Medtronic MC1AVR1 Reference Guide

Micra™ AV MC1AVR1
MR Conditional dual chamber transcatheter pacing system with SureScan™ technology (VDD)
Reference Manual Caution: Federal law (USA) restricts this device to sale by or on the order of a physician.
Micra™ AV MC1AVR1
Reference Manual
This manual describes the operation and intended use of the features of the Micra AV Model MC1AVR1 MR Conditional dual chamber transcatheter pacing system with SureScan™ technology (VDD).
Capture Management, CareLink, CareLink Encore, Medtronic, Medtronic CareLink, Micra, Quick Look, SureScan
Medtronic
Micra™ AV MC1AVR1
Contents
1 Introduction .......................................................... 6
1.1 About this manual .................................................... 6
2 Patient follow-up guidelines ........................................... 8
2.1 Follow-up appointments .............................................. 8
2.2 Optimizing device longevity ........................................... 10
3 Diagnostic data features ............................................. 12
3.1 Quick Look II summary data .......................................... 12
3.2 Rate histograms .................................................... 13
3.3 Battery and device performance data .................................. 14
4 Configuring pacing therapies ......................................... 19
4.1 Sensing ........................................................... 19
4.2 Single chamber pacing therapies ..................................... 31
4.3 Dual chamber modes ............................................... 33
4.4 Rate-responsive pacing ............................................. 43
4.5 Pacing with Capture Management .................................... 50
4.6 Rate Hysteresis .................................................... 57
Glossary ................................................................. 59
Index .................................................................... 63
Reference Manual 5
Medtronic
Micra™ AV MC1AVR1

1 Introduction

1.1 About this manual

This manual describes the operation and intended use of the features of the Micra AV Model MC1AVR1 MR Conditional dual chamber transcatheter pacing system with SureScan™ technology (VDD).
Throughout this manual, the word “device” refers to the implanted Micra AV device.
The report images, labels, and navigation instructions in this manual apply to the Medtronic Model SW044 software running on a Medtronic Model 2090 Programmer or a Medtronic Encore Model 29901 Programmer. The details of the user interface are provided for reference only and may not match those of other applications.
Navigation paths to software screens or programmable parameters are shown with a > character between each step in the path (for example, Params > Capture Management… > Amplitude Safety Margin).

1.1.1 Product literature

Before implanting the device, it is recommended that you take the following actions:
Read the product literature for information about prescribing, implanting, and using the device and for conducting a patient follow-up session.
Thoroughly read the technical manuals for the other system components.
Discuss the device and implant procedure with the patient and any other interested parties, and give them any patient information materials packaged with the device.
The following manuals and documents contain additional information about the programmer and implanted device:
Device manual – This manual provides summaries of device features, indications and contraindications, warnings and precautions, instructions for implanting the device, quick reference specifications, parameter tables, and an explanation of package symbols.
MRI technical manual – This manual provides MRI-specific procedures and warnings and precautions.
Programming guide – This manual explains how to use the programmer software to conduct a patient session.
Programmer reference manual – This manual provides a detailed description of the Medtronic programmer. There is a programmer reference manual for the Medtronic
6 Reference Manual
Medtronic
CareLink Model 2090 programmer and a programmer reference manual for the Medtronic CareLink Encore 29901 programmer.
Radio regulatory compliance information – This document provides compliance information related to the radio components of the device.
Micra™ AV MC1AVR1

1.1.2 Technical support

Medtronic employs highly trained representatives and engineers located throughout the world to serve you and, upon request, to provide training to qualified hospital personnel in the use of Medtronic products.
In addition, Medtronic maintains a professional staff of consultants to provide technical consultation to product users.
For more information, contact your local Medtronic representative, or call or write Medtronic at the appropriate address or telephone number listed on the back cover.
Reference Manual 7
Medtronic
Micra™ AV MC1AVR1

2 Patient follow-up guidelines

2.1 Follow-up appointments

Schedule regular patient follow-up sessions during the service life of the device. The first follow-up session should occur within 72 hours of implant so that the patient can be checked for device dislodgment, wound healing, and postoperative complications.
During the first few months after implant, the patient may require close monitoring. Schedule follow-up sessions at least every 3 months to monitor the condition of the patient and the device and to verify that the device is configured appropriately for the patient.

