Medtronic MUS0130069X6 Instructions for Use

Page 1
Proximal Cerebral Protection Device
INSTRUCTIONS FOR USE
Caution: Federal (USA) law restricts this product for sale by or on the order of a physician.
Medtronic, Inc., Minneapolis, MN
Page 2
INSTRUCTIONS FOR USE
x3
10
1
9
8
7
6
6
5
4
3
2
x1
13
x1
12
x2
11 1
x1
Figure 1.
1. Y-piece with hemostatic valve and flexible extension tubing (17 cm length)
2. Exit port of the working channel
3. Hollow mandrel, 0.035" guidewire compatible, to insert inside the working channel during Mo.Ma Ultra introduction
4. Shaft with 9F outer diameter
5. 3-way stopcock and flexible extension tubing (1)
6. Balloon inflation/deflation ports with attached 1-way stopcocks (11)
7. Distal balloon (ECA), with central radiopaque marker, occlusion range up to Ø 6 mm
8. Proximal balloon (CCA), with central asymmetric marker, occlusion range up to Ø 13 mm
9. Distal tip valve ("fish-mouth-valve")
10. 3x filter baskets
11. 2x 1-way stopcocks for connection to balloon inflation/ deflation ports (6)
12. 30 cc self-locking syringe
13. T-Safety Connector
0110090-4
Page 2 of 8
Medtronic, Inc., Minneapolis, MN
Manufacturer
Page 3
INSTRUCTIONS FOR USE

DESCRIPTION

The Mo.Ma™ Ultra Proximal Cerebral Protection Device, an embolic protection device, consists of a tri-lumen shaft (one working channel and two inflation-deflation lumens), two compliant balloons, and a handle at the proximal end. Cerebral protection with the Mo.Ma Ultra is achieved by proximal blood flow blockage at the carotid bifurcation by occluding the common carotid artery (CCA) and the external carotid artery (E CA) with the two compliant balloons (proximal and distal, respectively), which can be independently inflated. Debris removal is ultimately achieved by manual (syringe) blood aspiration through the working channel. Balloons are mounted on the distal portion of the shaft with the distal balloon close to the tip and the proximal balloon 6 cm from the distal balloon, just before the exit port of the working channel. The proximal CCA balloon can be inflated up to 13 mm while the distal ECA balloon can be inflated up to 6 mm. The hollow mandrel is intended for insertion into the working channel to improve trackability over a stiff 0.035” guidewire. The system is compatible with a 9F introducer sheath, its usable length (distance measured from the handle to the tip of the catheter) is 95 cm. The working channel, with a 0.083” internal diameter, serves as a guiding catheter and as a blood aspiration lumen. The exit-port of the working channel is located between the two balloons allowing easy access to the ICA.
Radiopaque markers, indicating the middle of each balloon, assist with device positioning. In addition, the asymmetry of the proximal marker assists with correct orientation of the exit-port of the working channel towards the ICA in order to facilitate guidewire introduction.
A one way tip-valve located at the distal end of the device prevents backflow from the ipsilateral ECA to the brain through the ICA.

CONTENTS

One Mo.Ma Ultra Proximal Cerebral Protection Device, one hollow mandrel, one hemostatic valve with 3-way stopcock and extension tubing, three 40 µm filter baskets, one 30 cc syringe with male Luer, one T-safety connector, two 1-way stopcocks.

INDICATIONS FOR USE

The Mo.Ma Ultra Proximal Cerebral Protection Device is indicated as an embolic protection system to contain and remove embolic material (thrombus/debris) while performing angioplasty and stenting procedures involving lesions of the internal carotid artery and/or the carotid bifurcation. The reference diameter of the external carotid artery should be between 3-6 mm and the reference diameter of the common carotid artery should be between 5-13 mm.

