Medtronic MiniMed 780G System User's User guide

Page 1
MiniMed™ 780G
Includes technology developed by
SYSTEM USER GUIDE
Page 2
Page 3
MiniMed™ 780G
SYSTEM USER GUIDE
Page 4
Copyright & trademarks
© 2020 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further, Together are trademarks of Medtronic. ™* Third party brands are trademarks of their respective owners. All other brands are trademarks of a Medtronic company.
Bolus Wizard™, CareLink™, Dual Wave™, Guardian™, MiniMed™, Mio™, Quick-set™, Silhouette™, SmartGuard™, Square Wave™
, and
Sure-T™
are trademarks of Medtronic MiniMed, Inc.
ACCU-CHEK
DreaMed Diabetes includes technology developed by DreaMed Diabetes.
Wi-Fi™*
Bluetooth™*
Humalog™*
NovoLog™*
NovoRapid™*
Android™*
iOS™*
ACCU-CHEK GUIDE
and
is a trademark of
are trademarks of
DreaMed Diabetes, Ltd
Roche
.
. The MiniMed™ 780G system algorithm
M003192C021_A RF: M994838A001
MMT-1885/MMT-1895
Page 5
Contacts:
Africa: Medtronic South Africa and
Southern Africa Office Reception Tel: +27(0) 11 260 9300 Diabetes: 24/7 Helpline: 0800 633 7867 Sub-Sahara 24/7 Helpline: +27(0) 11 260 9490
Albania: Net Electronics Albania
Tel: +355 697070121
Argentina: Corpomedica S.A.
Tel: +(11) 4 814 1333 Medtronic Directo 24/7: +0800 333 0752
Armenia: Exiol LLC
Tel: +374 98 92 00 11 or +374 94 38 38 52
Australia: Medtronic Australasia Pty.
Ltd. Tel: 1800 668 670
Bangladesh: Sonargaon Healthcare
Pvt Ltd. Mobile: (+91)-9903995417 or (+880)-1714217131
Belarus: Zarga Medica
Tel: +375 29 625 07 77 or +375 44 733 30 99 Helpline: +74995830400
België/Belgique: N.V. Medtronic
Belgium S.A. Tel: 0800-90805
Bosnia and Herzegovina:
"Novopharm“ d.o.o. Sarajevo Tel: +387 33 476 444 Helpline: 0800 222 33
Epsilon Research Intern. d.o.o.
Tel: +387 51 251 037 Helpline: 0800 222 33
Brasil: Medtronic Comercial Ltda.
Tel: +(11) 2182-9200 Medtronic Directo 24/7: +0800 773 9200
Bulgaria: RSR EOOD
Tel: +359 888993083 Helpline: +359 884504344
Canada: Medtronic Canada ULC
Tel: 1-800-284-4416 (toll free/sans­frais)
Chile: Medtronic Chile
Tel: +(9) 66 29 7126 Medtronic Directo 24/7: +1 230 020 9750 Medtronic Directo 24/7 (From Santiago): +(2) 595 2942
China: Medtronic (Shanghai)
Management Co., Ltd. Landline: +86 800-820-1981 Mobile Phone: +86 400-820-1981 Calling from outside China: +86 400-820-1981
Colombia: Medtronic Latin America
Inc. Sucursal Colombia Tel: +(1) 742 7300 Medtronic Directo 24/7 (Landline): +01 800 710 2170 Medtronic Directo 24/7 (Cellular): +1 381 4902
Croatia: Mediligo d.o.o.
Tel: +385 1 6454 295 Helpline: +385 1 4881144
Medtronic Adriatic d.o.o.
Helpline: +385 1 4881120
Danmark: Medtronic Danmark A/S
Tel: +45 32 48 18 00
Deutschland: Medtronic GmbH
Geschäftsbereich Diabetes Telefon: +49 2159 8149-370 Telefax: +49 2159 8149-110 24-Stdn-Hotline: 0800 6464633
Eire: Accu-Science LTD.
Tel: +353 45 433000
España: Medtronic Ibérica S.A.
Tel: +34 91 625 05 42 Fax: +34 91 625 03 90 24 horas: +34 900 120 330
Estonia: AB Medical Group Estonia Ltd
Tel: +372 6552310 Helpline: +372 5140694
Europe: Medtronic Europe S.A. Europe,
Middle East and Africa HQ Tel: +41 (0) 21-802-7000
France: Medtronic France S.A.S.
Tel: +33 (0) 1 55 38 17 00
Hellas: Medtronic Hellas S.A.
Tel: +30 210677-9099
Hong Kong: Medtronic International
Ltd. Tel: +852 2919-1300 To order supplies: +852 2919-1322 24-hour helpline: +852 2919-6441
India: India Medtronic Pvt. Ltd.
Tel: (+91)-80-22112245 / 32972359 Mobile: (+91)-9611633007 Patient Care Helpline: 1800 209 6777
Indonesia: Medtronic International Ltd.
Tel: +65 6436 5090 or +65 6436 5000
Israel: Medtronic
Tel. (orders): +9729972440, option 3 + option 1 Tel. (product support): +9729972440, option 2 Helpline (17:00 – 08:00 daily/weekends – Israel time): 1-800-611-888
Italia: Medtronic Italia S.p.A.
Tel: +39 02 24137 261 Fax: +39 02 24138 210 Servizio assistenza tecnica: Nº verde: 800 60 11 22
Japan: Medtronic Japan Co. Ltd.
Tel: +81-3-6776-0019 24 Hr. Support Line: 0120-56-32-56
Kazakhstan: Medtronic BV in
Kazakhstan Tel: +7 727 311 05 80 (Almaty) Tel: +7 717 224 48 11 (Astana) Круглосуточная линия поддержки: 8 800 080 5001
Kosovo: Yess Pharma
Tel: +377 44 999 900 Helpline: +37745888388
Latin America: Medtronic, Inc.
Tel: 1(305) 500-9328 Fax: 1(786) 709-4244
Latvija: RAL SIA
Tel: +371 67316372 Helpline (9am to 6pm): +371 29611419
Page 6
Lithuania: Monameda UAB
Tel: +370 68405322 Helpline: +370 68494254
Macedonia: Alkaloid Kons Dooel
Tel: +389 23204438
Magyarország: Medtronic Hungária
Kft. Tel: +36 1 889 0688
Malaysia: Medtronic International Ltd.
Tel: +603 7946 9000
Middle East and North Africa:
Regional Office Tel: +961-1-370 670
Montenegro: Glosarij d.o.o.
Tel: +382 20642495
México: Medtronic Servicios S. de R. L.
de C.V. Tel (México DF): +(11) 029 058 Tel (Interior): +01 800 000 7867 Medtronic Directo 24/7 (from México DF): +(55) 36 869 787 Medtronic Directo 24/7: +01 800 681 1845
Nederland, Luxembourg: Medtronic
B.V. Tel: +31 (0) 45-566-8291 Gratis: 0800-3422338
New Zealand: Medica Pacifica
Phone: 64 9 414 0318 Free Phone: 0800 106 100
Norge: Medtronic Norge A/S
Tel: +47 67 10 32 00 Fax: +47 67 10 32 10
Philippines: Medtronic International
Ltd. Tel: +65 6436 5090 or +65 6436 5000
Poccия: ООО «Медтроник»
Tel: +7 495 580 73 77 Круглосуточная линия поддержки: 8 800 200 76 36
Polska: Medtronic Poland Sp. z o.o.
Tel: +48 22 465 6934
Portugal: Medtronic Portugal Lda
Tel: +351 21 7245100 Fax: +351 21 7245199
Puerto Rico: Medtronic Puerto Rico
Tel: 787-753-5270
Republic of Korea: Medtronic Korea,
Co., Ltd. Tel: +82.2.3404.3600
Romania: Medtronic Romania S.R.L
Tel: +40372188017 Helpline: +40 726677171
Schweiz: Medtronic (Schweiz) AG
Tel: +41 (0)31 868 0160 24-Stunden-Hotline: 0800 633333 Fax Allgemein: +41 (0)318680199
Serbia: Epsilon Research International
d.o.o. Tel: +381 113115554
Medtronic Serbia D.o.o
Helpline: +381 112095900
Singapore: Medtronic International
Ltd. Tel: +65 6436 5090 or +65 6436 5000
Slovenija: Zaloker & Zaloker d.o.o.
Tel: +386 1 542 51 11 24-urna tehnična pomoč: +386 51316560
Slovenská republika: Medtronic
Slovakia, s.r.o. Tel: +421 26820 6942 HelpLine: +421 26820 6986
Sri Lanka: Swiss Biogenics Ltd.
Mobile: (+91)-9003077499 or (+94)-777256760
Suomi: Medtronic Finland Oy
Tel: +358 20 7281 200 Help line: +358 400 100 313
Sverige: Medtronic AB
Tel: +46 8 568 585 20 Fax: +46 8 568 585 11
Taiwan: Medtronic (Taiwan) Ltd.
Tel: 02-21836000 Toll free: +886-800-005285
Thailand: Medtronic (Thailand) Ltd.
Tel: +662 232 7400
Türkiye: Medtronic Medikal Teknoloji
Ticaret Ltd. Sirketi. Tel: +90 216 4694330
USA: Medtronic Diabetes Global
Headquarters 24-Hour Technical Support: +1-800-646-4633 To order supplies: +1-800-843-6687
Ukraine: Med Ek Service TOV
Tel: +380 50 3311898 or +380 50 4344346 Лінія цілодобової підтримки: 0 800 508 300
United Kingdom: Medtronic Ltd.
