Medtronic MIDAS REX MR7 Instruction Manual

Page 1
MIDAS REX® MR7
PTSC224
Pneumatic High-Speed System Pneumatyczny system wysokoobrotowy Pneumatický vysokorychlostní systém Nagy sebességĦ pneumatikus rendszer
Instruction Manual Instrukcja obsáugi Návod Használati útmutató
Rx Only
Page 2
Symbols
PTSC224
The following symbols may appear within this manual, on product labeling, or on the product itself:
Attention, see Instructions for Use. Tube Control
STERILE
United States federal law restricts this device to sale by or on the order of a physician.
Reference Number Use with
Lot Number Multi-Use Disposable
Serial Number
Quantity
Instrument Case
Air Pressure Relief
Attachment Instrument Case
Lubricant/Diffuser
Lubricant/Diuser
For single patient use only. Do not re-use, re-process, or re-sterilize this product. Re-use, re-processing or re-sterilization may compromise the structural integrity of the device and/ or create a risk of contamination of the device, which could result in patient injury, illness, or death.
Approximately equal to
Non-Sterile
Sterilized by Gamma Irradiation
Dissecting Tool
Attachment
Control Unit
Refurbished
Dissecting Tool
Attachment
Control Unit
Refurbished
+
Use by date
Accessory
Accessory
Date of manufacture
REGULATOR
Temperature Limitations
Unlock
Lock
|
On
Off USA Only
Finger-Operated Control Manufacturer
Foot-Operated Control Do not dispose to unsorted
Compliant with European Council Directive MDD 93/42/EEC.
Bone Mill
MOTOR
Brush
Adapter
Regulator
Bone Mill
Motor
Brush
Adapter
municipal waste.
Tool Control
EC REP
Authorized Representative in the European Community
Consult Instructions for Use
Page 3
Contents
PTSC224
General Information ...........................................................................................................................1
Indications for Use ..............................................................................................................................1
Contraindications ....................................................................................................................1
Special Notices .........................................................................................................................1
General Safety Precautions .....................................................................................................2
No Latex Policy ........................................................................................................................2
System Components ...........................................................................................................................3
Non-Disposable Components ...............................................................................................3
Disposable Components .........................................................................................................7
Legend® Dissecting Tools ........................................................................................................8
General Guidelines F
or Attachment and Tool Applications .............................................. 9
Setting up the Operating Room ..................................................................................................... 12
Power Source Requirements ............................................................................................... 12
Pneumatic Connections....................................................................................................... 12
Setting up the MR7 System ............................................................................................................. 13
Installing the Oiler Cartridge .............................................................................................. 13
Connecting the Motor ......................................................................................................... 14
Installing an Attachment and Tool ................................................................................................ 15
Straight At
tachments ............................................................................................................ 15
Specialized Attachments ...................................................................................................... 15
Activating the Motor ....................................................................................................................... 16
Disassembling the MR7 System ..................................................................................................... 17
Depressurize the System ...................................................................................................... 17
Disconnect Hoses ................................................................................................................. 17
Discard the Lubricant/Di user Cartridge ......................................................................... 17
Remove the Attachment and Tool ...................................................................................... 17
Cleaning & Sterilizing the MR7 System ........................................................................................ 18
MR7 Motor ............................................................................................................................ 18
Legend At
tachments / Tubes ............................................................................................... 19
i
Page 4
MR7 Pneumatic Control Unit / Regulator Hose /
PTSC224
Triton Adapter / Instrument Case ...................................................................................... 21
Transmissible Spongiform Encephalopathies (TSE) Return Policy............................... 21
Troubl eshooting ............................................................................................................................... 22
Refurbishing or Repairs .................................................................................................................. 25
Preventative Maintenance ............................................................................................................... 26
Limited Warran
ty ............................................................................................................................. 27
Appendix A—Specialized Attachments ........................................................................................ 28
Angled Attachments ............................................................................................................. 28
Angled Double Lock Attachments ..................................................................................... 29
Curved Bur Attachments ..................................................................................................... 30
Variable Exposure Attachments ......................................................................................... 31
Footed Attachments ............................................................................................................. 32
Rotating Footed Attachments ............................................................................................. 33
Contra-Angle Attachment (16-MF) ................................................................................... 34
Metal Cutting Attachments ................................................................................................. 35
Telescoping Attachments ..................................................................................................... 36
Perforator Driver At
tachments ........................................................................................... 37
Jacobs® Chuck Attachments ................................................................................................ 38
Bone Mill Attachment .......................................................................................................... 39
ii
Page 5
General Information
PTSC224
Read and understand this manual before use of the MR7 System.
The Midas Rex® MR7 system is designed for use by medical professionals familiar with powered surgical instrumentation. The surgeon is responsible f
be harmful. It is strongly recommended that the surgeon and dedicated operating room personnel are knowledgeable with the use of this equipment by being trained in Medtronic Midas Rex Hands-On Workshops or by one of the local authorized
representatives.
The MR7 system consists of the following components:
MR7 or MR7 T
MR7 Pneumatic Control Unit with Various Connectors
MR7 Regulator Hose
MR7 Lubricant/Di user Cartridge
MR7 Triton® Adapter (optional)
• Legend® Attachments*
• Legend® Dissecting Tools*
*The MR7 syst
ouch Motor
em uses the same attachments and dissecting tools as the Legend® Pneumatic High-Speed System.
or learning the proper techniques in the use of this system, as inappropriate use may potentially
General Information
Indications for Use
The Medtronic Midas Rex MR7 System is a pneumatically operated surgical instrument system. The pneumatic motors provide power to operate removable rotating surgical cutting tools and their accessories intended for use in neurosurgery, including craniotomy and spinal surgery; as well as Ear, Nose and Throat (ENT), orthopedic and general surgical applications including maxillofacial, craniofacial and sternotom
y surgeries.
Contraindications
None
Special Notices
The words warning, caution and note have special meanings in this manual, and should be carefully reviewed:
WARNING: A warning indicates that the personal safety of the patient or physician may be involved. Disregarding this information could result in injury to the patient or physician.
CAUTION: A caution indicates that there is a risk of damaging equipment.
NOTE: A note is intended to provide additional information, which may be useful
procedure.
, but is not essential to complete the
1
Page 6
General Information
PTSC224
General Safety Precautions
WARNINGS:
• Do not use the Midas Rex MR7 System before proper cleaning and sterilization.
• Do not operate the Midas Rex MR7 System in the presence of Magnetic Resonance Imaging devices.
• Do not use damaged, faulty, or modi ed Midas Rex MR7 S damage prior to each use:
Check the motor’s exhaust hose for cracks or tears.
• Visually inspect attachments and tools. Do not use bent or damaged tools.
• I
nstall attachment and dissecting tool, then brie y run motor.
* Check motor for overheating and leaking lubricant.
* Check attachment for overheating.
* Check dissecting tool for  ail.
• Do not operate the Midas Rex MR7 S
Motors and attachments which fail due to extended use may allow a component to detach and fall from the motor or attachment, and may cause patient injury.
• Heavy side loads and/or long operating periods may cause the device to overheat. If overheating occurs:
• Never place an overheated motor on the patient or draping during surgery.
• Discontinue use and rest the motor by using intermitt
sterile towel.
• If the motor is passed o , the receiver should grasp the motor by the proximal end close to the motor hose.
• T
o avoid injury to the patient or user, do not place the handpiece on the patient or in an unsecured location, when not
in use.
ystem without eye protection.
ystem components. Inspect the Midas Rex MR7 System for
ently, or wrap the motor/attachment interface with a moist
Midas Rex MR7 motors should only be operated when the attachment is in the
If a dissecting tool package is opened, but the tool is not used or contaminated, the tool can be re-sterilized by steam sterilization. Remove tool from all original packaging and place into an approved autoclave package. Steam sterilize as follows:
High Vacuum Steam: 270° F (132° C) for 5 minutes
Gravity Displacement: 270° F (132° C) for 15 minutes
The re-sterilized tool must be used promptly following re-sterilization. If rust or corrosion is encountered after re-sterilization, do not use the re-sterilized tool
.
position.
No Latex Policy
Legend and MR7 products, packaging materials, labels, package inserts, and similar items manufactured by and/or for Medtronic Powered Surgical Solutions (MPSS) do not contain latex.
2
Page 7
System Components
PTSC224
Non-Disposable Components
MR7 Motor
The MR7 motor is a high-speed, high-torque motor used to dissect bone and biomaterials.
Figure 1: Motor Components
System Components
1
5
1. Collet
2. Motor Case
3. Swivel
4. Hose
5. Finger Control Lever
3
2
4
6
6. Safety Slide (MR7 Touch Only)
In addition to the components listed in Figure 1, each MR7 motor has a lubricant/di user housing at the end of the motor hose, as seen in F
igure 2.
Figure 2: Lubricant/Di user Housing
3
Page 8
System Components
PTSC224
WARNING: Use only Medtronic Midas Rex Legend or MR7 devices with an MR7 motor. Use of other devices may cause injury or damage equipment, and will void the manufacturer’s warranty.