CONTRAINDICATIONS

The Mo.Ma Ultra Proximal Cerebral Protection Device is contraindicated for use in:
patients in whom antiplatelet and/or anticoagulation therapy is contraindicated
patients with severe disease of the ipsilateral common carotid artery
patients who are unable to respond to external questions and stimuli, or to exert a pressure with the contralateral hand
patients who have severe peripheral vascular disease preventing femoral access, hemorrhagic or hyper­coagulable status and/or inability to obtain hemostasis at the site of the femoral puncture
patients with severe vascular tortuosity or anatomy that would preclude the safe introduction of the Mo.Ma Ultra device, a stent system or other procedural devices
patients with uncorrected bleeding disorders

WARNINGS

The safety and effectiveness of the Mo.Ma Ultra Proximal Cerebral Protection Device has not been demonstrated with carotid stent systems other than the ACCULINK RX ACCULINK® Carotid Stent, XACT® Carotid Stent, PRECISE® Carotid Stent, PROTÉGÉ® RX Carotid Stent, and WALLSTENT® Carotid Stent.
®
and
The safety and effectiveness of the Mo.Ma Ultra Proximal Cerebral Protection Device has not been evaluated in vasculatures other than the carotid.
Antiplatelet and anticoagulation therapy should be administered pre- and post-procedure at a dose deemed appropriate by a physician.
Only interventionalists who have sufficient experience should carry out carotid artery angioplasty and stenting aided by proximal flow blockage cerebral protection devices. A thorough understanding of the technical principles, clinical applications and risks associated with carotid artery angioplasty and stenting is necessary before using this product.
The Mo.Ma Ultra Proximal Cerebral Protection Device is not recommended in patients who cannot tolerate contrast agents necessary for intraoperative imaging or who have chronic renal insufficiency.
This device is designed and intended for single use only. DO NOT RESTERILIZE AND/OR REUSE. Reuse or resterilization may create a risk of contamination of the device and/or cause patient infection or crossinfection, including, but not limited to, the transmission of infectious disease(s) from one patient to another. Contamination of the device may lead to injury, illness, or death of the patient. Reuse or resterilization may compromise the structural integrity of the device and/or lead to device failure, which, in turn, may result in patient injury, illness, or death. Medtronic will not be responsible for any direct, incidental, or consequential damages resulting from resterilization or reuse.
Perform balloon purging as described in this In struction for Use, before inserting the Mo.Ma Ultra Proximal Cerebral Protection Device inside the patient.
Exercise care during handling and avoid acute bends of the device before and during the balloon purging procedure.
Avoid positioning the Mo.Ma Ultra Proximal Cerebral Protection Device without the hollow mandrel.
Place the ECA balloon sufficiently deep into the ECA, so that the distal balloon (ECA balloon) cannot slip back into the CCA. However, do not place the ECA balloon deeper than 1.5 cm (measured from the ostium into the ECA) to avoid CCA proximal balloon interference with the stent placement.
The Mo.Ma Ultra balloons should not be over-inflated.
When inflating the occlusive balloons, inflation
confirmation must be made by angiographic visual estimation of balloon cylindrical shape deformation (not by pressure).
After inflating the two balloons immediately perform angiographic check of bloo d flow blockage as described in the CHECK OF FLOW BLOCKAGE section in the Instructions for Use, and check patient’s clinical tolerance to occlusion.
Should patient intolerance to occlusion occur during the procedure, immediately remove all debris performing syringe blood aspirations and deflate the proximal (CCA) balloon.
Before deflating the two occlusion balloons always verify that no more debris is retrieved in the aspirated blood as observed in the filter basket.
ICA lesion must not be crossed by guidewires or any other interventional catheters before the two occlusion balloons have been inflated and before having checked that blood flow has been effectively blocked.
When the Mo.Ma Ultra Proximal Cerebral Protection Device is exposed to the vascular system, it should be manipulated while under high-quality fluoroscopic observation.
Do not place the occlusion balloons into highly calcified vessel segments of the CCA or ECA.
Do not manipulate the Mo.Ma Ultra Proximal Cerebral Protection Device when the occlusive balloons are inflated.
If resistance occurs during manipulation do not force or continue to manipulate. The reason for the resistance must first be ascertained by fluoroscopy, road mapping or DSA before the Mo.Ma Ultra Proximal Cerebr al Protection Device is moved backwards or forwards.
0110090-4
Page 3 of 8
Medtronic, Inc., Minneapolis, MN
Manufacturer
Page 4
INSTRUCTIONS FOR USE
Use only a mixture of contrast medium and saline solution (50/50) to fill the occlusion balloons. Never use air or any gaseous medium or pure contrast to inflate the occlusion balloons.
Do not use with Lipiodol or Ethidiol contrast media, or other such contrast media, which incorporate the components of these agents.
Do not expose the Mo.Ma Ultra Proximal Cerebral Protection Device to organic solvents, e.g. acetone, alcohol.
Use the Mo.Ma Ultra Proximal Cerebral Protection Device prior to the Use Before date specified on the package.