Tel: +44 1923-205167
Österreich: Medtronic Österreich
GmbH Tel: +43 (0) 1 240 44-0 24 – Stunden – Hotline: 0820 820 190
Česká republika: Medtronic Czechia
s.r.o. Tel: +420 233 059 111 Non-stop helpLine (24/7): +420 233 059 059 Zákaznický servis (8:00 - 17:00): +420 233 059 950
Page 7
Contents
 ■   Safety and indications
3 Using this guide
3Conventions
4Emergency kit
5User safety
5Intended use
6 Contraindications
6Risks and side effects
9General warnings
9Pump
13 Reservoir and infusion sets
14 Sensor and serter
15 Transmitter
15 Meter
15 Exposure to magnetic fields and radiation
16 General precautions
16 Waterproof capabilities
17 Electrostatic discharge
17 Extreme temperatures
18 Skin care products
18 Infusion sets and sites, sensor, transmitter, and meter
18 Adverse reactions
18 Security precautions
Contents
v
Contents
Page 8
20 Insulin guidelines
20 Consumables
22 Other MiniMed 780G system devices
22 Accessories
 ■   Pump basics
26 Using the buttons
27 Sleep mode
28 Pump delivery system
28 Infusion set
29 Reservoir
29 Pump
29 Inserting the battery
31 Startup settings
32 Home screen in Manual mode
33 Shortcuts from the Home screen
34 Status icons
38 Menu screen
39 Menu map
41 Sound & Vibration screen
41 Status screen
42 Viewing basal delivery information
43 Settings covering a 24-hour period
 ■   Setting up insulin delivery
49 Setting up basal insulin
49 Basal rate
49 Max basal rate
51 Basal patterns
52 Temp basal rates
54 Suspending all insulin delivery and resuming basal insulin delivery
57 Entering a BG value
Page 9
57 Setting up bolus delivery
57 About bolus deliveries
58 Bolus delivery options
58 Max bolus
59 Bolus Wizard feature
65 Normal bolus
68 Stopping a normal bolus delivery
 ■   Reservoir and infusion set
73 Setting up the reservoir and infusion set
74 Removing the reservoir and rewinding the pump
76 Filling the reservoir and connecting it to the infusion set tubing
80 Placing the reservoir into the pump and filling the tubing with insulin
84 Inserting the infusion set into the body
85 Filling the cannula
86 Disconnecting the infusion set
86 Reconnecting the infusion set
 ■   Paired devices
89 Setting up the Accu-Chek Guide Link meter
90 Pairing the pump and meter
92 Pairing the pump and transmitter
95 MiniMed Mobile app
95 Uploading device data to CareLink software
 ■   Continuous glucose monitoring
99 CGM overview
100 Home screen with CGM in Manual mode
101 SG alert settings
101 High SG settings
102 Low SG settings
109 Setting up CGM
Contents
Contents vii
Page 10
109 Turning on the Sensor feature
110 Setting up the High SG settings
113 Setting up the low SG settings
115 Manually resuming basal insulin delivery during a Suspend before low or
Suspend on low event
116 Inserting the sensor
117 Connecting the transmitter to the sensor
117 Starting the sensor
118 Calibrating the sensor
119 Reconnecting the sensor
120 Deactivating the Sensor feature
120 Using CGM
121 The sensor graph when using CGM
122 Silencing sensor alerts
 ■   SmartGuard
127 Introduction
128 Auto Basal
129 Auto Correction
129 Giving a bolus when the SmartGuard feature is active
129 Preparing to set up the SmartGuard feature
130 Setting up the SmartGuard feature
131 Conditions to activate the SmartGuard feature
132 Suspend before low and Suspend on low features while using the
SmartGuard feature
132 SmartGuard Checklist
134 Home screen with the SmartGuard feature
134 Using the SmartGuard feature
134 The sensor graph with the SmartGuard feature
136 Entering a BG value in the SmartGuard feature
136 Delivering a bolus in the SmartGuard feature
141 Setting a temp target
Page 11
142 Staying in the SmartGuard feature
144 Exiting the SmartGuard feature
144 Returning to the SmartGuard feature after an exit
145 Using Block mode with the SmartGuard feature
145 Alert silence feature
 ■   General settings
149 Time and date
149 Display options
150 Block mode
151 Self Test
152 Manage Settings
153 Saving the settings
153 Restoring the settings
154 Clearing the settings
154 Clearing the active insulin
155 Viewing the pump setting history
156 Auto suspend
156 Language
 ■   History and graph
159 Introduction
159 History & Graph menu
159 History
159 Summary screen
160 Understanding the Summary screen
163 Daily History screen
164 Alarm History screen
164 SG Review screen
166 Graph screen
166 Time in Range screen
Contents
Contents ix
Page 12
 ■   Notifications and reminders
171 Notifications in the MiniMed Mobile app
171 Reminders
172 Personal reminders
173 Bolus BG Check reminder
173 Missed Meal Bolus reminder
174 Low Reservoir reminder
175 Set Change reminder
175 Calibration reminder
176 Alarms, alerts, and messages
177 Alarms
178 Alerts
179 Messages
179 Pump alarms, alerts, and messages
 ■   Additional basal features
183 Preset temp basal rates
185 Starting a preset temp basal delivery
186 Canceling a temp basal or preset temp basal
186 Additional basal patterns
186 Adding an additional basal pattern
187 Editing, copying, or deleting a basal pattern
188 Changing from one basal pattern to another
 ■   Additional bolus features
191 Bolus types
192 Bolus type example
192 Bolus settings
193 Bolus increment
193 Bolus speed
194 Changing the Bolus Wizard settings
Page 13
194 Changing the carb ratio
194 Changing the insulin sensitivity factor
195 Changing the BG target
195 Changing the active insulin time
196 Square Wave bolus
196 Turning the Square Wave bolus feature on or off
197 Delivering a Square Wave bolus using the Bolus Wizard feature
198 Delivering a Square Wave bolus using the Manual bolus feature
200 Dual Wave bolus
200 Turning the Dual Wave bolus feature on or off
200 Delivering a Dual Wave bolus using the Bolus Wizard feature
202 Delivering a Dual Wave bolus using the Manual bolus feature
204 Easy bolus
204 Setting up the Easy bolus feature
205 Delivering a bolus using the Easy bolus feature
206 Preset bolus
206 Setting up and managing preset bolus deliveries
208 Delivering a preset bolus
209 Stopping a Square Wave or Dual Wave bolus delivery
 ■   Troubleshooting
215 Pump issues
219 Sensor issues
 ■   Maintenance
223 Pump maintenance
223 Cleaning the pump
224 Storing the pump
226 Pump disposal
226 Meter maintenance
226 Unpairing a meter from the pump
227 Deleting the pump from a meter
Contents
Contents xi
Page 14
227 Transmitter and sensor maintenance
227 Unpairing the transmitter from the pump
229 Disconnecting the transmitter from the sensor
229 Removing the sensor
229 Cleaning the transmitter
229 Storing the transmitter
229 Removing the battery
 ■   Appendix A: List of alarms, alerts, and messages
233 Pump alarms, alerts, and messages
251 CGM (sensor) alarms, alerts, and messages
263 SmartGuard feature alerts and messages
268 CareLink software alert and message
 ■   Appendix B: Product specifications
273 Specifications and default settings
273 Alarm and alert escalation
274 Altitude range
274 Backlight
274 Basal delivery
275 BG meter reading
275 Bolus delivery
275 Bolus Wizard feature default settings
276 Bolus Wizard feature specifications
279 Carb ratios
279 Delivery accuracy
280 Easy bolus feature
281 Environmental conditions
281 Essential performance
281 Filling the infusion set and cannula
282 Infusion pressure
282 Insulin delivery default settings
Page 15
283 Low Reservoir reminder
283 Max bolus
284 Normal bolus
284 Occlusion detection
284 Percent temp basal
284 Program safety checks
285 Pump dimensions
285 Pump memory
285 Pump weight
285 Sensor default settings
287 Sound frequency
288 IEC60601-1-2:4th Edition notice
288 IEC60601-1-2:4th Edition; Special EMC Precautions for Medical Electrical
Equipment
288 IEC60601-1-2:4th Edition; 5.2.1.1
288 Guidance and manufacturer's declaration
293 Wireless communication
293 Open Source Software disclosure
294 Icon table
 ■   Glossary
 ■   Index
Contents
Contents xiii
Page 16
Page 17
Safety and indications
Page 18
Page 19
Safety and indications
This user guide describes the operation of the MiniMed 780G system with smart device connectivity and SmartGuard technology. SmartGuard technology adjusts insulin delivery based on sensor glucose (SG) readings. The MiniMed 780G insulin pump operates in Manual mode when the SmartGuard feature is inactive.
Consult a healthcare professional before starting insulin pump therapy.
Using this guide
Use the table of contents at the beginning of the user guide and the index at the end of the user guide to locate specific information.
For the glossary, which has definitions of terms and acronyms used, see page 297.
Conventions
Convention Definition
Select
Press
to activate a screen item, accept a value, or initiate an
action.
Select and hold
Press and hold
to perform an action.
Press Press and release a button.
Press and hold
Press and hold a button.
Bold text Indicates screen items and buttons, such as "Select Next to
continue."
Safety and indications 3
Safety and indications
Page 20
Convention Definition
X Indicates a value that might appear differently on the pump
screen.
Note
Note: A note provides helpful information.
Caution
CAUTION: A caution informs of a potential hazard which, if not avoided, might result in minor or moderate injury, or damage to the equipment.
WARNING
WARNING: A warning informs of a potential safety hazard which, if not avoided, may result in serious injury or death. It may also describe potential serious adverse reactions.
For instructions about setting up devices on the MiniMed 780G system, such as a sensor or infusion set, refer to the user guide for the related device.
Emergency kit
Keep an emergency kit available at all times to confirm that necessary supplies are ready. Tell a family member or friend where to find the emergency kit.
When traveling, test blood glucose (BG) more frequently to accommodate for changes in activity levels and meal times.
Include the following items in the emergency kit:
Rapid-acting glucose tablets
BG monitoring supplies
Urine or blood ketone monitoring supplies
Extra MiniMed infusion set and MiniMed reservoir
Extra new AA lithium or alkaline batteries, or fully charged NiMH batteries
4
Chapter 1
Page 21
Insulin syringe and rapid-acting U-100 insulin (with dosage instructions from a healthcare professional)
Adhesive dressing
•Glucagon
WARNING: Do not use the Bolus Wizard feature to calculate a bolus for a period of time after giving a manual injection of insulin by syringe or pen. Manual injections are not accounted for in the active insulin amount. Using the Bolus Wizard feature too soon after a manual injection may result in over-delivery of insulin and may cause hypoglycemia. Consult a healthcare professional for how long to wait after a manual injection before using the Bolus Wizard feature.
WARNING: Do not use the SmartGuard feature for a period of time after giving a manual injection of insulin by syringe or pen. Manual injections are not accounted for when the SmartGuard feature is active. Using the SmartGuard feature too soon after a manual injection may result in over-delivery of insulin and may cause hypoglycemia. Consult a healthcare professional for how long to wait after a manual injection before using the SmartGuard feature.
User safety
Intended use
MiniMed 780G system
The MiniMed 780G insulin pump is indicated for use by patients age 7-80 years with Type 1 diabetes, whose total daily dose of insulin is 8 units per day or more.
The MiniMed 780G system is intended for the continuous delivery of basal insulin at selectable rates, and the administration of insulin boluses at selectable amounts. The system is also intended to continuously monitor glucose values in the fluid under the skin. The MiniMed 780G system includes SmartGuard technology, which can be programmed to provide an automatic adjustment of insulin delivery based
Safety and indications
Safety and indications 5
Page 22
on continuous glucose monitoring (CGM) and can suspend the delivery of insulin when the SG value falls below, or is predicted to fall below, predefined threshold values.
Contraindications
Pump therapy is not recommended for people whose vision or hearing does not allow for the recognition of pump signals, alerts, or alarms.
Insulin pump therapy is not recommended for persons who are unwilling to perform at least four BG meter readings per day.
The SmartGuard feature cannot be used for people who require less than eight units or more than 250 units of insulin a day.
Pump therapy is not recommended for people who are unwilling or unable to maintain contact with their healthcare professional.
Pump therapy is not recommended for people with a significant cognitive or physical impairment that affects their ability to safely operate the pump, including blindness, hearing impairment, or lack of physical dexterity.
Pump therapy is not recommended for children who are not under the care of a parent or caregiver who is capable of safely operating the pump for the patient.
Risks and side effects
Risks related to insulin administration and pump use
Risks related to insulin infusion and potential interruptions of insulin delivery include:
•Hypoglycemia
Hyperglycemia
Diabetic ketoacidosis
•Seizure
•Coma
•Death
Risks related to insulin pump infusion set
Risks related to insulin pump infusion set use include:
Localized infection
6
Chapter 1
Page 23
Skin irritation or redness
Bruising
Discomfort or pain
Bleeding
Irritation
•Rash
Occlusions that may interrupt insulin delivery and lead to hyperglycemia and diabetic ketoacidosis
Follow the instructions in the provided user guides for the insertion and care of infusion sets. If an infusion site becomes irritated or inflamed, dispose of the infusion set in a sharps container, and select a different location to insert a new infusion set.
