Medtronic PS210-030P Instructions for Use

PlasmaBlade™ PLUS
Single-Use Device Designed for Use with Qualied Generators
Instructions For Use
© 2018 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further, Together are trademarks of Medtronic. All other brands are trademarks of a Medtronic company.
PlasmaBlade™ PLUS
Device Description
The PlasmaBlade™ PLUS is a single-use, monopolar RF device. It is designed to be used with the qualied Generators as part of the Surgery System. It can be operated with the integrated handswitch or a qualied Footswitch. The PlasmaBlade™ PLUS consists of a single bendable blade and telescoping shaft that can be congured in both standard and extended length. The nger grip also incorporates a suction lumen for the evacuation of smoke and uids. A CoagCap (ball electrode) attaches to the nger grip to allow a broader application of energy. For a full list of qualied generators, see IFU insert card.
Yellow Button (CUT)
Locking Mechanism
Blue Button (COAG)
Cable
Handpiece
Cable Clips
Bendable Blade
Finger Grip
CoagCap (Ball Electrode)
Suction Opening
Indications
The Surgery System is indicated for cutting and coagulation of soft tissue during General, Plastic and Reconstructive (including but not limited to skin incisions and development of skin aps), ENT, Gynecologic, Orthopaedic, Arthroscopic, Spinal and Neurological procedures.
Contraindications
The PlasmaBlade™ should not be used on small appendages or body parts, as in circumcision.
Adverse Events
As a consequence of electrosurgery, damage to surrounding tissue through iatrogenic injury could occur.
Warnings
• Observe re precautions at all times. An electrosurgical device may provide an ignition source due to sparking and heating.
• Do not use in the presence of ammable anesthetics or oxidizing gases such as nitrous oxide and oxygen. Do not activate the device until vapors from alcohol-based skin prepping agents have dissipated. Naturally occurring gases that accumulate in body cavities can also be an ignition source.
• Do not touch the electrode tip of the PlasmaBlade™ device while power is being applied as this may result in user injury.
• Do not contact metal objects and instruments with the PlasmaBlade™ device while power is being applied as unintended tissue damage and electrode tip damage could occur.
• Activation of the handpiece when not in contact with target tissue may cause capacitive coupling.
• Do not allow patient contact with grounded metal objects, as such contact may result in patient or user injury.
• Activation of the handpiece simultaneously while aspirating uid may alter the path of electrical energy away from target tissue.
• Activation of the PlasmaBlade™ device outside the eld of view could cause patient injury.
• Inadvertent patient contact may result in burns. When not in use, place the device in a dry and nonconductive area away from the patient.
• Ensure that only the active tip of the device is in contact with the patient during use.
• Do not insert the device into tissue past the blade, as damage to the device and unintended injury may occur.
• Monopolar devices require a Patient Return Electrode. The unit must detect proper Patient Return Electrode impedance before Generator output can be active. The impedance is continuously monitored while in Monopolar Mode. The unit presents audible and visible alarms if the unit detects improper impedance with the Patient Return Electrode in Monopolar Mode and will disable Generator output. Refer to the manufacturer’s instructions for application site and placement procedures when applying the Patient Return Electrode. Do not rely entirely on the impedance sensing feature as it can be aected by a damaged (shorted) Patient Return Electrode. It is recommended that the operator verify appropriate placement and contact of the Patient Return Electrode. Inadequate contact of the Patient Return Electrode may result in patient alternate site burns or injury. Inadequate contact of the Patient Return Electrode may result in patient alternate site burns or injury.
1
• Do not reuse, resterilize or reprocess the PlasmaBlade™ device as it is supplied sterile and intended for Single Use Only. A device that has been resterilized or reprocessed may not perform properly and may result in patient or user injury.
• When bending the blade, do not exceed a 45° angle. Do not bend more than three times. Excessive bending of the shaft may compromise performance of the device or cause device failure, which could result in patient or user injury. Use nger force to bend the blade; do not use forceps as this could damage the device.
• Avoid uid contact with the handle and its interfaces and connections as this may result in patient or user injury, or device failure.
