Medtronic MAP-1000 Instructions for Use

Riptide™
Aspiration Pump
TABLE OF CONTENTS
Riptide™ Aspiration Pump
English . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Symbol Glossary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2
English en
CANISTER HOLDER
TUBING
ASPIRA
COLLECTION CANISTER
User’s Manual
Riptide™ Aspiration Pump 110-115VAC, 60Hz 100VAC, 50/60Hz
VACUUM CONTROL KNOB
VACUUM GAUGE
ON/OFF BUTTON
POWER CORD CONNECTION
Figure 1: Riptide™ Aspiration Pump Features and Controls
TION PUMP
Figure 2: Riptide™ Aspiration Pump and Riptide™ Collection Canister with
DESCRIPTION
The Riptide™ Aspiration Pump is an externally powered electromechanical device capable of generating a vacuum designed for drawing uids and small particles into the Riptide™ Collection Canister during neurovascular interventional procedures. It is intended for clinical and catheter lab use and is not intended for transport or eld applications.
The recommended operating range for the Riptide™ Aspiration Pump is between 20-25 inHg (68-85 kPa).
The Riptide™ Aspiration Pump is intended for use as a component of the Riptide™ Aspiration System. The Riptide™ Aspiration System includes the Arc™ Catheter, React™ 68 Catheter, React™ 71 Catheter, Riptide™ Aspiration Tubing, Riptide™ Large Bore Aspiration Tubing, Riptide™ Aspiration Pump, and Riptide™ Collection Canister with Intermediate Tubing.
Federal law (USA) restricts this device and other components of the Riptide™ Aspiration System to sale by or on the order of a physician.
The Riptide™ Aspiration Pump package includes the following: o Riptide™ Aspiration Pump o Region specic power cords for connection to earthed receptacle. Select correct
cord from included kit. o Riptide™ Aspiration System Instructions for Use o Riptide™ Aspiration Pump User’s Manual
Required for use with Riptide™ Aspiration System (packaged separately) o Arc™ Catheter, Applied Part o React™ 68 Catheter, Applied Part o React™ 71 Catheter, Applied Part o Riptide™ Aspiration Tubing, Applied Part o Riptide™ Large Bore Aspiration Tubing, Applied Part o Riptide™ Aspiration Pump o Riptide™ Collection Canister with Intermediate Tubing
Intermediate Tubing
POTENTIAL EQUALIZATION PLUG
VACUUM INLET
COOLING AIR INTAKE
INTERMEDIATE
INDICATIONS AND CONTRAINDICATIONS
Refer to the Riptide™ Aspiration System Instructions for Use for indications and contraindications.
OPERATING DESCRIPTION
VACUUM GAUGE
VACUUM CONTROL KNOB
Figure 3: Riptide™ Aspiration Pump Internal Composition
The Riptide™ Aspiration Pump uses an industry proven oilless rocking piston type vacuum pump. A single, permanent split capacitor type electric motor powers a two-stage pump arrangement that provides high vacuum and ecient ow through the Riptide™ Collection Canister. In a reciprocating motion, a exible cup mounted on top of a wristless piston and connecting rod in each stage creates vacuum as the cup creates a vacuum as a result of the rocking motion. The vacuum is drawn in through the vacuum inlet and exhausted internally in the enclosure where it is mixed with cooling air and discharged through a vent located on the bottom.
ON/OFF BUTTON
VACUUM INLET
EXHAUST
TWO-STAGE PUMP
PUMP MOTOR
COOLING AIR EXHAUST
WARNINGS
The Riptide™ Aspiration System in contact with patients should only be used by physicians who have received appropriate training in interventional neuro­endovascular techniques and treatment of acute ischemic stroke. The Riptide™ aspiration tubing, Riptide™ Aspiration Pump, and Riptide™ Collection Canister with Intermediate Tubing are designed for setup by clinical support sta under the supervision of a trained physician.
Do not use the Riptide™ Aspiration System with components other than the Arc™ Catheter, React™ 68 Catheter, React™ 71 Catheter, Riptide™ Aspiration Tubing, Riptide™ Large Bore Aspiration Tubing, Riptide™ Aspiration Pump, and Riptide™ Collection Canister with Intermediate Tubing.
The Riptide™ Aspiration Pump is a Class I Medical Equipment. To avoid the risk of electric shock, this equipment must only be connected to a supply mains with protective earth.
