Caution: Federal Law (USA) restricts this device
to sale by or on the order of a physician (or
properly licensed practitioner).
Page 2
Page 3
Medtronic Kappa® 700 Series
Pacemaker Product Information
Manual
Models KSR701, KSR703, KSR706
3
Page 4
This Product Information Manual is primarily intended as an
implantation manual. For programming information see the
Pacemaker Reference Guide that accompanies the programmer
software. It is primarily intended as a follow-up manual, and contains
further information on therapeutic and diagnostic features,
troubleshooting information, follow-up precautionary information,
and complete reference information.
Device Description 8
Indications and Usage 9
Contraindications 9
Warnings and Precautions 10
Co-implantation with an Implantable Defibrillator 19
Adverse Events 21
Clinical Studies 25
Chapter 2 - Implanting the Pacemaker 29
Implantation Procedures 30
Implant Documentation 40
Parameter Programming at Implant 41
Medical Therapy Interactions 45
Assistance 48
Chapter 3 - Description 49
Pacing Mode Operations 50
Rate Responsive Pacing 51
Timing Operations 53
Pacing and Sensing Operations 54
Special Therapy Options 62
Chapter 4 - Pacemaker Follow-up 63
Pacemaker Telemetr y 64
Other Operations 67
5
Page 6
Diagnostics 70
General Recommendations 72
Appendix 73
NBG Codes 74
Special Notice 75
Pacemaker Specifications 77
Lead Requirements, Compatibility 78
Radiopaque Identification 79
Emergency Parameter Settings 80
Shipping and Nominal Parameter Settings 81
Electrical Reset Parameter Settings 84
Elective Replacement Indicator 87
Magnet Mode Conditions 87
Longevity Projections 88
Programmable Parameters 90
Nonprogrammable Parameters 97
Temporary Parameters 98
Tel e m e t r y Ma r k e r s98
Electrograms 99
Automatic Diagnostics 99
Clinician-Selected Diagnostics 100
Patient Information 100
Battery and Lead Telemetered Information 101
Battery Parameters 102
Mechanical Dimensions 102
Index 103
6
Page 7
Chapter 1 - Prescribing the Pacemaker
Device Description8
Indications and Usage 9
Contraindications 9
Warnings and Precautions 10
Co-implantation with an Implantable Defibrillator19
Adverse Events21
Clinical Studies25
7
Page 8
Device Description
Device Description
Medtronic Kappa 700 Series pacemakers (KSR700 Series) are
multiprogrammable, rate responsive, implantable pacemakers,
intended for a variety of bradycardia pacing applications. Rate
response is controlled through an activity-based sensor. The
following models are available:
ModelPolarityPrimary Leads
K
SR701Bipolar/UnipolarIS-1
KSR703Bipolar/UnipolarLow-profile 3.2 mm bipolar
K
SR706UnipolarUnipolar 5 or 6 mm
a
IS-1 refers to an International Connector Standard (see Document No. ISO 5841-3;
1992).
SR700 Series pacemakers are programmed using the
K
Medtronic.Vision software Model 9953 and a Medtronic Model 9790
programmer. For programming instructions refer to the Pacemaker Programming Guide that accompanies Medtronic Kappa 700 Series
software.
a
or IS-1
BI
a
BI
8
Page 9
Indications and Usage
Indications and Usage
KSR700 Series pacemakers are indicated for the following:
■
Rate adaptive pacing in patients who may benefit from
increased pacing rates concurrent with increases in activity.
■
Accepted patient conditions warranting chronic cardiac pacing
which include:
– Symptomatic paroxysmal or permanent second or
third-degree AV block.
– Symptomatic bilateral bundle branch block.
– Symptomatic paroxysmal or transient sinus node
dysfunctions with or without associated AV conduction
disorders.
– Bradycardia-tachycardia syndrome to prevent
symptomatic bradycardia or some forms of symptomatic
tachyarrhythmias.
Contraindications
KSR700 Series pacemakers are contraindicated for the following
applications:
■
Asynchronous pacing in the presence (or likelihood) of
competitive paced and intrinsic rhythms.
■
Unipolar pacing for patients with an implanted
cardioverter-defibrillator (ICD) because it may cause
unwanted delivery or inhibition of ICD therapy. See “Coimplantation with an Implantable Defibrillator” on page 19.
9
Page 10
Warnings and Precautions
Warnings and Precautions
■
Rate responsive modes – Do not use rate responsive modes
in those patients who cannot tolerate pacing rates above the
programmed Lower Rate.
■
Single chamber atrial modes – Do not use single chamber
atrial modes in patients with impaired AV nodal conduction
because ventricular capture cannot be assured.
Pacemaker Dependent Patients
■
Diagnostic modes – Never program diagnostic modes (OVO
and OAO) for pacemaker-dependent patients. For such
patients, use the programmer’s inhibit function for brief
interruption of outputs.
■
Electrogram (EGM) of the patient’s intrinsic activity should be
obtained with care since the patient is without pacing support
when using the programmer’s inhibit function.
■
Polarity override – Overriding the bipolar verification prompt
with bipolar polarity when a unipolar lead is connected results in no pacing output. See “Manually Programming Polarity” on
page 57 for further information.
■
A false bipolar pathway on a unipolar lead, a possible
occurrence with bipolar 3.2 mm connector pacemakers, may
result in a loss of output. See the warning on “Automatic
Polarity Configuration” on page 54 for further information.
10
Page 11
Warnings and Precautions
■
Loss of capture during threshold margin test (TMT) at a
20% reduction in amplitude indicates that the stimulation
safety margin is inadequate. Consider increasing the pacing
amplitude and/or pulse width. See “Magnet Operation” on
page 67 for further information on the Threshold Margin Test.
■
Capture Management will not program ventricular outputs
above 5.0 V or 1.0 ms. If the patient needs a pacing output
higher than 5.0 V or 1.0 ms, program Amplitude and Pulse
Width manually. See “Capture Management” on page 60 for
further information.
Medical Therapy
■
Therapeutic Diathermy can cause fibrillation, burning of the
myocardium, and irreversible damage to the pulse generator
due to induced currents.
■
Magnetic resonance imaging of pacemaker patients has
resulted in significant adverse effects. See “Magnetic
Resonance Imaging (MRI)” on page 15 for further information.
Storage and Resterilization
Medtronic pacemakers are intended for single use only. Do not
resterilize and re-implant explanted pacemakers.
The chart below gives recommendations on handling and storing the
package. Medtronic has sterilized the pacemaker with ethylene oxide
prior to shipment. Resterilizing the pacemaker is necessary if the
seal on the sterile package is broken. Resterilization does not affect
the “Use Before” date.
11
Page 12
Warnings and Precautions
Handling and Storage: AcceptableUnacceptable
Store and transport within
Environmental Temperature limits:
°F (- 18°C) to + 131°F (55°C).
0
Note: A full or partial electrical reset
condition may occur at temperatures
below 0°F (- 18°C). See “Electrical
Reset Parameter Settings” on
page 84.
Resterilization: AcceptableUnacceptable
Do not implant the device if it
has been dropped on a hard
surface from a height of
12 inches (30 cm) or more.
Resterilize if the sterile package
seal is broken. Place the device in
an ethylene oxide permeable
package and resterilize with
ethylene oxide. Allow the device to
aerate ethylene oxide residues.
Refer to sterilizer instructions for
details. Use an acceptable method
for determining sterility, such as
biological indicators.
Do not resterilize the device or the
torque wrench using:
Do not exceed 140
17 psi (103 kPa) when sterilizing.
Do not resterilize the device more
than twice.
°F (60°C) or
Lead Evaluation and Lead Connection
■
Connector compatibility – Do not use any lead with this
pacemaker without first verifying connector compatibility.
Using incompatible leads can damage the connector or result
in a leaking or intermittent connection.
■
Pacing and sensing safety margins – Consider lead
maturation when choosing pacing amplitudes, pacing pulse
widths, and sensing levels. See “Manual Programming” on
page 42.
12
Page 13
Warnings and Precautions
■
Hex wrench – Do not use a hex wrench with a blue handle or
a right-angled hex wrench. These wrenches have torque
capabilities greater than is designed for the lead connector.
See “Connecting Leads to the Pacemaker” on page 33 for lead
connection instructions.
Programming and Pacemaker Operation
■
Epicardial leads – Ventricular epicardial leads have not
been determined appropriate for use with the Capture
Management feature. Therefore, Capture Management should
be programmed Off if epicardial leads are implanted with
SR700 Series pacemakers.
K
■
Shipping values – Do not use shipping values for pacing
amplitude and sensitivity without verifying that they provide
adequate safety margins for the patient.
■
Constant current devices – To test the performance of the
lead, Medtronic recommends using a constant voltage device
such as the Medtronic Model 5311B (or equivalent) Pacing
System Analyser (PSA). Medtronic does not recommend using
a constant current device such as the Medtronic Model 5880A
or 5375 External Pacemaker because the Kappa SR
pacemaker has constant voltage output circuits.
■
Elective Replacement Indicator (ERI) – When ERI is set, the
pacemaker must be replaced within three months. See
“Elective Replacement Indicator” on page 69 for more
information.
■
Full electrical reset is indicated by VVI pacing at a rate of
65 ppm without the elective replacement indicator set. See
“Electrical Reset” on page 69 for more information.
■
Lead Monitor – If the Lead Monitor detects out-of-range lead
impedance, investigate lead integrity more thoroughly.
13
Page 14
Warnings and Precautions
Rate Increases
■
Twiddler’s syndrome, i.e., patient manipulation of the device
after implant, may cause the pacing rate to increase
temporarily if the pacemaker is programmed to a rate
responsive mode.
■
Muscle stimulation, e.g., due to unipolar pacing, may result
in pacing at rates up to the Upper Sensor Rate in rate
responsive modes.
Unipolar Sensing
■
Continuous myopotentials cause reversion to asynchronous
operation when sensed in the refractory period. Sensing of
myopotentials is more likely when atrial sensitivity settings of
0.5 through 1.0 mV and ventricular sensitivity settings of
1.0 and 1.4 mV are programmed.
Environmental and Medical Therapy Hazards
Patients should be directed to exercise reasonable caution in
avoidance of devices which generate a strong electric or magnetic
field. If the pacemaker inhibits or reverts to asynchronous operation
at the programmed pacing rate or at the magnet rate while in the
presence of electromagnetic interference (EMI), moving away from
the source or turning it off will allow the pacemaker to return to its
normal mode of operation.
Hospital and Medical Environments
■
Electrosurgical cautery could induce ventricular arrhythmias
and/or fibrillation, or may cause asynchronous or inhibited
pacemaker operation. If use of electrocautery is necessary, the
current path and ground plate should be kept as far away from
the pacemaker and lead as possible. See “Electrosurgical
Cautery” on page 47 for more information.
■
External defibrillation may damage the pacemaker or may
result in temporary and/or permanent myocardial damage at
14
Page 15
Warnings and Precautions
the electrode-tissue interface as well as temporary or
permanent elevated pacing thresholds. Attempt to minimize
current flowing through the pacemaker and lead system by
following these precautions when using external defibrillation
on a pacemaker patient:
– Position defibrillation paddles as far from the pacemaker as
possible (minimum of 5 inches [13 cm]). Attempt to
minimize current flowing through the pacemaker and leads
by positioning the defibrillation paddles perpendicular to
the implanted pacemaker/lead system.
