Conformité Européenne (European Conformity). This symbol means that
the device fully complies with the European Council Directive 98/79/EC
(27 October 1998, on in vitro diagnostic medical devices).
Attention, See Instructions for Use
Consult Instructions for Use
Fuse
On (power)
Off (power)
Alternating Current
Medical equipment with respect to electric shock,
hazards only in accordance with IEC 61010-1:2001 and CSA C22.2
No. 61010-1.
In Vitro Diagnostic Device
For US Audiences Only
Quantity
Serial Number
e and mechanical
fir
Catalog Number
Authorized Representative in the European Community
Fragile, Handle with Care
This Way Up
Main Menu
Quality Control Menu
Stop
Date of Manufacture
1
Biological Risks
Nonionizing Electromagnetic Radiation
Manufactured I
Cancel
Enter
Clear
Temperature Limitation
Humidity Limitation
Manufacturer
Do not dispose of this product in the unso
Dispose of this product according to local regulations. See
http://recycling.medtronic.com for instructions on proper disposal of this
product.
China RoHS Standard (SJ/T11364-2006) Electronic Information Products
Pollution
be used before it must be recycled (environmental protection use period).
n
rted municipal waste stream.
C
ontrol Symbol. The number represents the years the device can
View Test Results for the Last 20 Tests Performed 24
Transmit Patient and Quality Control (QC) Test Results 25
Transmission Using Floppy Drive 25
Transmission Using USB Port 25
Set Instrument Parameters 26
Quality Control (QC) Menu 27
General Quality Control (QC) Functions 27
Select the Control Type 28
Enter Control Lot Number, Expiration Date, and Range 28
Manual Entry 28
Bar Code Entry 29
To verify/record the ACT Plus
®
actuator heat block temperature 30
To view the due status for quality control testing 30
QC Manager Functions 30
Select the location for the ACT Plus
®
instrument: 32
Add/delete User IDs: 32
Add a User ID: 32
Delete a User ID: 32
Enable User Lockout: 32
4 English ACT Plus® Operator’s Manual
Change the QC Manager Password: 33
Download User IDs: 33
Transfer User IDs from ACT Plus
Instruments: 33
Select the Time Interval for Clearing Patient and User IDs: 33
Select the QC Lockout Function: 33
Set the QC Interval for Liquid and Electronic Controls: 34
To set the liquid QC interval: 34
To set the ACTtrac® e
lectronic QC interval: 34
Enable or Disable Test Cartridge Types: 34
Download All of the Instrument Settin
Device: 34
Set the Permanent Record or the Mode by Which the ACT Plus®
Instrument Maintains a Record of Downloaded Data: 34
Upload Software Upgrades from a Floppy Disk or USB Storage
Device: 35
Section 7: Test Methods 36
General 36
Activated Clotting Times: High Range
and Recalcified (RACT) Cartridges 37
Results 37
High Range Heparina
se (HR-HTC) 38
®
Instrument to Multiple
Floppy Disk or USB Storage
gs to a
(HR
-ACT), Low Range (LR-ACT),
Section
8: Quality Assurance 39
Instrument Self-Test 39
Liquid QC 39
Electronic QC (ACTtrac®) 39
Section 9: Maintenance 41
Instrument Case Cleaning 41
Actuator Assembly Cleaning 41
Actuator Heat Block Temperature 42
Temperature Verification Cartridge 42
Verification with Thermometer 43
Electromagnetic Emissions and Immunity Declarations 53
Appendix A: Optional Bar Code Scanner LS1900 57
Introduction 57
Parts of the LS1900 Series Scanner 57
Scanning in Hand-Held Mode 58
Setting Bar Code Format Options 58
Code 39 58
Interleave 2 of 5 59
Codabar 60
Troubleshooting 60
Defaults 60
Procedure for setting the defaults: 61
Appendix B: Optional Bar Code Scanner LS2208 65
Introduction 65
Parts of the LS2208 Scanner 65
Scanning in Hand-Held Mode 66
Setting Bar Code Format Options 66
6 English ACT Plus® Operator’s Manual
Bar Code Scanner Installation Instructions 70
Appendix C: Warranties 71
EQUIPMENT LIMITED WARRANTY 71
EQUIPMENT LIMITED WARRANTY 72
ACT Plus® Operator’s Manual English 7
8 English ACT Plus® Operator’s Manual
Section 1: System Description
The Medtronic ACT Plus® Automated Coagulation Timer is a
microprocessor-controlled electromechanical coagulation instrument intended for
in vitro determination of coagulation endpoints in fresh and citrated whole blood
samples. The ACT Plus
Medtronic disposable cartridges: High Range ACT, Low Range ACT, Recalcified
ACT, and High Range Heparinase.
Intended Use
The ACT Plus® instrument is intended for in vitro diagnostic testing in either a
hospital laboratory setting or a point of care (decentralized) setting (eg, in the
operating room, cardiac catheterization lab, intensive care unit, or clinic, etc.).
Principle of Clot Detection
Test reactions occur at 37°C ±0.5°C in single use cartridges placed in the Actuator
Heat Block Assembly. Fibrin formation is the endpoint of tests performed on the
ACT Plus
the plunger-flag mechanism in each cartridge channel. The plunger assembly falls
rapidly at programmed timed intervals through an unclotted sample. The fibrin web
formed during clotting impedes the fall rate of the plunger and is detected by a
photo-optical system located in the ACT Plus® instrument’s actuator assembly.
Tests are performed in duplicate, and the clotting time results are displayed on the
red Light Emitting Diode (LED) display and on the Liquid Crystal Display (LCD)
screen of the ACT Plus
®
instrument. Fibrin formation is detected by measuring the rate of fall of
Data Storage
The ACT Plus® stores a total of 1000 test results, any combination of patient and
quality control tests. Test data is stored in the order performed. Patient and quality
control test data may be transferred to the ACT Plus EDM (external data manager)
using the floppy drive or USB port. Test data may be transferred to a laboratory or
hospital information system using the serial communication port with vendor
neutral connectivity software.
®
instrument is designed for use with the following
®
instrument.
Test Cartridge
Test cartridges for the ACT Plus® instrument consist of the reagent chamber, the
reaction chamber, and the plunger assembly.
ACT Plus® Operator’s Manual English 9
Figure 1. Test Cartridge
1
2
3
4
5
6
7
8
9
1. Flag
2. Plunger
3. Tang
4. Reaction Chamber
5. Daisy
6. Reagent Chamber
7. Bottom Plug
8. Upper Fill Line
9. Lower Fill Line
Reagent Chamber
The reagent chamber contains the activator and other reagents which initiate and
contribute to formation of a clot. The reagent chamber is enclosed on the top by the
“daisy” and on the bottom by a flexible plug.
Reaction Chamber
The reaction chamber is located above the reagent chamber. The test sample is
introduced into the reaction chamber. Upon test initiation the plunger assembly is
lifted and the bottom plug of the reagent chamber is pushed up. This action delivers
the contents of the reagent chamber into the reaction chamber and mixes the
reagents with the test sample.
10 English ACT Plus® Operator’s Manual
Plunger Assembly
The plunger assembly consists of the “daisy”, the plunger and the “flag”.
The “daisy” provides the upper seal for the reagent chamber and is the mechanical
sensing eleme
“daisy” moves up and down through the test sample/reagent mixture. Clot
formation impedes the movement of the “daisy” through the sample.
The “flag” is located at the top of the plunger. A photo-optical system senses motion
of the plunger tang through the reaction mixture. When a decrease in the plunger’s
fall rate is detected, the timer stops and the clotting time results are displayed.
nt for the formation of a clot. As the plunger rises and falls, the
ACT Plus® Operator’s Manual English 11
Section 2: Precautions
General Precautions
■
The ACT Plus® instrument is intended for in vitro diagnostic use only.
■
For use by healthcare professionals.
■
The ACT Plus® instrument is intended for use while positioned on a level
surface.
■
To obtain valid results, the ACT Plus® instrument’s operating parameters
must be set up appropriately.
■
The screen values shown in this manual are intended for illustration only; they
are not intended to match actual test results.
■
Cartridge channels must be filled to the proper level for testing. Refer to the
specific cartridge package insert for details.
■
Dropping, soaking, or otherwise misusing the ACT Plus® instrument may
prevent it from functioning properly.
No Unapproved Instrument Use
Do not use the ACT Plus® instrument in any manner not specified in the operator’s
manual or other applicable labeling. The ACT Plus
accordance with all the instructions and labeling provided by Medtronic. Any other
use may adversely affect the safe and effective use of the instrument.
Explosive Anesthetics
The instrument must not be used in the presence of explosive gases or
anesthetics.
®
instrument must be used in
Main Power Off and Disconnect Access
Do not locate the ACT Plus® instrument in a place that restricts quick and easy
access to the main-power switch or the power cord. The main-power switch (on/off)
at the rear of the instrument and the power cord must be accessible at all times,
because they are the means to turn off or disconnect the instrument from its power
source.
Electric Shock
Electric shock hazard exists; do not remove the instrument cover. Refer all
servicing to Medtronic Technical Service or other authorized service
representative.
12 English ACT Plus® Operator’s Manual
Proper Maintenance and Calibration
It is important that the ACT Plus® be maintained in good working order and
calibrated on a regular basis by a Medtronic-qualified service technician.
Otherwise, a reduction in the accuracy of a measuring function may occur.
No User-Serviceable Parts
There are no user-serviceable parts inside the ACT Plus® instrument case. Only a
Medtronic Service Representative or authorized representative should repair or
service the instrument.
Cleaning/Decontamination
Do not steam, autoclave, or immerse the ACT Plus® instrument. Refer to Section 9,
Maintenance, for cleaning instructions.
Biohazard
All blood samples, controls, blood filled cartridges, syringes, and needles should
be considered potentially hazardous and handled with universal precautions to
protect against blood-borne pathogens. Dispose of all blood samples and these
products in accordance with the guidelines established for the specific institution.
The user must wear protective gloves when handling potentially infectious
substances.
Use of Cartridges and Controls
Please refer to the specific cartridge and control package inserts for additional
detailed warnings and precautions regarding use. Treat all blood samples as
potentially infectious.
ACT Plus® Operator’s Manual English 13
Section 3: Instrument Components
Actuator Heat Block Assembly
The actuator heat block is the receptacle for test cartridges. The ACT Plus®
software maintains the actuator heat block temperature at 37.0°C ±0.5°C. The
actuator heat block rotates between the open and closed positions. Cartridges
can be inserted into and removed from the actuator heat block when it is in the
open position. To initiate a clotting test, the operator rotates the actuator heat block
to the closed position. At termination of a test, the actuator heat block will
automatically rotate to the open position, and the results will be displayed. A test
can be terminated manually and the actuator heat block rotated to the open
position by depressing the Stop key on the front panel.
14 English ACT Plus® Operator’s Manual
Section 4: Installation
Unpacking
1. The ACT Plus® instrument is shipped with the following items:
■
Operator's manual
■
Power cord
■
Temperature verification cartridge
■
Point of care quick start envelope
■
Quick reference guide
■
Actuator assembly cleaning kit
■
USB storage device
2. Check the box for damage, and report any damage to the carrier.
3. Remove the protective bag.
4. Set the ACT Plus® instrument on a level surface.
5. Check the ACT Plus® instrument for visible damage.
6. Retain all shipping materials.
Power ON: Initial Checkout Procedure
1. Connect the power cord to an AC power source with the same voltage rating
that is listed on the serial plate located on the bottom of the ACT Plus
instrument.
2. Rotate the actuator heat block to the closed position.
3.
