Changing the Quickset® Infusion Set Using a Revel™ Insulin Pump
plunger
rod
transfer
guard
reservoir
To Rewind the Piston in Pump:
1. Wash your hands. Remove old infusion set from your body and the
reservoir from pump.
2. Go to Reservoir + Set in MAIN MENU, then select Reservoir Setup to
open REWIND screen. Press
To Fill the Reservoir:
1.
Remove reservoir
from package.
Pull plunger rod down to
2.
fill reservoir with air
3. After wiping top
of insulin vial
with alcohol,
hold blue
transfer guard
and press it
down onto
insulin vial.
4. Use thumb to
push air from
reservoir into
vial. Continue to
hold plunger rod
down.
5. Continue to hold
plunger rod
down as you
turn vial over
(vial should
now be on top
and reservoir on
bottom). Release
plunger. Then
slowly pull
plunger rod down
to fill reservoir with
insulin.
6. Check reservoir
for air bubbles.
Tap side of
reservoir to force
bubbles to rise
to top of
reservoir. Push
plunger rod
up to move
air bubbles into
insulin vial.
7. After air bubbles are
removed, pull plunger
rod back down to fill
reservoir with enough
insulin to last 2-3 days.
8. Flip vial over so it is now
on bottom. Hold blue
transfer guard with one
hand. With other hand
turn reservoir counter-
clockwise and pull it
straight up until
disconnected from
blue transfer guard.
Disconnect blue transfer
guard and dispose into a
sharps container.
ACT
to rewind.
![](/html/1c/1c92/1c9228a5a86abd30ee6420f4477e2ebb5a57b2db7c72c1da81edd2cf30b84b8a/bg2.png)
Changing the Quickset® Infusion Set Using a Revel™ Insulin Pump
To Connect the Tubing:
White
Button
Quickserter®
Blue
Button
White
Button
Tubing slot
Reservoir connector
Needle guard
Blue hub
Remove
1.
Quickset from
package. Hold
reservoir in one
hand. With
other hand
gently push
connector onto
top of reservoir
and turn
clockwise until it
slides and locks
into place.
Tap side of reservoir to
2.
force any remaining air
bubbles to rise to top.
Push up on plunger rod
until bubbles are out
and some insulin is in
tubing.
Turn plunger rod
3.
counter-clockwise until it
unscrews from reservoir.
Be careful not to pull out
plunger rod from
reservoir or insulin will
spill out.
Quickset®
To Fill the Tubing:
1. Press
ACT
so that
REWIND COMPLETE
appears. Insert newly
filled reservoir into
pump’s reservoir
compartment.
2. Turn reservoir
clockwise until it
locks into place.
3. Press
ACT
to go to
FILL TUBING
screen. Be sure
infusion set is NOT
connected to your
body. Then select
Yes and press
ACT
.
ACT
4. Press and hold
to fill tubing. The
pump will beep
as tubing fills.
Release
ACT
when
you see drops at tip
of connector needle.
Then press
ESC
once.
5. If you see drops at end of
Quickset® needle and
there are no air bubbles
in the tubing, select Yes
and press
ACT
.
6. When the FILL
CANNULA screen
appears, you are
ready to insert
the Quickset®
infusion set.