– Use the lowest clinically appropriate energy output (watt
seconds).
– Confirm pacemaker function following any defibrillation.
■
High energy radiation sources such as cobalt 60 or gamma
radiation should not be directed at the pacemaker. If a patient
requires radiation therapy in the vicinity of the pacemaker,
place lead shielding over the device to prevent radiation
damage and confirm its function after treatment.
■
Lithotripsy may permanently damage the pacemaker if the
device is at the focal point of the lithotripsy beam. If lithotripsy
must be used, program the pacemaker to a nonrate responsive
mode (VVI/AAI or VOO/AOO) prior to treatment; and keep the
pacemaker at least 1 to 2 inches (2.5 to 5 cm) away from the
focal point of the lithotripsy beam.
■
Magnetic resonance imaging (MRI). Pacemaker patients
subjected to MRI should be closely monitored and
programmed parameters should be verified upon cessation of
MRI. MRI of pacemaker patients should be carefully weighed
against the potential adverse affects. Clinicians should
carefully weigh the decision to use MRI with pacemaker
patients. Limited studies of the effects of MRI on pacemakers
have shown that:
15
Page 16
Warnings and Precautions
– Magnetic and radio frequency (RF) fields produced by MRI
may adversely affect the operation of the pacemaker and
may inhibit pacing output.
– Magnetic fields may activate magnet mode operation and
cause asynchronous pacing.
– Reported1 effects of MRI on pacing include increased
ventricular pacing beyond the rate limit.
■
Radiofrequency ablation procedure in a patient with a
SR700 Series pacemaker may cause any of the following:
K
– Asynchronous pacing above or below the programmed
rate.
– Reversion to an asynchronous operation.
– Pacemaker electrical reset.
– Premature triggering of the elective replacement indicator.
RF ablation risks may be minimized by:
1. Programming a non-rate responsive, asynchronous pacing
mode prior to the RF ablation procedure.
2. Avoiding direct contact between the ablation catheter and the
implanted lead or pacemaker.
3. Positioning the ground plate so that the current pathway does
not pass through or near the pacemaker system, i.e., place the
ground plate under the patient's buttocks or legs.
4. Having a Medtronic programmer available for temporary
pacing.
5. Having defibrillation equipment available.
1
Holmes, Hayes, Gray, et al. The effects of magnetic resonance imaging on implantable
pulse generators. PA C E. 1986; 9 (3): 360-370.
16
Page 17
Warnings and Precautions
Home and Occupational Environments
■
High voltage power transmission lines may generate
enough EMI to interfere with pacemaker operation if
approached too closely.
■
Communication equipment such as microwave transmitters,
linear power amplifiers, or high-power amateur transmitters
may generate enough EMI to interfere with pacemaker
operation if approached too closely.
■
Commercial electrical equipment such as arc welders,
induction furnaces, or resistance welders may generate
enough EMI to interfere with pacemaker operation if
approached too closely.
■
Home appliances which are in good working order and
properly grounded do not usually produce enough EMI to
interfere with pacemaker operation. There are reports of
pacemaker disturbances caused by electric hand tools or
electric razors used directly over the pacemaker implant site.
■
Electronic article surveillance (EAS) equipment such as
retail theft prevention systems may interact with pacemakers.
Patients should be advised to walk directly through and not to
remain near an EAS system longer than is necessary.
Cellular Phones
SR700 Series pacemakers have been tested to the frequency
K
ranges used by the cellular phones included in Table 1. Based on this
testing, these pacemakers should not be affected by the normal
operation of such cellular phones.
17
Page 18
Warnings and Precautions
These pacemakers contain a filter that allows usage, without
interaction, of all cellular phones having one of the transmission
technologies listed in Table 1. These transmission technologies
represent most of the cellular phones in use worldwide. Patients can
contact their local cellular phone service provider to confirm that the
provider uses one of these technologies.
Table 1. Cellular Phone Transmission Technologies
Transmission TechnologyFrequency Range
Analog
FM (Frequency Modulation)824 - 849 MHz
Digital TDMA
North American Standards
TDMA - 11 Hz806 - 821 MHz
NADC
PCS
International Standards
GSM
DCS
a
b
(TDMA - 50 Hz)
c
1900
d
e
1800
824 - 849 MHz
1850 - 1910 MHz
880 - 915 MHz
1710 - 1785 MHz
Digital CDMA
CDMA - DS
a
Time Division Multiple Access
b
North American Digital Cellular
c
Personal Communication System
d
Global System for Mobile Communications
e
Digital Cellular System
f
Code Division Multiple Access - Direct Sequence
f
18
824 - 849 MHz
Page 19
Co-implantation with an Implantable Defibrillator
Co-implantation with an Implantable
Defibrillator
■
An implantable defibrillator may be implanted concomitantly
with a bipolar pacemaker.
– The use of unipolar-only Model KSR706 and the KSR700
Series bipolar models implanted with unipolar leads is
contraindicated for patients having an implantable
defibrillator.
– Follow implant protocol and precautions for pacemaker
and defibrillator lead placement. Ensure the pacemaker is
configured to be compatible with the defibrillator.
Programming Considerations
Note the following programming considerations for patients who have
an implantable defibrillator.
■
Only bipolar pacing should be used with these patients. In
some cases, pacing in the unipolar configuration may cause
the defibrillator either to deliver inappropriate therapy or to
withhold appropriate therapy.
■
Polarity is automatically configured during Implant Detection
(see “Automatic Polarity Configuration” on page 3-14 of the
Product Information Manual). If lead integrity is suspect,
confirmation of the automatically programmed polarities
should be made after completion of Implant Detection in order
to assure that bipolar polarities have been programmed
appropriately.
■
The implantable cardiac defibrillator (ICD) should be turned off
during pacemaker implantation procedures until lead polarities
have been configured and confirmed. This is to prevent
possible back-up unipolar paces from triggering the ICD.
■
Lead Monitor should not be programmed to Adaptive. When a
prevalence of out-of-range lead impedance paces is detected,
19
Page 20
Co-implantation with an Implantable Defibrillator
the monitor automatically reprograms the selected lead(s) to
unipolar polarity. Pacing in the unipolar configuration may
cause the defibrillator either to provoke inappropriate therapy
or to withhold appropriate therapy.
■
Transtelephonic Monitor should be programmed to Off. If it is
programmed On, the pacing polarity is temporarily set to
unipolar when the magnet is applied. Pacing in the unipolar
configuration may cause the defibrillator either to provoke
inappropriate therapy or to withhold appropriate therapy.
■
Although these pacemakers are designed to be compatible
with implantable defibrillators, the potential does exist for a
defibrillation pulse to reset them.
– If a partial electrical reset occurs, these pacemakers
implanted with bipolar leads will retain atrial and ventricular
bipolar pacing polarities.
– If a full electrical reset occurs, these pacemakers
implanted with bipolar leads will reset to Implant Detection.
If lead integrity is suspect, confirmation of bipolar polarity
should be made after completion of Implant Detection.
20
Page 21
Adverse Events
Adverse Events
The Medtronic Kappa 700 Series devices were evaluated in a
multicenter prospective study (43 investigational centers, 15 centers
in the US) of the adaptive features and rate response of the device.
Clinical study of the Medtronic Kappa 700 Series of pacemakers
included 288 devices implanted in 285 patients worldwide.
Note: Clinical studies were not performed on the single chamber
SR700 Series models due to similarity in design and function to the
K
DR700 Series models. The clinical data collected for the KDR700
K
Series models therefore supports the safety and efficacy claims for
SR700 Series models and is included here for reference
the K
purposes.
There were a total of six deaths in the study; all were reviewed and
judged to be non-device related by a clinical events committee
comprising clinical investigators and Medtronic clinical evaluation
managers. Two were attributed to ventricular arrhythmia, one to
respiratory failure, the fourth to respiratory insufficiency due to
chronic obstructive pulmonary disease, the fifth to a mesotelioma,
and the sixth to multi-system organ failure.
Eight devices were explanted: three due to pocket infection, one due
to infection of the electrode, one from lead/connector mismatch, one
patient had a psychosomatic disorder, one patient required the
implant of a dual chamber ICD, and one patient continued with
vasovagal symptoms and the therapy did not provide sufficient
benefit.
Observed Adverse Events
A total of 355 adverse events were reported. The device-related
events (182 events) are listed in descending order of frequency in
Ta bl e 2 .
21
Page 22
Adverse Events
Tab le 2. Adverse Events Reported in Four or More Patients- Complicationsa
(Comps) and Observations
b
(Obs)
All patients implanted (n=288 devices in 285 patients, 133 device years)
c
% of
Patients
with
Comps
(n=285)
Comps
per
Device-
Year
(n=133)
% of
Patients
with Obs
(n=285)
Obs per
Device-
Year
(n=133)
EventTotal
Number
of
Events
(Patients)
Any adverse
355 (168)17.2%0.4552%2.22
events
Any device-related
182 (118)10.9%0.3134%1.06
events
Pain at pocket site32 (31)——10.9%0.24
Other23 (21)1.1%0.026.3%0.15
Inappropriate
11 (11)——3.9%0.08
programming
Atrial lead
11 (10)3.6%0.08——
dislodgment
Programmer/
software anomaly
11 (8)——2.8%0.08
d
Pocket infection7 (6)0.7%0.021.4%0.03
Intermittent
6 (6)0.7%0.021.4%0.03
undersensing
Palpitations6 (6)——2.1%0.05
Pocket hematoma6 (6)0.4%0.011.8%0.04
Ventricular lead
6 (6)2.1%0.05——
dislodgment
Elevated pacing
4 (4)0.7%0.020.7%0.02
thresholds
Syncope4 (4)——1.4%0.03
22
Page 23
Adverse Events
a
Complications included those adverse events which required invasive measures to
correct (e.g., surgical intervention), and were related to the presence of the pacing
system or procedure.
b
Observations included those adverse events which did not require invasive measures
to resolve, and were related to the presence of the pacing system or procedure.
c
Where present, a number in parentheses indicates the number of patients with the
event.
d
Programmer software anomalies observed: screen lock-ups while saving data to
diskette (8); problems printing reports outside of a patient session (2); and an incorrect
parameter setting on a printout (1), which occurred in an earlier version of the
software—software changes were made to eliminate reoccurrence.
The following other adverse events were reported, but occurred in
three or fewer patients: angina pectoris; atrial flutter (paroxysmal)/
atrial fibrillation; bipolar short circuit pathway; chest pain; chest pain
(non-specific); dizziness; dyspnea/shortness of breath; exit block;
failure to capture/loss of capture; false negative capture detection; far
field R-wave sensing; fatigue/tiredness; hypotension; inadequate
lead/pacemaker connection; infection of electrode; lead/connector
mismatch; lead insertion route problem; lead insulation failure;
migration of pulse generator; myopotential interference; other
oversensing; pacemaker mediated tachycardia; pacemaker
syndrome; pectoral muscle stimulation; penetration of myocardium
by lead; phrenic nerve/diaphragm muscle stimulation;
pneumothorax; swelling pocket site; tachycardia (atrial); thrombus
formation at lead; ventricular ectopy; ventricular tachycardia
(non-sustained); ventricular tachycardia (sustained).