Pneumatic Control Unit
The pneumatic control unit (Figure 3) provides variable speed motor control controls through a foot pedal. It also allows the user to switch between  nger and foot control of the motor (if applicable).
Regulator
The regulator (Figure 4) controls the delivery pressure of compressed gas to the pneumatic control unit. The pressure gauges monitor cylinder pressure (right gauge) and delivery pressure (left gauge).
Note: Outlet pressure gauge accurate to +/- 12 psi.
Figure 3: Pneumatic Control Unit
Instrument Case
The instrument case (Figure 5) is used to organize equipment.
Regulator Hose
Connects from the gas source to the pneumatic control unit to deliver compressed gas.
Figure 4: Regulator
Figure 5: Instrument Case
4
Page 9
System Components
PTSC224
N2 DISS to Male Schrader Adapter
The N2 DISS to male Schrader adapter (Figure 6) allows for the regulator hose to be attached to a female Schrader in-house gas connection. A N2 DISS to female Schrader adapter is also available f hose to a male Schrader in-house gas connection.
or connection of the regulator
N2 DISS to Air DISS Adapter
The N2 DISS to air DISS adapter (Figure 7) allows for the regulator hose to be attached to an Air DISS in-house gas connection.
N2 DISS to WF4 Adapter
The N2 DISS to WF4 adapter (Figure 8) allows for the regulator hose to be attached to a Midas Rex safety valve regulator previously used for Midas Rex Classic or Midas
Rex III motors. The in-line oiler must be removed from the safety valve regulator.
Motor Wrench
The motor wrench (Figure 9) is used to align arrows on motor colletats prior to installation of a Legend attachment.
Figure 6: N2 DISS to Male Schrader Adapter
Figure 7: N2 DISS to Air DISS Adapter
Figure 8: N2 DISS to WF4
Adapter
Triton Adapter
The Triton adapter (Figure 10) allows the Triton handpiece to be driven by the MR7 pneumatic control unit. It functions much the same way as the Triton port on the Legend pneumatic control unit, except that it is connec rather than being integrated into the control unit
ted between the control unit and the gas source,
Figure 9: Motor Wrench
Figure 10: Triton Adapter
5
Page 10
System Components
PTSC224
Legend Attachments
Legend motor attachments are available in various designs to facilitate a variety of surgical procedures. Attachments vary in length, diameter, and overall design. They are marked and color-coded to correspond with their associated dissecting tools. A few of the Legend attachments available are listed in the table below.
Attachment Example Other Details/Options
Standard Straight Attachments AS09
Standard Angled Attachments AA14
Straight Variable Exposure Attachments AVS07
Angled Variable Exposure Attachments AVA07
Fixed Footed Attachments AF01
Rotating Footed Attachments AF01R
Telescoping Attachments A
Contra-Angle Attachment AC16
Metal Cutting Attachment ASMC
Perforator Attachment AD01 Available in 800 RPM or 1000 RPM form.
5/32” Jacobs Chuck Attachment AD02
Bone Mill Attachment BM100
T10 (base)
TT12A (tube)
The telescoping attachment requires the use of the AT10
attachment base, as well as a telescoping tube. Tubes are available in straight, cur
ved, or hooded form.
NOTE: Angled and straight attachments with the same length, mark Curved and straight telescoping tubes with the same length, marking, and color band also share the same dissecting tool.
Example: The 14-AM straight and 14-AM angled attachments are 14 cm long, marked 14-AM and have a green color band. All dissecting tools with the pre x 14 (14MH30) may be used in either the 14-AM straight or 14-AM angled attachment.
Be sure to match the color code and nomenclature on the Legend Dissecting Tool packaging with the color band and nomenclature on the Legend Attachment.
ing, and color band share the same dissecting tool.
6
Page 11
Disposable Components
PTSC224
WARNING: Use only Medtronic Midas Rex Legend or MR7 devices with an MR7 motor. Use of other devices may cause injury or damage equipment, and will void the manufacturer’s warranty.
Lubricant/Diffuser Cartridge
The lubricant/di user cartridge (Figure 11) provides lubrication to the motor and  lters oil from exhausted air.
Telescoping T ubes
Telescoping tubes (Figure 12) provide support to the rotating dissecting tool. Telescoping tubes are disposable following multiple uses and should be discarded when heat or excessive vibration is noticed or insertion of tools becomes di cult.
Cleaning Brushes
Cleaning brushes (Figure 13) are used to clean debris from lumen of attachments and telescoping tubes. Sized for an internal bore diameter of 3.2 mm, 2.4 mm or 1.2 mm in Legend Attachments and Telescoping Tubes.
System Components
Figure 11: Lubricant/Di user Cartridge
Figure 12: Telescoping Tube
NOTE: Cleaning brushes will not pass thr
contra-angle, metal cutting, perforator, or Jacobs Chuck attachments, because they are not cannulated.
ough angled,
Figure 13: Cleaning Brushes
3.2mm
2.4mm
1.2mm
7
Page 12
System Components
PTSC224
Legend® Dissecting Tools
Legend dissecting tools are sterile cutting tools, intended for cutting bone and biomaterials.
Dissecting Tool Nomenclature
Part numbers for Legend dissecting tools follow a standard naming convention, which is described in the diagram below. A basic part number consists of the associated attachment length, the tool head shape, and the tool head diameter. Part numbers may also include a variety of pre xes to identify speci c attachment types, as well as a variety of su xes to provide
additional information about the dissec additional “-MN” su x.
ting tool. Tools that use a design taken from the Mednext line are designated by an
Associated
Attachment
Length
Tool
Head Diameter
(x.x millimeters)
Optional
Pre¿ x
Tool
Head Shape
Optional
Suf¿ x
Tool Number Pre¿ xes (not all inclusive)
F... For use with footed attachments
MC For use with metal cutting attachments
T For use with telescoping attachments
Tool Head Shapes (not all inclusive)
AC Acorn MH Match Head
BA Ball OV Oval
CY Cylinder RT Reverse Taper
HM Hole Maker TA Tapered
HS Hole Saw TD Twist Drill
Tool Number Suf¿ xes (note that more than one of the su xes listed may be combined in a single part number)
L Long S Spiral
D Diamond SH Short
X Extra DC Diamond Coarse
F Fine DX Diamond Extra Coarse
C Carbide MN Mednext Tool Design
8
Page 13
System Components
PTSC224
WARNINGS:
• Dissecting tools are for single-use only. Do not attempt to sterilize them. The dissecting tools are packed sterile and are not intended for repeat use. To prevent contamination, use only once.
• Do not use an attachment and dissec
• Do not attempt to remove a tool while the motor is running.
• Do not attempt to remove a tool from an overheated motor or attachment.
• Do
not use the device if the package is opened or damaged.
ting tool combination that results in tool  ail or excessive vibration.
General Guidelines For Attachment and Tool Applications
These are general guidelines for dissecting tool applications and are not an all-inclusive listing.
WARNING: Be sure to match the color code and nomenclature on the Legend Dissecting Tool packaging with the color band and nomenclature on the Legend Attachment. Failure to do so could result in injur sta .
Surgical
Application
Spine 8-B, 9-M, 14-AM,
Commonly
Used
Attachments
Match Head
15-A
Telescoping Match Head
Footed, Straight Tap er ed
Elongated spherical design allows controlled, delicate dissection. For entry hole, nerve decompression, osteophyte removal, sinus dissection, etc.
Ball
Helical cutting  utes dissect bone or cement e ec from a wide variety of approach angles. For debridement, decortication, sinus dissection, etc.
Oval
Helical cutting  utes and curved design blend acorn and ball styles to vary dissection e ciency with appr
decortication, laminotomy, entry hole, nerve decompression, osteophyte removal, etc.
Hole Maker/Saw
Matched sets of Hole Makers and Hole Saws are e cient and e ective for interbody fusion.
Cylinder
E ec
tive bone sculpting and planing. For graft shaping,
debridement, corpectomy, decortication, interbody fusion, fusion takedown, etc.
Acorn
Curved design varies dissection e ciency with varied approach angles. F
debridement, corpectomy, decortication, fusion takedown, etc.
Elongated spherical design allows controlled, delicate dissection. For entr removal, sinus dissection, etc.
Slender design for precise dissection with minimal bone loss. For transection, osteotomy, graft harvesting, bone shaping, entr
y hole, suture hole, midface advancement, etc.
Commonly Used
Dissecting Tools
or entry hole, laminotomy, bone shaping,
y hole, nerve decompression, osteophyte
y to the patient or operating room
Suggested
Motor(s)
MR7, MR7 Touch
tively
oach angle. For
MR7, MR7 Touch
9
Page 14
System Components
PTSC224
Surgical
Application
Neurosurgical– Cranial
General Surgery and Plastic Surgery (Craniofacial/ Maxillofacial/ Sternotomy)
Ear, Nose, and Throat (Otology, Neurootology)
Commonly
Used
Attachments
7-6ST, 8-B, 9-M, 10-9ST, 14-AM, 15-A
Telescoping Match Head
Footed Ta pe re d
7-6ST, 8-B, 9-M, 10-9ST, 14-AM
7-6ST, 10-9ST Ball
Match Head
Elongated spherical design allows controlled, delicate dissection. For entry hole, nerve decompression, osteophyte removal, sinus dissection, etc.
