Medtronic LIFEPAK CR-T Instruction Manual

Page 1
SETUP INSTRUCTIONS
LIFEPAK® CR–T
AED Trainer
After receiving the LIFEPAK CR–T AED Trainer, carefully remove the Trainer and the operating instructions from the shipping container. Examine the Trainer and accessories for any sign of damage during shipping.
Shipping container contents include:
LIFEPAK CR–T AED Trainer
Remote control
Remote control cable
Yellow plastic shock key cover
CHARGE-PAK™ Trainer
Quick reference card
If any components are missing or the Trainer appears to be damaged, contact your dealer or local Medtronic representative right away. Save the container for reshipping the Trainer if necessary.
Always place the Trainer in a location separate from the defibrillator to avoid confusion in real emergencies.
QUIK-PAK™ training electrodes
2 AAA alkaline batteries (for remote control)
4 D-cell alkaline batteries (for Trainer)
Operating instructions
Lid labels (international trainers only)
Soft case with identification card
Accessory Bag
Accessory
Bag
D-Cell
+
D-Cell
+
+
+
AAA
AAA
+
D-Cell
D-Cell
+
Page 2
Soft Case
Medtronic, Inc.
11811 Willows Road NE Redmond, WA 98052-2003 USA Telephone: 425.867.4000 Toll Free (USA only): 800.442.1142 Fax: 425.867.4121 Internet: www.medtronic-ers.com
Medtronic and LIFEPAK are registered trademarks of Medtronic, Inc. CHARGE-PAK and QUIK-PAK are trademarks of Medtronic, Inc. Specifications are subject to change without notice. © 2002-2007 Medtronic, Inc. All rights reserved.
Medtronic Europe S.A.
Rte. Du Molliau 31 Case postale 84 1131 Tolochenaz Switzerland Telephone: 41.21.802.7000 Fax: 41.21.802.7900
Page 3
LIFEPAK® CR–T
AED Trainer
Operating Instructions
Page 4
Page 5
OPERATING INSTRUCTIONS
LIFEPAK® CR–T
AED Trainer
Page 6
Text Conventions
Throughout these operating instructions, special text characters are used to indicate labels and voice prompts:
Operating control labels:
Voice prompts:
CAPITAL ITALICIZED LETTERS such as CALL FOR HELP NOW.
CAPITAL LETTERS such as LID RELEASE/ON-OFF.
Version History
These operating instructions describe the LIFEPAK CR–T AED Trainer, MIN 3201804-005 or later.
Medtronic , Inc.
11811 Willows Road NE Redmond, WA 98052-2003 USA Telephone: 425.867.4000 Toll Free (USA only): 800.442.1142 Fax: 425.867.4121 Internet: www.medtronic-ers.com
Medtronic and LIFEPAK are registered trademarks of Medtronic, Inc. CHARGE-PAK and QUIK-PAK are trademarks of Medtronic, Inc. Specifications are subject to change without notice. ©2002-2007 Medtronic, Inc. All rights reserved.
ii LIFEPAK CR–T Trainer Operating Instructions
Medtronic Europe S.A.
Rte. Du Molliau 31 Case postale 84 1131 Tolochenaz Switzerland Telephone: 41.21.802.7000 Fax: 41.21.802.7900
MIN 3201838-003 / CAT. 26500-001156
Page 7
TABLE OF CONTENTS 1
1 Getting Started
Introduction ............................................................................................................................................. 2
Unpacking and Inspecting the LIFEPAK CR–T AED Trainer .................................................................. 2
Controls, Indicators, and Labels ............................................................................................................. 3
Outside Controls, Indicators, and Labels ........................................................................................3
Remote Control and Remote Control Cable ................................................................................... 7
Installing and Replacing Batteries .......................................................................................................... 8
Installing and Replacing Trainer Batteries ....................................................................................... 8
Installing and Replacing Remote Control Batteries......................................................................... 8
2 Using the LIFEPAK CR–T AED Trainer
Voice Prompts and Tones ..................................................................................................................... 10
Using the Remote Control .................................................................................................................... 10
Remote Control Functions ............................................................................................................ 10
Using the Remote Control Cable ..................................................................................................12
Preparing for a Training Session ........................................................................................................... 12
Conducting a Training Session ............................................................................................................. 13
Following a Training Session ................................................................................................................ 14
3 Setup
Setup Configuration and Operating Settings ........................................................................................ 16
Changing Options for Operating Settings ............................................................................................. 18
Using the Remote Control and the LED Display ........................................................................... 18
Procedures for Changing Operating Settings................................................................................ 18
Shock Protocol .............................................................................................................................. 18
Selecting Shock Protocols.................................................................................................... 18
Configuring Shock Protocols ................................................................................................ 19
General Operating Settings........................................................................................................... 20
LIFEPAK CR–T Trainer Operating Instructions iii
©2002-2007 Medtronic, Inc.
Page 8
CPR Time...................................................................................................................................... 21
Language Option........................................................................................................................... 21
Setup .................................................................................................................................... 22
Resetting Default Settings............................................................................................................. 22
4 Caring for the LIFEPAK CR–T AED Trainer
Replacing Batteries ............................................................................................................................... 24
Troubleshooting ..................................................................................................................................... 24
Cleaning ................................................................................................................................................ 26
Product Recycling Information .............................................................................................................. 26
Recycling Assistance .................................................................................................................... 26
Preparation.................................................................................................................................... 26
Recycling of Training Electrodes ................................................................................................... 26
Packaging...................................................................................................................................... 26
Warranty ............................................................................................................................................... 26
Accessories .......................................................................................................................................... 27
Specifications ........................................................................................................................................ 27
A LIFEPAK CR–T AED Trainer Voice Prompts and Tones ....................................................... 29
B LIFEPAK CR–T AED Trainer Quick Start Reference Card .................................................... 33
C Declaration of Conformity .................................................................................................... 37
Index............................................................................................................................................. 41
iv LIFEPAK CR–T Trainer Operating Instructions
Page 9
LIST OF FIGURES
Figure 1 LIFEPAK CR–T AED Trainer Outside Controls, Indicators, and Labels (Top and Bottom)........ 3
Figure 2 LIFEPAK CR–T AED Trainer Outside Controls, Indicators, and Labels (Back and Side).......... 4
Figure 3 LIFEPAK CR–T AED Trainer Inside Features ........................................................................... 5
Figure 4 Inside Features after Opening the Electrode Packet................................................................. 6
Figure 5 Remote Control and Remote Control Cable.............................................................................. 7
Figure 6 LIFEPAK CR–T AED Trainer and Remote Control Battery Placement...................................... 8
Figure 7 Connecting the Training Electrode Packet to the LIFEPAK CR–T AED Trainer....................... 13
Figure 8 Placing Training Electrodes on the Manikin............................................................................. 14
Figure 9 LED Display............................................................................................................................. 18
Figure 10 Custom #1 and #2 Options...................................................................................................... 19
Figure 11 General Operating Setting Options .........................................................................................20
Figure 12 CPR Time Options ..................................................................................................................21
Figure 13 Language Options ................................................................................................................... 22
LIST OF TABLES
Table 1 Outside Controls, Indicators, and Labels (Top and Bottom) ..................................................... 3
Table 2 Outside Controls, Indicators, and Labels (Back and Side) ....................................................... 4
Table 3 Inside Features......................................................................................................................... 5
Table 4 Inside Features after Opening the Electrode Packet................................................................ 6
Table 5 Remote Control Features (Setup Functions)............................................................................ 7
Table 6 Remote Control Button Functions .......................................................................................... 10
Table 7 Trainer Operating Setting Options .......................................................................................... 16
Table 8 Shock Protocol Options.......................................................................................................... 19
Table 9 General Operating Setting Options ........................................................................................20
Table 10 Language Groups................................................................................................................... 21
Table 11 Voice Prompts and Tones Troubleshooting............................................................................. 24
Table 12 LED Display Troubleshooting ................................................................................................. 24
Table 13 Setup Options Troubleshooting .............................................................................................. 25
Table 14 Remote Control Troubleshooting ............................................................................................ 25
Table 15 Training Electrodes Troubleshooting....................................................................................... 25
Table 16 LIFEPAK CR–T AED Trainer Accessories.............................................................................. 27
Table 17 LIFEPAK CR–T AED Trainer Specifications ........................................................................... 27
Table 18 Remote Control Specifications ............................................................................................... 28
Table 19 Voice Prompts and Tones....................................................................................................... 31
LIFEPAK CR–T Trainer Operating Instructions v
©2002-2007 Medtronic, Inc.