2.1.1 Follow-up process

The process for conducting a follow-up evaluation includes the following steps:
1. Review the patient’s presenting rhythm, including an ECG.
2. Verify the status of the implanted device.
3. Verify the clinical effectiveness of the implanted device.
4. During a follow-up appointment, adjust device parameters as necessary.

2.1.2 Reviewing the presenting rhythm

The presenting rhythm may indicate the presence of undersensing and oversensing or ventricular loss of capture. These are pacing issues that can affect the delivery of pacing therapy. These issues can often be resolved by making programming changes.
Review the presenting rhythm by viewing the Live Rhythm Monitor and recording the EGM, ECG, and marker annotations. If you identify issues with the patient’s presenting rhythm, review the device settings and reprogram the device to values that are appropriate for the patient.

2.1.3 Verifying the status of the implanted device

Perform the following tasks to verify that the device is functioning correctly:
Assess the battery status.
Review the device status information on the Quick Look II screen.
Review any Observations on the Quick Look II screen.
8 Reference Manual
Medtronic
Micra™ AV MC1AVR1
2.1.3.1 Assessing the battery status
To assess the status of the device battery, review the Remaining Longevity estimate on the Quick Look II screen. If the device battery has reached a replacement threshold, the associated indicator is displayed. For more detail about the battery status, including battery voltage, review the Battery and Device Measurements data: tap the next to the Remaining Longevity field on the Quick Look II screen.
Warning: When the battery voltage reaches the End of Service (EOS) condition, the device permanently deactivates pacing and sensing and switches to the Device Off mode. The EOS symbol appears on the programmer screen.
If the Recommended Replacement Time (RRT) indicator or Elective Replacement Indicator (ERI) is displayed or if the battery voltage is at or below the displayed RRT voltage, contact your Medtronic representative and schedule an appointment with the patient to implant a new device. For more information about the replacement indicators, see the Micra AV SW044 Programming Guide.
2.1.3.2 Assessing the performance of the device
Follow-up appointment – During a follow-up appointment, you can check the performance
of the device by reviewing the electrode impedance, capture threshold, and sensing trends on the Quick Look II screen. To view this information, tap Data > Quick Look II. For a more detailed history of each measurement, tap the next to the appropriate trend graph. For more information about the automatic collection of these trends, see Section 3.3, “Battery and device performance data”, page 14.
If you also want to gather real-time information about the performance of the device, you can perform the following tests:
Impedance test: Compare the results of the test to previous electrode impedance measurements to see if there have been significant changes since the last follow-up session.
Threshold test: Perform this test to check the patient’s Capture Management thresholds.
Sensing test: Compare the test results to previous R-wave amplitude measurements.
For more information about performing these tests, refer to the Micra AV SW044 Programming Guide.

2.1.4 Verifying the clinical effectiveness of the implanted device

You can use the information available from the Quick Look II screen and in printed reports to assess whether the device is providing adequate clinical support for the patient.
Reference Manual 9
Medtronic
Micra™ AV MC1AVR1
2.1.4.1 How to assess effective pacing therapy
1. Interview the patient to confirm that the patient is receiving adequate cardiac support for daily living activities.
2. Review the pacing percentages on the Quick Look II screen and the Data – Rate Histograms window. Print a Rate Histograms Report. You can use the data on the rate histogram window and in the report to assess the patient’s pacing and sensing history. For more information about rate histograms, see Section 3.2, “Rate histograms”, page 13.
3. You can use the data on the A4 Amplitude Trend to assess changes in atrial sensing over time. Review the A4 Amplitude Trend data. Tap Data > Diagnostics > A4 Amplitude Trend > Open Data.

