Reprocessing Instructions for Endoskeleton™ TCS Torque Limiting Handles (P/N 5210-1046):
The recommended cleaning parameters for P/N 5210-1046 are detailed below. All torque drivers (P/N
5210-1046) are shipped in a non-sterile condition and must be cleaned and sterilized prior to use.
Torque drivers must not be disassembled for any reason. Consult the manufacturer’s specifications
for detailed instructions.
• Immediately after use, this device should be presoaked in an enzymatic detergent bath safe
for medical devices for five (5) minutes. The time and temperature of this soak should follow
the manufacturer’s instruction.
• Scrub the submerged device(s) with a soft bristle brush. Agitate the device while scrubbing.
Actuate moving parts. Use a small cytology brush to clean internal channels.
• Clean the cannulated devices using a small brush through the cannulation.
• Ultrasonically clean the instrument(s) for ten (10) minutes in a neutral pH detergent (or
acceptable alternative).
• Prepare the detergent according to the manufacturer’s recommendation.
• After presoak, the device must be rinsed in demineralized water.
Cleaning (Manual):
• Use a cleaning brush to remove any remaining soil or debris, paying special attention to small
grooves, crevices, and cannulation.
• Rinse thoroughly with warm (38-49˚ C (100-120˚F)) demineralized water. Agitate the device
for at least three (3) minutes.
• Check this device for any visible soil and repeat the cleaning process if soil is found. (A longer
presoak may be necessary.)
• Allow water to run through device. If the water leaving the device appears to be soiled, repeat
the cleaning procedure.
• Dry the exterior of the device(s) with a clean, lint-free cloth.
• DO NOT use any abrasive detergents, brushes, or cleaning pads. All brushes used should be
designed for use on medical devices.
Cleaning (Automated):
• An automatic cleaning device such as an ultrasonic cleaner is always preferred over manual
cleaning when available.
• Run the Automatic wash cycle with temperatures of 40°C and ultrasonic waves of 35 kHz for
10 minutes. These times and temperatures will vary depending on the equipment and
detergent used.
• Following cleaning, thoroughly rinse the devices with warm (38-49˚ C (100-120˚F))
demineralized water for at least one (1) minute.
• Allow water to run through device. If the water leaving the device appears to be soiled, repeat
the cleaning procedure.
• Dry the exterior of the device(s) with a clean, lint-free cloth.
• Only use an automatic cleaner that has been properly maintained and cleaned.
Steam Sterilization:
After a steam sterilization cycle the device must be allowed to cool for a minimum of 40 minutes in
(stagnant) open air or 30 minutes under forced air prior to being used. Failure to allow for proper
cooling can adversely affect torque performance.
Reprocessing Instructions for Endoskeleton™ TCS 3/16 Square, Torque Driver (P/N 5210-1076):
The recommended cleaning parameters for P/N 5210-1076 are detailed below. All torque drivers (P/N
5210-1076) are shipped in a non-sterile condition and must be cleaned and sterilized prior to use.
Torque-Limiting Drivers and Handles are not designed for disassembly before cleaning. To avoid
damage (pitting, rusting), avoid highly aggressive agents (NaOH, NaOCL), and salt solutions.
Surgical instruments should only be used and reprocessed by qualified healthcare practitioners.
Consult the manufacturer’s specifications for detailed instructions.