Medtronic PM35MN01 Operator's Manual

Capnostream™35
0482
Portable Respiratory Monitor
PM35MN, with Microstream™ EtCO2 and Nellcor™ SPO Technologies
Operator’s Manual
PN: PT00140239B
2
ii
Capnostream™35 Portable Respiratory Monitor
Copyright Information
©2021-11 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further, Together are trademarks of Medtronic. TM* Trademark of its respective owner. All other brands are trademarks of a Medtronic company.
Covidien, Oridion Medical 1987 Ltd. and Nellcor Puritan Bennett LLC are Medtronic companies.
Notice: Purchase of this product confers no express or implied license under any Oridion Medical 1987 Ltd. patent to use the instrument with any accessory that is not manufactured or licensed by Oridion Medical 1987 Ltd.
Possession or purchase of this device does not convey an express or implied license to use the device with unauthorized sensors or cables which would, alone, or in combination with this device, fall within the scope of one or more of the patents relating to this device.
US Patents: www.covidien.com/patents
Exemptions: Oridion Medical 1987 Ltd.'s liability under this warranty does not include any transportation damage or other charges or liability for direct, indirect or consequential damages or delay resulting from improper use or application of the product or the substitution upon it of parts or accessories not approved by Oridion Medical 1987 Ltd.
All information in this manual is believed to be correct. Oridion Medical 1987 Ltd. shall not be liable for errors contained herein with the performance or use of this manual.
Capnostream™35 Portable Respiratory Monitor
iii
Table of Contents
1. Introduction 14
1.1. Introduction 14
1.2. Safety Information 14
1.2.1. General 14
1.2.2. MRI Scanning 19
1.2.3. Alarms 20
1.2.4. Fire Hazard 22
1.2.5. Electrical 23
1.2.6. Electro-magnetic Interference 25
1.2.7. Connectivity 26
1.3. Definitions 27
1.4. Contacting Technical Support 27
1.5. Symbols 28
1.6. Who Should Read This Manual 31
1.7. Warranty and Disclaimers 31
1.7.1. Warranty 31
1.7.2. Disclaimers 32
2. Product Overview 34
2.1. Overview 34
2.2. Intended Use 35
2.3. Buttons, Indicators and Connections 36
2.3.1. Monitor Front Panel 36
2.3.2. Monitor Back Panel 39
2.3.3. Monitor Right and Left Panels 40
2.3.4. Display Screens 43
2.3.5. Monitoring Display Screen Options 43
2.3.6. Monitor Turn-off 60
2.3.7. Screen Navigation 60
2.3.8. Date, Time, Language, and Other Options 61
2.3.9. Capnostream™35 Monitor Operational Check Sheet 64
iv
Capnostream™35 Portable Respiratory Monitor
3. Installation 67
3.1. Unpacking and Inspecting the Monitor 67
3.2. Batteries 68
3.2.1. Batteries 68
3.2.2. Removable Battery Pack Installation 69
3.2.3. Battery and Power Usage 71
3.2.4. Battery Hot Swap 72
3.2.5. Battery and AC Connections Indicators 72
3.2.6. Handling the Battery Pack 74
3.2.7. Storing the Battery 74
3.2.8. Disposing of the Battery 75
3.2.9. Internal Battery Pack 75
3.3. Monitor Mounting Plate 75
3.4. Operation in EMS Environments 75
3.5. Periodic Maintenance 77
4. Operation 78
4.1. Turning on the Monitor 78
4.2. Monitor Preparation for a Patient 80
4.2.1. Patient Types 81
4.3. Patient Cases and Patient ID Numbers 83
4.4. Patient Events 83
4.5. Alarm and Pulse Volumes and Alarm Delay 85
4.5.1. Alarm Volume 86
4.5.2. Pulse Tone and Button Click Volume 86
4.5.3. Alarm Volume Default Options 87
4.5.4. Alarm Delay 87
4.6. Use of Scavenging System 88
4.7. Use of Pump Off Mode 89
4.8. Additional Monitor Modes 90
4.8.1. Demo Mode 90
4.8.2. Parameter Standby Mode 91
Capnostream™35 Portable Respiratory Monitor
v
4.8.3. EMS Mode 95
4.8.4. Reminder Signal 96
4.9. Monitor Screen Menu Reference Chart 97
4.10. Capnography with the Monitor 98
4.10.1. Microstream™ EtCO2 Consumables 98
4.10.2. FilterLine™ Sampling Line Connection 101
4.10.3. CO2 Data Displayed by the Monitor 102