Risks related to sensor use
Risks related to sensor use include:
Skin irritation
Allergic reaction
Bruising
•Discomfort
•Redness
Bleeding
•Pain
•Rash
•Infection
•Raised bump
Appearance of a small "freckle-like" dot where needle was inserted
Fainting secondary to anxiety or fear of needle insertion
Soreness or tenderness
Swelling at insertion site
Sensor fracture, breakage or damage
Safety and indications
Safety and indications 7
Page 24
Minimal blood splatter associated with sensor needle removal
Residual redness associated with adhesive, tape, or both
Scarring
Specific risks related to sensor use
Taking medications that contain paracetamol, including, but not limited to fever reducers and cold medicine, while wearing the sensor, may falsely raise sensor glucose readings and result in an over-delivery of insulin. The level of inaccuracy depends on the amount of paracetamol active in the body and may be different for each person. If paracetamol is taken, stop the use of the medication before using SG readings to make treatment decisions. Use additional BG meter readings to verify glucose levels, turn off the Auto correction feature, and consider turning off the SmartGuard feature. Check the label of any medication to confirm whether paracetamol is an active ingredient.
Do not use SG values to make therapy treatment decisions while the pump is in Manual mode. SG values can only be used to make therapy treatment decisions while the SmartGuard feature is active. SG and BG values may differ. If the SG reading is low or high, or there are symptoms of low or high glucose, confirm the SG reading with a BG meter prior to making therapy decisions to avoid hypoglycemia or hyperglycemia.
For persons seven to thirteen years of age, sensor insertion is approved for in the abdomen and buttocks. Do not insert the sensor into any other location.
For persons fourteen years and older, sensor insertion is approved for the abdomen and back of the upper arm. Do not insert the sensor into any other location.
Risks related to meter use
For the most current risks, see the User's Manual that came with the device.
Risks related to serter use
The serter contains small parts and may pose a choking hazard that can result in serious injury or death.
Side effects include discomfort and skin irritation at the insertion site.
8
Chapter 1
Page 25
Risks related to the MiniMed 780G system
•Hypoglycemia
Hyperglycemia
Diabetic ketoacidosis
•Seizure
•Coma
•Death
Removing the pump for temporary storage
If there is a need or desire to remove the pump, use the following guidelines:
Write down the current basal rates and use the Save Settings feature. For more information, see Saving the settings, on page 153.
Remove the battery. For more information, see Storing the pump, on page 224.
If the pump is disconnected for less than one hour, an insulin adjustment may not be required. If the pump is disconnected for more than one hour, consult a healthcare professional to determine an alternate method of insulin delivery.
General warnings
Pump
Do not use the pump in the presence of anesthetic mixtures that include oxidizing agents such as oxygen or nitrous oxide. Exposure to these conditions may damage the pump and result in serious injury.
Always use the fingertip for blood samples intended for sensor calibration. Do not use blood samples from the palm to calibrate the sensor. The palm has not been studied for use with the SmartGuard feature and the performance of the system using such blood samples is not known.
When the SmartGuard feature is active, SG readings are used to calculate basal insulin delivery and correction boluses. Do not use SG readings to make treatment decisions while the pump is in Manual mode. SG and BG values may differ. Sensor performance may occasionally vary from sensor to sensor and in different situations for a sensor, such as on the first day of use.
A BG meter reading is required in the following situations:
Safety and indications
Safety and indications 9
Page 26
Before a correction bolus is given in Manual mode.
The SG reading is lower than expected.
The SG reading is higher than expected.
Suspected hypoglycemia or symptoms of hypoglycemia.
Suspected hyperglycemia or symptoms of hyperglycemia.
Suspected diabetic ketoacidosis or symptoms of diabetic ketoacidosis.
Do not use SG readings to make treatment decisions while the pump is in Manual mode.
Do not rely on the pump tones or vibrations to navigate the pump screens or menus. Relying on pump tones or vibrations may result in incorrect menu or setting selection. Always view the pump screen when selecting menus and entering information into the system.
Only use rapid-acting U-100 insulin (Humalog, NovoLog, and NovoRapid) prescribed by a healthcare professional for use with an infusion pump. Use of any other drug or medication in the reservoir can cause serious injury.
Confirm that the infusion set is disconnected from the body before rewinding the pump or filling the infusion set tubing. Never insert the reservoir into the pump while the tubing is connected to the body. Doing so may result in an accidental infusion of insulin, which may cause hypoglycemia.
Do not insert the reservoir before rewinding the pump. Doing so may result in an accidental infusion of insulin, and may result in hypoglycemia.
Do not use the MiniMed 780G insulin pump or additional system devices next to other electrical equipment, which may cause interference. This includes mobile communication devices such as cell phones, GPS navigation systems, anti-theft systems, and any electrical equipment that has an output transmitter power greater than 1W. The recommended separation distance between the insulin pump and common RF emitters is 30 cm (12 inches). For more information about recommended separation distance guidelines between the insulin pump and common RF emitters, see Guidance and manufacturer's declaration, on page 288. Other electrical equipment that may compromise normal system operation has been contraindicated. For more information, see Exposure to magnetic fields and radiation, on page 15.
10
Chapter 1
Page 27
Do not unscrew or retighten the tubing connector on the reservoir while the infusion set is connected to the body. Doing so may result in an accidental infusion of insulin, and may cause hypoglycemia.
Do not use standard Luer sets with the MiniMed 780G system. Only use MiniMed reservoirs and infusion sets that are specifically designed for use with the MiniMed 780G system.
Do not change or modify the MiniMed reservoir or the MiniMed infusion set. Modification of these components may cause serious injury, interfere with device operation, and void the warranty.
Do not rely on preset pump alarms or reminders alone to check BG levels. Set additional reminders on other devices, such as a cell phone.
Do not change or modify the internal RF transmitter or antenna. Doing so may interfere with the safe operation of the equipment.
The MiniMed 780G system is only approved for use with the Guardian Link (3) transmitter with Bluetooth wireless technology (MMT-7911WW). The Guardian Link (3) transmitter can be identified by the “GL3” marking on the top of the device. Use of a transmitter not approved for communication with the pump may cause damage to system components and may result in inaccurate SG readings.
•If other devices that employ radio frequencies are in use, such as cell phones, cordless phones, walkie-talkies, and wireless networks, they may prevent communication between the transmitter and the insulin pump. This interference does not cause any incorrect data to be sent and does not cause any harm to devices. Moving away from, or turning off, these other devices may enable communication. Contact a local Medtronic support representative if RF interference continues.
Special Precautions regarding Electromagnetic Compatibility (EMC): This body­worn device is intended to be operated within a residential, domestic, public or work environment, where common levels of radiated “E” (V/m) or “H” fields (A/m) exist. Technologies that emit these fields include: cellular phones, wireless technology, electric can openers, microwaves, and induction ovens. The MiniMed 780G system generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the provided instructions, may cause harmful interference to radio communications.
Safety and indications
Safety and indications 11
Page 28
Portable and mobile RF communications equipment can affect the operation of the MiniMed 780G system. If interference occurs, move away from the RF transmitter.
The MiniMed 780G insulin pump can generate, use, and radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. If the MiniMed 780G insulin pump does cause interference to radio or television reception, try to correct the interference by one or more of the following measures:
Decrease the distance between the transmitter and the insulin pump to
1.8 meters (6 feet) or less.
Decrease the distance between the meter and the insulin pump to 1.8 meters (6 feet) or less.
Increase the separation between the transmitter and the device that is receiving/emitting interference.
The safety of the MiniMed 780G system has not been studied in persons with impaired kidney function. Persons with kidney disease should consult a healthcare professional to determine if the potential benefits of pump therapy outweigh the risks.
Monitor for diabetic retinopathy. During the beginning of insulin pump therapy, rapid improvement in glucose control and reduction in A1c may result in worsening of existing diabetic retinopathy. Use of the MiniMed 780G system has been associated with rapid improvement in glucose control. Monitor for diabetic retinopathy with retinal eye examinations and if necessary adequate treatment must be performed by your healthcare professional before beginning a treatment with the MiniMed 780G insulin pump.
The safety of the MiniMed 780G system has not been studied in pregnant women, persons with type 2 diabetes, or in persons using other anti­hyperglycemic therapies that do not include insulin. Persons in these situations should consult a healthcare professional to determine if the potential benefits of pump therapy outweigh the risks.
12
Chapter 1
Page 29
The safety of using the Suspend before low and Suspend on low features in patients who have no pump experience is not known. The Suspend before low and Suspend on low features should not be used if insulin pump settings have not been previously established. Insulin pump settings include basal rates, insulin to carb ratio, and insulin sensitivity factors. Consult a healthcare professional before using the Suspend before low or Suspend on low features.
Reservoir and infusion sets
See the user guides that came with the device for the most current warnings related to the reservoir and infusion set.
If insulin, or any other liquid, gets inside the tubing connector, it can temporarily block the vents that allow the pump to properly fill the infusion set. This may result in the infusion of too little or too much insulin, and may result in hyperglycemia or hypoglycemia. If this occurs, start over with a new reservoir and infusion set.
If a BG level reading is unexpectedly high during the infusion of insulin or if an occlusion alarm occurs, check the infusion set for clogs and leaks.
If in doubt, change the infusion set in case the soft cannula is dislodged, crimped, or partially clogged. Consult a healthcare professional to create a plan for rapid insulin replacement in the event this occurs. Check BG to confirm that the appropriate amount of insulin has been administered.
Only use reservoir and infusion sets manufactured or distributed by Medtronic Diabetes. The pump has been tested to operate when used with compatible reservoirs and infusion sets. Medtronic Diabetes cannot guarantee appropriate operation if the pump is used with reservoirs or infusion sets offered by third parties. Medtronic Diabetes is not responsible for any injury or pump malfunction that may occur in association with the use of incompatible components.
The low SG alert functionality is distinct from the automated insulin dosing function of the MiniMed 780G system. When using the SmartGuard feature, the MiniMed 780G system has been shown to be safe and effective for its intended use in this population. However, do not rely solely on the use of a low SG value for “Alert on Low” or “Alert before Low” for alerts set at 2.8 mmol/L and 3.3 mmol/L. A low SG alert may not reflect the user’s true BG at these levels, or may not alert. Do not ignore symptoms of hypoglycemia. If
Safety and indications
Safety and indications 13
Page 30
symptoms don't match SG readings, confirm SG with a BG meter reading. Solely relying on these SG alerts and readings for treatment decisions could result in missing severe hypoglycemic (low BG) events.
Sensor and serter
For the most current warnings, see the user guide that came with the device.
Keep the sensor and serter away from children. These products contain small parts and may pose a choking hazard that may result in serious injury or death.
A retractable needle is attached to the sensor, and minimal blood splatter may occur during insertion. For healthcare professionals and caregivers, cover the sensor with sterile gauze to minimize contact with blood. Keep as much distance as possible from the patient when removing the needle.
Do not attempt to remove the sensor if the sensor is broken. While there is no evidence of a sensor breaking in a patient’s body, sensor breakage can result in serious injury. If the sensor breaks while inserted, contact a healthcare professional for assistance to safely remove the sensor.
