Medtronic PTEYE-1 Instructions for Use

D121255_EN Rev F PTeye-1 Instructions for Use Insert P a g e | 1 of 4
PARATHYROID DETECTION SYSTEM PROBE
MODEL PTEYE-1
INSTRUCTION FOR USE INSERT
DESCRIPTION The Parathyroid Detection (PTeye) System aids surgeons in confirming suspected parathyroid tissue that has been
located visually by the surgeon during common surgical procedures using a probe assembly. The probe assembly includes one glass fiber optic element that emits non-ionizing radiation at 785nm in the near IR range (NIR) and one fiber optic detector elements that collect and transmit the fluorescence emitted by the tissue to a photo detector. Tissue detection is based on the ratio of the fluorescent response of the parathyroid to the thyroid tissue; the fluorescence of thyroid tissue is much lower than the parathyroid.
INDICATIONS FOR USE
The AiBiomed Parathyroid Detection System (Model PTeye) is an adjunctive tool intended to aid in the identification of parathyroid tissue by confirming parathyroid tissue already visually located by the surgeon.
CONTRAINDICATIONS
There are no known contraindications for use for the AiBiomed Parathyroid Detection System (Model PTeye).
DEVICE WARNINGS and PRECAUTIONS
This equipment is designed for use by medical professionals familiar with its required techniques and these instructions for use. Read and follow all warning and cautionary notices and instructions included in this IFU.
The words WARNING, PRECAUTION, and NOTE carry special meanings and they should be read carefully.
WARNING: The safety and/or health of the patient, user, or a third party is at risk. Comply with this warning to avoid injury to the patient, user, or third party.
PRECAUTION: This contains information concerning the intended use of the device or accessory. Damage to the equipment is possible if these instructions are not followed.
NOTE: A note is added to provide additional, focused, information.
Warnings
1. Due to limitations in parathyroid detection of autofluorescence by the PTeye System in certain disease states,
the device is not recommended for use in patients with secondary hyperparathyroidism and in patients with parathyroid cysts.
2. The PTeye System is intended to be an aid in the identification of parathyroid tissue and not as a parathyroid
locator. The use of this device has not been evaluated as a parathyroid tissue locator.
3. This device shall only be used with original and manufacturer’s accessories and replacement parts. Use of
other parts or materials may degrade safety.
D121255_EN Rev F PTeye-1 Instructions for Use Insert P a g e | 2 of 4
4. NO modifications of this equipment is allowed.
5. Do not reuse the probe assembly. Discard probe assembly after each surgery.
6. Do not use the device if the environmental conditions do not meet the environmental requirements of the
device
Precautions
1. Do not use the device with incompatible equipment or peripherals that are not authorized by AiBiomed.
Doing so may void existing certifications and/or warranties.
2. Prior to each use, visually inspect the PTeye system, including console, probe assembly, external power
supply, power cord, and foot pedal to ensure there are no visible cuts, cracks, or other damage. If any damage is found, do not use the system; contact your AiBiomed customer service representative for assistance.
3. Do not excessively bend or kink the flexible portion of the probe assembly.
4. Store the device and all peripherals in a protective container to prevent the possibility of damage when not in
use. Do not store the device in a location where it will be exposed to temperatures exceeding 140°F (+ 60°C).
5. Handle all equipment carefully. If the fiber optic probe assembly is dropped, it can be damaged, which
renders the system unusable.
6. This device should only be used in compliance with its intended use.
7. Due to a small sample size, limited clinical data is available regarding the safety and effectiveness of the
PTeye System for rare disease states such as tertiary hyperparathyroidism, concomitant thyroid-parathyroid diseases, malignant parathyroid diseases, or other circumstances when prophylactic thyroidectomies are performed in individuals at high-risk for certain diseases such as MEN2A.
STERILIZATION
The PTeye-1 Probe is provided sterile. Do not re-sterilize Compromised performance and cleaning hazards are imminent.
STORAGE AND HANDLING
Store at room temperature. Avoid storing the PTeye-1 Probe at conditions of excessive heat or humidity. Do not use after expiration date on label. All components should be handled carefully to avoid damaging the device. Inspect all parts for damage prior to use, do not use any damaged parts. Do not reuse device.
PROCEDURE
Verifying Probe Assembly Function
The probe assembly contains delicate fiber optic cables. To verify that these have not been damaged during the shipment process, follow these steps.
1. Inspect the probe sterile package for any signs of damage. If any damage is found, discontinue using the
probe and return the probe to AiBiomed.
2. Inspect the probe for any signs of damage, particularly paying attention to the probe tip. If any damage is
found, discontinue using the probe and return the probe to AiBiomed.
Note: AiBiomed recommends that you purchase additional probe assemblies so that a backup is available in case one is dropped or becomes unsterile during an operation.
Follow the IFU instruction for the PTeye Device (D121254_EN) in order to use the PTeye-1 Probe, make sure you have read the entire IFU and pay special attention to the warnings section.