[Optional] Attach the optional bar code scanner.
4. Turn the power switch, located on the rear panel, to the ON position. The
actuator heat block should rotate to the open position, followed by
illumination of all clotting time display segments, with the LCD indicating
“Self-Test Pass.” An audible tone indicating the end of the self-test should
accompany this sequence. The version of the ACT Plus
®
software will display
during this step.
5. From the Main Menu, select [Ca
rtridge Lot]. Scan the bar code label on the
cartridge package. The cartridge lot number and expiration date should be
entered in the appropriate cartridge-type fields.
Note: For manual data entry, refer to Section 6, Data Entry. For data entry
using the bar code scanner, refer to Appendix A or Appendix B.
Note: The lot number and expiration date must be entered before a cartridge
is u
ed; if no bar code scanner is used, see Section 6, Data Entry, for detailed
s
instructions.
6.
From the Main Menu, select [Cartridge Ty
pe]. Select the cartridge type
entered in Step 5, and press Enter.
7. Insert an empty cartridge into the actuator heat block.
®
ACT Plus® Operator’s Manual English 15
8. Rotate the actuator heat block to the closed position. The instrument should
begin a test.
9. Terminate the test by pressing the Sto
rotate to the open position.
10. Insert the temperature verification cartridge. After 10 minutes, check both the
erature indicator on the front panel and the temperature verification
temp
cartridge to confirm that both show a temperature of 37°C ±0.5°C. Adjust the
temperature, if necessary (see Section 9, Maintenance, “Temperature
Verification Cartridge”, for detailed instructions).
11. Insert a floppy disk or a USB storage device into the appropriate drive.
12. From the Main Menu, select [T
Unsent Patient Tests]. The results of the previous test should be sent to the
floppy disk or USB storage device.
13. If the instrument does not function properly, contact a Medtronic Service
Representative (see Section 10, Service and Troubleshooting).
r
p key. The actuator heat block should
ansmit Test Results], then [Transmit
Setup
Medtronic suggests taking the following steps prior to running tests with the ACT
®
Plus
instrument. Consult your Point of Care Coordinator for details.
1. Set Instrument Parameters (see page 26).
2. Set QC Manager parameters (see page 30):
a. Enter Instrument Location.
b. Enter User IDs.
c. Select QC Interval.
d. Set QC Lockout.
e. Enable Cartridge Type.
f. Set Permanent Record.
3. Enter Cartridge Lot Information (see page 23).
4. Enter Control Lot Information (see page 28).
5. Run Controls.
16 English ACT Plus® Operator’s Manual
Section 5: Instrument Controls
1
2
3
4
5
6
7
8
9
10
3
User Interface
Figure 2 identifies the User Interface components. A bar code scanner (not shown)
is optional.
Figure 2. User Interface
1. Clot Time Display
2. Data Entry Display
3. Variable Function Keypad
4. Quality Control Key
5. Cancel Key
6. Enter Key
7. Numeric Keypad
8. Clear Key
9. Stop Key
10. Main Menu Key
ACT Plus® Operator’s Manual English 17
1. The Clot Time Display is a red, seven-segment display of the clotting times
7
9
8
6
1
2
3
4
5
for Channels 1 and 2.
2. The Data
Entry Display, an LCD screen, displays clotting times. It also is
used to select a test, enter data, and navigate through the screens (see
Section 6, Data Entry, for detailed instructions).
3. The Variable Function Keypad is used to select the desired field in the
various screen options.
4. The Quality Control Key is used to access the Quality Control Menu.
5.
The Cancel Key is used to cancel any currently active function.
6. The Enter Key is used to accept entered numbers or data in an active field.
7. The Numeric Keypad is used to enter data into acceptable fields (ie, lot
numbers, user identification numbers [UIDs], and patient identification
numbers [PIDs]).
8. The Clear Key is used to clear the field.
9. The Stop Key is used to stop a test in progress.
10. The Main Menu Key is used to access the Main Menu.
Rear Panel
Figure 3 identifies the Rear Panel components.
Figure 3. Rear Panel Components
1. Handle
2. Fuse Holder
3. Main Power Switch
4. Power Inlet Module
(ON/OFF)
5. Equal Potential Terminal
18 English ACT Plus® Operator’s Manual
6. Serial Port (RS-232)
7. Serial Port (USB)
8. Bar Code Scanner Port
9. Floppy Drive
1. Handle.
2. The Fuse Holder is located
Maintenance, for fuse replacement.
3. The Main Power Switch is a rocker-type s
indicates OFF. It is located in the Power Inlet Module.
4. The Powe
instrument.
5. The Equal Potential Terminal
Plus
6. The Serial Port (RS-232) is a female DB-9 connector (see Section 11,
Specifications, for details).
7. The USB Port is u
8. The Bar Code Scanner Port is the connection for the optional bar code
scanner (see Section 11, Specifications, for details).
9. The Floppy Drive
information.
r Inlet Module is used to connect the power cord to the ACT Plus®
®
instrument to other instruments that may be used in a clinical setting.
sed to transfer and store patient and QC test information.
is used to transfer and store patient and QC test
in the Power Inlet Module. See Section 9,
witch: “|” indicates ON and “O”
is used to connect the ground of the ACT
Instrument Bottom
The Serial Number Label, which contains the serial number and safety agency
approvals, is located on the bottom of the ACT Plus
®
instrument.
Optional Bar Code Scanner
See Appendix A, Optional Bar Code Scanner LS1900, and Appendix B, Optional
Bar Code Scanner LS2208, for additional information.
ACT Plus® Operator’s Manual English 19
Section 6: Data Entry
MAIN MENU
View Test
Results
Transmit Test
Results
Instrument
Parameters
Enter ID
Cartridge Type
[HR-ACT]
Cartridge Lot
9710000772
Tem p : 3 7 . 1° C
PID:
Main Menu
The following variable function keys appear on the Main Menu: [Enter ID],
[Cartridge Type], [Cartridge Lot], [View Test Results], [Transmit Test Results],
and [Instrument Parameters].
The Main Menu can be accessed from any screen by pressing the [Main Menu]
n the left-hand side, below the display screen.
key o
Figure 4. Main Menu Screen
The [Enter ID] variable function key is used to enter patient (PID) and user (UID)
identification numbers. This information may be alphanumeric and is entered using
the numeric keypad.
The [Cartridge Type]
of cartridge for the test that is being performed (HR-ACT, LR-ACT, HTC, RACT,
GPC, or ACTtrac
and to the package insert for the specific cartridge.
The [Cartridge Lot]
and expiration date. Up to 2 lots of each cartridge type may be entered in the
database at a time.
The [View Test Results] va
last 20 tests performed including date, time, patient ID and test results.
variable function key is used to select the appropriate type
®
). For test method instructions, refer to Section 7 of this manual
variable function key is used to enter the cartridge lot number
riable function key is used to display the results of the
20 English ACT Plus® Operator’s Manual
The [Transmit Test Results] variable function key is used to transmit patient and
Quality Control (QC) results to an external location. Results may be exported to a
1.44 MB, 3.5-inch PC-formatted diskette using the floppy drive on the lower right
side of the instrument panel, a USB storage device using the USB port or results
may be exported via the serial port to a secure network interface (SNI) for
transmission to a Laboratory Information System (LIS) using vendor neutral
connectivity software. A message will display the number of test records
transmitted.
The [Instr
instrument settings: current date, current time, audio tone, language, screen
contrast adjustment, and output location.
ument Parameters] vari
able function key is used to enter the following
Enter Patient and User IDs
A Patient ID must be entered before a test can be run. The Patient ID can include
up to 12 alpha-numeric characters. Non alpha-numeric characters will be removed
when the Patient ID is scanned.
A User ID must be entered if the User Locko
User ID can include up to 15 alpha-numeric characters. Non alpha-numeric
characters will be removed when the User ID is scanned.
Note: Up to 600 user IDs can be entered.
Manual Entry
From the Main Menu:
1. Select [Enter ID]. The
Patient and User ID numbers.
To enter a Patient ID:
1. Select [Pat
2. Enter the Patient ID using the numerical keypad.
3. For an alpha character, press the decimal point “.” An alpha character will
appear in the Patient ID
4. To change the character, press the [A…Z] and [Z…
character is shown.
5. To accept the character select the decimal point
another alpha character or enter from the numeric keypad.
6. To continue entering numbers, select the
7. After entering the last alpha character or number in the Patient ID, press
[Enter] to confirm the ID.
8. The Patient ID will appear in the upper right-hand corner of the Enter ID
screen.
ent ID].
i
Enter ID screen will appear with options for entering
field.
ut option has been turned ON. The
A] keys until the desired
if you would like to enter
number using the numerical keypad.
ACT Plus® Operator’s Manual English 21
To enter a User ID:
1. Select [User ID].
2.
Enter the User ID using the numerical keypad.
3. For alpha entry, press the decimal point “.” An alpha character will appear in
the User ID field.
4. To select a character, press the [A…Z] and [Z…
character is shown.
5. Select the decimal point again if you would like to enter another alpha
character or enter from the numeric keypa
6. After entering the last alpha character or number of the User ID, press [Enter]
confirm the ID.
to
7. A message will appear, “Load Cartridge, Close Actuator.”
8. To return to the Main Menu, press [Main Menu]. A messag
top of the screen, “Verify Patient and User ID.”
A] keys until the desired
d.
e will appear at the
Bar Code Entry
From the Main Menu:
1. Select [Enter ID]. The
Patient and User ID numbers.
To enter a Patient ID:
1. Select [Pati
2. Scan the Patient ID from the patient band.
3. The Patient ID will appear in the upper right-hand corner of the Enter ID
screen.
To enter a User ID:
1. Select [User ID].
2
. Scan the User ID from the user's badge.
3. A message will appear, “Load Cartridge, Close Actuator.”
4. To return to the Main Menu, press [Main Menu]. A messag
top of the screen, “Verify Patient and User ID.”
ent ID].
Enter ID screen will appear with options for entering
e will appear at the
Select Cartridge Type
The cartridge type (including the electronic control) should be selected or verified
before each test.
From the Main Menu:
1. Select [Ca
2. To view the complete list of choices, [HR-ACT], [LR-ACT], [HTC], [RACT],
[GPC], and [ACTtrac
3. To confirm the selection, press [Enter].
22 English ACT Plus® Operator’s Manual
rtridge Type].
®
], continue to press [Cartridge Type].
Enter ACTtrac® Serial Number
From the Main Menu:
1. Select [Cartridge Lot].
2. To navigate through the list, use the arrows on either side of the box.
®
When the ACTtrac
appear: [Add Lot Number] and [Exit to Main Menu].
1. To enter the serial number, select [Add Lo
2. If the serial number has an “AT” at the beginning of it, press the “.” key and
the “AT” will appear in the serial number field.
3. Enter the rest of the serial number using the numeric keypad.
4. To confirm the entry, press [Enter].
cartridge type is activated, the following variable function keys
t Number].
Enter Cartridge Lot Number and Expiration Date
Manual Entry
From the Main Menu:
1. Select [Ca
2. To navigate through the list, use the arrows on either side of the box.
When there are no lot numbers for a cartridge type, the following variable function
keys appear: [Add Lot Numbe
1. To enter the lot number, select [Add Lot
Enter the cartridge lot number using the numeric keypad.
2.
3. To confirm the entry, press [Enter].
4. To enter the expiration date for the selected lot number, select [Add Exp Date]. The date format is yy-mm-dd.
5. Enter the cartridge expiration date.
6. To confirm the entry, press [Enter].
When a cartridge type has a single lot number entered and a second lot is to be
entered, the following variable function keys will appear: [Add Lot], [Remove Lot],
[Edit Lot/Exp Date] and [Exit to Main Menu].
