głębokiej stymulacji mózgu.
Zawartość opakowania
Neurostymulator
Klucz dynamometryczny
Dokumentacja produktu
Karta gwarancyjna (tylko w USA)
Formularz rejestracyjny
Karta identyfikacji pacjenta
Karta identyfikacji pacjenta
Karta identyfikacji pacjenta znajduje się w opakowaniu urządzenia. Pacjentowi należy doradzić, aby zawsze nosił przy sobie kartę identyfikacji.
! USA
Dołączona do urządzenia karta identyfikacji pacjenta jest tymczasowa; karta stała zostanie przesłana do pacjenta drogą pocztową po otrzymaniu przez firmę Medtronic formularza rejestracyjnego.
Formularz rejestracyjny implantu służy do zarejestrowania gwarancji oraz utworzenia wpisu dotyczącego wszczepionego urządzenia w bazie danych implantów firmy Medtronic.
Dane techniczne urządzenia
Neurostymulator jest wieloprogramowym ładowalnym urządzeniem do stymulacji za pomocą jednego lub dwóch odprowadzeń. Ustawienia stymulacji są zapisywane jako programy. Program to konkretny zbiór ustawień szerokości impulsu, częstotliwości oraz amplitudy związany z oddziaływaniem na daną kombinację elektrod. Maksymalnie cztery programy mogą zostać połączone w grupę i maksymalnie 2 programy mogą zostać zastosowane na każde odprowadzenie. W przypadku stosowania więcej niż jednego programu impulsy są dostarczane sekwencyjnie — najpierw impuls pochodzący z jednego programu, a następnie impuls z następnego programu.
Szerokość impulsu, amplituda oraz biegunowość elektrod dla każdego programu w ramach grupy mogą mieć inne wartości. Częstotliwość, limity częstotliwości, dostrajanie i tryb pracy cyklicznej dla każdego programu wewnątrz grupy mają takie same wartości.
37612  Polski 5
2007-09
Filename Date Time UC200xxxxxx EN
4.625 x 6 inches (117 mm x 152 mm)
Medtronic Confidential
ImplantManual.xsl - IPGTemplate.fm
Version: 08-14-2008
M927170A005
Tabela 1. Wartości robocze parametrów dla neurostymulatora Activa RC model 37612
Parametry programowalne
Zakres roboczy i przyrost
a
Liczba zdefiniowanych grup
1 do 4
b
Liczba programów na grupę
1 do 4
b
Konfiguracja elektrod Do 4 elektrod jako anoda, katoda lub wyłączona (OFF)
Obudowa zdefiniowana jako anoda lub wyłączona (OFF) Amplituda (tryb napięcia) 0 do 10,5 V z przyrostem 0,05 V lub 0,1 V Amplituda (tryb prądu) 0 do 25,5 mA z przyrostem 0,1 mA Amplituda — górna granica
regulacji przez pacjenta
Wg półkuli:
Granica śledzenia: +0 do +2 (z przyrostem 0,2); +2 do +4,5
(z przyrostem 0,5) Amplituda — dolna granica
regulacji przez pacjenta
Wg półkuli:
Granica śledzenia: –0 do –2 (z przyrostem 0,2); –2 do –4,0
(z przyrostem 0,5); pełen zakres
c
Szerokość impulsu 60 do 450 µs (z przyrostem 10 µs) Szerokość impulsu — górna
granica regulacji przez pacjenta
Granica śledzenia: +0 do +100 µs (z przyrostem 10 µs)
Szerokość impulsu — dolna granica regulacji przez pacjenta
Granica śledzenia: –0 do –100 µs (z przyrostem 10 µs)
Częstotliwość (tryb napięcia) 2 do 250 Hz (z przyrostem: 1 Hz od 2 Hz do 10 Hz, 5 Hz od
10 Hz do 250 Hz)
d
Częstotliwość (tryb prądu)
30 do 250 Hz (z przyrostem 5 Hz)
d
Częstotliwość — górna granica regulacji przez pacjenta
Granica śledzenia: +0 do +50 Hz (z przyrostem 10 Hz)
Częstotliwość — dolna granica regulacji przez pacjenta
Granica śledzenia: –0 do –50 Hz (z przyrostem 10 Hz)
Płynne uruchamianie i zatrzymywanie
OFF (WYŁ.), ON (WŁ.): czas dostrajania 1, 2, 4 lub
8 sekund Tryb cykliczny OFF (WYŁ.), ON (WŁ.): 0,1 s do 24 godz.(z przyrostem:
0,1 s od 0,1 s do 1 s, 1 s od 1 s do 59 s, 1 min od 1 min do
59 min, 1 godz. od 1 godz. to 24 godz.)
a
Blokady uniemożliwiają użycie niektórych zestawów parametrów.
b
W ramach 4 grup można zdefiniować nie więcej niż 16 programów.
c
Pełen zakres = –10,5 V (tryb napięcia), –25,5 mA (tryb prądu)
d
Częstotliwość ograniczona do 125 Hz, gdy na jednym odprowadzeniu aktywne są dwa programy.
6 Polski  37612
2007-09
Filename Date Time UC200xxxxxx EN
4.625 x 6 inches (117 mm x 152 mm)
Medtronic Confidential
ImplantManual.xsl - IPGTemplate.fm
Version: 08-14-2008
M927170A005
Tabela 2. Charakterystyka fizyczna neurostymulatora Activa RC model 37612
a
Opis Wartość
Typ złącza Ośmiobiegunowe, wbudowane z odstępem 2,8 mm Wysokość 54 mm Długość 54 mm Grubość
obudowa 9 mm
złącze 11 mm Masa 40 g Objętość
22 cm
3
Żywotność baterii 9 lat
Źródło zasilania
b
Ładowalna bateria litowo-jonowa
Temperatura przechowywania od -18°C do +52°C
Oznaczenie numeru seryjnego modelu
c
NKG
Kod identyfikatora radiocieniującego NKG
a
Wszystkie wymiary podano w przybliżeniu.
b
Baterie fabrycznie nowego neurostymulatora nie są w pełni naładowane. Aby zapoznać się z instrukcją ładowania neurostymulatora, należy skorzystać z podręcznika użytkownika systemu ładowania.
c
Numerem seryjnym jest liczba poprzedzona oznaczeniem modelu. Na ekranie programatora klinicznego wyświetlany jest cały numer seryjny rozpoczynający się od oznaczenia modelu.
37612  Polski 7
2007-09
Filename Date Time UC200xxxxxx EN
4.625 x 6 inches (117 mm x 152 mm)
Medtronic Confidential
ImplantManual.xsl - IPGTemplate.fm
Version: 08-14-2008
M927170A005