The following symbols may appear within this manual, on product labeling, or on the product itself:
Attention, see Instructions for Use.Tube Control
STERILE
United States federal law restricts this
device to sale by or on the order of a
physician.
Reference NumberUse with
Lot NumberMulti-Use Disposable
Serial Number
Quantity
Instrument Case
Air Pressure Relief
Attachment
Instrument Case
Lubricant/Diffuser
Lubricant/Diuser
For single patient use only. Do not
re-use, re-process, or re-sterilize this
product. Re-use, re-processing or
re-sterilization may compromise the
structural integrity of the device and/
or create a risk of contamination of the
device, which could result in patient
injury, illness, or death.
Approximately equal to
Non-Sterile
Sterilized by Gamma Irradiation
Dissecting Tool
Attachment
Control Unit
Refurbished
Dissecting Tool
Attachment
Control Unit
Refurbished
+
Use by date
Accessory
Accessory
Date of manufacture
REGULATOR
Temperature Limitations
Unlock
Lock
|
On
OffUSA Only
Finger-Operated ControlManufacturer
Foot-Operated ControlDo not dispose to unsorted
Compliant with European Council
Directive MDD 93/42/EEC.
Bone Mill
MOTOR
Brush
Adapter
Regulator
Bone Mill
Motor
Brush
Adapter
municipal waste.
Tool Control
EC REP
Authorized Representative in the
European Community
Consult Instructions for Use
Page 3
Contents
PTSC224
General Information ...........................................................................................................................1
Indications for Use ..............................................................................................................................1
Bone Mill Attachment .......................................................................................................... 39
ii
Page 5
General Information
PTSC224
Read and understand this manual before use of theMR7 System.
The Midas Rex® MR7 system is designed for use by medical professionals familiar with powered surgical instrumentation.
The surgeon is responsible f
be harmful. It is strongly recommended that the surgeon and dedicated operating room personnel are knowledgeable with
the use of this equipment by being trained in Medtronic Midas Rex Hands-On Workshops or by one of the local authorized
representatives.
The MR7 system consists of the following components:
• MR7 or MR7 T
• MR7 PneumaticControl Unit with Various Connectors
• MR7 Regulator Hose
• MR7 Lubricant/Di user Cartridge
• MR7 Triton® Adapter (optional)
• Legend® Attachments*
• Legend® Dissecting Tools*
*TheMR7 syst
ouch Motor
em usesthesameattachmentsand dissectingtoolsasthe Legend® Pneumatic High-Speed System.
or learningthe proper techniques in the use of thissystem, as inappropriate use may potentially
General Information
Indications for Use
The Medtronic Midas Rex MR7 System is a pneumatically operated surgical instrument system. The pneumatic motors
provide power to operate removable rotatingsurgical cuttingtoolsand their accessories intended for use in neurosurgery,
includingcraniotomy and spinal surgery; as well as Ear, Nose and Throat (ENT), orthopedic and general surgical applications
including maxillofacial, craniofacial and sternotom
y surgeries.
Contraindications
None
Special Notices
The words warning, caution and note have special meanings in this manual, and should be carefully reviewed:
WARNING: A warning indicatesthatthe personal safety of the patient or physician may be involved. Disregarding this
information could result in injury to the patient or physician.
CAUTION: A caution indicatesthatthere isa risk of damagingequipment.
NOTE: A note is intended to provideadditional information, which may be useful
procedure.
, but is notessential to completethe
1
Page 6
General Information
PTSC224
General Safety Precautions
WARNINGS:
• Do not use the Midas Rex MR7 System before proper cleaning and sterilization.
• Do not operate the Midas Rex MR7 System in the presence of Magnetic Resonance Imaging devices.
• Do not use damaged, faulty, or modi ed MidasRex MR7 S
damage prior to each use:
• Check the motor’sexhaust hose for cracks or tears.
• Visually inspect attachments and tools. Do not use bent or damaged tools.
• I
nstall attachment and dissecting tool, then brie y run motor.
* Check motor for overheatingand leaking lubricant.
* Check attachment for overheating.
* Check dissectingtool for ail.
• Do not operate the Midas Rex MR7 S
• Motorsand attachments which fail dueto extended use may allow acomponentto detach and fall from the motor or
attachment, and may cause patient injury.
• Heavy side loads and/or long operating periods may cause the device to overheat. If overheating occurs:
• Never place an overheated motor on the patient or draping during surgery.
• Discontinue use and rest the motor by using intermitt
steriletowel.
• If the motor is passed o , the receiver should grasp the motor by the proximal end close to the motor hose.
• T
o avoid injury to the patient or user, do not place the handpiece on the patient or in an unsecured location, when not
in use.
ystem withouteye protection.
ystem components. InspecttheMidasRex MR7 System for
ently, or wrap the motor/attachment interface with a moist
• MidasRex MR7 motorsshould only be operated when theattachment is in the
If a dissecting tool package is opened, but the tool is not used or contaminated, the tool can be re-sterilized by steam sterilization. Removetool from all original packagingand place into an approved autoclave package. Steam sterilizeas
follows:
High Vacuum Steam: 270° F (132° C) for 5 minutes
Gravity Displacement: 270° F (132° C) for 15 minutes
The re-sterilized tool must be used promptly following re-sterilization. If rust or corrosion isencountered after
re-sterilization, do not usethe re-sterilized tool
.
position.
No Latex Policy
Legend and MR7 products, packaging materials, labels, package inserts, and similar items manufactured by and/or for
Medtronic Powered Surgical Solutions (MPSS) do not contain latex.
2
Page 7
System Components
PTSC224
Non-Disposable Components
MR7 Motor
The MR7 motor is a high-speed, high-torque motor used to dissect bone and biomaterials.
Figure 1: Motor Components
System Components
1
5
1. Collet
2. Motor Case
3. Swivel
4. Hose
5. Finger Control Lever
3
2
4
6
6. Safety Slide (MR7 Touch Only)
In addition to the components listed in Figure 1, each MR7 motor hasa lubricant/di user housing at the end of the motor
hose, asseen in F
igure 2.
Figure 2: Lubricant/Di user Housing
3
Page 8
System Components
PTSC224
WARNING: Use only Medtronic Midas Rex Legend or MR7
devices with an MR7 motor. Use of other devices may
cause injury or damageequipment, and will void the
manufacturer’s warranty.
Pneumatic Control Unit
The pneumatic control unit (Figure 3) provides variable
speed motor control controls through a foot pedal. It also
allows the user to switch between nger and foot control
of the motor (if applicable).
Regulator
The regulator (Figure 4) controls the delivery pressure
of compressed gasto the pneumaticcontrol unit. The
pressuregauges monitor cylinder pressure (rightgauge)
and delivery pressure (leftgauge).
Note: Outlet pressure gauge accurate to +/- 12 psi.
Figure 3: Pneumatic Control Unit
Instrument Case
The instrument case (Figure 5) is used to organize
equipment.
Regulator Hose
Connects from thegassourceto the pneumaticcontrol
unitto deliver compressed gas.
Figure 4: Regulator
Figure 5: Instrument Case
4
Page 9
System Components
PTSC224
N2 DISS to Male Schrader Adapter
The N2 DISS to male Schrader adapter (Figure 6) allows
for the regulator hoseto beattached to a female Schrader
in-housegasconnection. A N2 DISS to female Schrader
adapter isalso available f
hoseto a male Schrader in-housegasconnection.
or connection of the regulator
N2 DISS to Air DISS Adapter
The N2 DISS to air DISS adapter (Figure 7) allows for the
regulator hoseto beattached to an Air DISS in-housegasconnection.
N2 DISS to WF4 Adapter
The N2 DISS to WF4 adapter (Figure 8) allows for the
regulator hoseto beattached to aMidasRex safety valve
regulator previously used for MidasRex Classic or Midas
Rex III motors. The in-line oiler must be removed from the
safety valve regulator.
Motor Wrench
The motor wrench (Figure 9) is used to align arrows
on motor collet ats prior to installation of a Legend
attachment.
Figure 6: N2 DISS to Male Schrader Adapter
Figure 7: N2 DISS to Air DISS Adapter
Figure 8: N2 DISS to WF4
Adapter
Triton Adapter
The Triton adapter (Figure 10) allows the Triton handpiece
to be driven by the MR7 pneumatic control unit. It
functions much the same way as the Triton port on
the Legend pneumatic control unit, except that it is
connec
rather than being integrated into the control unit
ted between the control unit and the gas source,
Figure 9: Motor Wrench
Figure 10: Triton Adapter
5
Page 10
System Components
PTSC224
Legend Attachments
Legend motor attachmentsareavailable in various designsto facilitatea variety of surgical procedures. Attachments vary in
length, diameter, and overall design. They are marked and color-coded to correspond with their associated dissecting tools.