Sensors are sterile and non-pyrogenic, unless the package has been opened or damaged. Do not use the sensor if the sterile package has been opened or damaged. Use of an unsterile sensor may cause site infection.
Instructions for using the One-press serter (MMT-7512) are different from other Medtronic insertion devices. Failure to follow directions, or using a different serter, may result in improper insertion, pain, or injury.
Keep the needle housing within sight at all times to avoid an accidental needlestick or puncture.
Confirm that the sensor is securely placed in the serter to avoid improper insertion, pain, or minor injury.
Watch for bleeding at the insertion site (under, around, or on top of the sensor). If bleeding occurs, do the following:
a. Apply steady pressure, using sterile gauze or a clean cloth placed on top
of the sensor, for up to three minutes. The use of unsterile gauze can cause site infection.
14
Chapter 1
Page 31
b. If bleeding stops, connect the transmitter to the sensor. If bleeding does
not stop, do not connect the transmitter to the sensor because blood can get into the transmitter connector, and may damage the device.
If bleeding continues, causes excessive pain or discomfort, or blood is significantly visible in the plastic base of the sensor, do the following:
a. Remove the sensor and continue to apply steady pressure until the
bleeding stops. Discard the sensor in a sharps container.
b. Check the site for redness, bleeding, irritation, pain, tenderness, or
inflammation. Treat based on instructions from a healthcare professional.
c. Insert a new sensor in a different location.
The safety of sensor use in critically ill patients is not known. Sensor use in critically ill patients is not recommended.
Transmitter
See the user guide included with the device for the most current warnings related to transmitter use.
Do not allow children to put small parts in their mouth. This product poses a choking hazard for young children.
Meter
For the most current warnings, see the User's Manual that came with the device.
Use the fingertip for blood samples collected for sensor calibration while the SmartGuard feature is active. Only use blood samples collected from the fingertip to calibrate the sensor. Blood samples from other locations, such as the palm or forearm, have not been studied, and the accuracy of these samples is unknown.
Exposure to magnetic fields and radiation
Do not expose the pump, transmitter, or sensor to MRI equipment, diathermy devices, or other devices that generate strong magnetic fields (for example, x­ray, CT scan, or other types of radiation). Strong magnetic fields can cause the system to malfunction, and result in serious injury. If the pump is exposed to a strong magnetic field, discontinue use and contact a local Medtronic support representative for further assistance.
Safety and indications
Safety and indications 15
Page 32
Magnetic fields, and direct contact with magnets, may affect the accurate functioning of the system which may lead to health risks such as hypoglycemia or hyperglycemia.
Remove the pump, sensor, transmitter, and meter before entering a room with x-ray, MRI, diathermy, or CT scan equipment. The magnetic fields and radiation in the immediate vicinity of this equipment can make the devices nonfunctional or damage the part of the pump that regulates insulin delivery, possibly resulting in over-delivery and severe hypoglycemia.
Do not expose the pump to a magnet, such as pump cases that have a magnetic clasp. Exposure to a magnet may interfere with the motor inside the pump. Damage to the motor can cause the device to malfunction, and result in serious injury.
Do not send the pump or transmitter through an x-ray scanning machine. The radiation can damage the pump components that regulate insulin delivery, and may result in over-delivery of insulin and hypoglycemia.
All system components, including the pump, transmitter, and sensor, must be removed prior to being screened with a full-body scanner. To avoid system removal, request an alternative screening method, if necessary.
Carry the Medical emergency card provided with the device when traveling. The Medical emergency card provides critical information about airport security systems and pump use on an airplane. Not following the guidance on the Medical emergency card may result in serious injury.
General precautions
Check BG levels at least every 12 hours. Pump alarms do not notify the patient of leaks in the infusion set or degradation of insulin. If BG is out of range, check the pump and the infusion set to confirm that the necessary amount of insulin is being delivered.
Waterproof capabilities
The pump is waterproof at the time of manufacture and when the reservoir and tubing are properly inserted. It is protected against the effects of being underwater to a depth of up to 3.6 meters (12 feet) for up to 24 hours.
16
Chapter 1
Page 33
If the pump is dropped, hit against a hard object, or otherwise damaged, the waterproof characteristics of the outer casing of the pump may be compromised. If the pump is dropped or might be damaged, carefully inspect it to confirm that there are no cracks before exposing the pump to water.
This waterproof capability rating applies only to the pump.
If water may have entered the pump or other pump malfunction is observed, check BG and treat high BG as necessary using an alternative source of insulin. Contact a local Medtronic support representative for further assistance, and consult a healthcare professional about high or low BG levels or with any other questions about care.
Electrostatic discharge
Very high levels of ESD can result in a reset of the pump's software and a pump error alarm. After clearing the alarm, confirm that the pump is set to the correct date and time, and that all other settings are programmed to the desired values. Following a pump reset, the SmartGuard feature will be unavailable for five hours to allow active insulin to be updated.
For more information on pump alarms, see Pump alarms, alerts, and messages, on page 233. Contact a local Medtronic support representative with any problems entering pump settings.
Extreme temperatures
Exposure to extreme temperatures can damage the device. Avoid the following conditions:
Temperatures above 40°C (104°F) or below 5°C (41°F).
Insulin solutions freeze near 0°C (32°F) and degrade at temperatures higher than 37°C (98.6°F). In cold weather, wear the pump close to the body and cover it with warm clothing. In a warm environment, take measures to keep the pump and insulin cool.
Do not steam, sterilize, autoclave, or otherwise heat the pump.
Safety and indications
Safety and indications 17
Page 34
Skin care products
Some skin care products, such as lotion, sunscreen, and insect repellents, can damage the plastic in the pump case. After using skin care products, wash hands prior to handling the pump. If a skin care product comes into contact with the pump, wipe it off as soon as possible with a damp cloth and mild soap. For instructions on cleaning the pump, see Cleaning the pump, on page 223.
Infusion sets and sites, sensor, transmitter, and meter
Refer to the corresponding device user guide for all warnings, precautions, and instructions relating to the device. Failure to reference the corresponding device user guide can result in minor injury, or damage to the device.
Adverse reactions
Refer to the sensor user guide for adverse reactions related to sensor use. Failure to reference the sensor user guide may result in minor injury, or damage to the sensor.
Security precautions
The MiniMed 780G insulin pump system is designed with security features to help keep the system and the data secure. These security features in the insulin pump system are set in the factory and ready to use when the insulin pump is received. For example, when the pump communicates with other devices in the system, such as the BG meter, transmitter, or compatible mobile device, the data that it is sending and receiving is encrypted and protected by cyclic redundancy checks. This helps prevent other people from being able to see system data, or to interfere with insulin pump therapy.
To help keep the system secure, follow these instructions:
Do not leave the insulin pump or the paired devices unattended.
Do not share the pump, transmitter, or BG meter serial number.
Do not connect the pump to any third-party devices not authorized by Medtronic.
Do not use any software not authorized by Medtronic to control the system.
Be attentive to pump notifications, alarms, and alerts because they may indicate that someone else is trying to connect to or interfere with the device.
18
Chapter 1
Page 35
Disconnect the Blue Adapter from the computer whenever it is not being used.
Use good cyber security practices; use anti-virus software and keep computer software up to date.
Refer to the MiniMed Mobile App User Guide for information on how to keep the compatible mobile device safe to use with the Medtronic devices.
The pump only communicates with paired devices. The short time that it takes to pair the pump with other devices is a sensitive time for security. During this time, it is possible for an unintended device to pair with the pump. While Medtronic has designed security features into the system to prevent this, to keep the system safe during pairing always follow these instructions:
Pair the transmitter, BG meter, or the compatible mobile device with the pump away from other people and devices.
When the transmitter successfully pairs with the pump, the green LED on the transmitter stops blinking. If the green LED on the transmitter continues to blink for several minutes or more after it is successfully paired, it may have been paired with an unintended device. See Unpairing the transmitter from the pump, on page 227 to delete the transmitter from the pump and then follow the steps to pair it again.
After pairing the BG meter or the compatible mobile device with the pump, make sure that the BG meter or compatible mobile device indicates that pairing was successful.
Consult a healthcare professional if there are symptoms of severe hypoglycemia or diabetic ketoacidosis, or suspect that the insulin pump settings, or insulin delivery changed unexpectedly.
If there is a concern that someone else is trying to connect to or interfere with the device, stop using it and contact a local Medtronic support representative immediately.
Safety and indications
Safety and indications 19
Page 36
Insulin guidelines
WARNING: Do not insert an insulin-filled reservoir into the pump, or connect an insulin-filled infusion set into the body, when training with the system. Doing so may result in the unintentional infusion of insulin, which may result in hypoglycemia. Start insulin therapy only when directed by a healthcare professional.
The MiniMed 780G system has been studied with, and is intended for use with, the following rapid-acting U-100 insulins:
•U-100 NovoLog
•U-100 Humalog
•U-100 NovoRapid
The use of any other insulin in the MiniMed 780G system has not been tested and is contraindicated for use with this device.
WARNING: Only use rapid-acting U-100 insulin (Humalog, NovoLog, and NovoRapid), as prescribed by a healthcare professional, in the MiniMed 780G system. Use of the incorrect type of insulin, or insulin with a greater or lesser concentration, may result in over-delivery or under-delivery of insulin, which may result in hypoglycemia or hyperglycemia. Consult a healthcare professional with any questions about the type of insulin that is compatible with the pump.
Consumables
The pump uses disposable, single-use, MiniMed reservoirs and infusion sets for insulin delivery.
20
Chapter 1
Page 37
WARNING: Only use reservoir and infusion sets manufactured or distributed by Medtronic Diabetes. The pump has undergone extensive testing to confirm appropriate operation when used with compatible reservoirs and infusion sets manufactured or distributed by Medtronic Diabetes. Medtronic Diabetes cannot guarantee appropriate operation if the pump is used with reservoirs or infusion sets offered by third parties and therefore Medtronic Diabetes is not responsible for any injury or malfunctioning of the pump that may occur in association with such use.
Reservoirs–Use the MiniMed reservoir MMT-332A, 3.0 mL (300-unit) or MiniMed reservoir MMT-326A, 1.8 mL (180-unit).
Infusion sets–Contact a healthcare professional for help in choosing a Medtronic Diabetes infusion set. Change the infusion set every two to three days per the infusion set manufacturer's instructions.
The following table lists the compatible infusion sets. The MMT numbers may change if other compatible infusion sets become available.
Type
MMT number
a
MiniMed Quick-set infusion set MMT-386, MMT-387, MMT-394, MMT-396,
MMT-397, MMT-398, MMT-399
MiniMed Silhouette infusion set MMT-368, MMT-377, MMT-378, MMT-381,
MMT-382, MMT-383, MMT-384
MiniMed Sure-T infusion set MMT-862, MMT-864, MMT-866, MMT-874,
MMT-876, MMT-884, MMT-886
MiniMed Mio infusion set MMT-921, MMT-923, MMT-925, MMT-941,
MMT-943, MMT-945, MMT-961, MMT-963, MMT-965, MMT-975
MiniMed Mio 30 infusion set MMT-905, MMT-906
MiniMed Mio Advance infusion set MMT-211, MMT-212, MMT-213, MMT-231,
MMT-232, MMT-233, MMT-242, MMT-243, MMT-244
a. All MMT numbers include "a" versions. For example, MMT-386 and MMT-386a.