Ball
Helical cutting  ute from a wide variety of approach angles. For debridement, decortication, sinus dissection, etc.
Twist Drill
Helical design with stop produces a hole with a precise depth. Ideal for plating.
A
corn
Curved design varies dissection e ciency with varied approach angles. For entry hole, laminotomy, bone shaping,
debridement, corpectomy, decortication, fusion takedown, etc.
Elongated spherical design allo dissection. For entry hole, nerve decompression, osteophyte removal, sinus dissection, etc.
Slender design for precise dissection with minimal bone loss. For transection, osteotomy, graft harvesting, bone shaping, entr
y hole, suture hole, midface advancement, etc.
Match Head
Elongated spherical design allo dissection. For entry hole, nerve decompression, osteophyte removal, sinus dissection, etc.
Ball
Helical cutting  utes dissect bone or cement e ectively from a wide variety of approach angles. F decortication, sinus dissection, etc.
Tap er ed
Slender design for precise dissection with minimal bone loss. For transection, osteotomy, graft harvesting, bone shaping, entry hole, suture hole, midface advancement, et
Twist Drill
Helical design with stop produces a hole with a precise depth. Ideal for plating.
Helical cutting  utes dissect bone or cement e ec from a wide variety of approach angles. For debridement, decortication, sinus dissection, etc.
Commonly Used Dissecting Tools
s dissect bone or cement e ectively
ws controlled, delicate
ws controlled, delicate
or debridement,
c.
tively
Suggested
Motor(s)
MR7
MR7, MR7 Touch
MR7
10
Page 15
System Components
PTSC224
Surgical
Application
Orthopaedics 8-B, 9-M,
Biometals/ Bioceramics/ Biomaterials
Commonly
Used
Attachments
14-AM, 21-TU, 26-R, Footed, Telescoping
Footed Ta pe re d
MC Metal Cutter
Commonly Used
Dissecting Tools
Ball
Helical cutting  utes dissect bone or cement e ectively from a wide variety of approach angles. For debridement, decortication, sinus dissec
Tap er ed
Slender design for precise dissection with minimal bone loss. For transection, osteotomy, graft harvesting, bone shaping, entry hole, suture hole, midface advancement, etc.
Acorn
Curved design varies dissec approach angles. For entry hole, laminotomy, bone shaping, debridement, corpectomy, decortication, fusion takedown, etc.
Cylinder
Eective bone sculpting and planing. For graft shaping, debridement, corpec fusion takedown, etc.
Slender design for precise dissection with minimal bone loss. For transection, osteotomy, graft harvesting, bone shaping, entr
y hole, suture hole, midface advancement, etc.
Cutting  utes or diamond wheel design remove metals, ceramics and other biomaterials e ec of approach angles. For cutting rods, pins, plates, implants, screws, etc.
tion, etc.
tion e ciency with varied
tomy, decortication, interbody fusion,
tively from a variety
Suggested
Motor(s)
MR7, MR7 Touch
MR7, MR7 Touch
11
Page 16
Setting up the Operating Room
PTSC224
Setting up the Operating Room
Power Source Requirements
Required Operating
(Dynamic) Pressure
80–120 psi 100 psi 12 cubic feet/min. Nitrogen or Dry-Filtered
5.5–8.3 bar 6.9 bar 340 liters/min.
CAUTION: Do not run the motor at an operating pressure above or below the required operating pressure range. Operating pressure belo may damage or reduce the life of the motor.
w 80 psi (5.5 bar) may not provide proper lubrication to the motor. Operating pressure above 120 psi (8.3 bar)
Nominal Operating (Dynamic) Pressure
Approximate Flow Rate
Required
Gas Type
Compressed Air
Pneumatic Connections
Figure 14: Gas Connection Options
1. Regulator Hose (N2 DISS)
2. N2 DISS to Air DISS Adapter Gas Source
3. N2 DISS to Male Schrader Adapter Gas Source
4. Gas Source (N2 DISS)
5. Regulator
6. DISS/WF4 Adapter
7. Regulator
CAUTION: If you are using the Midas Rex Safety Valve Regulator instead of the Legend Regulator, you must replace the in-line oiler with the DISS/WF4 adapter before use.
12
Page 17
Setting up the MR7 System
PTSC224
Setting up the MR7 System
Installing the Oiler Cartridge
WARNING: Do not use the MR7 system with the Midas Rex in-line oiler. The MR7 motor is su ciently lubricated by the lubricant/di user on the motor hose, and will be
over-lubricated if the Midas Rex in-line oiler is used.
1. Set the non-running (static) pressure to 80–120 psi (5.5–8.3 bar) at the gas source. Operating (dynamic) pressure may be adjust
2. Hold the lubricant/di user cartridge perpendicular
to the housing (F circular  tting onto the housing’s circular receptacle
(Figure 16), breaking the foil seal.
3. Rotate the cartridge down until it clicks into plac
4. Verify that the with the notch on the housing (Figure 17).
WARNINGS:
• Failure to properly secure the lubricant/di user cartridge may cause injury to operator and/or
operating room sta .
• Do not attempt to remo cartridge while the system is pressurized.
symbol on the cartridge is lined up
ed later.
igure 15), and press the cartridges
e.
ve the lubricant/di user
Figure 15: Aligning the Lubricant/Di user Cartridge with the
Housing
Figure 16: Pressing the Cartridge onto the Housing
CAUTIONS:
• Do not use an MR7 motor without a lubricant/ di user installed.
• Do not use a lubricant/di user cartridge for more than one hour of drill time.
• Do
• Do not attempt to re ll a used lubricant/di user
• Do not use a lubricant/di user cartridge if it appears
not re-use a lubricant/di user cartridge. It is a
single-use product.
cartridge.
to be damaged
, or if the inner foil seal is punctured.
Figure 17: Correctly I
nstalled Lubricant/Di user Cartridge
13
Page 18
Setting up the MR7 System
PTSC224
Connecting the Motor
Connect the motor hose to the motor port on the top of the pneumatic control unit, by swinging the port cover to the side and pressing the end of the hose into the port (Figure18).
ARNING: Do not pinch, kink, obstruct, cut, tear, or step
W
on the motor/exhaust hose. This may cause the hose to burst, potentially injuring the patient or user.
NOTES:
• If using the MR7 Touch motor, slide the contr on the pneumatic control unit to the position (Figure 19). This will automatically depress and lock the foot pedal. The control will not lock into the position unless the motor hose is connected into the motor port. When the motor hose is removed from the motor port, the foot pedal will return to normal position.
• If using the Triton Power Surgical Instrumentation Syst
em in conjunction with the MR7 motor, use the
optional Triton adapter to connect the Triton hose.
Refer to the documentation accompanying the adapter for connection instructions.
ol slide
Figure 18: Connecting the Motor Hose to the Motor Port
• The motor’s exhaust hose may have an oily  lm on the external sur temperature dierentials following sterilization. Wipe the exhaust hose with a sterile cloth prior to use. If motor continues to have oil on the exhaust hose, return the motor to MPSS f
WARNING: To avoid injury to the patient or user, do not use the pneumatic control unit to operate systems other than the MR7, Legend
Prior to installation of a Legend attachment and dissecting tool, ensure that the arrows on the motor
collet  ats are aligned (Figure 20). If the arrows are not aligned, closest to the motor case until its arrow is aligned with the arrow on the other collet  at.
WARNING: To avoid injury when using the MR7 Touch motor installing the attachment and tool.
use the motor wrench to turn the collet  at
, ensure the safety slide is in the “O” position before
face from pressure and/or
or refurbishing.
, and Triton systems.
Figure 19: Finger/Foot Control Slide
Figure 20: Aligning the collet  ats
14
Page 19
Installing an Attachment and Tool
PTSC224
Installing an Attachment and Tool
WARNING: Dissecting tool  utes are sharp and may perforate surgical gloves. Tools may be grasped with a hemostat to aid in installation and removal.
Straight Attachments
Installation:
1. Slide the attachment over the motor collet, aligning the triangular markers (Figure 21). You will feel and hear the attachment click into place when it is fully seated.
2. Insert the dissecting tool into the attachment with a slight rotational motion (F hear the tool click into place when it is fully seated in the attachment.
3. Turn the attachment to the case (Figure 23). Gently pull on the shaft of the dissecting tool to verify proper installation.
Removal:
Removal is the reverse of installation.
igure 22). You will feel and
position on the motor
Figure 21: Sliding the Attachment over the Motor Collet
Figure 22: Inserting the Dissecting Tool into the Attachment
Specialized Attachments
See Appendix A—Specialized Attachments for installation and removal instructions for other attachments.