PRECAUTIONS

Allergic reactions to contrast medium should be identified, and if possible, treated before the procedure.
The general technical requirements for catheter insertion must be observed at all times. This includes purging the balloons, flushing the components with sterile, isotonic saline solution prior to use and the usual prophylactic, systemic heparinization.
Confirm the compatibility of other devices (guidewires, balloon dilatation catheters, stent delivery systems, etc.) with the Mo.Ma Ultra Proximal Cerebral Protection Device before use.
Do not use any part of the Mo.Ma Ultra Proximal Cerebral Protection Device if the package is opened or damaged.
ADVERSE EVENTS SUMMARY OF CLINICAL RESULTS (G060247)
The objective of the ARMOUR Clinical trial was to assess the safety and effectiveness of the Mo.Ma Ultra Proximal Cerebral Protection Device for embolic protection during carotid artery angioplasty and stenting (CAS) procedures in high surgical risk patients. The serious adverse events that were reported by the investigational sites during the ARMOUR Clinical Trial are summarized in Table 1. A serious adverse event is any undesirable event that resulted in death, was life-threatening, required inpatient hospitalization or prolongation of hospitalization, required intervention to prevent permanent impairment/damage, or resulted in persistent or significant disability.
System Organ Class/Preferred Term ITT Population
Any Serious Adverse Event 16.4% (37) 29.7% (11)
Blood and lymphatic system disorders 1.8% (4)
Cardiac disorders 2.2% (5) 5.4% (2)
Gastrointestinal disorders 1.8% (4) 5.4% (2)
General disorders and administration site conditions
Infections and infestations 0.9%(2)
Table 1 . Site - Reported Serious Adverse Events
(N=225)
Anaemia 1.8% (4)
Coagulopathy 0.4% (1)
Bradycardia 0.9% (2) 2.7% (1)
Cardiac failure congestive 0.4% (1)
Cardio-respiratory arrest 0.4% (1)
Coronary artery disease 0.4% (1)
Ventricular tachycardia 2.7% (1)
Diverticular perforation 2.7% (1)
Gastrointestinal haemorrhage 0.9% (2)
Haematemesis 0.4% (1)
Upper gastrointestinal haemorrhage 0.4% (1) 2.7% (1)
0.9%(2)
Catheter site haemorrhage 0.4% (1)
Multi-organ failure 0.4% (1)
Pneumonia 0.9% (2)
Roll-In (N=37)
Table 1. Site - Reported Serious Adverse Events
System Organ Class/Preferred Term ITT Population
Injury, poisoning and procedural complications
Hip fracture 0.4% (1)
Lumbar vertebral fracture 0.4% (1)
Investigations 1.8% (4)
Blood urea increased 0.4% (1)
Ejection fraction abnormal 0.4% (1)
Haemoglobin decreased 0.9% (2)
Musculoskeletal and connective tissue disorders
Neck pain 0.4% (1)
Neoplasms benign, malignant and unspecified (incl cysts and polyps)
Lung neoplasm malignant 0.4% (1)
Nervous system disorders 3.6% (8) 5.4% (2)
Carotid artery stenosis 0.4% (1)
Cerebral hyperperfusion syndrome 0.4% (1)
Cerebrovascular accident 0.9% (2)
Confusional state 2.7% (1)
Convulsion 2.7% (1)
Embolic stroke 0.4% (1)
Ischaemic stroke 0.4% (1)
Loss of consciousness 0.4% (1)
Syncope 0.9% (2)
Renal and urinary disorders 2.7% (6)
Renal failure 1.3% (3)
Renal failure acute 0.4% (1)
Urethral stenosis 0.4% (1)
Urinary retention 0.4% (1)
Respiratory, thoracic and mediastinal disorders
Chronic obstructive pulmonary disease 2.7% (1)
Vascular disorders 4.0% (9) 18.9% (7)
Femoral artery occlusion 2.7% (1)
Hypotension 2.7% (6) 18.9% (7)
Iliac artery occlusion 2.7% (1)
Orthostatic hypotension 0.4% (1)
Vascular pseudoaneurysm 0.9% (2)