Safety and indications
Safety and indications 21
Page 38
Other MiniMed 780G system devices
Accu-Chek Guide Link meter–The MiniMed 780G pump is compatible with the Accu-Check Guide Link meter. The meter pairs with the pump, allowing the BG meter to send meter readings to the pump. This device may not be available in all countries.
Guardian Link (3) transmitter (MMT-7911WW)–The transmitter pairs with the pump, collects data measured by the sensor, and wirelessly sends this data to monitoring devices. This device is required for CGM.
Guardian Sensor (3) (MMT-7020)–The sensor is a disposable, single-use device inserted just below the skin to measure glucose levels in interstitial fluid. This device is required for CGM. The Guardian Sensor (3) (MMT-7020) glucose sensor is the only sensor compatible with the MiniMed 780G insulin pump and Guardian Link (3) transmitter.
Accessories
The following accessories may be used with the MiniMed 780G system.
Pump clip–attaches to a belt and can be used to open the battery compartment.
Activity guard (ACC-1520)–helps to prevent the reservoir from being rotated or removed from the pump during physical activities.
MiniMed Mobile app (MMT-6101 for Android or MMT-6102 for iOS)– provides a secondary display of insulin pump data and CGM, and uploads system data to CareLink software. The app can be installed on multiple mobile devices, but only one device can be paired with the pump at a time.
Blue Adapter (ACC-190)–uploads system data to CareLink software through a USB port on a computer. Refer to the CareLink software user guide for setup and operation of the Blue Adapter.
22
Chapter 1
Page 39
Pump basics
Page 40
Page 41
Pump basics
This chapter provides information about the basic features, buttons, and screens of the MiniMed 780G insulin pump.
Pump basics
25
Pump basics
Page 42
Using the buttons
u
p
bac
k
not
ification
light
lef
t
graph
;
sleep
m
ode
right
sel
e
c
t
d
o
w
n
CAUTION: Do not use sharp objects to press the pump buttons. The use of sharp objects can damage the pump.
Note: The notification light flashes when the pump has an alarm or alert. The notification light is not visible unless it flashes.
26
Chapter 2
Page 43
The following table describes how to use the pump buttons.
Button Description
Press this button to show the Menu screen from the Home screen, and to select the currently highlighted menu option.
, , , or Press these buttons to scroll up or down, to move left or right on certain
screens, to select the icons on the Menu screen, and to increase or decrease the value of a setting.
Press this button to go back to the previous screen.
Press and hold this button to return to the Home screen.
Press this button to access the Graph screen.
Press and hold this button to put the pump in Sleep mode.
Sleep mode
The pump enters Sleep mode after two minutes to conserve battery power. Sleep mode does not affect insulin delivery. Press any button to wake up the pump. Press and hold
for two seconds to manually enter Sleep mode.
Pump basics
Pump basics 27
Page 44
Pump delivery system
The following diagram shows the parts of the pump delivery system, including the infusion set*, reservoir, and pump.
tubing
adhesive
infusion set
cannula
reservoir connector
insertion
site section
pump
reservoir
compartment
reservoir
*Quick-set infusion set shown in illustration.
Infusion set
The infusion set consists of the following components:
The tubing carries insulin from the reservoir into the body.
28
Chapter 2
Page 45
The tubing connector attaches to the reservoir.
The insertion piece attaches to the body.
The cannula is a small, flexible tube inserted into the body. Some infusion sets use a small needle instead of a cannula.
Adhesive holds the infusion set in place.
Change the infusion set according to the user guide provided with the infusion set.
Reservoir
The reservoir stores insulin for delivery and is inserted into the pump reservoir compartment.
Pump
Underneath the reservoir compartment, a piston pushes up on the bottom of the reservoir to move insulin into the tubing, through the cannula, and into the body.
The pump delivers small doses of insulin, as low as 0.025 units. The piston inside the pump must be rewound each time a newly filled reservoir is inserted into the reservoir compartment.
Inserting the battery
The pump requires one new AA (1.5 V) battery. For best results, use a new AA lithium (FR6) battery. The pump also accepts an AA alkaline (LR6) or a fully charged AA NiMH (HR6) nickel-metal hydride rechargeable battery.
CAUTION: Do not use a carbon zinc battery in the pump. Carbon zinc batteries are not compatible with the pump and can cause the pump to report inaccurate battery levels.
Note: Do not use cold batteries because the battery life may incorrectly appear low. Allow cold batteries to reach room temperature before they are inserted into the pump.
The battery cap is located in the pump box with the accessories.
Pump basics
Pump basics 29
Page 46
battery cap
battery
pump clip
To insert the battery:
1. Insert a new or fully charged AA battery. Be sure to insert the negative end
(−) first.
2. Place the battery cap onto the pump. Use the bottom edge of the pump clip
or a coin to tighten the cap.
CAUTION: Do not overtighten or undertighten the battery cap. A battery cap that is too tight can cause damage to the pump case. A battery cap that is too loose can prevent detection of the new battery. Turn the battery cap clockwise until the cap slot is aligned horizontally with the pump case, as shown in the following example.
30 Chapter 2
Page 47
The first time a battery is inserted into the pump, the Startup Wizard begins. Any other time a battery is inserted into the pump, the Home screen appears and the pump resumes basal delivery.
Startup settings
The Startup Wizard appears after a battery is inserted for the first time. Use the Startup Wizard to set the language, time format, current time and date, and to rewind the pump. To re-enter these settings later, see Pump issues, on page 215.
To use the Startup Wizard:
1. On the Select Language screen, select a language, and then press .
The Select Time Format screen appears.
2. Select a time format, and then press
.
3. Enter the current time, and then select Next.
Pump basics
Pump basics 31
Page 48
The Enter Date screen appears.
4. Enter the current date, and then select Next.
A "Rewinding" message appears. The piston returns to its start position in the reservoir compartment. This may take several seconds.
When rewinding is complete, a message appears to confirm the startup is complete.
5. Select OK to go to the Home screen.
Home screen in Manual mode
The Home screen appears after the battery is changed, when the pump wakes from Sleep mode, and when another screen is not actively being used.
32
Chapter 2
Page 49
Note: This example shows the Home screen in Manual mode when the Sensor feature is turned off. For information about the Home screen when the Sensor feature is turned on, see Home screen with CGM in Manual mode, on page 100. For information about the Home screen with the SmartGuard feature, see Home screen with the SmartGuard feature, on page 134.
Active Insulin
09:00
1.0 U
mmol/L
BG
BG reading
active insulin
current time
status icons
The following items appear on the Home screen:
Item Description
Status icons The status icons show a quick status of the pump system. For more
information, see Status icons, on page 34.
Current time For details on setting the time, see Time and date, on page 149.
BG readings The current blood glucose (BG) reading is shown. BG is either entered
manually or received from a paired Accu-Chek Guide Link meter.
Active insulin Active insulin is the amount of bolus insulin that continues to lower BG
levels. For more details on active insulin, see Bolus Wizard settings, on page 60.
Shortcuts from the Home screen
The following table describes shortcuts that can be used to quickly access certain pump functions. These shortcuts only work on the Home screen.
Shortcut Description
Press this button to access the Status screen.
When the Easy bolus feature is turned on, press and hold this button to start the Easy bolus. The MiniMed 780G pump must be in Manual mode and Sleep mode first.
Pump basics
Pump basics 33
Page 50
Shortcut Description
Press this button to access the Time in Range screen when the sensor is turned on.
Press this button to access the Bolus screen. The Bolus screen that appears varies depending on the bolus feature that is currently active.
Status icons
The status icons on the Home screen provide the current status of the pump system. For information on viewing detailed status screens, see Status screen, on page 41.
Icon name Description
Battery The color and fill level of the icon indicate the charge level of the
pump battery. As the battery is used, the icon changes from solid green in the following order:
The battery is full.
The battery is low.
The battery can be used for less than 30 minutes and needs to
be replaced.
34 Chapter 2
Page 51
Icon name Description
Reservoir The reservoir icon shows the fill status of the MiniMed reservoir
MMT-332A, 3.0 mL (300-unit).
Approximately 85%–100% of the insulin remains in the reservoir.
Approximately 71%–84% of the insulin remains in the reservoir.
Approximately 57%–70% of the insulin remains in the reservoir.
Approximately 43%–56% of the insulin remains in the reservoir.
Approximately 29%–42% of insulin remains in the reservoir.
Approximately 15%–28% of the insulin remains in the reservoir.
Approximately 1%–14% of the insulin remains in the reservoir.
The amount of insulin remaining in the reservoir is unknown.
Note: The icon fill level is based on the 3.0 mL reservoir. When a full 1.8 mL reservoir is used, the yellow icon for approximately 43%-56% is shown.
Connection The connection icon shows the following information:
The sensor feature is on and communicating.
The sensor feature is on, but the transmitter is not
communicating with the pump.
Temporary network connection
The temporary network connection icon shows when the pump
is temporarily connected to a remote upload device.
Pump basics
Pump basics 35
Page 52
Icon name Description
Calibration The calibration icon shows the amount of time remaining until the
next sensor calibration is needed. These icons only appear when the Sensor feature is on.
• The color and the circle around the icon indicate the status.
• When the sensor is recently calibrated, the icon has a solid green circle around it. As the time for the next sensor calibration approaches, the green circle around the icon becomes smaller and the color of the icon changes.
• When the icon turns red, a sensor calibration is required.
• If the time until the next sensor calibration is unavailable, the icon has a solid blue circle around a question mark.
• When the sensor is not ready for a calibration, the circle shows three dots. This occurs when a new sensor is connected or within 15 minutes of a Calibration not accepted alert.
Trend arrows The trend arrows indicate the rate at which the most recent SG
readings are rising or falling. Glucose readings may trend up or down during certain activities, such as eating, giving a bolus, or when exercising. These icons appear only when the sensor feature is turned on.
or : SG has been rising or falling at a rate of 1.11-2.22
mmol/L over the last 20 minutes, or 0.06-0.11 mmol/L per minute.
or : SG has been rising or falling at a rate of 2.22-3.33
mmol/L over the last 20 minutes, or 0.11-0.17 mmol/L per minute.
or : SG has been rising or falling at a rate of more
than 3.3 mmol/L over the last 20 minutes, or more than 0.17 mmol/L per minute.
36 Chapter 2
Page 53
Icon name Description
Sensor life The number on the sensor life icon indicates the number of days that
remain in the life of the sensor. The icon appears on the Status screen and only when the sensor feature is turned on. After a new sensor is inserted, the icon is solid green. When one day remains in the life of the sensor, the icon turns red.
If the number of days that remain in the life of sensor is unavailable, the sensor life icon appears with three dots.
Block mode
The Block mode icon
shows that the pump is locked. For more
information about Block mode, see Block mode, on page 150.
Suspend
The suspend by sensor icon
indicates when insulin delivery is
suspended due to a Suspend before low or Suspend on low event. For more information, see The Suspend before low feature, on page 103 and The Suspend on low feature, on page 106.
The suspend unavailable icon
indicates that the Suspend before
low or Suspend on low feature is unavailable. For more information, see When the Suspend before low or Suspend on low features are unavailable, on page 107.
Alert silence
The Alert silence icon
indicates that the Alert Silence feature is
turned on and some alerts will not make a sound or vibration. Sensor alerts can be silenced for a specific duration using the Alert silence feature. For more information, see Silencing sensor alerts, on page 122.
