24965
Patient Connector
Electromagnetic Compatibility Declaration
Caution: Federal law (USA) restricts this device to sale by or on the
order of a physician.

The following list includes trademarks or registered trademarks of Medtronic in the United States and possibly in other
countries. All other trademarks are the property of their respective owners.
Medtronic

1 Electromagnetic compatibility declaration
The following list of accessories were included in the system demonstrating
compliance with the requirements of IEC 60601-1-2 2007 clauses 7 and 8:
Accessory Maximum length
MENB1020A0500XXX Power Supply
(Medtronic re-order number: 249651)
Use of accessories other than what is specifically listed may result in increased
emissions or decreased immunity of the 24965 Patient Connector.
Medical Electrical Equipment needs special precautions regarding
electromagnetic compatibility (EMC) and needs to be installed and used according
to the EMC information provided in the accompanying documents.
The 24965 Patient Connector should not be used adjacent to or stacked with other
equipment. If adjacent or stacked use is necessary, the 24965 Patient Connector
should be observed to verify normal operation in the configuration in which it will be
used.
The 24965 Patient Connector contains RF transmission and receiving capabilities.
Consequently, it is possible that other equipment may interfere with the 24965
Patient Connector even if that other equipment complies with CISPR emission
requirements. The following is a technical summary of the RF communication
properties:
Transmitting and receiving:
• Frequency of operation: 150 kHz to 200 kHz, 2400 MHz to 2483.5 MHz
• Modulation characteristics: 18K0M1D or 74K4F1D, FHSS, DTS
• Field strength: less than 30 dBµA/m @ 10 m, less than 10 mW EIRP
Guidance and manufacturer’s declaration—electromagnetic emissions
The 24965 Patient Connector is intended for use in the electromagnetic environment specified below. The customer or the user of the 24965 Patient Connector
should assure that it is used in such an environment.
Emissions test Compli-
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/flicker emissions
IEC 61000-3-3
ance
Group 1 The 24965 Patient Connector uses RF
Class B
Class A The 24965 Patient Connector is suitable for
Complies
Electromagnetic environment—guidance
energy only for its internal function. Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby
electronic equipment.
use in all establishments, including domestic establishments and those directly connected to the public, low-voltage power supply network that supplies buildings used for
domestic purposes.
2 m
3