Medtronic MSW001 Product Information

24965
Patient Connector
Electromagnetic Compatibility Declaration
Caution: Federal law (USA) restricts this device to sale by or on the order of a physician.
The following list includes trademarks or registered trademarks of Medtronic in the United States and possibly in other countries. All other trademarks are the property of their respective owners.
Medtronic
The following list of accessories were included in the system demonstrating compliance with the requirements of IEC 60601-1-2 2007 clauses 7 and 8:
Accessory Maximum length
MENB1020A0500XXX Power Supply (Medtronic re-order number: 249651)
Use of accessories other than what is specifically listed may result in increased emissions or decreased immunity of the 24965 Patient Connector.
Medical Electrical Equipment needs special precautions regarding electromagnetic compatibility (EMC) and needs to be installed and used according to the EMC information provided in the accompanying documents.
The 24965 Patient Connector should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is necessary, the 24965 Patient Connector should be observed to verify normal operation in the configuration in which it will be used.
The 24965 Patient Connector contains RF transmission and receiving capabilities. Consequently, it is possible that other equipment may interfere with the 24965 Patient Connector even if that other equipment complies with CISPR emission requirements. The following is a technical summary of the RF communication properties:
Transmitting and receiving:
• Frequency of operation: 150 kHz to 200 kHz, 2400 MHz to 2483.5 MHz
• Modulation characteristics: 18K0M1D or 74K4F1D, FHSS, DTS
• Field strength: less than 30 dBµA/m @ 10 m, less than 10 mW EIRP
Guidance and manufacturer’s declaration—electromagnetic emissions
The 24965 Patient Connector is intended for use in the electromagnetic environ­ment specified below. The customer or the user of the 24965 Patient Connector should assure that it is used in such an environment.
Emissions test Compli-
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions IEC 61000-3-2
Voltage fluctua­tions/flicker emissions IEC 61000-3-3
ance
Group 1 The 24965 Patient Connector uses RF
Class B
Class A The 24965 Patient Connector is suitable for
Complies
Electromagnetic environment—guid­ance
energy only for its internal function. There­fore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
use in all establishments, including domes­tic establishments and those directly con­nected to the public, low-voltage power sup­ply network that supplies buildings used for domestic purposes.
2 m
3