A few of the Legend attachmentsavailableare listed in thetablebelow.
AttachmentExampleOther Details/Options
Standard Straight AttachmentsAS09
Standard Angled AttachmentsAA14
Straight Variable Exposure AttachmentsAVS07
Angled Variable Exposure AttachmentsAVA07
Fixed Footed AttachmentsAF01
Rotating Footed AttachmentsAF01R
Telescoping AttachmentsA
Contra-Angle AttachmentAC16
Metal Cutting AttachmentASMC
Perforator AttachmentAD01Available in 800 RPM or 1000 RPM form.
5/32” JacobsChuck AttachmentAD02
Bone Mill AttachmentBM100
T10 (base)
TT12A (tube)
The telescoping attachment requires the use of the AT10
attachment base, as well as a telescoping tube. Tubes are
available in straight, cur
ved, or hooded form.
NOTE: Angled and straightattachments with thesame length, mark
Curved and straight telescoping tubes with the same length, marking, and color band also share the same dissecting tool.
Example: The 14-AMstraightand 14-AMangled attachmentsare 14 cm long, marked 14-AMand have a green color band.
All dissecting tools with the pre x 14 (14MH30) may be used in either the 14-AMstraight or 14-AMangled attachment.
Besureto match thecolor codeand nomenclature on the Legend Dissecting Tool packaging with the color band and
nomenclature on the Legend Attachment.
ing, and color band share the same dissecting tool.
6
Page 11
Disposable Components
PTSC224
WARNING: Use only Medtronic Midas Rex Legend or
MR7 devices with an MR7 motor. Use of other devices
may cause injury or damage equipment, and will void
the manufacturer’s warranty.
Lubricant/Diffuser Cartridge
The lubricant/di user cartridge (Figure 11) provides
lubrication to the motor and lters oil from exhausted air.
Telescoping T ubes
Telescoping tubes (Figure 12) provide support to the
rotating dissectingtool. Telescopingtubesare disposable
following multiple usesand should be discarded when
heat or excessive vibration is noticed or insertion of toolsbecomes di cult.
Cleaning Brushes
Cleaningbrushes (Figure 13) are used to clean debris
from lumen of attachmentsand telescopingtubes. Sized
for an internal bore diameter of 3.2 mm, 2.4 mm or 1.2
mm in Legend Attachmentsand Telescoping Tubes.
System Components
Figure 11: Lubricant/Di user Cartridge
Figure 12: Telescoping Tube
NOTE: Cleaningbrushes will not passthr
contra-angle, metal cutting, perforator, or Jacobs Chuck
attachments, because they are not cannulated.
ough angled,
Figure 13: Cleaning Brushes
3.2mm
2.4mm
1.2mm
7
Page 12
System Components
PTSC224
Legend® Dissecting Tools
Legend dissectingtoolsaresterilecuttingtools, intended for cuttingboneand biomaterials.
Dissecting Tool Nomenclature
Part numbers for Legend dissectingtools follow astandard namingconvention, which is described in the diagram below.
A basic part number consists of theassociated attachment length, thetool head shape, and thetool head diameter. Part
numbers may also includea variety of pre xes to identify speci cattachmenttypes, as well asa variety of su xes to provide
additional information about the dissec
additional “-MN” su x.
ting tool. Tools that use a design taken from the Mednext line are designated by an
Associated
Attachment
Length
Tool
Head Diameter
(x.x millimeters)
Optional
Pre¿ x
Tool
Head Shape
Optional
Suf¿ x
Tool Number Pre¿ xes (not all inclusive)
F...For use with footed attachments
MCFor use with metal cuttingattachments
TFor use with telescopingattachments
Tool Head Shapes (not all inclusive)
ACAcornMHMatch Head
BABallOVOval
CYCylinderRTReverse Taper
HMHoleMakerTATapered
HSHole SawTDTwist Drill
Tool Number Suf¿ xes (notethat morethan one of thesu xes listed may be combined in a single part number)
LLongSSpiral
DDiamondSHShort
XExtraDCDiamond Coarse
FFineDXDiamond ExtraCoarse
CCarbideMNMednext Tool Design
8
Page 13
System Components
PTSC224
WARNINGS:
• Dissecting tools are for single-use only. Do not attempt to sterilize them. The dissecting tools are packed sterile and are
not intended for repeat use. To prevent contamination, use only once.
• Do not use an attachment and dissec
• Do not attempt to remove a tool while the motor is running.
• Do not attempt to remove a tool from an overheated motor or attachment.
• Do
not use the device if the package is opened or damaged.
tingtool combination that results in tool ail or excessive vibration.
General Guidelines For Attachment and Tool Applications
These are general guidelines for dissecting tool applications and are not an all-inclusive listing.
WARNING: Be sure to match the color code and nomenclature on the Legend Dissecting Tool packaging with the color
band and nomenclature on the Legend Attachment. Failure to do so could result in injur
sta .
Surgical
Application
Spine8-B, 9-M, 14-AM,
Commonly
Used
Attachments
Match Head
15-A
TelescopingMatch Head
Footed, StraightTap er ed
Elongated spherical design allowscontrolled, delicate
dissection. For entry hole, nerve decompression, osteophyte
removal, sinus dissection, etc.
Ball
Helical cutting utes dissectbone or cemente ec
from a wide variety of approach angles. For debridement,
decortication, sinus dissection, etc.
Oval
Helical cutting utes and curved design blend acorn and ball
styles to vary dissection e ciency with appr
decortication, laminotomy, entry hole, nerve decompression,
osteophyte removal, etc.
Hole Maker/Saw
Matched sets of Hole Makers and Hole Saws are e cient and
e ective for interbody fusion.
Cylinder
E ec
tive bone sculpting and planing. For graft shaping,
debridement, corpectomy, decortication, interbody fusion,
fusion takedown, etc.
Acorn
Curved design varies dissection e ciency with varied
approach angles. F
debridement, corpectomy, decortication, fusion takedown, etc.
Elongated spherical design allowscontrolled, delicate
dissection. For entr
removal, sinus dissection, etc.
Slender design for precise dissection with minimal bone loss.
For transection, osteotomy, graft harvesting, boneshaping,
entr
y hole, suture hole, midfaceadvancement, etc.
Commonly Used
Dissecting Tools
or entry hole, laminotomy, boneshaping,
y hole, nerve decompression, osteophyte
y to the patient or operating room
Suggested
Motor(s)
MR7, MR7
Touch
tively
oach angle. For
MR7, MR7
Touch
9
Page 14
System Components
PTSC224
Surgical
Application
Neurosurgical–
Cranial
General
Surgery and
Plastic Surgery
(Craniofacial/
Maxillofacial/
Sternotomy)
Ear, Nose, and
Throat (Otology,
Neurootology)
Commonly
Used
Attachments
7-6ST, 8-B, 9-M,
10-9ST, 14-AM,
15-A
TelescopingMatch Head
FootedTa pe re d
7-6ST, 8-B, 9-M,
10-9ST, 14-AM
7-6ST, 10-9STBall
Match Head
Elongated spherical design allowscontrolled, delicate
dissection. For entry hole, nerve decompression, osteophyte
removal, sinus dissection, etc.
Ball
Helical cutting ute
from a wide variety of approach angles. For debridement,
decortication, sinus dissection, etc.
Twist Drill
Helical design with stop producesa hole with a precise depth.
Ideal for plating.
A
corn
Curved design varies dissection e ciency with varied
approach angles. For entry hole, laminotomy, bone shaping,
debridement, corpectomy, decortication, fusion takedown, etc.
Elongated spherical design allo
dissection. For entry hole, nerve decompression, osteophyte
removal, sinus dissection, etc.
Slender design for precise dissection with minimal bone loss.
For transection, osteotomy, graft harvesting, boneshaping,
entr
y hole, suture hole, midfaceadvancement, etc.
Match Head
Elongated spherical design allo
dissection. For entry hole, nerve decompression, osteophyte
removal, sinus dissection, etc.
Ball
Helical cutting utes dissectbone or cemente ectively
from a wide variety of approach angles. F
decortication, sinus dissection, etc.
Tap er ed
Slender design for precise dissection with minimal bone loss.
For transection, osteotomy, graft harvesting, boneshaping,
entry hole, suture hole, midfaceadvancement, et
Twist Drill
Helical design with stop producesa hole with a precise depth.
Ideal for plating.
Helical cutting utes dissectbone or cemente ec
from a wide variety of approach angles. For debridement,
decortication, sinus dissection, etc.