Figure 23: Attachment in the Locked Position
15
Page 20
Activating the Motor
PTSC224
Activating the Motor
NOTE: In order to activate the MR7 Touch motor, the safety slide on the  nger control switch must be in the | position, and the control slide on the foot control must be in the
motor is connected to the motor port.
1. Activate the motor by pressing on the foot control pedal (Figure 24), or by pressing on the  nger control lever (MR7 Touch motor only).
2. Adjust operating pressure as needed at the compressed gas source until supply pressure gauge on pneumatic contr unit reads within a range of 80–120 psi (5.5–8.3 bar) as required. Operating pressure (with motor running) will decrease slightly from the non-running (static) pressure setting when the motor is ac
Foot Control Pedal
position. The control slide will not lock in the position unless a
ol
tivated.
Figure 24: Foot Control Pedal
Air Pressure Gauge
Control Slide
Figure 25: MR7 Finger Control Switch
Finger Control Lever
WARNINGS:
• Do not use excessive force to pry or push bone with the attachment or tool during dissection. This could cause the tool to break and cause injur
• Use adequate irrigation during dissection, to prevent thermal necrosis.
MR7 motors should only be operated when the attachment is in the
y to the patient or operating room sta .
position.
NOTE: To decrease pressure, turn down the in-house compressed gas source or loosen the pressure handle on the regulator. Push down on the pressure relief at the pneumatic control unit to exhaust excess pressure in the hoses. T re-adjust pressure as needed.
16
hen
Page 21
Disassembling the MR7 System
PTSC224
Depressurize the System
1. Turn othe compressed gas at the source.
2. Press the pressure relief button on the pneumatic control unit, to release remaining gas.
Disassembling the MR7 System
WARNING: Do not disassemble equipment before the gas is released fr
om control unit.
Disconnect Hoses
Release the motor hose from the control unit, by holding the hose  rmly and pressing the motor port connection button.
Figure 26: Motor Port Connection and Pressure Relief Buttons
Motor Port Connection Button
Pressure Relief Button
Discard the Lubricant/Diffuser Cartridge
Remove the lubricant/di user cartridge from the housing and discard it.
CAUTION: Do not re-use a lubricant/di user cartridge. It is a single-use product.
Remove the Attachment and Tool
1. Follow the applicable removal instructions in the Installing an Attachment and Tool section of this manual to remove the attachment and tool from the motor.
WARNING: Dissecting tool  utes are sharp and may perforate surgical gloves. Tools may be grasped with a hemostat to aid in installation and removal.
2. Discard used dissecting tools in an appropriate container.
17
Page 22
Cleaning & Sterilizing the MR7 System
PTSC224
Cleaning & Sterilizing the MR7 System
MR7 Motor
Warnings and Precautions
Limitations Verify functionality prior to re-use.
INSTRUCTIONS
Point of Use No particular requirements. Follow hospital procedures.
Containment and Transportation
Preparation for Decontamination
Cleaning: Automated
(Do NOT use ultrasonic washer)
Cleaning: ManualWipe all external surfac
Disinfection No particular requirements. Follow hospital procedures.
Packaging For sterilization, place devices in instrument tray. Devices may be unwrapped, or wrapped with up to two layers of 1-ply
Sterilization
(Temperatures are minimum required, times are minimum required)
Do not soak/submerge MR7 devices.
Do not use ultrasound to clean MR7 devices.
Do not use chlorine based or corrosive cleaning agents such as bleach, lye, acetone, sodium hypochlorite/bleach, sodium h
ydroxide, formic acid, or solutions containing glutaraldehyde.
The use of a washer-disinfector for cleaning may cause a pre-mature degradation in performance.
Allow an adequate cooling period after st
Use ONLY nylon cleaning brushes. Non-nylon cleaning brushes leave residue that may prevent the tool from being secured properly in the handpiece.
It is recommended that devices are reprocessed as soon as is practical following use.
No particular requirements. F
Review the washer-disinfector warning above, before using this cleaning method. Remove devices from instrument trays before placing into washer baskets. Orient devices following recommendations of the washer/disinfector manufacturers. Verify that devices are visually clean after automated cleaning.
Recommended Washer Cycle
Pre-Wash: Cold tap water, 2 min.
Wash: 66oC, 5 min. using a neutral enzymatic detergent, pH 6.0–8.0
Rinse: Hot tap water, 1 min.
enzymatic detergent, pH 6.0–8.0.
Brush motor case and collet with a nylon brush dampened with a neutral enz be sure to brush under the  nger control lever.
Rinse motor thoroughly under running water, collet end pointed down. Dry collet and motor with lint free towel.
Verify that device
polypropylene wrap
Steam Sterilization:
Cycle:
Temperature:
Time:
Drying:
s are visually clean after manual cleaning.
ollow hospital procedures.
es of the motor and hose, and wipe inner surface of oiler housing with a cloth dampened with a neutral
Gravity
132°C
25 min.
15 minutes
eam sterilization.
Pre-vac
132°C
4 min.
15 minute
ymatic detergent. If using the MR7 Touch motor,
Pre-vac (FR/WHO)
134°C
18 min.
s
20 minutes
Pre-vac (UK)
134°C
3 min.
10 minutes
*Items contaminated with Transmissible Spongiform Encephalopathies (TSE) agents may be decontaminated using steam autoclave at a temperature of 134–137°C f referenced in NHS Estates HTM 2010 parts 4 & 6: Appendix 2, Items Contaminated With TSE Agents and WHO Infection Control Guidelines for Transmissible Spongiform Encephalopathies. Medtr suspected or con rmed with a TSE diagnosis be incinerated.
STERRAD Sterilization: Do not use low temperature hydrogen peroxide gas plasma sterilization due to lumen internal diameter and length restrictions.
100% EtO Sterilization Parameters:
Preconditioning: 51–59°C, 70 ±5% relative humidity, 30 min.
Temperature: 51–59°C
Relative humidity: 70 ±5%
Ethylene oxide concentration: 725 ± 25mg/L
Gas exposure time (full-cycle): 4 hours
Aeration: 18 hours at 51–59°C
Steris: Do not use liquid perac
18
or a single cycle of 18 minutes or repeated for a total of six 3-minute cycles as
onic recommends that all products used on a patient
etic acid sterilization due to immersion procedure.
Page 23
Cleaning & Sterilizing the MR7 System
PTSC224
Maintenance, Inspection and Testing
Storage Store with other sterile devices.
Inspect devices for any damage before and after each use. If damage is observed, do not use the device until it is repaired.
After cleaning and sterilization, verify functionality prior to re-use.
NOTE: The instructions provided above have been validated by the manufacturer as being CAPABLE of preparing the product for re-use. It remains the responsibility of the processor to ensure that the reprocessing as actually performed, using equipment, materials and personnel in the reprocessing facility, achieves the desired result. This normally requires validation and routine monitoring of the pr
ocess.
Legend Attachments / Tubes
Warnings and Precautions
Limitations Verify functionality prior to re-use.
INSTRUCTIONS
Point of Use No particular requirements.
Containment and Transportation
Preparation for Decontamination
Cleaning: Automated
(Do NOT use ultrasonic washer)
Cleaning: ManualW
Disinfection No particular requirements Follow hospital procedures.
Packaging F
Do not soak/submerge Legend devices.
Do not use ultrasound to clean Legend devices.
Do not use chlorine based or corrosive cleaning agents such as bleach, lye, acetone, sodium hypochlorite/bleach, sodium hydroxide, formic acid, or solutions containing glutaraldehyde.
The use of a washer-disinfector for cleaning may cause a pre-mature degradation in performance.
Allow an adequate cooling period after st
Use ONLY nylon cleaning brushes. Non-nylon cleaning brushes leave residue that may prevent the tool from being secured properly in the handpiece.
It is recommended that devices are reprocessed as soon as is practical following use.
No particular requirements. F
Review the washer-disinfector warning above, before using this cleaning method. Manually rinse attachments/tubes under tap water before placing into washer baskets. Orient devices following recommendations of the washer/disinfector manufacturers.
Recommended Washer Cycle
Pre-Wash: Cold tap water, 2 min.
Wash: 66°C, 5 min. using a neutral enzymatic detergent, pH 6.0–8.0
Rinse: Hot tap water, 1 min.
Immerse the head of Contra-Angle attachments in surgical instrument cleaning solution and run the motor for 1 minute.
Other attachments and tubes may be mechanically agitated in cleaning solution, but not soaked or immersed.
A nylon brush dampened with a surgical instrument cleaning solution may be used to clean the external sur connecting surfaces of the attachments and tubes.
Straight attachments, footed attachments and telescoping straight tubes have special cleaning brushes sized to the
attachment’s or telescoping tube’s internal diamet the attachment or telescoping tube from rear to front to loosen and remove debris trapped inside.
Move any moveable parts back and forth to allow solution to thoroughly clean attachment, e.g., sleeve on footed attachment, perforator attachment.
Rinse thoroughly with tap water.
Thoroughly dry attachments. An air gun may be used to blow moisture out from rear to front of attachment.
