Medtronic PM7100_Operaters_Manual_en_PT00097962B00 INVOS Patient Monitor Operator’s Manual

Operator’s Manual
TM
INVOS
Patient Monitor
©2019 Medtronic. All rights reserved. Medtronic and Medtronic logo are trademarks of Medtronic. ™* Third party brands are trademarks of their respective owners. All other brands are trademarks of a Medtronic company.

Table of Contents

1.1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
1.2 Intended Audience . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
1.3 Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
1.3.1 Safety Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
1.3.2 Explosion, Shock, and Toxicity Hazards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
1.3.3 Patient Monitoring and Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
1.3.4 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-4
1.3.5 Sensors, Cables, and Other Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-5
1.3.6 Electromagnetic Interference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-5
1.3.7 Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
1.4 Obtaining Technical Assistance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
1.4.1 Technical Services . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-7
1.4.2 Related Documents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-8
1.5 Warranty Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-9
2 Product Overview
2.1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
2.2 Product Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
2.3 Indications for Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
2.4 Product Views . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
2.4.1 Full System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
2.4.2 Monitor Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
2.5 Product and Carton Label Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-13
3 Installation
3.1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
3.2 Safety Reminders . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
3.3 Unpacking and Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
3.4 Power Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
3.4.1 AC Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
3.4.2 Battery Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-5
3.5 Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
3.5.1 Install the Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-7
3.5.2 Insert the Monitor into the Docking Station . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9
3.5.3 Connect the Preamplifier(s) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-11
3.5.4 Connect the Reusable Sensor Cables (RSCs) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-12
3.5.5 Apply Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-13
3.5.6 Set Institutional Defaults . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-16
3.5.7 Power Off the Monitoring System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-17
3.5.8 Position the Monitoring System in the Clinical Setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-17
Operator’s Manual iii
Table of Contents
4 Operation
4.1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
4.2 Safety Reminders . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
4.3 Quick Start . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
4.4 Operational Defaults . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-7
4.5 Menu Structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-9
4.6 Monitoring System Memory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-9
4.7 Set Up for Patient Monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-11
4.8 Optional Set-Up Tasks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-16
4.8.1 Changing the Sensor Placement Sequence . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-16
4.8.2 Repositioning On-Screen Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-17
4.8.3 Assigning or Modifying the Patient ID . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-19
4.8.4 Changing the On-Screen Sensor Labels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-21
4.8.5 Changing the Screen Brightness . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-23
4.9 Baselines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-24
4.10 Trend View Management . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-27
4.10.1 About the Trend View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-28
4.10.2 Changing the Trend View - Two Graphs or One Graph . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-29
4.10.3 Turning Trend Line Averaging On/Off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-30
4.10.4 Zooming the Trend View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-32
4.10.5 Reviewing Data that has Scrolled Off the Trend View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-32
4.10.6 Viewing Previous rSO2 Values on the Trend Graph . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-33
4.11 Alarm Management . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-35
4.11.1 Alarm Indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-35
4.11.2 Setting Alarm Limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-38
4.11.3 Changing the Alarm Volume . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-40
4.11.4 Silencing or Pausing Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-41
4.11.5 Dismissing an Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-43
4.12 Event Marks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-44
4.12.1 About Event Marks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-44
4.12.2 Marking Events . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-45
4.12.3 Viewing an Event Mark Label . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-45
4.12.4 Renaming an Event . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-46
4.12.5 Changing the Event List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-47
4.13 Area Under the Curve (AUC) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-49
4.13.1 About Area Under the Curve (AUC) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-49
4.13.2 Changing the AUC Threshold . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-50
4.13.3 Resetting AUC Collection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-51
4.14 Finish Monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-51
4.15 Case Histories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-52
4.15.1 About Case Histories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-52
4.15.2 Viewing Case Histories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-52
4.15.3 Exporting Case Histories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-55
iv Operator’s Manual
Table of Contents
5 Data Management
5.1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
5.2 Safety Reminders . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
5.3 Displaying Monitoring System Screens on an External Monitor . . . . . . . . . . . . . . . . . 5-2
5.4 Downloading Case Histories to a USB Drive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
5.5 Transmitting Monitoring System Data to External Devices via the Serial Port . . . . 5-3
5.5.1 Serial Port Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-3
5.5.2 Transmitting Real-Time Data to a Philips IntelliBridge™* and VueLink™*
Open Interface (IVOI) Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5
5.5.3 Transmitting Case Data to a Computer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-6
5.6 Data Formats . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-8
5.6.1 Case Histories Downloaded via USB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-8
5.6.2 Case Data Downloaded via Serial Port . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-10
5.6.3 Event Codes for Data Downloads . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-13
5.6.4 Status Codes for Data Downloads . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-17
6 Performance Considerations
6.1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
6.2 Safety Reminders . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
6.3 Patient Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
6.4 Sensor Use Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
6.5 EMI (Electromagnetic Interference) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
7 Product Maintenance
7.1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
7.2 Safety Reminders . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
7.3 Cleaning the Monitoring System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-2
7.3.1 Materials . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-2
7.3.2 Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-3
7.4 Maintenance Schedule . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-3
7.5 Service and Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-4
7.6 Service Life . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-4
7.7 Software and Firmware Updates . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-4
7.8 Recycling and Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-5
8 Alarms and Troubleshooting
8.1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-1
8.2 Alarm Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-2
8.3 Error Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-7
8.4 Product Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-11
Operator’s Manual v
Table of Contents
9 Accessories
9.1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-1
9.2 Safety Reminder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-1
9.3 Accessories/Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-2
10 Theory of Operations
10.1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-1
10.2 Theoretical Principles . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-1
11 Product Specifications
11.1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-1
11.2 Physical Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-2
11.3 Electrical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-3
11.4 Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-3
11.5 Environmental Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-4
11.6 Alarm Audio and Visual Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-4
11.6.1 Alarm Audio Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11-5
11.6.2 Alarm Visual Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11-5
11.6.3 POST Pass Tone . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11-6
11.7 Sensor Operating Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-6
11.8 Equipment Classifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-6
11.9 Biocompatibility Testing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-7
11.10 Manufacturer’s Declaration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-7
11.10.1 Electromagnetic Compatibility (EMC) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11-9
11.11 Essential Performance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-14
A Clinical Studies
A.1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-1
A.2 Hypoxia Study . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-1
A.3 Interventional Studies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-3
A.4 Validation Studies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-4
vi Operator’s Manual

List of Tables

Table1-1. Safety Symbol Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Table2-1. Label Symbol Descriptions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-13
Table3-1. Monitoring System - Standard Shipped Items . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Table3-2. AC Power Indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
Table3-3. Battery Status Indicators - Monitoring System Operating on Battery Power . . . . . . . . . . . 3-6
Table3-4. Institutional Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-16
Table4-1. Quick Start Steps. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
Table4-2. Operational Defaults . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-7
Table4-3. Menu Structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-9
Table4-4. System Memory - Common Situations and Recommended Actions . . . . . . . . . . . . . . . . . 4-10
Table4-5. Alarm Audio Button States . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-36
Table4-6. Audible and Visual Alarm Indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-37
Table4-7. Alarm Limits Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-38
Table5-1. Serial Port Pin-Out Descriptions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
Table5-2. Hardware Requirements for Philips VueLink™* and IntelliBridge™* Communication . . . 5-5
Table5-3. Case History Downloads - Data Fields . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-10
Table5-4. Case Data Downloads (PC LINK 1 and PC LINK 2) - Data Fields . . . . . . . . . . . . . . . . . . . . . . . 5-12
Table5-5. Data Downloads - Event Codes 1 to 40 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-13
Table5-6. Data Downloads - Event Codes 41 to 80. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-14
Table5-7. Data Downloads - Event Codes 81 to 120 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-15
Table5-8. Data Downloads - Event Codes 121 to 254 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-16
Table5-9. Data Downloads - Status Codes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-17
Table7-1. Maintenance Schedule. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-3
Table8-1. Alarm Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-2
Table8-2. Error Conditions and Resolutions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-7
Table9-1. Monitoring System Accessory Part Numbers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-2
Table11-1. Transport, Storage, and Operating Condition Ranges . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11-4
Table11-2. Alarm Audio Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11-5
Table11-3. Average Alarm Sound Pressure Levels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-5
Table11-4. Alarm Visual Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-5
Table11-5. POST Pass Tone Characteristics. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-6
Table11-6. INVOS™ Adult rSO
Table11-7. Electromagnetic Emissions Guidelines and Compliance . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11-9
Table11-8. Electromagnetic Immunity Guidelines and Compliance . . . . . . . . . . . . . . . . . . . . . . . . . . . .11-10
Table11-9. Recommended Separation Distance Calculations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11-11
Table11-10. Test Specifications for Enclosure Port Immunity to RF Wireless Communications
Equipment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11-13
Table11-11. Recommended Separation Distances . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11-14
TableA-1. Performance Characteristics: Accuracy Bias, Error (SD), RMSD, and Correlation (R
for rSO
and fSO2, 20 Subjects. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-2
2
TableA-2. Trend Bias, Error (SD), and Correlation (R
Sensor Operating Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-6
2
2
)
2
) for ΔrSO2 and ΔfSO2, 20 subjects . . . . . . . . . .A-3
Operator’s Manual vii
Page Left Intentionally Blank
viii