23
Page 24
Adverse Events
The following adverse events were deemed not device related
(173 events were reported): angina pectoris; atrial fibrillation; atrial
flutter (paroxysmal); atrial flutter (persistent); atrial tachycardia; chest
pain; chest pain (non-specific); congestive heart failure; dizziness;
dyspnea/shortness of breath; fatigue/tiredness; hypertension;
hypotension; insufficient cardiac output; myocardial infarction
(acute); palpitations; syncope; ventricular ectopy; ventricular
fibrillation; ventricular tachycardia (non-sustained); ventricular
tachycardia (sustained).
Potential Adverse Events
Adverse events (in alphabetical order), including those reported in
Table 2, associated with pacing systems include:
The Medtronic Kappa 700 Series devices were evaluated in a
multicenter prospective study (43 investigational centers, 15 centers
in the US) of the adaptive features and rate response of the device.
Note: Clinical studies were not performed on the single chamber
SR700 Series models due to similarity in design and function to the
K
KDR700 Series models. The clinical data collected for the KDR700
Series models therefore supports the safety and efficacy claims for
SR700 Series models and is included here for reference
the K
purposes.
Methods
This study compared the following features of the Medtronic Kappa
700 Series pacemakers to historical controls:
■
Rate Response
■
Capture Management
■
Automatic Polarity Configuration
■
Sensing Assurance
25
Page 26
Clinical Studies
Patient data were collected at implant, pre-discharge, two weeks,
one month, two and/or three months, and six months post implant.
Patients were evaluated utilizing a modified version of the Minnesota
Pacemaker Response Exercise Protocol (MPREP
1
) at their one
month visit. Evaluation of rate response performance for the
Medtronic Kappa 700 Series pacemaker was conducted using the
Metabolic Chronotropic model described by Wilkoff as applied by
2
. Automatic polarity configuration data were collected at implant.
Kay
Sensing Assurance and Capture Management data were collected at
each follow-up.
Description of Patients
Patients enrolled in the study represented a general dual chamber
pacing population.
Results of the Study
Table 3 summarizes the results of the clinical study. The incidence of
complications was found to be similar to that experienced by similar
devices. The performance of the automatic polarity configuration,
Capture Management, Sensing Assurance, and rate response
features were found to meet study objectives.
The slope of the exercise rate response (1.0 target slope) was less
than 0.65 for 26 of 87 (30%) of patients.
1
Benditt, David G. M, Editor, Rate Adaptive Pacing, Blackwell Scientific Publications,
Boston. 1993: 63-65.
2
Kay, Neal G., “Quantitation of Chronotropic Response: Comparison of Methods for
Rate-Modulating Permanent Pacemakers”, JACC 20(7):1533-41, Dec 1992.
26
Page 27
Clinical Studies
All patients implanted (n=288 devices in 285 patients, 133 device years)
Primary Objectives
Automatic Polarity Configuration (n with loss of output / N leads)
Total Leads0% (0/546)[0%, 0.55%]≤5%
Unipolar0% (0/107)[0%, 2.8%]≤5%
Bipolar0% (0/439)[0%, 0.7%]≤5%
Sensing Assurance
(n with loss of sensing or oversensing / N device years)
Atrial13.5% (18/133)[8.8%, 20.5%]≤35.7%
Ventricular0.8% (1/133)[0.2%, 4.1%]≤9.2%
Capture Management (n with all causes loss of capture / N device years)
Loss of capture5.3% (7/133)[2.6%, 10.5%]≤10.7%
Slope of MPREP rate response at 1 month (n=87 patients)
Mean0.81[0.76, 0.86][0.65, 1.35]
Table 3. Effectiveness Analysis
Percent of
events
% (n/N)
95%
Confidence
interval
Criteria:
Upper
95% CI
The Medtronic Kappa 700 Series pacemaker’s Rate Profile
Optimization (RPO) governs sensor indicated rate (SIR) output.
Figure 1 shows the SIR vs. the Wilkoff predicted heart rate achieved
using the RPO feature during the MPREP tests at 1 month.
27
Page 28
Clinical Studies
All patients reaching Anaerobic Threshold, N=87
Expected (Wilkoff) rate, mean and 95% CI
100
90
80
70
60
50
40
30
SIR (normalized)
20
10
0
0 204060
U
CI
%
5
9
r
pe
p
n
a
e
M
Lower 95% CI
I
S
e
t
c
e
p
x
E
)
7
8
=
N
(
R
MPREP Workload (normalized)
Figure 1. Sensor Indicated Rate (SIR) vs. Expected Rate at One Month
)
f
f
o
k
l
i
W
(
d
80
R
I
S
100
28
Page 29
Chapter 2 - Implanting the Pacemaker
Implantation Procedures 30
Implant Documentation40
Parameter Programming at Implant 41
Medical Therapy Interactions 45
Assistance48
29
Page 30
Implantation Procedures
Implantation Procedures
Testing Leads and Pacemaker
Equipment to Test the Pacemaker and Lead
To test the performance of the lead, Medtronic recommends using a
constant voltage device such as the Medtronic Model 5311B (or
equivalent) Pacing System Analyzer (PSA).
Caution: Do not use constant current devices (such as the
Medtronic external pacemaker Models 5880A, 5375, 5348, or 5346)
to test lead performance. They may damage the pacemaker’s
constant voltage output circuits.
For further procedures on determining thresholds and analyzing
pacemaker operation, consult the PSA technical manual.
Note: Wait at least 15 minutes after implanting screw-in leads
(nonsteroid) before measuring final stimulation thresholds and
intracardiac sensing potentials.
Determining Stimulation Threshold
While testing the lead system prior to implant, Medtronic
recommends taking and verifying threshold measurements in both
the unipolar and bipolar polarities for all of the pacemaker models.
Stimulation thresholds less than 1.0 V at a 0.5 ms pulse width are
recommended for acute ventricular leads. Atrial leads may have
slightly higher stimulation thresholds.
Test for the following configurations:
■
Lead tip to case (for unipolar pacing and sensing).
■
Lead tip to lead ring (for bipolar pacing and sensing).
If the stimulation threshold of a mature chronic lead exceeds 2.5 V at
a pulse width of 0.5 ms, consider replacing the lead.
Note: Low-profile 3.2 mm bipolar and IS-1 BI leads can be
connected to the pacemaker in only one way, i.e., distal (tip)
30
Page 31
Implantation Procedures
electrode to the negative terminal. Because of this, the negative clip
of the PSA connecting cable must be connected to the lead
connector pin for correct threshold measurement. (See PSA
technical manual for more information.)
Determining Lead Impedance
Acute lead impedance or pacing impedance for standard leads
should typically fall between 250 and 1000 ohms, although it may fall
outside this range for short periods of time.
Acute lead impedance for leads designed with impedance at or
greater than 1000 ohms should typically fall between 500 and
2500 ohms. High impedances with these leads are acceptable as
long as stimulation and sensing thresholds are acceptable.
If the lead or pacing impedance remains outside of the
recommended range, reposition or replace the lead.
Determining Intracardiac Sensing Potentials
The intracardiac signal must be of sufficient strength to be sensed by
the pacemaker. Intracardiac signals on chronic leads have a lesser
magnitude and slew rate.
Note: Acceptable stimulation thresholds are not an indication of
adequate sensing potentials.
31
Page 32
Implantation Procedures
Analyzing Pacemaker Operation
Before connecting the leads, check the pacemaker for proper
operation with a Medtronic Model 5311B (or equivalent) Pacing
System Analyzer (PSA). Note that specifications for device operation
apply at body temperature. As a result, some variance may occur if
measurements are taken at room temperature.
During this test, the setscrew(s) in the connector assembly must
contact the PSA’s cable probes. Use the appropriate cable probes to
match the connector ports for K
SR700 Series models.
32
Page 33
Implantation Procedures
Connecting Leads to the Pacemaker
The Model KSR701 connector assembly accepts IS-1 BI leads. The
Model K
IS-1 BI leads. The Model K
or IS-1 leads with the appropriate sizing sleeves.
K
aligned behind one another. (See Figure 2.)
SR703 accepts either low-profile 3.2 mm bipolar leads or
SR706 accepts 5 or 6 mm unipolar leads
SR700 Series pacemakers have a tip and ring connector contact
Tip Electrode
Ring Electrode
Figure 2. Example of Bipolar Connector Assembly for
Model K
SR701
Model KSR701 uses a setscrew to contact the lead tip electrode, but
does not use a setscrew to contact the ring electrode. Model K
SR703
contacts the lead with setscrews at the tip and ring electrodes.
33
Page 34
Implantation Procedures
Connection Procedure
Lead connection consists of three steps:
1. Insert the provided torque wrench into the bisected grommet
as shown in Figure 3.
a. Check that the setscrew is retracted from the connector
port by visual inspection of the opening. If a setscrew is
obstructing the pathway, rotate the setscrew
counterclockwise one or two revolutions until the opening
is no longer obstructed.
Note: Counterclockwise rotation beyond this point may
disengage the setscrew from the connector block.
b. Eliminate the “piston effect” (difficulty in inserting the lead
due to compression of trapped air in the connector port)
when the lead is inserted by leaving the torque wrench in
the distal grommet until the lead is secure. This allows a
pathway for venting trapped air when the lead is inserted.
The rubber connector seal maintains the torque wrench in
position.
34
Page 35
Implantation Procedures
b
a
Figure 3. Connector Port Setscrew Preparation
2. Push the lead into the connector port until the lead pin is
clearly visible in the pin viewing area as indicated by arrows in
Figure 4 and described in the table below.
Note: Lubricating the lead end with sterile water may ease
lead insertion.
Lead-Pin Tip Placement In Pin Viewing Area
SR701IS-1 BIVisible at end of area
Model K
Model K
SR703IS-1 BIJust becomes visible
Low-profile
3.2 mm bipolar
Model K
SR7065 or 6 mm unipolarVisible at end of area
Visible at end of area
35
Page 36
Implantation Procedures
KSR701
SR706
K
SR703
K
(IS-1 BI leads)
Figure 4. Lead Insertion Viewing Area
(with 3.2 mm bipolar leads)
SR703
K
3. Tighten the setscrew by turning clockwise until the torque
wrench clicks as indicated in Figure 5. The setscrew(s) must
be engaged for pacing to begin. Do not pull on the lead until all
setscrews have been tightened.
36
Page 37
Implantation Procedures
KSR706KSR703
SR701
K
Figure 5. Tightening Setscrews
Caution: Do not overtighten the setscrew(s). Do not use blue
handled or right angled hex wrenches, since they have torque
capabilities greater than is designed for this connector. The
torque wrench supplied with the pacemaker is designed to slip
when overtorqued to prevent damage to the socket or the
setscrews. Bending the wrench may break it.