Page 10
Page 11
1 GETTING STARTED
Getting Started
This section introduces the LIFEPAK® CR–T Automated External Defibrillator (AED) Trainer and provides an orientation to its features.
Introduction page 2 Unpacking and Inspecting the LIFEPAK CR–T
2
AED Trainer Controls, Indicators, and Labels 3 Installing and Replacing Batteries 8
LIFEPAK CR–T Trainer Operating Instructions 1
©2002-2007 Medtronic, Inc.
Page 12
Getting Started
INTRODUCTION
The LIFEPAK CR–T AED Trainer provides the necessary interface for practical skill demonstration during a simulated cardiac arrest. The Trainer is designed to assist the instructor with realistic training on the use of the defibrillator without the actual charge and discharge of electrical energy.
The Trainer simulates voice prompts, audible tones, and readiness display features of all LIFEPAK CR defibrillator models. These features help guide trainees through simulated analysis, simulated shock delivery, and prompted CPR time.
An infrared, wireless, remote control allows the instructor to operate the Trainer during training sessions. Labeled buttons on the remote control are used to define operating settings for training scenarios and, while training is in progress, pause Trainer operation or interrupt preset training scenarios. Optional operating settings offer flexibility and a variety of potential setup configurations.
The Trainer and remote control are intended for indoor use under normal room temperature and lighting conditions.
Note: The Trainer does not provide training for the defibrillator data management functions.
UNPACKING AND INSPECTING THE LIFEPAK CR–T AED TRAINER
After receiving the LIFEPAK CR–T AED Trainer, carefully remove the Trainer and the operating instructions from the shipping container. Examine the Trainer and accessories for any sign of damage during shipping.
Shipping container contents include:
LIFEPAK CR–T AED Trainer
•QUIK-PAK
training electrodes
Yellow plastic shock key cover
Remote control
Remote control cable
•CHARGE-PAK
Trainer
If any components are missing or the Trainer appears to be damaged, contact your dealer or local Medtronic representative right away. Save the container for reshipping the Trainer if necessary.
Note: Always place the Trainer in a location separate from the defibrillator to avoid confusion in real emergencies.
Operating instructions
Quick reference card
Soft case with identification card
Lid labels (international trainers only)
4 D-cell alkaline batteries (for Trainer)
2 AAA alkaline batteries (for remote control)
2 LIFEPAK CR–T Trainer Operating Instructions
Page 13
Getting Started
CONTROLS, INDICATORS, AND LABELS
There are several important controls, indicators, and labels on the LIFEPAK CR–T AED Trainer with which you should become familiar.
Outside Controls, Indicators, and Labels
The outside controls, indicators, and labels are identified in Figure 1 and Figure 2 and described in
Ta bl e 1 and Tab le 2 .
1
2
3
Getting Started
6
4
5
To p B o tt o m
Figure 1 LIFEPAK CR–T AED Trainer Outside Controls, Indicators, and Labels (Top and Bottom)
Tab le 1 Outside Controls, Indicators, and Labels (Top and Bottom)
1 Carrying handle The carrying handle is used to transport the Trainer.
2 Readiness display This display simulates whether the defibrillator is ready for use or needs
attention. The remote control and CHARGE-PAK activate the following indicators:
OK
OK indicator. Simulates that the defibrillator is ready for use.
CHARGE-PAK indicator. Simulates that the CHARGE-PAK needs to be replaced or is not installed.
Attention indicator. Simulates that the internal battery of the defibrillator is low.
Wrench indicator. Simulates that there is a condition that prevents or could prevent normal operation of the defibrillator. Service is required.
7
3 Lid The top of the Trainer. After the lid is open, the Trainer begins training
scenarios. When the lid is closed, the Trainer simulates the defibrillator in OFF status.
LIFEPAK CR–T Trainer Operating Instructions 3
©2002-2007 Medtronic, Inc.
Page 14
Getting Started
Tab le 1 Outside Controls, Indicators, and Labels (Top and Bottom) (Continued)
4 CHARGE-PAK Trainer The CHARGE-PAK Trainer is not functional. It is used only to
demonstrate how to remove and replace the CHARGE-PAK.
5 Lid release/On-Off
button
Button that opens the Trainer lid and simulates turning on the defibrillator. Pressing this button for approximately 2 seconds after the lid is open simulates turning off the defibrillator.
6 Storage compartment Storage compartment for the remote control cable and shock button
cover.
7 Battery compartment Compartment that holds four D-cell alkaline batteries.
8 9
10
Figure 2 LIFEPAK CR–T AED Trainer Outside Controls, Indicators, and Labels (Back and Side)
Tab le 2 Outside Controls, Indicators, and Labels (Back and Side)
8 Remote control signal
sensor
Back Side
14
13
Sensor used for wireless remote control operation. The signal from the remote control must be detected at this sensor for proper communication.
11
12
9 Light emitting diode
(LED) display
Display containing ten LED bars that are viewed to confirm selections when operating settings are changed. The LED display is also used to verify volume level when the volume is adjusted.
10 Remote control
Storage compartment for the remote control.
compartment
11 Software download
Connection point for loading software to the Trainer.
receptacle
12 Remote control
Connection point for the remote control cable.
receptacle
13 Master power
14 CHARGE-PAK Trainer
compartment
4 LIFEPAK CR–T Trainer Operating Instructions
switch Press and release switch that provides power to the Trainer.
Compartment that accommodates the removable CHARGE-PAK Trainer.
Page 15
Getting Started
Inside Features
When the Trainer lid is opened, you will see the electrode packet and the release handle shown in
Figure 3. Figure 4 shows the features you see after releasing the electrode packet. Ta b l e 3 and Tabl e 4
describe the inside features shown in Figure 3 and Figure 4.
15
16
17
Getting Started
19
Figure 3 LIFEPAK CR–T AED Trainer Inside Features
Tab le 3 Inside Features
18
15 Quick reference card An illustrated and condensed guide for using the Trainer.
16 Use By date Use By date shown: yyyy-mm-dd. This simulated date is visible when
the plastic lid is open or closed.
17 QUIK-PAK electrode
Packet that holds a set of training electrode pads.
packet
18 Electrode packet release
handle
19 Electrode packet anchor
Pulling this red handle tears open the training electrode packet and releases the training electrode pads.
Pin that secures the training electrode packet to the Trainer.
pin
LIFEPAK CR–T Trainer Operating Instructions 5
©2002-2007 Medtronic, Inc.
Page 16
Getting Started
25
20
21
24
23
22
Figure 4 Inside Features after Opening the Electrode Packet
Tab le 4 Inside Features after Opening the Electrode Packet
20 Speaker Speaker that projects the voice prompts and tones.
21 Electrode pad indicators Indicators that automatically flash red when the Trainer lid is opened,
alerting the trainee that the electrode pads are NOT applied to the training manikin.
Indicators turn solid green when the instructor presses the
button to indicate the electrode pads were properly applied during
PADS
ELECTRODE
the training scenario.
Refer to page 10 for information about using the
ELECTRODE PADS
button.
22 Training electrode pads Training electrode pads to be applied to the training manikin.
23 Blue plastic Blue plastic liner that protects the adhesive until the electrode pads are
used.
24 Training shock button Pressing this red flashing button simulates the delivery of a shock by the
defibrillator semiautomatic model. (The trainee cannot deliver a simulated shock unless a Trainer voice prompt instructs them to do so.)
When training simulates the defibrillator automatic model, cover this button with the yellow shock button cover.
25 Electrode connector Connects the electrode pads to the Trainer. The connection is not
functional—simulated connection/disconnection is initiated by means of the remote control.