• Position the cable to avoid creating a tripping hazard.
• Position the cable to avoid patient contact to protect against high frequency
current paths to the patient as this may result in patient or user injury.
• The use of electrosurgery in the presence of internal or external active implants is potentially hazardous.
• Interference from the electrical current can cause device malfunction. Consult the active implant manufacturer for further information before proceeding with the surgery.
• Direct contact with implanted leads can cause physical damage to the leads. Exercise caution around leads associated with any active implant; particularly those with thin insulation.
• Do not use monopolar electrosurgery on small appendages, such as in nger surgery, as it can cause thrombosis or other unintended injury to tissue proximal to the surgical site.
• The device should not be used near electrocardiograph electrodes as it can cause interference.
• Ensure the ball electrode is securely attached to the nger grip before use. Do not continue to activate the device if the ball electrode tip becomes loose as injury might occur.
Precautions
• The PlasmaBlade™ device should only be used by qualied medical personnel possessing training in the surgical procedures to be performed.
• Take care when handling the electrode tip to prevent possible damage to the tip and to prevent user injury.
• Use the lowest power setting and the shortest activation time possible to achieve the desired end eect.
• Unless the product is being used to spot coagulate a vessel, it is recommended to keep the electrode tip in motion while activated to avoid excessive eschar buildup. Excessive eschar buildup can compromise device performance.
• Do not use sharp or abrasive instruments or materials to clean eschar buildup on the electrode tip as this may damage the tip.
• When not in use, the device should be placed in the holster securely fastened to the surgical drape.
• Prior to initial use, ensure that all package inserts, including warnings, cautions, instructions for use, as well as the selected Generator Operator’s Manual, are read and understood.
• Prior to use, carefully inspect the package before opening. Do not use the device if the package appears damaged in any way. Return damaged packages to Medtronic Advanced Energy LLC.
• Prior to use, inspect the PlasmaBlade™ device for any defects. Do not use if insulation or connectors are damaged.
• The PlasmaBlade™ device is intended for use only with a qualied Generator.
• Prior to use, make sure that the locking mechanism is in the locked position
and unlocked only when adjusting the device shaft length.
Equipment List
• PlasmaBlade™ Compatible Generator
• Optional Generator Compatible Footswitch
• Patient Return Electrode, compatible with selected Generator
Instructions For Use
Read all instructions carefully. Failure to comply may lead to electrical or thermal injury, or cause device malfunction.
Please refer to the Generator Operator’s Manual for step-by-step instructions regarding initial setup and preparation for surgery. The PlasmaBlade™ device may only be used with a qualied Generator.
Before Surgery
1. Inspect the PlasmaBlade™ package to ensure that it is unopened and undamaged. If it is damaged or the sterile seal is broken, do not use the device.
2. Remove the PlasmaBlade™ device and holster from packaging using aseptic technique.
3. Inspect the device for any signs of damage. If the device is damaged it must not be used.
4. Attach the holster to the sterile drape using a clamp or tape. The PlasmaBlade™ cable can be placed into the cutouts of the holster for convenience.
NOTE: Ensure that the locking mechanism is in the locked position prior to use and unlocked only when adjusting the device shaft length.
2
5. For detailed instructions on the selected Generator use, please refer to the operator’s manual. Do not use the PlasmaBlade™ device until reading and understanding the directions for the selected Generator.
6. Switch on the Generator to perform the initial setup and self-test.
7. Plug the PlasmaBlade™ device into the Monopolar Connector of the Generator. Upon connection, the unit will alarm if the device is not recognized as a PlasmaBlade™ device or if it fails a unit test. In event of an alarm, replace the device.
8. If using suction, connect the suction tubing from either a central or portable suction line to the connector on the handpiece.
The Generator will display default power level settings for the PlasmaBlade™ device upon connection. These settings must be veried and adjusted by the user to ensure the appropriate settings are selected. The lowest power setting possible should be used to achieve the desired end eect.