PRECAUTIONS
The Riptide™ Aspiration Pump and the Riptide™ Collection Canister with Intermediate Tubing are supplied non-sterile and are intended for use outside of the sterile eld only.
Do not use the Riptide™ Aspiration Pump without the Riptide™ Collection Canister with Intermediate Tubing connected to the vacuum inlet.
The Riptide™ Aspiration Pump is intended for use only with the Riptide™ Collection Canister with Intermediate Tubing. The lid is equipped with integral overow protection and a biological lter required for safe and correct operation in the Riptide™ Aspiration System. Using non-validated canisters, lters, or systems may result in improper patient connections, uid overow, biological contamination, and non­repairable damage of the Riptide™ Aspiration Pump.
Do not allow aspirated uids, overow, or other liquids into the Riptide™ Aspiration Pump. This will cause non-repairable damage to the pump and create biological contamination. Pumps so contaminated must be removed from service and disposed of in accordance with standard biological waste disposal procedures in eect.
The Riptide™ aspiration tubing and Riptide™ Collection Canister with Intermediate Tubing are single use only.
The Riptide™ Aspiration Pump does not contain any user serviceable part inside. Do not attempt to open the enclosure.
Do not use in presence of explosive atmospheres, ammable liquids or anesthetic mixtures, and nitrous oxide.
Do not use in oxygen enriched environment.
SETUP
Remove the Riptide™ Aspiration Pump from storage and/or its packaging (if present) and visually inspect for damage.
Place the Riptide™ Aspiration Pump on a rigid, stable, at, horizontal surface outside of the sterile eld, within reach of the Riptide™ aspiration tubing and providing sucient tubing slack to the patient table.
The Pump should be placed so that the tubing connections, controls and gauges are visible and accessible to the operator. The Riptide™ Aspiration Pump must never be positioned in a manner that makes it dicult to operate.
Locate the Riptide™ Aspiration Pump such that a minimum of 6 inches clearance exists in all directions to adjacent objects or other common cath lab equipment, powered or un-powered.
Do not block the air vents located on the back and bottom of the enclosure. Do not place the pump on thick or plush absorbent table covering and do not cover the pump with poly or other sheeting during operation. Blocked air ow will result in pump overheating.
Attach the correct end of the supplied power cord to the power connection on the back of the enclosure. Press in rmly and conrm its solid positioning.
If required, attach the potential equalization connector to the potential equalization plug identied by a green/yellow disk located at the back of the pump.
Conrm that the mains power is available and the pump’s nameplate ratings are compatible.
The device power cord is the isolation means and is detachable from the Riptide™ Aspiration Pump. The cord’s appliance connector is an IEC60320 C13 type and dierent from the mains connector. Use only the correct power cord intended for the region. To connect the device, push the cord’s C13 plug into the C14 type appliance power inlet plug located at the back of the enclosure. At the other cord end, insert the mains plug into the designated mains power receptacle. NOTE: The Riptide™ Aspiration Pump should not be positioned such that it is dicult to disconnect the supplied power cord.
The Riptide™ Aspiration Pump’s power cable should not be bundled or held in proximity with other power cables that may carry very large surges and transients such as welders and large electric motors. In addition, separation from data cables and communications lines should follow EN50174-2:2008. In general, approximate separation from unshielded lines should be no less than 2 feet, separation for shielded cables 2 inches. Failure to follow these guidelines may result in electromagnetic interference due to radiated or re-radiated electrical noise.
Check the basic operation of the Riptide™ Aspiration Pump by pressing the ON/OFF power switch on the front panel. The switch should light up green and pump motor should start. The presence of vacuum can be conrmed by temporarily blocking the vacuum inlet with a nger and operating the pump’s vacuum control knob while observing the vacuum gauge. Note that the tethered cap is not a vacuum seal and may not seal air tight without additional nger pressure on top. Turn the pump o by pressing the ON/OFF power switch before proceeding.
Do not use the Riptide™ Aspiration Pump for aspiration without the Riptide™ Collection Canister with Intermediate Tubing connected to the vacuum inlet.
Obtain a new Riptide™ Collection Canister with Intermediate Tubing and perform the setup in accordance with the Riptide™ Aspiration System Instructions for Use and the Riptide™ Collection Canister with Intermediate Tubing Quick Reference Guide.