6 LIFEPAK CR–T Trainer Operating Instructions
Page 17
Getting Started
Remote Control and Remote Control Cable
The remote control is stored in the remote control compartment (refer to page 4). The dual-function keys are used to configure the Trainer operating settings prior to a training session and to alter simulation during training scenarios. The remote control communicates with the Trainer through signals to the infrared sensor located in the back of the Trainer.
The remote control cable, stored in the storage compartment (refer to pages 3 and 4), can be used as an alternative to the infrared sensor. Refer to Section 2, beginning on page 9, for information about connecting the remote control cable.
Figure 5 shows the remote control and remote control cable with the features identified in Tab l e 5 .
Remote control button functions are described in Section 2, beginning on page 10.
2
1
4
3
5
6
Getting Started
8
7
9
14
10
12
Figure 5 Remote Control and Remote Control Cable
Tab le 5 Remote Control Features (Setup Functions)
11
13
1 Feature not available (F1) 8 Pause button (C2)
2 Electrode pads button (F2) 9 Volume control button (OP) (L)
3 No advised shock button (T1) 10 Green select button
4 CHARGE-PAK button (F3) 11 Red select button
5 Service button (F4) 12 Up arrow button
6 Attention button (T2) 13 Right arrow button
7 Motion button (C1) 14 Remote control cable
LIFEPAK CR–T Trainer Operating Instructions 7
©2002-2007 Medtronic, Inc.
Page 18
Getting Started
INSTALLING AND REPLACING BATTERIES
The following paragraphs provide the procedures for installing or replacing LIFEPAK CR–T AED Trainer and remote control batteries.
Installing and Replacing Trainer Batteries
Four D-cell alkaline batteries provide power to the Trainer. The batteries are installed in a compartment in the bottom of the Trainer. Fresh batteries provide approximately 100 hours of continuous Trainer operation.
To install the batteries:
1 Verify that power to the Trainer has been turned off (using the
2 Carefully turn the Trainer upside down.
3 Remove the battery compartment cover by depressing both latches on the cover and lifting the
cover.
4 Insert four fresh, identical batteries, taking care to align the (+) and (-) terminals correctly. The
correct battery placement is shown in Figure 6.
5 Replace the cover and press down until you hear the latches snap into place.
MASTER POWER switch).
Note: Always replace all of the batteries at the same time. Always use four identical batteries.
Note: For permanent storage (longer than six months), remove the batteries from the battery
compartment. If the Trainer is used infrequently (approximately once every six months), the batteries can remain in the battery compartment.
Installing and Replacing Remote Control Batteries
Two AAA alkaline batteries provide power to the remote control. The batteries are located in a compartment on the bottom of the remote control. (See Figure 6.)
To replace the remote control batteries:
1 Turn the remote control upside down.
2 Open the battery compartment by sliding the battery compartment cover down with firm thumb
pressure.
3 Insert two fresh, identical batteries, taking care to align the (+) and (-) terminals correctly.
4 Replace the battery compartment cover, sliding it up until it snaps securely into place.
+–
D-Cell
-
–+
D-Cell
+–
D-Cell
–+
D-Cell
AAA
AAA
Figure 6 LIFEPAK CR–T AED Trainer and Remote Control Battery Placement
8 LIFEPAK CR–T Trainer Operating Instructions
Page 19
2 USING THE LIFEPAK CR–T AED TRAINER
This section provides information and instructions for using the LIFEPAK CR–T AED Trainer.
Voice Prompts and Tones page 10 Using the Remote Control 10 Preparing for a Training Session 12 Conducting a Training Session 13 Following a Training Session 14
Using the LIFEPAK CR–T AED Trainer
LIFEPAK CR–T Trainer Operating Instructions 9
©2002-2007 Medtronic, Inc.
Page 20
Using the LIFEPAK CR–T AED Trainer
VOICE PROMPTS AND TONES
The LIFEPAK CR–T AED Trainer voice prompts provide the trainee with step-by-step instructions for responding to training scenarios. For a complete list of voice prompts, refer to Appendix A.
USING THE REMOTE CONTROL
The remote control allows the instructor to:
Preset operating settings in a setup configuration (refer to Section 3, beginning on page 15).
Pause LIFEPAK CR–T AED Trainer operation.
Interrupt preset routines with
BATTERY
.
Select any one of the six available shock protocols.
The remote control communicates with the Trainer by using wireless infrared signals or the remote control cable. The signal indicator acknowledges remote control signals received, both with and without the remote control cable.
One red flash indicates that a signal was received from the remote control and that the Trainer is responding accordingly.
Three red flashes indicate that an invalid remote control key was pressed and that Trainer operation is unchanged.
A continuous red light indicates that the Trainer is in pause mode.
To use the remote control without the cable, the instructor must stand behind the Trainer. During use, the instructor must hold the remote control level rather than aiming it directly at the Trainer.
NO SHOCK ADVISED, CHECK PADS FOR GOOD CONTACT, MOTION, and LOW
Remote Control Functions
Remote control button functions are described in Ta bl e 6 . The Trainer is designed for maximum flexibility. Your defibrillator may not have all of the functions identified in Ta b le 6 .
Refer to Section 3, Setup Configuration and Operating Settings or a description of the alternate remote control functions, for example, F1.
Tab le 6 Remote Control Button Functions
Button Function Operation and Description
F1
F2
Feature not available
Simulates electrode pads application
Refer to Section 3, Setup Configuration and Operating Settings.
The
ELECTRODE PADS button is used to indicate whether or not
the electrode pads are properly applied to the manikin.
After the trainee applies the electrode pads to the manikin, press the indicators to solid green. Press the second time to initiate the voice prompt and to change the
ELECTRODE PADS button to change the ELECTRODE PAD
ELECTRODE PADS button a
CHECK PADS FOR GOOD CONTACT
ELECTRODE PAD indicators to
flashing red, indicating that the training electrodes were not applied properly or were disconnected after they were properly applied. Press the
ELECTRODE PADS button a third time after the
training electrodes are properly applied to continue the training scenario.
10 LIFEPAK CR–T Trainer Operating Instructions
Page 21
Using the LIFEPAK CR–T AED Trainer
Tab le 6 Remote Control Button Functions (Continued)
Button Function Operation and Description
T1
F3
F4
T2
Change shock to no shock advised
Simulates the CHARGE-PAK needs to be replaced
Simulates a call for service situation
Attention
The NO SHOCK ADVISED button can be used only during simulated analysis.
Press the a shock protocol to
The simulated OFF status (
RELEASE/ON-OFF
Press the
NO SHOCK ADVISED button to change any shock within
NO SHOCK ADVISED.
CHARGE-PAK button can be used only when the Trainer is in
MASTER POWER switch ON, LID
button OFF).
CHARGE-PAK button once to activate the CHARGE-PAK
indicator in the readiness display.
The
CHARGE-PAK indicator automatically appears on the
readiness display after simulating a patient use.
To clear the
CHARGE-PAK indicator from the readiness display,
remove and replace the CHARGE-PAK. The instructor can also clear the indicator by pressing the button a second time.
The
WRENCH button can be used only when the Trainer is in
simulated OFF status.
Press the
WRENCH button once to activate the wrench indicator
in the readiness display.
To clear the
WRENCH indicator from the readiness display, press
the button a second time.
ATTENTION button can be used when the Trainer is ON and
The when it is in simulated OFF status.
Press the
ATTENTION button once to activate the ATTENTION
indicator in the readiness display. This simulates a low internal battery condition (simulating a situation in which the actual defibrillator would have a minimum of 6 shocks or 42 minutes of monitoring time available for use).
To clear the press the
ATTENTION indicator from the readiness display,
ATTENTION button a second time.
Using the LIFEPAK CR–T AED Trainer
C1
Start or stop simulated motion
MOTION button is used only during simulated analysis.
The
Press the interrupts simulated analysis and delivers
MOTION button to start simulated motion. The Trainer
MOTION DETECTED
prompts for 10 seconds or until the instructor presses the
MOTION button again.
C2
Pause
PAUSE button suspends the Trainer in its present state and
The causes the LED display to illuminate red until operation is resumed. This allows the instructor to interject other instruction, if necessary. The Trainer will pause for approximately 30 minutes before it turns completely off.