Cable cutouts
Electrode cleaning slot
Table 1: PlasmaBlade™ PLUS Settings
Tissue Effect
CUT
low
hemostasis,
low collateral
damage
Level
(LCD
display)
Mode
1
2 2W
Low Cut
3 6W
Power
[W-RMS]
0.5W
4 10W
5 20W*
high
hemostasis,
higher
collateral
damage
COAG
low
hemostasis,
low collateral
damage
6
Medium Cut
7 35W
8 50W**
High Cut
9
(Blend 1)
High Cut
10
(Blend 2)
1
2
Low Coag
3
4
5
high
hemostasis,
higher
collateral
damage
6
7
High Coag
8
9
10
* When Cut level is changed from 5 to 6, there is a change in Cut mode from Low Cut to
Medium Cut.
** When Cut level is changed from 8 to 9, Cut mode changes from Medium Cut to High Cut
(Blend 1). Hence, output power will increase when the Cut level is changed from 9 to 8.
*** When Coag level is changed from 5 to 6, Coag mode changes from Low Coag to High
Coag. Hence, output power will increase when the Coag level is changed from 6 to 5.
**** The listed Maximum Output Voltage values pertain to device utilization with the AEX
generator. The overall Maximum Output Voltage is 2750 Vpk in accordance with device utilization on the Pulsar II generator.
20W
25W
50W
15W
20W
25W
30W
35W***
30W
35W
40W
45W
50W
****
Maximum
Output
Voltage (Vpk)
1000
1100
950
1400
1400
500
550
550
1500
1400
1400
1400
1400
1400
1600
2400
2400
2550
2600
2600
9. If using the optional footswitch, plug the wireless receiver into the Footswitch
Connector of the selected Generator. The footswitch can only be used to activate output on a monopolar device if a bipolar device is not connected.
10. Position the Patient Return Electrode on the patient and then connect
the Patient Return Electrode to the Patient Connector receptacle of the qualied Generator.
11. Monopolar devices require a Patient Return Electrode. The Generator must
detect proper Patient Return Electrode impedance before it can be active. The operator should conrm the presence of a properly connected Patient Return Electrode.
12. The qualied Generators have an Impedance indicator display that
indicates the type and impedance level of the Patient Return Electrode in monopolar mode. The indicator will remain green with proper return electrode impedance. The indicator will turn red when the unit does not detect proper return electrode impedance. Ensure proper return electrode impedance before starting surgery. Refer to the Generator Operator’s Manual for more information.
13. Remove the electrode tip protector before activation of device.
3
Manipulating the Device
1. Ensure power is not activated while manipulating the device or handling
the electrode tip, as this may result in user injury.
2. To extend the shaft, rmly grasp the
nger grip and pull away from the handpiece.
3. To collapse the shaft, rmly grasp
the nger grip and push into the handpiece. Be careful not to pinch anything between the nger grip
Collapse (push)
Rotate
Bend
Extend (pull)
and the shaft when collapsing.
4. If needed, use nger force to bend the blade to a maximum of 45°. Inspect
the blade for damage after bending and do not use if damaged. Do not bend more than three times. Excessive bending of the shaft may cause device failure and could result in patient or user injury.
5. The tip can be rotated to any angle. Firmly grasp the nger grip and rotate
to the desired position.
6. The shaft and nger grip may be
locked into place at the desired position by rotating the locking mechanism counterclockwise. The shaft has been locked into place
Locked
when the arrow lines up with the horizontal bar. Do not tighten beyond visual indicator.
7. To unlock the shaft and nger grip,
turn the locking mechanism until
Unlocked
loose.
Using the CoagCap (Ball Electrode)
1. Ensure power is not activated while handling the tip, as this may result in
user injury.
2. Remove the ball electrode from the package by peeling open the ends of
the pouch.
3. Inspect the tip for
damage. Do not use if damaged.
4. The blade of the device
must be straight for the
Attach
Detach
ball electrode to securely attach.
5. Collapse the shaft of the
PlasmaBlade™ device or hold the shaft to prevent it from moving. If the shaft is extended, it may move while you attach the ball electrode.