The Riptide™ Aspiration Pump is now ready to connect to Riptide™ aspiration tubing per the Riptide™ Aspiration System Instructions for Use.
OPERATION
Refer to the Riptide™ Aspiration System Instructions for Use and the Riptide™ Collection Canister with Intermediate Tubing Quick Reference Guide to conrm complete setup for aspiration.
With the Arc™ Catheter or React™ 68 Catheter, or React™ 71 Catheter, Riptide™ Aspiration Tubing, Riptide™ Large Bore Aspiration Tubing, Riptide™ Aspiration Pump, and Riptide™ Collection Canister with Intermediate Tubing connected, the Riptide™ Aspiration Pump is ready to begin supplying vacuum to the connected components.
Conrm that the ow control switch on the Riptide™ aspiration tubing is in the OFF position.
The ON/OFF power switch on the control panel is a push button that requires moderate force inward to activate and de-activate. When depressed to the ON position, the button will lock in the recessed position, the green symbol will light, and the pump will begin to operate. When the button is depressed again, the button will return to the original ush position, the green symbol will go dark, and pump will cease to run.
At any time when the power switch is turned OFF after the Riptide™ Aspiration Pump is run, any vacuum present in the system will be released and the Riptide™ Collection Canister with Intermediate Tubing, and Riptide™ aspiration tubing proximal to the ow control switch will return to ambient conditions in a few seconds.
Press the ON/OFF power switch on the Riptide™ Aspiration Pump control panel and conrm that the power switch button lights, the pump motor starts and produces a steady hum.
Allow it to run for at least one minute. Conrm normal operation and the presence of vacuum as indicated on the gauge.
Using the vacuum control knob on the Riptide™ Aspiration Pump, adjust the vacuum to a reading of minimum 20 inHg but not exceeding 25 inHg on the gauge. Adjust the vacuum by rotating the vacuum control. To increase the vacuum, turn the knob clockwise. To decrease the vacuum, turn the knob counter-clockwise.
Note that it is normal for the indicator needle of the gauge to oscillate or vibrate in response to pump pulsation. Allow the range of the needle to stabilize and adjust the vacuum control so that the midpoint of the motion aligns with the desired setting.
The Riptide™ Aspiration System is now ready to aspirate.
During the procedure, monitor the vacuum level indicated on the gauge. It is normal for the needle to drift about the desired set point as use conditions change. The vacuum level may drop as a result of turning the Riptide™ aspiration tubing ow control switch on, opening a tubing connection or exchanging a catheter. The vacuum level will recover to its set point when the ow control switch is returned to OFF or when the tubing or catheter connection is restored.
During the procedure, monitor the uid level that may have collected in the canister. It is recommended that the uid not be allowed to ll more than about 75% of the capacity. Should this occur, the physician should have the Riptide™ Collection Canister with Intermediate Tubing changed out.
At the conclusion of the aspiration procedure, turn o the Riptide™ Aspiration Pump by pressing the ON/OFF power switch. The green light on the switch will go out and the pump will stop producing a humming noise.
The Arc™ Catheter, React™ 68 Catheter, React™ 71 Catheter, Riptide™ Aspiration Tubing, Riptide™ Large Bore Aspiration Tubing and Riptide™ Collection Canister with Intermediate Tubing are single use only. Remove and discard in accordance with standard biological waste disposal procedures in eect.
Disconnect the power cord from the mains.
CLEANING
Place the attached tethered cap over the vacuum inlet prior to cleaning.
Disconnect the power cord from the pump and clean separately if required.
The outside of the Riptide™ Aspiration Pump enclosure may be cleaned by wiping down with 70% IPA and common non-bleach, non-solvent, clinical surface cleaning agents, e.g. Diversey Virex Tb (ready-to-use) or equivalent. Wipe dry before storage.
Do not use cleaning agents containing petroleum base solvents, acids, caustics, chlorinated solvents.
The Riptide™ Aspiration Pump is not compatible with any sterilization procedures including ETO, e-beam, gamma, and autoclave.
3
Do not attempt to disassemble the pump other than removing the cord and Riptide™ Collection Canister with Intermediate Tubing.
Do not immerse the Riptide™ Aspiration Pump in water or cleaning solutions or use a ooding rinse as a cleaning method.