Note: Status icons provide limited information. For example, the reservoir icon may indicate the reservoir is low on insulin. The Status screen shows more detailed information about how many units are left. For more information about the status screens, see Status screen, on page 41.
Pump basics
Pump basics 37
Page 54
Menu screen
Use the menu to go to screens that show various features and functions of the system. Press
from the Home screen to go to the menu. The selected menu
option appears in color. All other menu options appear in black and gray.
Use the menu to go to the following screens:
Menu selection Menu
icon
Description
Insulin
Deliver a bolus, set up and deliver basal insulin, suspend insulin delivery, and stop bolus during bolus delivery.
History & Graph
View history, sensor glucose review, graph, and time in range.
SmartGuard
Set up the SmartGuard feature.
Sound & Vibration
Set sound, vibrate, and volume options for notifications.
Reservoir & Set
Set up a new reservoir and infusion set, and fill a cannula.
Blood Glucose
Enter a BG value.
Status
View the status of the pump and other system features.
Paired Devices
Pair devices or CareLink software.
38 Chapter 2
Page 55
Menu selection Menu
icon
Description
Settings
Set up device settings, delivery settings, and alert settings.
Menu map
The following diagram provides a map to the screens and features that are available from the Main menu.
Pump basics
Pump basics 39
Page 56
Insulin
Bolus
• Bolus Wizard
• Manual Bolus
• Delivery Settings
Basal
• Temp Basal
• Basal Patterns
• Basal Pattern Setup Suspend All Delivery/Resume Basal Delivery Settings
SmartGuard
SmartGuard Checklist Temp Target SmartGuard Settings
• Target
• Auto Correction
SmartGuard on and off
Settings
Alert Settings
• High Alert
• Low Alert
• Snooze High & Low
• Reminders
Delivery Settings
• Bolus Wizard Setup
• Basal Pattern Setup
• Max Basal/Bolus
• Dual/Square Wave
• Bolus Increment
• Bolus Speed
• Preset Bolus Setup
• Preset Temp Setup
Device Settings
• Sensor on and off
• Time & Date
• Device Info
• Display
• Block Mode
• Self Test
• Review Settings
• Manage Settings
• Easy Bolus
• Auto Suspend
• Language
Sound & Vibration
Silence Sensor Alerts Volume Sound Vibration Alert Settings
Blood Glucose
BG
History & Graph
History
• Summary
• Daily History
• Alarm History Sensor Glucose Review
• High Limit
• Low Limit
• Days to Average Graph Time in Range
Status
Suspend All Delivery SmartGuard Checklist Pump Sensor
Paired Devices
Pair New Device Pair CareLink Mobile Meter Sensor
Reservoir & Set
New Reservoir Fill Cannula
Key
= shortcut
40 Chapter 2
Page 57
Sound & Vibration screen
The sound and vibration options are set on the Sound & Vibration screen. Sensor alerts can also be temporarily silenced. For information about silencing alerts, see Silencing sensor alerts, on page 122. A status icon on the Home screen indicates when alerts are silenced. For more information, see Status icons, on page 34.
To adjust the sound and vibration settings:
1.
From the Home screen, press , and then select .
2. Adjust the volume:
a. Select Volume.
b. Press .
c. Press , , , or , and then press .
3. Select Sound, and then press to turn the sound on or off.
4. Select Vibration, and then press to turn the vibration on or off.
Status screen
The Status screen provides access to information about the pump and information about the sensor, if applicable. The Status screen also provides the option to suspend all insulin delivery or resume basal insulin delivery.
Use the Status screen to access the following screens or options:
Screen or option Description
Stop Bolus This option appears when a bolus delivery is in progress. Select
Stop Bolus to stop the active bolus.
Suspend All Delivery or Resume Basal
This option indicates whether insulin delivery is currently suspended. Select Suspend All Delivery to suspend insulin delivery. Select Resume Basal to resume basal insulin delivery. For more information see Suspending all insulin delivery and resuming basal insulin delivery, on page 54.
SmartGuard Checklist screen
This screen shows a list of the required conditions before the pump can use the SmartGuard feature. For more information, see SmartGuard Checklist, on page 132.
Pump basics
Pump basics 41
Page 58
Screen or option Description
Pump screen This screen shows a detailed view of the pump status, the
reservoir status, battery status, pump serial number, pump name, model number, and other pump details.
Sensor screen This screen appears when the sensor feature is enabled. The
Sensor screen indicates any silenced alerts, status of calibration, transmitter battery, and shows the serial number and version number of the transmitter.
To view the status screens:
1. From the Home screen, press , and then select .
2. Press or to select a status screen, and then press .
Viewing basal delivery information
To view the current basal rate:
1.
From the Home screen, press , and then select .
2. Select Basal.
The current basal rate appears at the top of the screen.
To view basal patterns:
1.
From the Home screen, press , and then select .
2. Select Basal Patterns.
The Basal Patterns screen shows a list of configured basal patterns and the 24-hour insulin total for each basal pattern. A check mark appears next to the active basal pattern.
42
Chapter 2
Page 59
3. To view details for a basal pattern, select the basal pattern.
For more information about basal patterns, see Basal patterns, on page 51.
Settings covering a 24-hour period
Some pump functions allow settings to change over a 24-hour period.
The procedure in this section shows how to set multiple values over a 24-hour period. These steps apply to the following tasks:
Setting up basal patterns.
See Adding a new basal pattern, on page 51.
Setting up high SG settings.
See Setting up the High SG settings, on page 110.
Setting up low SG settings.
See Setting up the low SG settings, on page 113.
Setting up carb ratios, insulin sensitivities, and BG targets in the Bolus Wizard feature.
See Setting up the Bolus Wizard feature, on page 61.
For example, in the screenshot below, a basal pattern is set to allow the rate of basal insulin to change depending on the time of day.
Note: The following procedure uses the screens from editing a basal pattern as an example.
To set up values over a 24-hour period:
Pump basics
Pump basics 43
Page 60
1. On the appropriate settings screen, select the End time and enter the end
time for the first time period. In this example, the first desired time period is 8 hours. The start time always begins at 00:00. To set an 8-hour period, the end time of 08:00 is entered.
2. Enter the unit value for the first time period.
3. Press .
The start time for the next time period appears.
4. Enter the end time for the next time period.
44
Chapter 2
Page 61
5. Enter the unit value for the next time period.
6. Press .
The start time for the next time period appears.
7. Repeat steps 3-5 for every desired time period until the end time of 24:00 is
reached. This completes the 24-hour duration.
Pump basics
Pump basics 45
Page 62
46 Chapter 2
Page 63
Setting up insulin delivery
Page 64
Page 65
Setting up insulin delivery
This chapter explains how to use different types of insulin delivery.
Setting up basal insulin
Basal insulin is the "background" insulin that the body needs throughout the day and night to maintain target BG readings when food is not eaten. Basal insulin accounts for approximately one half of daily insulin requirements. The MiniMed 780G insulin pump simulates a pancreas by delivering insulin continuously over 24 hours.
Basal insulin is delivered according to a basal pattern.
Basal rate
Basal rate is the specific amount of basal insulin that the pump continuously delivers each hour. While some people use one basal rate all day, others require different rates at different times of the day.
Basal rates are set in one or more basal patterns. Each basal pattern covers 24 hours. For specific information about basal patterns, see Basal patterns, on page 51.
Max basal rate
The Max basal rate is the maximum amount of basal insulin that the pump can deliver per hour. Set the Max basal rate as indicated by a healthcare professional. It is not possible to set a basal rate, a temp basal rate, or a preset temp basal rate that would exceed the Max basal rate limit. After the basal patterns or preset temp basal rates are set, the Max basal rate cannot be lower than any of the existing basal rates. The Max basal rate can be set from 0 to 35 units per hour.
Setting up insulin delivery
49
Setting up insulin delivery
Page 66
To set the Max basal rate:
1.
From the Home screen, press , and then select .
2. Select Delivery Settings > Max Basal/Bolus.
The Max Basal/Bolus screen appears.
3. Select Max Basal.
4. To continue to the Max Basal Rate screen, select Continue.
5. Select Max Basal, and then set the maximum number of basal insulin units
per hour.
6. Select Save.
50
Chapter 3
Page 67
Basal patterns
The basal pattern determines the amount of basal insulin delivered throughout the day and night. A basal pattern is made up of one to 48 basal rates that are set to cover a full 24-hour period. Because basal insulin needs can vary, up to eight basal patterns can be set.
The following example represents one basal pattern with three basal rates set for three different time periods.
00:00 6:00
12:00 18:00
24:00
00:00 – 8:00
8:00 – 18:00
18:00 – 24:00
24 Hours
1.000
0.500
U/hr
0.650 U/hr
0.900 U/hr 0.900 U/hr
A healthcare professional determines the best basal rate.
Adding a new basal pattern
This procedure shows how to add a new basal pattern for the first time. To add an additional basal pattern, see Adding an additional basal pattern, on page 186.
To add a new basal pattern:
1.
From the Home screen, press , and then select .
2. Select Delivery Settings > Basal Pattern Setup.
3. Select Basal 1.
4. Select Options, and then select Edit.
Setting up insulin delivery
Setting up insulin delivery 51
Page 68
5. For one basal rate, the End time does not need to change. Press on the
24:00.
Note: For instructions on setting up multiple basal rates over a 24­hour period, see Settings covering a 24-hour period, on page 43.
6. Select Done.
Review the basal pattern. Press to return to the previous screen to make changes.
Note: If is pressed and Done is not selected, the changes are not saved.
7. Select Save.
Temp basal rates
The temp basal feature helps set and start a temporary basal rate that can be used immediately to manage BG for short-term activities or conditions.
Preset temp basal rates can be set for recurring short term situations. For more information on Preset temp basal rates, see Preset temp basal rates, on page 183. The duration of the temp basal rate can range from 30 minutes to 24 hours. After
52
Chapter 3
Page 69
the temp basal rate delivery is completed or canceled, the programmed basal pattern resumes. The temp basal rates and preset temp basal rates can be defined using either a percentage of the current basal pattern or by setting a specific rate, as described in the following table:
Temp basal rate type Description
Percent Percent delivers a percentage of the basal rates
programmed in the active basal pattern for the duration of the temp basal rate. The temp basal amount is rounded down to the next 0.025 units if the basal rate is set at less than 1 unit per hour, or to the next 0.05 units if the basal rate is set at more than 1 unit per hour.
Temp basal rates can be set to deliver from 0% to 200% of the scheduled basal rate. The percentage used is based on the largest basal rate scheduled during the temp basal rate duration and is limited by the Max basal rate.
Rate Rate delivers a fixed basal insulin rate in units per hour
for the duration of the temp basal rate. The amount set is limited by the Max basal rate.
Starting a temp basal rate
When a temp basal rate starts, basal delivery changes to the temp basal rate for the set duration. When the duration completes, the basal insulin automatically returns to the active basal pattern.
To start a temp basal rate:
1.
From the Home screen, press , and then select .
2. Select Basal > Temp Basal.
3. Set the Duration.
Setting up insulin delivery
Setting up insulin delivery 53
Page 70
4. Select Next.
5. Select Type to select Rate or Percent.
6. Depending on the type selected, do one of the following:
Enter a percentage.
•Enter a basal rate.
Select Review to review the temp basal setting.
7. Select Begin to start the temp basal rate.
The Temp Basal banner appears on the Home screen during delivery.