Pacemaker Implantation
Proper placement can help facilitate lead wrap and prevent muscle
stimulation and device migration.
Pacemaker Orientation
The pacemaker may be used in either right or left pectoral implants.
Either side of the shield can face the skin to facilitate excess lead
wrap. However, pacemakers with an insulative coating must be
implanted with the uncoated side facing the patient’s skin.
37
Page 38
Implantation Procedures
Avoiding Muscle Stimulation
If the patient experiences muscle stimulation while being paced in the
unipolar configuration, program the device to lower output settings,
such as a lesser Amplitude and/or narrower Pulse Width, without
compromising the patient’s stimulation safety margin.
Securing the Pacemaker (Suture Hole)
A suture placed through the suture hole located in the connector
assembly area helps secure the pacemaker in the subcutaneous
pocket, thus minimizing post-implant rotation and migration of the
device. Use normal surgical needles to penetrate the suture hole.
Replacing an Implanted Pacemaker
When replacing a pacemaker:
1. Program the pacemaker to a nonrate responsive mode prior to
explantation if previously programmed to a rate responsive
mode. This avoids any potential rate increase while handling
the pacemaker.
2. Insert the proper wrench through each slit in the rubber
grommet and loosen each setscrew by turning
counterclockwise.
3. Gently retract the lead from the pacemaker connector.
4. If the lead pin shows signs of pitting or corrosion, the entire
lead should be discarded and replaced with a new lead to
assure the integrity of the pacing system.
5. Measure stimulation thresholds and sensing potentials.
6. Pacemaker-dependent patients will be without pacing
support when the lead is disconnected. Therefore, have a
temporary external pacing instrument available for use.
38
Page 39
Implantation Procedures
Caution: Electrosurgical cautery could induce ventricular
arrhythmias and/or fibrillation, or may cause asynchronous or
inhibited pacemaker operation. Refer to “Electrosurgical Cautery” on
page 47 for information on electrocautery use.
When replacing a Medtronic pacemaker or a pacemaker from
another manufacturer, the physician may require a lead adaptor kit to
connect the pacemaker to the chronic lead. Use the appropriate hex
wrench or screwdriver to disconnect the chronic lead. Medtronic
provides such wrenches in adaptor kits for use with pacemakers
manufactured by Biotronik, Coratomic, CPI, Intermedics, Pacesetter,
Telectronics/Cordis, and Vitatron. See the Medtronic publication
Pacemaker and Connector Encyclopedia under “Adaptors and
Accessories” for the appropriate adaptor kits.
For questions about lead and pacemaker compatibility, consult your
Medtronic representative, or call Medtronic Technical Services.
Disposal of Pacemaker
When replacing the pacemaker (due to battery depletion or
explanting the pacemaker at the death of the patient who is to be
cremated), return the pacemaker to Medtronic for analysis and
disposal. Medtronic recommends cleaning and sterilizing the
pacemakers before returning. See the back cover for mailing
addresses.
39
Page 40
Implant Documentation
Implant Documentation
Pacemaker Registration
A registration form is included in the shipping package for each
Medtronic implantable pacemaker. Upon completion of the
registration form by the clinician, this form serves as a permanent
record of facts related to the implanted device. A copy of this form
should be returned to Medtronic, where vital information will be
transferred to a wallet-size, plastic-coated pacemaker Identification
Card which will be mailed directly to the patient. A temporary
identification card is included in the shipping package for use by the
patient until the permanent card arrives. Refer to the back cover for
mailing addresses.
Establishing a Patient Record
At the time of implant the clinician should establish a record of the
patient’s pacemaker programmed parameter settings for later
reference. A patient record can give the clinician an accurate account
of the patient’s medical history at subsequent follow-ups when
evaluating the pacemaker’s performance. The registration form
should not serve this purpose since the form is intended for
registering the patient and pacemaker with Medtronic.
40
Page 41
Parameter Programming at Implant
Parameter Programming at Implant
Shipping Parameters
The pacemaker is configured to a set of shipping parameters at the
time of manufacture. At implant the device operates at these settings
unless reprogrammed prior to implant. Refer to“Shipping and
Nominal Parameter Settings” on page 81 for specific shipping
parameter settings.
In the event that the pacemaker is not configured to shipping settings
at implant, the device may have changed to partial or full electrical
reset settings due to extreme temperatures or intense
electromagnetic interference.
Caution: If, at implant, a programmer interrogation indicates the
pacemaker has been reset to VVI, 65 ppm full electrical reset
settings, use the programmer to reset the device and reprogram the
previous mode and pacing operations. If, after a reset, a message
appears on the programmer screen/printout indicating the Date/Time
memory has been altered, contact your Medtronic representative for
further information.
Implant Detection
Implant Detection is a 30-minute period, beginning at lead insertion,
during which the pacemaker verifies that a lead has been connected
by measuring lead impedance. After 30 minutes of continuous lead
connection, the pacemaker completes Implant Detection and
activates the pacemaker’s automatic features mentioned in
“Automatic Operation” on page 41.
Automatic Operation
At the completion of Implant Detection, these features are
automatically enabled or begin adaptation from their shipping
settings.
41
Page 42
Parameter Programming at Implant
■
Operating polarities, which are determined during Implant
Detection, are confirmed at its completion.
■
Mode becomes VVIR (though shipped in the VVIR mode, the
device effectively operates in the VVI mode until Implant
Detection is completed).
■
Rate Profile Optimization is enabled.
■
Capture Management is enabled, and Ventricular Amplitude
and Pulse Width become adaptive.
■
Sensing Assurance is enabled, and Sensitivity becomes
adaptive.
■
Diagnostics are enabled.
If the physician manually programs Implant Detection to Off/
Complete during the 30-minute detect time, the automatic operation
of the above features will be activated at that time. If a programming
session is initiated at any time during Implant Detection, the
pacemaker will reset Implant Detection.
Manual Programming
If pacing rates, pacing outputs, and sensitivities are manually
programmed at implant, the physician should be aware of the
following programming considerations:
Pacing Rates
Program the Lower Rate to a setting that permits the patient to sleep
comfortably or program the Sleep Function On. Program the ADL
Rate to provide adequate cardiac output during typical patient
activity. Program the Upper Sensor Rate to provide cardiac output
that meets the patient’s metabolic demand during exercise without
provoking symptoms (e.g., angina, palpitations, etc.).
Note: For patients with stable angina, program the Upper Sensor
Rate to a value below the rate at which angina occurs.
42
Page 43
Parameter Programming at Implant
Pacing Outputs
Strive for acute ventricular stimulation thresholds of 1.0 V amplitudes
at 0.5 ms pulse widths. Atrial stimulation thresholds may be slightly
higher. Do not use a chronic atrial or ventricular lead if the stimulation
threshold exceeds 2.5 V amplitude at 0.5 ms pulse width.
The clinician should consider programming the pacemaker’s atrial or
ventricular output to allow a 2:1 voltage safety margin above the
measured stimulation threshold values. Capture Management can
also be programmed to provide ongoing ventricular capture
monitoring and adjustment. (Programming to an atrial mode disables
Capture Management.)
Sensitivity
Intracardiac signal amplitudes and slew rates decrease during the
lead maturation process. The clinician should consider 2:1 or 3:1
atrial or ventricular sensitivity safety margins when selecting
sensitivity settings, especially with acute leads. After establishing the
intracardiac sensing thresholds, the clinician may wish to program
atrial or ventricular sensitivity to a more sensitive setting (i.e., a lower
numerical setting than the measured sensing threshold). Weigh this
lower setting against the possibility of oversensing noise and
electromagnetic interference, especially with unipolar sensing.
Note: Adequate intracardiac signals should be present in both the
unipolar and bipolar polarities when using bipolar leads. Measure
sensing potentials in the unipolar polarity (lead tip to case) and
bipolar polarity (lead tip to lead ring) prior to connecting the leads.
Self-Inhibition
Self-inhibition results from the sensing of the pacing output pulse.
This is more likely to occur with high amplitudes, wide pulse widths,
and low sensitivity settings. Reprogramming the Sensitivity to a less
sensitive setting (higher numeric value) will usually resolve the
situation.
43
Page 44
Parameter Programming at Implant
In addition, inhibition can occur when far-field R-waves are sensed in
AAIR and AAI modes. This can be prevented by proper programming
of the Atrial Refractory Period and Atrial Blanking.
Muscle Stimulation with Unipolar Pacing
Under certain circumstances (e.g., high output settings, etc.),
pacemaker-induced muscle stimulation may occur at the pocket site
with the pacemaker programmed to ventricular unipolar pacing.
Pacemaker-induced muscle stimulation may be effectively controlled
and/or eliminated by programming Pulse Width to a narrower setting
or programming a lower Amplitude.
Adapting Other Parameters
In addition to programming those parameters discussed above, the
physician should consider adjusting other nominal parameters based
on the patient’s history, medical condition, lifestyle, and
hemodynamic potential and need.
■
Rate response parameters: Rate Profile Optimization (ADL
Response and Exertion Response), Activity Threshold, Activity
Acceleration, and Activity Deceleration.
■
Refractory and blanking periods: Ventricular Refractory
Period, Atrial Refractory Period, and Atrial Blanking.
In addition, the physician should consider programming Sleep
Function and Single Chamber Hysteresis for other pacing operations
during inactivity.
44
Page 45
Medical Therapy Interactions
Medical Therapy Interactions
This section provides details on possible interactions between
pacemaker therapy and other medical therapies.
Implantable Defibrillators
Some pacemaker patients may also require therapy from an
implantable defibrillator. Only bipolar pacing should be used with
these patients. In some cases, pacing in the unipolar configuration
may cause the defibrillator either to deliver inappropriate therapy or
to withhold appropriate therapy.
Regarding use of K
implantable defibrillator, note the following:
■
The unipolar-only Model KSR706 and the bipolar Models
SR701, KSR703 implanted with a unipolar lead should not be
K
used in patients having an implantable defibrillator.
■
The Models KSR701, KSR703 implanted with a bipolar lead
may be used in patients having an implantable defibrillator.
With these pacemakers, however, polarity is automatically
configured during Implant Detection. (See “Automatic Polarity
Configuration” on page 54.) If lead integrity is suspect,
confirmation of the automatically programmed polarity should
be made after completion of Implant Detection in order to
assure that bipolar polarity has been programmed
appropriately.
■
The implantable cardiac defibrillator (ICD) should be turned off
during pacemaker implantation procedures until lead polarities
have been configured and confirmed. This is to prevent
possible back-up unipolar paces from triggering the ICD.
■
Although these pacemakers are designed to be compatible
with implantable defibrillators, the potential does exist for a
defibrillation pulse to reset them.
SR700 Series models in patients having an
45
Page 46
Medical Therapy Interactions
– If a partial electrical reset occurs, these pacemakers
implanted with a bipolar lead will retain atrial or ventricular
bipolar pacing polarity.
– If a full electrical reset occurs, these pacemakers
implanted with a bipolar lead will reset to Implant
Detection. If lead integrity is suspect, confirmation of
bipolar polarity should be made after completion of Implant
Detection.