POTENTIAL COMPLICATIONS/ADVERSE EFFECTS

The complications that may result from a carotid balloon dilatation and stenting procedure, aided by a proximal flow blockage cerebral protection device, include but are not limited to:
Puncture / Access Site related:
Local hematoma
Local haemorrhage
Local or distal thromboembolic episodes
Thrombosis
Arterio-venous fistula
Pseudoaneurysm
Local infections
Procedure related:
Bradycardia
(N=225)
0.9%(2)
0.4% (1)
0.4% (1)
Roll-In (N=37)
2.7% (1)
0110090-4
Page 4 of 8
Medtronic, Inc., Minneapolis, MN
Manufacturer
Page 5
INSTRUCTIONS FOR USE
Hypotension
Carotid artery spasm
Dissection of the carotid arteries
Air emboli
Cerebrovascular accident (Stroke [ischemic,
hemorrhagic], TIA)
Acute myocardial infarction
Unstable angina
Intravascular stent migration
Angiography related:
Hyper- /Hypotension
Pain and tenderness
Arrhythmias
Sepsis/infection
Systemic embolization
Endocarditis
Short-term hemodynamic deterioration
Death
Drug reactions
Allergic reaction to contrast medium
Pyrogenic reaction

SUMMARY OF CLINICAL RESULTS

The results of this clinical trial support the safety and effectiveness of the
Mo.Ma Ultra Proximal Cerebral Protection Device.

Objective

The objective of the Invatec ARMOUR trial was to evaluate the safety and effectiveness of the Mo.Ma Ultra Proximal Cerebral Protection Device in hig h surgical risk subjects undergoing carotid artery stenting.

Tri al De si gn

Pivotal, prospective, multi-center, non-randomized trial with sequential enrollment of all qualified subjects undergoing carotid interventional procedures. All subjects who provided informed consent and met inclusion/exclusion criteria, underwent percutaneous revascularization of the carotid artery using the
TM
MO.MA
device and a stent approved by the FDA for carotid artery stenting. Follow-up took place at pre-discharge and at 30 days post-procedure. Results were compared to a performance goal of 13% for the 30-day major adverse cardiac and cerebrovascular events (MACCE) composite rate, which was derived from previous carotid stenting trials.

Trial Enrollment

A total of 262 subjects (37 roll-in, 225 ITT/pivotal) were enrol led at 25 sites in the United States (20) and the European Union (5) between September 25, 2007 and February 5, 2009.