Pressing the
PAUSE button a second time resumes Trainer
operation.
LIFEPAK CR–T Trainer Operating Instructions 11
©2002-2007 Medtronic, Inc.
Page 22
Using the LIFEPAK CR–T AED Trainer
Tab le 6 Remote Control Button Functions (Continued)
Button Function Operation and Description
OP L
Adjust volume of voice prompts
GREEN SELECT button
RED SELECT button
UP ARROW button
Each time the Trainer is turned on, the volume level defaults to the level used during the previous session.
Three volume levels are available (low, medium, high). Press the
VOLUME CONTROL button to select the volume level desired. As
you press the button, the volume increases. View the LED display to observe volume changes. When the highest volume is reached, the next press returns the volume to low.
The
GREEN SELECT button is used to change the shock protocol
and general operating setting options.
When the
GREEN SELECT button is pressed, the LED display
flashes green to indicate setup mode is activated.
Press the
GREEN SELECT button twice to display the shock
protocol.
For more information, refer to Section 3, beginning on page 15.
The
RED SELECT button is used to change CPR Time 1,
CPR Time 2, and the Language Option.
When the
RED SELECT button is pressed, the LED display flashes
red to indicate setup mode is activated.
For more information, refer to Section 3, beginning on page 15.
UP ARROW button is used to select a setting option.
The
The
RIGHT ARROW button
RIGHT ARROW button is used to advance through the
settings.
Using the Remote Control Cable
To use the remote control cable, insert the cable’s right-angle connector into the remote control receptacle. Insert the connector on the opposite end of the cable into the receptacle located on the side of the Trainer (see Figure 2, page 4).
Disconnect the remote control cable by grasping the connector on the side of the Trainer and gently pulling away from the Trainer. Disconnect the right-angle connector from the remote control in the same way.
PREPARING FOR A TRAINING SESSION
1 Connect the electrode packet to the Trainer as follows:
– Press the
– Insert the electrode pads connector into the Trainer cable receptacle until it snaps into place
(see Figure 7).
– Insert the black plastic anchor pin on the electrode packet into the anchor receptacle at the upper
left corner of the Trainer (see Figure 7).
– Ensure that the lower edge of the packet is under the fin on the lower front part of the Trainer.
– After the electrode packet is in place, close the Trainer lid.
LID RELEASE/ON-OFF button to open the Trainer lid.
12 LIFEPAK CR–T Trainer Operating Instructions
Page 23
Using the LIFEPAK CR–T AED Trainer
2 Slide and release the MASTER POWER switch to provide power to the Trainer.
3 Check to confirm that the existing setup configuration is appropriate for the training scenario. If not,
change the setup configuration as desired (refer to Section 3, beginning on page 15 for setup information).
4 If preparing for a training session on the automatic model of the defibrillator, place the yellow plastic
cover, stored in the storage compartment (refer to pages 3 and 4), over the
SHOCK button.
Two Quick Start Reference Cards are provided in Appendix B, beginning on page 33. These cards include a condensed description of the steps to complete when preparing the Trainer for a training scenario. The cards are perforated for easy removal and may be photocopied.
Using the LIFEPAK CR–T AED Trainer
Figure 7 Connecting the Training Electrode Packet to the LIFEPAK CR–T AED Trainer
CONDUCTING A TRAINING SESSION
1 Verify that power to the Trainer is on and the OK indicator appears in the readiness display.
2 Proceed with the training session.
LIFEPAK CR–T Trainer Operating Instructions 13
©2002-2007 Medtronic, Inc.
Page 24
Using the LIFEPAK CR–T AED Trainer
Note: Ensure that the trainee does not touch the adhesive on the QUIK-PAK training electrodes or
allow a training electrode to fold over on itself. Also ensure that the training electrodes are correctly oriented on the manikin as shown on the electrode packet (see Figure 8).
Figure 8 Placing Training Electrodes on the Manikin
Note: If the QUIK-PAK training electrodes accidentally detach from the manikin or disconnect from the Trainer, the press the
CHECK PADS FOR GOOD CONTACT voice prompt will not sound. The instructor must
ELECTRODE PADS button on the remote control to indicate that the electrode pads are
disconnected.
3 Upon ending a training scenario, press the lid release/ON-OFF button or close the Trainer lid to
simulate turning off the defibrillator.
The CHARGE-PAK indicator automatically appears on the
readiness display. To clear this indicator, the trainee must remove and reinsert the CHARGE-PAK or the instructor must press the
CHARGE-PAK button (F3).
FOLLOWING A TRAINING SESSION
1 Remove the training electrodes from the manikin by grasping the electrode pad at the cable end and
peeling it away from the manikin at a 45° angle. Do not remove the training electrodes from the manikin by pulling on the cable.
2 Reassemble the QUIK-PAK training electrode packet for the next training session.
3 Slide and release the
off after 15 minutes of inactivity.)
4 Care for the Trainer as appropriate (refer to Section 4, beginning on page 23).
Note: Always place the Trainer in a location separate from the defibrillator to avoid confusion in real emergencies.
MASTER POWER switch to turn off the Trainer. (The Trainer automatically turns
14 LIFEPAK CR–T Trainer Operating Instructions
Page 25
3 SETUP
This section provides setup configuration instructions for the LIFEPAK CR–T AED Trainer. It also identifies default settings and describes the setting options, including the procedures for changing those setting options.
Setup Configuration and Operating Settings page 16 Changing Options for Operating Settings 18
Setup
LIFEPAK CR–T Trainer Operating Instructions 15
©2002-2007 Medtronic, Inc.
Page 26
Setup
SETUP CONFIGURATION AND OPERATING SETTINGS
Table 7 lists the operating settings for the LIFEPAK CR-T AED Trainer setup configuration, describes the options for each setting, and identifies the preset default for each setting. Your defibrillator may not have all of the functions identified in Table 7.
Tab le 7 Trainer Operating Setting Options
Operating Settings Options/Description Default Setting
SHOCK PROTOCOL
There are six shock protocol options available:
Custom #1
Four protocols (F1 through F4) are preset and cannot be modified during setup.
Two protocols (C1 and C2) can be user-defined or customized by the instructor.
Note regarding Shock Protocol and Stack Shocks settings:
When Stack Shocks is OFF, each shock/no shock decision is followed by an interval of CPR (interval = time).
When Stack Shocks is ON, each series of consecutive shocks will be followed by an interval of CPR (refer to the CPR time settings on page 3-17.
F1 - Fixed #1 One
SHOCK ADVISED followed by three NO SHOCK ADVISED
decisions.
F2 - Fixed #2 Two
SHOCK ADVISED followed by two NO SHOCK ADVISED
decisions.
F3 - Fixed #3 One
F4 - Fixed #4 Four
NO SHOCK ADVISED, followed by one SHOCK ADVISED,
followed by two
NO SHOCK ADVISED decisions.
NO SHOCK ADVISED decisions.
C1 - Custom #1 Defined by instructor (refer to page 3-19). Nine consecutive
SHOCK ADVISED
decisions
C2 - Custom #2 Defined by instructor (refer to page 3-19). Nine consecutive
NO SHOCK ADVISED
decisions
START PROMPT
When the Trainer lid is opened, two beeping tones confirm
Voice prompt on that the Trainer is turned ON and that a training scenario has begun. A
CALL FOR HELP voice prompt can be added
following the beeping tones.
PULSE PROMPTS
Three signs of life voice prompt options are available.
CHECK FOR
PULSE
CHECK FOR PULSE This prompt is for the more advanced trainee to check for a pulse.
CHECK FOR
BREATHING
CHECK FOR SIGNS
OF CIRCULATION
16 LIFEPAK CR–T Trainer Operating Instructions
This prompt is for the lay trainee to check for signs of breathing.
This prompt is for the lay trainee to check for signs of circulation.
Page 27
Setup
Tab le 7 Trainer Operating Setting Options (Continued)
Operating Settings Options/Description Default Setting
AUTOMATIC/ SEMIAUTOMATIC MODEL
The Trainer can be set up to simulate a defibrillator, automatic model (no model (
SHOCK button).
