Medtronic ML-0714 Instructions for Use

CONTROLLED
Type:
Training Document
Reprocessing Instructions for Titan Devices
Document No. 70-0015
Revision No. 16
DCR No. 5520
Proprietary and confidential property of Medtronic
Page 1 of 12
Manufacturer: Titan Spine a division of Medtronic
Method: Manual and Automated Cleaning with Steam Sterilization Devices: All Titan devices provided non-sterile.
This document contains instructions for cleaning, disinfection, and sterilization of Titan product provided non-sterile, including implants, instruments, caddies and trays. Reusable instruments may enter normally sterile parts of the human body and are considered critical items. The cleaning and sterilization methods described below have been validated for use at an accredited laboratory per standards ANSI/AAMI TIR12, ANSI/AAMI TIR30, and ISO 17665-1. Cleaning and sterilization validations are summarized in the Additional Information section of this document. Any deviation from the suggested reprocessing methods should be evaluated for effectiveness using an approved method or standard and must be validated by the end user. Hospital approved methods should be followed when handling implants or instruments contaminated with blood, tissue, and/or bodily fluid. Proper personal protection equipment (PPE) per hospital procedures must be donned when handling product. Only trained personnel shall conduct reprocessing tasks. Any Titan product that appears to be damaged or corroded should not be used and a Medtronic representative should be notified immediately in order that product may be returned; damaged product should be reprocessed prior to shipment. Aesculap® SterilContainersTM should be reprocessed according to the manufacturer’s specifications.
WARNINGS
The World Health Organization (WHO) recommends special CJD (Creutzfeldt - Jakob disease) inactivation processing procedures. Consult a WHO representative for further information, if you have reason to believe a device has been contaminated.
Instrument surfaces can be damaged by improper handling; handle with care.
Take care while handling sharp instruments to avoid injury.
Abrasive cleaners or cleaning tools, such as steel wool, are to be avoided.
Improper concentrations of cleaning agents can damage stainless steel, titanium and aluminum
device’s protective surface leading to corrosion, discoloration, unrecognizable markings and other surface conditions. When using such products, be sure to follow manufacture recommended concentration, exposure time, temperature and material compatibility. Detergents with a pH between 7 and 9.5 are recommended. Certain plastics may become brittle or discolored due to excessive reprocessing. Follow manufacture recommended instructions when using disinfection units, cleaning agents and all other mechanical cleaning devices.
Do not exceed 137°C (279°F) during reprocessing steps.
Note: The cleaning and disinfecting equipment should comply with the requirements of ISO 15883.
Follow special instructions of the manufacturer for the automated washing device. Devices must be sterilized in a properly functioning, calibrated steam sterilizer.
Limitations on Reprocessing
Repeated reprocessing of medical devices has minimal effect on the devices. The end of life of Titan devices will be determined through inspection processes of each implant, instrument, caddy and tray after reprocessing. All devices are to be examined for functional integrity, corrosion and damage (proper assembly, scratches and dents, material residue, discoloration, impurities, etc.). Damaged devices, those with unrecognizable markings, signs of corrosion or dull edges should not be used and must be properly cleaned and sterilized before notifying a Medtronic representative and returning the product to Medtronic.
Reprocessing Instructions
The instructions provided have been validated by the manufacturer of the medical device as being capable of preparing a medical device for reuse. It remains the responsibility of the processor to ensure that the processing, as actually performed using equipment, materials and personnel in the processing facility, achieves the desired result. This requires verification and/or validation and routine monitoring of the process.
Initial treatment at the point of use
Instruments should be wiped clean of gross soil and kept moist to prevent drying before transferring instruments to the central processing department. Do not place used instruments back into the tray before cleaning. Transport used instruments in a closed puncture-proof container to the reprocessing area to avoid potential environmental contamination. Fragile instruments may be transported in a separate container to avoid damaging the instruments. Clean instruments as soon as possible after use. Adequate cleaning must be carried out prior to sterilization.
CONTROLLED
Type:
Training Document
Reprocessing Instructions for Titan Devices
Document No. 70-0015
Revision No. 16
DCR No. 5520
Proprietary and confidential property of Medtronic
Page 2 of 12
Preparation before cleaning
The general instructions for cleaning and decontamination described in this document shall be used for all Titan devices provided non-sterile, except those listed below. The cleaning instructions for products listed below will be described in sections that follow, as identified by their product information. Note: Aesculap® SterilContainersTM should be reprocessed according to the manufacturer’s specifications.