Commonly Used
Dissecting Tools
s dissect bone or cement e ectively
wscontrolled, delicate
wscontrolled, delicate
or debridement,
c.
tively
Suggested
Motor(s)
MR7
MR7, MR7 Touch
MR7
10
Page 15
System Components
PTSC224
Surgical
Application
Orthopaedics8-B, 9-M,
Biometals/
Bioceramics/
Biomaterials
Commonly
Used
Attachments
14-AM, 21-TU,
26-R, Footed,
Telescoping
FootedTa pe re d
MCMetal Cutter
Commonly Used
Dissecting Tools
Ball
Helical cutting utes dissectbone or cemente ectively
from a wide variety of approach angles. For debridement,
decortication, sinus dissec
Tap er ed
Slender design for precise dissection with minimal bone loss.
For transection, osteotomy, graft harvesting, boneshaping,
entry hole, suture hole, midfaceadvancement, etc.
Acorn
Curved design varies dissec
approach angles. For entry hole, laminotomy, bone shaping,
debridement, corpectomy, decortication, fusion takedown, etc.
Cylinder
E ectivebonesculptingand planing. For graftshaping,
debridement, corpec
fusion takedown, etc.
Slender design for precise dissection with minimal bone loss.
For transection, osteotomy, graft harvesting, boneshaping,
entr
y hole, suture hole, midfaceadvancement, etc.
Cutting utes or diamond wheel design remove metals,
ceramics and other biomaterials e ec
of approach angles. For cutting rods, pins, plates, implants,
screws, etc.
tion, etc.
tion e ciency with varied
tomy, decortication, interbody fusion,
tively from a variety
Suggested
Motor(s)
MR7, MR7
Touch
MR7, MR7
Touch
11
Page 16
Setting up the OperatingRoom
PTSC224
Setting up the Operating Room
Power Source Requirements
Required Operating
(Dynamic) Pressure
80–120 psi100 psi12 cubic feet/min.Nitrogen or Dry-Filtered
5.5–8.3 bar6.9 bar340 liters/min.
CAUTION: Do not run the motor atan operating pressureabove or below the required operating pressure range. Operating
pressurebelo
may damage or reducethe life of the motor.
w 80 psi (5.5 bar) may not provide proper lubrication to the motor. Operating pressureabove 120 psi (8.3 bar)
Nominal Operating
(Dynamic) Pressure
Approximate Flow Rate
Required
Gas Type
Compressed Air
Pneumatic Connections
Figure 14: GasConnection Options
1. Regulator Hose (N2 DISS)
2. N2 DISS to Air DISS Adapter Gas Source
3. N2 DISS to Male Schrader Adapter Gas Source
4. Gas Source (N2 DISS)
5. Regulator
6. DISS/WF4 Adapter
7. Regulator
CAUTION: If you are usingtheMidasRex Safety Valve Regulator instead of the Legend Regulator, you must replace the
in-line oiler with the DISS/WF4 adapter before use.
12
Page 17
Setting up the MR7 System
PTSC224
Setting up theMR7 System
Installing the Oiler Cartridge
WARNING: Do not usetheMR7 system with theMidas
Rex in-line oiler. The MR7 motor is su ciently lubricated
by the lubricant/di user on the motor hose, and will be
over-lubricated if theMidasRex in-line oiler is used.
1. Set the non-running (static) pressure to 80–120 psi
(5.5–8.3 bar) atthegassource. Operating (dynamic)
pressure may beadjust
2. Hold the lubricant/di user cartridge perpendicular
to the housing (F
circular tting onto the housing’s circular receptacle
(Figure 16), breaking the foil seal.
3. Rotatethecartridge down until itclicks into plac
4. Verify that the
with the notch on the housing (Figure 17).
WARNINGS:
• Failure to properly secure the lubricant/di user
cartridge may cause injury to operator and/or
operating room sta .
• Do not attempt to remocartridge whilethesystem is pressurized.
symbol on thecartridge is lined up
ed later.
igure 15), and pressthecartridge’s
e.
vethe lubricant/di user
Figure 15: Aligningthe Lubricant/Di user Cartridge with the
Housing
Figure 16: PressingtheCartridge onto the Housing
CAUTIONS:
• Do not use an MR7 motor without a lubricant/
di user installed.
• Do not use a lubricant/di user cartridge for more than one hour of drill time.
• Do
• Do not attempt to re ll a used lubricant/di user
• Do not use a lubricant/di user cartridge if it appears
not re-use a lubricant/di user cartridge. It is a
single-use product.
cartridge.
to be damaged
, or if the inner foil seal is punctured.
Figure 17: Correctly I
nstalled Lubricant/Di user Cartridge
13
Page 18
Setting up theMR7 System
PTSC224
Connecting the Motor
Connect the motor hose to the motor port on the top of
the pneumatic control unit, by swinging the port cover
to the side and pressing the end of the hose into the port
(Figure18).
ARNING: Do not pinch, kink, obstruct, cut, tear, or step
W
on the motor/exhaust hose. This may cause the hose to
burst, potentially injuring the patient or user.
NOTES:
• If using the MR7 Touch motor, slidethecontr
on the pneumaticcontrol unitto the position
(Figure 19). This will automatically depressand lock
the foot pedal. Thecontrol will not lock into the
position unlessthe motor hose isconnected into the
motor port. When the motor hose is removed from
the motor port, the foot pedal will return to normal
position.
• If using the Triton Power Surgical Instrumentation
Syst
em in conjunction with the MR7 motor, use the
optional Triton adapter to connectthe Triton hose.
Refer to the documentation accompanying the
adapter for connection instructions.
ol slide
Figure 18: ConnectingtheMotor Hoseto theMotor Port
• The motor’s exhaust hose may havean oily lm
on theexternal sur
temperature di erentials followingsterilization.
Wipe the exhaust hose with a sterile cloth prior to
use. If motor continuesto have oil on theexhaust
hose, return the motor to MPSS f
WARNING: To avoid injury to the patient or user, do not
usethe pneumaticcontrol unitto operatesystems other
than theMR7, Legend
Prior to installation of a Legend attachmentand
dissectingtool, ensurethatthearrows on the motor
collet ats are aligned (Figure 20). If the arrows are not
aligned,
closest to the motor case until its arrow is aligned with
the arrow on the other collet at.
WARNING: To avoid injury when usingtheMR7 Touch
motor
installingtheattachmentand tool.
use the motor wrench to turn the collet at
, ensurethesafety slide is in the “O” position before
face from pressureand/or
or refurbishing.
, and Triton systems.
Figure 19: Finger/Foot Control Slide
Figure 20: Aligning the collet ats
14
Page 19
Installing an Attachment and Tool
PTSC224
Installing an Attachment and Tool
WARNING: Dissectingtool utesaresharp and may
perforatesurgical gloves. Tools may be grasped with a
hemostatto aid in installation and removal.
Straight Attachments
Installation:
1. Slide the attachment over the motor collet, aligning
the triangular markers (Figure 21). You will feel and
hear the attachment click into place when it is fully
seated.
2. Insert the dissecting tool into the attachment with a
slight rotational motion (F
hear the tool click into place when it is fully seated in
the attachment.
3. Turn the attachment to the
case (Figure 23). Gently pull on the shaft of the
dissecting tool to verify proper installation.
Removal:
Removal isthe reverse of installation.
igure 22). You will feel and
position on the motor
Figure 21: Slidingthe Attachment over theMotor Collet
Figure 22: Inserting the Dissecting Tool into the Attachment
Specialized Attachments
See Appendix A—Specialized Attachments for installation
and removal instructions for other attachments.
Figure 23: Attachment in the Locked Position
15
Page 20
ActivatingtheMotor
PTSC224
Activating the Motor
NOTE: In order to activate the MR7 Touch motor, thesafety slide on the nger control switch mustbe in the | position, and
the control slide on the foot control must be in the
motor isconnected to the motor port.
1. Activatethe motor by pressing on the footcontrol pedal (Figure 24), or by pressing on the nger control lever (MR7
Touch motor only).
2. Adjust operating pressure as needed at the compressed gas source until supply pressure gauge on pneumatic contr
unit reads within a range of 80–120 psi (5.5–8.3 bar) as required. Operating pressure (with motor running) will decreaseslightly from the non-running (static) pressuresetting when the motor isac
Foot Control Pedal
position. Thecontrol slide will not lock in the position unlessa
ol
tivated.
Figure 24: Foot Control Pedal
Air Pressure Gauge
Control Slide
Figure 25: MR7 Finger Control Switch
Finger Control Lever
WARNINGS:
• Do not use excessive force to pry or push bone with the attachment or tool during dissection. This could cause the tool
to break and cause injur
• Use adequate irrigation during dissection, to prevent thermal necrosis.
• MR7 motorsshould only be operated when theattachment is in the
y to the patient or operating room sta .
position.
NOTE: To decrease pressure, turn down the in-housecompressed gassource or loosen the pressure handle on the
regulator. Push down on the pressure relief atthe pneumaticcontrol unitto exhaustexcess pressure in the hoses. T
re-adjust pressureas needed.