SHOCK button) or semiautomatic
Automatic The Trainer automatically delivers a simulated shock. The shock button cover is stored in the storage compartment (refer to pages 3 and 4).
Semiautomatic The trainee must press the Trainer
SHOCK button to deliver
a simulated shock.
CPR TIME
CPR Time is the time interval allowed between Trainer prompts to perform CPR. Seven CPR time intervals (in seconds) are available for CPR Time 1 and CPR Time 2: 15, 30, 45, 60, 90, 120, 180.
CPR Time 1 With Stack Shocks OFF: Occurs after every shock.
With Stack Shocks ON: Occurs after a series of consecutive shocks (stack) and after the first NSA that follows a shock series.
CPR TIme 2 Occurs after all other NSA conditions.
STACK SHOCKS
Stack Shocks can be set to ON or OFF. OFF *
ON The Trainer follows a traditional stacked shock protocol and delivers up to three consecutive
SHOCK ADVISED, as
necessary, without interposed CPR.
Semiautomatic
120 seconds *
OFF Allows for an increased emphasis on CPR by prompting the
PULSE CHECK
responder to provide CPR after each
Pulse Check can be set to NEVER, AFTER EVERY NSA, AFTER
SECOND NSA
, or ALWAYS.
SHOCK ADVISED.
NEVER *
NEVER Increases the time allowed for CPR by removing the prompt
AFTER EVERY
NSA
AFTER SECOND
NSA
to perform a pulse check after a
NO SHOCK ADVISED.
Prompts to perform a pulse check after every
ADVISED
.
Only prompts to perform a pulse check after the second
NO SHOCK ADVISED and every NO SHOCK ADVISED after that.
SHOCK ADVISED and after a
NO SHOCK
ALWAYS Prompts for a pulse check to be performed after every
SHOCK ADVISED and after every NO SHOCK ADVISED
decision.
LANGUAGE OPTION
This option allows you to select the language spoken for
English
voice prompts. Refer to Ta b l e 1 0 for available settings.
*These default settings are consistent with the 2005 AHA Guidelines for Cardiopulmonary Resuscitation and Emergency Cardiac Care:
• CPR time = 2 minutes
• One-shock strategy, followed by CPR
• Removal of Pulse Check prompt
Setup
LIFEPAK CR–T Trainer Operating Instructions 17
©2002-2007 Medtronic, Inc.
Page 28
Setup
CHANGING OPTIONS FOR OPERATING SETTINGS
The following paragraphs describe how to change the operating settings identified in Ta bl e 7 . Included are paragraphs that discuss using the remote control and LED display when making changes and paragraphs that explain the procedural steps for completing the changes.
Using the Remote Control and the LED Display
To make changes to the operating settings, you must use the remote control (shown previously in
Figure 5, page 7) and the Trainer LED display.
Before you perform any of the procedures for changing operating settings, complete these steps:
1 If the Trainer is completely turned off, press the
provide power to the Trainer.
2 Stand behind the Trainer and hold the remote control level, unless the remote control cable is
connecting the remote control and the Trainer.
3 As you complete the steps in the procedures that follow, view the LED display to ensure that the
correct bars and colors illuminate.
MASTER POWER switch forward and release it to
Note: The LED bars illuminate until you save the setting and continue to the next action.
Figure 9 shows the LED display with its ten LED bars blank. Each bar corresponds to a setting function and illuminates when you initiate a change.
1234567 89 0
Figure 9 LED Display
Procedures for Changing Operating Settings
The following paragraphs provide the procedural steps for changing operating settings. The procedures address Shock Protocol, General Operating Settings (including Start Prompt, Pulse Prompts, and Automatic/Semiautomatic Model), CPR Time, and the Language Option.
Shock Protocol
For most training scenarios, it is easiest to set up the shock protocol before the training session when the Trainer is in simulated OFF status (
Selecting Shock Protocols
To select one of the fixed shock protocols, F1 through F4, or custom shock protocols, C1 or C2:
1 Press the
GREEN SELECT button.
MASTER POWER switch ON, LID RELEASE/ON-OFF button OFF).
2 Refer to Ta bl e 8 to determine which protocol you want to enter and then press the appropriate
button.
3 Press the
GREEN SELECT button to save the selection. The signal sensor flashes, indicating the
selection is saved, and the Trainer enters a simulated OFF status.
18 LIFEPAK CR–T Trainer Operating Instructions
Page 29
Setup
Tab le 8 Shock Protocol Options
Button Shock Protocol Order LED Display*
F1 Fixed #1 SNNN
1234567 8
F2 Fixed #2 SSNN
1234567 8
F3 Fixed #3 NSNN
1234567 8
F4 Fixed #4 NNNN
1234567 8
C1 Custom #1 SSSSSSSSS
1234567 8
C2 Custom #2 NNNNNNNNN
1234567 8
9 0
9 0
9 0
9
9
9 0
0
0
* = Shock (Green) = No Shock (Red) = End Scenario (Blank)
Configuring Shock Protocols
You can change the Custom #1 and Custom #2 shock protocols to any configuration of
NO SHOCK ADVISED prompts, up to nine selections.
SHOCK and
To change Custom #1 and Custom #2 shock protocols:
1 Press the
2 Press either the
3 Press the
GREEN SELECT button.
MOTION button (C1) for Custom #1 or the PAUSE button (C2) for Custom #2.
UP S ARROW button to select shock or no shock. The first bar can be set only to shock or
no shock.
4 Press the
RIGHT X ARROW button to advance to the next bar and press the UP S ARROW button to
select shock or no shock.
5 Repeat Step 4 until you have set the scenario you want to demonstrate and then press the
ARROW button until remaining LED bars turn blank.
6 Press the
GREEN SELECT button to save the change. The signal sensor flashes, indicating the change
UP S
has been saved. The Trainer enters simulated OFF status.
Figure 10 shows the LED displaying the following shock protocol: Bar 1, no shock (red); Bar 2, shock (green); Bar 3, no shock (red); Bar 4, shock (green); Bar 5, end scenario (blank).
Setup
Bar Number
Figure 10 Custom #1 and #2 Options
LIFEPAK CR–T Trainer Operating Instructions 19
©2002-2007 Medtronic, Inc.
1234567890
Page 30
Setup
General Operating Settings
The procedure for changing the general operating settings (Start Prompt, Stack Shocks, Pulse Check, Pulse Prompts, and Automatic/Semiautomatic Model) is the same for each. Table 9 identifies each general operating setting, indicates the LED bar that corresponds to the setting, and describes the options available.
Tab le 9 General Operating Setting Options
Operating Setting Bar Number Options
Automatic/Semiautomatic
1 Green = Semiautomatic; Red = Automatic
Model
Start Prompt 2 Green = Voice ON; Red = Voice OFF
Stack Shocks 3 Green = OFF; Red = ON
Pulse Check 4, 5 Green, Green = NEVER
Red, Red = ALWAYS Yellow, Yellow = AFTER 2ND NSA Green, Yellow = AFTER EVERY NSA
Pulse Prompts 6 Green =
CHECK FOR PULSE
Red = CHECK FOR BREATHING Yellow = CHECK FOR SIGNS OF CIRCULATION
To change the general operating settings:
1 Press the
2 Press the
3 Press the
4 Press the
GREEN SELECT button.
VOLUME button (OP, L).
RIGHT X ARROW button to advance to the desired operating setting (LED bar number).
UP S ARROW button to select the desired option (LED bar color).
5 Repeat steps 3 and 4 until you have changed all of the desired operating settings.
6 After making your changes, press the
GREEN SELECT button to save the selections. The signal sensor
flashes, indicating the selections have been saved. The Trainer enters simulated OFF status.
Figure 11 shows the LED displaying the following option settings:
Bar 1, Automatic/Semiautomatic Model: Semiautomatic (green)
Bar 2, Start Prompt: Voice ON (green)
Bar 3, Stack Shocks ON (red)
Bars 4 and 5, Pulse Check OFF (green, green)
Bar Number
Figure 11 General Operating Setting Options
20 LIFEPAK CR–T Trainer Operating Instructions
1234567890
Page 31
Setup
CPR Time
To change CPR Time 1 or CPR Time 2:
1 Press the
2 Press the
–or–
Press the
3 Press the
illuminates red as you advance to it.)