Subject Demographics

The Mo.Ma Ultra Proximal Cerebral Protection Device could be used in subjects eligible for carotid angioplasty and stenting with stenoses involving the ICA and/or the carotid bifurcation, with an external carotid artery (ECA) reference vessel diameter 3-6 mm, and reference diameter of the common carotid artery (CCA) 5-13 mm. Subjects were required to be at least 18 years old and at high surgical risk for carotid endarterectomy procedures based on one or more of the criteria listed in Table 2. A target lesion stenosis 80% was required for asymptomatic subjects and 50% stenosis for symptomatic subjects. A symptomatic subject was defined as carotid stenosis associated with ipsilateral transient or visual TIA evidenced by amaurosis fugax, ipsilateral hemispheric TIAs or ipsilateral ischemic stroke within 6 months prior to enrollment. Baseline demographics and clinical characteristics are summarized in Table 3.
Patient Characteristics ITT Population
Clinical High Surgical Risk Criteria 80.1% (177/221)
Age >75 56.1% (124/221)
CCS angina class 3-4 or unstable angina 4.1% (9/2 21)
Table 2 . High Surgical Risk Criteria
(N=225)
Table 2 . High Surgical Risk Criteria
Patient Characteristics ITT Population
CHF Class III-IV 2.7% (6/221)
LVEF <35% 8.1% (18/221)
MI <6 weeks 0.5% (1/221)
Coronary artery disease with >=2 vessel disease in major vessel & history of angina
Severe pulmonary disease 4.5% (10/221)
Permanent contralateral cranial nerve injur y 0.0% (0/221)
Anatomical High Surgical Risk Criteria 33.0% (73/221)
High cervical lesion 3.2% (7/221)
Tandemlesions >70% 0.0% (0/221)
Hostile neck 5.0% (11/221)
CEA restenosis 7.7% (17/221)
Cervical immobility due to fusion or arthritis 2.7% (6/221)
Bilateral carotid stenosis, both requiring treatme nt 15.8% (35/221)
Table 3. Baseline Demographics and Clinical Characteristics
Patient Characteristics ITT Population
Age (years)
Mean ± SD (N) 74.7±8.54 (225)
(Min, Max) (42.4, 89.9)
Gender
Male 66.7% (150/225)
Symptomatic 15.1% (34/225)
Asymptomatic 84.9% (191/225)
Age > 75 years 56.1% (12 4/221)
Current Smoker 15.1% (33/219)
Diabetes Mellitus (DM) 37.1% (83/224)
Hypertension 87.1% (196/225)
Hyperlipidemia 81.9% (176/215)
History of renal failure 10.4% (23/221)
Family History of Premature CAD 20.7% (34/164)
Family History of Peripheral Vascular Disease 7.8% (12/153)
Family History of Stroke 25.0% (44/176)

Primary Endpoint and Secondary Endpoints

The ARMOUR trial primary endpoint (MACCE) was a composite endpoint of any myocardial infarction, stroke, or death at 30 days post-procedure. Endpoint results are summarized in Table 4.
Endpoints ITT Population
Primary Endpoint
Compositerate of MACCE to 30 days post-procedure
Any myocardial infarction 0.0% (0/220) 0.0% (0/37) 0.0% (0/257)
Stroke 2.3% (5/220) 0.0% (0/37) 1.9% (5/257)
Tabl e 4. : Primary and Secondary Endpoints
(N=220)
2.7% (6/220) 0.0% (0/37) 2.3% (6/257)
Major Ipsilateral Stroke 0.9% (2/220) 0.0% (0/37) 0.8% (2/257)
Major Contralateral Stroke 0.0% (0/220) 0.0% (0/37) 0.0% (0/257)
Minor Ipsilateral Stroke 1.4% (3/220) 0.0% (0/37) 1.2% (3/257)
(N=225)
14.5% (32/221)
(N=225)
Roll-In (N=37)FA(N=257)
0110090-4
Page 5 of 8
Medtronic, Inc., Minneapolis, MN
Manufacturer
Page 6
INSTRUCTIONS FOR USE
Table 4. : Primary and Secondary E ndpoints
Endpoints ITT Population
Minor Contralateral Stroke 0.0% (0/220) 0.0% (0/37) 0.0% (0/257)
Death 0.9% (2/220) 0.0% (0/37) 0.8% (2/257)
Compositerate of MACCE during procedure
Compositerate of MACCE at discharge
Secondary Endpoints
Device Success 98.2% (221/
Technical Success 94.6% (210/
Procedural Success 93.2% (207/
Restenosis at 30 days 1.6% (3/190) 5.9% (2/34)
Target Lesion Revascularization at 30 days
Access Site Complications 3.1% (7/225) 5.4% (2/37)
Table 5 . Distribution of Stents Used
Manufacturer/Model ITT Population (N of
Cordis PRECISE Nitinol Stent System 35.5% (82/231)
Abbott Acculink and RX Acculink Car otid Stent System
Abbott XACT Carotid Stent System 34.6% (80/231)
ev3 Protege RX Carotid Stent System 6.1% (14/231)
Boston Scientific Carotid WALLSTENT Monorail Endoprothesis
(N=220)
1.8% (4/225) 0.0% (0/37) 1.5% (4/262)
1.8% (4/225) 0.0% (0/37) 1.5% (4/262)
225)
222)
222)
0.0% (0/220) 0.0% (0/37)
Roll-In (N=37)FA(N=257)
100.0% (37/
37)
94.4% (34/
36)
91.7% (33/
36)
patients=225, N of
stents=231)
20.3% (47/231)
3.5% (8/231)