16
hen
Page 21
Disassembling the MR7 System
PTSC224
Depressurize the System
1. Turn o the compressed gas at the source.
2. Press the pressure relief button on the pneumatic control unit, to release remaining gas.
DisassemblingtheMR7 System
WARNING: Do not disassembleequipmentbeforethegas is released fr
om control unit.
Disconnect Hoses
Release the motor hose from the control unit, by holding the hose rmly and pressing the motor port connection button.
Figure 26: Motor PortConnection and PressureRelief Buttons
Motor Port Connection Button
Pressure Relief Button
Discard the Lubricant/Diffuser Cartridge
Remove the lubricant/di user cartridge from the housing and discard it.
CAUTION: Do not re-usea lubricant/di user cartridge. It isasingle-use product.
Remove the Attachment and Tool
1. Follow the applicable removal instructions in the Installing an Attachment and Tool section of this manual to remove the
attachment and tool from the motor.
WARNING: Dissectingtool utesaresharp and may perforatesurgical gloves. Tools may be grasped with a hemostat to aid
in installation and removal.
2. Discard used dissecting tools in an appropriate container.
17
Page 22
Cleaning & SterilizingtheMR7 System
PTSC224
Cleaning & Sterilizing the MR7 System
MR7 Motor
Warningsand
Precautions
LimitationsVerify functionality prior to re-use.
INSTRUCTIONS
Point of UseNo particular requirements. Follow hospital procedures.
PackagingFor sterilization, place devices in instrumenttray. Devices may be unwrapped, or wrapped with up to two layers of 1-ply
Sterilization
(Temperatures are
minimum required,
timesare minimum
required)
Do not soak/submerge MR7 devices.
Do not use ultrasound to clean MR7 devices.
Do not use chlorine based or corrosive cleaning agents such as bleach, lye, acetone, sodium hypochlorite/bleach, sodium
h
ydroxide, formicacid, or solutionscontainingglutaraldehyde.
The use of a washer-disinfector for cleaning may cause a pre-mature degradation in performance.
Allow an adequate cooling period after st
Use ONLY nylon cleaningbrushes. Non-nylon cleaningbrushes leave residuethat may preventthetool from beingsecured
properly in the handpiece.
It is recommended that devicesare reprocessed assoon as is practical following use.
No particular requirements. F
Review the washer-disinfector warning above, before using this cleaning method. Remove devices from instrument trays
before placing into washer baskets. Orient devices following recommendations of the washer/disinfector manufacturers. Verify
that devices are visually clean after automated cleaning.
Recommended Washer Cycle
Pre-Wash: Cold tap water, 2 min.
Wash: 66oC, 5 min. usinga neutral enzymatic detergent, pH 6.0–8.0
Rinse: Hot tap water, 1 min.
enzymatic detergent, pH 6.0–8.0.
Brush motor case and collet with a nylon brush dampened with a neutral enz
be sure to brush under the nger control lever.
Rinse motor thoroughly under running water, collet end pointed down. Dry collet and motor with lint free towel.
Verify that device
polypropylene wrap
Steam Sterilization:
Cycle:
Temperature:
Time:
Drying:
sare visually clean after manual cleaning.
ollow hospital procedures.
es of the motor and hose, and wipe inner surface of oiler housing with acloth dampened with a neutral
Gravity
132°C
25 min.
15 minutes
eam sterilization.
Pre-vac
132°C
4 min.
15 minute
ymatic detergent. If usingtheMR7 Touch motor,
Pre-vac (FR/WHO)
134°C
18 min.
s
20 minutes
Pre-vac (UK)
134°C
3 min.
10 minutes
*Itemscontaminated with Transmissible Spongiform Encephalopathies (TSE) agents may be decontaminated usingsteam
autoclaveatatemperature of 134–137°C f
referenced in NHS Estates HTM 2010 parts 4 & 6: Appendix 2, ItemsContaminated With TSE Agentsand WHO Infection Control
Guidelines for Transmissible Spongiform Encephalopathies. Medtr
suspected or con rmed with a TSE diagnosisbe incinerated.
STERRAD Sterilization: Do not use low temperature hydrogen peroxidegas plasmasterilization dueto lumen internal diameter
and length restrictions.
100% EtO Sterilization Parameters:
Preconditioning: 51–59°C, 70 ±5% relative humidity, 30 min.
Temperature: 51–59°C
Relative humidity: 70 ±5%
Ethylene oxideconcentration: 725 ± 25mg/L
Gasexposuretime (full-cycle): 4 hours
Aeration: 18 hoursat 51–59°C
Steris: Do not use liquid perac
18
or asinglecycle of 18 minutes or repeated for atotal of six 3-minutecyclesas
onic recommendsthatall products used on a patient
etic acid sterilization due to immersion procedure.
Page 23
Cleaning & Sterilizing the MR7 System
PTSC224
Maintenance,
Inspection and Testing
StorageStore with other sterile devices.
Inspect devices for any damage before and after each use. If damage is observed, do not use the device until it is repaired.
After cleaningand sterilization, verify functionality prior to re-use.
NOTE: The instructions provided above have been validated by the manufacturer as being CAPABLE of preparingthe
product for re-use. It remainsthe responsibility of the processor to ensurethatthe reprocessingasactually performed,
usingequipment, materialsand personnel in the reprocessing facility, achievesthe desired result. This normally requires
validation and routine monitoring of the pr
Do not use chlorine based or corrosive cleaning agents such as bleach, lye, acetone, sodium hypochlorite/bleach, sodium
hydroxide, formicacid, or solutionscontainingglutaraldehyde.
The use of a washer-disinfector for cleaning may cause a pre-mature degradation in performance.
Allow an adequate cooling period after st
Use ONLY nylon cleaningbrushes. Non-nylon cleaningbrushes leave residuethat may preventthetool from beingsecured
properly in the handpiece.
It is recommended that devicesare reprocessed assoon as is practical following use.
No particular requirements. F
Review the washer-disinfector warning above, before using this cleaning method. Manually rinse attachments/tubes under
tap water
before placing into washer baskets. Orient devices following recommendations of the washer/disinfector manufacturers.
Recommended Washer Cycle
Pre-Wash: Cold tap water, 2 min.
Wash: 66°C, 5 min. usinga neutral enzymatic detergent, pH 6.0–8.0
Rinse: Hottap water, 1 min.
Immerse the head of Contra-Angle attachments in surgical instrument cleaning solution and run the motor for 1 minute.
Other attachmentsand tubes may be mechanically agitated in cleaningsolution, but notsoaked or immersed.
A nylon brush dampened with asurgical instrumentcleaningsolution may be used to clean theexternal sur
connectingsurfaces of theattachmentsand tubes.
Straightattachments, footed attachmentsand telescopingstraighttubes havespecial cleaningbrushessized to the
attachment’s or telescoping tube’s internal diamet
the attachment or telescoping tube from rear to front to loosen and remove debris trapped inside.
Move any moveable parts back and forth to allow solution to thoroughly clean attachment, e.g., sleeve on footed attachment,
perforator attachment.
Rinse thoroughly with tap water.
Thoroughly dry attachments. An air gun may be used to blow moisture out from rear to front of attachment.
Usingan aerosol spray lubricant (such as Pana Spray), per
polypropylene wrap.
, until no visiblesoil is noticed, before placingthem into theautomatic washer. Remove devices from instrumenttrays
ipe all attachments and telescoping tubes with a cloth, dampened with a surgical instrument cleaning solution.
Holdingthecan approximately 10–15 cm (3–6 in.) away from theattachment, spray all componentsthat move, rotate, or slide with three quick squirts.
Articulate movable components to ensure proper lubrication.
Remove excess lubricant with a clean cloth.
or sterilization, place devices in instrumenttray. Devices may be unwrapped, or wrapped with up to two layers of 1-ply
ollow hospital procedures.
eam sterilization.
facesand internal
er. Push thebrush wet with surgical instrumentcleaningsolution through
form the following steps to lubricateattachments:
19
Page 24
Cleaning & SterilizingtheMR7 System
PTSC224
Sterilization
(Temperatures are
minimum required,
timesare minimum
required)
Maintenance,
Inspection and Testing
StorageStore with other st
Steam Sterilization:
Cycle:
Temperature:
Time:
Drying:
*Itemscontaminated with TSE agents may be decontaminated usingsteam autoclaveatatemperature of 134–137°C for asinglecycle of 18 minutes or repeated for atotal of six 3-minutecyclesas referenced in NHS Estates HTM 2010 parts 4 & 6:
Appendix 2, Items Contaminated W
Encephalopathies. Medtronic recommendsthatall products used on a patientsuspected or con rmed with a Tbe incinerated.