Suspending all insulin delivery and resuming basal insulin delivery
Use this feature to suspend all active basal and bolus insulin deliveries. A reminder that insulin is not being delivered occurs every 15 minutes. The pump will beep, vibrate, or both depending on the sound and vibration settings.
54
Chapter 3
Page 71
Note: The first reminder occurs 15 minutes after the pump display times out. If a button is pressed to wake up the pump, the reminder does not occur until 15 minutes after the pump display times out again. To adjust the timeout setting, see Display options, on page 149.
To continue basal insulin delivery, use the Resume Basal feature. The pump starts the programmed basal pattern but does not start any previously programmed bolus deliveries.
Note: To stop a bolus delivery without stopping the basal delivery, see Stopping a normal bolus delivery, on page 68.
WARNING: If insulin delivery is suspended during a bolus, check the pump daily history to determine the amount of insulin that was delivered, before insulin delivery is resumed. Bolus delivery and fill cannula do not restart when insulin delivery is resumed. If needed, program a new bolus or fill the cannula. Failure to resume basal insulin delivery can result in hyperglycemia and diabetic ketoacidosis.
WARNING: Do not rely solely on the sound or vibration notifications when using the sound or vibrate options. These notifications may not occur as expected if the speaker or vibrator in the pump malfunctions. A missed notification may result in the delivery of too much or too little insulin. This is most common when using the Easy bolus feature or when the pump is in manual suspend. Contact the local Medtronic support representative with any concerns.
To suspend all insulin delivery:
1.
From the Home screen, press
, and then select .
2. Select Suspend All Delivery.
A confirmation message appears.
Setting up insulin delivery
Setting up insulin delivery 55
Page 72
3. Select Yes to suspend all insulin delivery.
The pump functions are limited until insulin delivery is resumed.
The Delivery Suspended banner appears on the Home screen while insulin is suspended.
To resume basal insulin delivery:
1. While insulin delivery is suspended, from the Home screen press , and then
select
.
2. Select Resume Basal.
A confirmation message appears.
3. To resume basal insulin delivery, select Yes.
If a temp basal was active when the pump was suspended, it resumes, provided the time is still within the duration set.
Note: If a bolus delivery that was in progress before delivery was suspended is needed, check the Daily History screen for the actual bolus units delivered and the intended bolus amount. Then set up a new bolus amount as needed. For details about using the Daily History screen, see Daily History screen, on page 163.
56
Chapter 3
Page 73
Entering a BG value
The BG screen allows manual entry of a BG reading. Previously entered manual or meter BG readings do not appear on the BG screen. A BG reading received from a linked meter appears in a separate BG meter screen that requires confirmation.
To manually enter BG readings:
1.
From the Home screen, press , and then select .
2. Enter a BG value.
3. Select Save.
If entering a BG value between 2.2 mmol/L and 22.2 mmol/L, the sensor calibrates using the entered value.
Note: If entering a BG value outside that range, the sensor does not calibrate.
To confirm a BG reading from a BG meter:
1. When the BG Meter screen with the message Confirm BG? shows, select Yes
to confirm the BG meter value.
The BG received message shows.
Setting up bolus delivery
A bolus is given for two reasons: to cover food that contains carbohydrates or to correct glucose levels that are above the target range.
About bolus deliveries
A bolus can be delivered using either the Manual bolus feature or the Bolus Wizard feature. Multiple types of bolus deliveries are also available, including normal bolus, Square Wave bolus, and Dual Wave bolus. The bolus type depends on individual insulin needs. Discuss these options with a healthcare professional to determine what is best. For details about the different types of bolus deliveries available, see Bolus types, on page 191.
Setting up insulin delivery
Setting up insulin delivery 57
Page 74
Bolus delivery options
The following table describes how to deliver a bolus using the Bolus Wizard feature or Manual bolus feature.
Delivery method Description
Bolus Wizard feature Enter the BG meter value or the amount of carbs
expected from a meal, or both. Then the Bolus Wizard feature calculates an estimated bolus amount based on the individual settings.
The Bolus Wizard feature is only available in Manual mode.
For details about using the Bolus Wizard feature, see Bolus Wizard feature, on page 59.
Manual bolus feature Calculate and manually enter the bolus amount.
For details about using the Manual bolus feature, see
Delivering a normal bolus using the Manual bolus feature, on page 67.
Max bolus
The Max bolus setting limits the amount of insulin that can be delivered in a single bolus. The pump prevents single bolus insulin deliveries that exceed the Max bolus amount. The Max bolus can be set from 0 to 25 units. Set the Max bolus as indicated by a healthcare professional.
If the Max bolus is set up after the preset bolus deliveries are set, the Max bolus cannot be set lower than any of the existing preset bolus amounts.
The Max bolus setting applies to boluses delivered in Manual mode and delivered with the SmartGuard feature.
To set the Max bolus:
1.
From the Home screen, press , and then select .
2. Select Delivery Settings > Max Basal/Bolus.
The Max Basal/Bolus screen appears.
58
Chapter 3
Page 75
3. Select Max Bolus.
4. To continue to the Max Bolus screen, select Continue.
5. Select Max Bolus, and then set the maximum number of insulin units the
pump can deliver in one bolus.
6. Select Save.
Bolus Wizard feature
The Bolus Wizard feature uses Bolus Wizard settings to calculate an estimated bolus amount based on the BG readings and carbs that are entered.
After the Bolus Wizard feature is set up, use a normal bolus to deliver a food bolus, a correction bolus, or a food plus correction bolus. For more information, see Delivering a normal bolus with the Bolus Wizard feature, on page 65.
Setting up insulin delivery
Setting up insulin delivery 59
Page 76
The Bolus Wizard feature can also be used to deliver a Dual Wave bolus or a Square Wave bolus. For more information, see Bolus types, on page 191.
Bolus Wizard settings
To use the Bolus Wizard feature, consult a healthcare professional to determine the personal settings that should be used. The carb ratio, insulin sensitivity factor, BG target, and the active insulin time are needed to complete the setup. Always consult a healthcare professional before changes are made to the Bolus Wizard settings. The setup procedure begins on page 61.
Setting Description
Carb Ratio The carb ratio setting is used for food bolus calculations.
The number of carb grams that are covered by 1 unit of insulin.
Insulin Sensitivity Factor
The insulin sensitivity factor setting is used to calculate correction bolus amounts.
The insulin sensitivity factor is the amount that BG is reduced by 1 unit of insulin.
BG Target The Bolus Wizard feature calculates the estimated bolus
based on the BG target range. The high and low values set are the values to which the BG is corrected. To use a single target value rather than a range, set the same value for the high and low value of the BG target.
If the BG reading is above the high target value, a correction dose is calculated. If the BG reading is below the low target value, a negative correction is calculated and subtracted from the food bolus.
60 Chapter 3
Page 77
Setting Description
Active Insulin Time Active insulin is the bolus insulin that has been delivered
by the pump and is still working to lower glucose levels. The pump uses the active insulin time setting to determine if any active insulin is still in the body from prior boluses. This may help prevent hypoglycemia caused by overcorrection of high BG.
The current active insulin amount appears on the Home screen and includes only the bolus insulin received.
Consult a healthcare professional to get the active insulin time that best represents the insulin type used and the physiological insulin absorption rate.
Setting up the Bolus Wizard feature
To use the Bolus Wizard feature to calculate a bolus, first turn on the Bolus Wizard feature and enter the Bolus Wizard settings.
To set up the Bolus Wizard feature:
1.
From the Home screen, press , and then select .
2. Select Delivery Settings > Bolus Wizard Setup.
The Bolus Wizard Setup screen appears.
3. Select Bolus Wizard to turn on the feature.
If this is the first time the Bolus Wizard feature has been turned on, the following screen appears.
Setting up insulin delivery
Setting up insulin delivery 61
Page 78
4. Confirm the values needed are ready to be entered, then select Next.
The Carb Ratio 1/4 screen appears.
5. Select Next.
The Edit Carb Ratio 1/4 screen appears.
6. To enter one carb ratio, enter the g/U, and then press .
Note: For instructions on setting up more than one carb ratio over a 24-hour period, see Settings covering a 24-hour period, on page 43.
7. Select Next.
Note: If the values are outside of the value range, a message asks to confirm the settings.
62
Chapter 3
Page 79
The Sensitivity 2/4 screen appears.
8. Select Next.
The Edit Sensitivity 2/4 screen appears.
9. For one sensitivity factor, enter the mmol/L per U, and then press .
Note: For instructions on setting up more than one sensitivity factor over a 24-hour period, see Settings covering a 24-hour period, on page 43.
10. Select Next.
The BG Target 3/4 screen appears.
11. Select Next.
Setting up insulin delivery
Setting up insulin delivery 63
Page 80
The Edit BG Target 3/4 screen appears.
12. For one BG target range, enter the Lo and Hi target, and then press .
Note: For instructions on setting up more than one BG target range over a 24-hour period, see Settings covering a 24-hour period, on page 43.
13. Select Next.
The Active Insulin Time 4/4 screen appears.
14. Select Next.
The Active Insulin Time 4/4 screen appears.
15. Enter the Duration of the active insulin time, and then press .
16. Select Save.
64
Chapter 3
Page 81
The Bolus Wizard feature setup is now complete.
Turning the Bolus Wizard feature off
The Bolus Wizard feature can be turned off at any time. The Bolus Wizard settings remain in the pump. When the Bolus Wizard feature is turned off, the Bolus Wizard menu selection does not appear on the Bolus screen, and the insulin sensitivity factor or BG target settings can not be edited from the Bolus Wizard Setup screen.
To turn the Bolus Wizard feature off:
1.
From the Home screen, press , and then select .
2. Select Delivery Settings > Bolus Wizard Setup.
3. Select Bolus Wizard to turn the feature off.
WARNING: Do not use the Bolus Wizard feature to calculate a bolus for a period of time after giving a manual injection of insulin by syringe or pen. Manual injections are not accounted for in the active insulin amount. Therefore, the Bolus Wizard feature may ask to deliver more insulin than needed. Too much insulin can cause hypoglycemia. Consult a healthcare professional for how long to wait after a manual injection of insulin before the active insulin calculation of the Bolus Wizard feature can be relied on.
Normal bolus
A normal bolus provides a single immediate dose of insulin. Use a normal bolus to cover food intake or to correct a high BG meter reading.
Note: The pump can deliver a normal bolus while a Square Wave bolus or the Square portion of a Dual Wave bolus is being delivered.
Delivering a normal bolus with the Bolus Wizard feature
The Bolus Wizard screen shows the most recent glucose reading, if available. The table indicates the different ways that the Bolus Wizard screen shows the glucose reading.
Setting up insulin delivery
Setting up insulin delivery 65
Page 82
Bolus Wizard screen Glucose reading information
The icon indicates that a recent BG meter reading
is used by the Bolus Wizard feature to calculate a correction bolus.
The BG appears as dashes when no BG is available for the Bolus Wizard feature to calculate a correction bolus.
To deliver a normal bolus using the Bolus Wizard feature:
1. For a correction bolus or a food bolus with a correction, use the BG meter to
check BG.
Note: For more information on how to manually enter the BG meter reading, see Entering a BG value, on page 57.
2.
From the Home screen, press , and then select .
3. Select Bolus > Bolus Wizard.
The Bolus Wizard screen appears.