List of Figures

Figure1-1. Start-up Screen Showing Code Version. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
Figure2-1. INVOS™ Patient Monitor Components. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
Figure2-2. Monitor - Front . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
Figure2-3. Monitor - Left Side . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
Figure2-4. Monitor - Right Side. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6
Figure2-5. Monitor - Back . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7
Figure2-6. Docking Station - Bottom View. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8
Figure2-7. Docking Station - Connector for Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-9
Figure2-8. Sample Monitoring Screen Elements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-10
Figure3-1. Power Input Connection at Bottom of Docking Station . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
Figure3-2. Docking Station Power Indicator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
Figure3-3. Battery Location at Back of Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
Figure3-4. Battery Slot and Latch at Back of Monitor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8
Figure3-5. Battery Installed in Monitor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9
Figure3-6. Monitor Mounted in Docking Station. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-10
Figure3-7. Connecting the Preamplifier Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-11
Figure3-8. Connecting the RSC to the Preamplifier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-12
Figure3-9. Power Input Connection at Bottom of Docking Station . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-13
Figure3-10. Power Button on Top of Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-14
Figure3-11. Set-Up Screen - Preamplifier(s) Connected . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-15
Figure3-12. Set-Up Screen - No Preamplifier(s) Connected. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-15
Figure3-13. Monitoring System on Stand with Preamplifier Hung on Rail . . . . . . . . . . . . . . . . . . . . . . . . 3-17
Figure4-1. Set-Up Screen - Prompt for Sensor Connection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-12
Figure4-2. Monitoring Screen - New Case - No Baselines Set. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-14
Figure4-3. Default Sensor Placement Sequence . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-16
Figure4-4. Optional Sensor Placement Sequences. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-17
Figure4-5. Highlighting an On-Screen Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-18
Figure4-6. Sensor Icons Next to CURRENT Values (Press to Locate Sensor on Patient) . . . . . . . . . . . 4-18
Figure4-7. Patient Icon. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-19
Figure4-8. Entering a Patient ID Before Monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-20
Figure4-9. Patient Icon. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-20
Figure4-10. Entering a Patient ID During Monitoring. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-21
Figure4-11. Sensor Label Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-22
Figure4-12. SETTINGS - GENERAL - SCREEN BRIGHTNESS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-23
Figure4-13. Monitoring Screen - New Data Set - No Baselines Set . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-25
Figure4-14. SET BASELINES - No Baselines Previously Set. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-25
Figure4-15. SET BASELINES - Baselines Previously Set . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-26
Figure4-16. Monitoring Screen with Baselines Set . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-27
Figure4-17. Trend View. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-28
Figure4-18. Trend View - One Graph (Left, Default) and Two Graphs (Right) . . . . . . . . . . . . . . . . . . . . . .4-29
Figure4-19. SETTINGS - TRENDS - TWO AXIS VIEW vs. ONE AXIS VIEW. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-30
Figure4-20. Trend Line Averaging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-31
Figure4-21. SETTINGS - TRENDS - TREND LINE AVERAGING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-31
Operator’s Manual ix
List of Figures
Figure4-22. Trend View Review Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-33
Figure4-23. Trend Graph Pop-Up. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-34
Figure4-24. Monitoring Screen - Trend Graph Pop-Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-34
Figure4-25. Visual Alarm Indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-36
Figure4-26. SET ALARM LIMITS - Default Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-39
Figure4-27. SET ALARM LIMITS - Upper Limits On. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-39
Figure4-28. SETTINGS - ALARMS - ALARM VOLUME . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-41
Figure4-29. Dismissible Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-43
Figure4-30. Events Marked in Trend View. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-44
Figure4-31. ADD EVENT Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-45
Figure4-32. Event Mark Label. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-46
Figure4-33. EDIT EVENT Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-47
Figure4-34. SETTINGS - EVENTS (Screen 1) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-48
Figure4-35. AUC SUMMARY Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-49
Figure4-36. CASE HISTORY List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-53
Figure4-37. Case History Example . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-54
Figure5-1. VGA Port on Docking Station. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
Figure5-2. Serial Port on Docking Station. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Figure5-3. Serial Port Pin-Outs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
Figure5-4. Null Modem Cable Diagram for Connection to Serial Port (Two Options) . . . . . . . . . . . . . 5-7
FigureA-1. Accuracy rSO FigureA-2. Trend rSO
and fSO2, 20 Subjects . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-2
2
and fSO2, 20 Subjects . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-3
2
x Operator’s Manual
1 Introduction

1.1 Overview

This manual contains information for operating the INVOS™ patient monitor (the “monitoring system”).
This manual applies to the following product:
PM7100
Intended Audience, page 1-1
Safety Information, page 1-2
Obtaining Technical Assistance, page 1-7
Warranty Information, page 1-9

1.2 Intended Audience

This manual provides information to health-care professionals in a hospital setting for operation and maintenance of the monitoring system. Refer to the institution for any additional training or skill requirements beyond those identified here for opera­tion and maintenance of the monitoring system.
Before operating the monitoring system, carefully read this manual, the Instructions for Use for the accessories, and all precautionary information and specifications.
1-1
Introduction

Safety Information

1.3
This section contains important safety information for use of the monitoring system.

1.3.1 Safety Symbols

Table1-1.Safety Symbol Definitions
Symbol Definition
WARNING
Warnings alert users to potential serious outcomes (death, injury, or adverse events) to the patient, user, or environment.
Caution Cautions alert users to exercise appropriate care for safe and effective use of the product.
Note
Notes provide additional guidelines or information.

1.3.2 Explosion, Shock, and Toxicity Hazards

WARNING:
Explosion hazard — Do not use the monitoring system in the presence of flammable anesthetics.
WARNING:
Explosion hazard — When replacing the battery, do not use the incorrect type. Use only the battery available from Medtronic. See Accessories/Parts List, page 9-2.
WARNING:
Shock hazard — Ensure the monitoring system is properly grounded when operating on AC power.
WARNING:
Shock hazard — When connecting the monitoring system to any instrument, verify proper operation before clinical use. Any equipment connected to the data interface must be certified according to the latest IEC/EN 60950-1 standard for data-
1-2 Operator’s Manual
processing equipment, the latest IEC/EN 60601-1 standard for electromedical equipment, or the latest IEC/EN safety standards relevant to that equipment. All combinations of equipment must be in compliance with Requirements for Medical Electrical Systems within IEC/EN Standard 60601-1. Anyone who connects equipment to the data interface is configuring a medical system and, therefore, is responsible for ensuring the system complies with the Requirements for Medical Electrical Systems IEC/EN Standard 60601-1 and the electromagnetic compatibility IEC/EN Standard 60601-1-2. Performance may degrade if it is connected to secondary I/O devices when the equipment is not connected to earth reference.
WARNING:
Toxicity hazard — The LCD panel (screen) contains toxic chemicals. Do not touch broken LCD panels. Physical contact with a broken LCD panel can result in transmission or ingestion of toxic substances.

1.3.3 Patient Monitoring and Safety

Safety Information
WARNING:
The monitoring system should not be used as the sole basis for diagnosis or therapy. It is intended only as an adjunct in patient assessment. It must be used in conjunction with clinical signs and symptoms.
WARNING:
Always disconnect and remove the monitoring system and sensors during magnetic resonance imaging (MRI) scanning. Attempting to use the monitoring system during an MRI procedure could cause burns or adversely affect the MRI image or the monitoring system's performance.
WARNING:
As with all medical equipment, carefully route patient cabling to reduce the possibility of patient entanglement or strangulation.
WARNING:
Do not set alarm limits to extreme values that render the monitoring system ineffective. Ensure alarm limits are appropriate for each patient.
WARNING:
Do not silence, pause, or decrease the volume of audible alarms if patient safety could be compromised.
Operator’s Manual 1-3
Introduction
WARNING:
WARNING:
Caution:
An external multi-parameter system will not generate an alarm or error message if remote communication between the multi-parameter system and the monitoring system has been broken. During this period of no remote communication, the monitoring system will continue to monitor, generate alarms, and display status messages. The multi-parameter system operator should not rely on the multi­parameter system for generating alarms.
Choking hazard — The reusable sensor cables (RSCs) include a strain-relief clip that, if detached, may pose a choking hazard.
If two sensors are placed in close proximity to each other on a patient, the same preamplifier should be connected to both sensors to avoid poor performance.

1.3.4 Operation

WARNING:
Inspect the monitoring system and all accessories before use to ensure there are no signs of physical damage or improper function. Do not use if damaged.
WARNING:
If you do not hear a tone at system start-up, discontinue use of the monitoring system and contact Medtronic or a local Medtronic representative.
WARNING:
Monitoring system readings can be affected by certain patient conditions. Refer to
Patient Conditions, page 6-2.
WARNING:
Explosion hazard — When replacing the battery, do not use the incorrect type. Use only the battery available from Medtronic. See Accessories/Parts List, page 9-2.
WARNING:
To ensure proper performance, avoid shock, and prevent device damage or failure, do not expose the monitoring system to extreme moisture, such as direct exposure
1-4 Operator’s Manual
to rain. Do not immerse in water, solvents, or cleaning solutions, since the monitoring system and connectors are not waterproof.
Caution:
Dispose of the battery in accordance with local guidelines and regulations.

1.3.5 Sensors, Cables, and Other Accessories

WARNING:
The use of accessories, sensors, and cables other than those specified may result in poor performance of the monitoring system and increased electromagnetic emissions or decreased electromagnetic immunity of the monitoring system.
WARNING:
Failure to cover the sensor site with opaque material when operating under high ambient light conditions may result in poor performance.
Safety Information
Caution:
Make sure all connectors are fully engaged and free from moisture. Moisture intrusion may cause poor performance or no readings at all.

1.3.6 Electromagnetic Interference

WARNING:
Electromagnetic emissions from the monitoring system may interfere with other critical devices.
WARNING:
Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of the monitoring system, including cables. Otherwise, degradation of monitoring system performance may result.
WARNING:
The use of accessories, sensors, and cables other than those specified may result in poor performance of the monitoring system and increased electromagnetic emissions or decreased electromagnetic immunity of the monitoring system.
Operator’s Manual 1-5
Introduction
WARNING:
WARNING:
WARNING:
The monitoring system is intended for use by healthcare professionals only. It may cause radio interference or may disrupt the operation of nearby equipment. Mitigation for such disruption may require re-orienting or relocating the monitoring system or shielding the location.
Any radio frequency transmitting equipment or other nearby sources of electrical noise may result in disruption of the monitoring system.
The monitoring system is designed for use in environments in which the signal can be obscured by electromagnetic interference. During such interference, measurements may seem inappropriate or the monitoring system may not seem to operate correctly.
WARNING:
EMI disruption can cause cessation of operation or other incorrect functioning.
WARNING:
The monitoring system should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is necessary, observe the monitoring system to verify normal operation in the desired configuration. Technical alarms may indicate that the configuration is not appropriate for the monitoring system.
Caution:
This device has been tested and found to comply with the limits for medical devices related to IEC 60601-1-2: 2007 and IEC 60601-1-2:2014. These limits are designed to provide reasonable protection against harmful interference in a typical medical installation.
Caution:
When operating medical electrical equipment, special precautions related to electromagnetic compatibility (EMC) are required. Install the monitoring system according to the EMC information included in this manual.
1-6 Operator’s Manual
Caution:
The monitoring system generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to other devices in the vicinity. If interference is suspected, move monitoring system cables away from the susceptible device.
Caution:
The use of an electrosurgical or electrocautery instrument in the vicinity of the monitoring system may interfere with the signal and cause poor performance or no readings at all.

1.3.7 Cleaning

Obtaining Technical Assistance
Caution:
Do not autoclave or gas sterilize any components of the monitoring system.
Caution:
To prevent device damage or failure, do not expose the monitor to isopropyl alcohol.

1.4 Obtaining Technical Assistance

1.4.1 Technical Services

For technical information and assistance, contact Medtronic or a local Medtronic representative.
Medtronic Technical Services
5870 Stoneridge Drive, Suite 6
Pleasanton, CA 94588 USA
1.800.635.5267 or 1.925.463.4635
or contact a local Medtronic representative
www.medtronic.com
When calling Medtronic or a local Medtronic representative, have the monitoring system serial numbers and software versions available. Serial numbers are located
Operator’s Manual 1-7
Introduction
on the back of the monitor and the preamplifiers. The software version for the mon­itoring system is displayed on the start-up screen at power-on.
Figure1-1.Start-up Screen Showing Code Version
Note:
An authorized technician can view serial numbers and software versions through the monitoring system’s service mode. Refer to the monitoring system’s service manual.

