MINIMED™ 670G SYSTEM |
CHANGING THE MINIMED™SURE-T® INFUSION SET QUICK REFERENCE GUIDE
START HERE
1
5
2
1 2 3 4
6
3 4
Wash your hands.
Press
. Select Options.
Select Reservoir & Tubing. Select New Reservoir.
Select Rewind.
Remove the infusion set you have been using by
loosening the adhesive and pulling away from body.
Remove the used reservoir from the pump.
Remove from package. Make sure
insulin vial is at room temperature
to reduce the risk of air bubbles.
Pull plunger down to
the amount that you
plan to ll with insulin.
Wipe vial with alcohol. Place
vial on table. Firmly press the
blue transfer guard onto vial.
Push and hold
plunger down.
Plunger
rod
Plunger
Reservoir
Insulin vial
Transfer
guard
FILL RESERVOIR & CONNECT TO THE INFUSION SET TUBING
F
ollow the next steps to ll reservoir with insulin and connect to the infusion set tubing.
Continued on next page
![](/html/e0/e075/e0751c28660d208c4ff9c95d164091ad5a90134689fd1c9c84d876308b66b2f9/bg2.png)
MINIMED™ 670G SYSTEM |
CHANGING THE MINIMED™SURE-T® INFUSION SET QUICK REFERENCE GUIDE
Remove the infusion set from package.
Press the edges of the set release
connector. With the other hand, pull
the cap o.
Slide the set release
connector into the
set connector.
5
4
1
6
6
7
2
8
3
With your thumb still on the plunger,
ip over so vial is on top. Release
thumb and pull plunger down to ll
with insulin.
Tap the reservoir to move
air bubbles to top of
reservoir. Push plunger
up to move air into vial.
If needed, pull plunger
back down to amount
of insulin needed for
2-3 days.
To avoid getting insulin on the top of the
reservoir, turn vial over so it is upright. Hold
transfer guard and turn reservoir counterclockwise and remove from transfer guard.
WARNING: If insulin or any liquid gets inside the tubing connector, it can temporarily block the vents that allow the pump to properly fill
the infusion set. This may result in the delivery of too little or too much insulin, which could cause hyperglycaemia or hypoglycaemia.
CONNECT RESERVOIR TO INFUSION SET
Y
ou will place the tubing connector onto the end of the infusion set to the lled reservoir.
Gently push the tubing
connector onto reservoir.
Turn clockwise until locked.
You will hear a click.
If you see air bubbles, tap
reservoir to move them to
top. Push plunger just a bit
to move them into tubing.
Twist plunger counterclockwise to loosen
and remove.
5
Continued on next page
Set connector
Set release
Tubing
connector
Sure-T
insertion piece