Changing the Quickset® Infusion Set Using a Revel™ Insulin Pump
plunger rod
transfer guard
reservoir
To Rewind the Piston in Pump:
1. Wash your hands. Remove old infusion set from your body and the reservoir from pump.
2. Go to Reservoir + Set in MAIN MENU, then select Reservoir Setup to open REWIND screen. Press
To Fill the Reservoir:
1.
Remove reservoir
from package.
Pull plunger rod down to
2. fill reservoir with air
3. After wiping top of insulin vial with alcohol, hold blue transfer guard and press it down onto insulin vial.
4. Use thumb to push air from reservoir into vial. Continue to hold plunger rod down.
5. Continue to hold plunger rod down as you turn vial over (vial should now be on top and reservoir on bottom). Release plunger. Then slowly pull plunger rod down to fill reservoir with insulin.
6. Check reservoir for air bubbles. Tap side of reservoir to force bubbles to rise to top of reservoir. Push plunger rod up to move air bubbles into insulin vial.
7. After air bubbles are removed, pull plunger rod back down to fill reservoir with enough insulin to last 2-3 days.
8. Flip vial over so it is now on bottom. Hold blue transfer guard with one hand. With other hand turn reservoir counter- clockwise and pull it straight up until disconnected from blue transfer guard. Disconnect blue transfer guard and dispose into a sharps container.
ACT
Changing the Quickset® Infusion Set Using a Revel™ Insulin Pump
connector
To Connect the Tubing:
White Button
Quickserter®
Blue
Button
White Button
Tubing slot
Reservoir connector
Needle guard
Blue hub
Remove
1. Quickset from package. Hold reservoir in one hand. With other hand gently push connector onto top of reservoir and turn clockwise until it slides and locks into place.
Tap side of reservoir to
2. force any remaining air bubbles to rise to top. Push up on plunger rod until bubbles are out and some insulin is in tubing.
Turn plunger rod
3. counter-clockwise until it unscrews from reservoir. Be careful not to pull out plunger rod from reservoir or insulin will spill out.
Quickset®
To Fill the Tubing:
1. Press
ACT
so that REWIND COMPLETE appears. Insert newly filled reservoir into pump’s reservoir compartment.
2. Turn reservoir clockwise until it locks into place.
3. Press
ACT
to go to FILL TUBING screen. Be sure infusion set is NOT connected to your body. Then select Yes and press
ACT
.
ACT
4. Press and hold to fill tubing. The pump will beep as tubing fills. Release
ACT
when you see drops at tip of connector needle. Then press
ESC
once.
5. If you see drops at end of Quickset® needle and there are no air bubbles in the tubing, select Yes and press
ACT
.
6. When the FILL CANNULA screen appears, you are ready to insert the Quickset® infusion set.