2.1.5 Adjusting device parameters

Adjust the pacing and diagnostic data parameters as needed to address any issues identified during the follow-up appointment.
Note: Use caution when reprogramming the sensing parameters to ensure that appropriate sensing is maintained. For more information, see Section 4.1, “Sensing”, page 19.

2.2 Optimizing device longevity

Optimizing device longevity is a desirable goal because it may reduce the frequency of device replacement for patients. Optimizing device longevity requires balancing the benefit of device therapy and diagnostic features with the energy requirements placed on the battery for the operation of these features.
To view the Remaining Longevity estimate for the device, refer to the Quick Look II screen. For information about the longevity of the device, see the Micra AV MC1AVR1 Device Manual.
The following sections describe strategies that can help to reduce the energy requirements placed on the battery.

2.2.1 Managing pacing outputs

Capture Management – The Capture Management feature provides the device with
automatic monitoring and follow-up capabilities for managing pacing thresholds in the right ventricle. This feature is designed to monitor the pacing threshold and, optionally, to adjust the pacing outputs to maintain capture. Programming the Capture Management feature to the Adaptive mode allows the device to set the pacing amplitude just high enough to maintain
10 Reference Manual
Medtronic
capture while preserving battery energy. For more information, see Section 4.5, “Pacing with Capture Management”, page 50.
Manual optimization of amplitude and pulse width – If you choose to program the Capture Management feature to Monitor or Off, you can optimize the patient’s pacing output parameters manually. Perform a pacing threshold test to determine the patient’s pacing threshold. Select amplitude and pulse width settings that provide an adequate safety margin above the patient’s pacing threshold. An adequate safety margin decreases the pacing outputs and conserves battery energy. Refer to the Micra AV SW044 Programming Guide for more information about performing a pacing threshold test.
Pacing rate – The more paced events that are delivered, the more battery longevity is reduced. Make sure that you have not programmed an unnecessarily high pacing rate for the patient. Carefully consider using features that increase the bradycardia pacing rate. Use features such as Rate Response only for patients who can receive therapeutic benefit from the feature.
Micra™ AV MC1AVR1

2.2.2 Other longevity considerations

Atrial mechanical sensing – Atrial mechanical sensing (the VDD, VDI, or ODO pacing
modes) decreases battery longevity. The AV Conduction Mode Switch feature disables atrial mechanical sensing during periods of intact AV conduction, reducing the longevity impact. Programming the sum of 3 vectors for atrial sensing (A. Sensing Vector) has a minor negative impact on battery longevity.
Holter telemetry – Extended use of the Holter telemetry feature substantially decreases the battery longevity. The Holter telemetry feature continues to transmit EGM and marker annotations for the programmed time duration, regardless of whether the programming head is positioned over the device.
Note: Do not program the Holter telemetry feature to On unless instructed to do so by a Medtronic representative. Use of this feature requires that the patient is equipped with a customized Holter monitor provided by Medtronic for monitoring EGM.
Reference Manual 11
Medtronic
Micra™ AV MC1AVR1

3 Diagnostic data features

3.1 Quick Look II summary data

At the start of a patient session, it is useful to quickly view summary information about device operation and the patient’s condition. This overview can help you to determine whether you need to look more closely at diagnostic data or reprogram the device to optimize therapy for the patient.
The Quick Look II data summarizes the most important indicators of system operation and the patient’s condition. These indicators include battery and device status, pacing therapy information, and system-defined observations.
You can view Quick Look II data on the Quick Look II screen, which is displayed on the programmer at the beginning of a patient session. To return to the Quick Look II screen from another screen, tap Data > Quick Look II. For more information about using the Quick Look II screen, refer to the Micra AV SW044 Programming Guide.

