Medtronic MiniMed Mio Infusion Set Instructions Manual

Changing the mio™ Infusion Set Using a Revel™ Insulin Pump
plunger rod
transfer guard
reservoir
To Rewind the Piston in Pump:
1. Wash your hands. Remove old infusion set from your body and the reservoir from pump.
2. Go to Reservoir + Set in MAIN MENU, then select Reservoir Setup to open REWIND screen. Press to rewind.
To Fill the Reservoir:
ACT
Pull plunger rod down to
2. fill reservoir with air
After wiping top
3. of insulin vial with alcohol, hold blue transfer guard and press it down onto insulin vial until fully pushed into place.
4. Use thumb to push air from reservoir into insulin vial. Continue to hold plunger rod down.
5. Continue to hold plunger rod down as you turn vial over (vial should now be on top and reservoir on bottom). Release plunger. Then slowly pull plunger rod down to fill reservoir with insulin.
6. Check reservoir for air bubbles. Tap side of reservoir to force bubbles to rise to top of reservoir. Push plunger rod up to move air bubbles into insulin vial.
7. After air bubbles are removed, pull plunger rod back down to fill reservoir with enough insulin to last 2-3 days.
8. Flip vial over so it is now on bottom. Hold blue transfer guard with one hand. With other hand turn reservoir counter- clockwise and pull it straight up until disconnected from blue transfer guard. Disconnect blue transfer guard and dispose into a sharps container.
Changing the mio™ Infusion Set Using a Revel™ Insulin Pump
connector
To Connect the Tubing:
Remove the mio™
1. infusion set from packaging by pulling down the red plastic tab. Then peel sterile paper from the bottom.
2. With one hand press the three raised markings on the sides of the mio™ lid. Then raise and remove lid.
Free tubing from slot that
3. holds it in place. Gently unwind tubing in counter clockwise direction.
To Connect the Tubing:
5. Tap side of reservoir to force any remaining air bubbles to rise to top. Push up on plunger rod until bubbles are out and small amount of insulin is in tubing.
Turn plunger rod
6. counter-clockwise until it unscrews from reservoir. Be careful not to pull plunger rod out from reservoir or insulin will spill out.
To Fill the Tubing:
Press
1. REWIND COMPLETE appears. Insert newly filled reservoir into pump’s reservoir compartment.
ACT
so that
4. Hold reservoir in one hand. With other hand gently push connector onto top of reservoir and turn clockwise until it slides and locks into place.
Turn reservoir
2. clockwise until it locks in place.