4.10.4. Adjustable CO2 Parameters 104
4.10.5. Monitoring CO2 during MRI Scanning 106
4.11. Pulse Oximetry with the monitor 108
4.11.1. Pulse Oximetry Features 108
4.11.2. Connecting an SpO2 Sensor to the monitor 109
4.11.3. Nellcor™ Respiration Rate 109
4.11.4. OxiMax™ SPD Alert Management Technology 114
4.11.5. Nellcor™ SpO2 Sensors 121
4.11.6. Nellcor™ SpO2 Sensor Selection 123
4.11.7. Nellcor™ Sensor Features 126
4.11.8. Performance Considerations 126
4.11.9. Nellcor™ Sensor Performance Considerations 128
4.11.10. Nellcor™ Sensor Technology 132
4.11.11. SatSeconds™ alarm management parameter 132
4.11.12. Data Update Period, Data Averaging, and Signal
Processing 137
4.11.13. Nellcor™ Pulse Oximetry with OxiMax™ Technology
Essential Performance 138
4.11.14. SpO2 Data Displayed by the Monitor 140
4.11.15. Adjustable SpO2 Parameters 143
4.11.16. SpO2 Alarm Limit Message 145
4.12. Integrated Pulmonary Index™ Value 146
4.12.1. Integrated Pulmonary Index™ Algorithm: Introduction
146
4.12.2. Integrated Pulmonary Index™ Algorithm Warnings 147
4.12.3. Integrated Pulmonary Index™ Algorithm Display 148
4.12.4. Integrated Pulmonary Index™ Algorithm Options 149
vi
Capnostream™35 Portable Respiratory Monitor
4.13. Apneas per Hour (A/hr) and the Oxygen Desaturation Index (ODI) 149
4.13.1. A/hr and ODI: Introduction 149
4.13.2. Apneas per Hour 151
4.13.3. Oxygen Desaturation Index (ODI) 152
4.13.4. Monitoring with A/hr and ODI 152
4.14. Alarms and Messages 154
4.14.1. Alarms Introduction 154
4.14.2. Alarm Display 156
4.14.3. Disabling Alarms 159
4.14.4. Alarm Review Screen 160
4.14.5. Audio Alarm Paused or Off 163
4.14.6. Reminder Signal 165
4.14.7. Message Priorities 165
4.14.8. Alarm Delay 166
4.14.9. Types of Alarms 166
4.14.10. Changing Alarm Limits 177
4.14.11. Testing Alarm Settings 180
4.14.12. Alarm Limits - Factory Defaults 182
4.15. Trends 182
4.15.1. Introduction 182
4.15.2. Trend Storage 182
4.15.3. The Trend Display Screens 183
4.15.4. Choosing Trend Parameters 192
4.15.5. Clearing Trend Memory 192
4.15.6. Trend Display Mode 193
4.15.7. Trends Configuration 193
5. Product Data Output 195
5.1. Report Options 195
5.2. Reports 196
5.3. Data Output 202
5.3.1. Clinical Report Option 211
5.3.2. Printing reports 212
Capnostream™35 Portable Respiratory Monitor
vii
5.4. Data Transfer 212
5.4.1. USB 215
5.4.2. SD 218
5.4.3. Adapter Cables 218
5.4.4. Ethernet Connectivity 220
5.4.5. Serial Connectivity 221
5.4.6. Connection to Lantronix® Device 223
5.4.7. Working with the Nellcor™ Analytics Tool (NAT) 225
5.4.8. Wireless Communication 228
5.4.9. Vital Sync™ System 238
5.4.10. Printed Reports 242
6. Preventive Maintenance 243
6.1. Introduction 243
6.2. Service Entry and Service Mode Screens 243
6.3. Required Service Activities 245
6.4. Monitor Service Password 245
6.5. CO2 Calibration 249
6.6. CO2 Calibration Check 250
6.6.1. Calibration Check Procedure 251
6.7. Software Support 252
6.8. Operations Log 252
6.9. Maintenance 253
6.10. Cleaning 254
7. Troubleshooting 256
7.1. Electrical Issues 256
7.2. CO2 Troubleshooting 257
7.3. SpO2 Troubleshooting 259
7.4. CO2 Calibration 261
7.5. Returning the Monitor 261
7.6. Technical Assistance 262
viii
Capnostream™35 Portable Respiratory Monitor
8. Accessories 263
8.1. Microstream™ EtCO2 Consumables 263
8.2. Available Accessories and Spare Parts 270
9. Theory of Operations 273
9.1. Introduction 273
9.2. Features 273
9.3. Technology Overview 274
9.3.1. What is Capnography? 274
9.3.2. What is Pulse Oximetry? 275
10. Product Specifications 279
10.1. Power Supply 279
10.2. Battery 279
10.3. Controls 280
10.4. Display 280
10.5. Sound Pressure Data 281
10.6. Microstream™ Capnography Technology 281
10.7. Nellcor™ Pulse Oximetry 283
10.8. Alarms 285
10.9. General Characteristics 286
10.10. Equipment Classification 287
10.11. Compliance 288