3.1.1 Quick Look II battery and device status information

The Quick Look II data includes the following information about the battery and device status:
Estimate of remaining battery longevity
Trends of the weekly average impedance, capture threshold, and R-wave amplitude measurements
Most recent measured values for impedance, capture threshold, and R-wave amplitude

3.1.2 Quick Look II pacing therapy information

The Quick Look II data includes the following information about pacing therapy:
Programmed values for the Mode, Lower Rate, and Upper Tracking Rate parameters
Percentage of time (% of Time) spent in sensing or pacing since the last patient session
Percentage of time spent in AV Conduction Mode Switch, if applicable
12 Reference Manual
Medtronic
Micra™ AV MC1AVR1

3.1.3 Quick Look II Observations

Observations are based on an analysis of programmed parameters and data collected since the last session. The following types of observations can occur:
Device status observations inform you of conditions that affect device operation and require attention. Examples of such conditions include Recommended Replacement Time (RRT) or the occurrence of a device reset.
Electrode status observations report any potential issues with the sensing integrity of the electrodes and abnormal Capture Management results. These observations can also warn you about possible inconsistencies in the performance of the device.
Diagnostic data observations report noteworthy events, such as conditions that prevent diagnostic data from being collected effectively.
Clinical status observations alert you to abnormal patient conditions, such as high pacing thresholds and a potential lack of effective VDD therapy.
On the Quick Look II screen, if you select one of the displayed observations and more information about the selected observation is available, the becomes active. You can use the to look at the relevant details.

3.2 Rate histograms

Information about heart rates recorded between patient sessions can help you to monitor a patient’s condition to assess the effectiveness of therapies. Rate histograms show the distribution of atrial mechanical sensed-ventricular paced (AM-VP) events, atrial mechanical sensed-ventricular sensed (AM-VS) events, ventricular sensed (VS) only events, and ventricular paced (VP) only events.
To access the rate histograms data, select Data > Diagnostics > Rate Histograms > [Open Data].

3.2.1 Information provided by rate histograms

Rate histograms report the atrial and ventricular event data stored by the device. There are histograms for 2 types of heart rate data: ventricular rate and atrial ventricular (AV) rate. The histograms include data from the current collection period. Data storage for rate histograms is automatic; no setup is required.
The rate histogram shows the percentage of time that the device was pacing and sensing within rate ranges. There are 20 rate ranges that are each 10 bpm in length. Rates slower than 40 bpm are included in the < to 40 range; rates faster than 220 bpm are included in the 220 to > range.
Reference Manual 13
Medtronic
Micra™ AV MC1AVR1
% of Time – This section shows the percentage of the total time that the device paced or sensed during the collection period. If atrial data was collected during this time, the device will also show the percent of events that were preceded by an atrial mechanical sense (AM-VP and AM-VS).
Atrial ventricular rate histogram – The atrial ventricular rate histogram shows the rate distribution of ventricular events, both ventricular sense (VS) and ventricular pace (VP), denoting which ventricular events were preceded by atrial mechanical sensing events (AM-VS and AM-VP). This rate histogram appears only when atrial data has been collected.
Ventricular rate histogram – The ventricular rate histogram shows the rate distribution of ventricular sensed and paced events.

3.3 Battery and device performance data

The device automatically measures and records battery and device performance data every day. This information can help you assess the status of the device battery and identify issues with performance. The device records the following types of performance data:
Remaining longevity estimate and replacement indicators
Electrode impedance trend
Sensing amplitude trends (R-wave amplitude and A4 amplitude)
Capture threshold trend
Sensing integrity counter
You can access battery and device performance data from several different screens on the programmer:
Quick Look II screen: Data > Quick Look II
Battery and device measurements window: Data > Diagnostics > Battery and Device Measurements > Open Data
Electrode impedance window: Data > Diagnostics > Electrode Impedance Trend > Open Data
Capture threshold window: Data > Diagnostics > Capture Threshold Trend > Open Data
R-wave amplitude window: Data > Diagnostics > R-Wave Amplitude Trend > Open Data
A4 amplitude window: Data > Diagnostics > A4 Amplitude Trend > Open Data
14 Reference Manual
Medtronic Micra™ AV MC1AVR1