Do not inject, spray or otherwise introduce water, IPA, or cleaning agents into the cooling air intake louvers or exhaust vents located on the bottom.
Never inject, spray, aspirate, or otherwise introduce cleaning or decontamination agents into the vacuum inlet. The pump is designed only for air contact and the introduction of any liquids or solids will cause non-repairable damage to internal components.
If the Riptide™ Aspiration Pump is known to have ingested uids during a procedure it should be considered as contaminated and treated in accordance with standard biological waste procedures in eect. No approved cleaning methods exist for internal decontamination.
STORAGE
Place the tethered cap over the vacuum inlet to prevent entry of dust and other contaminants.
The Riptide™ Aspiration Pump may be stored in a clean warehouse or storeroom commonly used for clinical and lab supplies. See characteristics below for environmental limits.
The pump does not have an applicable “Use-by date”. However, after extended storage it is recommended that it be cleaned, examined for damage, and tested for basic functionality prior to its introduction in to a clinical setting.
TROUBLESHOOTING
No user serviceable parts are inside the pump. Do not modify the pump or open the enclosure.
Use the troubleshooting chart below to determine common causes and recommended remedies.
Riptide™ Aspiration Pump Fuses – replace only with exact equivalent F 5A H 250V, UL listed 5 x 20mm.
PROBLEM POSSIBLE CAUSE ACTION
Pump has mechanical damage to enclosure, gauges, controls
Power cord or a plug on the cord damaged
Power connection on pump enclosure damaged
Pressing ON power switch causes mains breaker to trip or mains fuse to blow
ON power switch does not light when pushed, pump does not run
Improper handling, dropping, mechanical damage
Improper handling, misuse, mechanical damage
Improper handling, misuse, mechanical damage
Internal electrical failure Replace pump
Mains power not connected Plug pump into receptacle
Cord not connected to appliance power inlet
Loose cord at back of unit Press in or adjust cord in back of
Bad power cord Replace power cord
Mains power source is bad or incorrect voltage
Pump fuse blown Replace fuse(s)
Mains supply breaker not on or mains fuse blown
On power switch not pushed hard enough
Internal mechanical or electrical failure
Replace pump
Discard and do not repair a damaged power cord. Do not use cords intended for use in regions other than local or use electrical plug adapters or converters to achieve a connection. Doing so may result in damage to the Riptide™ Aspiration Pump and violate electrical safety requirements. Required replacement power cords vary by region and certied commercial assemblies are readily available. The cords must meet the applicable specications listed in the CHARACTERISTICS table herein.
Replace pump
Plug cord into enclosure’s appliance power inlet plug
enclosure
Plug pump into correct receptacle
Use only exact equivalent. F 5A H250V, UL listed 5 x 20mm.
Turn on mains supply or replace fuse
Press again
Replace pump
PROBLEM POSSIBLE CAUSE ACTION
ON power switch does not light but pump starts and appears operating normally
ON power switch lights when pushed but pump makes no noise
ON power switch lights and pump starts briey but switch does not remain depressed
ON power switch lights when pushed and cooling fan on bottom runs but pump does not run
Pump runs but has abnormal operation: unusually loud, high vibration, running very hot, burning smell, smoke
Gauge needle vibrates excessively
Gauge needle indicates vacuum greater than 30 inHg or is pegged fully counter­clockwise when running
Pump runs but no or low vacuum indicated at gauge, vacuum may be present at Riptide™ aspiration tubing
Faulty ON power switch Complete procedure if required but
Internal electrical failure or faulty ON power switch
Faulty ON power switch Replace pump
Overheated motor due to operation at full load for an extended period or blocked cooling vents
Pump contaminated by aspiration overow
Internal mechanical or electrical failure
Pump rocking - not sitting on level surface with all 4 feet
Cooling air intake and/or bottom exhaust is blocked
Pump contaminated by overow
Pump connected to improper mains voltage supply
Internal electrical or mechanical damage or failure
Vacuum pressure set too low – especially below 15 inHg.
Faulty gauge if needle does not stabilize at full vacuum.
Faulty or damaged gauge. Use vacuum control knob to
Vacuum control knob not adjusted
Faulty gauge Use vacuum control knob to
Flow control switch on Riptide™ aspiration tubing not set correctly
replace pump at rst opportunity
Replace pump
Turn pump o and allow time to cool down. Remove any absorbent padding or plastic sheeting blocking pump air vents.