PREPARATION TECHNIQUE

1. Remove the Mo.Ma Ultra Proximal Cerebral Protection Device from the packaging.
2. Prior to use, carefully examine the entire Mo.Ma Ultra Proximal Cerebral Protection Device for defects or damage during shipment.
3. Connect the hemostatic working channel valve to the proximal port of the connector (1).
4. Flush the working channel with saline solution through the hemostatic valve and extension tubing.
5. Flush the hollow mandrel with saline solution.
6. Insert the tip of the mandrel through the hemostatic valve and into the working channel. Advance the mandrel completely and verify that the tip reaches the exit port and is positioned at the mid-point of the exit port.
7. Close the hemostatic valve in order to lock the mandrel in the correct position.

PURGING

1. Prepare the supplied 30 ml syringe by filling with 10 ml contrast/saline mix (50/50).
2. Connect the syringe to the “distal balloon ECA” luer port located on the handle of the connector.
3. Keeping the syringe downward, draw vacuum and wait for at least 15 seconds, until no air bubbles rise from the device.
4. With keeping the syringe tip downward, return to ambient pressure (release the syringe plunger).
5. Detach the syringe from the device, leaving a drop of fluid at the luer port.
Repeat steps 1-5 for the CCA proximal balloon port. The Mo.Ma Ultra Proximal Cerebral Protection Device is now purged.

DEVICE INSERTION, POSITIONING, ORIENTATION

1. After standard retrograde, femoral access is obtained, dilate the puncture site for insertion of a 9F long (approximately 25 cm long) introducer sheath.
2. Under fluoroscopic control, using standard diagnostic technique, engage the ECA with a 0.035” super-stiff guidewire.
3. Carefully introduce the proximal section of the 0.035” guidewire into the white distal tip valve [”fish-mouth­valve”(9)], located on the distal tip of the Mo.Ma Proximal Cerebral Protection Device. Advance the Mo.Ma® Ultra system (including mandrel) under fluoroscopy over the wire through the 9F introducer until the distal marker (on the ECA balloon) is positioned in the ECA close to the carotid bifurcation and max. 1.5 cm distally from the bifurcation ostium.
WARNING: Place the ECA balloon sufficiently deep into the ECA, so that the distal balloon (ECA balloon) cannot slip back into the CCA. However, do not place the ECA balloon deeper than 1.5 cm (measured from the ostium into the ECA) to avoid CCA proximal balloon interference with the stent placement.
4. The proper orientation of the working channel exit port towards the ICA can be checked by the proximal marker band. This marker band should point towards the ICA ostium.
5. Once the Mo.Ma Ultra Proximal Cerebral Protection Device is in place, open the hemostatic valve, remove the mandrel leaving the 0.035” guidewire in place, then close the hemostatic valve.
BALLOON INFLATION, FLOW BLOCKAGE CAUTION: Use only the supplied 30 ml syringe with male Luer
and T-safety connector for inflation of the occlusion balloons. Do not use the Mo.Ma Ultra Proximal Cerebral Protection Device in ECA reference diameters > 6 mm and in CCA reference diameters > 13 mm.
CAUTION: IMPORTANT NOTE ON BALLOON INFLATION CCA and ECA balloons are highly compliant and are designed to achieve atraumatic vessel occlusion by changing their shape from circular to cylindrical. They feature a wide balloon-vessel contact
area and provide the operator with immediate angiographic feedback that vessel occlusion has been achieved. For this
reason a pressure guided operation of the balloons must not be performed. Therefore during inflation, the operator must carefully watch the fluoroscopic image in order to detect the cylindrical shape deformation of the balloons.
1. Secure the supplied 30 ml syringe filled with 10 cc contrast/saline mix (50/50) to the T-safety connector.
2. Join the T-safety connector with the 1-way stopcock.
3. Ensure that no air bubbles remain in the syringe, the T­safety connector and the 1-way stopcock by flushing the assembly.
4. Attach the assembly to the ECA distal balloon port previously prepared. Ensure the assembly is securely connected to the device. With the 1-way stopcock open, gently inflate the balloon. Confirm proper balloon inflation under fluoroscopy.
5. If overpressure occurs, a few drops of fluid will escape from the T-safety connector.
6. As soon as balloon inflation is completed, turn the 1-way stopcock 90 degrees to the closed-position.
7. With the 1-way stopcock closed, detach the syringe and T­safety connector.
8. Inject 5 cc contrast through the working channel to ensure the proper occlusion of the ECA.
9. After proper positioning of the distal balloon in the ECA, remove the 0.035” guidewire from the ECA.
10. Repeat the same inflation procedure from number 1 to 7 for the proximal CCA balloon.
®
Ultra
0110090-4
Page 6 of 8
Medtronic, Inc., Minneapolis, MN
Manufacturer
Page 7
INSTRUCTIONS FOR USE