3.3.1 Remaining Longevity estimate and replacement indicators

The device automatically measures the battery voltage once a day at 02:30. The automatic daily battery voltage measurement is displayed on the Data – Battery and Device Measurements screen.
The programmer is able to estimate the remaining device longevity (the number of years until the battery reaches RRT) after 11 days of the device manufacture date. Longevity estimates are based on a history of battery voltage measurements made by the device since the manufacture date.
The Battery and Device Measurements screen provides the mean (Estimated at), minimum, and maximum values for remaining longevity. These values are based on a statistical analysis of accelerated battery discharge data. The maximum and minimum remaining longevity estimates are 95th percentile values calculated from the distribution of this data. That is, approximately 95% of devices are expected to reach RRT before the reported maximum value and approximately 95% of devices are expected to reach RRT after the reported minimum value.
If the device longevity estimator determines that there are less than 6 months (180 days) until the End of Service (EOS), the programmer displays the RRT symbol and the date when the battery reached the RRT condition. If the programmer displays the RRT symbol, contact your Medtronic representative and schedule a patient appointment to implant a new device.
The expected service life of the device after RRT, defined as the Prolonged Service Period (PSP), is 6 months (180 days). After the first 90 days of the PSP have passed, the device reaches the Elective Replacement Indicator (ERI) and the programmer displays the ERI indicator.1 When the device reaches the ERI condition, it automatically changes the pacing mode to VVI and sets the pacing rate to 65 bpm, unless the device is programmed to a non-pacing mode. It also changes Rate Hysteresis to Off if this feature is programmed to On. When the ERI indicator is displayed on the programmer, implant a new device immediately. For more information, see the Micra AV SW044 Programming Guide. The device reaches End of Service (EOS) based on battery voltage and switches to the Device Off mode, permanently deactivating the pacing operation. The programmer displays the EOS symbol.
1
Note: After ERI, all pacing parameters can be programmed, including mode and rate. Reprogramming the pacing parameters may reduce the duration of the ERI to EOS period.
Warning: When the battery voltage reaches the EOS condition, the device permanently deactivates pacing and sensing and switches to the Device Off mode. The EOS symbol appears on the programmer screen.
1
ERI may be indicated before the end of 90 days, and EOS may be indicated before the end of 180 days if the actual battery usage exceeds the expected conditions during the Prolonged Service Period. For an explanation of these conditions, see the Micra AV MC1AVR1 Device Manual.
Reference Manual 15
Medtronic
Micra™ AV MC1AVR1

3.3.2 Electrode impedance trend

Every day at 02:30, the device delivers a ventricular pace and automatically measures the electrode impedance. If the intrinsic heart rate is faster than the programmed pacing rate, the device increases the pacing rate to be slightly faster than the intrinsic rate for 1 interval.
The daily automatic electrode impedance measurements are displayed on the Electrode Impedance Trend screen, which plots the data as a graph. The graph shows up to 15 of the most recent measurements and up to 80 weekly summary measurements, providing minimum and maximum values for each week. Significant or sudden changes in electrode impedance may indicate a problem with the pacing electrode.
If the device is unable to perform automatic electrode impedance measurements, gaps are present in the trend graph.

