OsteoCool™ Independent
M708348B460E Rev. C
Thermocouple Kit
2019-01-23
IMPORTANT INFORMATION ON THE OSTEOCOOL™ INDEPENDENT
THERMOCOUPLE KIT
Carefully read all instructions prior to use. Observe all contraindications, warnings, and precautions noted in these instructions.
Failure to do so may result in patient complications.
FOR US AUDIENCES ONLY
Caution: federal law (USA) restricts this device to sale by or on the order of a physician.
DEVICE DESCRIPTION
The OsteoCool™ Independent Thermocouple Kit (sterile, single use) contains one OsteoCool™ Independent Thermocouple
and one OsteoCool™ Independent Thermocouple Introducer.
The thermocouple (sterile, single use) on the distal tip of the device measures tissue temperature throughout an RF ablation
procedure. The temperature measured is displayed through the OsteoCool™ RF Generator.
The introducer (sterile, single use) provides a path for the OsteoCool™ Independent Thermocouple. The introducer is to be
used with the OsteoCool™ Independent Thermocouple only.
INDICATIONS FOR USE
The OsteoCool™ Independent Thermocouple Kit, in combination with other components of the OsteoCool™ RF Ablation
System, is intended for:
▪ Palliative treatment in spinal procedures by ablation of metastatic malignant lesions in a vertebral body.
▪ Coagulation and ablation of tissue in bone during surgical procedures including palliation of pain associated with metastatic
lesions involving bone in patients who have failed or are not candidates for standard therapy.
▪ Ablation of benign bone tumors such as osteoid osteoma.
CONTRAINDICATIONS
Use of the OsteoCool™ RF Ablation System device is contraindicated in vertebral body levels C1-C7.
WARNINGS
▪ Do not use in patients who have electronic implants such as cardiac pacemakers without first consulting a qualified
professional (e.g. cardiologist). A possible hazard exists because interference with the action of the electronic implant may
occur, or the implant may be damaged.
▪ The OsteoCool™ Independent Thermocouple Kit contains single use devices. They should not be re-sterilized or re-used.
Reuse can cause patient injury and/or the communication of infectious disease(s) from one patient to another.
▪ Take appropriate precautions for patients with blood coagulation disorders, anticoagulant use, neurological deficit, or
systemic infection or local infection in area of the procedure.
▪ The procedure is to be performed with minimal to moderate sedation to allow the patient to remain in a communicative state
for patient feedback.
▪ Adequate measures must be taken to minimize x-ray exposure while using fluoroscopy. This exposure can result in acute
radiation injury as well as increased risk for somatic and genetic effects.
▪ Discontinue use if inaccurate, erratic, or sluggish temperature readings are observed. Use of damaged equipment may
cause patient injury.
▪ For safe use of the OsteoCool™ Independent Thermocouple Kit with the OsteoCool™ RF Ablation System, the physician
should have specific training, experience, and thorough familiarity with the use and application of this product.
▪ Do not use the device in fractures due to prostatic cancer or other osteoblastic metastases to the spine.
▪ Do not use the products in dense (sclerotic) bone including traumatic high energy fracture as device damage resulting in
patient injury may occur.
▪ Do not use this device in patients without metastatic malignant lesions in bone or in the vertebral body.

▪ When using this device in the vertebral body, do not use this device in patients with multiple myeloma, solitary
plasmacytoma, or primary malignant lesions in the index vertebra.
▪ Do not use this device in more than one vertebral body.
▪ Do not touch the electrode tip while power is being applied.
▪ Do not withdraw the OsteoCool™ RF Ablation Probe while power is being applied to ablate lesions.
▪ Use standard electrosurgical cautions when using the OsteoCool™ RF Ablation System in the vicinity of nerves and nerve
roots.
▪ Do not make physical contact between electrically conductive surfaces of the OsteoCool™ Independent Thermocouple Kit
and the OsteoCool™ RF Ablation Probe during the procedure when power is applied.
PRECAUTIONS
▪ Do not attempt to use the OsteoCool™ Independent Thermocouple Kit before thoroughly reading the accompanying
Instructions for Use and the Instructions for Use for the other OsteoCool™ RF Ablation System components required to
perform the RF lesion procedure.
▪ Only physicians familiar with RF lesion techniques should use the OsteoCool™ Independent Thermocouple Kit.
▪ Use care when handling the devices.
▪ Disconnect the OsteoCool™ Independent Thermocouple by pulling the connector, not the cable.
▪ Do not modify the OsteoCool™ Independent Thermocouple or OsteoCool™ Independent Thermocouple Introducer.
▪ Do not bend the OsteoCool™ Independent Thermocouple or OsteoCool™ Independent Thermocouple Introducer.
▪ The physician must determine, assess, and communicate to each individual patient all forseeable risks of the RF lesion
procedure.
ADVERSE EVENTS
The OsteoCool™ Independent Thermocouple Kit is used with other components of the OsteoCool™ RF Ablation System in RF
lesion procedures. Adverse events associated with the use of this device are similar to those indicated for medicated and
anesthetic methods utilized in other surgical procedures.
▪ As a consequence of electrosurgery, damage to surrounding tissue through iatrogenic injury could occur.
▪ Nerve injury including thermal injury, puncture of the spinal cord or nerve roots potentially results in radiculopathy, paresis,
and paralysis.
▪ Pulmonary embolism.
▪ Hemothorax or pneumothorax.
▪ Infection including deep or superficial wound infection.
▪ Unintended puncture wound including vascular puncture and dural tear.
▪ Hemorrhage.
▪ Hematoma.
▪ Pain.
PRODUCT SPECIFICATIONS
Connector 4-pin
Approximate dimensions Thermocouple Shaft: 8” length, 28G diameter
Introducer Needle: 8” length, 20G diameter
Cable length 48”
Protective sheath length 10”
Sterilization process Ethylene oxide (supplied sterile)
Storage environment Cool, dry environment
Max # of Sterilization Cycles Single Use
INSPECTION PRIOR TO USE
Prior to using the devices, inspect the packaging to ensure the package has not been damaged and sterility has not been
compromised. Do not use damaged or defective equipment in the procedure.
Inspect the device for any signs of physical damage on the shaft, handle, cable, or connector. Do not use damaged or defective
equipment in the procedure.
EQUIPMENT
The Osteocool™ Independent Thermocouple Kit is for use in Osteocool™ RF procedures performed in specialized clinical
settings. The equipment in the OsteoCool™ RF Ablation System is as follows (Figure 1):
1. OsteoCool™ RF Generator.
2. OsteoCool™ Pump Unit and OsteoCool™ Pump Cable.
3. OsteoCool™ Connector Hub.
4. OsteoCool™ Cart or OsteoCool™ Desk Stand (not shown).
5. OsteoCool™ RF Ablation Probe Kit.
a. OsteoCool™ RF Ablation Probe.