10.12. Electromagnetic Immunity 289
11. Institutional Settings 296
11.1. Institutional Defaults 296
11.1.1. Changing Institutional Defaults 296
11.1.2. Reset to Factory Defaults 297
11.1.3. Institutional Defaults Export/Import 298
11.1.4. Monitor Alarms Settings 299
11.1.5. Monitor Trend Settings 308
11.1.6. Monitor Settings 312
Capnostream™35 Portable Respiratory Monitor
ix
11.1.7. Home Screen Configuration 318
11.1.8. CO2 Parameters 319
11.1.9. SpO2 Parameters 320
11.1.10. Parameters Settings – Set Colors 321
12. Appendix A: Pulse Oximetry Clinical Studies 323
12.1. Clinical Studies Introduction 323
12.2. Clinical Studies Methodology 323
12.2.1. Hypoxia Methodology (Accuracy, Low Saturation, and
Motion Studies) 323
12.2.2. Low Saturation Methodology (Low Saturation Study
Only) 324
12.3. Clinical Studies Results 325
12.3.1. Accuracy Results (No Motion) 325
12.3.2. Clinical Studies Accuracy Results (Low Saturation) 328
12.3.3. Clinical Studies Accuracy Results (Motion) 331
12.4. Nellcor™ Respiration Rate Clinical Studies 333
12.4.1. Clinical Studies 333
12.4.2. Methods 334
12.4.3. Study Population 334
12.4.4. Study Results 334
12.4.5. Adverse Events 337
12.5. Clinical Studies Conclusion 337
12.5.1. No Motion 337
12.5.2. Motion 337
12.5.3. Nellcor™ Respiration Rate 338
13. Appendix B: Monitor Service Password 339
13.1. Monitor Service Password 339
x
Capnostream™35 Portable Respiratory Monitor
List of Figures
Figure 1. Monitor Front Panel 37 Figure 2. Monitor Back Panel 39 Figure 3. Monitor Right View 40 Figure 4. Monitor Left View 42 Figure 5. Home Screen Display #1, Standard 45
Figure 6. Home Screen Display #1, with Nellcor™ RR and OxiMax™
SPD Alert 47
Figure 7. Home Screen Display #2 48 Figure 8. Home Screen Display #3 50 Figure 9. Home Screen Display #4 51 Figure 10. Home Screen Display #5 52 Figure 11. Home Screen Display #6 53 Figure 12. Home Screen Display #7 54 Figure 13. Home Screen Display #8 55 Figure 14. Home Screen Display #9 57 Figure 15. Home Screen Display #10 58 Figure 16. Sample Parameters Adjustment Screen 60 Figure 17. Monitor Bottom, with Battery Pack Installed 69 Figure 18. Trend Memory Message 80 Figure 19. Patient Type Drop-down List 82 Figure 20. Event Marking Screen 84 Figure 21. Monitor Menu Flow 97
Figure 22. CO2 Data on the Monitor 102 Figure 23. CO2 Setup Screen 104 Figure 24. Clinically Significant Desaturation Patterns 115 Figure 25. The SPD Icon 116 Figure 26. Home screen with SPD Alert 120 Figure 27. The SPD Dismiss Window 121
Figure 28. Series of SpO2 Events 133 Figure 29. First SpO2 Event: No SatSeconds™ Alarm 134
Figure 30. Second SpO2 Event: No SatSeconds™ Alarm 135
Figure 31. Third SpO2 Event: Triggers SatSeconds™ Alarm 136 Figure 32. SpO2 Data on Waveform Screen 141
Figure 33. SpO2 Data in Numeric Screen 142
Figure 34. SpO2 Setup Screen 144 Figure 35. IPI Trend Graph 148 Figure 36. ODI and A/hr on the Home Screen 150
Capnostream™35 Portable Respiratory Monitor
xi
Figure 37. High Priority Alarm Example (No Breath) 157 Figure 38. Medium Priority Alarm Example 158 Figure 39. Alarm Enabled Icon (at left) and Alarm Disabled Icon (at
right) 159
Figure 40. Example of Home Screen with Some Alarms Disabled 160 Figure 41. Alarm Review Screen 161 Figure 42. Alarm Setup Screen 178 Figure 43. Graphical Trend Screen - 1 Hour Display 185 Figure 44. Tabular Trend Screen – 30 Second Display 188 Figure 45. Events List 191 Figure 46. Report Selection 196 Figure 47. Vent and Desat Report A 197 Figure 48. Vent and Desat Report B 199 Figure 49. Parameter Statistics Report 200 Figure 50. Alarm Review Report 201 Figure 51. Sample Html Report: Tabular Trend 211 Figure 52. Typical Flash Memory Device 215 Figure 53. Adapter Cable LEDs 218 Figure 54. WiFi Connection Icon; WiFi off at left, on at right 229 Figure 55. Menu Connectivity Screen showing WiFi connected 230