Endoskeleton™ TCS Torque Limiting Handle, P/N 5210-1046
Endoskeleton™ TCS 3/16 Square, Torque Driver, P/N 5210-1076
Every implant or instrument, regardless if it has been soiled, should be processed using the following recommendations. Remove devices form trays and caddies for cleaning. Open instrument teeth, locks, thread locks and hinges. Multi-component instruments should be disassembled in the reprocessing area prior to cleaning. If disassembly is required, instructions are as indicated in product specific instructions or as described in the Additional Information section.
Sort out sharp instruments to avoid injury and use caution during manual cleaning of sharp or pointed instruments. Carefully clean long, narrow, blind holes and inner lumen of instruments. For heavily soiled instruments, soak, flush, rinse with cool water and wipe with cloth to prevent soil from drying, as needed.
Cleaning: Manual
A neutral cleaner with pH 7 – 9.5 is recommended; cleaner with a pH greater than 11 should be used only on compatible materials according to its respective data sheet. Follow manufacturer instructions to achieve the correct dilution, temperature, and contact time and water quality of the detergent. Implants, instruments and cases should be cleaned in a fresh, newly prepared solution. Avoid overloading the ultrasonic wash basket.
Manually clean and process devices using the following parameters.
Phase
Cleaning instruction
Minimum
Duration of Time
(if applicable)
Water Type and
Temperature
(if applicable)
Pre-wash
Immerse disassembled device in cold water for 5 minutes. Do not use a fixating agent or hot water as this can cause fixation of residue.
5 minutes
Cold water
Brush device with a soft bristled nylon brush. Continue until all visible residues are removed. Carefully clean long, narrow, blind holes and brush inner lumen of instruments with a suitable bottle brush. Actuate all movable and hinged parts and clean area with a brush.
N/A
Cold water
Ultrasonic
Place device in an ultrasonic bath with a clean, freshly prepared solution of warm (40°C), demineralized water (DI/PURW) and cleaning product(s)* for 5 minutes.
5 minutes
Warm DI/PURW
water (40°C)
Rinse
Rinse the device thoroughly, covering all geometry with demineralized (DI/PURW) water for 15 seconds, using a water jet pistol for small gaps and slits as necessary.
15 seconds
DI/PURW water
Drying
Dry device with a clean, soft, lint-free cloth and/or purified air.
N/A
N/A
*Follow the detergent manufacturer’s recommendations for the correct dilution, temperature, contact time and water quality for the detergent. A summary of the cleaning validation is included in the Additional Information section.
CONTROLLED
Type:
Training Document
Reprocessing Instructions for Titan Devices
Document No. 70-0015
Revision No. 16
DCR No. 5520
Proprietary and confidential property of Medtronic
Page 3 of 12
Reprocessing Instructions for Endoskeleton™ TCS Torque Limiting Handles (P/N 5210-1046):
The recommended cleaning parameters for P/N 5210-1046 are detailed below. All torque drivers (P/N 5210-1046) are shipped in a non-sterile condition and must be cleaned and sterilized prior to use. Torque drivers must not be disassembled for any reason. Consult the manufacturer’s specifications for detailed instructions.
Immediately after use, this device should be presoaked in an enzymatic detergent bath safe
for medical devices for five (5) minutes. The time and temperature of this soak should follow the manufacturer’s instruction.
Scrub the submerged device(s) with a soft bristle brush. Agitate the device while scrubbing.
Actuate moving parts. Use a small cytology brush to clean internal channels.
Clean the cannulated devices using a small brush through the cannulation.
Ultrasonically clean the instrument(s) for ten (10) minutes in a neutral pH detergent (or
acceptable alternative).
Prepare the detergent according to the manufacturer’s recommendation.
After presoak, the device must be rinsed in demineralized water.
Cleaning (Manual):
Use a cleaning brush to remove any remaining soil or debris, paying special attention to small grooves, crevices, and cannulation.
Rinse thoroughly with warm (38-49˚ C (100-120˚F)) demineralized water. Agitate the device for at least three (3) minutes.
Check this device for any visible soil and repeat the cleaning process if soil is found. (A longer presoak may be necessary.)
Allow water to run through device. If the water leaving the device appears to be soiled, repeat the cleaning procedure.
Dry the exterior of the device(s) with a clean, lint-free cloth.
DO NOT use any abrasive detergents, brushes, or cleaning pads. All brushes used should be
designed for use on medical devices.
Cleaning (Automated):
An automatic cleaning device such as an ultrasonic cleaner is always preferred over manual cleaning when available.
Run the Automatic wash cycle with temperatures of 40°C and ultrasonic waves of 35 kHz for 10 minutes. These times and temperatures will vary depending on the equipment and detergent used.
Following cleaning, thoroughly rinse the devices with warm (38-49˚ C (100-120˚F)) demineralized water for at least one (1) minute.
