Medtronic MINIMED SURE-T Quick Reference Manual

Page 1
1
MINIMED™ 670G SYSTEM |
CHANGING THE MINIMED™SURE-T® INFUSION SET QUICK REFERENCE GUIDE
START HERE
1
5
2
1 2 3 4
6
3 4
Wash your hands. Press
. Select Options.
Select Reservoir & Tubing. Select New Reservoir.
Select Rewind.
Remove the infusion set you have been using by loosening the adhesive and pulling away from body.
Remove the used reservoir from the pump.
Remove from package. Make sure insulin vial is at room temperature to reduce the risk of air bubbles.
Pull plunger down to the amount that you
plan to ll with insulin.
Wipe vial with alcohol. Place vial on table. Firmly press the blue transfer guard onto vial.
Push and hold plunger down.
Plunger rod
Plunger
Reservoir
Insulin vial
Transfer guard
FILL RESERVOIR & CONNECT TO THE INFUSION SET TUBING
F
ollow the next steps to ll reservoir with insulin and connect to the infusion set tubing.
Continued on next page
Page 2
2
MINIMED™ 670G SYSTEM |
CHANGING THE MINIMED™SURE-T® INFUSION SET QUICK REFERENCE GUIDE
Remove the infusion set from package.
Press the edges of the set release connector. With the other hand, pull
the cap o.
Slide the set release connector into the set connector.
5
4
1
6
6
7
2
8
3
With your thumb still on the plunger,
ip over so vial is on top. Release thumb and pull plunger down to ll
with insulin.
Tap the reservoir to move air bubbles to top of reservoir. Push plunger up to move air into vial.
If needed, pull plunger back down to amount of insulin needed for 2-3 days.
To avoid getting insulin on the top of the reservoir, turn vial over so it is upright. Hold transfer guard and turn reservoir counter­clockwise and remove from transfer guard.
WARNING: If insulin or any liquid gets inside the tubing connector, it can temporarily block the vents that allow the pump to properly fill
the infusion set. This may result in the delivery of too little or too much insulin, which could cause hyperglycaemia or hypoglycaemia.
CONNECT RESERVOIR TO INFUSION SET
Y
ou will place the tubing connector onto the end of the infusion set to the lled reservoir.
Gently push the tubing connector onto reservoir. Turn clockwise until locked. You will hear a click.
If you see air bubbles, tap reservoir to move them to top. Push plunger just a bit to move them into tubing.
Twist plunger counter­clockwise to loosen and remove.
5
Continued on next page
Set connector
Set release
Tubing connector
Sure-T insertion piece
Page 3
3
MINIMED™ 670G SYSTEM |
CHANGING THE MINIMED™SURE-T® INFUSION SET QUICK REFERENCE GUIDE
3 4
1 2 3
Place reservoir into pump. Select Next.
Select Load and keep holding
.
When you see this screen, select Next.
After you see drops, press
and select Next.
Select Fill and keep holding
until you see
drops at the end of tubing, then let go.
Turn clockwise until you feel reservoir lock into place.
Drops at end of tubing
21
LOAD RESERVOIR AND FILL TUBING
Fol
low these steps to load the reservoir and ll the tubing.
PLACE RESERVOIR INTO PUMP
Now place the lled reservoir into the reservoir compartment of the pump.
Select Load Reservoir from the menu.
Press
to open the menu. If the pump is locked, you will need to unlock the pump after pressing
.
Select Next.
Continued on next page
THE BACKLIGHT MAY HAVE TURNED OFF
Press any button to turn the screen back on.
Page 4
4
MINIMED™ 670G SYSTEM |
CHANGING THE MINIMED™SURE-T® INFUSION SET QUICK REFERENCE GUIDE
2
43 5 6
Peel o both pieces of
the paper backing from the insertion piece.
With one hand, pinch up skin at prepared site. With the other hand, insert the needle straight in at a 90 degree angle.
You do not need to ll
a cannula. Press
and
select Done.
Pull o the needle
guard.
Press adhesive securely to skin.
Choose an insertion site from the shaded areas shown here. Wipe with alcohol or other antiseptic.
Remove paper backing from the set connector and secure adhesive to skin. Be sure to place it on an area of your body that allows for easy disconnecting from tubing.
1
INSERT INFUSION SET
Next, follow the steps to insert the infusion set into your body
.
Hold infusion set insertion piece by
ipping the plastic wings upwards and
pressing them together.
MiniMed is a trademark and Sure-T is a registered trademark of Medtronic, MiniMed Inc. ©2017 Medtronic MiniMed, Inc. All rights reserved. Medtronic Australasia Pty Ltd. 2 Alma Road, Macquarie Park. NSW 2113. Australia. medtronic-diabetes.com.au 5360-102018
COMPLETE THE INFUSION SET CHANGE
Use the pump to nish the process.
Your infusion set change is now complete!
Loading...