Using an aerosol spray lubricant (such as Pana Spray), per
polypropylene wrap.
, until no visible soil is noticed, before placing them into the automatic washer. Remove devices from instrument trays
ipe all attachments and telescoping tubes with a cloth, dampened with a surgical instrument cleaning solution.
Holding the can approximately 10–15 cm (3–6 in.) away from the attachment, spray all components that move, rotate, or slide with three quick squirts.
Articulate movable components to ensure proper lubrication.
Remove excess lubricant with a clean cloth.
or sterilization, place devices in instrument tray. Devices may be unwrapped, or wrapped with up to two layers of 1-ply
ollow hospital procedures.
eam sterilization.
faces and internal
er. Push the brush wet with surgical instrument cleaning solution through
form the following steps to lubricate attachments:
19
Page 24
Cleaning & Sterilizing the MR7 System
PTSC224
Sterilization
(Temperatures are minimum required, times are minimum required)
Maintenance, Inspection and Testing
Storage Store with other st
Steam Sterilization:
Cycle:
Temperature:
Time:
Drying:
*Items contaminated with TSE agents may be decontaminated using steam autoclave at a temperature of 134–137°C for a single cycle of 18 minutes or repeated for a total of six 3-minute cycles as referenced in NHS Estates HTM 2010 parts 4 & 6: Appendix 2, Items Contaminated W Encephalopathies. Medtronic recommends that all products used on a patient suspected or con rmed with a T be incinerated.
STERRAD Sterilization: Do not use low temperature hydrogen peroxide gas plasma sterilization due to lumen internal diameter and length restrictions.
100% EtO Sterilization Parameters:
Preconditioning: 51–59°C, 70 ±5% relative humidity
Temperature: 51–59°C
Relative Humidity: 70 ±5%
Ethylene oxide concentration: 725 ± 25mg/L
Gas exposure time (full-cycle): 4 hours
Aeration: 18 hours at 51–59°C
Steris: Do not use liquid peracetic acid sterilization due to immersion pr
Inspect devices for any damage before and after each use. If damage is observed, do not use the device until it is repaired.
Verify functionality prior to re-use.
Gravity
132°C
25 min.
15 minutes
erile devices.
Pre-vac
132°C
4 min.
15 minutes
ith TSE Agents and WHO Infection Control Guidelines for Transmissible Spongiform
, 30 min.
Pre-vac (FR/WHO)*
134°C
18 min.
20 minutes
ocedure.
Pre-vac (UK)*
134°C
3 min.
10 minute
s
SE diagnosis
NOTE: The instructions provided above have been validated by the manufacturer as being CAPABLE of preparing the product for re-use. It remains the responsibility of the processor to ensure that the reprocessing as actually performed, using equipment, materials and personnel in the reprocessing facility, achieves the desired result. This normally requires validation and routine monitoring of the process.
20
Page 25
Cleaning & Sterilizing the MR7 System
PTSC224
MR7 Pneumatic Control Unit / Regulator Hose / Triton Adapter / Instrument Case
Warnings and Precautions
Limitations Verify functionality prior to re-use.
INSTRUCTIONS
Point of Use No particular requirements. Follow hospital procedures.
Containment and Transportation
Preparation for Decontamination
Cleaning: Automated
(Do NOT use ultrasonic washer)
Cleaning: ManualWipe the pneumatic control unit, regulator hose, Triton adapter,
Disinfection No particular requirements. Follow hospital procedures.
Packaging No particular requirements. F
Sterilization Do not sterilize pneumatic control unit, regulator hose, or Triton adapter.
Maintenance, Inspection and Testing
Storage Do not store with sterile devices.
Do not soak/submerge MR7 devices.
Do not use ultrasound to clean MR7 devices.
Do not use chlorine based or corrosive cleaning agents such as bleach, lye, acetone, sodium hypochlorite/bleach, sodium hydroxide, formic acid, or solutions containing glutaraldehyde.
Do not sterilize the MR7 pneumatic control unit, regulator hose, or Triton adapter.
It is recommended that devices are reprocessed as soon as is practical following use.
No particular requirements. F
Not validated.
instrument cleaning solution after each use.
Inspect devices for any damage before and after each use. If damage is observed, do not use the device until it is repaired.
Verify functionality prior to re-use.
ollow hospital procedures.
and instrument case with a cloth dampened with surgical
ollow hospital procedures.
NOTE: The instructions provided above have been validated by the manufacturer as being CAPABLE of preparing the product for re-use. It remains the responsibility of the processor to ensure that the repr
ocessing as actually performed, using equipment, materials and personnel in the reprocessing facility, achieves the desired result. This normally requires validation and routine monitoring of the process.
Transmissible Spongiform Encephalopathies (TSE) Return Policy
Medtronic Powered Surgical Solutions (MPSS) will not authorize or accept the return of MPSS products that directly contact patients or is contaminated with a patient’s body  uids who is suspected or con rmed with a Transmissible Spongiform Encephalopathies / Creutzfeldt-Jakob Disease (TSE/CJD) diagnosis. Futhermore, MPSS recommends that all MPSS products used on a patient con rmed with a TSE diagnosis be incinerated. Contact your Sales Representative for replacement of product incinerated under this policy or for temporary equipment while original equipment is quarantined. Contact MPSS Regulatory A airs Department for additional inf patient suspected of a TSE/CJD diagnosis must be incinerated.
If TSE/CJD is excluded as a diagnosis, the quarantined reusable equipment may be returned f cleaning, decontamination and sterilization. Hospital personnel should contact their infection control personnel for current procedures and policy for reusable equipment processing when suspect of contamination with Creutzfeldt-Jakob Disease (CJD) or other Transmissible Spongiform Encephalopathies (TSE).
Reusable devices that have been used on patients with suspected Creutzfeldt-Jakob Disease (CJD Spongiform Encephalopathies (TSE) should be quarantined and not reused until diagnosis is con rmed or excluded.
Reusable devices should be quarantined after having been cleaned, decontaminated, sterilized and packed in a ridged sealed container until  nal diagnosis.
ormation regarding TSE contamination. MPSS dissecting tools used on a
or use after appropriate
) or other Transmissible
21
Page 26
Trou bleshooting
PTSC224
Troubleshooting
NOTE: All Legend and MR7 devices returned for servicing or refurbishing should be properly cleaned and sterilized prior to shipping.
Motor not Running or Low on Power:
Possible Cause Solution
Hoses not properly connected. Make sure all connections are secure.
Operating pressure inadequate. Check gas supply pressure gauge. Increase pressure
Attachment not pr
Foot pedal on pneumatic control unit not functioning pr
operly.
Motor stalls. Manually spin the dissecting tool, then activate the
according to compressed gas requirements, if necessary.
operly installed and locked onto the motor. Remove and re-install attachment and tool to ensure
proper installation and locking of attachment onto motor.
Check for obstructions under the foot pedal. If foot pedal continues to fail, return the pneumatic control unit to MPSS to be refurbished.
motor. If the motor continues to stall, return it to MPSS to be refurbished.
Motor Continues to Run:
Possible Cause Solution
Pneumatic control unit is not functioning properly. Depressurize the system and return the pneumatic
control unit to MPSS to be refurbished.
Finger control is not functioning properly. Return motor to MPSS to be refurbished.
Pneumatic contr
ol unit is locked in the  nger control position. Move the  nger control lever to the foot control position.
System Makes an Abnormal Noise:
Possible Cause Solution
Inadequate lubrication. Check for proper installation of the lubricant/di user
cartridge. If the problem persists, replace the cartridge. If replacing the cartridge doesn’t  x the problem, return the motor to MPSS f
Motor’s exhaust hose is damaged, or internal pressure hose is
detached.
Worn bearings.Switch attachments to det
Attachment not properly installed and locked onto the motor. Remove and reinstall attachment and tool to ensure
Safety relief valve has been activated by high air pressure.Ensure that air operating/dynamic air pressure is no
Depressurize the system and return the motor to MPSS
to be refurbished.
are failing in the motor or in the attachment. Return the
failing component to MPSS to be refurbished.
proper installation and locking of the attachment onto the motor.
higher than 120 psi.
or refurbishing.
ermine whether the bearings
22
Page 27
Trou bleshooting
PTSC224
Motor is Too Hot to Touch/Hold:
Possible Cause Solution
Inadequate cool down period following sterilization. Motor must be allowed to cool down following steam
sterilization.
Inadequate lubrication. Check for proper installation of the lubricant/di user
cartridge.
Attachment transferring heat to the motor
Heavy side loading during dissec
Inadequate irrigation. Ensure adequate irrigation to surgical site during bone
tion. Discontinue use and rest the motor by using it
. Switch attachments to determine whether the heat is
being generated by the motor or the attachment. Return the failing component to MPSS for refurbishing.
intermittently or wrap the motor with a moist sterile towel. If overheating continues, return the motor to MPSS for refurbishing.
dissection.