1.4.2 Related Documents

INVOS™ Adult rSO
about sensor selection and use.
INVOS™ Preamplifier Instructions for Use — Provides instructions for connecting
the monitoring system’s preamplifiers.
INVOS™ Patient Monitor Service Manual —
nicians for use when servicing the monitoring system.
Sensor Instructions for Use —
2
Provides information to authorized tech-
Provides important information
1-8 Operator’s Manual

Warranty Information

1.5
To obtain product warranty information, contact Medtronic or a local Medtronic representative. See Technical Services, page 1-7.
The information contained in this document is subject to change without notice. Medtronic makes no warranty of any kind with regard to this material, including, but not limited to, the implied warranties or merchantability and fitness for a particular purpose. Medtronic shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material.
Warranty Information
Operator’s Manual 1-9
Introduction
Page Left Intentionally Blank
1-10 Operator’s Manual
2 Product Overview

2.1 Overview

This chapter contains basic information about the INVOS™ patient monitor (the “monitoring system”).
Product Description, page 2-1
Indications for Use, page 2-2
Product Views, page 2-3
Product and Carton Label Symbols, page 2-13

2.2 Product Description

The monitoring system provides continuous, noninvasive indications of changes in regional oxygen saturation of blood (rSO
) in cerebral and somatic tissues.
2
The monitoring system consists of:
A tablet-style monitor that can be operated on AC or battery power
A VESA* docking station for the monitor, with USB, Serial (RS-232), and VGA ports
An AC power supply and cord that attaches to the docking station
Up to two preamplifiers to accommodate up to four sensors (two per preamplifier)
Up to four reusable sensor cables (RSCs) to attach sensors to the preamplifiers
INVOS™ rSO
Additional accessories as described in Accessories, page 9-1
sensors
2
2-1
Product Overview
Monitoring system features include:
User configurable rSO
olds (page 4-49)
Physiological and technical alarm reporting (page 8-2)
baselines (page 4-24), alarm limits (page 4-38), and AUC thresh-
2
User configurable data display (rSO
, change from baseline, sensor labels, and trend
2
data) (page 2-10)
Alarm silencing (page 4-41)
Event marking (page 4-44)
Visual representations of sensor locations (page 4-17)
Sensor functional state (page 8-2):
- Sensor off
- Sensor disconnect
- Sensor fault
Case history storage and export (page 4-52)
Real-time data output to external devices such as a Philips multi-parameter system or
PC (page 5-3)
VGA, Serial (RS-232), and USB interfaces (pages 2-5, 2-8)

2.3 Indications for Use

The INVOS™ Patient Monitor, model PM7100, is a noninvasive cerebral/somatic oximetry system intended for use as an adjunct monitor of regional hemoglobin oxygen saturation of blood in the brain or in other tissue beneath the sensor. It is intended for use on individuals greater than 40 kg at risk for reduced-flow or no-flow ischemic states.
2-2 Operator’s Manual

Product Views

2.4

2.4.1 Full System

System Component Overview
Product Views
Figure2-1.INVOS™ Patient Monitor Components
1 Sensors 4 Monitor
2 Preamplifier 5 Reusable sensor cable (RSC)
3 Docking station
Operator’s Manual 2-3
Product Overview
2.4.2

Monitor Components

Monitor - Front
Figure2-2.Monitor - Front
1
VESA
*-compliant stand accessory
2 Power cord (connected to
bottom of docking station)
3 Docking station 6 Docking station power indicator (illuminates blue
2-4 Operator’s Manual
4 Monitor indicators (power/battery, hard disk activ-
ity, wireless)
5 Monitor
when the docking station is receiving AC power)
Monitor - Left Side
Product Views
Figure2-3.Monitor - Left Side
1 Lever for adjusting monitor angle 4 USB 2.0 port
2 I/O port cover with latch (slide to access ports inside) 5 Docking station
3 USB 3.0 port
Operator’s Manual 2-5
Product Overview
Monitor - Right Side
Figure2-4.Monitor - Right Side
1 Docking station 3 Lever for adjusting monitor angle
2 Connector for preamplifier (x2)
2-6 Operator’s Manual
Monitor - Back
Product Views
Figure2-5.Monitor - Back
1 Docking station 4 Programmable buttons (not used)
2 Connectors for preamplifiers 5 Lever for adjusting monitor angle
3 Power button 6 Speaker
Operator’s Manual 2-7
Product Overview
Docking Station
Figure2-6.Docking Station - Bottom View
1 Power input connection (DC in jack) 4 Serial port (RS-232)
2 LAN port (not used, disabled) 5 USB 2.0 port (x2)
3 VGA port
2-8 Operator’s Manual
Figure2-7.Docking Station - Connector for Monitor
Product Views
1 Docking station 2 Connector for monitor
Operator’s Manual 2-9
Product Overview
Monitoring Screen
Figure2-8. Sample Monitoring Screen Elements
1 Message area Provides information about alarm conditions. The background indicates the
severity of the condition. If alarm audio is silenced, an alarm audio silenced symbol appears next to the alarm message. See Alarm Management, page 4-
35. If a “?” appears next to the alarm message, press the “?” for additional information.
N/A Dismiss button
(not shown)
2 Sensor icons Each sensor icon has a label indicating its location on the patient. See Chang-
3 Date and time Current date in MM/DD/YYYY format (by default) and current time in 12-
2-10 Operator’s Manual
Available in the message area when an alarm is dismissible. See Dismissing
an Alarm, page 4-43.
ing the On-Screen Sensor Labels, page 4-21.
hour HH:MM:SS format (by default). The date and time format can be changed by an authorized technician.
Product Views
4 Battery status
indicator
Indicates the remaining battery capacity:
Charged battery —
The level in the icon decreases as battery power
is used.
Low battery —
A medium-priority alarm occurs when the monitor­ing system is operating on battery power and the battery has a charge of <33% remaining. The alarm message BATTERY LOW appears.
Critically low battery —
A medium-priority alarm occurs when the monitoring system is operating on battery power and the battery has a charge of <5% remaining. The alarm message BATTERY CRITICALLY LOW appears. When no charge remains, the monitoring system auto­matically shuts down.
5
Battery charging indicator
AC power indicator
Indicates that the monitoring system is connected to AC power and the battery is charging.
When the battery is fully charged and the monitoring system remains con­nected to AC power, the AC power indicator replaces the battery charging indicator.
6 CURRENT The most recent rSO2 reading from each sensor.
7 CHANGE The percent change between the current rSO
reading and the established
2
baseline for each sensor, color-coded to the corresponding sensor icon.
8 Sensor/readings in
alarm state
If a sensor alarm or loss of readings occurs, the corresponding sensor is high­lighted on the screen. If readings are lost, the CURRENT and CHANGE values are replaced by dashes. Check the message area for information about the condition.
9 Alarm limits The current alarm limits for each sensor, color-coded to the corresponding
sensor icon. A horizontal red line indicates each alarm limit in the trend view graph. See Setting Alarm Limits, page 4-38.
10 Event mark Event marks indicate significant occurrences during monitoring. Event marks
can be added to the trend graph at any time during monitoring and are dis­played as vertical lines with flags. See Event Marks, page 4-44.
11 Baseline values The current rSO
baseline established for each tissue region being moni-
2
tored, color-coded to the corresponding sensor icon. Dotted lines in the trend view graph also indicate the baseline values, color-coded to the corre­sponding sensor icon. See Baselines, page 4-24.
Operator’s Manual 2-11
Product Overview
12 Alarm Audio button Indicates whether alarm audio is on, silenced, or paused. Press to silence or
pause alarm audio or to turn alarm audio back on. When an alarm occurs and alarm audio is silenced or paused, the button’s color corresponds to the highest-level alarm. See Alarm Indicators, page 4-35, and Silencing or Pausing
Alarms, page 4-41.
13 MARK EVENT
button
Press to add an event mark to the trend graph to indicate a significant occur­rence during monitoring. Choose from a customizable list of events. See
Event Marks, page 4-44.
14 Patient button While monitoring, press to return to the Set-up screen, reposition on-screen
sensors, or assign or modify a patient ID. See Set Up for Patient Monitoring, page 4-11, Repositioning On-Screen Sensors, page 4-17, and Assigning or Mod-
ifying the Patient ID, page 4-19.
15 Trend view The trend view shows the progression of rSO
case. Data is color-coded to the corresponding sensor icon. rSO
values over the course of a
2
values are
2
presented on the vertical (y) axis. Time is presented on the horizontal (x) axis. See Trend View Management, page 4-27.
16 MENU button Press to access a variety of settings and functions depending on whether or
not sensors are connected and monitoring has begun. See Menu Structure, page 4-9.
2-12 Operator’s Manual

Product and Carton Label Symbols

2.5
Table2-1.Label Symbol Descriptions
Symbol Description Symbol Description
Reference code (part number) Serial number
Product and Carton Label Symbols
Lot number
Prescription only US federal law restricts this device to
sale by or on the order of a physician
Manufacturer Date of manufacture
Fragile Keep dry
Must consult instructions for use Consult instructions for use
Atmospheric pressure limits (see Envi-
ronmental Conditions, page 11-4)
Humidity limits (see Environmental
Conditions, page 11-4)
European Community (EC) authorized representative
Protection against fluid ingress: Protect­ed against vertically falling water drops when tilted to 15º
Temperature limits (see Environmental
Conditions, page 11-4)
Proper waste disposal for electrical and electronic equipment
MR unsafe - Do not use during magnet­ic resonance imaging
CSA – Canadian Standards Association certification mark
Not made with natural rubber latex
Operator’s Manual 2-13
Defibrillation-proof type BF applied part
CE – Conformité Européene authoriza­tion mark
0123 – TÜV SÜD Product Service GmbH (notified body)
Electromagnetic interference may occur in the vicinity of equipment marked with this symbol
Product Overview
Table2-1.Label Symbol Descriptions (Continued)
Symbol Description Symbol Description
Quantity included in package (exam­ple: 20)
Do not immerse
2-14 Operator’s Manual
3 Installation

3.1 Overview

This chapter contains information for the installation and set up of the INVOS™ patient monitor (the “monitoring system”) prior to first-time use.
Safety Reminders, page 3-1
Unpacking and Inspection, page 3-2
Power Options, page 3-3
Setup, page 3-7

3.2 Safety Reminders

WARNING:
Inspect the monitoring system and all accessories before use to ensure there are no signs of physical damage or improper function. Do not use if damaged.
WARNING:
Shock hazard — Ensure the monitoring system is properly grounded when operating on AC power.
WARNING:
If you do not hear a tone at system start-up, discontinue use of the monitoring system and contact Medtronic or a local Medtronic representative.
Caution:
When operating medical electrical equipment, special precautions related to electromagnetic compatibility (EMC) are required. Install the monitoring system according to the EMC information included in this manual.
3-1
Installation
Caution:
Make sure all connectors are fully engaged and free from moisture. Moisture intrusion may cause poor performance or no readings at all.

