Medtronic LS10 Specifications

Page 1
ESSENTIAL ENERGY
The Valleylab™ LS10 energy platform
delivers the benets of LigaSure
sealing technology in a single-channel,
Specication Guide
1
vessel
Page 2
PERFORMANCE
CHARACTERISTICS
General
Output
conguration
Cooling Natural and forced convection
Display System status indicator – circle LED
Mounting A Medtronic cart (UC8009) or a stable,
Isolated output
(12 blocks) shows system status
System error indicator – triangle LED with exclamation mark
Instrument status indicator – LED above the LigaSure
Usage limit indicator – 2 with a line through it
at surface
port
Dimensions and Weight
Width 300 mm (11.81 inches)
Depth 377 mm (14.84 inches)
Height 105 mm (4.13 inches)
Weight 5 Kg (11 lbs)
Transport and Storage
Ambient temperature range
Relative humidity 25% to 85% (non-condensing)
Atmospheric pressure
Duration of storage
-30 C to 65 C (-22 F to 149 F)
500 to 1060 millibars
If stored for more than one year, see service manual for instructions or contact Medtronic service for further information.
Duty Cycle
Under maximum-output settings and rated-load conditions (30 ohm load), the energy platform is suitable for activation times of 5 seconds on, 15 seconds o, for 1 hour. With lesser settings and loads, you can activate the energy platform for greater durations without generating excessive internal temperatures.
Internal Battery
Battery for RTC Battery type: 3 V lithium button cell
Battery life: ve years
Operating Parameters
Ambient temperature range
Relative humidity 30% to 75% non-condensing
Atmospheric pressure
Warm-up time If transported or stored at
10 C to 40 C (50 F to 104 F)
700 to 1060 millibars
temperatures outside the operating temperature range, allow one hour for the energy platform to reach room temperature before use.
Audio Volume
The stated audio levels are at a distance of one meter. Alert tones meet the requirements of IEC60601-2-2.
Activation Tone
Volume (adjustable)
Frequency (nominal)
Duration Continuous while the system is
45 dBa minimum
Sealing in process – 440 Hz
activated
Page 3
Alert Tone
Volume
(not adjustable)
Frequency Seal cycle incomplete alert:
65 dBa minimum
784 Hz (high), 587 Hz (low)
Low Frequency (50/60 Hz) Leakage Current
(IEC 60601-2-2)
Enclosure source current, ground open
< 300 μA
Seal cycle complete tone: 985 Hz
System error tone: 1421 Hz
Duration Seal cycle incomplete alert:
The LigaSure played for 150 ms each with no break between tones
The order and frequency of the tones is 784 Hz, 587 Hz, 784 Hz, 587 Hz (high, low, high, low)
Seal cycle complete tone:
Two tones played for 175 ms each at 985 Hz, with a 175 ms break between the tones
System error tone:
Three 200 ms tones, separated by 300 ms for each error/system-alert event
regrasp alert is four tones
Radio Frequency Identication (RFID)
The RFID module is located above the LigaSure™ port. It identies the inserted LigaSure™ instrument and congures the energy platform with the data included in the RFID tag.
Source current, patient leads, all
outputs
Sink current at high line, all inputs
Normal polarity, intact ground: < 10 μA
Normal polarity, ground open: < 50 μA
Reverse polarity, ground open: < 50 μA
Mains voltage on applied part: < 50 μA
< 50 μA
High Frequency (RF) Leakage Current
(IEC 60601-2-2)
Measured with leads
LigaSure leakage
recommended by Medtronic
<116 mA RMS < 100 mA RMS
Measured directly at
the system terminals
LigaSure™ Vessel Sealing Technology
Seal: 400 kHz sinusoid, continuous
Output power changes by less than 20% or 12W, whichever is greater, as the line voltage varies from 90–132 volts and 208–264 volts (at rated load)
Frequency range 13.56 MHz
RF output power 68.17 dBuV/m at three meters
Type of antenna Integral loop antenna
Modulation Amplitude-shift keying (ASK)
Mode of operation (simplex/duplex)
Contains transmitter module FCC ID
Contains IC ID 11355A-JDK1901
Duplex
2AAVI-JDK1901
Input Power Requirements
Operating range is 90 to 264 AC volts 48– 62 Hz Maximum current is 5 amperes
Meets ETL and CE Specications
The VLLS10GEN energy platform meets all the pertinent clauses of IEC 60601-1 third edition and IEC 60601-2-2 third edition
0123
Page 4
OUTPUT
CHARACTERISTICS
Maximum Output for LigaSure™ Mode
Power readouts agree with actual power into rated load to within 15% or 5 W, whichever is greater.
PRECAUTION: To avoid injury to the patient or surgical team, use only instruments rated for use at, or greater than, the maximum peak voltages listed below. For example, bipolar instruments must have voltage ratings of 250 V peak or greater, as shown in the “Open Circuit Peak Voltage (max)” column.
Mode
LigaSure
Open Circuit
Peak Voltage (max)
250 V 500 V 30 Ω 270 W N/A 5.5 A
Open Circuit
P–P Voltage (max)
Rated Load (max)
Power (max) Duty Cycle
Current R.M.S
(max)
Output Waveforms
LigaSure™ vessel-sealing technology, an automatic adjustment, controls all modes. As tissue resistance increases throughout the seal cycle, the energy platform modulates current and voltage until tissue resistance meets seal complete requirements as needed by the tissue-sensing technology.
Reference:
1. Based on a clinical usability validation lab conducted by Medtronic with 22 surgeons and 15 nurses. 100% of surgeons and nurses surveyed believe it is easy to understand how to operate the Valleylab
IMPO RTANT: Pleas e refer to the pa ckage inser t for comple te instruc tions, cont raindic ations, war nings and pr ecaution s.
©2021 Medt ronic. All r ights rese rved. Med tronic, Me dtronic log o and Furthe r, Togethe r are tradem arks of Medt ronic. ™*Third p arty bra nds are trad emarks of th eir respec tive owners . All other br ands are tra demarks o f aMedtronic co mpany. 20-weu-ls10-spec-guide-4715293
medtronic.com/covidien/uk
LS10. Report: R0046696_A. October 2014.