Medtronic MSW002 Errata

Reveal LINQ Mobile Manager MSW002
Software Errata Sheet
Instructions for Use
Caution: Federal law (USA) restricts this device to sale by or on the order of a physician.
Medtronic, Medtronic with rising man logo, and Medtronic logo are trademarks of Medtronic. Third-party trademarks (“TM*”) belong to their respective owners. The following list includes trademarks or registered trademarks of a Medtronic entity in the United States and/or in other countries.
CareLink™, LINQ II™, Marker Channel™, Medtronic CareLink™, MyCareLink Heart™, MyCareLink™, Reveal LINQ™
Software errata sheet
Condition Observed behavior and suggested corrective action
After upgrading the app from a 1.x version, data does not carry over to the new version.
Some episode durations shown on the app are different from the durations displayed on the Medtronic CareLink Network.
The app uses inaccurate symbols to define bra­dycardia and tachycardia episode parameters.
When connecting to a patient connector while Bluetooth is disabled in the Control Center in iOS 11, the app does not notify you that Bluetooth is disableda.
The app may terminate unexpectedly. Behavior
The app may hang when generating a report PDF.
Behavior
The following types of information are no longer available after upgrading to the new app version:
• App preferences and authentication information
• Clinic and physician lists
• Pending uploads, transmissions, device registrations, and reports
You are prompted to re-authenticate and re-register the tablet after upgrading.
Suggestion
Re-authenticate and re-register the tablet. Recreate clinic or physician lists. There is no corrective action available to restore other missing data.
Behavior
A discrepancy of 1 s is observed between an episode duration on the app and the same episode duration on the Medtronic CareLink Network.
Suggestion
There is no corrective action for you to perform. The app truncates episodes down to the nearest second, while the Medtronic CareLink Network rounds intervals greater than or equal to 0.5 s up to the next second. Refer to the applicable ECGs.
Behavior
The app uses the “≤” symbol to describe bradycardia episode parameters and the “≥” symbol to describe tachycardia episode parameters. The Medtronic CareLink Network uses “<” and “>” symbols to describe the same episode parameters.
Suggestion
There is no corrective action for you to perform. The Medtronic CareLink Network correctly defines the episode parameters.
Behavior
The patient connector cannot be paired with the app.
Suggestion
Try one or more of the following options to resolve the condition and pair the patient connector with the app.
• Re-enable Bluetooth in the Control Center.
• Turn off Bluetooth and then turn it on again in the Settings on the tablet.
• Restart the tablet.
The app terminates unexpectedly.
Suggestion
Restart the app and repeat the workflow.
• No data is lost.
• The data and reports are available for viewing when the app has been restarted.
• All data and reports can also be retrieved from the CareLink Network.
• Reports are available for 7 days.
• Any data pending on the tablet remains pending on the tablet for 7 days.
• Re-enter patient data if necessary.
Behavior
The app hangs after tapping VIEW AND SAVE REPORT on the Reports tab.
Suggestion
Restart the application. Re-interrogate the device and save the session using Save Session. Generate the PDF report. If the app hangs again, use the data available within the app in the active patient session or the saved session.
3
Condition Observed behavior and suggested corrective action
Issue code 3230 may display when the app times out and the clinician reconnects the patient connector with the app.
The patient connector is not listed in the patient connector selection pop-up window.
The app hangs on the SET UP FOLLOW-UP CLINIC LIST screen.
Some channel marker symbols are displayed incorrectly in the Episode ECG section in a report or in the Episodes tab in the app.
The app hangs after the INSERT DEVICE screen.
The Bluetooth connection between the app and the patient connector is interrupted; the app tries to reconnect but does not do so.
During the insertion workflow, while interrogat­ing the device, the user is unexpectedly redir­ected to the home screen.
Multiple episode-detection markers show up on an ECG.
Behavior
Issue code 3230 is displayed when the app times out and the clinician attempts to reconnect the patient connector with the app.
Suggestion
Tap CANCEL on the 3230 issue screen, and reconnect the patient connector with the app.
Behavior
The patient connector cannot be paired with the app.
Suggestion
Retry the workflow for selecting or pairing the app with the patient connector.
Behavior
The app hangs unexpectedly.
Suggestion
Restart the app and repeat the workflow.
Behavior
The following channel markers are displayed inconsistently in the locations indicated below.
• Report: the asystole rejected marker should be displayed as the letter “D” with a strikethrough, instead it is displayed as “D” with an underscore.
• Report: the ignored VS marker should be displayed as the letter “V” with a strikethrough, instead it is displayed as “V” with an underscore.
• Report and Episodes tab: all brady episodes should display the text “Ter­minated” along with a small black box at the end of the ECG. Instead these indicators are not displayed.
Suggestion
A “D” with an underscore should be interpreted as Asystole Rejection. A “V” with an underscore should be interpreted as Ignored VS. The missing brady termi­nated marker is not primary information used for diagnosis.
Behavior
The app hangs unexpectedly during the insertion workflow.