3.3 Unpacking and Inspection

The monitoring system ships in multiple cartons. Upon receipt of the monitoring system, examine the cartons for evidence of damage. Contact Medtronic Technical Services immediately if a carton appears damaged. See Technical Services, page 1-7.
Standard items shipped are listed in Table3-1. Quantities of some items may vary based on customer requirements. Check all items for any sign of damage.
If you need to return the monitoring system for any reason, contact Medtronic Tech­nical Services. See
Technical Services, page 1-7
.
Note:
Save the cartons and packing material for any future transport.
Table3-1.Monitoring System - Standard Shipped Items
Item Quantity
Patient monitor 1
Lithium-ion battery 1
Patient monitor stand 1
VESA
™*
docking station
Preamplifiers 1 or 2
Reusable sensor cables (RSCs) 2 or 4
INVOS
adult rSO2 sensors
AC power supply 1
Power cord 1
Operator’s manual (on USB flash drive)
1. The manual may be printed from the included USB drive. Order a printed Operator’s Manual or Service Manual
from Medtronic Technical Services or your local Medtronic representative. See Technical Services, page 1-7.
1
1
Box of 5, 10, or 20
1
3-2 Operator’s Manual

Power Options

3.4
The monitoring system operates on AC power or on a rechargeable internal battery. Medtronic recommends operating the monitoring system on AC power whenever possible. Use battery power only when necessary, such as while moving the patient within the facility or during AC power interruptions.

3.4.1 AC Power

WARNING:
Shock hazard — Ensure the monitoring system is properly grounded when operating on AC power.
The power input connection for the monitoring system is at the bottom of the docking station. Always use the provided AC power cord and power supply.
Power Options
Figure3-1.Power Input Connection at Bottom of Docking Station
Operator’s Manual 3-3
Installation
When the power supply is connected to the AC outlet and the docking station, the power indicator on the docking station illuminates blue (Figure 3-2).
Figure3-2.Docking Station Power Indicator
3-4 Operator’s Manual
AC power to the monitoring system is indicated as follows:
Table3-2.AC Power Indicators
Status Monitor LED Screen icon
AC connected (battery charging)
AC connected (battery fully charged)
Power Options
AC not connected

3.4.2 Battery Power

Note:
Whenever the monitoring system is connected to AC power, the battery will charge as needed.
The monitoring system includes a rechargeable lithium-ion battery located in the back of the monitor.
One of the following, depending on battery charge level. See Table 3-3, page 3-6.
Operator’s Manual 3-5
Installation
Figure3-3.Battery Location at Back of Monitor
Medtronic strongly recommends using AC power during continuous operation. Use battery power only when necessary, such as during transport within a facility. If the monitoring system is operating on battery power, alarms are generated if the battery reaches low or critical levels. When the battery is depleted, the monitoring system shuts down immediately.
Battery Status Indicators
If using the monitoring system on battery power, be sure to frequently check the battery status (Table 3-3).
Table3-3.Battery Status Indicators - Monitoring System Operating on Battery Power
Battery status Screen icon Audible alarm Message
Charged (33% - 100% remaining)
Low (<33% remaining)
Critically Low (<5% remaining)
N/A N/A
Medium priority
Medium priority
BATTERY LOW
BATTERY CRITICALLY LOW
Malfunctioning Low priority - no
audible alarm
3-6 Operator’s Manual
BATTERY FAILURE
Battery Life
Caution:
Dispose of the battery in accordance with local guidelines and regulations.
For maximum battery reliability, replace the battery on a regular basis. Medtronic recommends replacing the battery once every 2 years. If the battery will not sustain monitoring for more than approximately 1 hour when fully charged, replace the battery immediately.
Contact Medtronic or a local Medtronic representative for information about obtain­ing replacement batteries.

3.5 Setup

See
Technical Services, page 1-7.
Setup
Use the following instructions to prepare the monitoring system for first use. See
Operation, page 4-1 for additional steps to be performed by the clinician when mon-
itoring a patient.
Install the Battery, page 3-7
Insert the Monitor into the Docking Station, page 3-9
Connect the Preamplifier(s), page 3-11
Connect the Reusable Sensor Cables (RSCs), page 3-12
Apply Power, page 3-13
Set Institutional Defaults, page 3-16
Power Off the Monitoring System, page 3-17
Position the Monitoring System in the Clinical Setting, page 3-17

3.5.1 Install the Battery

The monitoring system’s lithium-ion battery is shipped with the monitor, but not installed. Install and fully charge the battery prior to clinical use (see Apply Power, page 3-13 for charging information).
To install the battery:
1. Remove the battery from its packaging.
Operator’s Manual 3-7
Installation
2. At the back of the monitor, make sure that the battery latch, located above the battery
slot, is to the right (Figure 3-4).
Figure3-4.Battery Slot and Latch at Back of Monitor
1 Battery 3 Battery latch
2 Battery slot
3. Insert the battery into the slot, bottom edge (with three tabs) first.
4. With the battery completely inserted, slide the battery latch to the left to lock the
battery in place (Figure 3-5).
3-8 Operator’s Manual
Figure3-5.Battery Installed in Monitor
Setup

3.5.2 Insert the Monitor into the Docking Station

The docking station is intended to hold the monitor during typical use of the mon­itoring system. You can mount the docking station on the provided stand or on other VESA™*-compatible equipment capable of supporting at least 5.1 lb (2.3 kg) (VESA™* FDMI MIS-D, 75).
Monitor stand use: If you intend to use the monitor stand, attach the docking station to the stand using the provided hardware. Refer to the instructions provided with the monitor stand.
To insert the monitor into the docking station:
1. Align the connector at the bottom of the monitor with the connector in the docking
station (Figure 3-6).
2. Press the monitor down into the docking station until the connectors engage. Pegs and
guide holes in the docking station and monitor ensure proper alignment.
3. Press the monitor back into the clip at the top of the docking station until it fully engag-
es.
Operator’s Manual 3-9
Installation
Figure3-6.Monitor Mounted in Docking Station
3-10 Operator’s Manual
3.5.3

Connect the Preamplifier(s)

Caution:
Make sure all connectors are fully engaged and free from moisture. Moisture intrusion may cause poor performance or no readings at all.
To connect the preamplifier(s) to the monitor:
1. Align the red dot on the preamplifier cable connector with the red marking on the
monitor input connection (Figure 3-7).
Setup
Figure3-7.Connecting the Preamplifier Cable
2. Insert the cable connector directly upward into the connection until the locking sleeve
rotates and clicks into place.
3. Repeat steps if using a second preamplifier.
Note:
To disconnect a preamplifier cable from the monitor, grasp the locking sleeve and rotate it in the direction indicated by the arrow while pulling downward.
Operator’s Manual 3-11
Installation
3.5.4

Connect the Reusable Sensor Cables (RSCs)

Each preamplifier allows connection of up to two sensors via reusable sensor cables (RSCs). Using two preamplifiers, the monitoring system can accommodate up to four RSCs and four sensors.
To connect the RSCs:
1. Align the RSC’s male connector with the connection slot on the preamplifier. The con-
nector and slot are keyed to guide insertion.
2. Press firmly until the connector snaps into place. Ensure that the clip on the connector
engages completely with the connection slot.
Figure3-8.Connecting the RSC to the Preamplifier
3. Repeat these steps for each RSC to be used.
4. Coil and secure the RSC cabling to avoid tangling.
3-12 Operator’s Manual
3.5.5

Apply Power

WARNING:
If you do not hear a tone at system start-up, discontinue use of the monitoring system and contact Medtronic or a local Medtronic representative.
Before applying power, refer to Power Options, page 3-3 to understand the require­ments for using AC power and battery power.
To power on the monitoring system:
1. Ensure that the AC outlet is properly grounded and supplies the specified voltage and
frequency. See Electrical, page 11-3 for voltage and frequency specifications.
2. Connect the power supply to the power input connection on the bottom of the
docking station.
Setup
Figure3-9.Power Input Connection at Bottom of Docking Station
3. Connect the AC power cord to the power supply and the AC outlet. Verify that the blue
power indicator on the docking station is illuminated (Figure 3-2 on page 3-4).
The battery indicator at the top left of the monitor indicates the charge status. The battery will charge as necessary when the monitoring system is connected to AC power.
Charging
Fully charged
4. Press the Power button on top of the monitor.
Operator’s Manual 3-13
Installation
Figure3-10.Power Button on Top of Monitor
While the monitoring system performs its power-on self-test (POST), a progress bar appears at the bottom of the screen.
5. Ensure that the POST pass tone sounds when POST completes. See POST Pass Tone,
page 11-6 for tone specifications.
The POST pass tone is an audible confirmation of proper speaker performance. If the speaker does not function, alarm warning sounds will not be audible.
Once POST is complete, the Set-up screen appears (Figure 3-11).
3-14 Operator’s Manual
Figure3-11.Set-Up Screen - Preamplifier(s) Connected
Setup
Note:
If no preamplifiers are connected to the monitoring system, the screen prompts you to connect them (Figure 3-12). See Connect the Preamplifier(s), page 3-11.
Note:
Make sure that the battery is fully charged prior to clinical use. Table 3-2, page 3-5, describes the charging status indicators.
Figure3-12.Set-Up Screen - No Preamplifier(s) Connected
Operator’s Manual 3-15
Installation
3.5.6

Set Institutional Defaults

Table 3-4 describes the institutional settings available for the monitoring system.
Institutional defaults must be set by an authorized technician. Refer to the monitor­ing system's service manual. Refer to Table 4-2 on page 4-7 for information about additional options that can be set or changed by the clinician.
Table3-4.Institutional Settings
Item Available settings Factory default
Silence alarms at startup
Alarm audio off/pause duration
Alarm reminder signal
Line frequency 50 Hz, 60 Hz
POST during sensor off ON, OFF
Date (Select from calendar)
Date format DD MM YYYY, YYYY MM DD, MM DD
Time (Select hh:mm)
Time format 24 hr, 12 hr
Language DANISH, DUTCH, ENGLISH, FRENCH,
YES, NO NO - Alarms will sound unless the user
INDEFINITE, 2 MINS INDEFINITE - When silenced by the
ON, OFF
YYYY
GERMAN, GREEK, HUNGARIAN, ITAL­IAN, NORWEGIAN, POLISH, PORTU­GUESE, ROMANIAN, RUSSIAN, SLOVAK, SPANISH, SWEDISH
presses the Alarm Audio button
user, alarms remain silent until the user unsilences them
OFF
60 Hz
OFF For system diagnostics only. Do not
set to ON when used in a clinical set­ting.
Coordinated Universal Time (UTC)
MM DD YYYY
N/A
12 hr
ENGLISH
Serial port OFF, PC LINK 1, PC LINK 2, CLINICAL
TEST SETUP, VUE LINK
3-16 Operator’s Manual
OFF
3.5.7

Power Off the Monitoring System

To power off the monitoring system:
1.
Press the Power button on top of the monitor.
2. Observe that the screen goes completely dark. The battery indicator on the monitor and
the power indicator on the docking station remain illuminated as long as power is con­nected.