■
Follow implant protocol and precautions for pacemaker and
defibrillator lead placement. Ensure the pacemaker is
configured to be compatible with the defibrillator.
Warnings :
■
Lead Monitor should not be programmed to Adaptive. When a
prevalence of out-of-range lead impedance paces is detected,
the monitor automatically reprograms the selected lead(s) to
unipolar polarity. Pacing in the unipolar configuration may
cause the defibrillator either to provoke inappropriate therapy
or to withhold appropriate therapy.
■
Programming Transtelephonic Monitor On is contraindicated
for patients with implantable defibrillators. When it is
programmed On, the pacing polarity is temporarily set to
unipolar when the magnet is applied. Pacing in the unipolar
configuration may either cause the defibrillator to provoke
inappropriate therapy or to withhold appropriate therapy.
46
Page 47
Medical Therapy Interactions
Electrosurgical Cautery
Electrocautery may cause permanent loss of output prior to
explanting the pacemaker and can induce ventricular fibrillation. The
electrocautery tip should never be used within 6 inches (15 cm) of an
implanted pacemaker/lead system.
The use of electrocautery may cause any of the following:
triggered or inhibited output, a temporary pause in output,
permanent loss of output, a rate in excess of the 200 ppm rate
limit, reversion to asynchronous operation, an electrical reset,
or premature triggering of the elective replacement indicator.
Because of these potential complications, alternatives to
electrocautery should be used when available.
Effects to the pacemaker vary considerably based on the type of
electrocautery unit, coagulation and cutting current settings, current
pathway from the cautery tip to the indifferent plate, and the
pacemaker/lead system. Electrocautery units could also interfere
with ECG monitoring equipment. Monitor cardiac activity via blood
pressure or pressure pulses when using electrocautery.
If electrocautery cannot be avoided, follow these steps to minimize
complications:
1. Program the device to the VOO/AOO (non-rate responsive)
mode beforehand and avoid direct contact with the pacemaker
or leads to avoid induction of tachyarrhythmias and/or
fibrillation.
2. Position the ground plate so that the current pathway does not
pass through or near the pacemaker system, i.e., place ground
plate under the patient’s buttocks or legs during abdominal
surgery.
3. Use short, intermittent, and irregular bursts at the lowest
feasible energy levels.
4. Use a bipolar electrocautery system, where possible.
5. Have temporary pacing and defibrillation equipment available.
47
Page 48
Assistance
X-ray and Fluoroscopy
X-ray and fluoroscopy testing of pacemakers similar to the KSR700
Series pacemakers by exposure to diagnostic X-ray or fluoroscopic
radiation has not affected those pacemakers.
Assistance
Assistance Information
Medtronic employs highly trained representatives and engineers
located throughout the world to serve you and, upon request, to
provide training to qualified hospital personnel in the use of
Medtronic products. Medtronic also maintains a professional staff to
provide technical consultation to product users. For medical
consultation, Medtronic can often refer product users to outside
medical consultants with appropriate expertise. For more
information, contact your local Medtronic representative or call or
write Medtronic at the appropriate address or telephone number
listed on the back cover.
48
Page 49
Chapter 3 - Description
Pacing Mode Operations50
Rate Responsive Pacing 51
Timing Operations 53
Pacing and Sensing Operations 54
Special Therapy Options62
49
Page 50
Pacing Mode Operations
Pacing Mode Operations
A brief description of programmable mode operations is given below.
Warnings :
■
Do not use rate responsive modes in those patients who
cannot tolerate pacing rates above the programmed
Lower Rate.
■
Do not use single chamber atrial modes in patients with
impaired AV nodal conduction because ventricular capture
cannot be assured.
VVIR / AAIR Modes
The ventricle is paced at the sensor-indicated rate (VVIR) in the
absence of intrinsic ventricular events. The atrium is paced at the
sensor-indicated rate (AAIR) in the absence of intrinsic atrial events.
VVI / AAI, VVT / AAT Modes
The ventricle is paced at the programmed Lower Rate (VVI) in
the absence of intrinsic ventricular events. The atrium is paced at
the programmed Lower Rate (AAI) in the absence of intrinsic
atrial events.
Sensed ventricular events in the VVT mode trigger pacing pulses.
Sensed atrial events in the AAT mode trigger pacing pulses.
Note: Programmed triggered pacing will not occur faster than
300 ms (200 ppm) from the previous paced event. Temporary
programmed triggered pacing is not limited to 300 ms (200 ppm).
50
Page 51
Rate Responsive Pacing
VOO / VOOR, AOO / AOOR Asynchronous Modes
The VOOR mode paces the ventricle at the sensor-indicated rate,
whereas the VOO mode paces at the programmed Lower Rate. The
AOOR mode paces the atrium at the sensor-indicated rate, whereas
the AOO mode paces at the programmed Lower Rate.
OVO / OAO Diagnostic Modes
The OVO mode senses only intrinsic ventricular events, whereas the
OAO mode senses only intrinsic atrial events.
Warning: These modes should never be programmed for
pacemaker-dependent patients. For such patients, the programmer’s
“Inhibit” function may be used for brief interruption of outputs.
Rate Responsive Pacing
The KSR700 Series pacemakers use activity-based pacing, which
varies the pacing rate in response to the patient’s detected physical
motion. In rate responsive modes, the pacing rate varies between the
Lower Rate and Upper Sensor Rate limits and is dependent on the
following programmable parameters:
■
Activity Threshold establishes the minimum signal amplitude
above which activity signals are detected by the activity circuit.
■
Activity Acceleration and Deceleration times determine how
quickly the rate will rise and fall as a result of changes in patient
physical activity. Deceleration can also extend the rate fall by
providing up to 20 minutes of rate deceleration based on the
intensity and duration of exercise.
Rate response is initialized and VVIR pacing becomes effective upon
completion of Implant Detection.
51
Page 52
Rate Responsive Pacing
Rate Profile Optimization
Rate Profile Optimization controls how rapidly and to what level the
sensor-indicated rate increases and decreases within submaximal
and maximal rate ranges:
■
Submaximal rates are moderate pacing rates achieved during
typical daily patient activities. These rates are at or near the
ADL (Activities of Daily Living) Rate.
■
Maximal rates are rates at or near the Upper Sensor Rate
achieved during vigorous activities.
Rate response for rates below the submaximal rate range (i.e., at or
near the Lower Rate) are also controlled by Optimization.
Rate response function is optimized in the submaximal and maximal
rate ranges by a daily comparison of data in two rate profiles:
■
Sensor rate profile is an actual rate distribution of the patient’s
averaged sensor-indicated rates.
■
Target Rate Profile is a programmable rate distribution of the
patient’s desired rates. The ADL Response and Exertion
Response parameters define the time spent pacing in the
submaximal and maximal ranges, respectively.
Once each day, the pacemaker evaluates the percent of time spent
pacing in both the submaximal and maximal rate ranges by
comparing the sensor rate profile against the target rate profile. From
this comparison, the pacemaker automatically adjusts rate response
in both rate ranges.
52
Page 53
Timing Operations
Timing Operations
Rates
The KSR700 Series pacemakers have three rates that determine the
patient's pacing rate.
■
The programmable Lower Rate is the minimum pacing rate for
a given mode.
■
ADL Rate (Activities of Daily Living Rate) is the average target
rate that the patient’s heart rate is expected to reach during
moderate exercise.
■
Upper Sensor Rate provides the upper limit for the sensordriven rate during physical activity, particularly during vigorous
exercise.
Refractory and Blanking Periods
The pacemaker has several refractory periods to avoid inappropriate
restarting of certain timing intervals.
■
The programmable Ventricular Refractory Period (VRP)
prevents sensing of T-waves or PVCs. The VRP occurs after
paced, sensed, and refractory sensed ventricular events in
ventricular modes.
The first portion of the VRP is a nonprogrammable Ventricular
Blanking period in which ventricular sensing is disabled
after paced, sensed, and refractory sensed ventricular events.
The Ventricular Blanking period varies dynamically from
50 to 100 ms based on the strength and duration of the
ventricular event.
53
Page 54
Pacing and Sensing Operations
■
The programmable Atrial Refractory Period (ARP) prevents
atrial inhibition due to sensed far-field R waves or noise. The
ARP occurs after paced, sensed, and refractory sensed atrial
events in atrial modes.
The first portion of the ARP is a programmable Atrial Blanking
period that disables atrial sensing after paced, sensed, and
refractory sensed atrial events.
Pacing and Sensing Operations
Polarity
Unipolar polarity uses the lead tip as the active electrode and the
pacemaker case as the common electrode. Bipolar polarity uses the
lead tip as the active electrode and the lead ring as the common
electrode.
With the bipolar Model K
sensing polarities are automatically or manually programmable. The
Model K
SR706 provides unipolar pacing and sensing only.
Automatic Polarity Configuration
With bipolar pacemakers, polarity settings are selected automatically
during the pacemaker’s 30-minute Implant Detection period that
begins at lead insertion. During Implant Detection, with either a
bipolar or unipolar lead implanted, and with the Configure or Adaptive
parameter under Lead Monitor active, the pacemaker continuously
measures impedance as follows:
■
The pacemaker issues a bipolar pace and immediately checks
the pace for high impedance.
– If the pace is within range, it is considered an acceptable
bipolar pace.
– If high impedance is found, the pacemaker immediately
follows the bipolar pace with a backup unipolar pace.
SR701, KSR703 pacemakers, pacing and
54
Page 55
Pacing and Sensing Operations
■
If a unipolar backup pace is issued, the pacemaker checks it
for high impedance.
– If the pace is within range, it is considered an acceptable
unipolar pace.
– If high impedance is found, the pacemaker assumes a lead
is not attached and restarts Implant Detection.
Note: During polarity configuration, the pacemaker also
checks for low impedance paces but does not follow them with
unipolar backup paces. To avoid possible loss of capture due
to continuous low impedance pacing, the pacemaker sets
unipolar polarity when 3 of 16 paces are detected as low
impedance during the first phase of configuration (described
below).
Warning: If the Implant Detection feature identifies a bipolar,
high-impedance path, backup unipolar pacing will be
delivered. This may be inappropriate for patients with an
implantable defibrillator (ICD). See “Notes” on page 3-10.
Polarity configuration for bipolar pacemakers takes place in two
phases:
■
Initial Configuration Phase – Five minutes after lead
connection, if a high impedance unipolar pace has not reset
Implant Detection, the pacemaker issues three asynchronous
paces at magnet rate and determines if they are bipolar or
unipolar.
– Two of three paces must be bipolar for initial pacing and
sensing polarities to be determined bipolar for the
Confirmation Phase.
– Two of three paces must be unipolar for initial pacing and
sensing polarities to be set to unipolar and remain unipolar
during and after Implant Detection.
55
Page 56
Pacing and Sensing Operations
■
Confirmation Phase – If bipolar polarity is determined at five
minutes, the remaining 25 minutes of Implant Detection is used
to confirm bipolar polarity.
– The pacemaker continues to monitor each pace as
described above, providing backup unipolar paces for high
impedance bipolar paces.
– The pacemaker checks for a programmed number of out-
of-range (high or low impedance) paces that indicate
bipolar pacing is in jeopardy.