Allow water to run through device. If the water leaving the device appears to be soiled, repeat the cleaning procedure.
Dry the exterior of the device(s) with a clean, lint-free cloth.
Only use an automatic cleaner that has been properly maintained and cleaned.
Steam Sterilization: After a steam sterilization cycle the device must be allowed to cool for a minimum of 40 minutes in (stagnant) open air or 30 minutes under forced air prior to being used. Failure to allow for proper cooling can adversely affect torque performance.
Reprocessing Instructions for Endoskeleton™ TCS 3/16 Square, Torque Driver (P/N 5210-1076):
The recommended cleaning parameters for P/N 5210-1076 are detailed below. All torque drivers (P/N 5210-1076) are shipped in a non-sterile condition and must be cleaned and sterilized prior to use. Torque-Limiting Drivers and Handles are not designed for disassembly before cleaning. To avoid damage (pitting, rusting), avoid highly aggressive agents (NaOH, NaOCL), and salt solutions. Surgical instruments should only be used and reprocessed by qualified healthcare practitioners. Consult the manufacturer’s specifications for detailed instructions.
CONTROLLED
Type:
Training Document
Reprocessing Instructions for Titan Devices
Document No. 70-0015
Revision No. 16
DCR No. 5520
Proprietary and confidential property of Medtronic
Page 4 of 12
Point of Use:
1. Wipe instruments to remove visible soil.
2. Keep instruments moist, preferably submerged in cold water (<40°C) immediately after use.
Manual Cleaning:
1. Submerge instruments in a pH neutral enzymatic cleaning agent or detergent (pH <8.5) solution for 10 to 15 minutes. Do not use aldehyde fixating detergents because they can cross-link the protein residues, making it difficult to remove soiling. Follow detergent manufacturer's instructions for mixing ratios and temperature.
2. Use a soft bristle brush if needed to remove visible soil from surfaces. Give special attention to uneven surfaces (such as knurled handles) and drilled holes/cannulation. Use flexible bottle brush, syringes, or aspiration for hard to reach places such as cannulation.
3. Thoroughly rinse with de-ionized or sterile, purified water, (<40°C), until it is no longer slippery to the touch, for a minimum of 2 minutes. Use a syringe to apply rinsing solution under pressure to hard to reach areas such as cannulations and drilled holes.
4. Visually inspect instruments for remaining debris, paying special attention to hard to reach areas. Repeat rinsing steps if needed until no visible soiling remains.
5. Drain instruments on single-use drying paper or lint-free towel.
Ultrasonic Cleaning:
1. Submerge instrument in ultrasonic bath with pH neutral enzymatic cleaning agent or detergent, (pH <8.5), solution. Follow detergent manufacturer's instructions for mixing ratios and temperature.
2. Soak at 25 to 45 kHz for 10 to 15 minutes. Visually ensure complete immersion of instruments.
3. Thoroughly rinse with de-ionized or sterile, purified water, (<40°C), until it is no longer slippery to the touch, for a minimum of 2 minutes. Use a syringe to apply rinsing solution under pressure to hard to reach areas such as cannulations and drilled holes.
4. Visually inspect instruments for remaining debris, paying special attention to hard to reach areas. Repeat rinsing steps if needed until no visible soiling remains.
5. Drain instruments on single-use blotting paper.
Dry & Inspect Instruments:
1. Thoroughly dry instruments with a soft, lint-free cloth, with single-use drying paper, or with medical grade, filtered, compressed air. Hard to reach areas such as drilled holes and cannulations should be dried with medical compressed air.
2. Visually inspect instruments before sterilization. This is typically done under normal lighting and without magnification.
3. Visually inspect for soiling, corrosion, cracks, and damage to components.
4. Functionally inspect adapters (if equipped) to ensure they connect onto mating shafts. Check function of ratcheting screwdrivers in both directions, (forward and reverse).
5. Notify appropriate personnel regarding damage and malfunctioning components.
Sterilization:
1. Sterilization testing has been performed in accordance with ISO 17665-1:2006 Sterilization of Health Care Products – Moist Heat – Part 1 a. Gravity: 250°F, 30 min exposure, 45 min dry time.
2. After dry time allow a minimum of ten minutes for equilibration period to allow instrument to cool to room temperature.
Cleaning: Automated
Instruments and trays should be washed separately. A wire basket provided by the mechanical washer manufacturer may be used to hold loose instruments and component devices. All design features must be accessible for cleaning and positioned to assist with drainage where liquid may settle. For example, cannulations and blind holes should be inclined downwards and hinges should be open to assist with drainage. Avoid overloading the wire baskets and washer. Process the devices using the following parameters.
Loading...
+ 8 hidden pages