3.5.8 Position the Monitoring System in the Clinical Setting

WARNING:
As with all medical equipment, carefully route patient cabling to reduce the possibility of patient entanglement or strangulation.
Setup
Users may choose from a variety of mounting configurations, including the provid­ed monitor stand or other VESA™*-compatible mounting options capable of sup­porting at least 5.1 lb (2.3 kg) (VESA™* FDMI MIS-D, 75). In addition, the preamplifiers have a fold-away swivel hook for attachment to rails or poles (Figure 3-13).
Figure3-13.Monitoring System on Stand with Preamplifier Hung on Rail
Operator’s Manual 3-17
Installation
Be sure to consider the following when positioning the monitoring system:
Docking station use — The docking station is intended to hold the monitor during
typical use and provides the power connection for the monitoring system. Make sure the docking station is accessible at all times. See Insert the Monitor into the Docking Sta-
tion, page 3-9.
Monitor stand use — When using the monitor stand, place the stand on a sturdy flat
surface. Use the lever on the back of the stand to adjust the angle of the monitor for best visibility.
Power outlet access and power cord position — Ensure that the power outlet used
for the monitoring system is easily accessible.
Cable routing — Position the monitoring system for ease of access to all cabling. You
can use the fold-away hooks on the preamplifiers to hang them from rails or poles. Ensure that cables are routed so that there is no risk of patient entanglement or stran­gulation.
Monitor access and visibility — Ensure that the operator can easily access and view
the monitor while monitoring. Choose a location that allows visual communication of rSO
values and alarms. See Physical Characteristics, page 11-2 for visibility specifications.
2
3-18 Operator’s Manual
4 Operation

4.1 Overview

This chapter explains how to use the INVOS™ patient monitor (the “monitoring sys­tem”) to view and collect patient regional oxygen saturation (rSO
This chapter assumes that the monitoring system has been installed at the location of use and tested by the institution. See Installation, page 3-1 for full installation instructions.
Safety Reminders, page 4-2
Quick Start, page 4-5
Operational Defaults, page 4-7
Menu Structure, page 4-9
Monitoring System Memory, page 4-9
) data.
2
Set Up for Patient Monitoring, page 4-11
Optional Set-Up Tasks, page 4-16
Baselines, page 4-24
Trend View Management, page 4-27
Alarm Management, page 4-35
Event Marks, page 4-44
Area Under the Curve (AUC), page 4-49
Finish Monitoring, page 4-51
Case Histories, page 4-52
4-1
Operation

Safety Reminders

4.2
WARNING:
The monitoring system should not be used as the sole basis for diagnosis or therapy. It is intended only as an adjunct in patient assessment. It must be used in conjunction with clinical signs and symptoms.
WARNING:
Inspect the monitoring system and all accessories before use to ensure there are no signs of physical damage or improper function. Do not use if damaged.
WARNING:
Explosion hazard — Do not use the monitoring system in the presence of flammable anesthetics.
WARNING:
Always disconnect and remove the monitoring system and sensors during magnetic resonance imaging (MRI) scanning. Attempting to use the monitoring system during an MRI procedure could cause burns or adversely affect the MRI image or the monitoring system's performance.
WARNING:
As with all medical equipment, carefully route patient cabling to reduce the possibility of patient entanglement or strangulation.
WARNING:
Do not set alarm limits to extreme values that render the monitoring system ineffective. Ensure alarm limits are appropriate for each patient.
WARNING:
Do not silence, pause, or decrease the volume of audible alarms if patient safety could be compromised.
WARNING:
If you do not hear a tone at system start-up, discontinue use of the monitoring system and contact Medtronic or a local Medtronic representative.
4-2 Operator’s Manual
WARNING:
Monitoring system readings can be affected by certain patient conditions. Refer to
Patient Conditions, page 6-2.
WARNING:
Failure to cover the sensor site with opaque material when operating under high ambient light conditions may result in poor performance.
WARNING:
Electromagnetic emissions from the monitoring system may interfere with other critical devices.
WARNING:
Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of the monitoring system, including cables. Otherwise, degradation of monitoring system performance may result.
Safety Reminders
WARNING:
The use of accessories, sensors, and cables other than those specified may result in poor performance of the monitoring system and increased electromagnetic emissions or decreased electromagnetic immunity of the monitoring system.
WARNING:
The monitoring system is intended for use by healthcare professionals only. It may cause radio interference or may disrupt the operation of nearby equipment. Mitigation for such disruption may require re-orienting or relocating the monitoring system or shielding the location.
WARNING:
Any radio frequency transmitting equipment or other nearby sources of electrical noise may result in disruption of the monitoring system.
WARNING:
The monitoring system is designed for use in environments in which the signal can be obscured by electromagnetic interference. During such interference, measurements may seem inappropriate or the monitoring system may not seem to operate correctly.
Operator’s Manual 4-3
Operation
WARNING:
EMI disruption can cause cessation of operation or other incorrect functioning.
WARNING:
The monitoring system should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is necessary, observe the monitoring system to verify normal operation in the desired configuration. Technical alarms may indicate that the configuration is not appropriate for the monitoring system.
Caution:
If two sensors are placed in close proximity to each other on a patient, the same preamplifier should be connected to both sensors to avoid poor performance.
Caution:
Make sure all connectors are fully engaged and free from moisture. Moisture intrusion may cause poor performance or no readings at all.
Caution:
The monitoring system generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to other devices in the vicinity. If interference is suspected, move monitoring system cables away from the susceptible device.
Caution:
The use of an electrosurgical or electrocautery instrument in the vicinity of the monitoring system may interfere with the signal and cause poor performance or no readings at all.
Refer to Safety Information, page 1-2 for additional warnings and cautions regarding use of the monitoring system.
4-4 Operator’s Manual

Quick Start

4.3
If you are familiar with operating the monitoring system, follow the steps in Table 4-
1 to set up the device and monitor a patient. Be sure you have reviewed the infor-
mation in Safety Reminders, page 4-2 before monitoring a patient.
If you are not familiar with operating the monitoring system, carefully read all instructions in this chapter, beginning with Set Up for Patient Monitoring, page 4-11.
Table4-1.Quick Start Steps
1
Make sure all components of the monitoring system are present:
Monitor, docking station, power supply, power cord, preamplifier(s), reusable sensor cables (RSCs), INVOS™ rSO
sensors
2
2
Attach up to four INVOS™ rSO2 sensors to the patient
See Set Up for Patient Monitoring, page 4­11
See the Instructions for Use provided with the sensors
Quick Start
3
Power on the monitoring system See Set Up for Patient Monitoring, page 4-
11
4
Optionally, change the sensor placement sequence See Changing the Sensor Placement
Sequence, page 4-16
5
Connect the sensors to the RSCs, following the on­screen prompts, and verify the sensor locations
6
Secure all cables to avoid patient skin injury or entanglement
7
Optionally, enter a patient ID See Assigning or Modifying the Patient ID,
8
Begin monitoring:
- For a new case, press NEW DATA SET
- To continue a case, press APPEND OLD DATA
9
Set baselines
Press MENU, SET BASELINES
10 Set or change alarm limits, as desired
Press MENU, ALARM LIMITS
11
Optionally, change the trend view:
- Number of graphs
- Trend-line averaging
- Time scale
See Set Up for Patient Monitoring, page 4­11
See Set Up for Patient Monitoring, page 4­11
page 4-19
See Set Up for Patient Monitoring, page 4­11
See Baselines, page 4-24
See Setting Alarm Limits, page 4-38
See Trend View Management, page 4-27
Operator’s Manual 4-5
Operation
Table4-1.Quick Start Steps (Continued)
12 Optionally, change the on-screen sensor labels See Changing the On-Screen Sensor Labels,
page 4-21
13 Optionally, check and modify the list of event
markers
14 Optionally, change Area Under the Curve (AUC)
thresholds
15
Optionally, adjust the screen brightness See Changing the Screen Brightness, page
16
Optionally, adjust the alarm volume See Changing the Alarm Volume, page 4-40
17
Respond to alarms, as necessary:
- Alarm indicators
- Alarm message lookup
- Silence or pause alarms
- Dismiss alarms
18
Mark events, as desired See Event Marks, page 4-44
19
Finish monitoring See Finish Monitoring, page 4-51
See Changing the Event List, page 4-47
See Changing the AUC Threshold, page 4­50
4-23
See:
- Alarm Indicators, page 4-35
- Alarm Messages, page 8-2
- Silencing or Pausing Alarms, page 4-41
- Dismissing an Alarm, page 4-43
4-6 Operator’s Manual

Operational Defaults

4.4
The following table lists default settings for the monitoring system. Some of these settings can be changed by the user, while others require authorized access, as indi­cated.
Table4-2.Operational Defaults
Item Available settings Default
Upper alarm limits OFF, ON OFF
See Setting Alarm Limits, page 4-38
Lower alarm limits MANUAL, AUTO MANUAL
See Setting Alarm Limits, page 4-38
Operational Defaults
Upper rSO
Lower rSO
Silence alarms at startup
Alarm audio off/ pause duration
Alarm reminder signal ON, OFF OFF - No reminder when alarms are silenced
Line frequency 50 Hz, 60 Hz 60 Hz
limit 20 - 95 If ON, default is 90
2
See Setting Alarm Limits, page 4-38
limit MANUAL: 15 - 90
2
AUTO: 5 - 30
YES, NO
INDEFINITE, 2 MINS
MANUAL: 40 AUTO: 20 See Setting Alarm Limits, page 4-38
NO - Alarm audio is on by default, but alarm audio may be silenced or paused by the user. See
Silencing or Pausing Alarms, page 4-41.
This setting may be changed by an authorized technician.
INDEFINITE - When silenced, alarms remain silent until the user unsilences them. See Alarm Silence
Duration, page 4-42.
This setting may be changed by an authorized technician.
This setting may be changed by an authorized technician.
This setting may be changed by an authorized technician.
Alarm volume 1 - 10 5
See Changing the Alarm Volume, page 4-40
Screen brightness 1 - 10 7
See Changing the Screen Brightness, page 4-23
Operator’s Manual 4-7
Operation
Table4-2.Operational Defaults (Continued)
Item Available settings Default
Time format 24 hr, 12 hr 12 hr
This setting may be changed by an authorized technician.
Date format DD MM YYYY, YYYY MM DD,
MM DD YYY
Language Danish, Dutch, English,
French, German, Greek, Hun­garian, Italian, Norwegian, Polish, Portuguese, Romanian, Russian, Slovak, Spanish, Swedish
Trend line averaging ON, OFF OFF
Trend axis view TWO AXIS VIEW, ONE AXIS
VIEW
Trend time scale (x-axis)
Sensor placement sequence
AUC threshold type FIXED
1, 2, 4, 8, 12, 24 hours
4, 3, or 2 sensors 4 sensors
% BELOW BASELINE
MM DD YYYY This setting may be changed by an authorized
technician.
English This setting may be changed by an authorized
technician.
See Turning Trend Line Averaging On/Off, page 4­30
ONE AXIS VIEW See Changing the Trend View - Two Graphs or One
Graph, page 4-29
1 hour See Zooming the Trend View, page 4-32
See Changing the Sensor Placement Sequence, page 4-16
FIXED See Changing the AUC Threshold, page 4-50
AUC threshold FIXED: 30 - 60
% BELOW BASELINE: 0 - 30%
Serial port OFF, PC LINK 1, PC LINK 2,
CLINICAL TEST SETUP, VUE LINK
4-8 Operator’s Manual
FIXED: 50 % BELOW BASELINE: 25% See Changing the AUC Threshold, page 4-50
OFF This setting may be changed by an authorized
technician.