1. To enter the second lot number, select [Add Lot].
2. To enter the lot number, select [Add Lot Number].
3. Enter the cartridge lot number using the numeric keypad.
4. To confirm the entry, press [Enter].
5. To enter the expiration date for the selected lot number, select [Add Exp Date]. The date format is yy-mm-dd.
6. Enter the cartridge expiration date.
7. To confirm the entry, press [Enter].
rtridge Lot].
r], [Add Exp Date], and [Exit to Main Menu].
Number].
ACT Plus® Operator’s Manual English 23
When there are 2 lot numbers entered for a cartridge type, the following variable
function keys will appear: [Toggl e Ac tive ], [Remove Lot], [Edit Lot/Exp Date] and
[Exit to Main Menu].
To change the lot number of the currently active cartridge:
1. Select [Toggle Active].
2. Move the “*” to the cartridge lot number that is currently in use.
To remove a cartridge lot:
1. Select [Rem
2. To select the lot to be removed, use the up/down arrows (↑↓) to toggle to the
appropriate lot number.
3. To delete the selected cartridge lot number, select [Remov
ove Lot].
e Selected Lot].
Bar Code Entry
From the Main Menu:
1. Select [Ca
will appear.
When there is no or 1 lot number for a cartridge type:
1. Scan the bar code on the cartridge box. T
will automatically populate their respective fields.
2. To return to the Main Menu, press [Main
When 2 lot numbers exist for a cartridge type:
1. Select [Remove Lot].
2. To remove 1 of the lot numbers, toggle to the appropriate lot number.
3. To delete the selected cartridge lot number, select [Remove Selected Lot].
4. Scan the bar code on the cartridge box. The lot number and expiration date
will automatically populate their respective fields.
5. To return to the Main Menu, select [Main Menu].
rtridge Lot]. Page 1 of the Cartridge Lot/Expiration Date screen
he lot number and expiration date
Menu].
View Test Results for the Last 20 Tests Performed
1. To display the results for the last 20 tests, select [View Test Results].
2. Use the up/down arrows (↑↓) on the left and right of the screen to scroll
through the test data.
3. The date, time, patient ID and test results will be displayed.
Note: If the ACT Plus® power is cycled, the ability to display previous results is lost;
however, they are not erased from the result log.
24 English ACT Plus® Operator’s Manual
Transmit Patient and Quality Control (QC) Test Results
Patient and QC test results can be transmitted using the floppy drive, USB port, or
serial port. The transmit destination is set from the Instrument Parameters Menu.
Transmission Using Floppy Drive
From the Main Menu:
1. Select [Transmit Test Results]. The following variable function keys will
appear: [Transmit All Patient Tests], [Transmit Unsent Patient Tests],
[Transmit By Patient ID], [Transmit All QC Tests], [Transmit Unsent QC Tests ], and [Exit to Main Menu].
2. Insert a 3.5-inch IBM formatted floppy disk.
3. To transmit all patient test results in the memory
Tests ].
4. To transmit the patient test results that h
selected output location, select [Transmit Unsent Patient Tests].
5. To transmit all results for a specific patient, select [Tr
Then enter the Patient ID and press [Enter].
6. To transmit all QC test results (liquid and electronic), select [TraTests ]. All QC tests in the memory will be sent to the selected output location.
7. To transmit all QC test results that have not been previously sent to the
selected
Note: When transmission is complete, the number of test records sent will be
displayed.
1. Select [Transmit Test Results]. The following variable function keys will
appear: [Transmit All Patient Tests], [Transmit Unsent Patient Tests],
[Transmit By Patient ID], [Transmit All QC Tests], [Transmit Unsent QC Tests ], and [Exit to Main Menu].
2. Insert a USB storage device.
3. To transmit all patient test results in the memory
Tests ].
4. To transmit the patient test results that h
selected output location, select [Transmit Unsent Patient Tests].
5. To transmit all results for a specific patient, select [Tr
Then enter the Patient ID and press [Enter].
6. To transmit all QC test results (liquid and electronic), select [TraTests ]. All QC tests in the memory will be sent to the selected output location.
7. To transmit all QC tests that have not been previously sent to the selected
output location, select [T
ransmit Unsent QC Tests].
, select [Transmit All Patient
ave not previously been sent to the
ansmit By Patient ID].
nsmit All QC
Menu].
, select [Transmit All Patient
ave not previously been sent to the
ansmit By Patient ID].
nsmit All QC
ACT Plus® Operator’s Manual English 25
Note: When transmission is complete, the number of test records sent will be
displayed.
Note: When transmission is complete, it is safe to remove the USB storage device.
8. To return to the Main Menu, press [Main Menu].
Set Instrument Parameters
Parameters for the ACT Plus® instrument are set at the time of installation. This
must be completed prior to any other data entry.
1. Select [Instrume
appear: [Date - yyyy-mm-dd], [Time – hh:mm], [Audio Tone], [Language],
[Go to Page 2 of 2], and [Exit to Main Menu].
2. To set the current date, select [Date - yyyy-mm-dd]. The following ranges are
acceptable for the year, month, and day entries: yyyy = year (accepted range,
1998 to 2097), mm = month (accepted range, 01 to 12), and dd = days
(accepted range, 01 to 31).
3. To confirm the date entry, press [Enter].
Note:
The current date must be entered before cartridge and/or control lot
rs and expiration dates are entered.
numbe
4. To set the current time, select [Tim
a 24-hour clock where hh = hours (accepted range, 00 to 23) and
mm=minutes (accepted range, 00 to 59).
5. To confirm the time entry, press [Enter].
6. To set the audio tone to ON or OFF, select [Audio Tone].
7. To confirm the selection, press [Enter].
8. To select a language, select [Language]. The options are English, French
(Français), Italian (Italiano), German (Deutsch), Spanish (Español), Swedish
(Svenska), Greek (Ελληνικά), Danish (Dansk), or Norwegian (Norsk).
9. To view the choices, continue to press [La
English.
10. To confirm the selection, press [Ente
The following variable function keys appear on p
Parameters screens: [Screen Contrast Adjustment], [Output Location], [Back to Page 1 of 2], and [Exit to Main Menu].
1. To adjust the contrast on the
Adjustment]. The (-) key will lighten the text, and the (+) key will darken the
text.
2. To select the output mode for transferring patient and QC data, select [Output Location].
3. To view the choices, [FlLocation].
4. To confirm the selection, press [Ente
nt Parameters]. The following variable function keys
e – hh:mm]. The current time is based on
uage]. The default language is
ng
r].
age 2 of the Instrument
LCD screen, select [Screen Contrast
oppy], [USB], or [Serial], continue to press [Output
r].
26 English ACT Plus® Operator’s Manual
5. To return to the Main Menu, press [Main Menu].
--->> QC <<---
Tem p : 3 7 . 1° C
QC Due Status
QC Manager Setup
Exit to
Main Menu
QUALITY CONTROL MENU
Control Type
[HR-NM]
Control Lot
46372985102
Temperature
Adjustment
Quality Control (QC) Menu
To access the Quality Control Menu select the [Quality Control] key on the right
below the data entry screen. If User Lockout is turned ON, a valid User ID will be
needed. Enter the valid User ID from the keypad or scan the user’s badge.
The following variable function keys appear on the Quality Control Menu: [Contro
Type], [Control Lot], [Temperature Adjustment], [QC Due Status], [QC
Manager Set Up], and [Exit to Main Menu].
Note: All QC testing (liquid and electronic) must be performed from the Quality
Cont
rol Menu in order to log the controls in the QC files of the Data Manager.
l
Figure 5. Quality Control Menu Screen
General Quality Control (QC) Functions
The [Control Type] variable function key is used to select the type of control to be
performed. The following control types are available and will be determined by the
test cartridge being used:
■
The ACTtrac® electronic control has 3 settings: 98-102, 190-204, and
490-510.
■
High Range ACT (HR-ACT) cartridge: Normal (HR-NM) and Abnormal
(HR-AB) controls.
■
Low Range ACT (LR-ACT) cartridge: Normal Citrated Whole Blood (CWB)
and Abnormal (LR-AB) controls.
■
High Range Heparinase (HR-HTC) cartridge: Normal (HR-NM) and
Heparinase (HTC) controls.
■
Recalcified ACT (RACT) cartridge: Normal Citrated Whole Blood (CWB) and
Abnormal (RACT-AB) controls.
ACT Plus® Operator’s Manual English 27
The [Control Lot] variable function key is used to enter, remove, or edit the lot
number, expiration date, and range of the selected control. Up to 2 lots of each
control type may be entered into the database at a time.
The [Temperature Adjustment] varia
heat block temperature (refer to Section 9, Maintenance, for additional
information).
The [QC Due Status]
electronic controls, including the date and time that the control tests must be
performed.
variable function key displays the status of liquid and
ble function key is used to verify the actuator
Select the Control Type
The correct control type must be selected before the control test is initiated.
From the Quality Control Menu:
1. To select the type of control to be performed, select [Control Type].
2. To view the complete list of choices, continue to press [Control Type].
3. To confirm the selection, press [Enter]. This selection will display only the
available control types for the test selected in the Main Menu.
Enter Control Lot Number, Expiration Date, and Range
Manual Entry
From the Quality Control Menu:
1. Select [Control Lot].
2. To select the desired control, highlight the control in the list.
When there are no lot numbers entered for a
keys appear: [Add Lot Number], [Add Exp Date], [Set Range], and [Exit to Quality Control Menu].
1. To enter the lot number, select [Add Lot Number].
2
. Enter the lot number using the keypad.
3. To confirm the selection, press [Ente
4. To enter the expiration date for the selected lot number, select [Add Exp Date]. The date format is yy-mm-dd.
5. Enter the expiration date.
6. To confirm the entry, press [Enter].
7. To enter the range for the control, select [Set Ra
the control range is “lll-hhh,” where “lll” is the low end and “hhh” is the high
end of the control range.
Note: If the low range is less than 100, a zero must precede the 2-digit
number
, eg, “099.”
8. Enter the range.
control, the following variable function
r].
nge]. The format for entering
28 English ACT Plus® Operator’s Manual
9. To confirm the entry, press [Enter].
10. To return to the “Add Lot Number” selection, select [Exit Add Selection].
When a control type has a single lot number entered and a second lot is to be
entered, the follo
[Remove Lot], [Edit Range] and [Exit to Quality Control Menu].
1. To add a second control lot for a cartridge type, select [Add Lot/EXP Date].
2. To enter the second lot number, see the instructions above (when there are
no lot numbers).
When there are 2 lot numbers entered for a control type, the following variable
nction ke
fu
Quality Control Menu].
To change the lot number of the active control:
1. Select [T
currently being used.
To remove a control lot:
1. Select [R
2. To select the lot to be removed, use the up/down arrows (↑↓) to toggle to the
appropriate lot number.
3. Select [Re
To edit the range of an existing control lot:
1. Select [Edit Range].
2. To select the lot of the control to be edited, use the arrows to change the
high
3. Enter the range for the control.
4.
To confirm the entry, press [Enter].
wing variable function keys appear: [Add Lot/EXP Date],
ys appear: [Toggle Active], [Remove Lot], [Edit Range], and [Exit to
oggle Active] to move the “*” to the control lot number that is
ve Lot].
emo
move Selected Lot] to delete the selected control lot number.
lighted lot.