Figure 56. Service Mode Connectivity Screen, with Connected
Network 231
Figure 57. WiFi Configuration 236 Figure 58. Service Entry Screen, bottom section 244 Figure 59. Service Mode Screen 244 Figure 60. Service Entry Screen 246 Figure 61. Service Mode Password Change 246
Figure 62. Service Mode Password Change, Entering a New
Password 247
Figure 63. Password Successfully Set Pop-up Window 248 Figure 64. Variations of the Pleth due to Respiration 277 Figure 65. Revert To Factory Defaults Screen 298 Figure 66. Institutional Defaults Alarm Setup 302
Figure 67. Modified Bland-Altman for SpO2 (All Data - No Motion):
SaO
vs. (SpO2 - SaO2) 327
2
Figure 68. Modified Bland-Altman for SpO2 (All Data - Low
Saturation): SaO
vs. (SpO2 - SaO2) 329
2
Figure 69. Modified Bland-Altman for SpO2 (All Data - Motion): SaO2
vs. (SpO
- SaO2) 332
2
xii
Capnostream™35 Portable Respiratory Monitor
Figure 70. Modified Bland-Altman Plot of Nellcor Respiration Rate
Software versus Clinician Overscored Respiration Rate 336
List of Tables
Table 1. Symbols that Appear on the Monitor and its Packaging 28 Table 2. Display Options Available on System Screen 62 Table 3. Event Markings 85 Table 4. Message and Alarm Status during Different Parameter
Standby Situations 93
Table 5. CO2 Parameters 105
Table 6. CO2 Waveform Setup Parameters 106 Table 7. Nellcor™ SpO2 Sensor Models and Patient Sizes 124 Table 8. Pulse Oximetry Performance 138
Table 9. Electromagnetic Immunity Compliance and Guidance, Not
Life-Supporting Equipment 139
Table 10. Adjustable SpO2 Parameters 145 Table 11. IPI Values 147 Table 12. Adjustable IPI Options 149 Table 13. Alarm Indications 155 Table 14. High/Medium Priority Alarms 166 Table 15. Advisory Messages 173 Table 16. Silent Advisory Messages 173 Table 17. Waveform Area Messages 176 Table 18. Trend Increments and Zoom Levels 187 Table 19. Report Options Access 195
Table 20. Ventilation and Desaturation and Parameter Statistics
Report Types 202
Table 21. Data Transfer Report Types 204 Table 22. Connectivity Icons 213 Table 23. Communication Protocols 214 Table 24. USB File Naming Convention 216 Table 25. USB Files Examples 217 Table 26. Serial Adapter Cable (RS232 and RS422) LEDs 219 Table 27. Ethernet (LAN) Adapter Cable LEDs 219 Table 28. NAT Alarms 227 Table 29. Wireless Specifications 237 Table 30. Microstream™ Consumables 263
Capnostream™35 Portable Respiratory Monitor
xiii
Table 31. Microstream™ Advance Consumables 265 Table 32. Monitor Accessories and Spare Parts 270 Table 33. Nellcor™ Adult Respiratory Sensor Operating Range and
Power Dissipation Power Dissipation 52.5 mW 285
Table 34. Guidance and Manufacturer’s Declaration -
Electromagnetic Emissions 289
Table 35. Guidance and Manufacturer’s Declaration –
Electromagnetic Immunity 290
Table 36. Recommended Separation Distances between Portable
and Mobile RF Communications Equipment and the Monitor 294
Table 37. Factory Default Alarm Limits 299 Table 38. Parameters Available on the Alarm Review Screen 306 Table 39. Alarm Delay Settings 307 Table 40. Parameters Available on the Trend Screens 311 Table 41. Percent Modulation During Motion 325 Table 42. SpO2 Accuracy Results (No Motion) 326
Table 43. RMSD of SpO2 per Decade (No Motion) 328
Table 44. RMSD of SpO2 per Plateau (No Motion) 328 Table 45. SpO2 Accuracy Results (60 to 80% SaO2, low saturation) 328
Table 46. RMSD of SpO2 per Decade (Low Saturation) 330
Table 47. RMSD of SpO2 per Plateau (Low Saturation) 330 Table 48. SpO2 Accuracy Results during Motion 331
Table 49. RMSD of SpO2 per Decade (Motion) 333
Table 50. RMSD of SpO2 per Plateau (Motion) 333 Table 51. Demographic Data 334
Table 52. Accuracy and Precision for Respiration Rate, Version 2.0 vs.
Capnography-Based Reference 335
Table 53. Local Bias with Respect to Reference Respiratory Rate 337
Capnostream™35 Portable Respiratory Monitor
14