Menu Structure

4.5
MENU ALARM LIMITS UPPER LIMITS
Menu Structure
Table4-3.Menu Structure
LOWER LIMITS
SETTINGS GENERAL SCREEN BRIGHTNESS
SENSOR PLACEMENT SEQUENCE
ALARMS ALARM VOLUME
TRENDS TREND LINE AVERAGING
TWO AXIS VIEW/ONE AXIS VIEW
EVENTS
CASE HISTORIES
SET BASELINES
AREA UNDER CURVE
1. Available only when no sensors are connected to the monitoring system.
2. Available only when monitoring has begun.

4.6 Monitoring System Memory

1
2
2
RESET AUC
THRESHOLD
The monitoring system stores the following information about the current case in memory:
Case ID
Current baselines (value and time baselines were obtained)
Current sensor positions:
The monitoring system remembers the last sensors that were connected, the loca-
tions they were assigned (for example, left cerebral), and their last baseline values.
The monitoring system also remembers the last reusable sensor cable (RSC) and
preamplifier port assignments (for example, which RSC and preamplifier the left cerebral sensor was connected to).
Operator’s Manual 4-9
Operation
The previously used sensor location overrides the last RSC/preamplifier port assign-
ment. For example, if you plug what was the right cerebral sensor into what was the left cerebral RSC/preamplifier port, it will show up on screen as the right cerebral sensor.
Because the monitoring system retains this information for an ongoing case, you can disconnect sensors from the system and reconnect the same sensors without having to reassign sensor locations or perform the sensor set-up routine.
Table 4-4 describes some common situations, the actions you should perform, and
how the system reacts.
Table4-4.System Memory - Common Situations and Recommended Actions
When this situation occurs... Perform this action...
You have started monitoring a patient but notice that you have the sensor locations swapped.
Sensors need to be disconnected from the monitoring system during a case (for example, to reroute the cables).
The monitoring system is powered off and back on during a case
A case has ended. Unplug and discard all sensors that were used during the case.
.
Navigate to the sensor Set-up screen (see Repositioning On-Screen
Sensors, page 4-17), drag and drop the sensors to the correct loca-
tions on the screen, confirm the new locations, then press APPEND OLD DATA to resume monitoring.
Disconnect and reconnect the affected sensors. The monitoring system automatically assigns the sensors back to the proper loca­tions.
No action is necessary. The monitoring system automatically assigns sensors to the proper location and resumes monitoring.
Dismiss the “Sensor Not Connected” alarm. The monitoring system returns to the Set-up screen.
4-10 Operator’s Manual

Set Up for Patient Monitoring

Set Up for Patient Monitoring
4.7
The following steps assume you are ready to begin monitoring a patient in a surgical setting or other hospital environment. Be sure you have reviewed the information in
Safety Reminders, page 4-2 before monitoring a patient.
To set up the monitoring system:
1. Make sure all components of the monitoring system are present:
Monitor
Docking station - Recommended for extended monitoring.
Power supply and power cord - AC power is recommended for extended moni-
toring. You can use battery power briefly, if necessary, such as when transporting a patient between locations within a hospital.
Preamplifiers - Use one or two preamplifiers, depending on the number of sites to
be monitored. Each preamplifier accommodates two sensors. If no preamplifiers are connected, see Connect the Preamplifier(s), page 3-11 for instructions.
Reusable sensor cables (RSCs) - Use one RSC per sensor, up to two per preampli-
fier. If no RSCs are connected, see Connect the Reusable Sensor Cables (RSCs), page 3­12, for instructions.
INVOS™ rSO
sensors - Before use, carefully read the sensor Instructions for Use,
2
including all warnings, cautions, and instructions.
2. Determine the sites you will monitor on the patient. Select the correct type of INVOS™
rSO
sensor for your patient; do not mix sensor types. Attach up to four sensors to the
2
patient. Refer to the Instructions for Use provided with the sensors for application instruc­tions.
3. Power on the monitoring system:
a. Make sure that the monitoring system’s power supply is plugged into the docking
station and a properly grounded hospital mains outlet (see AC Power, page 3-3). You may operate the monitoring system briefly on battery power if necessary, but AC power is recommended for extended monitoring. See Power Options, page 3-3 for additional information about powering the monitoring system.
b. Press the Power button on top of the monitor (see Apply Power, page 3-13). If oper-
ating on battery power, check the battery status indicator on the screen to deter­mine whether the battery needs charging (see Table 3-3, page 3-6).
Operator’s Manual 4-11
Operation
c. Ensure that the POST pass tone sounds when POST completes. See POST Pass Tone,
page 11-6 for tone specifications.
The POST pass tone is an audible confirmation of proper speaker performance. If the speaker does not function, alarm warning sounds will not be audible.
Once POST is complete, the Set-up screen appears (Figure 4-1).
Figure4-1.Set-Up Screen - Prompt for Sensor Connection
4. Optionally, you can set the sensor sequence for the number of sensors you are using. If
you are using fewer than four sensors, you can set the number to three or two for on­screen representations. See Changing the Sensor Placement Sequence, page 4-16.
5. Connect the RSCs to the sensors that have been applied to the patient:
Caution:
If two sensors are placed in close proximity to each other on a patient, the same preamplifier should be connected to both sensors to avoid poor performance.
a. Note the highlighted sensor location on screen (Figure 4-1). Locate the correspond-
ing sensor applied to the patient.
b. Look for the flashing blue light on the preamplifier and at the end of the corre-
sponding RSC.
c. Align the sensor’s male connector with the connection slot on the RSC. The connec-
tor and slot are keyed to guide insertion.
d. Press firmly until the connector snaps into place. The monitoring system indicates
proper connection by displaying an rSO
reading at the corresponding sensor loca-
2
4-12 Operator’s Manual
e. Look for the next highlighted sensor location on screen and the next flashing blue
f. Repeat these steps for each sensor applied to the patient.
g. Verify the placement of all sensors by briefly pressing each on-screen sensor loca-
Note:
As you connect the RSCs to sensors applied to the patient, readings are displayed on the Set-up screen. However, trends are not tracked and physiological alarms are disabled. Do not attempt to monitor the patient from the Set-up screen.
Set Up for Patient Monitoring
tion. If the sensor was previously used on the monitoring system, the last baseline obtained on the sensor is also displayed.
light on the preamplifier and RSC. Connect the next sensor to the corresponding RSC.
tion and noting the blue flashing light on the RSC cable. Make sure that the on­screen location matches the sensor location on the patient. If the location does not match, you can reposition the on-screen sensors rather than disconnecting the RSCs. See Repositioning On-Screen Sensors, page 4-17.
6. Check the location of all cables connected to the monitoring system. Make sure that the
patient is not lying on any cables or connectors. To prevent entanglement and pro­longed contact with patient skin, you can secure the RSCs with the strain-relief clips attached to the cables. Do not place the RSCs, preamplifiers, or cables connecting the preamplifiers to the monitor in contact with the patient.
7. Optionally, enter a patient ID. See Assigning or Modifying the Patient ID, page 4-19.
8. Begin monitoring by starting a new case or continuing an existing case:
For a new case, press NEW DATA SET.
To continue a case, press APPEND OLD DATA.
The monitoring screen appears.
Operator’s Manual 4-13
Operation
Figure4-2.Monitoring Screen - New Case - No Baselines Set
Note:
If necessary, you can return to the Set-up screen while monitoring by pressing the Patient button:
9. Set baselines. See Baselines, page 4-24.
10. Set or change alarm limits, as desired. See Setting Alarm Limits, page 4-38.
11. Optionally, change the trend view. You can view all sensor readings on one graph or
two, use trend line averaging, and change the time scale. See Trend View Management, page 4-27.
12. Optionally, change the on-screen sensor labels. See Changing the On-Screen Sensor
Labels, page 4-21.
13. Optionally, check the list of available event markers and change it if necessary. See
Changing the Event List, page 4-47
14. Optionally, change the Area Under the Curve (AUC) threshold. See Changing the AUC
Threshold, page 4-50.
15. Optionally, adjust the screen brightness. See Changing the Screen Brightness, page 4-23.
16. Optionally, adjust the alarm volume. See Changing the Alarm Volume, page 4-40.
4-14 Operator’s Manual
17. As you monitor the patient, you may need to perform the following tasks:
Respond to alarms - See Alarm Indicators, page 4-35, Alarm Messages, page 8-2,
Mark clinically significant events - See Event Marks, page 4-44.
Note:
In the event of defibrillation, the monitoring system will recover operation without intervention within 30 seconds after the defibrillation voltage is removed.
18. When the monitoring session is complete, follow the steps in Finish Monitoring, page 4-
51.
Set Up for Patient Monitoring
Silencing or Pausing Alarms, page 4-41, and Dismissing an Alarm, page 4-43.
Operator’s Manual 4-15
Operation

Optional Set-Up Tasks

4.8
Changing the Sensor Placement Sequence
Repositioning On-Screen Sensors
Assigning or Modifying the Patient ID
Changing the On-Screen Sensor Labels
Changing the Screen Brightness

4.8.1 Changing the Sensor Placement Sequence

By default, the monitoring system assumes that you will use two cerebral sensors and two somatic sensors per case. You are prompted to connect the sensors in the following order:
Figure4-3.Default Sensor Placement Sequence
If you monitor a different combination of sites, you can change the prompts to one of the following sequences:
4-16 Operator’s Manual
Figure4-4.Optional Sensor Placement Sequences
To change the sensor placement sequence:
1. Press MENU.
Optional Set-Up Tasks
2. Press SETTINGS.
3. Press GENERAL.
4. Under SENSOR PLACEMENT SEQUENCE, press the desired sequence.
5. Press CLOSE.

4.8.2 Repositioning On-Screen Sensors

If the on-screen sensor representation does not correspond to sensor placement on the patient, you can reposition the on-screen sensors before or during monitoring.
To reposition the on-screen sensors before monitoring:
1. At the Set-up screen, briefly press an on-screen sensor icon to determine the corre-
sponding sensor’s location on the patient. The on-screen sensor icon is highlighted. The blue LEDs on the RSC and preamplifier flash to indicate the corresponding sensor.
Operator’s Manual 4-17
Operation
Figure4-5.Highlighting an On-Screen Sensor
2. Drag the on-screen sensor to a new position and release.
3. Confirm the new position by touching each highlighted on-screen sensor.
To reposition the on-screen sensors while monitoring:
1. At the monitoring screen, briefly press a sensor icon next to the CURRENT values to
determine the corresponding sensor’s location on the patient. The blue LEDs on the RSC and preamplifier flash to indicate the corresponding sensor.
Figure4-6.Sensor Icons Next to CURRENT Values (Press to Locate Sensor on Patient)
2. Press the Patient button. The patient icon appears.
4-18 Operator’s Manual
Figure4-7.Patient Icon
Optional Set-Up Tasks
3. Press Sensor Set-up.
4. At the Set-up screen, drag the on-screen sensor to a new position and release.
5. Confirm the new position by touching each highlighted on-screen sensor.
6. Press APPEND OLD DATA to resume monitoring. Note that there will be a brief gap in
the data on the trend graph.