STERRAD Sterilization: Do not use low temperature hydrogen peroxidegas plasmasterilization dueto lumen internal diameter
and length restrictions.
Steris: Do not use liquid peraceticacid sterilization dueto immersion pr
Inspect devices for any damage before and after each use. If damage is observed, do not use the device until it is repaired.
Verify functionality prior to re-use.
Gravity
132°C
25 min.
15 minutes
erile devices.
Pre-vac
132°C
4 min.
15 minutes
ith TSE Agents and WHO Infection Control Guidelines for Transmissible Spongiform
, 30 min.
Pre-vac (FR/WHO)*
134°C
18 min.
20 minutes
ocedure.
Pre-vac (UK)*
134°C
3 min.
10 minute
s
SE diagnosis
NOTE: The instructions provided above have been validated by the manufacturer as being CAPABLE of preparingthe
product for re-use. It remainsthe responsibility of the processor to ensurethatthe reprocessingasactually performed,
usingequipment, materialsand personnel in the reprocessing facility, achievesthe desired result. This normally requires
validation and routine monitoring of the process.
20
Page 25
Cleaning & Sterilizing the MR7 System
PTSC224
MR7 Pneumatic Control Unit / Regulator Hose / Triton Adapter /
Instrument Case
Warningsand
Precautions
LimitationsVerify functionality prior to re-use.
INSTRUCTIONS
Point of UseNo particular requirements. Follow hospital procedures.
Containmentand
Transportation
Preparation for
Decontamination
Cleaning:
Automated
(Do NOT use ultrasonic
washer)
Cleaning: ManualWipe the pneumaticcontrol unit, regulator hose, Triton adapter,
SterilizationDo notsterilize pneumaticcontrol unit, regulator hose, or Triton adapter.
Maintenance,
Inspection and Testing
StorageDo notstore with sterile devices.
Do not soak/submerge MR7 devices.
Do not use ultrasound to clean MR7 devices.
Do not use chlorine based or corrosive cleaning agents such as bleach, lye, acetone, sodium hypochlorite/bleach, sodium
hydroxide, formicacid, or solutionscontainingglutaraldehyde.
Do not sterilize the MR7 pneumatic control unit, regulator hose, or Triton adapter.
It is recommended that devicesare reprocessed assoon as is practical following use.
No particular requirements. F
Not validated.
instrumentcleaningsolution after each use.
Inspect devices for any damage before and after each use. If damage is observed, do not use the device until it is repaired.
Verify functionality prior to re-use.
ollow hospital procedures.
and instrument case with a cloth dampened with surgical
ollow hospital procedures.
NOTE: The instructions provided above have been validated by the manufacturer as being CAPABLE of preparingthe
product for re-use. It remainsthe responsibility of the processor to ensurethatthe repr
ocessingasactually performed,
usingequipment, materialsand personnel in the reprocessing facility, achievesthe desired result. This normally requires
validation and routine monitoring of the process.
Medtronic Powered Surgical Solutions (MPSS) will not authorize or accept the return of MPSS products that directly contact
patients or is contaminated with a patient’s body uids who is suspected or con rmed with a Transmissible Spongiform
Encephalopathies / Creutzfeldt-Jakob Disease (TSE/CJD) diagnosis. Futhermore, MPSS recommends that all MPSS products
used on a patient con rmed with a TSE diagnosis be incinerated. Contact your Sales Representative for replacement of
product incinerated under this policy or for temporary equipment while original equipment is quarantined. Contact MPSS
Regulatory A airs Department for additional inf
patient suspected of a TSE/CJD diagnosis must be incinerated.
If TSE/CJD is excluded as a diagnosis, the quarantined reusable equipment may be returned fcleaning, decontamination and sterilization. Hospital personnel should contacttheir infection control personnel for current
proceduresand policy for reusableequipment processing when suspect of contamination with Creutzfeldt-Jakob Disease
(CJD) or other Transmissible Spongiform Encephalopathies (TSE).
Reusable devices that have been used on patients with suspected Creutzfeldt-Jakob Disease (CJD
Spongiform Encephalopathies (TSE) should be quarantined and not reused until diagnosis is con rmed or excluded.
Reusable devices should be quarantined after having been cleaned, decontaminated, sterilized and packed in a ridged
sealed container until nal diagnosis.
ormation regarding TSE contamination. MPSS dissectingtools used on a
or useafter appropriate
) or other Transmissible
21
Page 26
Trou bleshooting
PTSC224
Troubleshooting
NOTE: All Legend and MR7 devices returned for servicing or refurbishing should be properly cleaned and sterilized prior to
shipping.
Motor not Running or Low on Power:
Possible CauseSolution
Hoses not properly connected.Makesureall connectionsaresecure.
Foot pedal on pneumatic control unit not functioning
pr
operly.
Motor stalls.Manually spin the dissecting tool, then activate the
according to compressed gas requirements, if necessary.
operly installed and locked onto the motor.Removeand re-install attachmentand tool to ensure
proper installation and locking of attachment onto
motor.
Check for obstructions under the foot pedal. If foot pedal
continues to fail, return the pneumatic control unit to
MPSS to be refurbished.
motor. If the motor continues to stall, return it to MPSS to
be refurbished.
Motor Continues to Run:
Possible CauseSolution
Pneumatic control unit is not functioning properly.Depressurize the system and return the pneumatic
control unit to MPSS to be refurbished.
Finger control is not functioning properly.Return motor to MPSS to be refurbished.
Pneumatic contr
ol unit is locked in the nger control position.Move the nger control lever to the foot control position.
System Makes an Abnormal Noise:
Possible CauseSolution
Inadequate lubrication.Check for proper installation of the lubricant/di user
cartridge. If the problem persists, replace the cartridge.
If replacing the cartridge doesn’t x the problem, return
the motor to MPSS f
Motor’s exhaust hose is damaged, or internal pressure hose is
detached.
Worn bearings.Switch attachmentsto det
Attachment not properly installed and locked onto the motor.Removeand reinstall attachmentand tool to ensure
Safety relief valve hasbeen activated by high air pressure.Ensure that air operating/dynamic air pressure is no
Depressurizethesystem and return the motor to MPSS
to be refurbished.
are failing in the motor or in the attachment. Return the
failingcomponentto MPSS to be refurbished.
proper installation and locking of theattachment onto
the motor.
higher than 120 psi.
or refurbishing.
ermine whether thebearings
22
Page 27
Trou bleshooting
PTSC224
Motor is Too Hot to Touch/Hold:
Possible CauseSolution
Inadequate cool down period following sterilization.Motor must be allowed to cool down following steam
sterilization.
Inadequate lubrication.Check for proper installation of the lubricant/di user
cartridge.
Attachmenttransferring heatto the motor
Heavy side loading during dissec
Inadequate irrigation.Ensureadequate irrigation to surgical site duringbone
tion.Discontinue useand restthe motor by using it
.Switch attachmentsto determine whether the heat is
being generated by the motor or the attachment. Return
the failing component to MPSS for refurbishing.
intermittently or wrap the motor with a moist sterile
towel. If overheating continues, return the motor to
MPSS for refurbishing.
dissection.
Attachment Will not Properly Seat on the Motor:
Possible CauseSolution
Motor collet ats are not aligned.Use the Legend motor wrench to rotate the at closest to
the motor case until its marker is aligned with the marker
on the at farthestaway fr
om the motor case.
Tool is Dif¿ cult to Remove from Attachment:
Possible CauseSolution
Aging of attachment.Return to MPSS to be refurbished, or purchase new
Improper cleaning.
Use of reprocessed tools.
Use of an unauthorized refurbisher.
equipment.
16-MF Contra-Angle Attachment is Overheating:
Possible CauseSolution
The Contra-Angle attachment operates by a set of internal
gears to engage the drive shaft. It is normal for some heat to
be generated approximately 2cm from the distal end of the
attachment and at the right of the angle head.
Verify pressuresetting of 80 psi (5.5 bar).
If
heat continues or isexcessive, return the attachment to
MPSS to be refurbished or purchase new equipment.
Perforator is Running too Slow:
Possible CauseSolution
Pressure set incorrectly.Check the pressure setting at the foot control.
23
Page 28
Trou bleshooting
PTSC224
Dissecting Tool Flails:
Possible CauseSolution
A non-Legend dissectingtool isbeing used.Replace with a Legend dissectingtool.
Worn attachment or tubebearings.Try another attachment or tube to isolate the location
Attachment/tubeand tool are notcompatible.Match color code on the dissectingtool packagingto the
Motor is damaged.Return motor to MPSS to be refurbished.
Dissecatcertain speeds.
Dissecting Tool Vibrates Excessively:
Possible CauseSolution
Dissectingtool’ssizeand geometry may create excessive
vibration atcertain speeds.
tingtool’s sizeand geometry may contributeto ailing
of the problem. If theattachment is failing, return itto
MPSS. I
f thetube is failing, dispose of itand usea new
tube.
color code on the attachment/tube.
