Medtronic PARADIGM 522, PARADIGM 722 User Manual

PARADIGM® 522/722
Wallet Card
Pump modes
Troubleshooting
Read alarm text on your pump screen by pressing down arrow. Follow instructions. ESC/ACT to clear. Consult troubleshooting chapter located in your User Guide.
normal
The pump is in “normal” operation if it does not display an open or closed circle.
special
Indicates:
dual/square wave bolusbasal patterns (a or b)temporary basallow reservoirlow batteryblock
attention
Indicates:
suspendedrewindprimealarms
sensor icon
The Sensor Feature is turned on but the pump is not communicating with the transmitter.
sensor icon
The pump is communicating with the transmitter.
MAIN MENU
Bolus
Suspend
Basal
Prime
Utilities
BOLUS MENU
Set Bolus* Use Bolus Wizard* Manual Bolus* Bolus History
ACT
ACT
Bolus Wizard Setup
Dual/Square Bolus
Max Bolus
Easy Bolus BG Reminder
SUSPEND
BASAL MENU
Set/Edit Temp Basal Cancel Temp Basal* Select Patterns* Set/Edit Basal Basal Review
Patterns Temp Ba sa l Ty pe
Max Basal Rate
PRIME MENU
Manual Prime* Fixed Prime
Prime History
Rewind
UTILITIES MENU
Lock Keypad Alarm Daily Totals Time/ Dat e Alarm Clock Meter Options Remote Options Block Selftest User Settings Language
* Menu choice appears only if certain settings are On or Off.
Sensor
SENSOR MENU
Enter Meter BG* Sensor Update Hist.* Sensor Alarm History* Sensor Start* Sensor Setup
Press ACT to Stop Pump
ACT
ACT
ACT
ACT
Alarms and alerts
Follow instructions. ESC/ACT to clear.
Alarm Instructions
METER BG NOW
A meter BG is needed right away to calibrate your sensor and to keep receiving sensor glucose readings.
CAL ERROR
If Cal error occurs after entering a meter BG, wait until your BG is stable to enter a new meter BG for calibration.
LOW
The glucose value is lower than or equal to the set low glucose value.
HIGH
The glucose value is higher than or equal to the set high glucose value.
SENSOR END
The sensor has reached the end of its life. Replace your sensor. The sensor has a maximum life of 72 hours (3 days).
SENSOR ERROR
This alarm occurs when the sensor signals are either too high or too low. You do not need to change the sensor. Clear the alarm. If the alarm persists, refer to your user guide to test your transmitter.
WEAK SIGNAL
The pump has not received data for a period of time, as set in Missed Data. Move the pump closer to the transmitter, or move the transmitter and pump to a new location on your body.
LOST SENSOR
The pump has not received a signal from the transmitter for more than 40 minutes. Do NOT disconnect. Check the transmitter and sensor connection. If the alarm was due to a connection problem, use the Find Lost Sensor function to continue: Main Menu > Sensor
> Sensor Start > Find Lost Sensor. If the alarm was due to a
transmission problem, bring the pump closer to the sensor and transmitter. Use the Find Lost Sensor function to continue.
6025082-012 111606
Loading...