Bar Code Entry
From the Quality Control Menu:
1. Select [Cont
When there is no or 1 lot number for a control type:
1. Scan the bar code on the control box. The lot number and expiration date will
automatically popu
2. To enter the range for the control, select [Set Ra
the control range is “lll-hhh,” where “lll” is the low end and “hhh” is the high
end of the control range.
3. To confirm the entry, press [Enter].
4. To return to the Quality Control Menu, select [Exit to Qu
ACT Plus® Operator’s Manual English 29
rol Lot].
late their respective fields.
nge]. The format for entering
ity Control Menu].
al
When a control type has 2 lot numbers entered for a control type:
1. To remove one of the lot numbers for the control type, select [Remov
e Lot].
2. To delete the selected control lot number, toggle to the appropriate lot number
and select [Remove Selected Lot].
3. To enter the new lot number, scan the bar code on the control box. The lot
number a
4. To enter the range for the control, select [Set Ra
nd expiration date will automatically populate their respective fields.
nge]. The format for entering
the control range is “lll-hhh,” where “lll” is the low end and “hhh” is the high
end of the control range.
5. To confirm the entry, press [Enter].
6. To return to the Quality Control Menu, select [Exit to Quality Control Menu].
To verify/record the ACT Plus® actuator heat block
temperature
For detailed instructions on verifying and recording the actuator heat block
temperature, refer to Section 9, Maintenance.
To view the due status for quality control testing
From the Quality Control Menu:
1. Select [QC Due
electronic controls for the cartridges that have been enabled with the following
message options:
■
The date and time when the next control tests are due.
■
QC due, if the controls for a test are past due.
■
QC not performed, if controls have never been run for a cartridge type.
Status]. The ACT Plus® instrument will display all liquid and
QC Manager Functions
The QC Manager Setup Menu is accessed from the Quality Control Menu; it is a
password-protected area. The default password is provided in a separate
envelope. If this card is lost, contact Medtronic Technical Service for assistance
(800-328-3320). Non-US customers, please contact a local/regional Medtronic
sales office.
The following variable function keys appear in the QC
[Location], [User ID Setup], [Clear ID Interval], [QC Lockout], [QC Interval Setup], [Cartridge Enable Setup], [Download Settings], [Perm Record], and
[Instrument Upgrade].
Note: QC Manager parameters must be set before general QC information is
entered and before any tests are performed.
30 English ACT Plus® Operator’s Manual
Manager Setup Menu:
The [Location] variable function key is used to select the location of the ACT Plus®
--->> QC Manager <<---
--->> QC Manager <<---
Q
C
L
o
c
k
o
u
t
[
O
f
f
]
Q
C
I
n
t
e
r
v
a
l
S
e
t
u
p
G
o
t
o
P
a
g
e
2
o
f
2
QC MANAGER SETUP
(Page 1 of 2)
QC MANAGER SETUP
(Page 2 of 2)
Cartridge
Enable Setup
Perm Record
[USB]
Instrument
Upgrade
Exit to Quality
Control Menu
Download
Settings
Go to
Page 1 of 2
Location
[CVOR]
User ID Setup
Clear ID Interval
[20 minutes]
instrument. Location options include the following: CVOR (cardiovascular
operating room), Cath Lab (Cardiac Catheterization Lab), ECMO (Extracorporeal
Membrane Oxygenation), Dialysis, ICU/CCU (Intensive Care/Coronary Care
Units), Lab, and Other.
The [User ID Setup] variable function key displays the screens for the following
fu
changing the QC Manager Password, downloading User IDs to a floppy disk or
USB storage device, and uploading User IDs from a floppy disk or USB storage
device to the ACT Plus
The [Clear ID In
(none, 5 minutes, 10 minutes, 20 minutes, 60 minutes, or 120 minutes) that will
elapse before the Patient and User IDs are cleared. Once cleared, the PID and UID
will need to be reentered prior to testing.
The [QC Lockout] variable function key is used to set QC Lockout to 1 of the
following options: OFF, Warning, or ON.
The [QC Interval Setup] variable function key is used to set the required interval
for performing liquid and electronic QC tests. The options for liquid are the
following: None, 8 hours, or 7 days. The options for ACTtrac are None or 8 hours.
The [Cartridge Enable Setup] varia
the various cartridge types that may be run on the ACT Plus
The [Download Settings] va
parameters and settings to a floppy disk or USB storage device.
The [Per
the results that have been downloaded. The options are None, Floppy, USB, or
Electronic.
The [Instrument Upgrade] variab
ACT Plus
ACT Plus® Operator’s Manual English 31
Figure 6. QC Manager Setup Screen
nctions: adding and deleting User IDs, turning ON or OFF User Lockout,
®
user database.
terval] variable function key is used to select the time interval
ble function key is used to enable or disable
®
instrument.
riable function key is used to download all instrument
m Record] variab
®
software.
le function key designates the official mode of tracking
le function key is used to install upgrades to the
Select the location for the ACT Plus® instrument:
From the QC Manager Setup:
1. To change the desired location, select [Loca
2. Continue to press [Location] until the correct location is found.
3. To confirm the selection, press [Enter].
tion].
Add/delete User IDs:
Note: Up to 600 User IDs can be entered into the ACT Plus.
From the QC Manager Setup screen:
1. Select [User ID Setup].
From the User ID Setup screen:
1. Select [Add/Delete User IDs]. The following keys appear: [Add], [Delete],
and [Exit to User ID Setup].
Add a User ID:
From the Add/Delete User IDs screen:
1. Select [Add].
2. Enter the User ID using the numeric keypad. Press the decimal point “.” on
the keypad to activate alpha characters.
3. To select the desired character, use the [A…Z] and
4. To confirm the selection, press [Enter].
[Z…A] keys.
Delete a User ID:
From the Add/Delete User ID screen:
1. Select [Del
2. Use the arrows on either side of the displayed box to navigate up or down to
select the desired User ID.
3. To remove the User ID from the list, select [Delete Selection].
4
. To return to the User ID Setup screen, select [Exit to User ID Setup].
ete].
Enable User Lockout:
From the User ID Setup screen:
1. Select [User L
2. To view the choices, [ON] or [OFF], continue to press [User Lockout].
3. To confirm the selection, press [Enter].
32 English ACT Plus® Operator’s Manual
ockout].
Change the QC Manager Password:
From the User ID Setup screen:
1. Select [QC Mgr Password].
2.
Enter up to a 6-digit numeric password.
3. Enter the password a second time to verify.
4. To confirm, press [Enter].
5. Record the new password in a secure location.
Download User IDs:
From the User ID Setup screen:
1. Insert a formatted disk into the floppy drive or insert a USB storage device into
the USB port.
2.
Select [Down
to the floppy drive or USB storage device.
Note: When transmission is complete, it is safe
load User IDs]. All of the User IDs entered will be downloaded
to remove the USB storage device.
Transfer User IDs from ACT Plus® Instrument to Multiple
Instruments:
From the User ID Setup screen:
1. Insert either a formatted disk or a USB storage device containing downloaded
User IDs into the appropriate drive.
2. Select [Uplo
Note: When transmission is complete, it is safe
User IDs also can be uploaded to the ACT Plus
from the ACT Plus
for specific instructions.
ad User IDs].
to remove the USB storage device.
®
®
External Data Manager (EDM). Refer to the EDM Instructions
instrument using a file generated
Select the Time Interval for Clearing Patient and User IDs:
From the QC Manager Setup screen:
1. Select [Cl
2. To view the choices, [None], [5 minutes], [10 minutes], [20 minutes],
[60 minutes], and [12
3. To confirm the choice, press [Enter].
ear ID Interval].
0 minutes], continue to press [Clear ID Interval].
Select the QC Lockout Function:
From the QC Manager Setup screen:
1. Select [QC
2. To view the choices, [ON], [Warning], or [OFF], continue to press [QC Lockout].
3. To confirm the choice, press [Enter].
ACT Plus® Operator’s Manual English 33
Lockout].
Set the QC Interval for Liquid and Electronic Controls:
From the QC Manager Setup screen:
1. Select [QC Interval Setup]. T
[Liquid Interval], [ACTtrac
he following variable function keys appear:
®
Interval] and [Exit to QC Manager Setup].
To set the liquid QC interval:
1. Select [Liquid Interval].
2. To view the choices, [None], [8 hours] or [7 days], continue to press [Liquid Interval].
3. To confirm the choice, press [Enter].
To set the ACTtrac® electronic QC interval:
1. Select [ACTtrac® Interval].
2. To view the choices, [None] or [8 hours], continue to press [ACTtrac®
Interval].
3. To confirm the choice, press [Enter].
Enable or Disable Test Cartridge Types:
From page 2 of the QC Manager Setup screens:
1. Select [Ca
2. To select the desired cartridge, use the arrows on either side of the display of
cartridges to navigate up or down.
3. Select the desired cartridge.
4. To view the choices, [ON] or [OFF],
5. To confirm the choice, press [Enter].
rtridge Enable Setup].
continue to press the desired cartridge.
Download All of the Instrument Settings to a Floppy Disk or USB
Storage Device:
From page 2 of the QC Manager Setup screens:
1. Insert a formatted disk or USB Storage Device into the appropriate drive.
2. Select [Down
load Settings].
Set the Permanent Record or the Mode by Which the ACT Plus®
Instrument Maintains a Record of Downloaded Data:
From page 2 of the QC Manager Setup screens:
1. Select [Perm Record].
2.
To view the choices, [None], [Floppy], [USB] or [Electronic], continue to
press [Perm Record].
34 English ACT Plus® Operator’s Manual
3. To confirm the choice, press [Enter].
Note: This does not affect the output location previously selected in
Instrument Parameters. This selection will warn only if test records will be
overwritten.
Upload Software Upgrades from a Floppy Disk or USB Storage
Device:
From page 2 of the QC Manager Setup screens:
1. Insert the floppy disk or USB Storage Device containing the
Medt
ronic-released software upgrade for the ACT Plus
appropriate drive.
2. Select [Instrument Upgrad
e].
3. Enter the Upgrade Password sent with the software upgrade.
4. Select [Start Software Upgrade].
5. When the software upgrade is complete, power OFF then ON.
®
instrument into the
ACT Plus® Operator’s Manual English 35
Section 7: Test Methods
General
Notes:
■
Lot numbers and expiration dates for cartridges and controls must be entered
prior to running a test (refer to Section 6, Data Entry for detailed instructions).
■
The User ID and Patient ID numbers must be entered prior to running a test
(refer to Section 6, Data Entryfor detailed instructions).
1.
Power ON the ACT Plus
actuator heat block to reach a temperature of 37.0°C ±0.5°C.
Note: When the heat block temperature is less than 36.0°C the software will
not allo
patient test to be initiated. The message “Heat Block Lockout
w a
<36.0°C” is displayed.
2. From the Main Menu, select [Ca
the correct cartridge is selected. Press [Enter] to confirm.
3. Prewarm the cartridges for at least 3 minutes before col
sample. Cartridges for HR-ACT, LR-ACT, and RACT may be prewarmed for
up to 12 hours without affecting performance. For specific information, consult
the package insert that accompanies each cartridge.
4. Tap the cartridge to resuspend the reagent in the chamber before adding the
test sample.
5. Fill each cartridge chamber with the appropriate sample volume. The sample
leve
should be between the upper and lower fill lines in each reaction
l
chamber. Allow the sample to flow down the back of the chamber, taking care
to avoid getting the sample on the plunger tang (see Figure 1).
6. Insert the cartridge into the actuator heat block and rotate the block to the
closed
position.