4.8.3 Assigning or Modifying the Patient ID

By default, cases are identified by date and start/end time, but do not have patient IDs assigned. As an option, you can assign an alphanumeric patient ID to a case before or during monitoring. You can also modify a patient ID at any time while monitoring.
Note:
Follow institutional protocols for assigning patient IDs. Avoid using patient-sensitive information in IDs.
To assign a patient ID before monitoring:
1. Set up the monitoring system as described in Set Up for Patient Monitoring, but do not
begin monitoring.
2. Press PATIENT ID. A keyboard appears.
Operator’s Manual 4-19
Operation
Figure4-8.Entering a Patient ID Before Monitoring
3. Type the patient ID.
4. Press CLOSE to dismiss the keyboard.
5. Begin the case by pressing NEW DATA SET.
To assign or modify a patient ID while monitoring:
1. After monitoring begins, press the Patient button. The patient icon appears.
Figure4-9.Patient Icon
2. Press the patient icon. A keyboard appears.
4-20 Operator’s Manual
Figure4-10.Entering a Patient ID During Monitoring
Optional Set-Up Tasks
3. Type the patient ID, or modify the existing ID.
4. Press CLOSE to dismiss the keyboard.
5. Press the Patient button to dismiss the patient icon.

4.8.4 Changing the On-Screen Sensor Labels

By default, sensors are labeled as follows on screen:
Left cerebral sensor
Right cerebral sensor
Right somatic sensor
Left somatic sensor
You can change the sensor labels to any alphabetic value (A - Z) or S1 - S4.
Operator’s Manual 4-21
Operation
Note:
You can verify the location of a sensor by briefly pressing the sensor label next to the CURRENT value on the monitoring screen. The blue LEDs on the RSC light to indicate the corresponding sensor.
To change a sensor label:
1. After monitoring begins, press and hold the sensor label you want to change. A sensor
label menu appears.
Figure4-11.Sensor Label Menu
2. Scroll through the choices by pressing anywhere within the list and dragging up or
down. When you see the choice you want, press it to highlight it. Note that duplicate labels are not permitted.
3. Press CLOSE.
4. Verify that the sensor label has changed.
4-22 Operator’s Manual
4.8.5

Changing the Screen Brightness

By default, the screen brightness is set to 7 with a range of 1 to 10.
To change the screen brightness:
1. Press MENU.
2. Press SETTINGS.
3. Press GENERAL.
Figure4-12.SETTINGS - GENERAL - SCREEN BRIGHTNESS
Optional Set-Up Tasks
4. Press an arrow key to increase or decrease the screen brightness.
5. Press CLOSE.
Operator’s Manual 4-23
Operation

Baselines

4.9
The monitoring system requires an rSO2 baseline for each tissue region being mon­itored so that changes from the baseline can be reported. Changes in rSO2 values of >20% from baseline are considered clinically significant and cause for concern and
possible interventions.
When measured rSO system issues an alarm. You can use the monitoring system’s default alarm limits or
set custom limits, as described in Setting Alarm Limits, page 4-38.
It is recommended that you obtain baselines while the patient is stable and awake (for example, prior to surgical induction). You can set baselines for all sensors at once or for individual sensors. If necessary, you can retake baselines at any time during monitoring.
Automatic Baselines: If you do not actively set baselines, the monitoring system automatically sets them approximately five minutes after monitoring begins. You can use these automatic baselines, or retake the baselines. Be aware that sudden large changes in a patient’s saturation values during the initiation of monitoring can result in unrepresentative automatically calculated baseline values. If you use auto­matic baselines, be sure to check the values to make sure they are appropriate. Man­ually retake the baselines, if necessary.
Note:
If you reposition or replace a sensor during monitoring, be sure to retake the baseline for that sensor to ensure a valid representation of rSO
To set or retake baselines:
1. Set up the monitoring system and begin monitoring as described in Set Up for Patient
values are below or above specific limits, the monitoring
2
at that location.
2
Monitoring, page 4-11. While baselines have not yet been set, the monitoring system
displays rotating arrows for the CHANGE values next to the CURRENT rSO
readings.
2
4-24 Operator’s Manual
Figure4-13.Monitoring Screen - New Data Set - No Baselines Set
Baselines
2. Press MENU.
3. Press SET BASELINES. The SET BASELINES screen either indicates that no baselines have
been set (no value next to “BL” as in Figure 4-14), or, if automatic baselines have taken effect, indicates the values and times they were taken (Figure 4-15).
Figure4-14.SET BASELINES - No Baselines Previously Set
Operator’s Manual 4-25
Operation
Figure4-15.SET BASELINES - Baselines Previously Set
4. Set baselines for an individual sensor or all sensors at once:
Individual sensor - Press the on-screen circle representing the sensor. The baseline
is set to the current rSO
reading at that sensor site. The new baseline value is dis-
2
played with the date and time it was taken.
All sensors - Press RETAKE ALL BASELINES. The baselines are set to the current rSO
readings at all sensor sites. The new baseline values are displayed with the date and time they were taken.
5. Press CLOSE.
The baseline values are indicated to the left of the trend graph, color-coded to the corresponding sensor label. An event marker in the trend graph indicates the point at which the baselines were taken.
The CHANGE values reflect the difference between the baselines and the CURRENT readings.
2
4-26 Operator’s Manual
Figure4-16.Monitoring Screen with Baselines Set
Trend View Management
Note:
If you disconnect and re-connect the same sensor while monitoring, the baseline for that sensor is maintained. If you replace a sensor with a new sensor while monitoring, the baseline calculated with the previous sensor is applied to the new sensor. If the monitoring system is powered off and back on while monitoring, baselines are maintained for all sensors connected to the system.

4.10 Trend View Management

About the Trend View
Changing the Trend View - Two Graphs or One Graph
Turning Trend Line Averaging On/Off
Zooming the Trend View
Reviewing Data that has Scrolled Off the Trend View
Viewing Previous rSO2 Values on the Trend Graph
Operator’s Manual 4-27
Operation
4.10.1

About the Trend View

“Trend view” refers to the data graph on the monitoring screen.
The trend view shows the progression of rSO
values over the course of a case. It pro-
2
vides a visual reference for on-going rSO2 readings in relationship to baseline values and alarm limits, as well as a means to reference significant clinical events during the
case.
The trend view is stored in the monitoring system’s memory and can be reviewed at a later time. See Case Histories, page 4-52.
Figure4-17.Trend View
1
rSO2 scale (y-axis)
2 Current baselines (color-coded to sensor
labels).
3 Trend line (color-coded to sensor labels) 7 Event mark
4 Alarm limits (color-coded to sensor labels)
4-28 Operator’s Manual
5 Alarm limit line
6 Time scale (x-axis)
4.10.2

Changing the Trend View - Two Graphs or One Graph

By default, the monitoring system displays all trends on one graph (Figure 4-18, left). If desired, you can display trends on two graphs - one for cerebral sensors, one for somatic sensors (Figure 4-18, right).
Figure4-18.Trend View - One Graph (Left, Default) and Two Graphs (Right)
Trend View Management
To change the trend view (two graphs vs. one graph):
1. Press MENU.
2. Press SETTINGS.
3. Press TRENDS.
Operator’s Manual 4-29
Operation
Figure4-19.SETTINGS - TRENDS - TWO AXIS VIEW vs. ONE AXIS VIEW
4. Press TWO AXIS VIEW or ONE AXIS VIEW, as desired. Your choice is highlighted by a
white square.
5. Press CLOSE.
Note:
The trend view setting remains in effect across power cycles.

4.10.3 Turning Trend Line Averaging On/Off

Trend line averaging provides a 60-minute rolling average of rSO2 values. Viewing a rolling average can be useful in situations where there is frequent and wide variabil-
ity in rSO the real-time rSO2 values in the graph. The trend line is the same color as the real­time values for each sensor. The numerical rSO2 values and percent change from baseline continue to be displayed in real-time.
values. The averaged data is displayed as a bold line superimposed over
2
4-30 Operator’s Manual
Figure4-20.Trend Line Averaging
Trend View Management
By default, trend line averaging is off.
To turn trend line averaging on/off:
1. Press MENU.
2. Press SETTINGS.
3. Press TRENDS.
Figure4-21.SETTINGS - TRENDS - TREND LINE AVERAGING
4. Press ON or OFF next to TREND LINE AVERAGING, as desired.
Operator’s Manual 4-31
Operation
5. Press CLOSE.
Note:
The trend line averaging setting remains in effect across power cycles.

4.10.4 Zooming the Trend View

By default, the trend view shows 1 hour of data at a time. You can zoom the trend view to see different time intervals. Available intervals are 1, 2, 4, 8, 12, and 24 hours.
Note that only the horizontal axis (time) zooms, not the vertical axis (rSO
To zoom the trend view:
1. To zoom out and see a longer time range, place two fingers within the trend graph, 1-
2 inches apart horizontally, and slide together.
2. To zoom in and see a shorter time range, place two fingers within the trend graph, side-
by-side horizontally, and slide apart.
Note:
The zoom setting remains in effect across power cycles.