1. Introduction

1.1. Introduction

Carefully read this operator’s manual and the Directions for Use that accompany Microstream™ etCO henceforth referred to as sampling lines) and the SpO2 sensors, in order to use the Capnostream™35 portable respiratory monitor PM35MN, with Microstream™ EtCO to as the Capnostream™35 monitor or the monitor) correctly and safely. Use of the monitor requires full understanding and strict observance of these instructions, the precautionary information, and the specifications.
and Nellcor™ SPO2 technologies (henceforth referred
2

1.2. Safety Information

1.2.1. General

sampling lines (FilterLine™ sampling lines,
2
WARNING:
If uncertain about the accuracy of any measurement, first check the patient’s vital signs by alternate means, and then make sure the monitor is functioning correctly.
WARNING:
Patient monitoring via a network connected to the monitor is intended to supplement and not replace any part of the hospital's device monitoring. Do not rely on the remote patient monitoring as the sole source of patient data.
In order to assure a timely response to changes in patient status, a clinician must be within visual and/or audible range of the monitoring device. In order to provide medical
Safety Information
Capnostream™35 Portable Respiratory Monitor
15
intervention, a clinician must interact with the device at the bedside.
WARNING:
The monitor should not be used as an apnea monitor.
WARNING:
The monitor should be considered an early warning device. As a trend towards patient deoxygenation is indicated, blood samples should be analyzed by a laboratory co­oximeter to completely understand the patient's condition.
WARNING:
To ensure patient safety, do not place the monitor in any position that might cause it to fall on the patient. If the monitor is mounted, ensure that is securely mounted.
WARNING:
Carefully route patient cabling (SpO line) to reduce the possibility of patient entanglement or strangulation.
sensor and sampling
2
WARNING:
Do not lift the monitor by the SpO line, as they could disconnect from the monitor, causing the monitor to fall on the patient.
WARNING:
Do not obstruct the monitor so that the visual alarm indicators cannot be seen or the audio alarm indicators cannot be heard.
WARNING:
The monitor should not be used adjacent to or stacked with other equipment; if adjacent or stacked use is necessary, the
sensor cable or sampling
2
Safety Information
16
Capnostream™35 Portable Respiratory Monitor
monitor shall be observed to verify normal operation in the configuration in which it will be used.
WARNING:
Use of accessories, transducers and cables other than those specified or provided by the manufacturer of this equipment could result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and result in improper operation.
WARNING:
All cables and tubing, including SpO2 sensors and CO2 sampling lines, should be kept clear of the defibrillator and its electrodes, and should not run between, adjacent to, or overlapping the electrodes and the electrode wires, in order to reduce potential interference between the monitor and defibrillation equipment.
WARNING:
All SpO in order to enable use of a defibrillator with the monitor.
sensors must be completely intact and undamaged,
2
WARNING:
Re-use of single-use accessories could pose a cross­contamination risk to the patient or damage the functioning of the monitor.
WARNING:
readings, respiratory rate, pulse oximetry readings, and
CO
2
pulse signals can be affected by sensor application errors, certain ambient environmental conditions, including electromagnetic interference, and certain patient conditions.
WARNING:
The monitor is a prescription device and is to be operated by qualified healthcare personnel only.
Safety Information
Capnostream™35 Portable Respiratory Monitor
17
WARNING:
No modification of this equipment is allowed.
WARNING:
If calibration does not take place as instructed in the relevant service manual, the monitor may be out of calibration. A monitor that is out of calibration may provide inaccurate results.
WARNING:
Do not use any monitoring system, sensor, cable, connector, or screen that appears damaged. Remove any damaged equipment from service for inspection by a qualified service technician.
WARNING:
Do not perform service or maintenance on the device while it is in use.
WARNING:
The device is not sterile. Do not autoclave or sterilize this device.
WARNING:
Do not transport damaged or defective lithium cells and batteries via air transport.
WARNING:
Charge the monitor fully before your first use of the monitor, by attaching the monitor to AC power with the removable battery installed. See 3.2.5 Battery and AC
Connections Indicators on page 72 for more information.
Safety Information
18
Capnostream™35 Portable Respiratory Monitor
WARNING:
The unit should always be operated with the removable battery installed in order to provide back-up power in the event of a momentary or temporary power outage.
WARNING:
The characteristics of a breath are calculated differently for the different patient types. Setting the correct patient type is therefore very important. Incorrect setting will result in inaccurate monitoring of the patient's respiration, and could lead to incorrect alarm limits or produce incorrect patient IPI data.
WARNING:
Changing the institutional default settings might adversely affect the monitoring of patients. Changes to the Institutional settings must only be made by authorized personnel.
Caution:
Storage or transport of the monitor under environmental conditions beyond those mentioned in the specification will affect monitor performance and damage the monitor.
Note:
Temperature sensors will turn the monitor off when it exceeds the permitted temperature.
Note:
Dispose of calibration gases according to local regulations for disposal of gases.
Note:
Any serious incident related to device use that may occur
Safety Information
Capnostream™35 Portable Respiratory Monitor
19
should be reported immediately to the manufacturer, the local competent authority, and any other regulators as required.
Devices connected to the monitor must be medical grade only.
The accurate display of the following parameters is required in order to fill the essential performance of the monitor: Carbon dioxide levels in expired breath (CO saturation of blood (SpO2) and pulse rate when monitoring with pulse oximetry. If the patient is being monitored with both functions, all of these parameters will be displayed. If the patient is being monitored with either or both functions, alarm volume is included in the monitor's essential performance.
) when monitoring with capnography, and arterial oxygen
2

1.2.2. MRI Scanning

WARNING:
Do not use oximetry sensors during magnetic resonance imaging (MRI) scanning. Conducted current could cause burns. The sensors may affect the MRI image, and the MRI unit may affect the accuracy of oximetry measurements.
WARNING:
Do not use the FilterLine™ H Set Infant/Neonatal set, the VitaLine™ H Set Infant/Neonatal set, or any of the Microstream™ Advance Neonatal-Infant Intubated CO Line Extended Duration products during magnetic resonance imaging (MRI) scanning. Using these products during MRI scanning could harm the patient.
WARNING:
During MRI scanning, the monitor must be placed outside the MRI suite. When the monitor is used outside the MRI suite, etCO FilterLine™ XL or the MRI XL sampling line. Reference 4.10.5
Monitoring CO2 during MRI Scanning on page 106.
monitoring can be implemented using the
2
Filter
2
Safety Information
20
Capnostream™35 Portable Respiratory Monitor
Caution:
Use of a CO2 sampling line with H in its name (indicating that it is for use in humidified environments) during MRI scanning may cause artefacts on the MRI image. The use of non-H sampling lines is advised. For a list of sampling lines, reference
8.1 Microstream™ EtCO2 Consumables on page 263.