Replace pump. Units contaminated with bio uids should be disposed of in accordance with standard hospital biohazard procedures.
Replace pump
Re-position pump on at, rigid, horizontal surface
Re-position pump. Do not place pump on plush absorbent padding or cover with splash sheeting.
Replace pump. Units contaminated with bio uids should be disposed of in accordance with standard hospital biohazard procedures.
Replace pump with unit that matches mains power input
Replace pump
Increase vacuum between 20-25 inHg. It is normal for the indicator needle of the gauge to oscillate or vibrate in response to pump pulsation, especially at minimal vacuums. Allow the range of the needle to stabilize and adjust the vacuum control so that the midpoint of the motion aligns with the desired setting.
Procedure may be continued at physician’s discretion but replace pump at rst opportunity
attempt to set vacuum to recommended range of 20-25 inHg. Procedure may be continued at physician’s discretion but replace pump at rst opportunity
Adjust knob to set vacuum to desired level
attempt to set vacuum to recommended range of 20-25 inHg. Procedure may be continued at physician’s discretion but replace pump at rst opportunity
Adjust ow control switch to OFF
PROBLEM POSSIBLE CAUSE ACTION
Pump runs with vacuum indicated at gauge, but no vacuum at Riptide™ aspiration tubing distal end
Pump runs but low suction or insucient vacuum in Catheter
Pump contaminated by overow, pump seizing
Internal electrical or mechanical damage or failure
Loose, leaking, or missing Intermediate Tubing
Flow control switch in OFF position
Riptide™ Collection Canister lid loose or drain port open
Riptide™ Collection Canister not vertical and secured in Riptide™ Aspiration Pump receptacle
Riptide™ Collection Canister bio-lter plugged
Plugged or kinked tubing Straighten or replace tubing
Broken canister, defective connector or vacuum tubing
Riptide™ Collection Canister full or foaming. Overow protection valve has activated and stopped ow
Riptide™ aspiration tubing ow control switch set incorrectly
Defective or broken Riptide™ aspiration tubing ow control switch
Loose, leaking, or missing Intermediate Tubing
Loose canister lid, vacuum leak
Canister dump/vent port is open or leaking
Pinched or kinked connector or suction tubing
Incorrect vacuum control knob setting
Incorrect ow control switch setting
Flow control switch on tubing defective
Canister bio lter is clogged or overow valve is closed
Vacuum leak internal to pump enclosure
Canister is full and overow protection working normally
Pump contaminated by canister overow or liquid ingress from previous procedure
Replace pump
Replace pump
Install tubing correctly and check ttings
Move switch to ON position
Tighten lid, close port with tethered cap
Turn o the Riptide™ Aspiration Pump and place the Riptide™ Collection Canister in the receptacle, and restart the Riptide™ Aspiration Pump
Replace the Riptide™ Collection Canister. Do not attempt to clean or bypass lter.
Replace with a new Riptide™ Collection Canister with Intermediate Tubing and Riptide™ aspiration tubing
Replace the Riptide™ Collection Canister. Do not attempt to bypass overow valve.
Set ow switch to desired setting
Replace the Riptide™ aspiration tubing
Install tubing and check ttings
Reposition and press lid closed
Close dump/vent port with tethered cap on lid
Straighten connector or suction tubing
Adjust vacuum control knob to proper setting
Adjust ow control switch to proper setting – OFF to build vacuum, ON to direct vacuum to catheter
Replace the Riptide™ aspiration tubing
Replace the Riptide™ Collection Canister. Do not attempt to clean or bypass lter.
Replace pump
Replace the Riptide™ Collection Canister. Do not attempt to clean or bypass canister. Dispose of full canister in accordance with standard hospital biohazard procedures.
Replace pump. Do not attempt to clean or decontaminate pump. Pump units contaminated with bio uids should be disposed of in accordance with standard hospital biohazard procedures.
4
PROBLEM POSSIBLE CAUSE ACTION
Vacuum indicated on gauge does not respond to change in vacuum control setting
Pump spontaneously stops running or is intermittent during use
Pump does not shut o when switch is pushed
Riptide™ Collection Canister does not t in the receptacle of the Riptide™ Aspiration Pump
Tubing does not t correctly on Riptide™ Aspiration Pump vacuum inlet
MAINTENANCE
Operating life: The Riptide™ Aspiration Pump has an operating life of 500 hours.