4.10.5 Reviewing Data that has Scrolled Off the Trend View

During extended monitoring, trend data scrolls off the trend view to the left. You can review this trend data while monitoring by swiping the trend view to the right.
Current rSO trends are not displayed until you swipe the trend view back to the left. As an indi-
cation that current trends are not shown, “REVIEW MODE” appears on the trend view.
values are still displayed while you review older trends, but current
2
values).
2
4-32 Operator’s Manual
Figure4-22.Trend View Review Mode
Trend View Management
To review data that has scrolled off the trend view:
1. Briefly touch within the trend graph and swipe to the right. The data shifts back in time
as indicated by the time scale. The message “REVIEW MODE” appears.
2. Repeat swipes until the data you want to view appears. Note that the amount of shift
corresponds to the length of the swipe.
3. To return to current trend data, swipe to the left until the message “REVIEW MODE” no
longer appears.
4.10.6 Viewing Previous rSO
Values on the Trend Graph
2
While monitoring a patient, you can view rSO2 values that were recorded earlier in the case. Values appear in a pop-up corresponding to the point of interest (Figure 4-
23).
Operator’s Manual 4-33
Operation
Figure4-23.Trend Graph Pop-Up
To view previous rSO2 values on the trend graph:
1. Press and hold the point of interest on the trend graph. The values at that point appear
in a pop-up that indicates the time they were recorded.
Figure4-24.Monitoring Screen - Trend Graph Pop-Up
2. Slide your finger in either direction on the graph to see values at different times.
3. Release when finished viewing values.
4-34 Operator’s Manual

Alarm Management

4.11
Alarm Indicators
Setting Alarm Limits
Changing the Alarm Volume
Silencing or Pausing Alarms
Dismissing an Alarm
Note:
Refer to Table 8-1 on page 8-2 for a complete list of alarm messages, priorities, and resolutions.
Alarm Management

4.11.1 Alarm Indicators

The monitoring system uses audio and visual indicators to identify alarms.
The message area at the top of the monitoring screen indicates active alarms (Figure 4-25). If multiple alarms occur, the message area shows the color of the highest priority alarm and indicates the total number of alarms currently active. By pressing the arrow in the message area, you can expand the list and view all active alarms.
When an rSO
value crosses an alarm threshold, the background of the affected rSO2
2
reading flashes yellow.
The Alarm Audio button indicates alarm audio status. When alarms are silenced or paused, the Alarm Audio button also indicates alarm status (Table 4-5 on page 4-36).
Operator’s Manual 4-35
Operation
Figure4-25.Visual Alarm Indicators
1 Alarm Audio button 3 rSO2 reading in alarm state
2 Message area
Table4-5.Alarm Audio Button States
Alarm Audio
button
Alarm audio
on, silenced, or paused
On Alarm of any status or no alarms
Silenced No alarms
Alarm status
Low- or medium-priority alarm
4-36 Operator’s Manual
Table4-5.Alarm Audio Button States (Continued)
Alarm Management
Alarm Audio
button
Alarm audio
on, silenced, or paused
Alarm status
Paused (2 minutes) No alarms
Low- or medium-priority alarm
Table 4-6 summarizes the monitoring system’s alarm indicators.
Table4-6.Audible and Visual Alarm Indicators
Priority Audible
1
tone
Medium 3 pulse
burst
Visual indication
Message area: Flashing yellow bar with text message
Battery icon: For battery alarm, flashing yellow; oth­erwise, no change
readings area: Yellow flashing background on
rSO
2
affected reading for sensor alarms (dashes appear if readings are lost)
Alarm Audio button: Yellow if alarms silenced or paused
2
Example messages
BATTERY LOW LOW RSO2
3
Low 2 pulse
burst
Message area: Yellow bar with text message (not flashing)
TREND DATA LOST
Battery icon: No change
readings area: No change
rSO
2
Alarm Audio button: Yellow if alarms silenced or paused
1. By default, alarm audio is on, but may be silenced or paused by the user. See Silencing or Pausing Alarms, page 4-41. See Table 11-
2 on page 11-5 for alarm tone specifications.
2. See Table 11-4 on page 11-5 for alarm visual specifications.
3. See Table 8-1 on page 8-2 for a complete list of alarm messages.
Operator’s Manual 4-37
Operation
4.11.2

Setting Alarm Limits

WARNING:
Do not set alarm limits to extreme values that render the monitoring system ineffective. Ensure alarm limits are appropriate for each patient.
The monitoring system provides default alarm limit settings, as listed in Table4-7.
You can change the alarm limits before or after setting baselines, or you can use the existing alarm limits. Alarm limits are retained across power cycles, so check the current settings to determine if they are appropriate for the patient.
Alarm limits item Available settings Default
Table4-7.Alarm Limits Settings
UPPER LIMITS OFF - No alarm sounds for high rSO2 values
Upper limit range (when ON)
LOWER LIMITS MANUAL - Specify a low rSO
Lower limit range when set to MANUAL
Lower limit range when set to AUTO
Note:
Upper and lower alarm limits for each sensor are not allowed to cross.
To set alarm limits:
1. Press MENU.
2. Press ALARM LIMITS. The SET ALARM LIMITS screen shows the most recent limits set (or
ON - Specify a high rSO
20 - 95 90
AUTO - Specify a percentage below baseline that will generate an alarm
15 - 90 40
-30% to -5% -20%
value that will generate an alarm
2
value that will generate an alarm
2
the defaults if limits have not previously been set).
OFF
MANUAL
4-38 Operator’s Manual
Figure4-26.SET ALARM LIMITS - Default Settings
Alarm Management
Note:
Four sensors are shown even if fewer are in use.
3. To set upper alarm limits:
a. Press ON for UPPER LIMITS. A second scale appears for each sensor.
Figure4-27.SET ALARM LIMITS - Upper Limits On
b. Next to a sensor, press and hold the slider for the UPPER limit, drag it to the desired
value, and release. You can set different values for each sensor.
Operator’s Manual 4-39
Operation
4. To set lower alarm limits:
a. Press MANUAL or AUTO for LOWER LIMITS, as desired.
b. Next to a sensor, press and hold the slider for the LOWER limit, drag it to the desired
value, and release. You can set different values for each sensor.
5. Press CLOSE. The new alarm limits take effect immediately.
Note:
Alarm limit settings remain in effect across power cycles.

4.11.3 Changing the Alarm Volume

WARNING:
Do not silence, pause, or decrease the volume of audible alarms if patient safety could be compromised.
When alarm audio is on, the default alarm volume is set to 5 with a range of 1 to 10. See Table 11-2 on page 11-5 for alarm tone specifications.
To change the alarm volume:
1. Press MENU.
2. Press SETTINGS.
3. Press ALARMS.
4-40 Operator’s Manual
Figure4-28.SETTINGS - ALARMS - ALARM VOLUME
Alarm Management
4. Press an arrow key to increase or decrease the alarm volume.
5. Press CLOSE.
Note:
The alarm volume setting remains in effect across power cycles.

4.11.4 Silencing or Pausing Alarms

WARNING:
Do not silence, pause, or decrease the volume of audible alarms if patient safety could be compromised.
Depending on your institutional default, alarm audio may be on or off when you start a case. When alarm audio is on, you can silence or pause the audio.
A white Alarm Audio button indicates alarm audio is on. You can silence or pause alarms by pressing the button. The button changes to indicate whether alarms are silenced or paused (see Table 4-5 on page 4-36). Whether an alarm is silenced or paused depends on institutional default settings (see Alarm Silence Duration, page 4-
42).
If an alarm condition occurs when alarm audio is silenced or paused, the Alarm Audio button changes color to match the current highest-priority alarm condition
Operator’s Manual 4-41
Operation
(see Table 4-5 on page 4-36).
If alarms are silenced or paused, you can turn alarm audio back on by pressing the Alarm Audio button. The button changes to the following:
Note:
When alarms are silenced or paused, the following alarms are not silenced:
- BATTERY CRITICALLY LOW (while monitoring system is operating on battery power)
- SYSTEM FAILURE (under certain circumstances)
Note:
After a power cycle, the last setting used (alarm audio on or silenced) is retained.
Alarm Silence Duration
Pressing the Alarm Audio button either silences alarms or pauses them temporarily, depending on institutional defaults. The factory default is permanent silence of alarms.
The pause option, which must be set by an authorized technician, is 2 minutes. If the pause option is implemented, a countdown timer appears below the Alarm Audio button when the button is pressed to pause an alarm. After 2 minutes, alarm audio resumes.
Alarm Reminder Signal
By default, there is no reminder signal to indicate that alarms are silenced or paused. Alarm silence reminders may be turned on by an authorized technician. See Alarm
Audio and Visual Characteristics, page 11-4 for alarm reminder signal specifications.
4-42 Operator’s Manual
4.11.5

Dismissing an Alarm

Some alarms, such as BATTERY LOW and SENSOR NOT CONNECTED, can be dis­missed without resolving the situation. Others, such as LOW RSO2 and CHECK SENSOR, clear only when the condition is resolved. Table 8-1 on page 8-2 indicates which alarms are dismissible and which are not.
To dismiss an alarm:
1. Note whether the DISMISS button is available in the alarm message area and whether
there are multiple alarms occurring. When there are multiple alarms, the DISMISS button applies only to dismissible alarms.
Figure4-29.Dismissible Alarm
Alarm Management
2. If there are multiple alarms, press the down arrow to view all current alarms and deter-
mine which are appropriate to be dismissed.
3. Press DISMISS next to the appropriate alarm. Dismissible alarms can only be dismissed
individually.
Operator’s Manual 4-43
Operation

Event Marks

4.12
About Event Marks
Marking Events
Viewing an Event Mark Label
Renaming an Event
Changing the Event List

4.12.1 About Event Marks

Use event marks to indicate significant occurrences during monitoring. Choose from a list of events that is customizable for specific cases. Event marks are displayed on the trend graph as vertical lines with flags and stored in memory for case history review. Note that when baselines are set (either automatically or manually), an event mark is automatically added to the trend graph.
Figure4-30.Events Marked in Trend View
4-44 Operator’s Manual
4.12.2

Marking Events

To mark an event:
1. While monitoring, press MARK EVENT. The ADD EVENT menu lists the available events.
Figure4-31.ADD EVENT Menu
Event Marks
2. Scroll through the list by pressing anywhere within the list and dragging up or down.
When you see the event you want, press it to highlight it. (If you don’t see the event you want, see Changing the Event List, page 4-47.)
3. Press OK. The event mark appears on the trend graph.

4.12.3 Viewing an Event Mark Label

Event mark labels indicate the event name and the time the event was recorded. You can view event mark labels while monitoring or reviewing a case history.
To view an event mark label:
1. At the trend graph, press the flag below an event mark line. The event name and time
appear for approximately 5 seconds.
Operator’s Manual 4-45
Operation
Figure4-32.Event Mark Label

4.12.4 Renaming an Event

If an event is mis-named, you can change it at any time while monitoring or review­ing a case history.
To rename an event:
1. At the trend graph, press the flag below an event mark line.
2. When the event mark label appears, release the flag and press the label or the flag again.
The EDIT EVENT menu appears.
4-46 Operator’s Manual
Figure4-33.EDIT EVENT Menu
Event Marks
3. Scroll through the list by pressing anywhere within the list and dragging up or down.
When you see the event you want, press it to highlight it. (If you don’t see the event you want, see Changing the Event List, page 4-47.)
4. Press OK.
5. Press the flag below the event mark line to verify the change.

4.12.5 Changing the Event List

A default list of common events is provided. You can change the list before or during a case.
Note:
Be aware that changes to the list are retained across power cycles, so the list you see may reflect a previous case. Make sure that the list of events is appropriate for your patient.
To change the event list:
1. Press MENU.
2. Press SETTINGS.
3. Press EVENTS. One of several screens of available events appears. An X next to an event
indicates that it currently appears in the ADD EVENT and EDIT EVENT menus.
Operator’s Manual 4-47
Operation
Figure4-34.SETTINGS - EVENTS (Screen 1)
4. Press NEXT or BACK to view all of the available events in alphabetical order.
5. To select or de-select an individual event, press the checkbox next to the event or the
event itself. If necessary, you can restore the list to its default values by pressing RESTORE.
6. Press CLOSE.
4-48 Operator’s Manual
Loading...