Medtronic ML-0714 Instructions for Use

GENERAL STERILIZATION AND CLEANING INSTRUCTIONS
Quality Statement
TeDan Surgical Innovations’, LLC (TSI) dedication to delivering state of the art surgical retractor systems has allowed TSI to become the leader in an ever advancing medical industry. TSI strives to meet our customer needs and market demands in a timely manner through our experienced and innovative management team.
Warranty
All reusable products oered by TeDan Surgical Innovations, LLC, are guaranteed by a Lifetime Warranty against defects in materials and workmanship from the date of purchase. TSI’s Lifetime Warranty does not cover damages caused by normal wear, failure to follow instructions, misuse and products that have been repaired or modied in ways not approved by TSI. Prepackaged sterile products are warranted to be free from defect until expiration date. Any product that proves to be defective during this period will be repaired or replaced at no charge. TeDan Surgical Innovations, LLC shall not be held responsible for consequential or indirect damage arising from the sale or use of any product.
Please note:
The color of TSI’s Titanium and Aluminum instruments may vary due to the anodizing process or alloy used. Shading or loss of color may also occur after sterilization. This is not a defect in the instrument or material and will not aect the performance of your high quality TSI instrument.
When loosening, do not force the knob of the articulating arm past the stop. Doing so could damage the ball joint and aect the rigidity of the articulating arm (Figure 1).
Do not over turn the pivoting mechanism located on the retractor frames. Forcing the pivoting mechanism past stop may cause damage to the device (Figure 2).
The procedure tested by TSI should be used for sterilization and cleaning. A hospital may choose to follow another procedure in the event that the procedure is validated. In either case, the eectiveness of the procedure should be established. TSI utilized biological indicators to establish the eectiveness of the procedure. Other methods can be found in ISO 17665-1:2006 Standard.
Cleaning Instructions:
Please note in preparation of cleaning:
1. Automated cleaning is not suitable for instruments with long lumens, ball joints, or stainless steel cables (e.g. suction tubes, surgical arms and exible surgical arms). Such instruments should undergo a manual cleaning prior to sterilization.
2. When applicable, instruments should be cleaned in the open or unlocked position.
3. Flexible Arms: Turn tightening knob counter-clockwise to loosen the internal cable to enable movement of the surgical arm beads to allow uid to ow between each link prior to placing arm in sonicator. Articulating Arms: Turn tightening knob clockwise to tighten ball joint prior to placing arm in sonicator.
4. Black coated and color anodized components may be negatively aected if aggressive cleaning mediums or appliances (e.g. extreme acidic/alkaline, abrasives) are used. A pH neutral cleaner, which may or may not contain enzymes, (such as Prolystica manufactured by STERIS Corp.) is recommended. Exposure to chlorides or hydrogen peroxide may negatively aect the coating or colorization of the components.
Manual cleaning instructions:
1. Rinse each instrument individually with a steady stream of tap water until gross contaminants are removed. Depending on the complexity of the device, this process should take approximately 1-2 minutes.
2. Place each instrument into a sonicator containing an enzymatic, pH neutral detergent solution prepared according to the manufacturer’s instructions and
Figure 1 Figure 2
TSI-0101001 1
© 2017 TEDAN SURGICAL INNOVATIONS/ Rev11
www.tedansurgical.com
sonicate for 10 minutes.
3. Prepare a wash solution using an enzymatic, pH neutral detergent, in a wash container with tap water using the concentration recommended by the detergent manufacturer.
4. Transfer each instrument to a manual wash container and fully immerse in the cleaning solution prepared in Step 3.
5. While still submerged, any visible contamination and debris should be removed by scrubbing each instrument with an appropriately-sized soft nylon bristle brush until visibly clean, paying particular attention to hard-to-clean areas such as crevices and joints. Depending on the complexity of the device, this process should take approximately 1-2 minutes. Using an appropriately sized syringe, ush cannulations at least three times with 50mL of cleaning solution each time. See below for additional instrument specic instructions.
a. For shafted instruments with cannulae, ush with cleaning solution
by inserting the ushing cannula and attaching an appropriately sized syringe lled with cleaning solution.
b. For applicable instruments, such as Suction Tubes, insert the supplied
mandrel into the suction tube lumen to dislodge potentially trapped debris.
c. For ex arms, turn the tightening knob counter-clockwise to loosen the
internal cable to enable movement of the surgical arm beads to allow uid
6. Rinse each instrument with deionized water for 30-60 seconds.
7. Dry each instrument using clean, absorbent, low lint wipes to remove excess rinse
Automated cleaning instructions:
Use only washer/disinfector machines that have been validated in accordance with ISO
15883.
1. Perform pre-cleaning to remove gross contaminants as follows:
2. While still submerged, remove visible soil by scrubbing with an appropriately sized
3. Rinse with owing deionized water for a minimum of 30 seconds for each
4. Load instruments into washer/disinfector in accordance with the manufacturer’s
5. Arrange instruments with curved surfaces and cannulations facing downward to
6. Operate the washer/disinfector cycle according to the manufacturer’s instructions.
Recommended minimal washer/disinfector parameters:
• Heated wash at 140°F (60°C) for 2 minutes
• Heated tap water rinse at 140°F (60°C) for 20 seconds
• Heated deionized water rinse at 180°F (82°C) for 1 minute
• Forced air drying at 240°F (116°C) for 9 minutes
Surgical arm inspection before use:
Articulating Arms
1. Inspect entire assembly for damage.
2. Hold arm assembly at column and turn central tightening knob clockwise.
3. Check to make sure that arm is rigid at all three joints.
4. Insert arm column into table clamp, turn column tightening lever clockwise and
Flexible Arms
1. Inspect entire assembly for damage.
2. Turn the ex arm tightening knob clockwise and ensure the arm is suciently rigid
to ow between each link.
water.
a. Prepare a wash solution using a pH neutral detergent, in a wash container
with tap water using the minimum concentration recommended by the
detergent manufacturer. b. Submerge and soak in wash solution for a minimum of 1 minute. c. Flush any cannulations that may be present two times with 50 ml of
cleaning solution each time.
soft nylon bristle brush for a minimum of 4 minutes until no visible soil is observed.
instrument.
instructions.
prevent pooling of water.
ensure that it holds securely.
© 2017 TEDAN SURGICAL INNOVATIONS/ Rev11 www.tedansurgical.com
TSI-01010012
+ 2 hidden pages