The enclosure does not contain any user serviceable parts inside and should not be opened.
Fuses: Located at the back of the unit. Replace only with exact equivalent F 5A H 250V, UL listed 5 x 20mm.
Sterilization: The Riptide™ Aspiration Pump is NOT designed to be sterilized by autoclaving, chemical means (such as ETO and other sterilant compounds), or radiation. Any attempt to do so will cause non-repairable damage to the unit.
DISPOSAL
Riptide™ Aspiration Pump with mechanical/electrical failure or normal wear-out: Dispose of non-functioning pumps in accordance with standard hospital procedures for electromechanical equipment.
Riptide™ Aspiration Pump with contamination by aspirated uid ingress: Dispose of in accordance with standard hospital biohazard procedures.
The Arc™ Catheter, React™ 68 Catheter, React™ 71 Catheter, Riptide™ Aspiration Tubing, Riptide™ Large Bore Aspiration Tubing and Riptide™ Collection Canister with Intermediate Tubing are single use only. Dispose of in accordance with standard hospital biohazard procedures.
Faulty vacuum control knob or faulty vacuum gauge
Internal electrical or mechanical damage or failure
Loss of mains power, breaker interruption, power surge, brownout
Power cord disturbed, loose Re-plug power cord securely to
Blown Riptide™ Aspiration Pump fuse
Pump overheated due to long continuous running in poor cooling
Riptide™ Collection Canister overowed and uid ingress has seized pump
Attempted use of non­validated canister without lters; uid ingress causing pump to seize
Incorrect tubing connection for the Riptide™ aspiration tubing connected directly to vacuum inlet of the Riptide™ Aspiration Pump bypassing the Riptide™ Collection Canister
Internal electrical or mechanical damage or failure
Switch not pushed hard enough
Faulty switch or internal electrical failure
Attempted use of non­validated canister system.
Attempted use of non­validated canister and/ or tubing
Attempted connection of the Riptide™ aspiration tubing directly to the vacuum inlet of the Riptide™ Aspiration Pump
Replace pump
Replace pump
Restore mains power or connect to uninterruptable power supply.
mains and pump enclosure
Replace fuse(s) Use only exact equivalent F 5A H 250V, UL listed 5 x 20mm.
Allow pump to cool, correct cooling air ow blockage, and restart
Replace Riptide™ Aspiration Pump and Riptide™ Collection Canister
Replace Riptide™ Aspiration Pump. Use correct Riptide™ Collection Canister Pump may be biological hazard
Replace pump. Pump may be biological hazard
Replace pump
Press again
Pull mains plug to un-power. Replace pump
Replace with Riptide™ Collection Canister with Intermediate Tubing
Replace with Riptide™ Collection Canister with Intermediate Tubing
Review the Riptide™ Aspiration System Instructions for Use. Install the Riptide™ Collection Canister and connect the Riptide™ aspiration tubing to the port labeled “PATIENT” on the Riptide™ Collection Canister.
CHARACTERISTICS
Characteristic Value
Model Riptide™ Aspiration Pump MAP-1000
Vacuum Range, Nominal 0-29 inHg (0-98kPa)
Vacuum Gauge +/- 5% of full scale value
Flow Rate, Nominal 0-0.8SCFM (0-23 LPM)
Power Input Rating 110-115VAC, 60Hz, 500VA max.
Overcurrent Protection (Fuses) F 5A H 250V, UL listed 5 x 20mm
Approximate Dimensions Length – 16.1” (40.9cm)
Weight Approximately 24lbs (11kg)
Duty Cycle Non-continuous
Operating Conditions Temperature: 65 to 75F (18 to 24ºC)
Storage Conditions Temperature: -25 to 125ºF (-32 to 52ºC)
Enclosure Ingress Rating IP21 (Drip Tight)
Power Cord, North America Rating: 10A, 125VAC
Power Cord, Japan Rating: 12A, 125VAC
100VAC, 50/60Hz, 700VA max.
