Changing the mio™ Infusion Set Using a Revel™ Insulin Pump
plunger
rod
transfer
guard
reservoir
To Rewind the Piston in Pump:
1. Wash your hands. Remove old infusion set from your body and the
reservoir from pump.
2. Go to Reservoir + Set in MAIN MENU, then select Reservoir Setup to
open REWIND screen. Press to rewind.
To Fill the Reservoir:
ACT
1. Remove reservoir
from package.
Pull plunger rod down to
2.
fill reservoir with air
After wiping top
3.
of insulin vial
with alcohol,
hold blue
transfer guard
and press it
down onto
insulin vial until
fully pushed
into place.
4. Use thumb to
push air from
reservoir into
insulin vial.
Continue to hold
plunger rod
down.
5. Continue to hold
plunger rod
down as you
turn vial over
(vial should now
be on top and
reservoir on
bottom). Release
plunger. Then
slowly pull
plunger rod down
to fill reservoir
with insulin.
6. Check reservoir
for air bubbles.
Tap side of
reservoir to force
bubbles to rise
to top of
reservoir. Push
plunger rod
up to move
air bubbles into
insulin vial.
7. After air bubbles are
removed, pull plunger
rod back down to fill
reservoir with enough
insulin to last 2-3 days.
8. Flip vial over so it is now
on bottom. Hold blue
transfer guard with one
hand. With other hand
turn reservoir counter-
clockwise and pull it
straight up until
disconnected from blue
transfer guard. Disconnect
blue transfer guard and
dispose into a sharps
container.

Changing the mio™ Infusion Set Using a Revel™ Insulin Pump
To Connect the Tubing:
Remove the mio™
1.
infusion set from
packaging by pulling
down the red plastic tab.
Then peel sterile paper
from the bottom.
2. With one hand press the
three raised markings on
the sides of the mio™ lid.
Then raise and remove
lid.
Free tubing from slot that
3.
holds it in place. Gently
unwind tubing in counter
clockwise direction.
To Connect the Tubing:
5. Tap side of reservoir to
force any remaining
air bubbles to rise to top.
Push up on plunger rod
until bubbles are out and
small amount of insulin is
in tubing.
Turn plunger rod
6.
counter-clockwise until it
unscrews from reservoir.
Be careful not to pull
plunger rod out from
reservoir or insulin will
spill out.
To Fill the Tubing:
Press
1.
REWIND COMPLETE
appears. Insert newly
filled reservoir into
pump’s reservoir
compartment.
ACT
so that
4. Hold reservoir in
one hand. With
other hand
gently push
connector onto
top of reservoir
and turn
clockwise until it
slides and locks
into place.
Turn reservoir
2.
clockwise until it
locks in place.

Changing the mio™ Infusion Set Using a Revel™ Insulin Pump
To Fill the Tubing:
3. Press
FILL TUBING
screen. Be sure
infusion set is
NOT connected to
your body. Then
select Ye s and
press
4. Press and hold
to fill tubing. The
pump will beep as
tubing fills. Release
when you see drops at
tip of connector needle.
Then press ESC.
5.
If you see drops at
end of mio™
needle and there
are no air bubbles
in the tubing,
select Ye s and
press
6. When the FILL
CANNULA screen
appears, you are
ready to insert
the mio™ infusion
set.
7. Carefully hold needle
guard by tip. To remove it,
gently turn needle guard
& pull away to expose
needle.
Place tubing in slot on
8.
side of inserter.
ACT
to go to
ACT
.
ACT
.
ACT
ACT
Choose Insertion Site:
The best areas of the body to insert infusion set are
shown in shaded areas of this drawing.
Clean the selected insertion site with alcohol or other
antiseptic wipe.
To Insert the mio™ Infusion Set:
1. Gently peel paper to
expose adhesive.
Turn mio™ over and hold
2.
it by the lined ridges on
the sides.
With other hand pull up
3.
on center of inserter until
it clicks and locks into
place.
Place mio™ against
4.
prepared site on body.
Press the two round
indentations on side of
mio™ to insert needle.
5. Push down on center of
serter to press adhesive
against skin.

Changing the mio™ Infusion Set Using a Revel™ Insulin Pump
To Insert the mio™ Infusion Set:
6.
To remove inserter, hold
the center handle and
gently pull it straight out
away from body. Press
adhesive securely against
skin.
Place lid back onto
7.
inserter to safely discard
needle.
To Fill Cannula:
1. At FILL CANNULA
screen, press
ACT
.
2. In order to fill
empty cannula
under skin
press to enter
0.3 of insulin (if you
are using a 6mm
cannula) or 0.5
units (for 9mm
cannula). Press
ACT
.
OR
9501417-011 20111005