7. Reagent is delivered into the reaction chamber, and the time to clot formation
is measured
.
8. When clot formation is detected, or if the test i
sounds, and the actuator heat block automatically rotates to the open
position.
®
instrument. Allow at least 10 minutes for the
dge Type]. Continue to press this key until
rtri
lecting the test
s terminated, an audible tone
36 English ACT Plus® Operator’s Manual
Activated Clotting Times: High Range (HR-ACT), Low
Range (LR-ACT), and Recalcified (RACT) Cartridges
Table 1 lists the Medtronic Activated Clotting Time Cartridges available for use with
the Medtronic ACT Plus
Table 1. Medtronic Activated Clotting Time Cartridges
ProductSample TypeChannel Sample
HR-ACTFresh drawn0.4 mLCardiovascular or
LR-ACTFresh drawn0.2 mLTherapeutic (0-1.5
RACTCitrated0.2 mLTherapeutic (0-1.5
®
instrument.
Vol ume
Intended Use
Cath Lab (1 U
Heparin/mL or more)
s Hep
Unit
Units Hep
nit
arin/mL)
arin/mL)
Table 2 displays the cartridge reagents (0.1 mL per channel) a
nd maximum
prewarm times for each cartridge type and cartridges with sample:
Duplicate channel results should fall within 10% of each other for baseline
(unheparinized) samples and within 12% of each other for extended or heparinized
samples.
ACT Plus® Operator’s Manual English 37
Sample Calculation:
Channel 1 clotting time210 seconds
Channel 2 clotting time200 seconds
Mean clotting time205 seconds
Difference10 seconds
12%25 seconds
The difference of 10 seconds is less than 12%,
which is 25 seconds. These results
are acceptable. The instrument’s operable range is 25 to 999 seconds.
High Range Heparinase (HR-HTC)
The HR-HTC cartridge is a modification of the High Range Activated Clotting Time
Cartridge used to identify the presence of heparin in a fresh whole blood sample.
One of the cartridge channels contains a purified heparinase, an enzyme specific
for heparin, which can rapidly neutralize up to 6 U/mL of heparin in a test sample.
For detailed instructions on using the HR-HTC Cartridge, refer to the Activated
Clotting Time Cartridges Instructions for Use.
38 English ACT Plus® Operator’s Manual
Section 8: Quality Assurance
Instrument Self-Test
The ACT Plus® instrument performs the following self-tests when powered ON:
1. The actuator assembly cycles and opens the a
it is closed.
2. An audio beep will sound, the Startup screen will appear, and the message,
“Self-Te
3. The message, “Self-Test Pass,” is displa
to the Main Menu.
Note: If any of the instrument’s self-tests
software will display the message “Self-Test: FAIL XXX,” where XXX is a
3-digit number indicating a System Error Code. Record this code (if
displayed), and notify the Medtronic Service Department or an authorized
representative before continuing (refer to Section 10, Service and
Troubleshooting).
4. The temperature will be displayed automatica
Note: Allow a minimum of 10 minutes of warm-up time for the actuator heat
bloc
actual warm-up time will depend on the instrument’s ambient temperature.
st in Progress” will be displayed.
k to reach the desired temperature before performing any tests. The
Liquid QC
Liquid controls are used to verify instrument and cartridge performance and
operator technique. When only liquid controls are used the minimum QC
requirements are 2 levels of control performed every 8 hours that the device is used
for patient testing. CLOTtrac
Low Range, Recalcified, and High Range Heparinase cartridges.
®
Coagulation Controls are available for High Range,
ctuator heat block assembly if
yed followed by automatic transition
fail or are not completed, the
lly.
Electronic QC (ACTtrac®)
The ACTtrac® Electronic Quality Control is an interactive, mechanical,
software-controlled verification device that includes both quantitative and
qualitative results. It interacts with the ACT Plus
emulating certain functions of a test cartridge. The ACTtrac
aspects of the ACT Plus
flag sensor function, reagent delivery pin height, lift wire height, and 3-level tests
(emulated clotting time ranges).
ACT Plus® Operator’s Manual English 39
®
instrument that relate to proper test cartridge function:
®
instrument by mechanically
®
checks the following
Notes:
■
The ACTtrac® may be used to supplement liquid quality control testing. Refer
to the current regulatory agency guidelines on the acceptability and use of
electronic controls.
■
Refer to the ACTtrac® Operator’s Manual for the specific instructions and
additional information.
40 English ACT Plus® Operator’s Manual
Section 9: Maintenance
1
1
2
3
4
Instrument Case Cleaning
The instrument case and exposed surfaces of the actuator and dispenser should
be kept clean. Clean the case routinely by wiping off dust and dried blood with a
cloth dampened with water or 1 of the following chemicals: isopropyl alcohol,
bleach, Liqui-Nox
or mild detergent. The ACT cleaning kit can be
areas in the dispenser. Cleaning kits can be obtained from Medtronic Customer
Service or an authorized representative.
Actuator Assembly Cleaning
The actuator assembly should be cleaned at least once a month and more
frequently if required. If blood should get into the actuator assembly (see Figure 7),
it is critical that the instrument is cleaned as soon as possible.
Liqui-Nox® is a registered trademark of Alconox, Inc.
2
Sani-Cloth® is a registered trademark of Nice-Pak Products Inc.
3
Clorox® is a registered trademark of the Clorox Company.
ACT Plus® Operator’s Manual English 41
An instrument cleaning kit (Catalog number 303-58) is provided with each ACT
®
Plus
instrument. Use the materials in the kit to perform the following cleaning
procedure:
®
1. Dip the swab provided in the enclosed packet in the Liqui-Nox
solution.
2. Swab the flag lift wire, removing all blood.
3. Swab inside the actuator assembly cover, especially the detector and emitter
areas of the photo-optical sensor.
®
4. Remove any excess Liqui-Nox
solution with a dry swab.
5. If blood should get into the detector of the lamp area and cannot be removed
with a swab, Error Code “4” may be displayed.
Note: Cleaning kits can be obtained from Medtronic Customer Service or an
authorized representative.
Actuator Heat Block Temperature
The actuator heat block temperature should be checked once a month to verify that
it is 37°C ± 0.5°C. The ACT Plus
into its temperature log.
Temperature Verification Cartridge
1. Turn the ACT Plus® instrument ON and allow the instrument to warm up for
15 minutes.
2. Place the Temperature Verification Cartridge (Catalog No. 313-11) into the
actuator heat blo
3. Wait for temperature equilibration to occur (minimum 10 minutes) and check
the T
emperature Verification Cartridge reading.
4. The instrument-displayed temperature
Cartridge temperature should both be within 36.5°C to 37.5°C. The
Temperature Verification Cartridge temperature should be within ± 0.5°C of
th
e instrument-disp
From the Quality Control Menu:
5. To change the temperature, if n
been verified, select [Temperature Adjustment].
6. Enter the temperature verification cartridge reading using the numeric keypad
(values
must be between 35°C and 39°C).
7. To accept the entry, press [Ente
temperature verification cartridge and the display will be recorded in the
temperature log.
Note: If the temperature is still not wit
adjustments can be repeated after a minimum of 10 minutes. To repeat
adjustmen
ts, select [Repeat Adjustment].
ck.
layed temperature.
®
instrument records all temperature adjustments
and the Temperature Verification
eede
d, or to record that the temperature has
r]. The time, date, and temperatures of the
hin the specified range (see Step 5),
42 English ACT Plus® Operator’s Manual
8. If you are unable to calibrate the display to match the temperature on the
Temperature Verification Cartridge, contact the Medtronic Service
Representative or an authorized distributor.
Verification with Thermometer
1. Turn the ACT Plus® instrument ON and allow it to warm for 15 minutes.
2. Use a cartridge, with the plunger assembly removed and filled with 0.2 to
0.3 mL of water. Insert the cartridge into th
3. Place a calibrated thermometer in one of the cartridge reaction chambers.
4. Wait for temperature equilibration to occur (minimum 10 minutes) and check
the thermometer reading.
5. The instrument-displayed temperature and thermometer-measured
temperature shou
thermometer-measured temperature should be within ± 0.5°C of the
instrument-disp
From the Quality Control Menu:
6. To change the temperature, if needed, or
been verified, select [Temperature Adjustment].
7. Enter the thermometer reading using the keypad (values must be between
35°C and 39°C).
8.
To accept the value, press [Enter]. The ti
thermometer and the display will be logged in the temperature log.
Note: If the temperature is sti
adjustments can be repeated after a minimum of 10 minutes. To repeat
adjustme
nts, select [Repeat Adjustment].
9. If you are unable to calibrate the display to match the thermometer-measured
temperature, con
distributor.
ld both read within 36.5°C to 37.5°C. The
ayed temperature.
l
ll not wit
tact the Medtronic Service Representative or an authorized
hin the specified range (see Step 5),
e actuator heat block.
record
that the temperature has
me, date, and temperatures of the
Fuse Replacement
Two fuses are located in the Power Inlet Module on the ACT Plus® instrument’s
rear panel. To replace or examine the fuses:
1. Disconnect the power cord.
2. Open the cover of the fuse holders using a small flathead screwdriver.
3. Slide the fuse holders out and, if necessary, replace the fuses with T2A 250-V
fuses.
4. Insert the fuse holders back into the Power Inlet Module.
5. Close the cover.
6. Power ON the instrument to
power failure.
ACT Plus® Operator’s Manual English 43
determine
if replacing the fuse(s) corrected the
Note: If fuses blow repeatedly, contact a Medtronic Service Representative or an
authorized representative.
Preventive Maintenance
Annual preventive maintenance is recommended for the ACT Plus® to ensure
accurate performance and reliability. A qualified Medtronic Service Representative
should perform all maintenance or service. The inspection and maintenance may
be performed at your site or at Medtronic.
44 English ACT Plus® Operator’s Manual
Section 10: Service and Troubleshooting
Service
For all repair requests in the United States, contact Medtronic CardioVascular
Service at 1-800-433-4311. Non-US customers, please contact a local/regional
Medtronic sales office.
Service training for Medtronic products is available to qualified hospital personnel
request. Detailed information, such as circuit diagrams, component parts
upon
lists, and calibration instructions, is available for qualified technical personnel.
Do not attempt to ser
period, as any manufacturer’s warranty will be voided.
WARNING: Do not adjust, modify, repair, or touch
actions could injure the operator or cause faulty operation of the ACT Plus
these actions may injure the operator or cause malfunction of the instrument.
Refer all repair requests to a Medtronic Instrument Service Representative. In the
United States: Medtronic CardioVascular. Refer to the back cover for contact
information.
Instrument Return
To return an ACT Plus® instrument for service, contact your Medtronic Service
Representative. Original shipping materials should be used to return the ACT
Plus® instrument; it should be enclosed within the foam inserts and placed in the
box. The actuator heat block should be in the closed position when it is shipped. If
the original box is no longer available, notify Medtronic Instrument Service or an
authorized representative to obtain shipping materials.
vice the ACT Plus® instrument during the warranty
the internal circuitry. These
®
as
End of Life Disposition
Do not dispose of this product in the unsorted municipal waste stream. Dispose of
this product according to local regulations. See
instructions on proper disposal of this product.
http://recycling.medtronic.com for
Troubleshooting
The ACT Plus® instrument self-diagnoses many error and precautionary
conditions, which are displayed on the LCD screen.
Message Types
The ACT Plus® instrument provides 3 types of messages to indicate its operating
condition: Information/Status; Alerts; and System Alarms. These messages, with
the exception of System Alarms, appear in the screen’s Status Message area.