4 When the bar corresponding to the desired time illuminates red, press the
the selection. The signal sensor flashes, indicating that the selection has been saved. The Trainer enters simulated OFF status.
Figure 12 shows the LED displaying the CPR Time set to 90 seconds.
RED SELECT button.
NO SHOCK ADVISED button (T1) for CPR Time 1 and then proceed with steps 3 and 4.
ATTENTION button (T2) for CPR Time 2 and then proceed with steps 3 and 4.
RIGHT X ARROW button to advance through the possible CPR time intervals. (Each bar
RED SELECT button to save
CPR Time
Bar Number
Figure 12 CPR Time Options
15 30 45 60 90 120 180
1234567890
Language Option
You can set the voice prompts to be delivered in any of up to six languages, depending on the language group ordered with your Trainer. There are nine language groups available. Table 10 lists the language groups and the languages available in each group.
Tab le 1 0 Language Groups
Configuration Bar 1 Bar 2 Bar 3 Bar 4 Bar 5 Bar 6
Group 1 English Norwegian Danish Swedish Finnish
Group 2 English German French Dutch
Group 3 English Polish Hungarian Czech Russian
Group 4 English Japanese Cantonese Mandarin
Group 6 English
Group 7 English Italian Spanish
Group 8 English Iberian
Portuguese
Spanish Icelandic
Setup
Group 9 English Romanian Hebrew Lithuanian Slovak Slovenian
To change the language option:
1 Press the
2 Press the
3 Press the
LIFEPAK CR–T Trainer Operating Instructions 21
©2002-2007 Medtronic, Inc.
RED SELECT button.
VOLUME CONTROL button (OP, L).
RIGHT X ARROW button until the desired bar illuminates red.
Page 32
Setup
4 Press the RED SELECT button to save the selection. The signal sensor flashes, indicating that the
selection has been saved. The Trainer enters simulated OFF status.
Setup
Figure 13 shows an example of a Trainer configured with Group 1, and the desired language is Finnish.
Option
Bar Number
Figure 13 Language Options
123456
1234567890
Resetting Default Settings
To reset all settings to the original factory default, hold the MASTER POWER switch to ON for at least five seconds. Four yellow bars will appear when the reset is complete.
22 LIFEPAK CR–T Trainer Operating Instructions
Page 33
4 CARING FOR THE LIFEPAK CR–T AED TRAINER
Caring for the LIFEPAK CR–T AED Trainer
This section provides information about caring for the LIFEPAK CR–T AED Trainer.
Replacing Batteries page 24 Troubleshooting 24 Cleaning 26 Product Recycling Information 26 Warranty 26 Accessories 27 Specifications 27
LIFEPAK CR–T Trainer Operating Instructions 23
©2002-2007 Medtronic, Inc.
Page 34
Caring for the LIFEPAK CR–T AED Trainer
REPLACING BATTERIES
When batteries for the LIFEPAK CR–T AED Trainer or the remote control become depleted or damaged, they should be replaced immediately. Refer to Section 1, page 8 for procedures detailing the steps for installing and replacing batteries in the Trainer and the remote control.
TROUBLESHOOTING
Ta bl e s 11 through 15 provide checklists for troubleshooting LIFEPAK CR–T AED Trainer operating problems. If you cannot correct a problem, discontinue using the Trainer until the problem is resolved. Contact your local Medtronic representative.
Note: Do not disassemble the Trainer. There are no operator-serviceable components and operator disassembly may void the warranty.
Tab le 1 1 Voice Prompts and Tones Troubleshooting
Observation Possible Cause Corrective Action
Low volume; Trainer remains functional
D-cell batteries are low
Speaker grill partially
Replace D-cell batteries.
Remove obstruction.
covered
Three beeping tones sound when the Trainer is turned OFF
No sound; Trainer remains functional
D-cell batteries are low • Replace D-cell batteries immediately.
Speaker grill covered
Speaker damaged
Remove blockage.
Contact your local Medtronic representative.
No sound; Trainer does not function
D-cell batteries depleted or damaged
Trainer damaged
Replace D-cell batteries.
Contact your local Medtronic representative.
Tab le 1 2 LED Display Troubleshooting
Observation Possible Cause Corrective Action
Intensity is fading (or inconsistent) • Low battery Replace D-cell batteries.
Unlit when pressed; Trainer does not function
MASTER POWER switch
D-cell batteries depleted or damaged
Trainer damaged
Replace D-cell batteries.
Contact your local Medtronic representative.
LED display scrolls yellow light back and forth
Trainer detects interference
Remote control operation not
Check for possible obstruction.
Install remote control cable to improve connection.
appropriate
24 LIFEPAK CR–T Trainer Operating Instructions
Page 35
Caring for the LIFEPAK CR–T AED Trainer
Tab le 1 3 Setup Options Troubleshooting
Observation Possible Cause Corrective Action
Caring for the LIFEPAK CR–T AED Trainer
Voice prompts are out of sequence • Incorrect operating
setting options
Review setup procedures in
Section 3, beginning on page 15.
selected
Voice prompts are incomplete Trainer damaged Contact your local Medtronic
representative.
CONNECT ELECTRODES sequence
repeats after training electrodes are connected to the Trainer
Instructor did not press remote control to begin scenario sequence
Damaged remote control
Infrared sensor not working properly
Press the ELECTRODE PADS button once on the remote control to initiate voice prompts.
Contact your local Medtronic representative.
Clean or clear the signal sensor, if obstructed.
Install remote control cable.
Contact your local Medtronic representative.
Tab le 1 4 Remote Control Troubleshooting
Observation Possible Cause Corrective Action
Trainer does not respond to remote control button presses
Infrared sensor is not receiving remote control signal
Reposition remote control relative to the Trainer.
Remove anything blocking the signal sensor.
Install remote control cable.
D-cell batteries
Replace D-cell batteries.
depleted or damaged
Remote control
Replace remote control.
damaged
Remote control
Replace remote control batteries.
batteries low
Tab le 1 5 Training Electrodes Troubleshooting
Observation Possible Cause Corrective Action
Training electrode pads do not adhere to manikin
Training electrode connector will not plug into the Trainer
LIFEPAK CR–T Trainer Operating Instructions 25
©2002-2007 Medtronic, Inc.
Adhesive contaminated
Training electrode connector damaged
Trainer electrode receptacle damaged
Replace training electrode packet.
Replace training electrode packet.
Contact your local Medtronic representative.
Page 36
Caring for the LIFEPAK CR–T AED Trainer
CLEANING
Clean the Trainer, remote control, and the remote control cable by wiping surfaces with one of the following:
A lint-free, slightly damp cloth moistened with mild soap and water.
A lint-free, slightly damp cloth moistened with denatured alcohol.
Observe the following precautions when cleaning:
Do not immerse or soak the Trainer, remote control, or the cable.
Do not use bleach, bleach dilution, or phenolic compounds.
Do not steam or gas sterilize.
PRODUCT RECYCLING INFORMATION
Recycle the Trainer and its accessories at the end of their useful life.
Recycling Assistance
Items should be recycled according to national and local regulations. Contact your local Medtronic representative for assistance.
Preparation
Items should be clean and contaminant-free prior to being recycled.
Recycling of Training Electrodes
After training electrodes have reached the end of their useful life, follow your local clinical procedures for recycling.
Packaging
Packaging should be recycled according to national and local regulations.
WARRANTY
In the USA, call 1.800.442.1142. Outside the USA, contact your local Medtronic representative for warranty information.
26 LIFEPAK CR–T Trainer Operating Instructions
Page 37
Caring for the LIFEPAK CR–T AED Trainer
ACCESSORIES
The following accessories are available for the LIFEPAK CR–T AED Trainer:
Tab le 1 6 LIFEPAK CR–T AED Trainer Accessories
Accessory Part Number
QUIK-PAK Training Electrode Packet 3201805
Shock Button Cover 3201801
Remote Control and Cable 3201799
Serial Cable 3206630
CHARGE-PAK Trainer 3201800
Quick Reference Card 3201508
Soft Case 3201803
Soft Case Cards 3201792
Caring for the LIFEPAK CR–T AED Trainer
International Lid Labels 3201755
SPECIFICATIONS
Table 17 lists the specifications for the LIFEPAK CR–T AED Trainer. Ta bl e 1 8 lists the specifications for the remote control.