Attachment Will not Properly Seat on the Motor:
Possible Cause Solution
Motor collet  ats are not aligned. Use the Legend motor wrench to rotate the  at closest to
the motor case until its marker is aligned with the marker
on theat farthest away fr
om the motor case.
Tool is Dif¿ cult to Remove from Attachment:
Possible Cause Solution
Aging of attachment. Return to MPSS to be refurbished, or purchase new
Improper cleaning.
Use of reprocessed tools.
Use of an unauthorized refurbisher.
equipment.
16-MF Contra-Angle Attachment is Overheating:
Possible Cause Solution
The Contra-Angle attachment operates by a set of internal gears to engage the drive shaft. It is normal for some heat to be generated approximately 2cm from the distal end of the attachment and at the right of the angle head.
Verify pressure setting of 80 psi (5.5 bar). If
heat continues or is excessive, return the attachment to
MPSS to be refurbished or purchase new equipment.
Perforator is Running too Slow:
Possible Cause Solution
Pressure set incorrectly. Check the pressure setting at the foot control.
23
Page 28
Trou bleshooting
PTSC224
Dissecting Tool Flails:
Possible Cause Solution
A non-Legend dissecting tool is being used. Replace with a Legend dissecting tool.
Worn attachment or tube bearings.Try another attachment or tube to isolate the location
Attachment/tube and tool are not compatible. Match color code on the dissecting tool packaging to the
Motor is damaged. Return motor to MPSS to be refurbished.
Dissec at certain speeds.
Dissecting Tool Vibrates Excessively:
Possible Cause Solution
Dissecting tool’s size and geometry may create excessive vibration at certain speeds.
ting tool’s size and geometry may contribute to  ailing
of the problem. If the attachment is failing, return it to MPSS. I
f the tube is failing, dispose of it and use a new
tube.
color code on the attachment/tube.
Adjust the speed by changing the pressure setting or fo
ot/ nger control. Do not use if  ailing persists. Change
dissecting tools.
Adjust the speed by changing the pressure setting or foot/ nger control. Change dissecting tools.
Dissecting Tool Will not Seat Properly in the Motor or Attachment Collet:
Possible Cause Solution
Debris in collet of attachment or motor. Clean the attachment or motor thoroughly according
to the instructions in this manual. If cleaning does not correct the problem, return the attachment or motor to MPSS to be refurbished.
A non-Legend dissec
ting tool is being used. Replace with a Legend dissecting tool.
Smoke is Generated by the Attachment or Motor:
Possible Cause Solution
Attachment is not in the locked position. Make sure the attachment is in the locked position.
24
Page 29
Refurbishing or Repairs
PTSC224
Refurbishing or Repairs
When the MR7 System requires servicing or refurbishing, contact Medtronic Powered Surgical Solutions Repair Services for a return authorization and instructions for returning the equipment. Medtronic Powered Surgical Solutions provides quality assured service by factory-trained personnel who will utilize genuine Midas Rex Legend parts as required. All items being returned for servicing or refurbishing should be properly cleaned and st
Contact:
Medtronic Powered Surgical Solutions Repair Services: (800) 335-9557 or (817) 788-6440 mmrcustomerservice@medtronic.com
Peak performance, reliability and maximum service life from your MR7 System may be assured by using only those Midas Rex Legend products for your MR7 System that are manufactured by and sold through Medtronic Powered Surgical Solutions, Fort Worth, Texas. While Medtronic Powered Surgical Solutions guarantees complete compatibility among its products within a speci c product line, the dissecting tools are designed for single-use only, and Medtronic disclaims any responsibility when reprocessed dissecting tools are used. If you would like more information about the patient and product risks associated with reprocessed tools, please contact the number or e-mail address list
Due to safety and environmental concerns, Medtronic Powered Surgical Solutions requests the return of pneumatic high
speed motors for proper disposal at the end of the product life cycles.
erilized prior to shipping.
ed above.
25
Page 30
Preventative Maintenance
PTSC224
Preventative Maintenance
The Midas Rex MR7 System Preventive Maintenance Manual has been developed to assist you in getting the greatest ownership value from your MR7 System, while helping to maximize its performance, safety and reliability. The scheduled preventive maintenance/serv preventive maintenance manual for the speci c steps necessary to maintain the MR7 System.
ice program is in addition to the required routine cleaning after each use. Please refer to the
26
Page 31
Limited Warranty
PTSC224
A. This Limited Warranty provides the following assurance to the purchaser of a Medtronic Midas ReMR7 Pneumatic
High Speed System. This Limited Warranty is extended only to the buyer purchasing the MR7 System directly from Medtronic or from its a liate or its authorized distributor or representative. The Midas Rex® MR7 Pneumatic High Speed System includes the motor, foot control, instrumentation cases and trays (hereafter referred to as Syst Components), straight and angled motor attachments (hereinafter referred to as “Attachments”), telescoping tubes (hereinafter referred to as Semi-reusable Components) and dissecting tools and other accessories not listed above and jointly referred to as MR7 Pneumatic High Speed System, unless speci cally noted.
(1) Should a System Component fail to function to Medtronic’s published speci cations during the term of
this Limited Warranty (one year fr date of sale of a refurbished or used System Component), Medtr Component or any portion thereof.
(2) Should an Attachment fail to function to Medtronic’s published speci cations during the term of this
Limited Warranty (90 days from the date of sale of a new Attachment), Medtronic will either repair or replace the Attachment or any portion thereof.
(3) Should a Semi-reusable Component fail to function to Medtronic’s published speci cations during
the term of this Limited Warranty (30 days fr Medtronic will replace the Semi-reusable Component or any portion thereof
(4) Should a Single Use Component fail to function to Medtronic’s published speci cations prior to its “use
by” date Medtronic will replace the Single Use Component.
B. To qualify for this Limited Warranty, these conditions must be met:
(1) The Product must be used on or before its “Use By” or “Use Before” date, if applicable.
(2) The Product must be used in accordance with its labeling and may not be altered or subjected to misuse,
abuse, ac
(3) Medtronic must be noti ed in writing within thirty (30) days following discovery of a defect.
(4) The Product must be returned to Medtronic within thirty (30) days of Medtronic receiving notice as
pr
(5) Upon examination of the Product by Medtronic, Medtronic shall have determined that: (i) the Product was
not repaired or altered by anyone other than Medtronic or its authorized representative, (ii) the P was not operated under conditions other than normal use, and (iii) the prescribed periodic maintenance
and services, if applicable, have been performed on the Product
C. This Limited Warranty is limited to its express terms. THIS LIMITED WARRANTY IS IN LIEU OF ALL OTHER
WARRANTIES, EXPRESSED OR IMPLIED WHETHER STATUTORY OR OTHERWISE, INCLUDING ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. In no event shall Medtronic be liable for any
consequential, incidental MR7 System, whether a claim for such damage is based upon the warranty, contract, negligence or otherwise.
D. The exclusions and limitations set out above are not intended to, and should not be construed so as to, contravene
mandatory provisions of applicable law. Users may benet from statutory warranty rights under legislation
governing the sale of consumer goods. If any part or term of this Limited Warranty is held by any court of competent jurisdiction to be illegal, unenforceable, or in con ict with applicable law, portion of the Limited Warranty shall not be a ected, and all rights and obligations shall be construed and enforced as if this Limited Warranty did not contain the particular part or term held to be invalid.
cident or improper handling.
ovided for in (3) above.
, prospective or other similar damage resulting from a defect, failure, or malfunction of the
om the date of sale of a new System Component or 90 days from the
onic will either repair or replace the Motor
om the date of sale of a new Semi-reusable Component),
.
the validity of the remaining
Limited Warranty
em
roduct
27
Page 32
Appendix A—Specialized Attachments
PTSC224
Appendix A—Specialized Attachments
Angled Attachments
Installation:
1. Slide the attachment over the motor collet, aligning the triangular markers (Figure 27). You will feel and hear the attachment click into place when it is fully seated.
2. Turn the attachment to the case (Figure 28).
3. With the “ tool lock” in the position, insert the dissecting tool into the attachment with a slight rotational motion (Figure 29). You will feel and hear the tool click into place when it is fully seated.
4. Turn the tool lock to the pull on the shaft of the dissecting tool to verify proper installation.
NOTE: A dissecting tool may be installed in the attachment before the angled attachment is placed on the motor.
CAUTION: Hold the handpiec attachment, so that the attachment does not
inadvertently loosen from the handpiece.
position on the motor
position (Figure 30). Gently
e assembly by the
Figure 27: Sliding the Attachment over the Motor Collet
Figure 28: Attachment in the Locked Position
Removal:
Removal is the reverse of installation.
Figure 29: Inserting the Dissecting Tool into the Attachment
Figure 30: Tool Lock in the Locked P
osition
28
Page 33
Appendix A—Specialized Attachments
PTSC224
Angled Double Lock Attachments
Installation:
1. Slide the attachment over the motor collet, aligning the triangular markers.
2. Pull the attachment towards the motor and turn the attachment to the
3. With the “ tool lock” in the position, insert the dissecting tool into the attachment with a slight rotational motion. You will feel the tool click into place when it is fully seated.