– The pacemaker confirms bipolar pacing and sensing
polarity at the end of 25 minutes by examining three
asynchronous paces as described above for the Initial
Configuration Phase.
Note: If a prevalence of out-of-range bipolar paces is detected
during the Confirmation Phase, the pacemaker reconfigures
pacing and sensing polarities to unipolar and the
asynchronous paces are not delivered.
Warning: When implanting a Model K
SR703 pacemaker, ensure that
both the tip and ring setscrews are properly engaged and all
electrical contacts are sealed in order to prevent possible electrical
leakage between the tip and ring contacts. Such leakage may cause
the pacemaker to falsely identify a unipolar lead as bipolar, resulting
in a loss of output. For the same reason, ensure that electrical
contacts are sealed when using lead extenders or adaptors with all
bipolar models.
Notes:
■
During Implant Detection, the unipolar-only Model KSR706
pacemaker examines unipolar pacing impedance to determine
lead connection.
■
When implanting a bipolar pacemaker in a patient with an
implantable defibrillator, the physician has the option to
manually program polarities to avoid the possible occurrence
of unipolar backup paces during Implant Detection.
56
Page 57
Pacing and Sensing Operations
Manually Programming Polarity
Pacing and Sensing Polarities for the Models KSR701, KSR703 can
also be manually programmed. However, the Configure selection
under Lead Monitor, which controls automatic configuration, must be
“turned off” by programming the Monitor Only selection instead.
When pacing and/or sensing polarity is manually programmed from
unipolar to bipolar, the pacemaker verifies whether a functioning
bipolar lead is connected by measuring lead impedance. If lead
impedance is outside the accepted programmed impedance range
for bipolar polarity (between 200 and 4000 ohms), the lead is
assumed to be unipolar and pacing and/or sensing polarity remains
set to unipolar.
Warning: If the clinician overrides the bipolar lead verification and
programs bipolar polarity when a unipolar lead is connected to the
pacemaker, no pacing output results.
Lead Monitor
The Lead Monitor feature enables the pacemaker to monitor lead
integrity and adapt bipolar pacing and sensing to unipolar when
bipolar lead integrity is in doubt. It also controls automatic
configuration of lead polarities at implant. Lead Monitor is available in
all pacing modes.
The Lead Monitor feature monitors lead integrity by looking for high
and low impedance on each pace. The Atrial or Ventricular Lead
Monitor can be programmed to operate as follows:
■
Configure – provides automatic polarity configuration of
bipolar pacemakers (bipolar pacemakers are shipped in
Configure).
57
Page 58
Pacing and Sensing Operations
■
Adaptive
– monitors bipolar paces and provides backup unipolar
paces when impedance is high
– switches pacing and sensing polarities from bipolar to
unipolar when a prevalence of out-of-range (high and low
impedance) paces is detected
– provides automatic polarity configuration if the pacemaker
is programmed to Adaptive prior to implant
■
Monitor Only – monitors unipolar and bipolar paces to
determine if they are out of range (high and low impedance).
Unipolar-only Model K
SR706 is shipped in and only
programmable to Monitor Only.
When Lead Monitor is set to Adaptive or Monitor Only, and the
pacemaker determines a lead is out of range, the pacemaker issues
a status warning that appears on the programmer screen at the next
interrogation. However, a status warning is not issued at any time
during the configuration process.
After automatic polarity configuration is complete, Lead Monitor is
automatically set to Monitor Only for bipolar and unipolar
configurations. If, however, the Adaptive setting was programmed
prior to implant, bipolar configurations are set to Adaptive, but
unipolar configurations are set to Monitor Only.
Warning: Lead Monitor should not be programmed to Adaptive for
patients with an implantable defibrillator. When a prevalence of outof-range lead impedance paces is detected, the monitor
automatically reprograms the selected lead(s) to unipolar polarity.
Pacing in the unipolar configuration may cause the defibrillator either
to provoke inappropriate therapy or to withhold appropriate therapy.
Caution: If the Lead Monitor detects out-of-range lead impedance,
investigate lead integrity more thoroughly.
58
Page 59
Pacing and Sensing Operations
Sensitivity
Sensitivity determines the minimum intracardiac signal that the
pacemaker can detect when intrinsic atrial or ventricular events
occur. With K
or manually programmable.
Sensing Assurance, the pacemaker's automatic sensing feature,
adjusts sensitivities within defined limits. At the completion of Implant
Detection, the pacemaker enables Sensing Assurance and begins
continual monitoring of the peak amplitude of sensed signals. In
response to monitoring, it automatically increases or decreases
Sensitivity to maintain an adequate sensing margin with respect to
the patient's sensed P and R waves.
During provocative myopotential testing of the Medtronic Kappa 700
Series pacemakers having Sensing Assurance On, 56% (28/50) of
patients experienced atrial oversensing while 7.3% (4/55)
experienced ventricular oversensing. The atrial oversensing rate is
comparable to that experienced by Medtronic’s Thera devices which
had an atrial oversensing rate of 52.4% (75/143) during provocative
testing. The ventricular oversensing rate of Medtronic Kappa 700
Series pacemakers compared favorably to the ventricular
oversensing rate experienced by Medtronic Thera pacemakers –
44.1% (63/143).
For guidelines regarding manually programming Sensitivity settings,
refer to “Parameter Programming at Implant” on page 41.
Atrial or Ventricular Sensitivity can also be programmed on a
temporary basis.
Note: Atrial Sensitivity with unipolar sensing polarity is restricted to
0.5 mV to limit oversensing of muscle noise or electromagnetic
interference.
SR700 Series pacemakers, sensitivity is automatically
59
Page 60
Pacing and Sensing Operations
Pacing Output
Amplitude and Pulse Width determine the pacing pulse strength
necessary to capture the atrium or ventricle. Amplitude and Pulse
Width are both manually programmable for the atrial and ventricular
channels or automatically programmable for the ventricular channel
through the Capture Management feature. Amplitude and Pulse
Width can also be programmed on a temporary basis. Refer to
“Parameter Programming at Implant” on page 41 for pacing output
guidelines.
Capture Management
The Capture Management feature provides automatic monitoring of
ventricular pacing thresholds and automatic adjustment of Amplitude
and Pulse Width to maintain capture.
When Capture Management is programmed to Monitor Only, the
pacemaker periodically causes paces to be delivered (affecting
pacemaker timing temporarily if necessary). The pacemaker then
monitors the paces by changing first amplitude and then pulse width
to find two points that lie on the strength duration curve that define
the boundary between settings that capture and those that do not.
How often the pacemaker performs this pacing threshold search is
determined by the programmable Capture Test Frequency
parameter. This parameter determines how often the pacing
threshold search will be initiated and provides for retry if the test
is delayed.
60
Page 61
Pacing and Sensing Operations
When Capture Management is programmed to Adaptive, the
pacemaker responds to monitoring by adapting ventricular amplitude
and pulse width using the following programmable parameters:
■
Amplitude Margin and Pulse Width Margin – the
pacemaker- determined threshold multiplied by the selected
safety factor.
■
Minimum Adapted Amplitude and Minimum Adapted
Pulse Width – the lower limit to which Amplitude and Pulse
Width can be set by the pacemaker during adaptation.
■
Acute Phase Days Remaining – time in days during which
the pacemaker will not decrease output settings below the
initially programmed settings. This parameter is used during
the lead maturation period.
Capture Management becomes operational automatically when
Implant Detection is completed.
Warning: Capture Management will not program ventricular outputs
above 5.0 V or 1.0 ms. If the patient needs a pacing output higher
than 5.0 V or 1.0 ms, program Amplitude and Pulse Width manually.
Note: In the event of partial or complete lead dislodgment, Capture
Management may not prevent loss of capture.
61
Page 62
Special Therapy Options
Special Therapy Options
Sleep Function
Sleep Function is a programmable On or Off feature that reduces the
paced rhythm during sleep. It replaces the programmed Lower Rate
with a slower Sleep Rate during a programmable sleep period. At the
programmed Bed Time, the pacemaker gradually lowers the
operating lower rate from the programmed Lower Rate to the
programmed Sleep Rate over a 30 minute period. The Sleep Rate
becomes the operating lower rate until the programmed Wake Time
occurs, when the pacemaker gradually raises the operating lower
rate from the Sleep Rate to the Lower Rate over a 30 minute period.
Note: For rate responsive modes, the Sleep Rate becomes the
minimum rate during the sleep period. However, in the presence of
sensor-detected activity, the sensor-indicated rate overrides the
Sleep Rate.
Single Chamber Hysteresis
Single Chamber Hysteresis allows tracking of an appropriate intrinsic
rhythm below the Lower Rate during extended periods of inactivity
such as sleep. When a nonrefractory sensed event occurs, sensed
events that follow are allowed to occur below the programmed Lower
Rate to the programmed Hysteresis Rate. As long as the intrinsic rate
is above the Hysteresis Rate, pacing is inhibited. When the intrinsic
rate drops to the Hysteresis Rate, escape pacing results and the
programmed Lower Rate returns as the operating lower rate until the
next nonrefractory sensed event occurs.
62
Page 63
Chapter 4 - Pacemaker Follow-up
Pacemaker Telemetry 64
Other Operations 67
Diagnostics 70
General Recommendations 72
63
Page 64
Pacemaker Telemetr y
Pacemaker Telemetry
Pacemaker telemetry is established by placing the programming
head over the implant site and moving the head along the axis shown
in Figure 6 until the head lights indicate telemetry is possible.
Pacemaker telemetry can be used for routine monitoring. A brief
description of the types of telemetry is given below.
Axis
Figure 6. Positioning the Programming Head
Parameter Summary
Upon interrogation by the programmer, the pacemaker telemeters
the programmed parameter settings along with battery status and
device identification.
Note: The parameters may change due to a full or partial electrical
reset, elective replacement indication, Rate Profile Optimization,
automatic Sensitivity, Capture Management, or Lead Monitor
operation. See the appropriate topic for more information.
64
Page 65
Pacemaker Telemetry
Battery and Lead Information
The pacemaker provides Battery and Lead Information for both
measured and calculated values of the pacing output pulse, the lead,
and the battery. The pacemaker’s Battery and Lead Information is
accessed via the programmer.
Telemetry Markers (Extended Markers)
Telemetry markers depict cardiac events that are useful in evaluating
pacemaker operations.
■
Standard Markers (Marker Channel Telemetry) annotate
pacemaker paced, sensed, and refractory sensed events in
any mode.
■
Therapy Trace Markers provide depictions of therapy
interventions that alter timing intervals.
Electrogram (EGM)
The Electrogram (EGM) is a real-time intracardiac waveform
detected by the atrial or ventricular lead. The waveform is
accompanied by standard Marker Channel telemetry annotation.
Warning: An EGM of the patient’s intrinsic activity should be
obtained with care since the patient is without pacing support when
using the programmer’s inhibit function.
Notes:
■
The simulated pacing pulses on the EGM should not be
interpreted as representing the actual timing or amplitude of
paced events. Use an ECG to determine such events.
■
Intracardiac amplitudes measured on the EGM should not be
used to assess the adequacy of the programmed Sensitivity
setting. Use the sensing test instead.