Tab le 1 7 LIFEPAK CR–T AED Trainer Specifications
General Device designed for low impact, low stress environments. For use in
temperature-controlled areas protected from precipitation. Capable of withstanding heavy usage and prolonged periods of operation.
Physical Characteristics Height: 10.7 cm (4.2 in.)
Width: 20.3 cm (8.0 in.)
Depth: 24.1 cm (9.5 in.), excluding handle
Weight: 2.1 kg (4.6 lb) with batteries
Batteries 6 volts/4 cells, range 3.4 Vdc ~ 6.5 Vdc
Four non-rechargeable D-cell alkaline batteries. New batteries will provide not less than 100 hours of continuous operation under normal indoor-environment conditions.
Buttons Master Power, lid release/ON-OFF, shock
Speaker Volume output: Hi, Med., Low
Operating Temperature 3° to 38° C (37° to 100° F)
Storage Temperature –10° to 60° C (14° to 140° F) (remove batteries for storage)
Water Resistance Spillage resistance per IEC 601-1, part 44.3
LIFEPAK CR–T Trainer Operating Instructions 27
©2002-2007 Medtronic, Inc.
Page 38
Caring for the LIFEPAK CR–T AED Trainer
Tab le 1 8 Remote Control Specifications
Transmission Wireless infrared signal. Range, 3 meters max
82-inch (2-meter) optional cable attachment
Size 4.57 cm x 3.38 cm x 1.65 cm (1.8 in. x 1.3 in. x .6 in.)
Weight (without batteries) 32 grams (.08 lb)
Batteries 3 volts supplied from two AAA alkaline batteries
Buttons 13-key membrane switch pad labeled for the following functions:
NO SHOCK, SERVICE REQUIRED, ELECTRODE PAD CONTACT, MOTION, REPLACE CHARGE-PAK, CPSS, PAUSE, VOLUME CONTROL, LOW INTERNAL BATTERY, SETUP NAVIGATION BUTTONS
Operating Temperature 3° to 38° C (37° to 100° F)
Storage Temperature –10° to 60° C (14° to 140° F) (remove batteries for storage)
Water Resistance Spillage resistance per IEC 601-1, part 44.3
28 LIFEPAK CR–T Trainer Operating Instructions
Page 39
Appendixes 0
APPENDIX A LIFEPAK CR–T AED TRAINER VOICE PROMPTS AND TONES A
Appendixes
LIFEPAK CR–T Trainer Operating Instructions 29
©2002-2007 Medtronic, Inc.
Page 40
Page 41
LIFEPAK CR–T AED Trainer Voice Prompts and Tones
Tab le 1 9 Voice Prompts and Tones
Voice Prompt To nes
CALL FOR HELP NOW
REMOVE ALL CLOTHING FROM PATIENT’S CHEST
PULL RED HANDLE TO OPEN BAG
LOOK AT PICTURES ON PADS
PEEL ONE PAD OFF BLUE PLASTIC
APPLY PAD TO BARE SKIN EXACTLY AS SHOWN IN THE PICTURE
PRESS PAD FIRMLY
PEEL OTHER PAD OFF BLUE PLASTIC
CHECK PADS FOR GOOD CONTACT
CHECK CONNECTOR
DO NOT TOUCH PATIENT – EVALUATING HEART RHYTHM
MOTION DETECTED
STOP MOTION
STAND BY PREPARING TO SHOCK
EVERYONE CLEAR
After the AED Trainer lid opens, two beeping tones sound. If the start option is set to voice prompts, this voice prompt will sound following the beeping tones.
Appendixes
Two beeping tones sound to simulate heart rhythm analysis.
PRESS FLASHING BUTTON
DO NOT TOUCH PATIENT – DELIVERING SHOCK
Semiautomatic model only. An alarm tone sounds until
SHOCK button is pressed.
the
An alarm tone sounds until the SHOCK button is pressed (semiautomatic model) or until a simulated shock is delivered (automatic model).
SHOCK DELIVERED
SHOCK NOT DELIVERED
NO SHOCK ADVISED
PROVIDE CHEST COMPRESSIONS AND RESCUE BREATHS
CHECK FOR PULSE
IF NO PULSE, PROVIDE CHEST COMPRESSIONS AND RESCUE BREATHS
LIFEPAK CR–T Trainer Operating Instructions 31
©2002-2007 Medtronic, Inc.
Page 42
LIFEPAK CR–T AED Trainer Voice Prompts and Tones
Tab le 1 9 Voice Prompts and Tones (Continued)
Voice Prompt To nes
CHECK FOR BREATHING
IF NOT BREATHING, PROVIDE CHEST COMPRESSIONS AND RESCUE BREATHS
CHECK FOR SIGNS OF CIRCULATION
IF NO SIGNS OF CIRCULATION, PROVIDE CHEST COMPRESSIONS AND RESCUE BREATHS
CONTINUE CARE
32 LIFEPAK CR–T Trainer Operating Instructions
Page 43
APPENDIX B LIFEPAK CR–T AED TRAINER QUICK START REFERENCE CARD B
Appendixes
LIFEPAK CR–T Trainer Operating Instructions 33
©2002-2007 Medtronic, Inc.
Page 44
Page 45
LIFEPAK® CR–T AED TRAINER Quick Start Reference Card
Quick Start Setup Procedure
1 Turn on the Trainer using the MASTER POWER switch.
2 Press the
GREEN SELECT button on the remote control.
3 Select the desired shock protocol (the default setting is C1) using the buttons referenced in the following table.
4 Press the
GREEN SELECT button again to confirm shock protocol. The Trainer enters simulated OFF status.
5 You are now ready to begin a training session.
6 When the electrode pads are properly placed on the manikin, press the
ELECTRODE PADS button (F2) on the
remote control.
7 Use the remote control to create troubleshooting situations. Refer to the opposite side of this card for details.
Remote Control Button AED Trainer Shock Protocol Shock Order*
F1
F2
F3
F4
C1**
C2**
LIFEPAK® CR–T AED TRAINER Quick Start Reference Card
Quick Start Setup Procedure
Fixed #1 SNNN
Fixed #2 SSNN
Fixed #3 NSNN
Fixed #4 NNNN
Custom #1 SSSSSSSSS
Custom #2 NNNNNNNNN
* S: Shock Advised (green LED)
N: No Shock Advised (red LED)
** Custom settings, C1 and C2, may be changed
(refer to operating instructions).
1 Turn on the Trainer using the MASTER POWER switch.
2 Press the
GREEN SELECT button on the remote control.
3 Select the desired shock protocol (the default setting is C1) using the buttons referenced in the following table.
4 Press the
GREEN SELECT button again to confirm shock protocol. The Trainer enters simulated OFF status.
5 You are now ready to begin a training session.
6 When the electrode pads are properly placed on the manikin, press the
ELECTRODE PADS button (F2) on the
remote control.
7 Use the remote control to create troubleshooting situations. Refer to the opposite side of this card for details.
Remote Control Button AED Trainer Shock Protocol Shock Order*
F1
F2
F3
F4
C1**
C2**
Fixed #1 SNNN
Fixed #2 SSNN
Fixed #3 NSNN
Fixed #4 NNNN
Custom #1 SSSSSSSSS
Custom #2 NNNNNNNNN
* S: Shock Advised (green LED)
N: No Shock Advised (red LED)
** Custom settings, C1 and C2, may be changed
(refer to operating instructions).
Page 46
LIFEPAK® CR–T AED TRAINER Quick Start Reference Card
Button Function Button Function
Press to change SHOCK ADVISED to NO SHOCK
ADVISED
. Function is active only during simulated
analysis.
Press to advance the training electrodes to ON or OFF.
Press once to activate the indicator in the readiness display. To clear, press the button again.
Press any time to suspend device operation. Press again to resume operation at point of interruption.
Press to advance
MOTION DETECTED to ON or OFF.
This feature is active during simulated analysis only.
