Medtronic MAC-1200 Instructions for Use

Riptide™
Aspiration System
TABLE OF CONTENTS
Riptide™ Aspiration System
English . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Symbol Glossary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2
English en Instructions for Use
Riptide™ Aspiration System
DEVICE DESCRIPTION
The Riptide™ Aspiration System is comprised of the following devices:
Arc™ Catheter
React™ 68 Catheter
React™ 71 Catheter
Riptide™ Aspiration Tubing
Riptide™ Large Bore Aspiration Tubing
Riptide™ Aspiration Pump
Riptide™ Collection Canister with Intermediate Tubing
The Riptide™ Aspiration System is designed to remove occlusive thrombus from the cerebral vasculature using continuous aspiration. The catheter shaft has a hydrophilic coating that spans the distal 40 cm to reduce friction during use. The catheter acts as a conduit for providing direct aspiration to the occlusion site generated by the Riptide™ Aspiration Pump. The catheter is introduced through a guide catheter or sheath and into the intracranial vasculature and guided over a neurovascular guidewire and/or microcatheter to the site of the primary occlusion. The catheter is connected to the Riptide™ Aspiration Pump through the aspiration tubing. The Intermediate Tubing connects the Riptide™ Collection Canister to the Riptide™ Aspiration Pump. The Riptide™ Aspiration System in this conguration may be used to aspirate thrombus from the occluded vessel.
Note: The catheter and aspiration tubing are Applied Parts.
INDICATION FOR USE
The Riptide™ Aspiration System is intended for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.
CONTRAINDICATION
There are no known contraindications.
WARNINGS
The Riptide™ Aspiration System should only be used by physicians who have received appropriate training in interventional neuro-endovascular techniques for treatment of acute ischemic stroke. The Riptide™ Aspiration System is designed for setup by clinical support sta under the supervision of a trained physician.
Do not advance, retrieve, or use any component of the Riptide™ Aspiration System against resistance without careful assessment of the cause using uoroscopy. If the cause cannot be determined, withdraw the device or system as a unit. Unrestrained torquing or forced insertion of the catheter against resistance may result in damage to the device or vessel.
For vessel safety, do not perform more than three attempts at aspiration in the same vessel using the Riptide™ Aspiration System.
Do not use devices in conjunction with or in place of those devices specied above when using the Riptide™ Aspiration System.
PRECAUTIONS
The catheter and the aspiration tubing are supplied sterile and are intended for single use only. Do not resterilize or re-use. Resterilization and/or re-use may result in ineective catheter coating lubrication, which may result in high friction and the inability to access the target neuro vasculature location.
Use the catheter, aspiration tubing, and Riptide™ Collection Canister with Intermediate Tubing prior to the “Use-by date”.
The Riptide™ Aspiration Pump and Riptide™ Collection Canister with Intermediate Tubing are supplied non-sterile and are intended for use outside of the sterile eld only.
The Riptide™ Collection Canister with Intermediate Tubing is intended for single use only. Dispose in accordance with standard biologic waste disposal procedures.
Do not use kinked or damaged devices. Do not use open or damaged packages.
Inspect the catheter before use to verify that its size and condition are suitable for the specic procedure.
Use the catheter under uoroscopic guidance to conrm placement of the catheter proximal to the thrombus prior to applying aspiration.
Maintain a constant infusion of appropriate ush solution.
When performing aspiration, ensure that the aspiration tubing ow switch is in the open position for only the minimum time needed to remove thrombus.
The recommended continuous duration of aspiration is a maximum of 90 seconds.
If repositioning of the catheter is necessary during the revascularization procedure, such reposition should be performed over an appropriate neurovascular guidewire and/or microcatheter using standard techniques.
Do not use automated high-pressure contrast injection equipment with the catheter because it may damage the device.
Medical management and post stroke care should follow AHA/ASA guidelines and institutional protocols.
As in all surgical interventions, monitoring of intra-procedural blood loss is recommended so that appropriate management may be instituted.
Possible malfunctions of the Riptide™ Aspiration System include, but are not limited to issues associated with vacuum for the aspiration tubing, Riptide™ Aspiration Pump,
ARC-132 REACT-68 REACT-71 AT-88-110 MAT-110-110 MAP-1000 MAC-1200
and Riptide™ Collection Canister with Intermediate Tubing that may lead to decreased therapeutic response during mechanical thrombectomy.