Insertion line
6. Connect the tip by holding the nger grip of the ball electrode and pushing
the tip over the blade and onto the nger grip. The ball electrode should be ush with the insertion line on the device and click into place.
7. Ensure the ball electrode is securely attached to the nger grip before use.
8. The ball electrode can be rotated to any direction and bent to a maximum
of 90°. To bend the ball electrode, apply force only in the bendable area.
9. To remove the ball electrode, grasp the nger grip and pull straight o.
Hold the shaft of the PlasmaBlade™ device to prevent it from extending.
During Surgery
Place the device in the holster when not in use.
1. To cut, press the yellow button/pedal on the handswitch or footswitch.
2. To coagulate, press the blue button/pedal on the handswitch or footswitch.
3. The ball electrode should be oriented perpendicular to the working surface
during activation. Coagulation with the ball electrode should be achieved with spot (vertical up and down) motion.
NOTE: The ball electrode can be used with both cut and coagulate modes for various tissue eects. See Tables 1 and 2 for more information.
NOTE: Eschar buildup on the tip can be removed with gauze pads or by inserting the tip into the slot at the front of the holster and drawing the device backwards through the slot. Inspect the device for any signs of damage after cleaning. Buildup on the ball electrode can be minimized by using it on moist tissue at the lowest eective setting.
Cleaning slot
After Surgery
1. Turn o the Generator.
2. Disconnect the PlasmaBlade™ device, Footswitch, and Patient Return
Electrode from the Generator.
3. Discard the PlasmaBlade™ device after use. It is intended for Single
Use Only. Do not reuse, resterilize, or reprocess. A device that has been resterilized or reprocessed may not perform properly and may result in patient or user injury.
4
How Supplied
The PlasmaBlade™ device is supplied sterile and is intended for Single Use Only. Contents are sterile unless the package is opened or damaged. Do not reuse, resterilize, or reprocess the PlasmaBlade™ device. The product is sterilized using Ethylene Oxide (EO).
Customer Service
Please call Medtronic Customer Service at +18667779400 if you have any device returns or questions about Medtronic’s device.
Limited Express Warranty
IF, PRIOR TO THE PRODUCT EXPIRATION DATE, THE STERILIZED DEVICE IS FOUND TO BE INOPERABLE DURING NORMAL AND PROPER USE IN ACCORDANCE WITH APPLICABLE INSTRUCTIONS, MEDTRONIC ADVANCED ENERGY LLC WILL REPLACE THE PRODUCT AT NO CHARGE.
This Limited Express Warranty does not extend to a product where the user has compromised the sterile integrity of the package or the product is used after the product expiration date. A sterile package that has been compromised should not be used.
Disclaimer of Implied Warranties and Consequential Damages
MEDTRONIC ADVANCED ENERGY LLC MAKES NO OTHER WARRANTIES WITH RESPECT TO THE PRODUCT AND EXPRESSLY DISCLAIMS ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, AS TO MERCHANTABILITY, FITNESS FOR A PARTICULAR PURPOSE OR ANY OTHER MATTER. IN NO EVENT SHALL MEDTRONIC ADVANCED ENERGY LLC BE LIABLE FOR ANY INDIRECT, CONSEQUENTIAL OR SPECIAL DAMAGES OF ANY KIND. THE REMEDIES SET FORTH IN THE LIMITED EXPRESS WARRANTY ARE THE EXCLUSIVE REMEDIES AVAILABLE TO CUSTOMER.
Symbols
Follow Instructions for Use
Use by Date
Do Not Reuse
Do Not Resterilize
Humidity Limitation
Contains DEHP (di-2-ethyl hexyl phthalate)
Temperature Limitation
Sterilized by Ethylene Oxide
Non-Pyrogenic
Catalog Number
Lot Number
Manufacturer
Caution: Federal law (USA) restricts this device to sale by or on the order of a
physician.
5
Medtronic Advanced Energy LLC 180 International Drive Portsmouth, NH 03801 USA +18667779400
medtronic.com
manuals.medtronic.com
© Copyright 2018 Medtronic Advanced Energy LLC 70-10-1453 Rev D 2018/10
Loading...