CHECK OF FLOW BLOCKAGE

1. After the two occlusion balloons have been inflated, connect a syringe prepared with contrast/saline mix (50/
50) to the side-port of the working channel.
2. Slowly aspirate 10 cc of blood into the syringe.
3. Under fluoroscopic control, slowly inject 5 cc of contrast/ saline mix and verify that contrast stagnation occurs at the carotid bifurcation.
NOTE: If the total flow blockage is achieved in the CCA and ICA, the contrast medium will stagnate in the occluded segment and can be used as a “road map” for the adequate positioning of any stent or dilation balloon catheter.
CAUTION: Ensure that the injection of contrast medium with a maximum of 5 cc is performed in a slow and controlled manner. Do not perform further injections until aspiration steps have been completed.
4. Once contrast stagnation has been verified, cerebral protection with the Mo.Ma Ultra Proximal Cerebral Protection Device is achieved, and the carotid artery procedure can be performed.

CHECK OF PATIENT TOLERANCE

Continuously monitor the patient when initiating the CCA occlusion balloon. Observe the patient’s mental and motor status. The patient should be able to hear commands, speak, and move all limbs.
Control and maintain patient systemic blood pressure. Should the patient manifest immediate or delayed intolerance after inflation of the occlusion balloons, perform the following: aspirate blood until no more debris appears in the filter basket to prevent cerebral embolism. Refer to the section “DEBRIS REMOVAL”.
Deflate the proximal balloon (CCA) in order to restore cerebral blood flow.
When the patient’s symptoms have resolved, re-inflate the proximal balloon in order to check persistence of intolerance. If intolerance persists, the procedure may be terminated or an alternative protection device may be used.
CAUTION: Should the patient not tolerate the occlusion of the CCA, the operator must first aspirate blood to prevent cerebral embolism, immediately followed by deflation of the proximal balloon CCA in order to restore cerebral blood flow.

CAROTID TREATMENT PROCEDURE

To treat the diseased ICA, the operator may use all materials of choice which are compatible with 6F introducer sheaths. The working channel of the Mo.Ma Ultra Proximal Cerebral Protection Device has an inner fully usable lumen diameter of 0.083” (2.12 mm) and is used as a guiding catheter or sheath for the advancement and retrieval of guidewires, pre- and post-dilatation balloon catheters, and carotid stent delivery systems.
Carotid angioplasty and stenting are to be performed according to standard technique and are left to the operator’s experience and skills.
It is mandatory to leave the ICA guidewire across the treated lesion while the Mo.Ma Ultra Proximal Cerebral Protection Device is in place.