Adjustthespeed by changingthe pressuresetting or
fo
ot/ nger control. Do not use if ailing persists. Change
dissectingtools.
Adjustthespeed by changingthe pressuresetting or
foot/ nger control. Change dissecting tools.
Dissecting Tool Will not Seat Properly in the Motor or Attachment Collet:
Possible CauseSolution
Debris in collet of attachment or motor.Clean the attachment or motor thoroughly according
to the instructions in this manual. If cleaning does not
correct the problem, return the attachment or motor to
MPSS to be refurbished.
A non-Legend dissec
ting tool is being used.Replace with a Legend dissecting tool.
Smoke is Generated by the Attachment or Motor:
Possible CauseSolution
Attachment is not in the locked position.Makesuretheattachment is in the locked position.
24
Page 29
Refurbishing or Repairs
PTSC224
Refurbishing or Repairs
When the MR7 System requires servicing or refurbishing, contact Medtronic Powered Surgical Solutions Repair Services for
a return authorization and instructions for returning the equipment. Medtronic Powered Surgical Solutions provides quality
assured service by factory-trained personnel who will utilize genuine Midas Rex Legend parts as required. All items being
returned for servicing or refurbishing should be properly cleaned and st
Peak performance, reliability and maximum service life from your MR7 System may beassured by using only those Midas Rex Legend products for your MR7 System thatare manufactured by and sold through Medtronic Powered Surgical
Solutions, Fort Worth, Texas. While Medtronic Powered Surgical Solutions guarantees complete compatibility among its
products within aspeci c product line, the dissectingtoolsare designed for single-use only, and Medtronic disclaimsany
responsibility when reprocessed dissectingtoolsare used. If you would like more information aboutthe patientand product
risks associated with reprocessed tools, please contact the number or e-mail address list
Due to safety and environmental concerns, Medtronic Powered Surgical Solutions requeststhe return of pneumatic high
speed motors for proper disposal at the end of the product life cycles.
erilized prior to shipping.
ed above.
25
Page 30
Preventative Maintenance
PTSC224
Preventative Maintenance
The Midas Rex MR7 System Preventive Maintenance Manual has been developed to assist you in getting the greatest
ownership value from your MR7 System, while helping to maximize its performance, safety and reliability. The scheduled
preventive maintenance/serv
preventive maintenance manual for thespeci csteps necessary to maintain theMR7 System.
ice program is in addition to the required routinecleaningafter each use. Please refer to the
26
Page 31
Limited Warranty
PTSC224
A. This Limited Warranty providesthe followingassuranceto the purchaser of aMedtronicMidasRex® MR7 Pneumatic
High Speed System. This Limited Warranty is extended only to the buyer purchasing the MR7 System directly from
Medtronic or from its a liate or its authorized distributor or representative. The Midas Rex® MR7 Pneumatic High
Speed System includes the motor, foot control, instrumentation cases and trays (hereafter referred to as Syst
Components), straight and angled motor attachments (hereinafter referred to as “Attachments”), telescoping tubes
(hereinafter referred to as Semi-reusable Components) and dissecting tools and other accessories not listed above
and jointly referred to as MR7 Pneumatic High Speed System, unless speci cally noted.
(1) Should a System Component fail to function to Medtronic’s published speci cations duringtheterm of
this Limited Warranty (one year fr
date of sale of a refurbished or used System Component), Medtr
Component or any portion thereof.
(2) Should an Attachment fail to function to Medtronic’s published speci cations duringtheterm of this
Limited Warranty (90 days from the date of sale of a new Attachment), Medtronic will either repair or
replacethe Attachment or any portion thereof.
(3) Should a Semi-reusable Component fail to function to Medtronic’s published speci cations during
the term of this Limited Warranty (30 days fr
Medtronic will replace the Semi-reusable Component or any portion thereof
(4) Should a Single Use Component fail to function to Medtronic’s published speci cations prior to its “use
by” date Medtronic will replace the Single Use Component.
B. To qualify for this Limited Warranty, these conditions must be met:
(1) The Product must be used on or before its “Use By” or “Use Before” date, if applicable.
(2) The Product mustbe used in accordance with its labelingand may notbealtered or subjected to misuse,
abuse, ac
(3) Medtronic mustbe noti ed in writing within thirty (30) days following discovery of a defect.
(4) The Product mustbe returned to Medtronic within thirty (30) days of Medtronic receiving noticeas
pr
(5) Upon examination of the Productby Medtronic, Medtronicshall have determined that: (i) the Product was
not repaired or altered by anyone other than Medtronic or itsauthorized representative, (ii) the P
was not operated under conditions other than normal use, and (iii) the prescribed periodic maintenance
and services, if applicable, have been performed on the Product
C. This Limited Warranty is limited to its express terms. THIS LIMITED WARRANTY IS IN LIEU OF ALL OTHER
WARRANTIES, EXPRESSED OR IMPLIED WHETHER STATUTORY OR OTHERWISE, INCLUDING ANY IMPLIED WARRANTY
OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. In no eventshall Medtronicbe liable for any
consequential, incidental
MR7 System, whether a claim for such damage is based upon the warranty, contract, negligence or otherwise.
D. The exclusions and limitations set out above are not intended to, and should not be construed so as to, contravene
mandatory provisions of applicable law. Users may bene t from statutory warranty rights under legislation
governing the sale of consumer goods. If any part or term of this Limited Warranty is held by any court of
competent jurisdiction to be illegal, unenforceable, or in con ict with applicable law,
portion of the Limited Warranty shall not be a ected, and all rights and obligations shall be construed and enforced
as if this Limited Warranty did not contain the particular part or term held to be invalid.
cident or improper handling.
ovided for in (3) above.
, prospective or other similar damage resulting from a defect, failure, or malfunction of the
om the date of sale of a new System Component or 90 days from the
onic will either repair or replacetheMotor
om the date of sale of a new Semi-reusableComponent),
.
the validity of the remaining
Limited Warranty
em
roduct
27
Page 32
Appendix A—Specialized Attachments
PTSC224
Appendix A—Specialized Attachments
Angled Attachments
Installation:
1. Slide the attachment over the motor collet, aligning
the triangular markers (Figure 27). You will feel and
hear the attachment click into place when it is fully
seated.
2. Turn the attachment to the
case (Figure 28).
3. With the “ tool lock” in the position, insertthe
dissectingtool into theattachment with aslight
rotational motion (Figure 29). You will feel and hear
thetool click into place when it is fully seated.
4. Turn the tool lock to the
pull on theshaft of the dissectingtool to verify
proper installation.
NOTE: A dissecting tool may be installed in the
attachment before the angled attachment is placed on
the motor.
CAUTION: Hold the handpiec
attachment, so that the attachment does not
inadvertently loosen from the handpiece.
position on the motor
position (Figure 30). Gently
eassembly by the
Figure 27: Slidingthe Attachment over theMotor Collet
Figure 28: Attachment in the Locked Position
Removal:
Removal is the reverse of installation.
Figure 29: Inserting the Dissecting Tool into the Attachment
Figure 30: Tool Lock in the Locked P
osition
28
Page 33
Appendix A—Specialized Attachments
PTSC224
Angled Double Lock Attachments
Installation:
1. Slide the attachment over the motor collet, aligning
the triangular markers.
2. Pull the attachment towards the motor and turn the
attachment to the
3. With the “ tool lock” in the position, insertthe
dissectingtool into theattachment with aslight
rotational motion. You will feel thetool click into
place when it is fully seated.
4. Turn the tool lock to the theshaft of the dissectingtool to verify proper
installation.
5. On Variable Exposure Attachments, use the TUBE adjustment ringto adjusttheexposure of the
dissectingtool. With thetool pointingaway fr
you, turn the ring to the right to increase the length
of thetube, thereby decreasingtheexposure of thetool. Turn the ringto the leftto decreasethe length
of thetube, thereby increasingtheexposure of thetool
.
position.
position. Gently pull on
om
Figure 31: Installing the attachment.
Figure 32: Inserting the Dissecting Tool into the Attachment
Removal:
1. To remove the attachment, hold the motor in the
palm of your hand, and push thesleeve on the
attachment distally while turning the attachment to
the
position.
2. Releasethesleeveand removetheattachment.
Figure 33: Removingthe Attachment
29
Page 34
Appendix A—Specialized Attachments
PTSC224
Curved Bur Attachments
WARNINGS:
• Use adequate irrigation and keep the cooling sleeve
soaked during dissection. Inadequate irrigation may
cause thermal necrosis.