4. Turn the tool lock to the the shaft of the dissecting tool to verify proper installation.
5. On Variable Exposure Attachments, use the TUBE adjustment ring to adjust the exposure of the dissecting tool. With the tool pointing away fr you, turn the ring to the right to increase the length of the tube, thereby decreasing the exposure of the tool. Turn the ring to the left to decrease the length of the tube, thereby increasing the exposure of the tool
.
position.
position. Gently pull on
om
Figure 31: Installing the attachment.
Figure 32: Inserting the Dissecting Tool into the Attachment
Removal:
1. To remove the attachment, hold the motor in the palm of your hand, and push the sleeve on the
attachment distally while turning the attachment to the
position.
2. Release the sleeve and remove the attachment.
Figure 33: Removing the Attachment
29
Page 34
Appendix A—Specialized Attachments
PTSC224
Curved Bur Attachments
WARNINGS:
• Use adequate irrigation and keep the cooling sleeve soaked during dissection. Inadequate irrigation may cause thermal necrosis.
• Do not modify the bur. Bending or prying may break the bur, causing harm to the patient or operating
oom sta .
r
• Excessive pressure applied to the bur may cause bur damage. If this occurs, use extreme care to ensure that all fragments of the bur are remo patient.
• Disposable devices are for single-use only. Do not attempt to sterilize disposable devices. The disposables are packed sterile and are not intended for repeat use. T once.
• Test for bur wobble (eccentricity) at the desired speed prior touse. Select a new bur or reduce speed if wobble is observed prior to use or during the procedure. Bur wobble may cause patient injur
CAUTIONS:
Curved burs are not designed for variable tool exposure. Do not attempt to adjust tool exposure, as this may damage the device.
o prevent contamination, use only
ved from the
y.
Cooling the Bur
1. Prior to initial use, soak the cooling sleeve by dipping it into a cup of saline or DI water, as shown below.
Wetting of the cooling sleeve prior to cutting
2. During use, maintain copious irrigation of the cooling sleeve and bur tip by dribbling saline or DI water along the entire length of the cooling sleeve and bur tip.
Operation
1. Activate the motor and gently press the curved bur against the bone to begin dissection.
2. Use a light sweeping motion to continue removing bone.
Removal
1. Turn both AT10 (V01) locking rings to the position and pull the bur out of the attachment.
2. Discard the curved bur, according to hospital procedures.
Reuse and Cleaning
Do not reuse.
• When operating or t
bur is properly inserted and locked into AT10 attachment.
esting the motor, ensure the
Installation
1. Turn the tube locking ring and the tool locking ring on the AT10 (V01) Telescoping Attachment into the
position.
2. Slide the curved bur into the AT10 (V01), until the hub is fully seated.
3. Gently press on the head of the dissecting tool to lock it into place. You will hear a click when the curved bur is fully seated.
4. Turn both locking rings to the AT10 (V01).
5. Gently pull on the curved bur to verify proper installation.
6. Refer to your motor’s Instruction Manual for information on setting up the AT10 (V01) attachment and motor.
7. Brie y run the motor with the curved bur installed, checking for bur w If either occurs, perform the corrective actions described in the Troubleshooting section of these instructions.
obble or excessive vibration.
position on the
Sterility
Each bur is gamma-sterilized and is not intended for repeated use. Do not attempt to re-sterilize.
Cutting Time
Continuous cutting for extended periods may cause the device to heat to an uncomfortable temperature. To avoid
this, limit cutting to the recommended times below:
Maximum
Continuous
Cutting Time
1 & 2 mm burs 2 minutes 12 minutes
3 & 4 mm burs 3 minutes 12 minutes
Maximum Total
Cutting Time
30
Page 35
Variable Exposure Attachments
PTSC224
CAUTION: The Legend Variable Exposure attachments can be distinguished from standard attachments by the dual color bands on the attachment. Match the color band on the attachment to the color code on the
ting tool packaging.
dissec
WARNINGS:
• Surgeons should familiarize themselves with the performance of dissecting tools before use, and
should explore the e ect of various levels of tool exposure on dissection stability. excessive chatter, vibration, or movement, decrease the tool exposure.
• Dissecting tool size and geometry may contribute to excessive vibration at certain speeds. I
decrease speed by adjusting the foot/ nger control, or by changing the operating pressure or console
speed setting. If necessary, use a di erent dissecting tool.
If the tool exhibits
ncrease or
Appendix A—Specialized Attachments
Figure 34: Adjusting Tool Exposure
Installation:
1. Refer to the appropriate section of this manual on installing  xed or angled attachments.
2. After installation, use the TUBE adjustment ring to adjust the exposure of the dissecting tool (Figure 34). With the tool pointing away fr ring to the right to increase the length of the tube, thereby decreasing the exposure of the tool. Turn the ring to the left to decrease the length of the tube, thereby increasing the exposure of the tool
WARNING: Do not use the Variable Exposure Attachment if the TUBE adjustment ring spins freely or fails to click into plac exposure may change without warning.
CAUTION: Make sure that the tool lock (angled attachments only) and the attachment lock are still in the
position after each adjustment of the tool exposure. Attempting to increase the exposure too far may result in the attachment becoming unlocked. Accidentally turning the tool lock may result in reduced speed and/or overheating of the attachment.
WARNING: Do not use the end of the tube as a depth gauge or depth stop.
e with each adjustment, as the
om you, turn the
.
Removal:
1. Removal is the reverse of installation.
2. When cleaning, clean the attachment completely,
 rst without adjusting the tube length, then with the tube fully extended, and with the tube fully retrac
ted.
31
Page 36
Appendix A—Specialized Attachments
PTSC224
Footed Attachments
Installation:
1. Insert the dissecting tool into the motor collet with a slight rotational motion (Figure 35). You will feel and
hear the tool click into place when it is fully seated.
2. Slide the footed attachment over the dissecting tool
, onto the motor, aligning the triangular markers
(Figure 36).
3. Pull the footed attachment towards the motor and turn the attachment to the
Removal:
1. To remove the Legend footed attachment, hold the motor in the palm of your hand, and push the sleeve on the attachment distally while turning the attachment to the position (Figure 38).
WARNING: Remove Legend Footed Attachments cautiously and slowly per instructions to avoid injury to the operator.
2. Release the sleeve and remove the attachment.
3. Pull the dissecting tool out of the motor collet and discard
the tool.
position (Figure 37).
Figure 35: Inserting the Tool into the Collet
Figure 36: Sliding the Attachment onto the Motor
Figure 37: Attachment in the Locked Position
32
Figure 38: Removing the Attachment
Page 37
Appendix A—Specialized Attachments
PTSC224
Rotating Footed Attachments
NOTE: Rotating and  xed footed attachments with the same length, marking, and color band share the same dissecting tools.
Installation:
1. Insert the dissecting tool into the motor collet with a slight rotational motion (Figure 39). You will feel and
hear the tool click into place when it is fully seated.
2. Slide the attachment over the dissecting tool and onto the motor (Figure 40). You will feel and hear the attachment click into place when it is fully seated.
3. Turn the attachment to the
NOTE: The footed end of the attachment now has 360˚ of unrestricted rotation.
Removal:
Removal is the reverse of installation.
, aligning the triangular markers
position (Figure 41).
Figure 39: Inserting the Tool into the Collet
Figure 40: Sliding the Attachment onto the Motor
Figure 41: Attachment in the Locked Position
33
Page 38
Appendix A—Specialized Attachments
PTSC224
Contra-Angle Attachment (16-MF)
CAUTION: Do not run the 16-MF attachment with operating pressure above 80 psi (5.5 bar). This may
cause over-heating and damage to internal gears of attachment.
Installation:
1. Decrease the pressure at the compressed gas source to 80 psi (5.5 bar).
2. Adjust the pressure as needed by lowering it at the gas source, then push down the pressure relief on the pneumatic contr
pressure in the hoses.
3. Slide the attachment over the motor collet, aligning the triangular markers (Figure 42). You will feel and hear the attachment click into plac seated.
4. Turn the attachment to the position (Figure 43).
5. Turn the attachment head lever to the open position (Figure 44).
6. Insert the dissecting tool and return the lever to the
closed position. Gently pull on the dissecting tool shaft to verify pr
ol unit to exhaust the excess
e when it is fully
oper installation.
Figure 42: Sliding the Attachment over the Motor Collet
Figure 43: Attachment in the Locked Position
Removal:
1. Removal is the reverse of installation.
2. Discard the dissecting tool after removing it from the attachment.
NOTE: A dissecting tool may be installed and locked in the attachment before the Contra-Angle attachment is
installed on the motor
.
Figure 44: Opening the Attachment Head Lever
34
Page 39
Metal Cutting Attachments
PTSC224
WARNING: For metal transection, observe the following safety guidelines:
• Wear eye protection.
• Irrigate well to cool the cutting surfaces.
• Protect the wound site from metal debris.