3.3.3 Sensing amplitude trend

After completing the electrode impedance measurement, which starts at 02:30 every day, the device begins to measure the amplitude of intrinsic ventricular sensed events. The device attempts to measure the amplitude of 5 normal intrinsic ventricular sensed events. After collecting these measurements, the device records their median value as the most recent R-wave amplitude measurements. If the device has not collected 5 amplitude measurements for the day by midnight, no measurement is recorded. The sensing amplitude trend graph shows a gap for that day.
Note: The A4 Amplitude Trend measures the maximum acceleration in the A4 window, which may not be the true A4 amplitude if the A4 signal does not occur in the A4 window.
The daily automatic sensing amplitude measurements are displayed on the R-Wave Amplitude Trend and A4 Amplitude Trend windows, which plot the data as a graph. The graph shows up to 15 of the most recent measurements and up to 80 weekly summary measurements, providing minimum and maximum values for each week. Significant or sudden changes in the R-wave amplitude may indicate a problem with the sensing electrode. Significant or sudden changes in the A4 amplitude may indicate a problem with effective VDD pacing therapy.
When operating in the VDD mode, the device periodically collects A4 amplitude measurements over the course of the day. The A4 amplitude is measured as the maximum acceleration in the A4 window. This measurement is collected regardless of whether an atrial sense occurred in that cycle. At the end of the day at midnight (00:00), the device stores the most common measured A4 amplitude of the day in the A4 Amplitude Trend. If the device has not collected enough measurements for the day, no measurement is recorded. The A4 Amplitude Trend graph shows a gap for that day. This gap can happen if the device does not spend enough time per day operating in the VDD mode.
16 Reference Manual
Medtronic
To view the R-wave sensing amplitude trend, tap Data > Diagnostics > R-Wave Amplitude Trend > Open Data.
To view the A4 sensing amplitude trend, tap Data > Diagnostics > A4 Amplitude Trend > Open Data.
Micra™ AV MC1AVR1

3.3.4 Capture threshold trend

If the Capture Management feature is programmed to Adaptive or Monitor, the device automatically performs daily pacing threshold searches and records the results in the Capture Threshold Trend data. In the adaptive mode, Capture Management also performs hourly pacing threshold confirmation checks. For more information about Capture Management, see Section 4.5, “Pacing with Capture Management”, page 50.
The results of the most recent daily pacing threshold measurements are displayed on the Data – Trends window on the capture threshold trend graph. The graph shows up to 15 of the most recent measurements and up to 80 weekly summary measurements, including minimum and maximum values for each week. The Capture Threshold window also shows programmed values for pacing output and Capture Management parameters, the last measured threshold value, and a link to a detailed view of the last 15 days of the threshold measurement data.
The Capture Management (Last 15 days detail) window shows daily results from the last 15 days of threshold measurements. These results include the dates, times, threshold measurements, pacing amplitude values, and notes describing the results of each pacing threshold search.
The Capture Threshold trend data provides a means to evaluate the operation of Capture Management and the appropriateness of the current pacing output values. In addition, sudden or significant changes in the pacing threshold may indicate a problem with the pacing electrode.
Note: It is possible for a high threshold observation to occur without a corresponding value shown on the Capture Threshold Trend graph. The observation occurs when a single Capture Management test is aborted due to a high threshold value. When a single Capture Management test is aborted due to a high threshold value, the device attempts a new Capture Management test an hour later. If the new test does not result in a high threshold value, the device stores this result in the Capture Threshold Trend for the day. If 3 consecutive Capture Management tests are aborted due to a high threshold, a threshold value of > 5.0 V is stored in the Capture Threshold Trend. The device does not attempt any more Capture Management tests for that day.
To view the capture threshold trend on the programming screen, tap Data > Diagnostics > Capture Threshold Trend > Open Data.
Reference Manual 17
Medtronic Micra™ AV MC1AVR1

3.3.5 Sensing Integrity Counter

The Sensing Integrity Counter records the number of short ventricular intervals that occur between patient sessions. A large number of short ventricular intervals may indicate oversensing.
To view the Sensing Integrity Counter on the programming screen, tap Data > Diagnostics > Battery and Device Measurements > Open Data.
18 Reference Manual
Medtronic
Micra™ AV MC1AVR1

4 Configuring pacing therapies

4.1 Sensing

Effective sensing is essential for the safe and effective use of the device. The device must sense the occurrence of intrinsic cardiac events while avoiding oversensing so that it can deliver therapies appropriately. Effective ventricular sensing can reduce the effects of long depolarizations after paced events, oversensing the same event, and sensing T-waves, noise, and interference. Effective atrial sensing can improve the efficacy of VDD pacing therapy.
Programmable blanking periods and atrial refractory periods help to screen out extraneous sensing or to prevent the device from responding to it. Both blanking periods and atrial refractory periods follow pacing pulses and sensed events. Sensing is inhibited during blanking periods. The device is able to sense events that occur during atrial refractory periods, but it marks them as refractory events. Refractory events generally have no effect on the timing of subsequent pacing events.

