CHANGING THE MINIMED™SURE-T® INFUSION SET QUICK REFERENCE GUIDE
START HERE
1
5
2
1 2 3 4
6
3 4
Wash your hands. Press
. Select Options.
Select Reservoir & Tubing. Select New Reservoir.
Select Rewind.
Remove the infusion set you have been using by loosening the adhesive and pulling away from body.
Remove the used reservoir from the pump.
Remove from package. Make sure insulin vial is at room temperature to reduce the risk of air bubbles.
Pull plunger down to the amount that you
plan to ll with insulin.
Wipe vial with alcohol. Place vial on table. Firmly press the blue transfer guard onto vial.
Push and hold plunger down.
Plunger rod
Plunger
Reservoir
Insulin vial
Transfer guard
FILL RESERVOIR & CONNECT TO THE INFUSION SET TUBING
F
ollow the next steps to ll reservoir with insulin and connect to the infusion set tubing.
Continued on next page
2
MINIMED™ 670G SYSTEM |
CHANGING THE MINIMED™SURE-T® INFUSION SET QUICK REFERENCE GUIDE
Remove the infusion set from package.
Press the edges of the set release connector. With the other hand, pull
the cap o.
Slide the set release connector into the set connector.
5
4
1
6
6
7
2
8
3
With your thumb still on the plunger,
ip over so vial is on top. Release thumb and pull plunger down to ll
with insulin.
Tap the reservoir to move air bubbles to top of reservoir. Push plunger up to move air into vial.
If needed, pull plunger back down to amount of insulin needed for 2-3 days.
To avoid getting insulin on the top of the reservoir, turn vial over so it is upright. Hold transfer guard and turn reservoir counter­clockwise and remove from transfer guard.
WARNING: If insulin or any liquid gets inside the tubing connector, it can temporarily block the vents that allow the pump to properly fill
the infusion set. This may result in the delivery of too little or too much insulin, which could cause hyperglycaemia or hypoglycaemia.
CONNECT RESERVOIR TO INFUSION SET
Y
ou will place the tubing connector onto the end of the infusion set to the lled reservoir.
Gently push the tubing connector onto reservoir. Turn clockwise until locked. You will hear a click.
If you see air bubbles, tap reservoir to move them to top. Push plunger just a bit to move them into tubing.
Twist plunger counter­clockwise to loosen and remove.
5
Continued on next page
Set connector
Set release
Tubing connector