Appendix A—Specialized Attachments
Figure 45: Sliding the Attachment over the Motor Collet
• Use a clamp or grasping device to contr fragments during transection of any metal component.
The Metal Cutting Attachment uses the tungsten carbide or diamond wheel dissecting tools.
ol loose
Installation:
1. Slide the attachment over the motor collet aligning the triangular markers (Figure 45). You will feel and hear the attachment click into place when it is fully seated.
2. Turn the attachment to the
3. Turn the tool lock towards the icon, and insert the dissecting tool into the attachment with a slight rotational motion (Figure 46). You will feel and hear the tool click into place when it is fully seated.
4. Turn the tool lock to the pull on the shaft of the dissecting tool to verify proper installation.
NOTE: A dissecting tool may be installed and locked in the attachment before the metal cutting attachment is
installed on the motor
.
position.
position (Figure 47). Gently
Figure 46: Inserting the Tool into the Attachment
Removal:
Removal is the reverse of installation.
WARNING: Do not use metal cutting dissecting tools on bone.
Figure 47: Turning the T
ool Lock to the Locked Position
35
Page 40
Appendix A—Specialized Attachments
PTSC224
Telescoping Attachments
Installation:
1. Slide the attachment over the motor collet aligning the triangular markers (Figure 48). You will feel and hear the attachment click into place when it is fully seated.
NOTE: A dissecting tool may be installed and locked in the attachment befo
installed on the motor.
re the telescoping attachment is
F
igure 48: Sliding the Attachment over the Motor Collet
2. Turn the attachment to the
3. Turn the TUBE locking ring towards the icon.
4. Insert the base of the telescoping tube into the attachment (Figure 49).
5. To lock the tube in place, turn the TUBE locking ring towards the
6. To insert the tool, make sure that the TOOL locking ring is in the tool into the top of the tube (Figure 50). You will feel and hear the tool click into place when it is fully seated.
7. Turn the TOOL locking ring to the pull on the shaft of the dissecting tool to verify proper installation.
8. If the tube position needs to be changed, turn the TUBE locking ring towards the tube, then turn the TUBE locking ring towards the icon.
icon. Do not over tighten.
position, and insert the dissecting
position.
position. Gently
icon, re-position the
Removal:
1. To remove the attachment, unlock both locking rings, and pull the telescoping tube and tool out of the attachment.
2. Turn the attachment to the from the motor.
position and remove it
Figure 49: Inserting the Telescoping Tube
Figure 50: Inserting the Dissecting Tool
NOTE: Telescoping tubes are disposable following
multiple uses, and should be discarded when heat or excessive vibration is noticed, or when insertion of the tool becomes di cult.
36
Figure 51: Locking Rings
TUBE Locking Ring
TOOL Locking Ring
Page 41
Appendix A—Specialized Attachments
PTSC224
Perforator Driver Attachments
The Perforator Attachment has a Hudson chuck to drive any device with a Hudson shank, i.e., cranial perforator device.
NOTE: MPSS does not provide cranial perforator devices.
Installation:
1. Slide the attachment over the motor collet aligning the triangular markers (Figure 52). You will feel and hear the attachment click into place when it is fully seated.
2. Turn the attachment to the
3. To install a cranial perforator device with a Hudson shank, pull back on the collar of the attachment (Figure 54).
4. Insert the device and release the collar.
NOTE: A cranial perforator device may be installed in the attachment befo
motor.
re the attachment is installed on the
position (Figure 53).
Figure 52: Sliding the Attachment over the Motor Collet
Figure 53: Attachment in the Locked Position
Removal:
1. To remove the cranial perforator device, pull back on the collar of the attachment.
2. To remove the attachment, turn it to the position and slide it o of the motor.
WARNING: Consult the cranial perforator device labeling for the recommended speed speci cations.
Maximum Speed of Perforator Attachments
Gas Pressure
(Dynamic)
80 psi 660 rpm 850 rpm
100 psi 820 rpm 1050 rpm
120 psi 890 rpm 1140 rpm
Model AD01
Output Speed
(Max.)
Model AD03
Output Speed
(Max.)
Figure 54: Pulling Back on the Collar
37
Page 42
Appendix A—Specialized Attachments
PTSC224
Jacobs® Chuck Attachments
The Jacobs Chuck attachment is a non-cannulated 5/32” chuck with key for drilling.
Installation:
1. Slide the attachment over the motor collet aligning the triangular markers (Figure 55). You will feel and hear the attachment click into place when it is fully seated.
2. Turn the attachment to the
3. To install a drill bit, turn the Jacobs key to open the collar.
4. Insert the drill bit and tighten the collar.
NOTE: A drill bit may be installed in the attachment before the Jacobs Chuck attachment is installed on the
motor
.
Removal:
Removal is the reverse of installation.
position (Figure 56).
Figure 55: Sliding the Attachment over the Motor Collet
Figure 56: Attachment in the Locked Position
38
Page 43
Bone Mill Attachment
PTSC224
The Legend Bone Mill attachment is composed of a non sterile base and gamma-sterilized, single use
disposable components. The gamma-sterilized single use disposable components consist of one bowl, one cap
OW and HIGH settings), and one spatula. A Legend
(with L or MR7 motor provides the power to drive the Bone Mill
attachment. The Bone Mill attachment is intended to mill bone into particles 1 to 5mm in size.
F
igure 57: Bone Mill Attachment
1. Cap
2. Bowl
3. Bone Mill Base
4. Motor
5. Spatula
Appendix A—Specialized Attachments
1
2
3
WARNINGS:
Do not operate the Bone Mill attachment without the bow
Do not come in contact with the interior of the disposable bowl and cap during bone milling.
Do not use the Bone Mill disposable components f more than one surgical procedure, as this may cause cross-contamination and a ect patient safety.
Use only Legend/MR7 motors and Legend Bone
Mill disposables in combination with the Bone Mill attachment. Use of other device
manufacturer’s warranty.
l, and cap secured in place.
s will void the
Using the Bone Mill Attachment
1. Place the bowl onto Bone Mill base and secure the latches.
2. Lay harvested bone specimens into the bowl.
CAUTIONS:
• The addition of any liquid to the milling process should be administered in small volumes to prevent
leakage.
or
5
4
Figure 58: Bone Mill Cap
HIGH
LOW
Figure 59: Inserting the Motor
• Do
• Do not use Bone Mill disposables to process more
3. Secure the cap onto the bowl by rotating
4. Use the HIGH cap position when starting with large
5. Insert the motor into the motor connection on the
6. Activate the motor to mill bone. Continue to run the
not use Bone Mill disposables to process more than 6 bone pieces approximately 1.5 cc in size at one time.
than a total of 40 cc of bone during any one surgical
pr
ocedure.
counterclockwise. Do not activate the Bone Mill Attachment without securing the cap.
piec
es of bone or large batch amounts. Use the LOW
cap position to  nish milling a batch or when starting
with small pieces or batch amounts (Figure 58).
Bone Mill base and r (Figure 59).
Bone Mill Attachment until desired bone particle size is achieved. Adjust motor speed to obtain desired particle size.
otate the motor to the position
39
Page 44
Appendix A—Specialized Attachments
PTSC224
NOTE: If bone does not process, deactivate the motor, rotate the motor to the position, and remove the cap. Reinsert the cap in the HIGH position and activate the motor. If bone spins without milling or jams, divide bone chunks into two batches and process separately.
7. Process the bone f to gauge quality of milling before continuing.
8. Continue to process bone in 5–15 second batches until the majority of particles reach desired size.
9. To remove bone particles stop operation of the motor and r
cap by rotating it clockwise. Use the spatula inside the bowl and cap to remove bone particles.
NOTES:
Milling time is dependent on the density of the patients bone, the ratio of cancellous and cortical
bone, the size of the individual particles and the batch amount of the bone to be pr check bone after 10–15 seconds to gauge quality of
milling.
• I
f the blade becomes inoperable, stop operation of
the motor and rotate the motor to the Remove the cap and use the spatula to free bone and/or blades.
• After removing bone particles from the bowl, the milling process may be repeated with additional
bone specimens, as needed bone may be processed by the Bone Mill Disposable
during any one surgical procedure.)
or 5–15 seconds and visually check
otate it to the position. Remove the
ocessed. Always
position.
. (Up to a total of 40 cc of
After Use
1. Rotate the motor to the position.
2. Unlatch the Bone Mill disposables from the base.
3. Discard the bowl, cap, and spatula per hospital procedure.
4. Clean the Bone Mill base according to the directions below.
5. Clean the motor according to its instruction manual.
Bone Mill bowl, cap, and spatula are single use
devices. Opened but unused Bone Mill disposables may not be re-sterilized under any sterilization method.
Preventative Maintenance
The Bone Mill Attachment should be returned for factory maintenance annually (approximately 100 procedures).
If the Bone Mill Attachment does not function according to these instructions, call your Medtronic Regional Distributor or Medtr If necessary, return the Bone Mill Attachment base to your Distributor or Sales Representative to be repaired.
onic Neurologic Sales Representative.
40
Page 45
PTSC224
Loading...