D121255_EN Rev F PTeye-1 Instructions for Use Insert P a g e | 2 of 4
4. NO modifications of this equipment is allowed.
5. Do not reuse the probe assembly. Discard probe assembly after each surgery.
6. Do not use the device if the environmental conditions do not meet the environmental requirements of the
device
Precautions
1. Do not use the device with incompatible equipment or peripherals that are not authorized by AiBiomed.
Doing so may void existing certifications and/or warranties.
2. Prior to each use, visually inspect the PTeye system, including console, probe assembly, external power
supply, power cord, and foot pedal to ensure there are no visible cuts, cracks, or other damage. If any damage
is found, do not use the system; contact your AiBiomed customer service representative for assistance.
3. Do not excessively bend or kink the flexible portion of the probe assembly.
4. Store the device and all peripherals in a protective container to prevent the possibility of damage when not in
use. Do not store the device in a location where it will be exposed to temperatures exceeding 140°F (+ 60°C).
5. Handle all equipment carefully. If the fiber optic probe assembly is dropped, it can be damaged, which
renders the system unusable.
6. This device should only be used in compliance with its intended use.
7. Due to a small sample size, limited clinical data is available regarding the safety and effectiveness of the
PTeye System for rare disease states such as tertiary hyperparathyroidism, concomitant thyroid-parathyroid
diseases, malignant parathyroid diseases, or other circumstances when prophylactic thyroidectomies are
performed in individuals at high-risk for certain diseases such as MEN2A.
STERILIZATION
The PTeye-1 Probe is provided sterile. Do not re-sterilize Compromised performance and cleaning hazards are
imminent.
STORAGE AND HANDLING
Store at room temperature. Avoid storing the PTeye-1 Probe at conditions of excessive heat or humidity. Do not use
after expiration date on label. All components should be handled carefully to avoid damaging the device. Inspect all
parts for damage prior to use, do not use any damaged parts. Do not reuse device.
PROCEDURE
Verifying Probe Assembly Function
The probe assembly contains delicate fiber optic cables. To verify that these have not been damaged during the
shipment process, follow these steps.
1. Inspect the probe sterile package for any signs of damage. If any damage is found, discontinue using the
probe and return the probe to AiBiomed.
2. Inspect the probe for any signs of damage, particularly paying attention to the probe tip. If any damage is
found, discontinue using the probe and return the probe to AiBiomed.
Note: AiBiomed recommends that you purchase additional probe assemblies so that a backup is available in case
one is dropped or becomes unsterile during an operation.
Follow the IFU instruction for the PTeye Device (D121254_EN) in order to use the PTeye-1 Probe, make sure you
have read the entire IFU and pay special attention to the warnings section.