Depth – 13.2” (33.5cm) Height – 12.3” (31.2cm)
97% (58.2 minutes on, 1.8 minutes o)
Humidity 20-80% non-condensing Air pressure: 24-31 inHg (81-105kPa)
Humidity: 5-90% non-condensing Air pressure: 20-31 inHg (68-105 kPa)
Length: 10 feet (3.05m) General: UL62, CSA 22.2.no.49 Type: Hospital grade, UL817, CSA C22.2 no. 21 Mains plug: NEMA 5-15 (IEC type B), green dot Appliance plug: IEC60320 C13 Cable type: SJT Conductors: 3, AWG18 minimum Conductor color code: IEC60446
Length: 2.5m (~8.2feet) General: JIC C 3005 Type: Hospital grade, JIS T 1021, “H” mark Mains plug: JIS8303 (IEC type B-Japan), green dot Appliance plug: IEC60320 C13 Cable type: VCTF Conductors: 3, 1.25mm minimum Conductor color code: IEC60446
ELECTRICAL SAFETY
General IEC60601-1:2012 General standard, Medical devices
Earthbond IEC60601-1:2012
Leakage Current IEC60601-1:2012
ELECTROMAGNETIC COMPATIBILITY
The Riptide™ Aspiration Pump is expected to be relatively unaected by common electromagnetic (EM) disturbances. In the event of unusual or extraordinary EM events beyond the limits, it is possible that device may exhibit a loss of Essential Performance (the ability to generate the necessary vacuum required for aspiration as set by the user). Should this occur, discontinue use of the device. The Riptide™ Aspiration Pump is intended for use in electromagnetic environments as specied below.
0.2 ohm max.
Type BF, Class I Medical Equipment
Guidance and Manufacturer’s Declarations
Guidance and Manufacturer’s Declaration –
The Riptide™ Aspiration Pump is intended for use in the electromagnetic environment specied below. The customer or the user of the pump should make sure that it is used in such an environment.
Emissions Test Compliance Electromagnetic Environment
RF Emissions CISPR 11
Electromagnetic Emissions
- Guidance
Group 1 The Riptide™ Aspiration Pump
may use RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
Emissions Test Compliance Electromagnetic Environment
RF Emissions CISPR 11
Harmonic Emissions IEC 61000-3-2
Voltage Fluctuations/ Flicker Emissions IEC 61000-3-3
Class B The Riptide™ Aspiration
Complies
Complies
- Guidance
Pump is suitable for use in all establishments other than domestic and those directly connected to the public low­voltage power supply network that supplies buildings used for domestic purposes.
Guidance and Manufacturer’s Declaration –
Electromagnetic Immunity
The Riptide™ Aspiration Pump is intended for use in the electromagnetic environment specied below. The customer or the user of the pump should make sure that it is used in such an environment.
Immunity
Test
Electrostatic Discharge (ESD)
IEC 61000-4-2
Electrical Fast Transient/ Burst IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage Dips, Short Interruptions, and Voltage Variations on Power Supply Lines IEC 61000­4-11
Power Frequency (50/60 Hz) Magnetic Field IEC 61000-4-8
NOTE: UT is the a.c. mains voltage prior to application of the test level.
Conducted RF IEC 61000-4-6
IEC 60601 Test Level
±6 kV contact ±8 kV air
±2 kV for power supply lines ±1 kV for input/ output lines
±1 kV line(s) to line(s) ±2 kV line(s) to earth
<5% UT (>95% dip in UT) for 0.5 cycle 40% UT (60% dip in UT) for 5 cycles 70% UT (30% dip in UT) for 25 cycle <5% UT (>95% dip in UT) for 5 seconds
3 A/m Complies Mains power quality should
3V rms 150 kHz to 80 MHz
Compliance
Level
Complies The Riptide™ Aspiration
Complies Mains power quality should
Complies Mains power quality should
Complies Mains power quality
Complies
Electromagnetic
Environment - Guidance
Pump should not be aected by electrostatic discharge that might occur under normal conditions of use.
It is recommended that oors should be wood, concrete or ceramic tile. If oors are covered with synthetic material, the relative humidity should be at least 30%.
be similar to that of a typical commercial or hospital environment.
be similar to that of a typical commercial or hospital environment.
should be similar to that of a typical commercial or hospital environment. If the user of the Riptide™ Aspiration Pump requires continued operation during power mains interruptions, it is recommended that the Riptide™ Aspiration Pump be powered from an uninterruptible power supply or a battery.
be similar to that of a typical commercial or hospital environment.