65
Page 66
Pacemaker Telemetr y
Transtelephonic Monitor
The Transtelephonic Monitor is a programmable On or Off feature
intended for use with remote pacemaker monitoring services. When
the Transtelephonic Monitor is programmed On, the Threshold
Margin Test is delayed for five seconds upon magnet application to
enhance communication with transtelephonic equipment. If the
pacing polarity is programmed to bipolar, it is temporarily set to
unipolar to provide improved ECG artifact detection for the remote
monitoring equipment. (The programmed polarity is restored when
the magnet is removed.) When Transtelephonic Monitor is
programmed Off, the Threshold Margin Test is not delayed upon
magnet application, and conventional transtelephonic monitoring
can occur.
Warning: Programming Transtelephonic Monitor On is
contraindicated for patients with implantable defibrillators. When it is
programmed On, the pacing polarity is temporarily set to unipolar
when the magnet is applied. Pacing in the unipolar configuration may
cause the defibrillator either to provoke inappropriate therapy or to
withhold appropriate therapy.
Extended Telemetry
Extended Telemetry provides frequent patient monitoring in the
programmed mode. When Extended Telemetry is programmed On,
the pacemaker automatically transmits the programmed telemetry
type (i.e., Electrogram, Marker Channel telemetry) when the
programmer head or monitoring receiver (with or without a magnet)
is positioned over the pacemaker. Extended Telemetry is
automatically programmed Off after 48 hours or can be programmed
Off via the programmer.
66
Page 67
Other Operations
Other Operations
Magnet Operation
Magnet operation includes an initial capture safety margin test, an
automated self test, and magnet mode operation with telemetry:
■
The Threshold Margin Test (TMT) provides a check for loss of
capture. It is performed at 100 ppm with pulse amplitude
reduced by 20% on the third pulse.
Warning: Loss of capture during TMT may indicate that the
stimulation threshold safety margin is inadequate. Perform a
pacing threshold test and reprogram output to establish a 2:1
voltage safety margin, if necessary.
■
The Fully Automated Self Test (FAST) is a pattern of pacing
pulses that indicates either of the following events has
occurred:
– Lead Monitor has detected an out-of-range lead, and, if
programmed to Adaptive, has switched polarities from
bipolar to unipolar.
– Capture Management has measured unusually high
ventricular thresholds.
Beginning on the cycle following TMT, the pacemaker delivers
two pacing pulses at 85 ppm followed by two pulses at 65 ppm
and then repeats the same sequence once again.
■
Magnet Mode forces the pacemaker’s programmed operation
to an asynchronous mode (VOO or AOO). Magnet Mode rate
is 85 ppm, however, the rate is 65 ppm if an electrical reset
occurs or the elective replacement indicator is set.
67
Page 68
Other Operations
Magnet Operation Using the Transtelephonic Magnet
When a transtelephonic magnet is placed over the pacemaker, a
TMT is automatically initiated. A FAST, if programmed On,
immediately follows the TMT if any of the events mentioned above
that initiate the test have occurred. When the FAST (or TMT when a
FAST is not initiated) is completed, pacing is forced to Magnet Mode
operation until the magnet is removed.
Magnet Operation Using the Programming Head
The programming head does not initiate a TMT when placed over the
pacemaker. The clinician must perform a Magnet Test in order to
obtain a TMT. A FAST, if programmed On, will follow the TMT if
initiated by any of the events described above. When the FAST (or
TMT when a FAST is not initiated) is completed, pacing is forced to
Magnet Mode operation until the Magnet Test is terminated or the
programming head is removed.
FAST Reporting Using the Remote Assistant
The FAST is available with the patient-activated hand-held Remote
Assistant when the Remote Assistant diagnostic is off, and FAST has
been programmed On.
With the Remote Assistant, the FAST results indicating a lead or
threshold problem are reported by two beeps in a low tone, and a
yellow light appears on the underside of the activator. When no
problems exist, the FAST results are reported by three beeps, and a
green light appears instead.
FAST with the Remote Assistant also reports if the ERI has been set
or an electrical reset has occurred (indicated by two beeps and a
yellow light from the activator).
Temporary Parameters
Several pacemaker parameters can be programmed temporarily
during a patient session. These include: Mode, Rate, Amplitude,
Pulse Width, and Sensitivity.
68
Page 69
Other Operations
Electrical Reset
A partial Electrical Reset can occur when the pacemaker is exposed
to cold temperatures, electrocautery, internal or external
defibrillation. Under these conditions, the pacemaker resumes
pacing at the previously programmed mode and preserves many of
the primary pacing parameter settings in operation prior to the partial
reset. See “Electrical Reset Parameter Settings” on page 84 for a
complete list of the preserved settings as well as full Electrical Reset
settings.
If the temperature or electrical disturbance is severe enough, the
pacemaker can reset to full Electrical Reset settings, in which case
the pacemaker reverts to VVI mode at a pacing rate of 65 ppm and
restarts Implant Detection. To restore the pacemaker to its previous
mode and pacing operations, it must be reset and manually
reprogrammed via the programmer.
Note: If after a reset a message appears on the programmer screen
or printout indicating the Date/Time memory has been altered,
contact your Medtronic representative for further information.
Elective Replacement Indicator
The Elective Replacement Indicator (ERI) forewarns when the
battery is nearing depletion. When the pacemaker detects that the
battery voltage has dropped below 2.59 V, the pacemaker sets the
ERI status and reverts to VVI operation at a rate of 65 ppm. Pacing
is maintained, in VVI mode, even in pacemakers previously
programmed to the AAIR, ADIR, AAI, ADI, AAT, AOOR, or
AOO mode.
Caution: When ERI is set and verified, the pacemaker must be
replaced within three months.
High-rate pacing over several years, or with the battery partially
depleted, may trigger the ERI prematurely.
Continuous pacing at high rates (i.e., greater than 100 ppm) and high
output settings (i.e., pulse width greater than 1.0 ms and/or
69
Page 70
Diagnostics
amplitude greater than 5.0 V) may cause the battery to deplete
rapidly and set ERI. While operating at ERI, the battery voltage may
rise above 2.59 V, allowing the pacemaker to be reset and
reprogrammed.
Before and after the use of high-rate pacing, the clinician should
interrogate the pacemaker to determine the battery status and ERI
condition.
Emergency Pacing
Emergency pacing provides VVI pacing at high output settings in
emergency situations for pacemaker-dependent patients. See
“Emergency Parameter Settings” on page 80.
Rate Limit
In the event of a single component failure, the absolute rate limit is
held to 200 ppm. Rate limit is automatically overridden in temporary
single chamber modes for high rate pacing.
Diagnostics
The pacemaker has automatic data summary diagnostics, some of
which have an accompanying clinician-selected diagnostic to collect
further detailed data. At the completion of Implant Detection, the
pacemaker enables diagnostic collection.
High Rate Episode – Automatically collects basic data on
atrial or ventricular high rates and, optionally, clinician-selected
70
Page 71
Diagnostics
detailed data of atrial or ventricular electrogram with a
rate trend.
■
Lead Diagnostics – Chronic Lead Impedance Trend
automatically collects long-term chronic lead impedance data.
The automatic Lead Monitor Counters present the impedance
status of every beat. Chronic lead impedance data is collected
in all pacing modes except for triggered pacing modes (AAT
and VVT). This data is not collected in non-pacing modes
(OAO and OVO).
■
Key Parameter History – Automatically records the status of
key parameters pertaining to the last eight programming
sessions.
■
Sensitivity Trend – Automatically collects the minimum and
maximum sensitivities used to operate automatic sensitivity.
■
Capture Management Trend – Automatically collects
information about the range of ventricular thresholds and
outputs over the life of the device and, optionally,
clinician-selected detailed data reporting the results of pacing
threshold tests.
■
Custom Rate Trend – Clinician-selected collection of heart
rate and percent-paced trends over a beat-to-beat or
24-hour period.
■
Lead Monitor Impedance Trend – Provides trend data on
lead integrity.
■
Sensitivity Detail – Provides a detailed view of Sensing
Assurance operation by recording sensitivity values at 1-hour
intervals.
1
1
If you selected Sensitivity Detail as one of the setup options for Capture Management
Detail, the recording interval is 2 hours rather than 1 hour.
71
Page 72
General Recommendations
■
Remote-Assistant Symptoms – Data for symptom
diagnosis is collected by the clinician-selected atrial or
ventricular beat-to-beat rate trend with optional EGM.
Data collection is initiated by the patient using the Remote
Assistant patient activator. The patient will hear a clicking
sound from the Remote Assistant when initiating a telemetry
link with the pacemaker. If the link is successful, the Remote
Assistant responds with three beeps, and a green light
appears on the underside of the activator. If communication is
unsuccessful, it responds with two beeps in a lower tone, and
a yellow light appears instead.
■
Remote-Assistant Exercise – Records heart rate data for a
trend graph during a particular exercise activity.
General Recommendations
During the first few months after receiving a new pacing system the
patient may require close monitoring. In the U.S.A., the physician
may wish to use the individual state’s or Medicare’s maximum
frequency table for dual chamber pacemakers under Guideline I,
i.e., every two months from the second month through three years
post-implant, and monthly thereafter (Ref. Carriers Manual, Part 3,
Claims Process, Transmittal No. 1051, or a later publication, if
available). Note that Guideline I applies to those pacing systems
(i.e., the pacemaker and the lead used) for which 5 years of clinical
data are not yet available.
72
Page 73
Appendix
Appendix
NBG Codes
Special Notice75
74
73
Page 74
NBG Codes
NBG Codes
The KSR700 Series pacemakers may be described as VVIR
pacemakers using the NBG
1
code (NBG code stands for “The North
American Society of Pacing and Electrophysiology [NASPE] and the
British Pacing and Electrophysiology Group [BPEG] Generic”). The
five-letter code, which supersedes the ICHD Code, describes the
operation of implantable pacemakers. For simplicity, this manual
uses only the first three or four letters, e.g., AAI, VVI, VVIR, etc. The
following chart describes the first four letters of the NBG code.
CHAMBER PACED
V
=
Ventricle
A
=
Atrium
D
=
Dual Chamber
S
=
Single Chamber
O
=
None
CHAMBER SENSED
V
=
Ventricle
A
=
Atrium
D
=
Dual Chamber
S
=
Single Chamber
O
=
None
VVIR
MODE OF RESPONSE
T
=
Triggered
I
=
Inhibited
D
=
Double (Both)
O
=
None
PROGRAMMABLE/
RATE RESPONSE
P
=
Programmable
M
=
Multiprogrammable
C
=
Communicating
R
=
Rate Responsive
O
=
None
1
Bernstein A., et al., The NASPE/BPEG Pulse Generator Code, Pace, 10 (4),
Jul-Aug 1987.