4.1.1 Ventricular sensing

The device senses electrically in the right ventricle. No sensing occurs in the left ventricle. Each sensitivity setting represents a threshold value that defines the minimum electrical amplitude recognized by the device as a sensed event in the right ventricle. Selecting a higher value for the sensing threshold reduces the sensitivity to lower amplitude signals.
4.1.1.1 Operation of ventricular sensing threshold
Ventricular sensing operates at a fixed, programmed ventricular sensitivity.
4.1.1.2 Operation of ventricular blanking periods
Ventricular blanking periods follow paced and sensed ventricular events. Ventricular blanking periods help to prevent the device from sensing pacing pulses, post-pacing depolarization, T-waves, and oversensing of the same event.
Reference Manual 19
Ventricular
blanking
Markers
Medtronic Micra™ AV MC1AVR1

Figure 1. Programmable blanking periods

1 For the duration of this ventricular blanking period, which is defined by the Blank Post VS
parameter, ventricular sensing is disabled after a sensed ventricular event.
2 For the duration of this ventricular blanking period, which is defined by the Blank Post VP
parameter, ventricular sensing is disabled after a paced ventricular event.
4.1.1.3 Noise reversion in the VVI and VVIR modes
The operation associated with continuous noise sensing is called noise reversion. During continuous noise reversion, ventricular sensing is disabled and ventricular pacing occurs at the sensor-indicated rate in the VVIR mode and at the programmed lower rate in the VVI mode.
4.1.1.4 Preventing noise sensing
Noise reversion may be caused by electromagnetic interference (EMI) or low sensitivity settings. You can reduce or eliminate noise reversion by one of the following actions:
Identify the source of EMI and increase the distance between the patient and the EMI source.
Reprogram Sensitivity to a less sensitive setting (higher numerical value).
Reprogram Blank Post VP and Blank Post VS to blank the T-waves.
4.1.1.5 Ventricular sensing parameters

Table 1. How to navigate to ventricular-electrical sensing parameters

Parameters Path
RV Sensitivity Params > Sensitivity
Blank Post VP Blank Post VS
Params > RV Blanking
20 Reference Manual
Medtronic
Micra™ AV MC1AVR1

4.1.2 Atrial mechanical sensing

Although the device is located entirely in the right ventricle, mechanically sensing the atrium from the ventricle makes it possible to provide AV synchronous pacing for patients with AV block. The device senses and paces electrically in the right ventricle. The device also uses an accelerometer to detect the mechanical vibrations produced by the atrial contraction, called A4. The ventricular contraction and relaxation also produces mechanical vibrations, called A1, A2, and A3. The timing of A1, A2, A3, and A4 is similar to the heart sounds S1, S2, S3, and S4, respectively.
Warning: Patient activities and environments which present mechanical vibrations to the patient can interfere with the mechanical sensing of atrial contractions. This can result in loss of AV synchrony.
The signals are defined as follows:
The A1 signal:
– Occurs at the beginning of ventricular systole and represents the closing of the mitral
valve and the tricuspid valve
– Occurs after the beginning of the QRS complex on an ECG
The A2 signal:
– Occurs at the completion of ventricular systole and represents the closing of the aortic
valve and the pulmonary valve
– Occurs near the end of the T-wave on an ECG
The A3 signal:
– Occurs during ventricular diastole
– Occurs after the T-wave on an ECG
– Corresponds in timing to the E-wave on a Doppler echo
– Represents the passive filling of blood from the atrium into the ventricle
The A4 signal:
– Occurs when the atrium contracts and pushes blood into the ventricle
– Occurs after the P-wave on an ECG
– Corresponds in timing to the A-wave on a Doppler echo
Reference Manual 21
Loading...
+ 47 hidden pages