Portable and mobile RF communications equipment should be used no closer to any part of the Riptide™ Aspiration Pump than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Recommended separation distance
5
Radiated RF IEC 61000-4-3
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is
aected by absorption, and reection from structures, objects and people.
a
Field strengths from xed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcasts and TV broadcast cannot be predicted theoretically with accuracy. To access the electromagnetic environment due to xed RF transmitters, an electromagnetic site survey should be considered. If the measured eld strength in the location in which the Riptide™ Aspiration Pump is used exceeds the applicable RF compliance level above, the pump should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the pump or the transmitting device from which it is receiving signals. b
Over the frequency range 150 kHz to 80 MHz, eld strengths should be less than 3 V/m.
3V/m 80 MHz to 6 GHz
Complies d = 1.2 √P 80 MHz-800 MHz
d = 2.3 √P 800 MHz-6 GHz
Where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m). Field strengths from xed RF transmitters, as determined by an electromagnetic site survey, a should be less than the compliance level in each frequency range. Interference may occur in the vicinity of equipment marked with the following symbol:
b
Recommended separation distances between portable and
mobile RF communications equipment and the Riptide™
Aspiration Pump
This section provides information on the recommended separation distance between portable and mobile RF communications equipment and the Riptide™ Aspiration Pump. The Riptide™ Aspiration Pump is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the pump users can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the pump as recommended below, according to the maximum output power of the communications equipment.
Rated Maximum Output Power of Transmitter (W)
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.74
1 1.2 1.2 2.3
10 3.8 3.8 7.4
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is
aected by absorption, and reection from structures, objects and people.
WARRANTY DISCLAIMER
Although this product has been manufactured under carefully controlled conditions, the manufacturer has no control over the conditions under which this product is used. The manufacturer therefore disclaims all warranties, both expressed and implied, with respect to the product including, but not limited to, any implied warranty of merchantability or tness for a particular purpose. The manufacturer shall not be liable to any person or entity for any medical expenses or any direct, incidental or consequential damages caused by any use, defect, failure or malfunction of the product, whether a claim for such damages is based upon warranty, contract, tort or otherwise. No person has any authority to bind the manufacturer to any representation or warranty with respect to the product. The exclusions and limitation set out above are not intended to, and should not be construed so as to contravene mandatory provisions of applicable law. If any part or term of this Disclaimer of Warranty is held illegal, unenforceable or in conict with applicable law by a court of competent jurisdiction, the validity of the remaining portions of this Disclaimer of Warranty shall not be aected, and all rights and obligations shall be construed and enforced as if this Disclaimer of Warranty did not contain the particular part or term held to be invalid.
Separation Distance According to the Frequency of
150 kHz to
80 MHz
d = 1.2√P
Transmitter (m)
80 MHz to
800 MHz
d = 1.2√P
800 MHz to
6 GHz
d = 2.3√P
6
S
www.medtronic.com/manuals
IP21
REF
N
ymbol Glossary
Caution: Federal (USA) law restricts this device to sale by or on the order of a physician
Consult instructions for use at this website.
Caution
Follow Instructions for Use
Keep dry
Type BF Applied Part
S
CONTENTS
NON
STERILE
Serial number
Date of manufacture
Contents of Package
Non-sterile
Equipotentiality
Dangerous voltage
Ingress Protection Marking (IP Code per IEC60529) Protection against vertically dripping water drops
Catalogue number
Manufacturer
E491532
Stand-by
Variability, rotational adjustment
UL Certication MEDICAL - GENERAL MEDICAL EQUIPMENT AS TO ELECTRICAL SHOCK, FIRE AND MECHANICAL HAZARDS ONLY IEC 60601-1 Edition 3.1 (2012)/EN 60601-1:2016 + A1:2013 + A12:2014 ANSI/AAMI ES 60601-1:A1: 2012, C1: 2009/(R) 2012 and A2: 2010/(R) 2012, CSA CAN/ CSA-C22.2 NO. 60601-1:14
7
Medtronic, Medtronic with logo and Medtronic logo are U.S. and internationally registered trademarks of Medtronic.
© 2019 Medtronic.
Micro Therapeutics, Inc. d/b/a ev3 Neurovascular 9775 Toledo Way Irvine, CA 92618 USA Tel: +1.949.837.3700
M998301ADOC2 Rev. A (09/2019)
Loading...