Press to change the speaker volume. The volume level scrolls up, and returns to lowest level after the highest is reached.
Press to activate
WRENCH indicator on the readiness
display (active only when Trainer is in simulated OFF
Press to enter the GREEN SELECT button operating settings.
status). To clear, press the button again.
Press to activate
CHARGE-PAK indicator on the
readiness display (active only when Trainer is in simulated OFF status). To clear, press the button again.
Feature not available
Press to enter the RED SELECT button operating settings.
Press the to the next operating setting.
Press the desired option.
ATTENTION
RIGHT ARROW button to advance
UP ARROW button to select the
LIFEPAK® CR–T AED TRAINER Quick Start Reference Card
Button Function Button Function
Press to change SHOCK ADVISED to NO SHOCK
ADVISED
. Function is active only during simulated
analysis.
Press to advance the training electrodes to ON or OFF.
Press once to activate the indicator in the readiness display. To clear, press the button again.
Press any time to suspend device operation. Press again to resume operation at point of interruption.
Press to advance
MOTION DETECTED to ON or OFF.
This feature is active during simulated analysis only.
Press to change the speaker volume. The volume level scrolls up, and returns to lowest level after the highest is reached.
Press to activate
WRENCH indicator on the readiness
display (active only when Trainer is in simulated OFF
Press to enter the GREEN SELECT button operating settings.
status). To clear, press the button again.
Press to activate
CHARGE-PAK indicator on the
readiness display (active only when Trainer is in simulated OFF status). To clear, press the button again.
Feature not available.
Press to enter the RED SELECT button operating settings.
Press the to the next operating setting.
Press the desired option.
ATTENTION
RIGHT ARROW button to advance
UP ARROW button to select the
Page 47
APPENDIX C DECLARATION OF CONFORMITY C
Appendixes
LIFEPAK CR–T Trainer Operating Instructions 37
©2002-2007 Medtronic, Inc.
Page 48
Page 49
Declaration of Conformity
Appendixes
LIFEPAK CR–T Trainer Operating Instructions 39
©2002-2007 Medtronic, Inc.
Page 50
Page 51
INDEX
A
Attention Button 7, 11 Attention Indicator 3 Automatic Defibrillator operating setting 17 Automatic/Semiautomatic Shock 17, 20
B
Back View of Trainer 4 Battery
Compartment for 4 Replacement of 24
Bottom View of Trainer 3 Buttons
Attention 7, 11 CHARGE-PAK 7, 11 Electrode Pads 6, 7 Green Select 7, 12 Lid Release/ON-OFF 4 Motion 7, 11 No Shock Advised 7, 11 Pause 7, 11 Red Select 7, 12 Remote Control 7, 10–12 Right Arrow 7, 12 Service 7, 11 Shock 6 Up Arrow 7, 12 Volume Control 7, 12 Wrench 7, 11
C
Carrying Handle 3 Changing
Custom Shock Protocols 19 Operating Settings 18
CHARGE-PAK
Button location 7 Button operation 11 Indicator location 3
Trainer description 4 Cleaning the AED Trainer 26 Clearing the CHARGE-PAK Indicator 14
Conducting a Training Session 13 Connecting the Training Electrode Packet 12 Controls, Indicators, and Labels 3–4 CPR Time 17
D
Default Operating Settings 16–17 Default Settings, resetting 22
E
Electrode
Connector location 6 Electrode (QUIK-PAK) Packet
Anchor Pin location 5
Connecting to AED Trainer 12
Release Handle location 5
Removing from manikin following a session 14
Index
LIFEPAK CR–T Trainer Operating Instructions 41
©2002-2007 Medtronic, Inc.
Page 52
Index
Electrode Pads
Button location 7 Button operation 10 Indicators description 6
F
Features inside the AED Trainer 5–6
G
General Operating Settings
Automatic/Semiautomatic Shock 17, 20 CPR Time 17 Pulse Check 17, 20 Pulse Prompts 16, 20 Shock Protocol 16 Stack Shocks 17, 20 Start Prompt 16, 20
Green Select Button
Location 7 Operation 12
H
Handle location 3 Handle, carrying 3
I
Inside Features 5, 6 Installing and Replacing Batteries 8, 24 Installing and Replacing Remote Control Batteries 8
L
Language Option 17, 21 LED
Display location 4 Display operation 18–21
Troubleshooting 24 Lid 3 Lid Release/ON-OFF Button 4 LIFEPAK CR–T AED Trainer
Accessories for 27
Back View of 4
Bottom View of 3
Cleaning the 26
Controls, Indicators, and Labels 3–4
Features inside of 5–6
Installing/Replacing Batteries in 8, 24
Operating/Default Settings of 16–17
Remote Control/Remote Control Cable features
7 Specifications for 27 Storing the 2, 14 Troubleshooting 24–25 Warranty 26
M
Master Power Switch location 4 Motion Button
Location 7 Operation 11
N
No Shock Advised Button
Location 7 Operation 11
O
OK Indicator location 3 Operating Settings Options 18 Operating Settings/Defaults
Automatic/Semiautomatic Shock 17, 20 CPR Time 17 Language option 17 Pulse Check 17, 20 Pulse Prompts 16, 20 Shock Protocol 16 Stack Shocks 17, 20 Start Prompt 16, 20
Outside Controls, Indicators, and Labels 3
P
Pause Button
Location 7
Operation 11 Power Switch, Master 4 Preparing for a Training Session 12 Product Recycling 26
Pulse Check 17, 20 Pulse Prompts 16, 20
Q
Quick Reference Card 35 QUIK-PAK Training Electrodes location 5
R
Readiness Display location 3 Red Select Button
Location 7
Operation 12 Remote Control
Buttons functions 10–12
Buttons locations 7
Cable connection description 7
Cable connection location 4
Cable connection operation 12
Communication description 10
Functions 10–12
Operation 10
Signal Sensor location 4
Specifications 28
Storage compartment location 4
Troubleshooting 25
42 LIFEPAK CR–T Trainer Operating Instructions
Page 53
Index
Replacing Batteries 24 Resetting Default Settings 22 Right Arrow Button
Location 7 Operation 12
S
Semiautomatic Defibrillator operating setting 17 Service Button
Location 7 Operation 11
Setting
CPR Time 21 General Operation 20 Language Option 21 Shock protocol 18
Setup Configuration and Operating Settings 16–17 Setup Options Troubleshooting 25 Shock
Button cover placement 13 Button location 6 Custom protocols 19 Fixed protocols 18
Protocol Operation 16, 18–19 Side View of Trainer 4 Speaker location 6 Specifications 27
Stack Shocks 17, 20 Start Prompt 16, 20 Storing the Trainer following Training Session 14
Using the Remote Control 10 Using the Remote Control Cable 12
V
Voice Prompts and Tones
Description 31 Troubleshooting 24
Volume Control Button
Location 7 Operation 12
W
Warranty 26 Wrench
Button 11 Indicator 3
T
Top View of Trainer 3 Training Electrode Placement 14 Training Electrodes (QUIK-PAK)
Connecting for Training Session 13
Connector location 6
Pads location (packet closed) 5
Pads location (packet open) 6
Placing (electrodes) during Training Session 13
Removing (electrodes) following Training
Session 14
Troubleshooting 25 Training Sessions
Conducting a 13
Following a 14
Preparing for 12 Training Shock Button location 6
Turning On the Trainer 13
U
Unpacking and Inspecting the Trainer 2 Up Arrow Button
Location 7
Operation 12 Use By Date
Location 5
Operation 5
Index
LIFEPAK CR–T Trainer Operating Instructions 43
©2002-2007 Medtronic, Inc.
Page 54
Page 55
Page 56
Medtronic, Inc.
11811 Willows Road NE Redmond, WA 98052-2003 USA Telephone: 425.867.4000 Toll Free (USA only): Fax: 425.867.4121 Internet: www.medtronic-ers.com
800.442.1142
Med tronic Europe S. A.
Rte. du Molliau 31 Case postale 84 1131 Tolochenaz Switzerland Telephone: 41.21.802.7000 Fax: 41.21.802.7900
MIN 3201838-003 / CAT. 26500-001156
Loading...