Operators should take all necessary precautions to limit X-radiation doses to patients and themselves by using sucient shielding, reducing uoroscopy times, and modifying X-ray technical factors where possible.
POTENTIAL ADVERSE EVENTS
acute occlusion
allergic reaction and anaphylaxis from contrast media
arrhythmia
complication at puncture site
complication of radiation exposure (e.g., alopecia, burns ranging in severity from skin reddening to ulcers, cataracts, and delayed neoplasia)
death
decreased therapeutic response
embolism
PROCEDURE
1. Refer to Warnings, Precautions, and Potential Adverse Events prior to use.
2. As each component of the Riptide™ Aspiration System is used, remove the component from the packaging and inspect for damage or for kinks.
3. Prepare a guide catheter or sheath according to the device manufacturer’s Instructions for Use.
4. Place the guide catheter or sheath into the appropriate artery for access to the occlusion site.
CATHETER PREPARATION AND USE
1. Conrm vessel diameter and ensure the catheter is appropriate for treatment of the vessel.
Recommended Sizing Guidelines
Model
ARC-132 ≥ 2.5
REACT-68 ≥2.7
REACT-71 ≥2.7
2. Prepare the catheter for use by ushing the packaging hoop and device with heparinized saline.
3. Remove the catheter from the packaging hoop.
4. Inspect the catheter for kinks or damage. If any damage is observed, replace with a new device.
5. Connect a rotating hemostasis valve to the proximal hub of a guide catheter or
sheath.
NOTE: When using the React™ 71 Catheter, the minimum inner diameter recommended for a sheath is 0.087” (2.21 mm).
6. Insert the catheter through the rotating hemostasis valve. If required, the provided split y-introducer may be used to facilitate the introduction of the catheter through the rotating hemostasis valve, inserting the tip through the introducer starting at the split end and penetrating the valve with the tubular introducer end. After the catheter has passed through the rotating hemostasis valve, the split y-introducer may be removed by backing it out of the valve, pulling gently on the split ends to separate the remaining portion. Discard the separated introducer.
7. Using conventional catheterization techniques under uoroscopic guidance advance the catheter into the target vessel over an appropriate neurovascular guidewire and/or microcatheter. Position the catheter proximal to the thrombus. Remove the guidewire and/or microcatheter from the catheter.
NOTE: Access to the site of occlusion may be aided by the triaxial use of a compatible inner microcatheter.
Riptide™ Aspiration Pump
Figure 1: Riptide™ Aspiration System Setup Diagram
stula
foreign body in patient
hypertension/hypotension
infection
inammation
intracranial hemorrhage
ischemia
nerve damage
neurological decits including stroke
organ failure/shock
pain
stenosis
vessel spasm, thrombosis, dissection, or perforation
Recommended Vessel Diameter (mm)
Minimum
Riptide™ Collection Canister with Intermediate Tubing
Aspiration Tubing
Catheter
Riptide™ Aspiration System PREPARATION AND USE
1. Refer to Figure 1 (above) for the correct arrangement of devices for proper setup of the Riptide™ Aspiration System.
2. Refer to the Riptide™ Aspiration Pump User’s Manual for pump setup, operating instructions and troubleshooting if required.
3
3. Refer to the Riptide™ Collection Canister Quick Reference Guide for assembly setup.
4. Place the Riptide™ Aspiration Pump on a stable, at horizontal surface outside of the sterile eld within reach of the aspiration tubing and providing sucient slack to the patient table. Locate the Riptide™ Aspiration Pump such that a minimum of 6 inches clearance exists in all directions to adjacent objects or other common cath lab equipment, powered or un-powered.
5. Attach the Riptide™ Aspiration Pump’s power cord to the designated mains nominal power specied in the Riptide™ Aspiration Pump User’s Manual. Conrm that the power cord is securely plugged into the back of the pump enclosure. If required, attach the potential equalization connector to the potential equalization plug.
6. The Riptide™ Collection Canister lid should be pre-assembled to the cup as it is removed from its packaging. If it is not or becomes separated, secure the lid to the cup by snapping it tightly on the lip.
7. The large drain port on the Riptide™ Collection Canister lid should be pre-assembled with a closure cap as it is removed from its packaging. If it is not or becomes separated, seal the port with the large tethered cap on the lid by pressing it in place rmly. The small caps attached to the lid may remain tethered and unused until needed.