DEBRIS REMOVAL

CAUTION: At the end of the procedure, perform debris removal by blood aspiration while the Mo.Ma Ultra Proximal Cerebral Protection Device system, with fully inflated occlusion balloons, is still in place.
Before performing blood aspiration, verify that the hemostatic valve in the proximal port of the working channel is completely closed.
1. Connect a 20 cc empty syringe (not supplied) to the 3-way stopcock of the side port of the hemostatic valve (5).
2. Open the stopcock and start to slowly aspirate blood. CAUTION: During aspiration, carefully monitor the patient
for any symptoms of intolerance. NOTE: This procedure can be done at any stage during
the intervention but it must be done at minimum, at the end of the procedure, i.e., following stent deployment or post­dilatation of the stented lesion.
3. Slowly aspirate at least 60 cc of blood (3 syringes of 20 cc). Check the last syringe for debris by flushing the blood content through the supplied filter basket(s). If debris is visible in the filter basket, aspirate an additional 20 cc of blood and continue to filter until no more debris can be detected.
WARNING: If no blood can be aspirated through the working channel, carefully advance a suitable guiding catheter through the working channel, slightly distal to the exit port of the working channel. Aspirate 60 cc through this guiding catheter as noted above in Step 3.

BALLOON DEFLATION

1. Prior to deflation, remove the T-safety connector from the 30 cc inflation syringe. Alternatively, any other syringe can be used for deflation.
2. When the absence of any residual debris has been ensured, deflate the distal balloon (ECA) under fluoroscopic control by connecting the syringe to the closed 1-way stopcock of the distal inflation port, opening the 1-way stopcock and applying negative pressure.
3. Deflate the proximal balloon (CCA) under fluoroscopy by connecting the syringe to the closed 1-way stopcock of the distal inflation port, opening the 1-way stopcock and applying negative pressure.
4. Perform controlled angiography of the carotid artery.

SYSTEM WITHDRAWAL

1. Retract the guidewire out of the working channel.
2. Gently retrieve the Mo.Ma Ultra Proximal Cerebral Protection Device under fluoroscopic control.

HOW SUPPLIED

The Mo.Ma Ultra Proximal Cerebral Protection Device is supplied sterile, non-pyrogenic, latex free, and intended for single use only. The Mo.Ma Ultra Proximal Cerebral Protection Device is sterilized by ethylene oxide gas. It will remain sterile as long as the packaging remains unopened and undamaged. It is supplied with a single pouch. Use product prior to Use Before date printed on the label.
CAUTION: Do not use if the inner package is open or damaged.

STORAGE

Store at controlled room temperature, in a dry place. Keep away from sunlight. Do not expose to organic solvents (e.g., acetone, alcohol). Rotate inventory so that the devices are used prior to the expiration date on package label.

DISPOSAL

After use, this product may be a potential biohazard. Handle and dispose of all such devices in accordance with accepted medical practice and applicable local, regional, and national laws and regulations.

WARRANTY DISCLAIMER

This product and each component of its system have been designed, manufactured, tested and packaged with all reasonable care. The warnings contained in Medtronic’s Instructions For Use are expressly to be considered as an integral part of this provision. Medtronic warranties the product until the expiry date indicated on the same. The warranty is valid provided that the use of the product was consistent with the Instructions For Use. Medtronic disclaims any warranty of merchantability or fitness for a particular purpose of the product. Medtronic is not liable for any direct, indirect, incidental or consequential damages caused by the product. Except in the case of fraud or grave fault on Medtronic’s part, compensation of any damage to the buyer will not, in any event, be greater than the invoice price of the disputed products. The guarantee contained in this provision incorporates and substitutes the legal guarantees for defects and compliance, and excludes any other possible liability of Medtronic, however originating, from this product supplied. These limitations of liability and warranty are not intended to contravene any mandatory provisions of law applicable. If any clause of the disclaimer is considered by a competent court to be invalid or to be in conflict with the applicable law, the remaining part of it shall not be affected
0110090-4
Page 7 of 8
Medtronic, Inc., Minneapolis, MN
Manufacturer
Page 8
INSTRUCTIONS FOR USE
and remain in full force and effect. The invalid clause shall be substituted by a valid clause which best reflects Medtronic’s legitimate interest in limiting its liability or warranty. No person has any authority to bind Invatec to any warranty or liability regarding this product.
*0110090-4*
© 2012 Medtronic, Inc.
0110090-4
Page 8 of 8
Manufacturer
Medtronic, Inc. 710 Medtronic Parkway Minneapolis, MN 55432
www.medtronic.com
Tel. (888) 283-7868 Fax (800) 838-3103
Medtronic, Inc., Minneapolis, MN
USA
Manufacturer
Loading...