ACT Plus® Operator’s Manual English 45
■
Information/Status: Screen message with no audio tone.
■
Alerts: Screen message with a 3-beep audio tone (audio is active when the
[Audio Tone] parameter is set to ON).
■
System Alarms: Screen message with a single, long audio tone (audio is
always active). System Alarms are presented when the ACT Plus
®
instrument’s self-testing mechanism detects a condition that may affect the
ability of the instrument to function properly. These messages are displayed,
if possible, and may be accompanied by a 3-digit Error Code, which service
personnel use to identify the system error.
Message Priorities
Messages are displayed in a priority structure. If multiple error or precautionary
conditions occur simultaneously, only the highest-priority message appears on the
display.
.
System MessagesCause/Resolution
Both lot numbers are filled.
OR
Both cartridge lots full.
Bar code is not a cartridge type.The user is trying to scan a control lot on the
Bar code is not a control type.The user is trying to scan a cartridge lot on
Bar code is not
valid.The bar code information was not valid as a
Expired cartridge lot.
OR
Expired control lot.
Cartridge lot already exists.
OR
Control lot already exists.
Cartridge is not enabled.The cartridge selected is not enabled in the
Cartridge for control is not enabled.The user is trying to scan in a lot number for
Expiration date edited. QC due.The expiration date for the cartridge lot has
The user is trying to add a lot number when
2 already exist for the selected cart
ridge.
Remove 1 of the lot numbers.
i
dge setup screen.
cartr
the control setup screen.
dge or control. Rescan the bar code.
cartri
Reprogram the bar code scanner.
The user is attempting to run a test with a
i
dge/control that is beyond the
cartr
expiration date. Make sure that the
expiration date on the packaging matches
the date in the cartridge lot entry screen.
The user is trying to scan a lot number when
that lot number
already exist
s for the
selected cartridge/control.
Man
ager Setup screen. Contact your QC
QC
Manager to enable the selected cartridge.
a control whe
n the cartridge th
at uses that
control is not enabled in the QC Manager
Setup. Contact your QC Manager to enable
the selected cartridge.
d; con
been change
trol tests must be
repeated for this lot. Repeat the controls for
this lot number.
46 English ACT Plus® Operator’s Manual
Expiration date is not valid.
OR
Invalid expiration date.
The expiration date is not in a valid format.
The proper format for the expiration date for
a cartridge or control lot is yy-mm-dd.
Test will overwrite saved test history.The current test will overwrite test results in
the per
manent r
ecord. Transmit all unsent
test records to the output source.
Invalid password.The password that was entered is not
Run ACTtrac
ct. Enter the correct p
®
from QC Menu.The user is attempting to run an ACTtrac®
corre
test from the Main Menu. Go to the QC Menu
screen to run ACTtrac
®
tests.
assword.
No cartridge lot number active.The user is attempting to run a test without
an active cartrid
ge lot number. Enter the
cartridge lot number for this test type.
Cartridge not recognized.
OR
All cartridges have been disabled. Have the
QC Manager
enable the app
ropriate tests.
There are no active cartridges to edit.
OR
There are no active cartridges.
No control lot number active.The user is attempting to run a control test
without an ac
tive control lot number. Enter a
valid lot number for the control.
There are no other active cartridges.Only the selected test is enabled. Have the
QC Manager
enable test
s.
No test results available.No test results are available to view on the
ew Current Test Results screen.
Vi
Patient ID required.Patient ID is required to run a patient test.
Enter a Patient
ID in the Enter ID screen.
Preventive maintenance due.Preventive maintenance is due. Call
onic Instrument Service to schedule a
Medtr
preventive maintenance check.
Value Out of Range. Enter 35 – 39.The Actuator Heat Block temperature cannot
be adjusted out of
the 35-to-
39 range. If the
Actuator Heat Block cannot be adjusted
within the 35-to-39 range, call Medtronic
Instrument Service.
There are no intervals set.The QC Manager has not set the intervals
liquid or electronic controls. Have the QC
for
Manager set QC intervals in the QC
Manager screen.
QC Due: HH:MM.The QC for this cartridge will run out at the
s
cribed time (hh = hour; mm = minute).
pre
QC not performed.The QC test for this cartridge has not been
performed or
has failed. Perform QC tests
for this cartridge.
ACT Plus® Operator’s Manual English 47
The passwords do not match.When changing the password for the QC
Manager, the Confirm Password screen did
not match the Change Password screen.
Reenter the same password in the Confirm
Password screen that was entered in the
Change Password screen.
QC past due.The QC for this cartridge has passed its
erform QC tests for this cartridge.
P
interval.
Test terminated before clot detect.The test stopped before clots were detected
in both channels. Repeat test.
No User ID
s to download.The QC Manager is trying to download User
IDs to the floppy disk or USB storage device,
User IDs are in the memory. Add User
but no
IDs to the memory and download to the
floppy drive or USB storage device.
No valid User IDs in file.The QC Manager is trying to upload User
the flo
IDs from
device, but there are no User IDs in the file.
Download User IDs from an ACT Plus
instrument or an ACT Plus
ppy disk or USB storage
®
External Data
®
Manager (EDM) to the floppy disk or USB
storage device, and use that floppy disk or
USB storage device to upload them.
Test requires valid User ID.The QC Manager has set User Lockout to
ID is r
ON so that a User
equired to run a
patient test. Enter a valid User ID in the
Enter ID screen.
Alerts
Bar code scan timeout.A bar code input sequence was detected,
> 10% spread error.The channels of the controls tested are not
> 12% spread error.The channels of the controls tested are not
Error - File not found.The ACT Plus
t it did not complete corr
bu
ectly. Rescan the
bar code.
thin 10%. Repeat the control test.
wi
i
thin 12%. Repeat the control test.
w
®
instrument is trying to read a
file on a floppy disk or USB storage device,
but cannot find it.
48 English ACT Plus® Operator’s Manual
Error Messages
Disk error - Disk is faulty.The ACT Plus® instrument is trying to read a
Serial port error.An error occurred in accessing the serial
Motor stall. Power OFF then ON (E9).The motor has been inhibited. To reset the
System error. Power OFF then ON.The system has become unstable. To reset
Bad Reference in Ch1 and Ch2.
OR
Bad Reference in Ch1.
OR
Bad Reference in Ch2.
Error: E1No cartridge is detected. Place a car
Error: E2 or E3A software error has occurred. Call
No flags.
OR
No flag in Ch1.
OR
No flag in Ch2.
OR
No flags in Ch1 or Ch2.
Insufficient data.The channels of the control timed out at 999.
Error: Invalid reply.Internal communication error. Call Medtronic
Light path failure in Ch1 and Ch2 (E4).
OR
Light path failure in Ch1 (E4).
OR
Light path failure in Ch2 (E4).
Motor stall.The motor is inhibited. Press STOP to reset.
tor stall, cycle power (E8).The motor is inhibited. To reset, power OFF
Mo
No reply.Internal communication error. Call Medtronic
floppy disk or USB storage device and
cannot read it. Insert a floppy disk or USB
storage device that meets the specification
listed in Section 11, Specifications.
port. Retry transfer
system, turn th
the system, tur
The ACT Plus
baseline drop time for Channel 1, Channel
and both channels. Repeat test.
the actuator assembly and retest.
t
ronic Instrument Service.
Med
The ACT Plus
flag in Channel 1, Channel 2, or both
channels. Repeat the test.
Repeat the
ument Ser
Instr
The flag sensor has failed. Clean the
or assembly cover per the instructions
t
actua
in the Maintenance section of this manual. If
the error is not corrected, call Medtronic
Instrument Service.
Repeat the
then ON.
ument Ser
Instr
.
e power OFF then ON.
n the power
®
instrument could not get a
®
instrument did not sense a
contr
ol.
vice.
test.
vice.
OFF then ON.
t
ridge in
2,
ACT Plus® Operator’s Manual English 49
Notes: If a System Alarm occurs that cannot be resolved by cycling the power OFF
then ON, record the 3-digit Error Code (if displayed), and contact Medtronic
Instrument Service for assistance.
Tempera t u re
Heat Block Temp XX.X°C.The current actuator heat block temperature
Heat Block Temp > 44°C (E99).The actuator heat block temperature is
Heat Block Temp < 5°C (E00).The actuator heat block temperature is less
(XX.X°C) is displayed when it is between
20°C and 41°C.
r
than 44°C.
greate
C.
than 5°
50 English ACT Plus® Operator’s Manual
Section 11: Specifications
Instrument
Number of Channels:
Timing Range (Seconds):
Actuator Heat Block Temperature:
Range:
Resolution:
Accuracy:
Electrical
Note: Technical data, features, and options referenced in this manual are based on
the latest information available at the time of printing. Medtronic reserves the right
to change specifications without notice.
Classification (IEC 61010):
2
6 - 999
37.0°C ± 2°C
0.1°C
± 0.5°C
Class I, IPX0, Continuous operation
Power:
Voltage:
Frequency:
Maximum current:
Fuse:
Power cord:
Length:
Type:
1.0 A (100 - 240)
T2A 250 V
2 wires plus a ground (earth) connector
3.1 m (10 ft.)
3-prong hospital grade (USA only)
IEC 61010-1: 2001
CSA C22.2 NO. 61010-1
IEC 60601-1-2: 2007
33 cm (13.0 in)
20 cm (8.0 in)
27 cm (11.0 in)
5.22 kg (11.5 lb)
14°C to 32°C (57°F to 90°F)
10% to 90%, noncondensing
0°C to 49°C (32°F to 120°F)
5% to 90%, noncondensing
Indoor
≤ 2000 meters
Not to exceed ± 10% of nominal voltage
Category II
2
ACT Plus® Operator’s Manual English 51
Data Ports:
Serial data port:
USB port:
Bar code scanner port:
Floppy drive:
19200 baud, 8 data bits, 1 stop bit, no parity
Used with Medtronic approved USB storage device,
M930703A001.
Used with the optional Bar Code Scanner LS1900 or
LS 2208.
PC compatible, 1.44 MB, 3.5-inch floppy disk.
Default Operating Parameters
The ACT Plus® instrument is shipped with the following default parameter settings:
Patient Parameters
Patient ID
User ID
Cartridge Lot Number
Cartridge Exp Date
Control Lot Number
Control Exp Date
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Group 1The ACT Plus
Class AThe ACT Plus
Class A
Complies
®
should assure that it is used in such an
®
function. Therefore, its RF emissions are very low and
are not likely to cause any interference in nearby
electronic equipment.
other than domestic and those directly connected to the
public low-voltage power supply network that supplies
buildings used for domestic purposes.
uses RF energy only for its internal
®
is suitable for use in all establishments
ACT Plus® Operator’s Manual English 53
IEC 60601-1-2 Table 202,
Guidance and manufacturer's declaration - electromagnetic immunity
The ACT Plus® is intended for use in the electromagnetic environment specified below.
The customer or the user of the ACT Plus
environment.
Immunity test IEC 60601
test level
Electrostatic
discharge
±6 kV contact
±8 kV air
(ESD)
61000-4-2
IEC
Electrical fast
ansient/bur
tr
IEC 61000-4-4
±2 kV for power
st
supply lines
±1 kV for
input/output lines
Surge
IEC 61000-4-5
±1 kV line(s) to
line(s)
±2 kV line(s) to earth
Voltage dips,
short
uptions
interr
and volt
age
variations on
power supply
input lines
IEC
-4-1
61000
<5% U
(>95% dip in UT) for
0.5 cycle
40% U
(60% dip in UT) for 5
cycles
70% U
(30% dip in UT) for
1
25 cycles
<5% U
(>95% dip in UT) for
T
T
T
T
5 seconds
Power
equency
fr
(5
0/60 Hz)
3 A/m3 A/mPower frequency magnetic
magnetic field
IEC 61000-4-8
Note: U
is the a.c. mains voltage prior to application of the test level.