74
Page 75
Special Notice
Special Notice
Medtronic implantable pacemakers and leads are implanted in the
extremely hostile environment of the human body. This environment
places severe limitations on the design and function of the
pacemaker and lead. These limitations unavoidably reduce the
potential performance and longevity of the pacemaker and lead
despite the exercise of due care in design, component selection,
manufacture and testing prior to sale. Reference is hereby made to
published data on predictable failure rates of the pacemaker and lead
and their implantation which have either been furnished by Medtronic
or are otherwise available to physicians. The pacemaker includes an
inseparable power source which will ultimately cease to function due
to exhaustion or premature failure thereby necessitating removal of
the pacemaker. The lead is necessarily very small in diameter and
very flexible, which unavoidably increases the likelihood of breakage
or breach of its insulation. Other reasons for failure of the pacemaker
or lead include, but are not limited to: body rejection phenomena;
change in performance characteristics due to component changes or
failures; unusual physiological variations in patients; lead
displacement; lead fracture; fibrotic tissue formation; elevated
thresholds; medical complications, such as myocardial infarction
occurring at the site of the electrode; erosion of the pacemaker or
lead through body tissue or interference from transmitters, tools,
appliances, instruments, equipment or other devices which use
electricity or electromagnetic wave transmission. Consequently, no
representation or warranty is made that failure or cessation of
function of the pacemaker or lead will not occur, that the body will not
react adversely to the implantation of the pacemaker or lead, that
medical complications will not follow the implantation of the
pacemaker or lead or that the pacemaker and lead will, in all cases,
restore adequate cardiac function.
75
Page 76
Page 77
Pacemaker Specifications
Pacemaker Specifications
Lead Requirements, Compatibility78
Radiopaque Identification 79
Emergency Parameter Settings 80
Shipping and Nominal Parameter Settings 81
Electrical Reset Parameter Settings 84
Elective Replacement Indicator 87
Magnet Mode Conditions 87
Longevity Projections 88
Programmable Parameters90
Nonprogrammable Parameters 97
Temporary Parameters 98
Telemetry Markers 98
Electrograms99
Automatic Diagnostics 99
Clinician-Selected Diagnostics100
Patient Information100
Battery and Lead Telemetered Information101
Battery Parameters 102
Mechanical Dimensions102
77
Page 78
Lead Requirements, Compatibility
Lead Requirements, Compatibility
The chart below lists the available models by polarity type with the
required lead for proper operation. Refer to “Connecting Leads to the
Pacemaker” on page 33.
ModelPolarityPrimary Leads
K
SR701Bipolar/UnipolarIS-1
KSR703Bipolar/UnipolarLow-profile 3.2 mm bipolar
K
SR706UnipolarUnipolar 5 or 6 mm
a
IS-1 refers to an International Connector Standard (see Document No. ISO 5841-3;
1992) whereby pulse generators and leads so designated are assured of meeting the
electrical and mechanical parameters specified in the IS-1 International Standard.
a
or IS-1
BI
a
BI
78
Page 79
Radiopaque Identification
Radiopaque Identification
Listed below are the pacemaker’s radiopaque identifications, which
are used to identify the pacemaker with fluoroscopy. An X-ray
showing the location of the radiopaque for Model K
Figure 7.
ModelsRadiopaque
SR701PHT
K
K
SR703PHW
K
SR706PHU
PHT
SR701 is given in
Figure 7. Model KSR701 X-Ray
79
Page 80
Emergency Parameter Settings
Emergency Parameter Settings
Listed below in Table 4 are the pacemaker’s emergency parameter
settings.
Table 4. Emergency Settings
ModeVVI
Pacing Rate70 ppm
Ventricular
Amplitude7.5 V
Pulse Width1.5 ms
Sensitivity2.8 mV
Pacing PolarityUnipolar
Sensing PolarityUnipolar
Lead MonitorMonitor Only
Capture ManagementOff
Ventricular Refractory Period330 ms
Single Chamber HysteresisOff
80
Page 81
Shipping and Nominal Parameter Settings
Shipping and Nominal Parameter Settings
Listed below in Table 5 are the pacemaker’s shipping and nominal
parameter settings.
Note: “Adaptive” indicates that the parameter value is adapted
during operation. “Unchanged” indicates that the parameter setting is
unchanged by nominal programming.
Table 5. Shipping and Nominal Settings
ParameterShippingMedtronic
a
ModeVVI (VVIR
)VVIR
Lower Rate60 ppm60 ppm
Upper Sensor Rate120 ppm120 ppm
ADL Rate95 ppm95 ppm
Rate Profile OptimizationOn
a
ADL Response3-Mod. Active3-Mod. Active
Exertion Response3-Mod. Frequent3-Mod. Frequent
Activity ThresholdMed. LowUnchanged
Activity Acceleration30 secUnchanged
Activity DecelerationExerciseUnchanged
Nominal
Unchanged
Ventricular Lead
Amplitude3.5 V (Adaptive
Pulse Width0.4 ms (Adaptive
Sensitivity2.8 mV (Adaptive
Sensing AssuranceOn
Pace PolarityConfigure
Sense PolarityConfigure
a
Unipolar
Unipolar
a
c
a
c
a
)3.5V
a
)0.4msb (Adaptive)
a
)2.8mVb (Adaptive)
On
or
or
Unchanged
Unchanged
b
(Adaptive)
81
Page 82
Shipping and Nominal Parameter Settings
Tab le 5. Shipping and Nominal Settings (Continued)
ParameterShippingMedtronic
Lead MonitorConfigured or
Monitor Only
c
Notify if <200 ohmsUnchanged
Notify if >4000 ohms4000 ohms
Monitor Sensitivity88
Capture ManagementAdaptive
a
Amplitude Margin1.5x (times)1.5x (times)
Pulse Width Margin1.5x (times)1.5x (times)
Minimum Adapted
2.5 V2.5 V
Amplitude
Minimum Adapted
0.4 ms0.4 ms
Pulse Width
Capture Test
Day at RestDay at Rest
Frequency
Acute Phase Days
112 daysUnchanged
Remaining
Refractory Period330 ms330 ms
Sleep OffOff
Single Chamber
OffUnchanged
Hysteresis
Implant DetectionOn/RestartUnchanged
Transtelephonic MonitorOffUnchanged
Extended TelemetryOffUnchanged
Extended MarkerStandardUnchanged
FAST IndicatorsOnUnchanged
Nominal
Unchanged
Adaptive
82
Page 83
Shipping and Nominal Parameter Settings
Tab le 5. Shipping and Nominal Settings (Continued)
ParameterShippingMedtronic
Serial NumberFactory SetUnchanged
Clinician-Selected
Diagnostic
a
At the completion of the 30-minute Implant Detection period:
– Mode becomes VVIR
– Rate Profile Optimization is enabled
– Polarities are automatically configured for Models K
– Capture Management is enabled, and Ventricular Amplitude and Pulse Width
become adaptive
– Sensing Assurance is enabled, and Sensitivity becomes adaptive
b
Value from which adaptive adjustment begins when nominals are programmed
c
Unipolar-only Model KSR706
d
Models KSR701 and KSR703l
Capture Management
Detail
Nominal
Unchanged
SR701 and KSR703.
83
Page 84
Electrical Reset Parameter Settings
Electrical Reset Parameter Settings
Listed below in Table 6 are the pacemaker’s full and partial electrical
reset parameter settings.
Note: “Unchanged” indicates that the parameter setting is
unchanged by a reset.
Battery/Lead InformationReplacement message and displayed
battery voltage on programmer
Magnet Mode Conditions
The pacemaker’s magnet mode is VOO or AOO at 85 ppm.
87
Page 88
Longevity Projections
Longevity Projections
Listed below in Table 8 are the longevity projections from implant to
the Elective Replacement Indicator (ERI) for the Model K
SR703 and KSR706 pacemakers. The estimated time varies based
K
SR701,
on the lead impedance and the percent paced.
Table 8. Projected Longevity from Implant to ERI
Programmed SettingsPercent
Mode: VVIRLead Impedance
Lower Rate: 60 ppm500
Nominal Outputs
Amplitude: 3.5 V
Pulse Width: 0.4 ms
Low Outputs
Amplitude: 2.5 V
Pulse Width: 0.4 ms
Lower Outputs
Amplitude 1.5 V
Pulse Width: 0.4 ms
Paced
100%
50%
100%
50%
100%
50%
Projected Longevity
(Years)
ohms
6.5
7.5
7.4
7.9
8.4
8.8
600
ohms
6.8
7.7
7.6
8.1
8.5
8.9
1000
ohms
7.5
8.2
8.0
8.3
8.8
9.1
Warning: When the Elective Replacement Indicator (ERI) is set, the
pacemaker will continue to operate at ERI conditions for a minimum
of three months until the battery depletes.
88
Page 89
Longevity Projections
Note: The values above are based on calculations using deliverable
battery capacity. These values are estimates of longevity projections
from implant to ERI and are intended to assist the clinician in
understanding the effects of various pacing conditions on battery
longevity. These values should not be interpreted as precise
projections.
Listed below in Table 9 are the longevity projections from ERI to
erratic pacing for the Model K
SR701, KSR703 and KSR706
pacemakers. At most programmed settings, virtually all (i.e., 99.9%)
of these pacemakers will achieve a minimum of three months
longevity after ERI is set.
Note: The average time in the table below refers to the projected
mean longevity of pacemakers at the stated conditions.
Table 9. Longevity from ERI to Erratic Pacing
Previous Programmed
Settings Before ERI
Lower Rate: 60 ppm
Amplitude: 3.5 V
Pulse Width: 0.4 ms
Mode: VVIR/AAIR100%3.58.5
Mode: VVI/AAI100%3.58.5
a
Programmed settings which lead to increased current demands in VVI, 65-ppm
operating condition after ERI may slightly decrease average projections. Increased
current demands post-ERI could be caused by an increase in the percent paced and/
or the increase in pacing rate.
b
Current data show that combinations of extreme pacing conditions (high output values
[≥ 5.0 V] and high pulse widths, and /or low impedance values [300 ohms] may
decrease the average projection in the AAI/AAIR and VVI/VVIR modes.
Percent
Paced
Projected Longevity
(Months)
Lead Impedance
500 ohms
MinimumAverage
a,b
89
Page 90
Programmable Parameters
Programmable Parameters
Listed below in Table 10 through Table 16 are the pacemaker’s
programmable modes and parameters.
Table 10. Pacing Modes and Rates
ParameterCapabilityNotes
Pacing ModeVVIR, VVI, VVT, VOOR,
Lower Rate30, 35, 40, … 175 ppm
Upper Sensor Rate80, 90, 95, … 180 ppm
VOO, AAIR, AAI, AAT,
AOOR, AOO, OVO, OAO
(except 65 and 85 ppm)
Rate responsive
modes
90
Page 91
Programmable Parameters
Table 11. Rate Response Parameters
ParameterCapabilityNotes
ADL Rate60, 65, 70, … 180 ppmRate responsive
modes
Rate Profile
Optimization
ADL Response1-Inactive
Exertion Response1-Infrequent
ADL Setpoint5, 6, 7, … 40,
UR Setpoint15, 16, 17, … 40,
Activity ThresholdLow, Medium Low,
Activity Acceleration15, 30, 60 sec
Activity Deceleration2.5, 5.0, 10 min, Exercise
On, Off Rate responsive
2-Less Active
3-Mod. Active
4-More Active
5-Very Active