• Do not modify the bur. Bending or prying may break
the bur, causing harm to the patient or operating
oom sta .
r
• Excessive pressure applied to the bur may cause bur
damage. If this occurs, useextremecareto ensurethatall fragments of thebur are remo
patient.
• Disposable devices are for single-use only. Do
notattemptto sterilize disposable devices. The
disposablesare packed sterileand are not intended
for repeat use. T
once.
• Test for bur wobble (eccentricity) at the desired speed prior touse. Selecta new bur or reducespeed
if wobble is observed prior to use or duringthe
procedure. Bur wobble may cause patient injur
CAUTIONS:
• Curved burs are not designed for variable tool
exposure. Do not attempt to adjust tool exposure, as
this may damage the device.
o prevent contamination, use only
ved from the
y.
Cooling the Bur
1. Prior to initial use, soak the cooling sleeve by dipping
it into acup of saline or DI water, asshown below.
Wetting of the cooling sleeve prior to cutting
2. During use, maintain copious irrigation of the cooling
sleeve and bur tip by dribbling saline or DI water
along the entire length of the cooling sleeve and bur
tip.
Operation
1. Activate the motor and gently press the curved bur
against the bone to begin dissection.
2. Use a light sweeping motion to continue removing
bone.
Removal
1. Turn both AT10 (V01) locking rings to the position
and pull the bur out of the attachment.
2. Discard the curved bur, according to hospital
procedures.
Reuse and Cleaning
Do not reuse.
• When operating or t
bur is properly inserted and locked into AT10
attachment.
esting the motor, ensure the
Installation
1. Turn the tube locking ring and the tool locking ring
on the AT10 (V01) Telescoping Attachment into the
position.
2. Slide the curved bur into the AT10 (V01), until the
hub is fully seated.
3. Gently press on the head of the dissecting tool to
lock it into place. You will hear aclick when thecurved bur is fully seated.
4. Turn both locking rings to the
AT10 (V01).
5. Gently pull on the curved bur to verify proper
installation.
6. Refer to your motor’s Instruction Manual for
information on setting up the AT10 (V01) attachmentand motor.
7. Brie y run the motor with the curved bur installed, checking for bur w
If either occurs, perform the corrective actions
described in theTroubleshootingsection of these
instructions.
obble or excessive vibration.
position on the
Sterility
Each bur isgamma-sterilized and is not intended for
repeated use. Do notattemptto re-sterilize.
Cutting Time
Continuouscutting for extended periods may causethe
deviceto heatto an uncomfortabletemperature. To avoid
this, limit cutting to the recommended times below:
Maximum
Continuous
Cutting Time
1 & 2 mm burs2 minutes12 minutes
3 & 4 mm burs3 minutes12 minutes
Maximum Total
Cutting Time
30
Page 35
Variable Exposure Attachments
PTSC224
CAUTION: The Legend Variable Exposure attachments
can be distinguished from standard attachments by
the dual color bands on the attachment. Match the
color band on the attachment to the color code on the
ting tool packaging.
dissec
WARNINGS:
• Surgeons should familiarize themselves with the
performance of dissectingtoolsbefore use, and
should explore the e ect of various levels of tool
exposure on dissection stability.
excessive chatter, vibration, or movement, decrease
the tool exposure.
• Dissecting tool size and geometry may contribute
to excessive vibration at certain speeds. I
decreasespeed by adjustingthe foot/ nger control,
or by changingthe operating pressure or console
speed setting. If necessary, use a di erent dissecting
tool.
If the tool exhibits
ncrease or
Appendix A—Specialized Attachments
Figure 34: Adjusting Tool Exposure
Installation:
1. Refer to theappropriatesection of this manual on
installing xed or angled attachments.
2. After installation, usethe TUBE adjustment ringto
adjusttheexposure of the dissectingtool (Figure 34).
With the tool pointing away fr
ring to the right to increase the length of the tube, thereby decreasingtheexposure of thetool. Turn the
ring to the left to decrease the length of the tube, thereby increasingtheexposure of thetool
WARNING: Do not usethe Variable Exposure
Attachment if the TUBE adjustment ringspins freely
or failsto click into placexposure may change without warning.
CAUTION: Make sure that the tool lock (angled
attachments only) and the attachment lock are still in the
position after each adjustment of thetool exposure.
Attemptingto increasetheexposuretoo far may result
in theattachmentbecoming unlocked. Accidentally
turningthetool lock may result in reduced speed and/or
overheating of theattachment.
WARNING: Do not usetheend of thetubeasa depth
gauge or depth stop.
e with each adjustment, as the
om you, turn the
.
Removal:
1. Removal isthe reverse of installation.
2. When cleaning, clean the attachment completely,
rst without adjusting the tube length, then with the tube fully extended, and with thetube fully retrac
ted.
31
Page 36
Appendix A—Specialized Attachments
PTSC224
Footed Attachments
Installation:
1. Insert the dissecting tool into the motor collet with a
slight rotational motion (Figure 35). You will feel and
hear thetool click into place when it is fully seated.
2. Slide the footed attachment over the dissectingtool
, onto the motor, aligningthetriangular markers
(Figure 36).
3. Pull the footed attachment towards the motor and turn theattachmentto the
Removal:
1. To remove the Legend footed attachment, hold
the motor in the palm of your hand, and push the
sleeve on the attachment distally while turning the
attachment to the position (Figure 38).
WARNING: Remove Legend Footed Attachments
cautiously and slowly per instructions to avoid injury to
the operator.
2. Releasethesleeveand removetheattachment.
3. Pull the dissecting tool out of the motor collet and
discard
the tool.
position (Figure 37).
Figure 35: Inserting the Tool into the Collet
Figure 36: Slidingthe Attachment onto theMotor
Figure 37: Attachment in the Locked Position
32
Figure 38: Removingthe Attachment
Page 37
Appendix A—Specialized Attachments
PTSC224
Rotating Footed Attachments
NOTE: Rotating and xed footed attachments with the
same length, marking, and color band share the same
dissecting tools.
Installation:
1. Insert the dissecting tool into the motor collet with a
slight rotational motion (Figure 39). You will feel and
hear thetool click into place when it is fully seated.
2. Slide the attachment over the dissecting tool and
onto the motor
(Figure 40). You will feel and hear theattachmentclick into place when it is fully seated.
3. Turn the attachment to the
NOTE: The footed end of theattachment now has 360˚
of unrestricted rotation.
Removal:
Removal isthe reverse of installation.
, aligningthetriangular markers
position (Figure 41).
Figure 39: Inserting the Tool into the Collet
Figure 40: Slidingthe Attachment onto theMotor
Figure 41: Attachment in the Locked Position
33
Page 38
Appendix A—Specialized Attachments
PTSC224
Contra-Angle Attachment (16-MF)
CAUTION: Do not run the 16-MF attachment with
operating pressureabove 80 psi (5.5 bar). This may
cause over-heating and damage to internal gears of
attachment.
Installation:
1. Decrease the pressure at the compressed gas source
to 80 psi (5.5 bar).
2. Adjust the pressure as needed by lowering it at the
gas source, then push down the pressure relief on
the pneumatic contr
pressure in the hoses.
3. Slide the attachment over the motor collet, aligning thetriangular markers (Figure 42). You will feel and
hear theattachmentclick into placseated.
4. Turn the attachment to the position (Figure 43).
5. Turn the attachment head lever to the open position
(Figure 44).
6. Insert the dissecting tool and return the lever to the
closed position. Gently pull on the dissecting tool
shaft to verify pr
ol unit to exhaust the excess
e when it is fully
oper installation.
Figure 42: Slidingthe Attachment over theMotor Collet
Figure 43: Attachment in the Locked Position
Removal:
1. Removal isthe reverse of installation.
2. Discard the dissecting tool after removing it from the attachment.
NOTE: A dissecting tool may be installed and locked in
the attachment before the Contra-Angle attachment is
installed on the motor
.
Figure 44: Openingthe Attachment Head Lever
34
Page 39
Metal Cutting Attachments
PTSC224
WARNING: For metal transection, observe the following
safety guidelines:
• Wear eye protection.
• Irrigate well to cool thecuttingsurfaces.
• Protectthe wound site from metal debris.
Appendix A—Specialized Attachments
Figure 45: Slidingthe Attachment over theMotor Collet
• Use a clamp or grasping device to contr
fragments duringtransection of any metal
component.
The Metal Cutting Attachment uses the tungsten carbide
or diamond wheel dissectingtools.
ol loose
Installation:
1. Slide the attachment over the motor collet aligning
the triangular markers (Figure 45). You will feel and
hear the attachment click into place when it is fully
seated.
2. Turn the attachment to the
3. Turn the tool lock towards the icon, and insertthe dissectingtool into theattachment with aslight
rotational motion (Figure 46). You will feel and hear
thetool click into place when it is fully seated.