T
®
should assure that it is used in such an
Compliance levelElectromagnetic
ment - guidance
environ
±6 kV contact
±8 kV air
Floors should be wood,
concrete or ceramic tile. If
e covered with
floors ar
synthetic material, the
relative humidity should be
at least 30%.
r power
±2 kV fo
supply lines
±1 kV for input/output
lines
±1 kV differential
mode
±2 kV common mode
Mains power quality should
that of a typical
be
commercial o
r hospital
environment.
If the user of the ACT Plus
requires continued operation
during power mains
interruptions, it is
recommended that the ACT
®
be powered from an
Plus
uninterruptible power supply
or a battery.
<5% U
T
(>95% dip in UT) for
0.5 cycle
40% U
T
(60% dip in UT) for 5
cycles
70% U
T
(30% dip in UT) for 25
cycles
<5% U
T
(>95% dip in UT) for 5
seconds
Mains power quality should
be that of a typical
commercial o
r hospital
environment.
If the user of the ACT Plus
requires continued operation
during power mains
interruptions, it is
recommended that the ACT
®
be powered from an
Plus
uninterruptible power supply
or a battery.
fields should be at levels
cteristic of a typical
chara
location in a typical
commercial or hospital
environment.
®
®
54 English ACT Plus® Operator’s Manual
IEC 60601-1-2: 2001 Table 204
Guidance and manufacturer's declaration - electromagnetic immunity
The ACT Plus® is intended for use in the electromagnetic environment specified below.
The customer or the user of the ACT Plus
environment.
Immunity testIEC 60601
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
evel
test l
3 Vrms
150 kHz to 80
z
MH
3 V/m
80 MHz to 2.5
GHz
®
should assure that it is used in such an
Compliance
Level
rmsd = 1.2
3 V
3 V/md = 1.2
Electromagnetic environment guidance
Portable and mobile RF
communications equipment should be
used no closer
®
, including cables, than the
Plus
recommended separation distance
calculated from the equation applicable
to the frequency of the transmitter.
Recommended separation distance
√P
√P 80 MHz to 800 MHz
√P 800 MHz to 2.5 GHz
d = 2.3
where P is the maximum output power
rating of the transmitter in watt
according to the transmitter
manufacturer and d is the
recommended separation distance in
meters (m).
Field strengths from fixed RF
smitters, as deter
tran
electromagnetic site survey,
less than the compliance level in each
frequency r
Interference may occur in the vicinity of
equipment marked with the following
symbol:
ange.
to any part of the ACT
s (W)
mined by an
a
should be
b
Note: At 80 MHz and 800 MHz, the higher frequency range applies.
Note: These guidelines may not apply in all situations. Electromagnetic propagation
is af
ed by absorption and reflection from structures, objects and people.
fect
ACT Plus® Operator’s Manual English 55
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the ACT Plus
exceeds the applicable RF compliance level above, the ACT Plus
verify normal operation. If abnormal performance is observed, additional measures may be
necessary, such as re-orienting or relocating the ACT Plus
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
®
should be observed to
®
.
®
is used
Recommended separation distances between portable and mobile RF
IEC 60601-1-2: 2001 Table 206
communications equipment and the ACT Plus®
The ACT Plus® is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the ACT Plus
electromagnetic interference by maintaining a minimum distance between portable and
mobile RF communications equipment (transmitters) and the ACT Plus
below, according to the maximum output power of the communications equipment.
Rated
imum
max
output power
of transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80
MHz
√P
d = 1.2
80 MHz to 800 MHz
√P
d = 1.2
0.010.120.120.23
0.10.380.380.73
11.21.22.3
103.83.87.3
100121223
For transmitters rated at a maximum output power not listed above, the recommended
aration dist
sep
ance d in meters (m) can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
Note: At 80 MHz and 800 MHz, the higher frequency range applies.
Note: These guidelines may not apply in all situations. Electromagnetic propagation
ed by absorption and reflection from structures, objects and people.
fect
is af
Cautions:
• Do not use non-Medtronic components with Medtronic in-line-pow
The use of non-Medtronic components may result in damage to Medtronic components,
increased emissions, or decreased electromagnetic immunity of the Medtronic devices
or systems.
• Do not use Medtronic in-line-powered external devices a
djacent to, o
other electronic devices. Using Medtronic devices in these configurations may result in
decreased electromagnetic immunity of the Medtronic devices or systems.
®
can help prevent
®
as recommended
800 MHz to 2.5 GHz
d = 2.3√P
ered external devices.
r stacked with,
56 English ACT Plus® Operator’s Manual
Appendix A: Optional Bar Code Scanner LS1900
1
2
3
4
Introduction
The Symbol®1 LS1900 Series scanner is compatible with the ACT Plus®
instrument. The scanner will read the following barcode formats:
■
Code 39
■
Code 39 With Check Digit (Mod 10)
■
Code 39 Full ASCII
■
Interleave 2 of 5
■
Interleave 2 of 5 with Check Digit (Mod 49)
■
Codabar
■
Code128
Parts of the LS1900 Series Scanner
Figure 8. Bar Code Scanner
1. Light Emitting Diode (LED)
2. Beeper
3. Trigger
4. Exit Window
1
Symbol® is a registered trademark of Symbol Technologies, Inc.
ACT Plus® Operator’s Manual English 57
Scanning in Hand-Held Mode
1. Connect the scanner cable to the ACT Plus® instrument. The scanner will
beep when power is on.
2. Aim the scanner at the bar code and press the trigger.
3. Be sure that the red line from the scann
Figure 9. Correct scan location
When the scanner has successfully decoded the bar code, the scanner will beep,
and the information will be entered into the ACT Plus
goes across the entire bar code.
er
®
instrument.
Setting Bar Code Format Options
The scanner is preset to read UIDs and PIDs in format Code 39 With Check Digit
(Mod 10), Interleave 2 of 5 with Check Digit (Mod 49), Codabar and Code 128. It is
also preset to read the cartridge/ control information.
To set additional barcode format options:
Code 39
Scan the “Enable Code 39 Check Digit” barcode if the UID/PID barcodes use Code
39 check digit to verify a correct scan.
Figure 10. Enable Code 39 Check Digit
Scan the “Disable Code 39 Check Digit” barcode if the UID/PID barcodes do not
use Code 39 check digit to verify a correct scan. The default setting is Code 39
Disabled, so this b
in the previous step.
Note: Disabling the Code 39 Check Digit will allow the scanner to scan a barcode
that
has the check digit and add it to the data.
58 English ACT Plus® Operator’s Manual
arcode should only be used if Code 39 was mistakenly enabled
Figure 11. Disable Code 39 Check Digit
Scan the "Code 39 Full ASCII Conversion" bar code to allow for use of Code 39 full
ASCII barcodes.
Figure 12. Enable Code 39 Full ASCII
Interleave 2 of 5
Scan the "Enable Interleave 2 of 5 Check Digit" barcode if the UID/PID barcodes
use Interleave 2 of 5 check digit to verify a correct scan.
Figure 13. USS Check Digit
Scan the "Interleave 2 of 5 Any Length" bar code for use of Interleave 2 of 5
barcodes with any lengths.
Figure 14. 2 of 5 - Any Length
Scan the "Disable Interleave 2 of 5 Check Digit" barcode if the UID/PID barcodes
do not use Interleave 2 of 5 check digit to verify a correct scan. The default setting
is Interleave 2 of 5 Disabled so this barcode should only be used if Interleave 2 of
5 was mistakenly enabled in the previous step.
Note: Disabling the Interleave 2 of 5 Check Digit will allow the scanner to scan a
barcode that has the check digit and add it to the data.
Figure 15. Disable
ACT Plus® Operator’s Manual English 59
Codabar
Scan the "Codabar" bar code to allow the use of Codabar for scanning User and
Patient IDs.
Figure 16. Enable Codabar
Scan the "NOTIS Editing Enable" barcode if the UID/PID barcodes use Codabar
barcode format. The NOTIS Editing Enable barcode will remove the Start/Stop
characters imbedded in the Codabar barcodes.
Figure 17. Enable NOTIS Editing
Troubleshooting
Nothing happens when you follow the
operating instructions.
The laser comes on, but the symbol does not
decode.
The symbol is decoded, but not transmitted
to the host.
Interface/power cables are loose. Check for
loose cable connections.
The scanner is not programmed for the
correct bar code type. Be sure
is programmed to read the type of bar code
you are scanning.
The bar code symbol is unreadable. Check
e symbol to make su
th
Try scanning test symbols of the same bar
code type.
The distance between the scanner and bar
code is incorr
to the bar code.
T
he scanner is not programmed for the
correct bar code type. Be sure the scanner
is programmed to read the type of bar code
being scanned.
ect. Move
the scanner
re it is not defaced.
the scanner closer
Defaults
If the bar code scanner does not read the bar code on the ACT Plus® disposable
packaging, reset the defaults before calling Medtronic Instrument Service. To reset
the defaults, follow this procedure, and try to rescan the packaging.
60 English ACT Plus® Operator’s Manual
Procedure for setting the defaults:
1. To ensure that the scanner is reset, first scan the Set All Defaults bar code.
Figure 18. Set All Defaults bar code
2. Scan the “Low Volume” bar code, unless a different volume is desired.
Figure 19. Low Volume bar code
3. Scan the "Transmit Code ID" barcode to allow the barcode scanner to
transmit the ID of Barcode format along with the data of the barcode that was
just scanned.
Figure 20. Transmit Code ID
4. Scan the “Enable UCC/EAN-128” bar code. The cartridges and controls use
this format.
Figure 21. Enable UCC/EAN-128 bar code
5. Scan the “Scan Prefix” bar code, then scan the following numbers: 1, 0, 0, 2.
Figure 22. Scan Prefix bar code
Figure 23. 1 bar code
ACT Plus® Operator’s Manual English 61
Figure 24. 0 bar code
Figure 25. 0 bar code
Figure 26. 2 bar code
6. Scan the “Scan Suffix” bar code, then scan the following numbers: 1, 0, 0, 3.
Figure 27. Scan Suffix bar code
Figure 28. 1 bar code
Figure 29. 0 bar code
Figure 30. 0 bar code
Figure 31. 3 bar code
62 English ACT Plus® Operator’s Manual
7. Scan the “Scan Options” bar code, then scan the "Prefix Data Suffix" bar
code.
Figure 32. Scan Options bar code
Figure 33. Prefix Data Suffix bar code
8. Scan the “Enter” bar code. This completes the procedure. The bar code
scanner will now work with the ACT Plus
Figure 34. Enter bar code
®
instrument.
ACT Plus® Operator’s Manual English 63
64 English ACT Plus® Operator’s Manual
Appendix B: Optional Bar Code Scanner LS2208
1
2
3
4
1. Light Emitting Diode (LED)
2. Beeper
3. Trigger
4. Exit Window
Introduction
The Symbol®1 LS2208 scanner is compatible with the ACT Plus® instrument. The
scanner will read the following bar code formats:
■
Code 39
■
Code 39 with Check Digit (Mod 10)
■
Code 39 Full ASCII
■
Interleave 2 of 5
■
Interleave 2 of 5 with Check Digit (Mod 49)
■
Codabar
■
Code128
Parts of the LS2208 Scanner
1
Symbol® is a registered trademark of Symbol Technologies, Inc.