Suggestion
End the session and restart the insertion workflow. Re-perform any insertion steps up to the point where you experienced the issue, and then resume the insertion workflow as normal.
Behavior
The BLUETOOTH INTERRUPTED message blinks on and off.
Suggestion
End the session and restart your workflow. If the 24967 patient connector has been left over the LINQ II device for 10 minutes or more, the LINQ II device will not communicate with the 24967 patient connector or the MyCareLink Heart app for 24 hours. If this occurs, you will see a lock icon next to the patient’s name in the SELECT A PATIENT list.
Behavior
The app unexpectedly returns to the home screen.
Suggestion
Restart the insertion workflow.
Behavior
An additional red triangle detection indicator shows up on an ECG.
Suggestion
There is no corrective action for you to perform. It is possible to have multiple detection indicators: the detection event for the current episode is indicated, and the additional detection indicator also corresponds to a real episode-detection event. Consider correlating the episode you are viewing with the corresponding detection marker.
4
Condition Observed behavior and suggested corrective action
ECG values are clipped when viewing reports. Behavior
When a report is viewed, some ECG values are partially clipped.
Suggestion
View the report on the Medtronic CareLink Network.
ECG data isn’t visible in the Available Epi- sodes list.
Marker Channel annotations and other episode ECG details overlap on a report.
The app crashes when attempting to connect to a patient connector.
Text in the Episode List on the Reports tab is clipped.
The app hangs if the tablet auto-locks while the app is searching for the device.
The Postal Code field displays an error mes­sage if non-numeric characters are entered.
The app and the Medtronic CareLink Network show different episode durations for the same episode.
Report labels display incorrectly in certain lan­guages.
Behavior
When an available episode is expanded in the Episodes tab, ECG data doesn’t appear where it is expected.
Suggestion
Tap the area where the missing data should be. The missing data will appear.
Behavior
Episode labels, including Marker Channel annotations, overlap each other on a report, making them difficult to read.
Suggestion
If the information is unreadable, view the report on the Medtronic CareLink Net­work.
Behavior
During a CHECK PATIENT session, the app crashes when attempting to con- nect to a patient connector.
Suggestion
Restart the app and the CHECK PATIENT session.
Behavior
Information from selected episodes is clipped when displayed on the Reports tab.
Suggestion
Generate the Session Report. The information will not be clipped in the report.
Behavior
During the insertion workflow, the app hangs if the tablet auto-locks while the patient connector is searching for the insertable device.
Suggestion
Force close the app and restart the app and the insertion workflow.
Behavior
If letters, hyphens, or other non-numeric characters are entered into any Postal Code field in the app, an error message is displayed and the user is prompted to correct the entered data. Patient auto-enrollment and orders of patient monitors are not completed.
Suggestion
If possible, enter only numbers into the Postal Code field. Otherwise, log into the appropriate clinic on the Medtronic CareLink Network to enroll the patient and order the appropriate patient monitor.
Behavior
The app lists an episode duration that is up to one second longer or shorter than the duration shown on the Medtronic CareLink Network.
Suggestion
View the Episode ECG in the app.
Behavior
When the tablet language is set to a language other than English, French, Italian, German, or Spanish, report labels that should say In Clinic or Remote display a software string instead.
Suggestion
Change the tablet language to English, French, Italian, German, or Spanish, then generate the report again.
5
Condition Observed behavior and suggested corrective action
Reports are unavailable after app update. Behavior
When the app is updated, the report from the previous session is unavailable from the Reports tab.
Suggestion
No action available. The report from the previous session is no longer accessi­ble. The information is available on the Medtronic CareLink Network.
Pending data transmissions disappear after app update and don’t show up on the Medtronic CareLink Network.
Behavior
When the app is updated, pending data transmissions disappear from the Pending Uploads screen, but don’t become available on the Medtronic CareLink Network.
Suggestion
No action available. The pending data transmissions are no longer available.
Re-authentication and tablet re-registration are required after app update.
Behavior
When the app is relaunched after updating, it prompts you to log in and to re-register the tablet.
Suggestion
To log in, enter your clinic’s Username and Password. To re-register the tablet, enter a new Tablet Name and select your Hospital or Clinic on the CREATE TABLET NAME screen. Contact your Medtronic representative if you have questions about or have for­gotten your clinic’s Username or Password.
a
The Bluetooth® word mark is a registered trademark of Bluetooth SIG, Inc. Any use of the word mark by Medtronic is under license.
6
Medtronic, Inc.
*M016249C001*
710 Medtronic Parkway Minneapolis, MN 55432 USA www.medtronic.com +1 763 514 4000
Medtronic B.V. Earl Bakkenstraat 10 6422 PJ Heerlen The Netherlands +31 45 566 8000
Canada
Medtronic Canada ULC 99 Hereford Street Brampton, Ontario L6Y 0R3 Canada 1 800 268 5346
日本メドトロニック株式会社
CRHF 事業部 〒 108-0075 東京都港区港南 1-2-70 03 6776 0047 medtronic.co.jp
© 2021 Medtronic M016249C001 A 2021-05-17
Loading...