66 Chapter 3
Page 83
4. For a food bolus, select Carbs to enter the carb count of the meal. For a
correction bolus where no food was eaten, leave the carbs value at 0.
5. The calculated bolus appears in the Bolus field.
If a change to the bolus amount is needed, select Bolus and modify the bolus amount.
6. Select Deliver Bolus to start the bolus.
The pump beeps or vibrates and a message appears when the bolus starts. The Home screen shows the bolus amount as it is being delivered. The pump beeps or vibrates when bolus delivery is complete.
Delivering a normal bolus using the Manual bolus feature
Note: To see a normal bolus, the Dual Wave bolus and Square Wave
bolus feature must be turned on, see Bolus types, on page 191.
The following procedure describes how to deliver a normal bolus using the Manual bolus feature.
To deliver a normal bolus using the Manual bolus feature:
1.
From the Home screen, press
, and then select .
2. Do one of the following:
Setting up insulin delivery
Setting up insulin delivery 67
Page 84
Select Bolus if the Bolus Wizard feature is turned off.
Select Bolus > Manual Bolus if the Bolus Wizard feature is turned on.
The Manual Bolus screen appears.
3. Select Bolus to set the bolus delivery amount in units.
4. Select Deliver Bolus to start the bolus.
Stopping a normal bolus delivery
These procedures describe how to stop a normal bolus.
Note: This procedure describes how to stop a bolus that is in progress. It does not stop basal insulin delivery. To stop all insulin delivery, use the Suspend All Delivery feature (press
, select , and select Suspend All
Delivery).
To stop a normal bolus delivery:
1.
While the pump delivers a normal bolus, press
and then select .
The Insulin menu appears.
2. Select Stop Bolus.
A message appears confirming if bolus delivery should be stopped.
68
Chapter 3
Page 85
3. Select Yes to confirm.
The Bolus Stopped screen appears and shows the amount of bolus delivered, and the original bolus amount set up.
4. Select Done.
Note: The delivered amount can be viewed in the insulin delivery history screen after the procedure is closed. For more information, see Daily History screen, on page 163.
Setting up insulin delivery
Setting up insulin delivery 69
Page 86
70 Chapter 3
Page 87
Reservoir and infusion set
Page 88
Page 89
Reservoir and infusion set
This chapter provides information about setting up the reservoir and infusion set.
Setting up the reservoir and infusion set
Confirm that the time and date on the pump are correct before insulin is used with the pump for the first time. For information about how to change the time and date on the pump, see Time and date, on page 149. Consult a healthcare professional to determine the appropriate pump settings before insulin is used with the pump.
The following items are needed:
MiniMed 780G insulin pump
•vial of rapid-acting U-100 insulin
MiniMed reservoir
MiniMed-compatible infusion set and its user guide
WARNING: Do not use the pump to deliver insulin for the first time until the active insulin has been cleared. If the pump has been used for training with bolus delivery before insulin is used, the active insulin value may be inaccurate. This may result in inaccurate insulin delivery, and serious injury. For details, see Clearing the active insulin, on page 154.
Reservoir and infusion set
73
Reservoir and infusion set
Page 90
Note: Different infusion sets may have different instructions for insertion
into the body. All the procedures in the sections within this chapter must be followed in order to change the reservoir and infusion set.
Removing the reservoir and rewinding the pump
If this is the first time a reservoir is inserted into the pump, proceed to the pump rewind instructions. For more information about the reservoir see the reservoir user guide.
WARNING: Always confirm that the infusion set is disconnected from the body before rewinding the pump or filling the infusion set tubing. Never insert the reservoir into the pump while the tubing is connected to the body. Doing so may result in an unintentional infusion of insulin, and may cause hypoglycemia.
When the pump rewinds, the piston in the reservoir compartment returns to its starting position and allows a new reservoir to be placed into the pump.
The piston is located in the reservoir compartment of the pump. It engages the reservoir and pushes insulin through the tubing.
piston
reservoir
To remove the reservoir:
1. Wash hands with soap and water.
2. Remove the infusion set by loosening the adhesive and pulling away from
the body.
Note: For instructions on how to remove the infusion set from the body refer to the user guide that came with the infusion set.
74
Chapter 4
Page 91
3. If the optional activity guard is attached to the reservoir compartment on the
pump, remove it now.
4. Turn the tubing connector a half-turn counter-clockwise, and then remove
the reservoir and the connector from the pump.
5. Dispose of the used reservoir and infusion set according to local regulations,
or contact a healthcare professional for disposal information.
To rewind the pump:
1.
From the Home screen, press , and then select .
2. Select New Reservoir.
The New Reservoir screen appears.
If an infusion set is still connected to the body, remove it now. For instructions on how to remove the infusion set from the body, refer to the user guide that came with the infusion set.
If the reservoir is still in the pump, remove it now. For instructions on how to remove the reservoir, see Removing the reservoir and rewinding the pump, on page 74.
3. Select Rewind.
Reservoir and infusion set
Reservoir and infusion set 75
Page 92
A "Rewinding" message appears as the piston in the reservoir compartment returns to its start position.
Another message appears when the pump finishes rewinding, and then the New Reservoir screen appears.
The reservoir may now be filled and connected to the infusion set tubing connector. Follow the steps in the next section to complete these actions before selecting Next on the pump screen.
Filling the reservoir and connecting it to the infusion set tubing
WARNING: Always allow the insulin to reach room temperature before use. Cold insulin may cause air bubbles in the reservoir and tubing, which may result in inaccurate insulin delivery.
The following procedures must be performed in the order presented.
76
Chapter 4
Page 93
To fill the reservoir and connect it to the infusion set tubing:
1. Remove the reservoir from the package and fully extend the plunger.
plunger
reservoir
transfer
guard
2. Swab the top of the insulin vial with alcohol (not shown).
3. Without pushing down on the plunger, firmly press the blue transfer guard
onto the vial.
transfer guard
insulin vial
4. Push and hold the plunger down. This pressurizes the vial. While still holding
down the plunger rod, flip the vial over so the vial is on top. Release the hold on the plunger rod and pull the plunger down to fill the reservoir with insulin.
plunger
insulin vial plunger
Reservoir and infusion set
Reservoir and infusion set 77
Page 94
5. Gently tap the side of the reservoir to make any air bubbles rise to the top of
the reservoir. Push the plunger up to move the air into the vial.
6. If needed, slowly pull the plunger back down to the amount of insulin
needed.
7. To avoid getting liquid on the top of the reservoir, flip the vial over so that it
is upright. Turn the reservoir counter-clockwise, then pull straight up to remove the reservoir from the transfer guard.
hold insulin vial upright
transfer guard
reservoir
78 Chapter 4
Page 95
WARNING: Do not use the reservoir or infusion set if any liquid gets on the top of the reservoir or inside the tubing connector, as shown in the image. Liquid may temporarily block the vents. This may result in the delivery of too little or too much insulin, which may cause hyperglycemia or hypoglycemia. If any liquid gets on the top of the reservoir or inside the tubing connector, start over with a new reservoir and infusion set.
Tubing connector
Top of
reservoir
The reservoir is now ready to be connected to the infusion set tubing.
8. Follow the instructions in the infusion set user guide to access the infusion
set tubing.
9. Place the infusion set tubing connector onto the reservoir. Turn the connector
clockwise, pressing gently against the reservoir until it slides in. Push in and continue turning the connector until the reservoir and the connector lock together with an audible click.
connector
10. If any air bubbles are present, tap the side of the reservoir to force the air
bubbles to the top of the reservoir. Then remove the air bubbles by pushing up on the plunger until insulin is seen in the tubing.
Reservoir and infusion set
Reservoir and infusion set 79
Page 96
11. Without pulling, turn the plunger counter-clockwise to remove it from the
reservoir.
12. Select Next from the New Reservoir screen on the pump display.
Note: The backlight may have turned off. Press any button to turn the screen back on.
The New Reservoir screen appears and indicates that the reservoir is ready to be placed into the pump.
Follow the steps in the next section to complete this action before selecting Next on the pump screen.
Placing the reservoir into the pump and filling the tubing with insulin
Do not place the reservoir into the pump until training has been received.
WARNING: Always rewind the pump before placing a new reservoir. Failing to rewind the pump may result in an unintentional infusion of insulin, which may cause hypoglycemia.
WARNING: Always confirm that the infusion set is disconnected from the body before rewinding the pump or filling the infusion set tubing. Never place the reservoir into the pump while the tubing is connected to the body. Doing so may result in an unintentional infusion of insulin, which may cause hypoglycemia.
80
Chapter 4
Page 97
To place the reservoir into the pump and fill the tubing with insulin:
1. Confirm that the pump has been rewound. For more information, see
Removing the reservoir and rewinding the pump, on page 74.
2. Place the filled reservoir into the reservoir compartment of the pump.
Note: If the pump is being used for the first time, remove the red shipping cap from the reservoir compartment.
3. Turn the tubing connector clockwise until the connector is locked. The tubing
connector should be aligned horizontally with the top of the pump.
4. Select Next from the New Reservoir screen.
The Load Reservoir screen appears.
Reservoir and infusion set
Reservoir and infusion set 81
Page 98
5. Select and hold Load until a checkmark appears on the screen and the pump
beeps or vibrates.
Note: If is pressed after the loading process begins, a Loading incomplete alarm occurs.
6. Select Next.
The Fill Tubing screen appears.
7. Select and hold Fill on the Fill Tubing screen. The pump beeps 6 times.
The pump continues to beep as it fills the tubing, and the amount of insulin used appears on the screen.
82
Chapter 4
Page 99
WARNING: Always check the tubing for air bubbles. Continue to press Fill until no bubbles remain in the tubing. Air bubbles may result in inaccurate insulin delivery.
8. Stop holding Fill when insulin droplets form on the tip of the infusion set
needle.
Note: The location of the infusion set needle may be different depending on the type of infusion set being used.
Note: If the Max Fill reached alarm occurs, it means more than 30 units of insulin were used to fill the tubing. For more information on the Max Fill reached alarm, see Pump alarms, alerts, and messages, on page 233.
9. Select Next on the Fill Tubing screen.
The Fill Cannula? screen appears.
Do not proceed with filling the cannula until after the infusion set has been inserted. Follow the steps in the next section to insert the infusion set into the body before proceeding with the steps on the pump screen.
Note: If an infusion set with a steel cannula is used, the cannula does not need to be filled, and Done may be selected.
Reservoir and infusion set
Reservoir and infusion set 83
Page 100
Inserting the infusion set into the body
Always refer to the infusion set user guide and the serter user guide, if needed, for instructions about how to insert an infusion set into the body.
WARNING: Do not remove the reservoir from the pump while the infusion set is connected to the body. Doing so may result in the delivery of too little or too much insulin, which may cause hyperglycemia or hypoglycemia.
The approved areas for infusion set insertion are shaded in the following example. Avoid the 5.0 cm (2-inch) area around the navel to help ensure a comfortable infusion site and to help with adhesion.
CAUTION: Do not use the same infusion set insertion site for an extended period of time. This may cause the site to become overused. Rotate the infusion set insertion sites regularly.
CAUTION: Always change the infusion set as indicated by the infusion set user guide. Using the same infusion set for an extended period of time may cause infusion set occlusion or site infection.
After the infusion set is inserted into the body follow the steps in the following section to fill the cannula.
84
Chapter 4
Loading...