8. Place the Riptide™ Collection Canister into the receptacle on the Riptide™ Aspiration Pump and press rmly into place. The horizontal port marked with the embossed label as labeled “PATIENT” should point away from the pump enclosure to enable unobstructed attachment of the aspiration tubing.
9. The Intermediate Tubing should be pre-assembled to the Riptide™ Collection Canister lid as it is removed from its packaging. If it is not or becomes separated, secure the connector at one end of the supplied Intermediate Tubing over the vertical central port of the Riptide™ Collection Canister marked with the embossed label as “VACUUM” by pressing down rmly. Either connector end of the Intermediate Tubing may be used.
10. Attach other end of the Intermediate Tubing to the vacuum inlet of the Riptide™ Aspiration Pump by pressing down rmly.
11. Open the aspiration tubing pouch with the contents accessible toward the sterile eld. Remove the tubing assembly from the pouch into the sterile eld direction. Place or hold the distal end with the ow control switch on the patient table and pass the proximal end of aspiration tubing from the sterile eld to the Riptide™ Collection Canister located on the Riptide™ Aspiration Pump.
12. Attach the connector at the proximal end of the aspiration tubing to the horizontal Riptide™ Collection Canister port marked with the embossed label as labeled “PATIENT”. Press the connector rmly onto the port.
13. Slide the actuator of the ow control switch on the aspiration tubing to OFF to close the tubing to ow.
14. Press the start button on the Riptide™ Aspiration Pump and allow it to run for at least one minute. Conrm normal operation and the presence of vacuum as indicated on the gauge.
15. Using the vacuum control knob on the Riptide™ Aspiration Pump, adjust the vacuum to a reading of minimum 20 inHg on the gauge but not exceeding 25 inHg. Adjust the vacuum by rotating the vacuum control knob until the gauge reads the desired vacuum level. To increase the vacuum, turn the knob clockwise. To decrease the vacuum, turn the knob counter-clockwise.
NOTE: It is normal for the indicator needle of the gauge to oscillate or vibrate in response to pump pulsation. Allow the range of the needle to stabilize and adjust the vacuum control so that the midpoint of the motion aligns with the desired setting.
16. The Riptide™ Aspiration System is now ready to aspirate.
17. Connect luer tting at the distal end of the aspiration tubing to the side port of the rotating hemostasis valve attached to the catheter.
18. Advance the distal tip of the catheter to engage with the proximal end of the thrombus or in accordance with the physicians preferred technique.
19. To begin aspiration, slide the ow control switch toward the ON position. To stop the aspiration ow, slide the actuator of the ow control switch toward the OFF position.
20. If necessary, additional passes with catheter may be used to further remove thrombus at the discretion of the physician.
21. After withdrawal from the patient, disconnect the aspiration tubing from the catheter; replace it with a 5cc or 10cc syringe, and aspirate approximately 5cc of blood from the catheter to remove any thrombus that may remain.
22. Obtain a post-treatment angiogram in accordance with the physicians preferred technique.
23. At the conclusion of the aspiration procedure, turn OFF the Riptide™ Aspiration Pump. The catheter, the aspiration tubing, and the Riptide™ Collection Canister with Intermediate Tubing are single use. Remove and discard in accordance with standard biological waste disposal procedures.
24. Clean and prepare the Riptide™ Aspiration Pump for storage as described in the User’s Manual.
WARRANTY DISCLAIMER
Although this product has been manufactured under carefully controlled conditions, the manufacturer has no control over the conditions under which this product is used. The manufacturer therefore disclaims all warranties, both expressed and implied, with respect to the product including, but not limited to, any implied warranty of merchantability or tness for a particular purpose. The manufacturer shall not be liable to any person or entity for any medical expenses or any direct, incidental or consequential damages caused by any use, defect, failure or malfunction of the product, whether a claim for such damages is based upon warranty, contract, tort or otherwise. No person has any authority to bind the manufacturer to any representation or warranty with respect to the product. The exclusions and limitation set out above are not intended to, and should not be construed so as to contravene mandatory provisions of applicable law. If any part or term of this Disclaimer of Warranty is held illegal, unenforceable or in conict with applicable law by a court of competent jurisdiction, the validity of the remaining portions of this Disclaimer of Warranty shall not be aected, and all rights and obligations shall be construed and enforced as if this Disclaimer of Warranty did not contain the particular part or term held to be invalid.
Loading...
+ 5 hidden pages