Figure 35.
ACT Plus® Operator’s Manual English 65
Scanning in Hand-Held Mode
1. Connect the scanner cable to the bar code scanner and the ACT Plus®
instrument. The scanner will beep when power is on.
2. Aim the scanner at the bar code and press the trigger.
3. Be sure that the red line from the scann
Figure 36. Correct scan location
When the scanner has successfully decoded the bar code, the scanner will beep,
and the information will be entered into the ACT Plus
goes across the entire bar code.
er
®
instrument.
Setting Bar Code Format Options
1. Scan the following bar codes in sequence. The scanner will beep when it
successfully reads a setup bar code. If you make an error, start over.
2. Scan the "Reset Default Values" bar code (Figure 37).
Figure 37. Reset Default Values bar code
3. Scan the "IBM PC/AT & IBM PC compatibles" bar code (Figure 38) to set
communication to a PC.
Figure 38. IBM PC/AT and IBM PC Compatibles bar code
4. Scan the "Low Volume" bar code (Figure 39) to set the volume mode to Low.
Figure 39. Low Volume bar code
5. Scan the "Interleave 2 of 5 to Any Length" bar code (Figure 40).
66 English ACT Plus® Operator’s Manual
Figure 40. 2 of 5 Any Length bar code
6. Scan the “Enable Codabar” bar code (Figure 41).
Figure 41. Enable Codabar bar code
7. Scan the "Enable NOTIS Editing" bar code to enable Notis editing for
Codabar format (Figure 42).
Figure 42. Enable Notis Editing bar code
8. Scan the "Transmit code ID" bar code to enable the Transmit Code ID function
(Figure 43).
Figure 43. Transmit Code ID bar code
9. To format UCC/EAN-128 for reading cartridges/controls, scan the "Scan
Prefix" bar code, then scan the following numbers: 1, 0, 0, 2.
Figure 44. Scan Prefix bar code
Figure 45. 1 bar code
ACT Plus® Operator’s Manual English 67
Figure 46. 0 bar code
Figure 47. 0 bar code
Figure 48. 2 bar code
10. To format UCC/EAN-128 for reading cartridges/controls, scan the "Scan
Suffix" bar code, then scan the following numbers: 1, 0, 0, 3.
Figure 49. Scan Suffix bar code
Figure 50. 1 bar code
Figure 51. 0 bar code
Figure 52. 0 bar code
68 English ACT Plus® Operator’s Manual
Figure 53. 3 bar code
11. To format UCC/EAN-128 for reading cartridges/controls, scan the "Scan
Options" and "Prefix Data Suffix" bar codes.
Figure 54. Scan Options bar code
Figure 55. Prefix Data Suffix bar code
12. Scan the "Enter" bar code to accept the entered values.
Figure 56. Enter bar code
ACT Plus® Operator’s Manual English 69
Bar Code Scanner Installation Instructions
1. Remove the 2 Phillips screws using a #2 screwdriver.
2. Install the bracket as shown in Figure 57 and secure with the 2 screws
removed in Step 1.
Figure 57.
3. Plug the interface modular connector into the cable interface port on the
bottom of the scanner handle.
4. Connect the other end of the cable to
Plus (Figure 58).
connector on the back of the ACT
the
Figure 58.
70 English ACT Plus® Operator’s Manual
Appendix C: Warranties
EQUIPMENT LIMITED WARRANTY
The following LIMITED WARRANTY Applies to U
1
nited States Customers
Only:
A. This LIMITED WARRANTY provides the follo
wing assurance to the
purchaser of the Medtronic® Automated Coagulation Timer Plus®, hereafter
referred to as the “Equipment”:
(1) Should the Equipment fail to function within normal tolerances due to a
n materials or workmanship within a period of one (1) year,
defect i
commencing with the delivery of the Equipment to the purchaser,
Medtronic will at its option: (a) repair or replace any part or parts of the
Equipment; (b) issue a credit to the purchaser equal to the Purchase
Price, as defined in Subsection A(2), against the purchase of the
replacement Equipment; or (c) provide a functionally comparable
replacement Equipment at no charge.
(2) As used herein, Purchase Price shall mean the lesser of the net invoiced
price of th
e
original, or current functionally comparable, or replacement
Equipment.
B. To qualify for the repair, replacement or credit set forth in Section A, the
follo
wing conditions must be met:
(1) The Equipment must be returned to Medtronic within thirty (30) days after
discovery of th
e defect (Medtronic may, at its option, repair the Equipment
on the Customer’s site).
(2) The Equipment must not have been repaired or altered either (i) outside
of Med
tronic's factory or (ii) by any person not authorized by Medtronic to
repair the Equipment in any way which, in the judgment of Medtronic,
affects its stability and reliability. The Equipment must not have been
subjected to misuse, abuse or accident.
C. This LIMITED WARRANTY is limited to its express terms. In particular:
(1) Except as expressly provided by this LIMITED WARRANTY,
MED
RONIC IS NOT RESPONSIBLE FOR ANY INDIRECT,
T
INCIDENTAL OR CONSEQUENTIAL DAMAGES BASED ON ANY
DEFECT, FAILURE OR MALFUNCTION OF THE Equipment, WHETHER
THE CLAIM IS BASED ON WARRANTY, CONTRACT, TORT OR
OTHERWISE.
(2) This LIMITED WARRANTY is made o
n
ly to the purchaser of the
Equipment. EXCEPT AS SET FORTH IN THIS LIMITED WARRANTY,
MEDTRONIC MAKES NO WARRANTY, EXPRESS OR IMPLIED,
INCLUDING, BUT NOT LIMITED TO, ANY IMPLIED WARRANTY OF
MERCHANTABILITY, OR FITNESS FOR A PARTICULAR PURPOSE
WHETHER ARISING FROM STATUTE, COMMON LAW, CUSTOM OR
1
This LIMITED WARRANTY is provided by Medtronic, Inc., 710 Medtronic Parkway, Minneapolis, MN
55432. It applies only in the United States. Areas outside the United States should contact their local
Medtronic representative for exact terms of the LIMITED WARRANTY.
ACT Plus® Operator’s Manual English 71
OTHERWISE. NO EXPRESS OR IMPLIED WARRANTY SHALL
EXTEND BEYOND THE PERIOD SPECIFIED IN A(1) ABOVE. THIS
LIMITED WARRANTY SHALL BE THE EXCLUSIVE REMEDY
AVAILABLE TO THE CUSTOMER.
(3) The exclusions and limitations set out above are not intended to, and
shou
ld not be construed
so as to, contravene mandatory provisions of
applicable law. If any part or term of this LIMITED WARRANTY is held to
be illegal, unenforceable or in conflict with applicable law by a court of
competent jurisdiction, the validity of the remaining portions of the
LIMITED WARRANTY shall not be affected, and all rights and obligations
shall be construed and enforced as if this LIMITED WARRANTY did not
contain the particular part or term held to be invalid. This LIMITED
WARRANTY gives the purchaser specific legal rights. The purchaser may
also have other rights which vary from state to state.
(4) No person has any authority to bind Medtronic to any representation,
cond
iti
on or warranty with respect to the Equipment other than as set forth
in this LIMITED WARRANTY.
EQUIPMENT LIMITED WARRANTY
1
(FOR REPAIRABLE EXTERNAL EQUIPMENT/OUTSIDE THE U.S.)
e following Limited Warranty Applies to Customers Outside the United
Th
ates:
St
A. This LIMITED WARRANTY provides the fo
wing assurance to the
llo
purchaser of the Medtronic® Automated Coagulation Timer Plus®, hereafter
referred to as the “Equipment,” that should the Equipment fail to function
within normal tolerances due to a defect in materials or workmanship within a
period of one (1) year, commencing with the delivery of the Equipment to the
purchaser, Medtronic will at its option: (a) repair or replace any defective part
or parts of the Equipment; (b) issue a credit equal to the original Equipment
purchase price (but not to exceed the value of the replacement Equipment),
against the purchase of replacement Equipment; or (c) provide functionally
comparable replacement Equipment at no charge.
B. To qualify for this repair, replacement or credit, the
following conditions must
be met:
(1) The Equipment must be returned to Medtronic within sixty (60) days after
discovery of th
e defect (Medtronic may, at its option, repair the Equipment
on site).
(2) The Equipment must not have been repa
ired or altered by someone other
than Medtronic in any way which, in the judgment of Medtronic, affects its
stability and reliability.
(3) The Equipment must not have been subjected to misuse, abuse or
accide
nt.
1
This LIMITED WARRANTY is provided by Medtronic, Inc., 710 Medtronic Parkway, Minneapolis, MN
55432. Areas outside the United States should contact their local Medtronic representative for exact
terms of the LIMITED WARRANTY.
72 English ACT Plus® Operator’s Manual
C. This LIMITED WARRANTY is limited to its express terms. In particular,
Medtronic is not responsible for any incidental or consequential damages
based on any use, defect or failure of the Equipment, whether the claim is
based on warranty, contract, tort or otherwise.
D. The exclusions and limitations set out above are not intended to, and should
not be construed so as to,
If any part or term of this LIMITED WARRANTY is held by any court of
competent jurisdiction to be illegal, unenforceable or in conflict with applicable
law, the validity of the remaining portion of the LIMITED WARRANTY shall not
be affected, and all rights and obligations shall be construed and enforced as
if this LIMITED WARRANTY did not contain the particular part or term held to
be invalid.
contravene mandatory provisions of applicable law.
ACT Plus® Operator’s Manual English 73
74 English ACT Plus® Operator’s Manual
Europe
Europe/Africa/Middle East Headquarters
Medtronic International Trading Sàrl
Route du Molliau 31
Case Postale 84
CH - 1131 Tolochenaz
Switzerland
Internet: www.medtronic.co.uk
Tel. 41-21-802-7000
Fax 41-21-802-7900
Authorized Representative in the
European Community
Medtronic Japan
Comodio Shiodome 5F
2-14-1 Higashi-Shimbashi, Minato-ku
Tokyo 105-0021
Japan
Tel. 81-3-6430-2011
Fax 81-3-6430-7140
Australia
Medtronic Australasia Pty. Ltd.
97 Waterloo Road
North Ryde NSW 2113
Australia
Tel. 61-2-9857-9000
Fax 61-2-9878-5100
Asia
Medtronic International Ltd.
Suite 1602 16/F, Manulife Plaza
The Lee Gardens, 33 Hysan Avenue
Causeway Bay
Hong Kong
Tel. 852-2891-4068
Fax 852-2591-0313
Americas
Latin America
Medtronic Latin America
3750 NW 87th Avenue
Suite 700
Miami, FL 33178
USA
Tel. 305-500-9328
Fax 786-709-4244
Canada
Medtronic of Canada Ltd.
99 Hereford Street
Brampton, Ontario L6Y 0R3
Canada
Tel. 905-460-3800
Fax 905-826-6620
Toll-free:
1-800-268-5346
United States
Manufacturer:
Medtronic, Inc.
710 Medtronic Parkway
Minneapolis, MN 55432
USA
Internet: www.medtronic.com
Tel. 763-514-4000
Fax 763-391-9100
Toll-free:
1-800-328-2518
(24-hour consultation service)
Medtronic Perfusion Systems or
Manufacturing Facility:
7611 Northland Drive
Minneapolis, MN 55428
Tel. 763-391-9000
Fax 763-391-9100
Customer Service and Product Orders
Toll-free: 1-800-854-3570
Visit Medtronic at
www.perfusionsystems.com