1.2.3. Alarms

WARNING:
Do not silence, pause or decrease the volume of the audible alarm if patient safety may be compromised.
WARNING:
Do not pause or turn off the audible alarms until you verify that the patient is being monitored by other means, such as direct observation.
WARNING:
Do not decrease the volume of the audible alarm to a level that is less than the ambient sound level. Doing so might impede operator recognition of the audible alarm, which might lead to patient harm during an alarm situation.
WARNING:
Always respond immediately to a device alarm since the patient may not be monitored during certain technical/caution alarm conditions.
WARNING:
Before each use, verify that the alarm enabled/disabled status and the alarm limits are appropriate for the patient being monitored.
Safety Information
Capnostream™35 Portable Respiratory Monitor
21
WARNING:
Do not set alarm limits to extreme values that render the monitoring system ineffective. Ensure alarm limits are appropriate for each patient.
WARNING:
Do not preset different or inappropriate alarm limits for the same or similar equipment in any single area, since this may compromise patient safety.
WARNING:
Remote alarm generation is intended to supplement and not replace any part of the hospital's device monitoring. Do not rely on the remote alarms as the sole source of alarms.
In order to assure a timely response to device alarms, a clinician must be within visual and/or audible range of the alarming device. In order to provide medical intervention, a clinician must interact with the device at the bedside.
WARNING:
Consider the following when setting delays on alarms that are being viewed with a remote alarm system: Connected devices report data to the application periodically, not continuously. Because of this, as well as delays caused by network bandwidth or hardware limitations or network loading, the true duration of any device alarm will be longer than the delay set in this screen for that alarm. Carefully consider these factors when choosing delay settings, and use the shortest delay settings that are practical to reduce nuisance alarms, to avoid undue delay in response to events actually requiring direct clinical intervention.
Safety Information
22
Capnostream™35 Portable Respiratory Monitor

1.2.4. Fire Hazard

WARNING:
When using the monitor with anesthetics, nitrous oxide or high concentrations of oxygen, connect the gas outlet to a scavenger system.
WARNING:
The monitor is not suitable for use in the presence of flammable anesthetic mixture with air, oxygen or nitrous oxide.
WARNING:
The sampling line may ignite in the presence of O directly exposed to laser, ESU devices, or high heat. When performing head and neck procedures involving laser, electrosurgical devices or high heat, use these devices with caution to prevent flammability of the sampling line or surrounding surgical drapes.
when
2
Caution:
Recharge the removable battery pack only in the monitor or the external battery charger (reference 8.2 Available Accessories on page 270) to avoid possible heating, burning or rupture of the battery pack.
Caution:
The battery used in this device may present a risk of fire or chemical burn if mistreated. Do not disassemble, operate the monitor beyond its environmental specifications (reference
10.9 General Characteristics on page 286), or incinerate. Replace
battery with Medtronic PN PM35BTY only. Use of another battery or a refurbished battery may present a risk of fire or explosion.
Safety Information
Capnostream™35 Portable Respiratory Monitor
23
Caution:
Replace the battery according to the instructions in this manual. Dispose of used battery promptly. Keep away from children. Do not open or short circuit the battery. Misuse may cause battery to ignite, explode, leak or get hot, causing personal injury.

1.2.5. Electrical

WARNING:
To protect against electric shock hazard, the monitor’s cover is to be removed only by qualified service personnel. There are no user-serviceable parts inside.
WARNING:
To ensure patient electrical isolation, connect only to other equipment with circuits that are electrically isolated.
WARNING:
Do not connect the monitor to a power supply other than that supplied with the monitor.
WARNING:
To avoid the risk of electric shock, the monitor must be connected only to a supply mains with protective earth.
WARNING:
Connect the monitor only to a hospital grade receptacle. The plug must be inserted into a properly wired receptacle; if a properly wired receptacle is not available, a qualified electrician must install one in accordance with the governing electrical code. Do not use extension cords or adapters of any type. The power cord and plug must be intact and undamaged.
Safety Information
24
Capnostream™35 Portable Respiratory Monitor
WARNING:
Do not connect to an electrical outlet controlled by a wall switch or a dimmer.
WARNING:
Measure the monitor’s leakage current whenever an external device is connected to the serial port. Leakage current must not exceed 100 microamperes.
WARNING:
Protection of the monitor against the effects of the discharge of a cardiac defibrillator is dependent upon the use of appropriate SpO
Nellcor™ SpO2 Sensor Models and Patient Sizes on page 124.
cables and sensors listed in Table 7.
2
Caution:
Any USB device (excluding a flash memory drive) or PC connected to the monitor must be running on a battery, or a IEC 60601-1 compliant power supply (Clause 16, ME Systems), or a IEC 60601-1 compliant isolation transformer (Clause 16, ME Systems).
Caution:
All signal input and output (I/O) connectors are intended for connection of devices complying with Clause 16, ME Systems, of IEC 60601-1 only. Connecting additional devices to the monitor may increase chassis or patient leakage currents. To maintain operator and patient safety, consider the requirements of Clause 16, ME Systems, of IEC 60601-1. Measure the leakage currents to confirm that no electric shock hazard exists.
Caution:
Electrical installation of the room or the building in which the monitor is to be used must comply with regulations specified by the country in which the equipment is to be used.
Safety Information
Capnostream™35 Portable Respiratory Monitor
25
Caution:
Keep power cord, plug and socket clear in case an urgent power supply disconnection is required.

