
Medtronic Navigated Manual
M708348B455E Rev. G
Reusable Instruments for Use with
the StealthStation™ System
2021-05-25
IMPORTANT INFORMATION FOR MEDTRONIC NAVIGATED MANUAL REUSABLE
INSTRUMENTS
Note: not all parts may be available in each geography.
DESCRIPTION
Medtronic Navigated Reusable Instruments are spine preparation instruments fabricated from high grade stainless steel. These
instruments are specifically designed for use in procedures where the use of stereotactic surgery may be appropriate. Placing
Medtronic single-use sterile spheres on each of the tracker passive stems allows a Medtronic computer-assisted surgery system
such as the StealthStation™ Image Guidance System to track the instruments in the surgical field. An instrument may
incorporate a measuring function which has uses as described on the label and the instrument.
Medtronic Navigated Reusable Instruments are compatible with various Medtronic spinal implant systems.
If the device passes the visual and functional inspection as described in the IFU, the instrument can be expected to perform as
intended for the perioperative period.
INTENDED USE
Medtronic Navigated Surgical Instruments are intended to be used during preparation and placement of Medtronic implants
during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive
procedures. Medtronic Navigated Reusable Instruments can be compatible with various Medtronic spinal implant systems.
Navigated surgical instruments are specifically designed for use with the StealthStation™ System, which is indicated for any
medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical
structure, such as a skull, a long bone, or vertebra, can be identified relative to a CT or MR-based model, fluoroscopy images,
or digitized landmarks of the anatomy. An instrument may incorporate a measuring function, which has uses as described on
the label and the instruments.
Do not implant the instruments.
WARNINGS
▪ Breakage, slippage, misuse, or mishandling of instruments, such as on sharp edges, may cause injury to the patient,
surgical personnel, or processing personnel.
▪ Improper maintenance, handling, or inadequate cleaning procedures can render instruments unsuitable for the intended
purpose and may present a danger to the patient, surgical personnel, or processing personnel.
▪ The surgeon should use extreme caution when working in close proximity to vital organs, nerves, or vessels. In addition,
excessive force should not be used when positioning instruments or implants, as it could cause injury to the patient.
▪ Immediate-Use Steam Sterilization is only intended for individual instruments and should only be considered under urgent
clinical situations and when approved by local policies.
▪ During navigation, it is important to frequently confirm navigational accuracy by touching the tip of the instrument on known
anatomical points, including accuracy checkpoints, and comparing the position of the instrument tip in the image with its
physical location.
▪ To avoid potential navigation inaccuracy, exercise caution to prevent the application of force on the tracker that could bias
the navigated instrument.
▪ When using the Navigated Elevate™ Inserter, only the collapsed position of the implant is navigated. Implant expansion is
not tracked.
▪ When using the Navigated T2 Stratosphere™ Inserter, only the collapsed position of the implant is navigated. Implant
expansion is not tracked.
Medtronic does not and cannot warrant the use of the instruments nor any component parts upon which repairs have been
made or attempted except as performed by Medtronic or an authorized Medtronic repair representative.

PRECAUTIONS
▪ Excessive force applied by instruments can dislodge implanted devices.
▪ Extreme care should be taken to ensure instruments remain in good working order. During the procedure, the proper
functioning of instruments is extremely important. Instruments should not be bent or damaged. Misuse of instruments
resulting in corrosion, binding, scratching, loosening, bending, or fracture of components of instruments may inhibit or
prevent proper function.
▪ Do not use instruments for any action for which they were not intended.
▪ Regularly review the operational state of all instruments and, if necessary, contact Medtronic or an authorized Medtronic
repair representative. To avoid injury, the instrument should be carefully examined for functionality or damage prior to use.
A damaged instrument should not be used. Additional back-up instruments should be available.
▪ Instruments should be carefully placed on trays, cleaned after each use, and stored in a dry environment.
▪ Preoperative and operating procedures, including knowledge of surgical techniques, are important considerations in
successful use of instruments. Proper selection and compliance of the patient greatly affect results. Read and follow all
product information supplied by the manufacturer of devices.
▪ Extra care should be taken when using instruments in pediatric patients since these patients can be more susceptible to
stresses required in use of instruments.
▪ Ensure instruments with a measuring function are not worn and any surface markings are clearly visible.
▪ Personnel should wear all appropriate Personal Protective Equipment (PPE) as recommended by the sterilizer
manufacturer.
▪ When explanting and/or disposing of a device, be sure to avoid exposure to bodily substances such as blood, tissue, etc.,
as contact could lead to infection or disease. Always wear and use proper equipment, taking special care with sharp objects
and needles. Follow your healthcare center’s policy regarding both the disposal of devices and any events of exposure.
POTENTIAL ADVERSE EFFECTS
▪ Nerve damage, paralysis, pain, and/or damage to soft tissue, visceral organs, or joints.
▪ Infection if instruments are not properly cleaned and sterilized.
▪ Dural leak in cases of excessive load application.
▪ Impingement of close vessels, nerves, and organs by slippage or misplacement of the instrument.
▪ Damage due to spontaneous release of clamping devices or spring mechanisms of certain instruments.
▪ Cutting of skin or gloves of operating staff.
▪ Bone fracture.
▪ Tissue damage to the patient, physical injury to surgical personnel, and/or increased operating time that may result from
accidental disassembly of multi-component instruments occurring during surgery.
▪ Methods of use are determined by the user's experience and training in surgical procedures. A successful result is not
always achieved in every surgical case. This fact is especially true in spinal surgery where other patient conditions may
compromise results.
Physician note: although the physician is the learned intermediary between the company and the patient, the important medical
information given in this document should be conveyed to the patient.
For US audiences only
Caution: Federal law (USA) restricts these devices to sale by or on the order of a physician.
This device should be used only by physicians familiar with the device, its intended use, and available surgical techniques.
PACKAGING
Medtronic Navigated Surgical Instruments are supplied non-sterile. Instruments may be shipped in stand-alone packaging or in
accessories. Stand-alone packaging, accessories, and contained medical devices should be intact upon receipt. Accessories
and contained medical devices should be checked for completeness and signs of damage prior to use. Damaged packages and
accessories should not be used and should be returned to Medtronic.
EXAMINATION
Instruments must always be examined pre/post cleaning and prior to surgery.
Examination should be thorough and must include a visual and functional inspection of the working surfaces, pivots, racks,
spring or torsional operation, cleanliness of location holes or cannulations, and the presence of any cracks, bending,
deformation, or distortion, and that all components are present.
Never use instruments or accessories with obvious signs of excessive wear or damage, or that are incomplete or otherwise nonfunctional.
Visual Inspection
Ensure the following:
▪ Laser marking, engravings, and other markings are legible.
▪ No cracks, deformation, or distortion are present.
▪ No Discoloration, corrosion, stains, or rust. are present. If present, wipe the device clean in accordance with the instructions
in the PROCESSING-GENERAL CONSIDERATIONS section of Medtronic Reusable Instruments and Accessories for
processing instructions (0380035).