4. Turn the tool lock to the
pull on theshaft of the dissectingtool to verify
proper installation.
NOTE: A dissecting tool may be installed and locked in
the attachment before the metal cutting attachment is
installed on the motor
.
position.
position (Figure 47). Gently
Figure 46: Insertingthe Tool into the Attachment
Removal:
Removal is the reverse of installation.
WARNING: Do not use metal cutting dissectingtools on
bone.
Figure 47: Turning the T
ool Lock to the Locked Position
35
Page 40
Appendix A—Specialized Attachments
PTSC224
Telescoping Attachments
Installation:
1. Slide the attachment over the motor collet aligning
the triangular markers (Figure 48). You will feel and
hear the attachment click into place when it is fully
seated.
NOTE: A dissecting tool may be installed and locked in
the attachment befo
installed on the motor.
re the telescoping attachment is
F
igure 48: Slidingthe Attachment over theMotor Collet
2. Turn the attachment to the
3. Turn the TUBE locking ring towards the icon.
4. Insert the base of the telescoping tube into the attachment (Figure 49).
5. To lock the tube in place, turn the TUBE locking ring towardsthe
6. To insert the tool, make sure that the TOOL locking
ring is in the tool into thetop of thetube (Figure 50). You will
feel and hear thetool click into place when it is fully
seated.
7. Turn the TOOL locking ring to the
pull on theshaft of the dissectingtool to verify
proper installation.
8. If the tube position needs to be changed, turn the
TUBE locking ringtowardsthetube, then turn the TUBE locking ringtowardsthe
icon.
icon. Do not over tighten.
position, and insertthe dissecting
position.
position. Gently
icon, re-position the
Removal:
1. To remove the attachment, unlock both locking rings,
and pull the telescoping tube and tool out of the
attachment.
2. Turn the attachment to the
from the motor.
position and remove it
Figure 49: Inserting the Telescoping Tube
Figure 50: Inserting the Dissecting Tool
NOTE: Telescoping tubes are disposable following
multiple uses, and should be discarded when heat or
excessive vibration is noticed, or when insertion of the
tool becomes di cult.
36
Figure 51: Locking Rings
TUBE Locking Ring
TOOL Locking Ring
Page 41
Appendix A—Specialized Attachments
PTSC224
Perforator Driver Attachments
The Perforator Attachment has a Hudson chuck to drive
any device with a Hudson shank, i.e., cranial perforator
device.
NOTE: MPSS does not provide cranial perforator devices.
Installation:
1. Slide the attachment over the motor collet aligning
the triangular markers (Figure 52). You will feel and
hear the attachment click into place when it is fully
seated.
2. Turn the attachment to the
3. To install a cranial perforator device with a Hudson shank, pull back on thecollar of theattachment
(Figure 54).
4. Insert the device and release the collar.
NOTE: A cranial perforator device may be installed in
the attachment befo
motor.
re the attachment is installed on the
position (Figure 53).
Figure 52: Slidingthe Attachment over theMotor Collet
Figure 53: Attachment in the Locked Position
Removal:
1. To remove the cranial perforator device, pull back on
the collar of the attachment.
2. To remove the attachment, turn it to the position
and slide it o of the motor.
WARNING: Consultthecranial perforator device labeling
for the recommended speed speci cations.
Maximum Speed of Perforator Attachments
Gas Pressure
(Dynamic)
80 psi660 rpm850 rpm
100 psi820 rpm1050 rpm
120 psi890 rpm1140 rpm
Model AD01
Output Speed
(Max.)
Model AD03
Output Speed
(Max.)
Figure 54: Pulling Back on theCollar
37
Page 42
Appendix A—Specialized Attachments
PTSC224
Jacobs® Chuck Attachments
The Jacobs Chuck attachment is a non-cannulated 5/32”
chuck with key for drilling.
Installation:
1. Slide the attachment over the motor collet aligning
the triangular markers (Figure 55). You will feel and
hear the attachment click into place when it is fully
seated.
2. Turn the attachment to the
3. To install a drill bit, turn the Jacobs key to open the collar.
4. Insert the drill bit and tighten the collar.
NOTE: A drill bit may be installed in theattachment
before the Jacobs Chuck attachment is installed on the
motor
.
Removal:
Removal isthe reverse of installation.
position (Figure 56).
Figure 55: Slidingthe Attachment over theMotor Collet
Figure 56: Attachment in the Locked Position
38
Page 43
Bone Mill Attachment
PTSC224
The Legend Bone Mill attachment is composed of
a non sterile base and gamma-sterilized, single use
disposablecomponents. Thegamma-sterilized single
use disposablecomponentsconsist of onebowl, onecap
OW and HIGH settings), and onespatula. A Legend
(with L
or MR7 motor providesthe power to drivethe BoneMill
attachment. The Bone Mill attachment is intended to mill
bone into particles 1 to 5mm in size.
F
igure 57: Bone Mill Attachment
1. Cap
2. Bowl
3. Bone Mill Base
4. Motor
5. Spatula
Appendix A—Specialized Attachments
1
2
3
WARNINGS:
• Do not operate the Bone Mill attachment without
the bow
• Do not come in contact with the interior of the
disposablebowl and cap duringbone milling.
• Do not use the Bone Mill disposable components f
morethan onesurgical procedure, asthis may causecross-contamination and aect patientsafety.
• Use only Legend/MR7 motors and Legend Bone
Mill disposables in combination with the Bone
Mill attachment. Use of other device
manufacturer’s warranty.
l, and cap secured in place.
s will void the
Using the Bone Mill Attachment
1. Place the bowl onto Bone Mill base and secure the
latches.
2. Lay harvested bone specimens into the bowl.
CAUTIONS:
• The addition of any liquid to the milling process
should be administered in small volumes to prevent
leakage.
or
5
4
Figure 58: Bone Mill Cap
HIGH
LOW
Figure 59: InsertingtheMotor
• Do
• Do not use Bone Mill disposables to process more
3. Secure the cap onto the bowl by rotating
4. Use the HIGH cap position when starting with large
5. Insert the motor into the motor connection on the
6. Activatethe motor to mill bone. Continueto run the
not use Bone Mill disposables to process more
than 6 bone pieces approximately 1.5 cc in size at
one time.
than a total of 40 cc of bone during any one surgical
pr
ocedure.
counterclockwise. Do notactivatethe BoneMill
Attachment withoutsecuringthecap.
piec
es of bone or large batch amounts. Use the LOW
cap position to nish milling a batch or when starting
with small pieces or batch amounts (Figure 58).
Bone Mill base and r
(Figure 59).
Bone Mill Attachment until desired bone particle size
isachieved. Adjust motor speed to obtain desired
particlesize.
otatethe motor to the position
39
Page 44
Appendix A—Specialized Attachments
PTSC224
NOTE: If bone does not process, deactivatethe motor,
rotate the motor to the position, and remove the cap.
Reinsert the cap in the HIGH position and activate the
motor. If bone spins without milling or jams, divide bone
chunks into two batches and process separately.
7. Process the bone fto gauge quality of millingbeforecontinuing.
8. Continueto processbone in 5–15 second batches
until the majority of particles reach desired size.
9. To remove bone particles stop operation of the
motor and r
cap by rotating it clockwise. Use the spatula inside
the bowl and cap to remove bone particles.
NOTES:
• Millingtime is dependent on the density of the
patient’sbone, the ratio of cancellousand cortical
bone, the size of the individual particles and the
batch amount of the bone to be pr
check bone after 10–15 seconds to gauge quality of
milling.
• I
f the blade becomes inoperable, stop operation of
the motor and rotate the motor to the
Remove the cap and use the spatula to free bone
and/or blades.
• After removingbone particles from thebowl, the
milling process may be repeated with additional
bone specimens, as needed
bone may be processed by the Bone Mill Disposable
during any one surgical procedure.)
or 5–15 secondsand visually check
otate itto the position. Removethe
ocessed. Always
position.
. (Up to atotal of 40 cc of
After Use
1. Rotatethe motor to the position.
2. Unlatch the BoneMill disposables from thebase.
3. Discard the bowl, cap, and spatula per hospital
procedure.
4. Clean the BoneMill baseaccordingto the directionsbelow.
5. Clean the motor accordingto its instruction manual.
Bone Mill bowl, cap, and spatula are single use
devices. Opened but unused BoneMill disposables
may notbe re-sterilized under any sterilization
method.
Preventative Maintenance
The Bone Mill Attachment should be returned for factory
maintenanceannually (approximately 100 procedures).
If the Bone Mill Attachment does not function according to these instructions, call your MedtronicRegional
Distributor or Medtr
If necessary, return the Bone Mill Attachment base to
your Distributor or SalesRepresentativeto be repaired.
onic Neurologic SalesRepresentative.
40
Page 45
PTSC224
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.