1.2.6. Electro-magnetic Interference

This monitor has been tested and found to comply with the requirements for medical devices according to the standard IEC 60601-1-2. This standard is designed to provide reasonable protection against harmful interference in a typical medical installation.
However, because of the proliferation of radio-frequency transmitting equipment and other sources of electrical noise in healthcare environments (for example: cellular phones, mobile two–way radios, electrical appliances), it is possible that high levels of such interference due to close proximity or strength of a source may result in disruption of performance of this device.
WARNING:
Operating high frequency electrosurgical equipment in the vicinity of the monitor can produce interference in the monitor and cause incorrect measurements. Do not use the monitor with nuclear spin tomography (MRT, NMR, NMT) as the function of the monitor may be disturbed.
WARNING:
EMI disruption can cause erratic readings, cessation of operation, or other incorrect functioning.
WARNING:
The monitoring system is intended for use by healthcare professionals only. It may cause radio interference or may disrupt the operation of nearby equipment. Mitigation for such disruption may require re-orienting or relocating the monitoring system or shielding the location.
Safety Information
26
Capnostream™35 Portable Respiratory Monitor
WARNING:
Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of this monitor, including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result.

1.2.7. Connectivity

WARNING:
Ensure the security of a WiFi network before connecting the device to the network; connection to an unsecure network may endanger patient data. Access of unauthorized personnel through an unsecure network may result in patient risk.
WARNING:
Use the appropriate configuration information to ensure proper connectivity.
WARNING:
If there are changes to the network with which you are connecting the device, verify that changes are reflected in the device to ensure connectivity.
WARNING:
The Capnostream™35 monitor has been approved for use with the Vital Sync™ system. Use with hospital data systems other than the Vital Sync™ system is under the user’s responsibility.
WARNING:
Only use Medtronic-approved hardware or remote monitoring software for data port connectivity.
Contacting Technical Support
Capnostream™35 Portable Respiratory Monitor
27

1.3. Definitions

Note:
A Note is inserted to point out procedures or conditions which may otherwise be misinterpreted or overlooked and to clarify apparently contradictory or confusing situations.
Caution:
A Caution is inserted to call attention to a procedure which, if not followed exactly, can lead to damage or destruction of the equipment.
WARNING:
A Warning is inserted to call attention to dangerous or hazardous conditions inherent to the operation, cleaning, and maintenance of the equipment which may result in personal injury or death of the operator or patient.

1.4. Contacting Technical Support

For any technical issue involving the monitor, contact your local Service Center or Covidien Technical Support, as follows:
North America: Tel: 1-888-ORIDION (674-3466), Fax: (781) 453-2722; Outside North America: Tel: + (972) 2-589-9104, Fax: + (972) 2-582-8868; E-mail:
Capnographytechnicalsupport@medtronic.com
Symbols
28
Capnostream™35 Portable Respiratory Monitor

1.5. Symbols

The following symbols appear on the body of the monitor.
Table 1. Symbols that Appear on the Monitor and its Packaging
Symbol Description
Monitor ON/OFF button
Battery/power indicator
Home button
Back button
Temporarily silence audible alarms
Enter button with directional arrows
Type BF Defibrillator Proof Protection
Gas inlet
Gas outlet
port indicator
CO
2
Symbols
Capnostream™35 Portable Respiratory Monitor
29
Micro SD card port
12VDC
Nellcor™ SpO2 connection
USB flash memory connection port
Monitor connection
Federal Communications Commission
For prescription use only
Caution, consult accompanying documents
Follow Instructions for Use (symbol appears in blue on monitor)
Directive on waste from electrical and electronic equipment
Non-ionizing electromagnetic radiation
Wireless regulation (Australia)
Symbols
30
Capnostream™35 Portable Respiratory Monitor
IPX rating
TUV Mark
CE Mark
Serial number
Catalog number
Date of manufacture
Manufacturer
EC Representative
Shipping symbols
Temperature limit for storage and transport
Humidity limit for storage and transport
Atmospheric pressure limit for storage and transport
+ 310 hidden pages